HBM QUANTUM X SCM-HV User manual

Quick Start Guide | Kurzanleitung | Guide rapide
English Deutsch Français
SCM-HV

Hottinger Baldwin Messtechnik GmbH
Im Tiefen See 45
D-64239 Darmstadt
Tel. +49 6151 803-0
Fax +49 6151 803-9100
www.hbm.com
Mat.: 7-2001.3615
DVS: A3615-4.0 HBM: public
11.2015
EHottinger Baldwin Messtechnik GmbH.
Subject to modifications.
All product descriptions are for general information only.
They are not to be understood as a guarantee of quality or
durability.
Änderungen vorbehalten.
Alle Angaben beschreiben unsere Produkte in allgemeiner
Form. Sie stellen keine Beschaffenheits- oder Haltbarkeits
garantie dar.
Sous réserve de modifications.
Les caractéristiques indiquées ne décrivent nos produits
que sous une forme générale. Elles n'impliquent aucune
garantie de qualité ou de durablilité.

Quick Start Guide | Kurzanleitung | Guide rapide
English Deutsch Français
SCM-HV

2A3615-4.0 HBM: public SCM-HV
English
1 Safety instructions 3........................................
2 Markings used 15............................................
2.1 Symbols on the signal conditioner 15............................
3 Introduction 16..............................................
4 Scope of supply 16..........................................
5 Accessories 17..............................................
6 Connecting 18...............................................
6.1 Preparatory measures 18......................................
6.2 Starting up 21................................................
6.3 Connection options 22.........................................
7 Dimensions 27..............................................
8 Specifications 28............................................

Safety instructions
SCM-HV A3615-4.0 HBM: public 3
1 Safety instructions
Intended use
The SCMHV (signal conditioner) is only used to mea
sure the differential voltages within the ratings stated in
the Specifications, in conjunction with a suitable Quan
tumX module.
The signal conditioner is designed for use in clean and
dry rooms, and cables must not be routed through out
door areas either. It must not be operated where dust
levels or the moisture content of the air are particularly
high, where there is a risk of explosion, or in an aggres
sive chemical environment.
See the specifications for the permissible ambient tem
perature, as well as the permissible storage and trans
portation temperatures. Should condensation have
formed during transportation, the SCMHV must be accli
matized for approx. 2 hours, before being put into opera
tion.
The signal conditioner must only be operated on suitable
QuantumX modules (see specifications) — it is not ac
ceptable to apply measuring voltage when the signal con
ditioner is not connected. The operating position is op
tional.
Everyone involved with siting, starting up, or operating
the signal conditioner must have read and understood
the operating manual and in particular the technical
safety instructions.
In the interests of safety, the signal conditioner should
only be operated by qualified personnel (see below) and
as described in the operating manual. It is also essential
to comply with the legal and safety requirements for the

Safety instructions
4A3615-4.0 HBM: public SCM-HV
application concerned during use. The same applies to
the use of accessories.
The signal conditioner is not intended for use as a safety
component. Please also follow the instructions in the “Ad
ditional safety precautions" section. Proper and safe op
eration requires proper transportation, correct storage,
siting and mounting, and careful operation.
Operating conditions
SProtect the signal conditioner against direct contact
with water.
SProtect the signal conditioner against moisture and
humidity, as well as weather conditions such as rain
or snow. The signal conditioner protection class is
IP20 (DIN EN 60529).
SDo not expose the signal conditioner to direct sunlight.
SComply with the maximum permissible ambient tem
peratures and maximum humidity figures stated in the
specifications.
SThe signal conditioner is designed for the CAT II mea
surement category (measuring circuits directly con
nected to sockets or similar low-voltage grid installa
tion connectors), with degree of pollution 2
(non-conductive pollution, temporary conductivity can
occur, as a result of condensation).
SIf the signal conditioner is operated outside the mea
surement category ratings specified in
EN 61010-1:2010 (not connected to the supply net
work, formerly CAT I), it is necessary to analyze the
operating voltage, the loop impedance, the occasional
overvoltage and the transient overvoltages of the cir
cuit to be measured. The SCMHV can only be used
once the analysis data has been compared with the

Safety instructions
SCM-HV A3615-4.0 HBM: public 5
maximum permissible values specified in the SCMHV
data sheet.
SThe signal conditioner must not be modified from the
design or safety engineering point of view except with
our express agreement. In particular, any repair or
soldering work on motherboards (replacement of com
ponents) is prohibited. When exchanging complete
modules, use only original parts from HBM.
SThe signal conditioner is delivered from the factory
with a fixed hardware and software configuration.
Changes can only be made within the possibilities
documented in the manuals.
SThe signal conditioner is maintenance free.
SPlease note the following when cleaning the housing:
- Disconnect the signal conditioner from all current
and voltage supplies.
- Also switch off the feeding QuantumX module.
- Clean the housing with a soft, slightly damp (not
wet!) cloth. Never use solvent as this could damage
the labeling or the housing.
- When cleaning, ensure that no liquid gets into the
signal conditioner or onto the connections.
- Give the SCM-HV sufficient time to dry before start
ing it up again.
SThe signal conditioner must be stored in dry, closed
rooms. If the signal conditioner has been transported
at extreme temperatures, you should wait for at least
2 hours to allow the signal conditioner to acclimatize,
before switching it on.
SIn accordance with national and local environmental
protection and material recovery and recycling regula
tions, old signal conditioners that can no longer be

Safety instructions
6A3615-4.0 HBM: public SCM-HV
used must be disposed of separately and not with nor
mal household garbage.
SThe SCMHV is intended solely for use in conjunction
with a QuantumX module. If the SCMHV is also inte
grated into a system, it is the responsibility of the sys
tem installer to keep the system safe.
Qualified personnel
Qualified persons means persons entrusted with the in
stallation, fitting, commissioning and operation of the
product who possess the appropriate qualifications for
their function.
SFor measurements covered by the low voltage direc
tive, the signal conditioner must only be connected by
a qualified electrician, or by someone with electrical
training under the supervision of a qualified electri
cian. (A qualified electrician is someone whose techni
cal training, know-how and experience, as well as
knowledge of the relevant requirements, allows them
to assess the work assigned to them and to recognize
potential risks, and who has been designated a quali
fied electrician by the contractor).
Working safely
SWhen unpacking the contents of the package, check
that everything is present. When it is unpacked, the
SCMHV should be inspected for mechanical dam
age. If damage has occurred during transit, the signal
conditioner must not be operated. Any shipping dam
age caused by inadequate packaging is excluded
from the warranty.
SError messages should only be acknowledged once
the cause of the error is removed and no further dan
ger exists.

Safety instructions
SCM-HV A3615-4.0 HBM: public 7
SThe signal conditioners and devices of automation
must be blocked in such a way that adequate protec
tion or locking against unintentional actuation is pro
vided (e.g. access checks, password protection, etc.).
SIf the SCMHV is operated on QuantumX modules,
additional safety precautions must be taken in terms
of both hardware and software for signal conditioners
working in networks, so that a line break or other in
terruptions to signal transmission do not cause unde
fined states or loss of data in the automation device.
SAfter making settings and carrying out activities that
are password-protected, you must make sure that any
controls that may be connected remain in a safe con
dition until the switching performance of the signal
conditioner has been tested.
Additional safety precautions
This signal conditioner is built and tested in accordance
with EN 61010, Safety Requirements for Electrical Equip
ment for Measurement, Control and Laboratory Use, and
was in perfect condition when it left production. To main
tain this condition and ensure risk-free operation (reduc
ing residual risks, EN 610101:2010 17 c), the user must
comply with the instructions and warning notices con
tained in this operating manual.
Additional safety precautions to meet the requirements of
the relevant national and local accident prevention regu
lations must be taken in plants where malfunctions could
cause major damage, loss of data or even personal in
jury.
The scope of supply and performance of the signal con
ditioner covers only a small area of measurement tech
nology. Before starting up the signal conditioner in a sys
tem, a project planning and risk analysis must first be

Safety instructions
8A3615-4.0 HBM: public SCM-HV
implemented, taking into account all the safety aspects of
measurement and automation technology, so that resid
ual dangers are minimized. This particularly concerns
personal and machine protection. In the event of a fault,
the relevant precautions must establish safe operating
conditions.
General dangers of failing to follow the safety
instructions
The signal conditioner is state-of-the-art and reliable. The
signal conditioner may give rise to residual dangers if it is
inappropriately installed or operated.
Warning signs and danger symbols
Important instructions for your safety are specifically
identified. It is essential to follow these instructions in or
der to prevent accidents and damage to property.
DANGER
This marking warns of an imminently threatening danger
ous situation which - if safety requirements are disre
garded - will result in death or serious physical injury.
WARNING
This marking warns of a potentially dangerous situation
which - if disregarded - can result in death or extremely
serious physical injury.

Safety instructions
SCM-HV A3615-4.0 HBM: public 9
CAUTION
This marking warns of a potentially dangerous situation
which - if disregarded - can result in slight or moderate
physical injury.
Notice
This marking draws attention to a situation which - if dis
regarded - can result in damage to property.
DANGER
When applying a dangerous contact voltage to the mea
suring connections, you must comply with all the related
safety regulations!
DANGER
Voltages with no energy limit that exceed one of the fol
lowing values are rated as dangerous according to
EN 61010:
1. AC voltage, 33 V rms value
2. AC voltage, 46 V peak value
3. DC voltage, 70 V
Higher voltages must only be applied by qualified person
nel who are familiar with the dangers involved! It is es
sential to comply with the related safety regulations!

Safety instructions
10 A3615-4.0 HBM: public SCM-HV
DANGER
If the signal conditioner is not used as intended, the pro
tection provided by the signal conditioner may be im
paired.
DANGER
If the signal conditioner is used outside the ratings, the
protection provided by the signal conditioner may be im
paired.
DANGER
In the measurement category II range, the SCMHV is
approved for a maximum voltage to ground potential of
300 V DC or 300 V rms AC.
DANGER
The protection provided by the signal conditioner can be
impaired if the signal conditioner is operated in ambient
air that is too humid or has too much condensation, or in
a damp or wet condition, such as after liquid spillages
(see specifications).

Safety instructions
SCM-HV A3615-4.0 HBM: public 11
DANGER
The protection provided by the signal conditioner can be
impaired if the signal conditioner is operated above the
permissible maximum operational height (see specifica
tions).
DANGER
The SCMHV housing is rated for a maximum impact
stress of 1 J, according to EN 610101:2010 8.2.2. The
protection provided by the device may be impaired if the
SCMHV is operated in situations where increased im
pact stress is to be expected. Adopt suitable measures
such as enclosures, a covering roof, etc., to protect all
the measurement technology against excessive impact
stress.
DANGER
The SCMHV is not intended for operation in potentially
explosive atmospheres.
DANGER
To avoid injury from coming into contact with danger
ously active potential, do not wear jewelry (necklaces,
etc.) that could pose a risk, and do not touch any live
parts. Comply with the safety regulations: isolate, secure,
check, ground and short-circuit, cover and safeguard.

Safety instructions
12 A3615-4.0 HBM: public SCM-HV
DANGER
Only trained and qualified personnel who are authorized
to do so by HBM are allowed to open the SCMHV. Be
fore opening the signal conditioner, the feeding Quan
tumX module must be switched off and the SCMHV dis
connected from all circuits. Operation when open is not
allowed.
DANGER
The signal conditioner, including the measurement cable
and the safety connector, must not be changed or modi
fied in any way.
DANGER
In the following situations, the signal conditioner must be
shut down and secured to prevent inadvertent operation:
Svisible damage to the signal conditioner
Sdamage to the connection cable, so that the white
insulation is visible
S(audibly) loose parts in the signal conditioner
Sthe signal conditioner no longer works

Safety instructions
SCM-HV A3615-4.0 HBM: public 13
DANGER
The maximum cable length, including the fitted 1 m long
measuring lead, must not exceed 30 m.
It is not acceptable to run cables through outdoor areas.
DANGER
The SCM is designed for CAT II applications. Measure
ment category II is applicable to test and measuring cir
cuits that are directly connected to the mains connections
(sockets) of the lowvoltage grid installation. The
SCMHV is not suitable for any other low-voltage grid
installation applications and its use for this purpose is
unacceptable.
DANGER
The following applies to test and measuring circuits with
out a rated measurement category (not directly con
nected to the supply network, formerly CAT I): Whether
the SCMHV is suitable for test or measuring circuit appli
cations not intended for direct connection to the supply
network, can only be determined by an accurate analysis
of the operating voltage, the peak voltage, the loop im
pedance, the occasional overvoltage and the transient
overvoltage of these circuits. The following characteris
tics apply to the SCMHV in this respect:
SOperating voltage: max. "415 V DC or 415 V rms AC
SDifferential voltage: max. "500 V DC or 500 V rms
AC

Safety instructions
14 A3615-4.0 HBM: public SCM-HV
SPeak voltage, common/differential mode: 700 V
SLoop impedance: at least 100 mΩ
(see EN 61010-2-030:2011 Table AA.1)
SOccasional overvoltage: none
SMaximum transient overvoltage: "2000 V
WARNING
In conjunction with a QuantumX module, the SCMHV is
a class A device. This device can cause radio interfer
ence in home premises. In this case, the operator may
be asked to take appropriate action.

Markings used
SCM-HV A3615-4.0 HBM: public 15
2 Markings used
2.1 Symbols on the signal conditioner
Take the details in the operating manual into
account
Read and comply with the operating manual.
CE mark
The CE mark enables the manufacturer to guarantee that
the product complies with the requirements of the rele
vant EC directives (the Declaration of Conformity can be
found at http://www.hbm.com/hbmdoc).
Statutory waste disposal mark
In accordance with national and local environmental pro
tection, material recovery and recycling regulations, old
signal conditioners that can no longer be used must be
disposed of separately and not with normal household
garbage.
If you need more information about disposal, please con
tact your local authorities or the dealer from whom you
purchased the product.

Introduction
16 A3615-4.0 HBM: public SCM-HV
3 Introduction
The SCMHV signal conditioner is used for the differen
tial measurement of high voltages with QuantumX mod
ules offering strain gauge full bridge input, for example
MX840A, MX440A and MX410/B.
Within the scope of measurement category II (supply net
work from the socket), the maximum operating voltage to
ground potential is 300 V DC or 300 V rms AC. Taking
the maximum acceptable rms value into account, the
peak value of the voltage must not at any time exceed
700 V between the connections (differential mode) and to
ground potential (common mode).
Its robustness in the face of high voltages and transients
also makes the SCMHV suitable input protection for
QuantumX modules, when the voltages to be measured
include high voltage peaks.
The 10 V measuring range allows low voltage values to
be measured at a high potential and currents to be mea
sured via an external shunt.
The SCMHV applies a safety limit to the measuring volt
age, which is electrically isolated from the adjacent chan
nels in the QuantumX module. This allows any number of
SCMHVs to be operated without mutual interference.
The signal conditioner can be automatically identified and
adjusted with TEDS (electronic data sheet).
4 Scope of supply
SQuantumX SCM-HV Voltage signal conditioner
Order no.: 1-SCM-HV
SQuick Start Guide

Accessories
SCM-HV A3615-4.0 HBM: public 17
5 Accessories
The SCM-HV Voltage signal conditioner is specified ac
cording to EN 61010 for 300 V CAT II. Only accessories
approved for at least this category can be used and con
nected at the measuring connections.
Recommended accessories
S4 mm test terminal, 1000 V red,
multi-contact 66.9416-22
S4 mm test terminal, 1000 V black,
multi-contact 66.9416-21
SRed test probe, multi-contact 24.0238-22
SBlack test probe, multi-contact 24.0238-21
SRed 4 mm coupling, multi-contact 66.9123-22
SBlack 4 mm coupling, multi-contact 66.9123-21

Connecting
18 A3615-4.0 HBM: public SCM-HV
6 Connecting
6.1 Preparatory measures
SConnect the SCM-HV signal conditioner to a suitable
QuantumX module.
10 V - 30 V DC
1-NTX001
300 V DC / 300 Vrms AC measuring voltage.
QuantumX voltage supply:
2
1
1
SLock the module by tightening the two knurled nuts (1)
SMake sure that the SCMHV is locked
SMake sure that all the disconnecting devices and the
emergency stop and emergency OFF devices are
easily accessible and can be reached by third parties
EN 610101:2010 5.4.4 b