Healy TimeWaver User manual

Healy
Electrostimulation device
Instructions for Use
Frequencies for Life
Frequencies for Your Life

2 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
Table of contents
1. General 4
2. Healy 4
2.1 Intended Use 4
2.2 Indications 4
2.3 Contraindications 4
2.4 Description 4
2.4.1 Hardware 5
Description 5
2.4.2 Software Description 5
2.5 General 5
2.6 Safety Notes 6
3. Preparations for Using the Healy 7
3.1 Installation 7
3.1.1 Smartphone requirements: 7
3.1.2 Android Operating System 7
3.1.3 iOS Operating System 7
3.1.4 Hardware 8
3.1.5 Pairing via Bluetooth 8
3.1.6 Connection of the Electrodes 8
4. Operating the Healy Device 10
4.1 Starting the Treatment Using the Smartphone 10
4.2 Handling of the Healy Hardware 10
4.3 Starting the Treatment without the Smartphone12
4.4 Stopping the Treatment 12
5. Cleaning, Disinfection and Maintenance 12
5.1 General 12
5.2 Cleaning 13
5.3 Disinfection 13
5.4 Storage between Applications 13
5.5 Maintenance 13
5.6 Troubleshooting 13
EN |
Healy Electrostimulation device Instructions for Use

6. System Components 14
7. Technical Data 14
7.1 General 14
7.2 Electrical Parameters 15
7.2.1 Power Source Requirements 15
7.2.2 Healy 15
7.2.3 Electromagnetic Compatibility 16
7.2.4 Radio Frequency Wireless Technology 16
7.2.5 FCC Statement 17
7.3 Mechanical Parameters 18
7.4 Environmental Conditions 18
7.4.1 Operation and Storage 18
7.5 Stimulation Settings (Standard Programs) 18
8. Disposal 19
9. Used Symbols and Icons 19
10. Revision status 22

4 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
1. General
Thank you for choosing Healy. Please read this manual carefully
and pay special attention to the warnings and cautions.
This manual is intended for the user. Intended users include
adults at home, and medical professionals in the healthcare or
professional environment. It aims to ensure safe handling and
enables users to become familiar with the medical device.
2. Healy
2.1 Intended Use
Healy is a microcurrent stimulation device used for pain man-
agement, using currents in the microampere range with dier-
ent frequencies. It can be applied to dierent areas of the body.
2.2 Indications
The device is designed to be used for temporary relief of pain
associated with sore and aching muscles in the shoulder, waist,
back, arms and legs due to strain from exercise or normal house-
hold work activities and for the symptomatic relief and manage-
ment of chronic, intractable pain and relief of pain associated
with arthritis.
2.3 Contraindications
Microcurrent treatment must not be performed when metallic
foreign bodies are in the area of treatment, pacemakers or im-
planted debrillators are present, bleeding, risk of bleeding or
embolism are present. Treatments must not be performed on
areas of the skin which have been treated radiotherapeutically
or have a sensory disorder or if the user has a fever.
• Not for use by pregnant women
• Not for use by those with epilepsy
• Do not place the electrodes on the head or neck.
Consult a medical professional before using if you have any of
these contraindications.
2.4 Description
The Healy is a portable device, internally powered and con-
trolled via Bluetooth by a software application that is installed
on a Smartphone. It has one stimulation output. The stimula-
tion output can be varied between -10 V and +10 V, 0 to 1 MHz,
0 – 4 mA.
The charging of the internal battery is done through the Micro-
USB-Port and the supplied power supply cable.
The software application “Healy APP“ is necessary for conguration and control of the Healy. The “Healy App“ is available in the Play S tore or the App St ore. Before in stallatio n verify the sm artphone re quirements i n section 4.1.1.

2.4.1 Hardware
Description
Figure 2.4.1.1: Healy top
view: Pos. 1 Icon and LED
for Bluetooth Connec-
tion; Pos. 2 On-O But-
ton; Pos. 3 Battery icon,
Pos. 4 LED for generator
action (if stimulator out-
put is active, then the LED
is blinking) and battery
indicator; Pos. 5 Genera-
tor icon; Pos. 6 Icons for
the positive and negative
electrode plugs; Pos. 7 USB-Connector icon; Pos. 8 Socket for
power supply through micro-USB–connector (not visible in this
gure); Pos. 9 Sockets for the electrodes (not visible in this gure)
2.4.2 Software Description
The “Healy APP“ is required for conguring and controlling the
hardware. It allows sending treatment programs to the device.
With the App you can start and stop the treatment. The “Healy
APP“ connects to your Healy hardware using Bluetooth. The
“Healy APP“ can connect only to one device at a time.
Figure 2.4.2.1: Overview in-
terface of the Healy APP. Pos.
1-Icon for the signal quality of
the Bluetooth connection to the
Healy; Pos. 2-Icon for the bat-
tery status of the Healy
2.5 General
The Healy is for use only to treat
conditions described in these
instructions. Intentional misuse
can be harmful or reduce the ef-
fectiveness of the device.
• Carefully read the instructions
for use before use.
• This product is intended only
for the purpose described in
the instructions for use.
• In this manual, the word DANGER warns of possible personal
injury!
• In this manual, the word CAUTION warns of possible damage
caused to property!

6 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
2.6 Safety Notes
Danger/Warning:
• Keep this device out of the reach of children.
• Do not immerse the device in liquids.
• Do not use while driving motor vehicles or operating heavy
machinery. The ability to drive or to operate heavy machinery
may be impaired immediately after treatment.
• Only charge with the supplied cable from an USB-Socket with
5V dc and max 2500 mA.
• In the case of a defect please contact the manufacturer. Any
attempt to repair the device will void the warranty.
• Use with care as the cables may cause a tripping or strangu-
lation hazard.
• Don’t use in presence of ammable gases or in an oxygen en-
riched environment.
• Stop treatment and remove the electrodes if any irritation,
sensation, allergic reaction or burning occurs at the electrode
contact points and contact us.
• Take care that the units output connectors are always point-
ed towards the oor. This way the device is protected against
damage by water or liquids.
• Do not place electrodes on the chest; this can cause ventric-
ular brillation.
• The device is designed for indoor use only.
• The device should only be stored and transported under the
conditions dened in these instructions.
• Do not use the device in shielded locations, such as intensive
care units in hospitals, airplanes or wherever the use of mobile
phones is prohibited.
• The device has no life-sustaining functions.
• The device must be operated only with supplied accessories
and together with compatible smartphones.
• Not for use by those with a pacemaker or heart disease.
• Fixed, portable, mobile HF-communication devices and
RF-emitters, e.g. mobile phones, RFID, diathermy and securi-
ty systems (e.g. electromagnetic anti-theft systems, metal de-
tectors), may inuence medical electrical devices. You should
keep a safety distance of 30 cm (12 inchs) from HF-communica-
tion devices and RF-emitters of parts and cables of the Healy,
when the device is in use. Don’t stack it with such devices or
put it next to other electrical equipment when used.
• Avoid damages to the electrode cables and the charging cables
caused by sharp objects. If the cables are damaged do not use
them and contact the manufacturer for a replacement. Don’t
use cables that are damaged.
• To protect the device and its accessories against contamina-
tion and damage keep it closed and away from children, dust,
sunlight, moisture, pets and vermin.

• Only use cables and electrodes supplied with the equipment.
Other accessories, cables or electrodes could result in high-
er electromagnetic eld emissions or less electromagnetic im-
munity and may lead to loss of function.
• Use of this equipment adjacent to or stacked with other equip-
ment should be avoided because it could result in improper
operation.
• Do not use the adhesive electrode if the gel is dry or visible
damage is present or the connecting cable is damaged.
• The equipment and the connecting leads used with this device
must all be in good condition.
• Do not use if recurrent episodes of skin irritation occur.
• Do not apply electrodes on skin aected by abrasions, rash
or wounds.
• Do not carry out maintenance while the equipment is in use.
• Operation in close proximity (i.e., less 3 feet) to a diathermy,
shortwave or microwave therapy ME equipment may produce
instability in the stimulator output.
• Simultaneous use together with RF-surgical equipment, elec-
trocautery or a debrillator is prohibited.
3. Preparations for Using the Healy
3.1 Installation
Before putting the Healy into operation you need to prepare the
hardware and install the “Healy APP“.
3.1.1 Smartphone requirements:
Bluetooth Bluetooth 4.0 or higher
Operating system iOS1 version 9.0 or higher
Android2version 4.4 or higher
1iOS is a registered trademark of the operating system for
smartphones and devices manufactured by Apple Inc.
2Android is a registered trademark of Google Inc. for the oper-
ating system for smartphones and other devices
3.1.2 Android Operating System
• Open the Play Store1
• Search for the “Healy APP“
• Choose “Healy APP“
• Download and install on your smartphone
3.1.3 iOS Operating System
• Open the Apple Store2
• Search for the “Healy APP“
• Choose “Healy APP“
• Download and install on your smartphone
1Play Store is a registered trademark of Google Inc.
2App-Store is a registered trademark of Apple Inc.

8 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
3.1.4 Hardware
• Unpack the Healy and check for damage
• Charge the batteries for 1.5 hours before rst use
• Switch on the Healy device by briey pressing the on-switch
and verify that the right LED is blinking
Note: The battery cannot be removed by the user.
3.1.5 Pairing via Bluetooth
After installation of the App, switch on the Healy by briey push-
ing the on-switch (see gure 2.4 .1.1) of the device.
After starting the App there is a short starting screen (see g-
ure 3.1. 5.1).
After the App is loaded it is scanning for Bluetooth devices (see
gure 3.1.5.2).
Within 20 seconds the App scans the environment and connects
to the Healy automatically. If your smartphone detects a Healy
gure 3.1.5.3 appears.
After you scan the QR-Code on the back of the Healy or enter
the serial number of the device you can connect to the Healy by
pressing the “Connect” button (see gure 3.1.5.4).
3.1.6 Connection of the Electrodes
The electrodes have been tested and evaluated together with
the Healy and are safe and eective for use, if the recommenda-
tions and advice in this manual are followed. Check the package
containing the electrodes and make sure the gel is intact; if the
package is visibly dented or damaged, do not use the product.
Put the plugs of the electrode cables into the sockets of the
Healy. Be sure to have the correct polarity. Place the electrodes
onto the desired treatment area.
Figure 3.1.5.2:
Scanning screen
Figure 3.1.5.1:
Starting screen

Important note on using the adhesive electrodes:
• Never place electrodes in the region of the carotid artery
• Do not place the electrodes on open wounds
• Before use clean the treatment area and if necessary shave
the area
• Choose an area free from hair, skin lesions or scars, folds in
the skin, or bony protuberances and which is not in the vicini-
ty of metallic implants.
• Clean and dry the application area before placing the elec-
trode on it.
• Avoid the use of solvents as these could cause skin irritation
and reduce the adherence of electrodes.
• If using alcohol to clean the area, make sure that it has com-
pletely dried before applying the electrodes.
• Connect the electrode to the cable of the electrostimulator.
• Check the functioning of the system before beginning the pro-
cedure.
• If the procedure is interrupted, check the functioning of the
system again before recommencing the session.
• Remove the liner and apply the electrode to the skin surface.
• Before switching on the electrostimulator, check the adher-
ence of the electrode to the skin surface.
WARNINGS:
• A reduced contact between the skin and the electrode may
cause the electric current to concentrate in one point and
cause burning.
• Interrupt the treatment if there is discomfort or a burning sen-
sation.
Figure 3.1.5.3:
Scan serial number screen
Figure 3.1.5.4:
Connect screen

10 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
• After the session, turn o the electrostimulator, disconnect
the cable from the electrode and gently remove the electrode
from the skin by lifting it up by one corner.
• Rapid removal of the electrode could cause skin traumas.
• Do not remove by pulling the connection part of the cable.
• After use, reposition the electrodes in the plastic liner and put
them back in the pouch.
• Electrodes may be reused: Reuse only for the same person.
• Verify the adhesion of the electrode before use. When elec-
trodes lose adhesion, gently rub one or two drops of water on-
to the gel surface. This may extend the usability. If there is no
improvement, replace with new electrode.
• Keep the packet closed in order to prevent the unused elec-
trodes from becoming dehydrated.
• Do not reposition the electrode during use.
4. Operating the Healy Device
You should become familiar with these frequently used opera-
tions:
• switch On-O (please refer to chapter 4.1.5 and 5.4)
• select treatment (please refer to chapter 5.1)
• start stimulation (please refer to chapter 5.1)
• connect the electrodes (please refer to chapter 4.1.6)
4.1 Starting the Treatment Using the Smartphone
After the installation of the App, pairing and connection of the
electrodes, the Healy is ready for operation. In the overview
screen you can select which program category you want to use
(see gure 4.1.1).
After selecting a program category the program page with the
icons of the dierent programs is loaded (see gure 4.1.2).
After you select a program from the program page screen the
information for placing the electrodes is displayed (see gure
4.1.3).
After selecting “Start” on the information page the treatment is
started and the treatment page is displayed (see gure 4.1.4). In
the treatment page you can set the intensity of the treatment by
using the plus and minus buttons.
When the electrical circuit is closed the circle, which is an indica-
tor for the time, will switch from black to white (see gure 4.1.5).
4.2 Handling of the Healy Hardware
During the treatment you should use the clip on the back of
the device to attach the Healy to your clothing. The electrodes
should be axed as shown in the App (please refer to chapter
10). Route the electrode cables to the Healy so that you feel com-
fortable.

Figure 4.1.1:
Overview interface
Figure. 4.1.4:
Treatment page
Figure 4.1.2:
Program page
Figure 4.1.3: Information page
for the placing the electrodes

12 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
4.3 Starting the Treatment without the Smartphone
The device always stores the last treatment program. By press-
ing the on-switch of the Healy hardware, the last program that
was sent from the smartphone to the device will start with an in-
tensity of 15%.
4.4 Stopping the Treatment
You can see the current progress of the program running on the
device by using the App.
To stop the treatment select one of the following:
• briey press the on/o-switch (pause mode)
• tap the icon in the treatment mode in the app (see gure 4.4.1)
• hold the on/o-switch for about 10 seconds until device hard-
ware resets
• in emergency situations it is possible to remove the electrodes
or disconnect the electrode cables
5. Cleaning, Disinfection and Maintenance
5.1 General
Maintenance activities increase the lifespan and safety of the
Healy. Cleaning and disinfection reduces the risks of cross con-
tamination and infections. Please read this section carefully.
Cleaning and disinfection should be performed once a week, or
if used by dierent patients prior to use.
Figure 4.1.5: Treatment page
with white time indicator –
electrical circuit is closed
Figure 4.1.6: Treatment page
with black time indicator –
electrical circuit is open

5.2 Cleaning
Switch o the Healy before
cleaning. For cleaning, com-
mercial cleaning products may
be used. The surfaces can be
wiped with a damp sponge or
cloth. Do not use too much liq-
uid and make sure that no liq-
uid penetrates beneath the
covers or enclosures.
Caution:
Do not use cleaning agents
with solvents (e.g. acetone,
nail polish, cleaners) or abra-
sive cleaners as they will dam-
age the surface.
5.3 Disinfection
Disinfection should be per-
formed in the same manner
as cleaning. Use only chem-
ical disinfectants in liquid
state cleared by the FDA. Use
a sponge or a cloth to wipe the
disinfectant over the device.
Caution:
Spray disinfection can cause a security risk. Only use disinfec-
tant wipes.
5.4 Storage between Applications
After the treatment unplug all cables and electrodes and put
them into the storage case and store in a cool and dry location.
Refer to the safety notes for environmental storage conditions.
5.5 Maintenance
There are no serviceable parts by the users. In a case of dam-
age, please contact the manufacturer. The following mainte-
nance activities are recommended and should be considered for
safe working.
Frequency By whom Maintenance activity
Prior to
each use
User Check housing, cables and elec-
trodes for damages. In case of
damage, contact the manufactur-
er or an authorized dealer
5.6 Troubleshooting
In the following you nd typical warnings or messages that are
related to the app or hardware of the device. If you face an event
that is not listed here or a problem still exists, please contact the
manufacturer.
Figure 4.4.1: Pause screen
By pressing the “X” in the
pause screen you cancel the
program and get back to the
overview interface.

14 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
Event or
message
Cause Actions required
No connection
from the smart-
phone to the
device
Device is
switched o
Switch on the device
and restart the App
No connection
from the smart-
phone to the
device
Bluetooth in
your smart-
phone is deac-
tivated.
Open connectivity
settings and activate
Bluetooth
Repeated short
blinking of right
LED
Batteries are
low
Charge the device
Device is not
responding
or
Undened
error/EM
disturbance
Hold the on-switch
for 10 seconds. This
triggers a hardware
reset.
No Stimulation
output
Undened
error/EM
disturbance
Contact your autho-
rized dealer or the ma-
nufacturer Healy GmbH
6. System Components
The Healy is delivered with a set of electrodes and a charging
cable.
Reference
number
Description Quantity
0006 Healy hardware device 1
108-4063 Press button adhesive electrodes
round Ø 32 mm
4
108-4062 Connection cable for electrodes
96 cm - press button on 2 mm
2
108-4068 Charging cable USB 0.15 m 1
7. Technical Data
7.1 General
Protection
against ingress of
water and parti-
culate matters
IP 42
Protected against particulate matter,
larger than 1mm. Protection against
dripping water at an angle of 15°.

EN | 15T. +49 30 54905949-1 | www.healy.eu | info@healy.de
7.2 Electrical Parameters
7.2.1 Power Source Requirements
The Healy can be connected to any power source with the sup-
plied USB-charging cable and the following specications:
Supply voltage range 4.4 V – 5.5 V DC
Output current Max. 2500 mA
Frequency DC
Socket type USB- Type A
Power type High powered 500 mA or higher
Insulation type Double or reinforced insulation
Compliance To IEC60950 or IEC60601-1
7.2.2 Healy
Input voltage Se e 7. 2.1
Input Current 20 mA
Protection class Internally powered
Applied type BF, the electrodes are in contact to
the patient’s body
Compliance IEC60601-1:2012; IEC60601-1-2
Battery capacity 145 mAh
Battery charging time 60 min - 90 min
Charging cycles Life
expectancy
500 times, more than 80% of initial
capacity at 20°C
Output voltage -
Stimulation -10 V to +10 V
Output current Max. 4 mA (voltage control) / ±2,6
mA (current control)
Accuracy voltage ±0.1 V with ±10 %
Accuracy current ±20 µA with ±10 %
Output frequency -
Stimulation DC up to 1 MHz
Signal form Rectangular biphasic, rectangular
monophasic or DC

16 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
7.2.3 Electromagnetic Compatibility
CISPR 11 -
emissions
EN55022
Group 2,
30 MHz – 1 GHz and 1 GHz – 6 GHz, classe B
The Healy is suitable for use in all establish-
ments, including the domestic area and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic
emissions
Class A acc. to IEC61000-3-2
Flicker Complies IEC61000-3-3
Immunity Complies IEC6061-1-2:2014 and EN61000-4-3
80 MHz – 2,7 GHz, 10 V/m
Immunity against wireless communication
equipment
28 V/m for MHz: 450, 810, 870, 930, 1720, 1845,
1970, 2450
27 V/m for MHz: 385
9 V/m for MHz: 710, 745, 780, 5240, 5500, 5785
ESD
protection
Complies IEC61000-4-2
8kV contact - 15 kV air
Power
frequency
magnetic
elds
Complies IEC61000-4-8
Test level 30 A/m for 50 Hz and 60 Hz
7.2.4 Radio Frequency Wireless Technology
Description Bluetooth Low Energy RF – Bluetooth V4
Compliance SIG TCRL Core - 2015-12-22; Core and Host Ver-
sion 2.0-4.2
(SIG = Bluetooth Special Interest Group)
ETSI EN301 489-1; ETSI EN301 489-17
Compati-
bility
Smartphones with Bluetooth Wireless techno-
logy and Bluetooth 4.0 or higher
Operating
frequency
2.4GHz
Congura-
tion
The Healy device starts the Bluetooth wireless
technology automatically when it is switched
on. Switch on the Bluetooth wireless techno-
logy on your smartphone and start the Hea-
ly App. During the rst scan, the app scans
and connects to the Healy device, see Section
3.1.5. The rst connection is called pairing.

Notes on
usage
After pairing, the Healy device is permanently
connected to the Healy App. The Healy App
can only control one Healy device at a time.
The wireless Bluetooth technology has a
limited range, therefore the smartphone
should be close to the Healy during use
(distances not more than 10 m [32 feet]).
If the connection is interrupted, the last
congured or started application will be
resumed. The pairing is not lost, only the
connection can be lost. It is not necessary
to perform the pairing again after each
shutdown or loss of connection with the Healy
device.
If you want to pair with another Healy device,
turn o the paired one or go to the settings
and select “Connect to another Healy”. Then
it is possible to pair again with another Healy
device, see section 3.1.5.
7.2.5 FCC Statement
Changes or modications not expressly approved by the party
responsible for compliance could void the user’s authority to op-
erate the equipment.
This device complies with part 15 of the FCC Rules. Operation is
subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including
interference that may cause undesired operation.
Note: This equipment has been tested and found to comply
with the limits for a Class B digital device, pursuant to part 15
of the FCC Rules. These limits are designed to provide reason-
able protection against harmful interference in a residential in-
stallation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance
with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interfer-
ence will not occur in a particular installation. If this equipment
does cause harmful interference to radio or television reception,
which can be determined by turning the equipment o and on,
the user is encouraged to try to correct the interference by one
or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit dierent
from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for
help.

18 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
7.3 Mechanical Parameters
Dimensions 55 x 55 x 7 mm
Weight 34 g
Cable length charging cables 0.15 m
Cable length electrode cables 96 cm (REF: 108-4062)
Output socket 2 x 2 mm pin
Input socket USB-Type B micro
7.4 Environmental Conditions
7.4.1 Operation and Storage
Temperature Range 5°C to 40°C / 41°F to 104°F
Relative humidity 30 % to 80 %
Atmospheric pressure 700 hPa to 1060 hPa
7.5 Stimulation Settings (Standard Programs)
The waveform is always a biphasic rectangle. One phase is de-
ned through one positive peak and one negative peak. By the
nature of the rectangle signal each current peak is considered as
a DC level. There is no oset (DC or AC) to the signal. The param-
eters are valid for a load resistance of 1 kOhm.
Name of the
program
Intensity cur-
rent or voltage
(amplitude for
100% setting)
Frequency
1 Phase
Impulse
width high
Impulse
delay/rest
Impulse
width low
Impulse
delay/rest
Pain +/- 600 µA 0.5 Hz 1 s 0 s 1 s 0 s
Pain locally +/- 600 µA 100 Hz 2.5 ms 2.5 ms 2.5 ms 2.5 ms

EN | 19T. +49 30 54905949-1 | www.healy.eu | info@healy.de
8. Disposal
The expected service life is 5 years. Ask your local authority for
recycling advice.
9. Used Symbols and Icons
The following icons are used on the device, the packaging and
the labelling:
Symbol for manufacturer
Electronic waste – do not
dispose. Recycle acc. to local
regulations or contact your
authorized dealer or the
manufacturer.
Recycling symbol
Use only inside In conformity with European
legislation Keep away from heat
Applied part type BF Read instructions for use
before use
Symbol for Bluetooth
Wireless connection
Caution: Read instructions
for use before use. Follow
safety advices in this manual.
SGS US CAN certication
Mark Symbol for Battery
Temperature limitations Bluetooth Symbol for on- and o-switch

20 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
Follow the instructions for
use Best before Symbol for USB-Connection
Keep dry Plus icon for positive
electrode socket
Icon for the bluetooth signal
strength between the Healy
App and Healy hardware
Hardware reference number One patient use Battery status of the Healy
hardware
Symbol for generator opera-
tion – stimulation output Icon for overview interface Icon for the trademark Healy
Minus icon for negative
electrode socket
The following symbols are used in the Healy – App
Pictograph for the placement of adhesive electrodes.
Appears after selection of the stimulation program
Note: the background colour may vary,
depending on the selected and congured
program parameter
REF 0006
Table of contents
Other Healy Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Dürr Medical
Dürr Medical CR 35 VETwin Installation and operating instructions

Fukuda Denshi
Fukuda Denshi DynaScope 7000 Series Service manual

Impulse Dynamics
Impulse Dynamics OPTIMIZER Integra CCM-D Patient manual

Huntleigh
Huntleigh Sonicaid FM800 Series Instructions for use

Auxilab
Auxilab MiniFuge-GD06 Operation and maintenance manual

ClearCount
ClearCount SmartSponge Operating procedures manual