Druck PTC200 User manual


Copyright 2022 Baker Hughes Company.
English–PTC Temperature Calibrators | i
Introduction
This technical manual provides operating instructions for the PTC Series temperature calibrators.
Scope
This technical manual contains a brief description, operation and testing procedures for the user
of this equipment.
Safety
The manufacturer has designed this equipment to be safe when operated using the procedures
detailed in this manual:
• Do not use this equipment for any other purpose than that stated. Incorrect use can prevent
the protection given by the equipment from working.
• Use suitably qualified1 Technicians and good engineering practice for all procedures in this
publication.
Maintenance
The equipment must be maintained using the manufacturer’s procedures and should be carried
out by authorized service agents or the manufacturer’s service departments.
Software Version
This manual contains operating instructions for PTC Series temperature calibrators with software
version 28.
Technical Advice
For technical advice contact Druck or subsidiary manufacturer of this product.
1. A qualified technician must have the necessary technical knowledge, documentation, special test equipment
and tools to carry out the required work on this equipment.

Copyright 2022 Baker Hughes Company.
ii | PTC Temperature Calibrators–English
Marks and Symbols on the Equipment
Symbol Description
This equipment meets the requirements of all relevant European safety directives. The
equipment carries the CE mark.
This equipment meets the requirements of all relevant UK Statutory Instruments. The
equipment carries the UKCA mark.
This symbol, on the equipment, indicates that the user should read the user manual.
This symbol, on the equipment, indicates a warning and that the user should refer to the user
manual.
This symbol warns the user of the danger of electric shock.
This symbol warns the user of the danger of hot surfaces.
Druck is an active participant in the UK and EU Waste Electrical and Electronic Equipment
(WEEE) take-back initiative (UK SI 2013/3113, EU directive 2012/19/EU).
The equipment that you bought has required the extraction and use of natural resources for
its production. It may contain hazardous substances that could impact health and the
environment.
In order to avoid the dissemination of those substances in our environment and to diminish
the pressure on the natural resources, we encourage you to use the appropriate take-back
systems. Those systems will reuse or recycle most of the materials of your end life
equipment in a sound way. The crossed-out wheeled bin symbol invites you to use those
systems.
If you need more information on the collection, reuse, and recycling systems, please contact
your local or regional waste administration.
Please visit the link below for take-back instructions and more information about this
initiative.
https://druck.com/weee

Copyright 2022 Baker Hughes Company.
English–PTC Temperature Calibrators | iii
Abbreviations
The following abbreviations are used in this manual; the abbreviations are the same in the
singular and plural.
Abbreviation Description
A Ampere
ac Alternating Current
cm Centimetre
cSt Centistokes
dc Direct Current
DUT Device Under Test
e.g. For example
etc. And so on
FS Full Scale
GND Ground
h Hour
i.e. That is
K Kelvin
lbs Pounds
LED Light Emitting Diode
m Minute
max Maximum
min Minimum or minute
mm Millimetre
n/a Not Applicable
PC Personal Computer
PID Proportional Integral Differential
PV Primary Variable
SP Set Point
Ref Reference
RH Relative Humidity
SP Set Point
TRRoom Temperature
USB Universal Serial Bus
V Volts
VA Volt Ampere
°C Degrees Celsius
°C/min Degrees Celsius per minute
°F Degrees Fahrenheit

Copyright 2022 Baker Hughes Company.
iv | PTC Temperature Calibrators–English

Copyright 2022 Baker Hughes Company.
English–PTC Temperature Calibrators | v
Contents
1. Description 1
1.1 Models 1
1.2 Unpacking 1
1.3 Intended Purpose 2
2. Safety Instructions 2
2.1 Qualified Personnel 3
2.2 General Safety Instructions 3
2.3 Special Safety Instructions 3
2.4 Calibration Liquid Safety Instructions 4
3. Construction and Function 5
3.1 Construction 5
3.1.1 Integrated Measurement Instrument 6
3.2 Function 7
3.2.1 Operation Procedure 8
3.3 Test Tasks 9
4. Commissioning and Operation 9
4.1 Operating Conditions 10
4.2 Electrical Connection 10
4.2.1 Electrical Connection Procedure 10
4.3 Calibrator Preparation 11
4.3.1 Adaptor Sleeve Insert 11
4.3.2 Air Shield Insert 12
4.3.3 Infrared Insert 13
4.3.4 Surface Insert 14
4.3.5 Tub Insert 15
4.3.6 Integrated Measuring Instrument 19
4.3.7 Measurement Interference Suppression 19
4.4 Switching On, Cool Down and Switching Off 20
4.4.1 Switching On 20
4.4.2 Cooling Down the Calibrator 21
4.4.3 Switching Off 21
5. User Interface 22
5.1 Main Window 22
5.2 Display Range 23
5.3 Toolbar 24
5.4 Remote Control (WebApp) 24
6. Test Task 25
6.1 Select Test Task 26
6.2 Configure Test Task 26
6.2.1 Name of Test Task 27
6.2.2 Data Acquisition 27
6.2.3 Functionality 27

Copyright 2022 Baker Hughes Company.
vi | PTC Temperature Calibrators–English
6.2.4 Device Under Test (DUT) 29
6.2.5 Test Points 30
6.2.6 Alarm Settings 33
6.2.7 Barcode 33
7. Calibrator Setup 33
7.1 Configure Network 34
7.2 Update Calibrator Software 34
7.3 Configure Presentation Format 35
8. Testing Process / Calibration 35
8.1 Before Starting 35
8.2 Starting the Testing 36
8.3 Canceling the Testing 36
8.4 After the Testing 37
9. Measurement Log Management 37
10. Troubleshooting 38
11. Maintenance, Cleaning and Transportation 38
11.1 Maintenance 38
11.1.1 Before Use 38
11.1.2 Annually 39
11.1.3 Recalibration 39
11.1.4 Calibration Liquid 39
11.1.5 Magnetic Stirrer 39
11.1.6 Mains Fuses 39
11.2 Calibration 39
11.2.1 Recalibration 39
11.2.2 Adjustment 40
11.3 Cleaning 40
11.3.1 Exterior Chassis 40
11.3.2 Vent Grilles for Inlet Air 40
11.3.3 Inserts 40
11.3.4 Tub Insert 40
11.4 Return Goods/Material Procedure 41
11.4.1 Safety Precautions 41
12. Decommissioning and Disposal 41
12.1 Before Decommissioning 41
12.2 Decommissioning 41
12.3 Disposal of Calibration Liquid 42
12.4 Disposal of Calibrator 42
13. Specification 42
13.1 Shared Specification 42
13.2 PTC200 Specification 43
13.3 PTC700 Specification 44

Copyright 2022 Baker Hughes Company.
viii | PTC Temperature Calibrators–English

Copyright 2022 Baker Hughes Company.
English–PTC Temperature Calibrators | 1
1. Description
The PTC Series calibrators are used to test and calibrate temperature measuring instruments
and temperature sensors, as well as for measuring temperatures. Typical applications include the
test and calibration of thermometers, temperature switches/thermostats, resistance
thermometers and thermocouples.
The portable instruments are of compact and robust construction so can be used on-site or in a
laboratory.
The PTC Series may be used for service purposes, or industrial and laboratory applications.
1.1 Models
1.2 Unpacking
Carefully unpack the unit to prevent any damage.
Model Temperature Range Function Integrated Measuring
Instrument
PTC200 -55°C to 200°C
(-67°F to 392°F) Dry Block
PTC700 TR to 700°C
(TR to 1290°F) Dry Block
PTC165 -35°C to 165°C
(-31°F to 329°F) Micro Bath
PTC255 TR to 255°C
(TR to 491°F) Micro Bath
PTC165i -35°C to 165°C
(-31°F to 329°F) Micro Bath
PTC255i TR to 255°C
(TR to 491°F) Micro Bath
INFORMATION The temperature calibrators are delivered in special
protective packaging. Save the packaging for returning the instrument safely to
the manufacturer for calibration or repair.
PTC165PTC200 PTC255iPTC700

Copyright 2022 Baker Hughes Company.
2 | PTC Temperature Calibrators–English
Check the completeness of the delivery based on the delivery note.
1.3 Intended Purpose
The PTC Series of calibrators may only be used for the testing and calibration of suitable
temperature measuring instruments, temperature sensors and for measuring temperatures.
The calibrators may not be used for warming up or heating materials or gases. The calibrators
have been designed for indoor use only.
The micro baths may only be used with a suitable calibration liquid. The only permitted liquids are
silicone oil, see Section 2.4.
Hazardous media (flammable or explosive liquids or gases) may not be used.
The operational safety of the calibrator is only guaranteed by intended use. The specified limits
may under no circumstances be exceeded. See Section 13.
It is your responsibility to select the instrument, which is suitable for your specific application, to
connect it correctly, to carry out the tests safely and to maintain all components.
2. Safety Instructions
Before you use the calibrator, read through this operating manual carefully. If the instructions are
not followed, in particular the safety guidelines, this could result in danger for people, the
environment, and the calibrator and the system it is connected to.
The calibrator represents state-of-the-art technology. This encompasses accuracy, operating
function and the safe operation of the calibrator.
Table 1: Delivery Checklist
Item PTC200
PTC700
PTC165
PTC255
PTC165i
PTC255i
Druck Premium Temperature Calibrator
External reference sensor.
Dry block insert.
Mains cable.
Network (Ethernet) cable.
Test certificate.
Safety manual with directions on how to
download the instruction manual.
Protective packaging and transportation
protection.
Insert exchange tool.
Work cover with five silicone plugs.
Tub insert.
Terminal connectors (4 × red, 4 × black
and 1 × white)
2 × Thermocouple adapters.
2 × Clamp-on ferrites.
2 × Ferrite keys (for opening the clamp-on
ferrites)

Copyright 2022 Baker Hughes Company.
English–PTC Temperature Calibrators | 3
In order to guarantee that the calibrator operates safely, the operator must act competently and
be conscious of safety issues.
Druck provides support for the use of its products either personally or via relevant literature. The
customer verifies that our product is fit for purpose based on our technical information. The
customer performs customer-specific and application-specific tests to ensure that the product is
suitable for the intended use. With this verification, all hazards and risks are transferred to our
customers; our warranty is not valid.
2.1 Qualified Personnel
• The personnel in charge of the installation, operation and maintenance of the calibrator must
hold a relevant qualification. This can be based on training or relevant instructions.
• Personnel must be aware of this operating manual and have access to it at all times.
2.2 General Safety Instructions
• In all work, the existing national regulations for accident prevention and safety in the
workplace must be complied with. Any internal regulations of the operator must also be
complied with, even if these are not mentioned in this manual.
• Ensure that the complete operating instructions are always available at the calibrator
installation site.
• Degree of protection according to EN 60529. Ensure that the ambient conditions at the site
of use does not exceed the requirements for the stated protection rating, see Section 13.1.
• Structural safety in accordance with EN 61010-1. The calibrator must be installed in such a
way that the requirements for structural safety are met.
• Only use the calibrator if it is in perfect condition. Damaged or faulty calibrators must be
checked without delay and, if necessary, replaced.
• If problems cannot be remedied, immediately shut down the calibrator and ensure that it
cannot be started up accidentally.
• Never leave the calibrator unattended when it is in operation or in the cooling phase.
• Do not remove or destroy type plates or other markings on the calibrator, or the warranty is
rendered null and void.
2.3 Special Safety Instructions
• Thermal protection fuse.
a. The calibrator is equipped with a thermal protection fuse that works independently. If
there is an over-temperature inside the housing, the power supply to the heating system
is cut off. The thermal protection fuse is non-resettable. The calibrator cannot be used
anymore.
b. After the calibrator has cooled, return the calibrator to Druck.
• Risk of injury from hazardous gases.
a. When liquids are heated, evaporation can result in dangerous gases being released.
• The calibrator may not be used in a potentially explosive atmosphere.
a. Remove all the easily flammable media from the vicinity of the calibrator.
b. Ensure that the calibrator cannot come in contact with easily flammable or explosive
media.
• Operate the calibrator only in the temperature range permissible for the DUT.
• Ensure that the DUT is held securely in the calibrator.
a. Use only suitable inserts.

Copyright 2022 Baker Hughes Company.
4 | PTC Temperature Calibrators–English
b. When doing so, also ensure that the structural safety of the calibrator is retained.
• Expert mode.
a. When selecting the function, DUT and testing tasks, you can switch into the expert mode
for administration and configuration. The settings that are made in this mode require
detailed knowledge of the calibration and method of working of calibrators. Contact
Druck for specific training.
b. If the settings are incorrect, the calibrator may be damaged.
2.4 Calibration Liquid Safety Instructions
Before using calibration liquids, read the entire safety data sheet thoroughly. Pay particular
attention to the information on the physical and chemical properties.
Only use calibration liquids that are suitable for the required temperature range, and which are
not flammable.
Calibrators are set for specific fluid types. Use only the optional Druck supplied or recommended
silicone oil. These oils have temperature limitations that must be adhered to.
Druck recommends the following calibration liquids:
• Use only the silicone oil recommended in Table 2.
• Always read the safety data sheet supplied with the silicone oil before using it.
• Always ensure adequate ventilation when working with silicone oil, since hazardous
substances can be released.
• Prevent silicone oil from coming into contact with your eyes.
• Silicone oil is hygroscopic. Use the work cover to seal the micro bath after use. Alternatively,
remove the tub insert from the calibrator and use the transport cover to seal the tub insert.
Table 2: Calibration Liquids
Model Calibration Liquid Calibration Range Flashpoint
PTC165(i) XIAMETER™ PMX-200 Silicone
Fluid 10 cSt
-35 °C to 155 °C
(-31 °F to 311 °F)
165 °C
(149 °F)
PTC255(i) XIAMETER™ PMX-200 Silicone
Fluid 50 cSt
50 °C to 270 °C
(122 °F to 518 °F)
280 °C
(536 °F)

Copyright 2022 Baker Hughes Company.
English–PTC Temperature Calibrators | 5
3. Construction and Function
3.1 Construction
Figure 1: General View
The calibrator consists of a robust steel housing (2) with integral carrying handle (1). The touch
screen display (3) is located on the front panel of the unit.
The mains inlet, switch, and fuses (4) are located on the front of the unit. The external reference
sensor is connected to (9). USB and network connections are available at (10).
During operation, cooling air is drawn in from the base of the unit (11, 12) and expelled through
grilles (6, 8). PTC200 and PTC700 expel air through the top of the unit (6). PTC165(i) and
PTC255(i) expel air through side grilles (8).
The rear section of the housing contains the insulated dry block or micro bath. Heating elements,
together with an integrated temperature sensor, are used to regulate the temperature of the dry
block / micro bath. Some models contain cooling elements. These allow the temperature to be
regulated below room temperature.
1 Carrying handle. 2 Steel housing.
3 Touch screen for operation and measurement value
display.
4 Mains inlet, switch with fuses.
5 Dry block. 6 Exhaust cooling air through upper housing grilles.
7 Micro bath tank. 8 Exhaust cooling air through side housing grilles.
9 External reference sensor connection. 10 USB ports and network connection.
11 Inlet air for housing cooling. 12 Inlet air for dry block / micro bath cooling.
8
9
1
2
3
4
5
6
7
11
12
10

Copyright 2022 Baker Hughes Company.
6 | PTC Temperature Calibrators–English
3.1.1 Integrated Measurement Instrument
The PTC165i and PTC255i are equipped with an integrated measurement instrument.
Figure 2: Integrated Measurement Instrument
Symbol Description
RTD A / B Resistance thermometers (2, 3 or 4-wire)
Switch test (socket 1 and 2)
TC A / B Thermocouple input.
+24V out Transmitter supply.
mA in Current measurement.
Vin Voltage measurement.
COM Signal common (ground).
Functional earth connection.
ext. Ref. External reference sensor connection.

Copyright 2022 Baker Hughes Company.
English–PTC Temperature Calibrators | 7
3.2 Function
The function of the calibrator is determined by the calibrator model and the installed insert.
Figure 3 shows an overview of the available inserts.
Figure 3: Insert Overview
Table 3 details the required insert diameter and material for each calibrator model. Use only
Druck supplied inserts. Druck inserts have the correct tolerances and materials for the calibrator
models.
Note: Inserts are not interchangeable between calibrator models with different bore sizes.
Table 4 details the required external reference sensor for each model. It also details the
recommended drilling size in the insert to accept the external reference sensor.
Note: The external reference sensor is calibrated for use with a specific calibrator. Do not use a
different external reference sensor with the calibrator. This will degrade the calibration accuracy.
Table 5 shows the compatibility of insert for each calibrator model. The table also indicates when
to use the external reference sensor, or the internal reference sensor. Where it is possible to use
1 Adaptor sleeve insert. 2 Air Shield insert.
3 Infrared insert. 4 Surface insert.
5 Tub insert.
Table 3: Insert Diameter and Material
Model Insert Diameter Insert Material
PTC200 28 mm (1.10″) Brass
PTC700 29 mm (1.14″) Aluminum Bronze Alloy
PTC165(i)
PTC255(i) 60 mm (2.36″) Aluminum
Table 4: External Reference Sensor
Model External Reference
Sensor Part Number Temperature Range Drilling Diameter
PTC200
PTC165(i)
PTC255(i)
IOPTC-EXSEN-1 -55 °C to 255 °C
(-67°F to 491 °F)
3.3 mm to 3.5 mm
(0.130″ to 0.138″)
PTC700 IOPTC-EXSEN-2 5°C to 700 °C
(41°F to 1292 °F)
4.8 mm to 5.0 mm
(0.189″ to 0.197″)
12 3 4 5

Copyright 2022 Baker Hughes Company.
8 | PTC Temperature Calibrators–English
either the external or internal reference sensor, ensure that the correct function is selected. See
Table 6.
Table 6 shows the function names as they appear in the “Select Function” window.
3.2.1 Operation Procedure
1. Install a suitable insert into the calibrator, see Section 4.3. Ensure that the insert has a
secure fit to ensure optimum heat transfer to the DUT.
2. Once all preparations have been completed, switch on the calibrator. See Section 4.4.1.
3. Select the relevant function in the calibrator software for the insert. See Table 6 on page 8.
Note: The inserts have different characteristics. The calibrator is supplied with a set of
predetermined functions tailored to each insert. These have been determined by the factory
and are write protected. They may be used as a basis to create your own custom functions.
4. The calibrator heats or cools the dry block, or calibration liquid, to the set point temperature.
As soon as the temperature is stable, the DUT can be calibrated.
Table 5: Insert Compatibility
Insert Reference Sensor PTC200 PTC700 PTC165(i) PTC255(i)
Adaptor Sleeve Internal
External
Air Shield External
Infrared Internal
Surface External
Tub External
Table 6: Function Names
Insert Reference Sensor Function Name
Adaptor Sleeve
Internal dryblock + int.reference
dryblock int.ref.
External dryblock + ext.reference
dryblock ext.ref.
Air Shield External dryblock + ext.reference
air shield insert
Infrared Internal infrared + int.reference
Surface External surface + ext.reference
Tub External tubinsert + ext.reference
silicone oila
a. Refer to Section 2.4 for the correct type of silicone oil to use with the calibrator.
INFORMATION For your safety, manual operation of the calibrator, i.e. direct
starting and staying at a temperature, is not possible.
The testing procedure is always started with a testing task, see Section 3.3. This
ensures that the calibrator is always started with a defined end of test behavior.
See “Behavior at Test End” on page 32.

Copyright 2022 Baker Hughes Company.
English–PTC Temperature Calibrators | 9
5. Repeat step 3 for all the required calibration set points.
3.3 Test Tasks
Test tasks are containers for defined test conditions. They are helpful for recurring testing
processes, for standardizing test sequences and for generating measurement logs.
All required settings and configurations for the calibration of a DUT are compiled in a test task.
The parameters of the testing task are saved and linked to the selected function and DUT. See
Section 6.
The operation concept of the calibrator is based on pre-installed and self-defined test tasks. They
are a central component of the function and operation of the calibrator.
The emphasis of the operating manual is on the use of test tasks during operation of the
calibrator.
The calibrator is shipped with pre-defined functions, test samples and testing tasks. These have
been defined in the factory, in which the basic settings of the calibrator have been stored.
The protected test tasks can be neither be deleted nor edited. They serve as templates for your
own self-defined test tasks. They can be copied and then modified. This task must not be
completed without specific training from Druck.
You can define your own test tasks for different DUT or test sequences. These test tasks are
directly saved in the calibrator and can be easily activated. This makes quick access to recurring
test tasks possible.
Upon switching on the calibrator, the first test task of the selection list is loaded automatically,
with the relevant parameters.
4. Commissioning and Operation
For safe operation of the calibrators, a formal commissioning procedure is necessary.
Commissioning includes the installation, electrical connections, preparation for the calibration as
well as correct switching on and off of the calibrator.
Further, a visual inspection for damage is required before use.
The required steps are described in the following sections.
INFORMATION The function and the DUT are independent. During the
configuration of test tasks (Section 6), the respective expert mode can be
entered. This allows administration and configuration when selecting the
function or DUT.
Note that changes to existing functions and DUTs always affect all linked test
tasks.
WARNING The calibrator may become very hot during operation. If the
calibrator is operated without supervision, third-party persons in the vicinity
could get injured. Moreover, flammable material could come into contact with
the calibrator and cause significant harm to personnel and/or damage to
property.
Never leave the calibrator unattended when it is in operation or in the cooling
down phase.

Copyright 2022 Baker Hughes Company.
10 | PTC Temperature Calibrators–English
4.1 Operating Conditions
Select a safe installation site for commissioning the calibrator.
Considerations for selecting a safe installation site and calibrator operating position:
• Only suitable for indoor use, do not use outdoors.
• Operate only in the vertical position on an even surface. The surface must be stable, clean,
and dry.
• If the operational conditions do not conform to the above, the structural safety and the
specified properties of the calibrator cannot be guaranteed.
• At high testing temperatures, it is recommended to use a sufficiently large, fire-resistant,
supporting surface.
• Ensure sufficient clearance around the calibrator. On the front side > 1m, behind and to the
sides > 0.5 m. Ensure that there is sufficient head clearance and sufficient clear space above
the calibrator.
• Ensure sufficient ventilation.
• Do not operate in the vicinity of flammable materials.
• Do not install in a cupboard or other similar confining location.
• The ventilation openings must not be blocked or covered.
• The calibrator must be installed so that it can be switched off at any time.
4.2 Electrical Connection
Check the following points before you connect the calibrator:
1. Check the specified operating mains supply voltage range for the calibrator. See Section 13.
2. Ensure that the mains voltage range is the same as that specified on the rating plate.
3. Only connect the calibrator to a properly installed and earthed 3-pole socket for mains plugs
with earthing contact.
4. Only use extension cables or adaptor plugs with a protective earth connection.
4.2.1 Electrical Connection Procedure
1. Insert the mains cable into the mains inlet socket of the calibrator.
2. Insert the plug of the mains cable in a suitable mains outlet with earthing contact.
INFORMATION The plug of the mains connecting cable serves as a
emergency stop device. Ensure that the plug is always easily accessible and
easy to reach. In an emergency, pull the plug so that the calibrator is isolated
from the mains.
RISK OF ELECTRIC SHOCK Calibrators that have been exposed to high
humidity for long periods have an initial higher than normal earth leakage
current. Ensure that the calibrator is always connected to a protective earth.
Always use a known good mains cable with an appropriate power plug.
Use only a Druck specified mains cable.
Ensure that the mains outlet is adequately rated and has a protective earth
connection.
Before turning on the power, make sure that the calibrator is properly connected
to the protective earth. The protective earth is connected to the calibrator
through the mains plug.
This manual suits for next models
3
Table of contents
Other Druck Test Equipment manuals

Druck
Druck UPS-III-IS User manual

Druck
Druck DPI620 Genii User manual

Druck
Druck TRX User manual

Druck
Druck DPI 620 Genii User manual

Druck
Druck ADTS 403R Use and care manual

Druck
Druck InstruMetrics ADTS542F Setup guide

Druck
Druck UPS-III User manual

Druck
Druck DPI 610E User manual

Druck
Druck DPI 841 User manual

Druck
Druck DPI 620 Genii Instruction Manual