Healy Wellness User manual

Healy Wellness
Instructions for Use
Frequencies for Life

66 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
Table of contents
1. General 68
2. Healy 68
2.1 Intended Use 68
2.2 Warnings 68
2.3 Description 68
2.3.1 Hardware Description 68
2.3.2 Software Description 68
2.4 General 69
2.5 Safety Notes 69
3. Preparations for using the Healy 70
3.1 Installation 70
3.1.1 Smartphone requirements 70
3.1.2 Android operating system 71
3.1.3 iOS Operating System 71
3.1.4 Hardware 71
3.1.5 Pairing via Bluetooth 71
3.1.6 Connection of the electrodes 72
4. Operating the Healy 73
4.1 Starting the application using the smartphone 73
4.2 Handling of the Healy hardware 75
4.3 Starting the application without the
smartphone 75
4.4 Stopping the application 75
5. Cleaning, disinfection and maintenance 76
5.1 General 76
5.2 Cleaning 76
EN |
Healy Instructions for Use
Disclaimer: While scientic research underlies Healy technology,
its connection to health and wellness has not been extensively
explored or demonstrated. The Healy is not intended to cure, treat,
mitigate, diagnose or prevent disease, but rather to support ener-
getic balance and enhance recovery, vitality and wellbeing.

EN | 67
T. +49 30 54905949-1 | www.healy.eu | info@healy.de
5.3 Storage between applications 76
5.4 Maintenance 76
5.5 Troubleshooting 76
6. System Components 77
7. Technical Data 77
7.1 General 77
7.2 Electrical parameters 78
7.2.1 Power source requirements 78
7.2.2 Healy 78
7.3 EC-Declaration of Conformity Under the
Directive 78
7.4 Mechanical parameters 79
7.5 Environmental conditions 79
7.5.1 Operation and storage 79
8. Disposal 79
9. Used Symbols and Icons 79
10. Revision status 82
11. Attachment instruction for use “Electrodes” REF:
108-4061 bracelet electrode (black, carbon) 82
11.1 Description 82
11.2 Instructions for use 82
11.3 Warnings 82
11.4 Product life 82
12. Attachment instruction for use “Electrodes” REF:
108-4004 ear electrodes (pair) 83
12.1 Description 83
12.2 Care and storage of the electrodes 83
12.3 Service life of the product 83
12.4 Specications 83
12.5 Important note on using the ear electrodes 83

68 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
1. General
Thank you for choosing Healy. Please read this manual carefully
and pay special attention to the warnings and cautions.
This manual is intended for the user. Intended users include
adults at home. It aims to ensure safe handling and enables
users to become familiar with the device.
2. Healy
2.1 Intended Use
Healy is a microcurrent device that uses specic frequencies to
promote energetic balancing. It can be applied to dierent parts
of the body.
2.2 Warnings
•Not for use by pregnant women
•Not for use by those with epilepsy
•Do not place the electrodes directly on the eyes, covering the
mouth, on the front of the neck (especially the carotid sinus), or
on the chest and the upper back or crossing through the heart.
2.3 Description
The Healy is a portable device, internally powered and
controlled via Bluetooth by a software application that is ins-
talled on a Smartphone. It has one output. The output can be va-
ried between -10V and +10 V, 0 to 1 MHz, 0 – 4 mA.
The charging of the internal battery is done through the Micro-
USB-Port and the supplied power supply cable.
The software application “Healy APP“ is necessary for congura-
tion and control of the Healy. The “Healy App“ is available in the
Play Store or the App Store. Before installation verify the smart-
phone requirements in section 4.1.1.
2.3.1 Hardware Description
Figure 2 .3.1.1: Healy top view:
Pos. 1 Icon and LED for Blue-
tooth Connection; Pos. 2 On-
O Button; Pos. 3 Battery
icon, Pos. 4 LED for genera-
tor action (if stimulator out-
put is active, then the LED is
blinking) and battery indica-
tor; Pos. 5 Generator icon;
Pos. 6 Icons for the positive
and negative electrode plugs;
Pos. 7 USB-Connector icon;
Pos. 8 Socket for power supply
through micro-USB–connector (not visible in this gure); Pos. 9
Sockets for the electrodes (not visible in this gure)
2.3.2 Software Description
The “Healy APP“ is required for conguring and controlling the
hardware. It allows sending application programs to the device.

EN | 69
T. +49 30 54905949-1 | www.healy.eu | info@healy.de
With the App you can start and
stop the application. The “Hea-
ly APP“ connects to your Healy
hardware using Bluetooth. The
“Healy APP“ can connect only to
one device at a time.
Figure 2.3.2.1: Overview in-
terface of the Healy APP. Pos.
1-Icon for the signal quality of
the Bluetooth connection to the
Healy; Pos. 2-Icon for the batte-
ry status of the Healy
2.4 General
•Intentional misuse can be har-
mful or reduce the eective-
ness of the device.
•Carefully read the instructions
for use before use.
•This product is intended only
for the purpose described in
these instructions for use.
•In this manual, the word DANGER warns of possible personal
injury!
•In this manual, the word WARNING warns of possible damage
caused to property!
2.5 Safety Notes
Danger/Warning:
•Keep this device out of the reach of children.
•Do not immerse the device in liquids.
•Do not use while driving motor vehicles or operating heavy
machinery. The ability to drive or to operate heavy machinery
may be impaired immediately after application.
•Only charge with the supplied cable from an USB-Socket with
5V dc and max 2500 mA.
•In the case of a defect please contact the manufacturer. Any
attempt to repair the device will void the warranty.
•Use with care as the cables may cause a tripping or strangu-
lation hazard.
• Don’t use in presence of ammable gases or in an oxygen en-
riched environment.
•Stop application and remove the electrodes if any irritation,
sensation, allergic reaction or burning occurs at the electrode
contact points and contact us. Pay attention when using the
device and currents higher than 2mA (ear-clip electrodes) or 4
mA (surface electrodes).
•Take care that the units output connectors are always pointed
towards the oor. This way the device is protected against da-
mage by water or liquids.
•The device should only be stored and transported under the
conditions dened in these instructions.
Figure 2.3.2.1

70 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
•The device is designed for indoor use only.
•Do not use the device in shielded locations, such as intensive
care units in hospitals, airplanes or wherever the use of mobile
phones is prohibited.
•The device must be operated only with supplied accessories
and together with compatible smartphones.
•Not for use by those with a pacemaker or heart disease
•Avoid damages to the electrode cables and the charging cables
caused by sharp objects. If the cables are damaged do not use
them and contact the manufacturer for a replacement. Don’t
use cables that are damaged.
•To protect the device and its accessories against contamina-
tion and damage keep it closed and away from children, dust,
sunlight, moisture, pets and vermin.
•Only use cables and electrodes supplied with the equipment.
Other accessories, cables or electrodes could result in higher
elec tromagnetic eld emissions or less electromagnetic immu-
nity and may lead to loss of function.
•Use of this equipment adjacent to or stacked with other equip-
ment should be avoided because it could result in improper
operation.
•Do not use the adhesive electrode if the gel is dry or visible da-
mage is present or the connecting cable is damaged.
•The equipment and the connecting leads used with this device
must all be in good condition.
•Do not use if recurrent episodes of skin irritation occur.
• Do not apply electrodes on skin aected by abrasions, rash or
wounds.
•Do not carry out maintenance while the equipment is in use.
• Operation in close proximity (i.e., less 3 feet) to a shortwave or
microwave therapy ME equipment may produce instability in
the stimulator output.
3. Preparations for using the Healy
3.1 Installation
Before putting the Healy into operation you need to prepare the
hardware and install the “Healy APP“.
3.1.1 Smartphone requirements
Bluetooth Bluetooth 4.0 or higher
Operating system iOS1version 9.0 or higher
Android2version 4.4 or higher
1iOS is a registered trademark of the operating system for
smartphones and devices manufactured by Apple Inc.
2
Android is a registered trademark of Google Inc. for the opera-
ting system for smartphones and other devices

EN | 71
T. +49 30 54905949-1 | www.healy.eu | info@healy.de
3.1.2 Android operating system
•Open the Play Store3
•Search for the “Healy APP“
•Choose “Healy APP“
•Download and install on your smartphone
3.1.3 iOS Operating System
•Open the Apple Store4
•Search for the “Healy APP“
•Choose “Healy APP“
•Download and install on your smartphone
3Play Store is the trademark of Google Inc.
4App-Store is the trademark of Apple Inc.
3.1.4 Hardware
•Unpack the Healy and check for damage
•Charge the batteries for 1.5 hours before rst use
• Switch on the Healy device by briey pressing the on-switch
and verify that the right LED is blinking
Note: The battery cannot be removed by the user.
3.1.5 Pairing via Bluetooth
After installation of the App, switch on the Healy by briey
pushing the on-switch (see gure 2.4.1.1) of the device..
After starting the App there is a short starting screen (see gure
3.1. 5.1).
After the App is loaded it is scanning for Bluetooth devices (see
gure 3.1.5.2).
Figure 3.1.5.2: Scanning screen
Figure 3.1. 5.1: Starting screen

72 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
Within 20 seconds the App scans the environment and connects
to the Healy automatically. If your smartphone detects a Healy
gure 3.1.5.3 appears.
After you scan the QR-Code on the back of the Healy or enter the
serial number of the device you can connect to the Healy by pres-
sing the “Connect” button (see gure 3.1.5.4).
3.1.6 Connection of the electrodes
The electrodes have been tested and evaluated together with
the Healy and are safe and eective for use, if the recommenda-
tions and advice in this manual are followed. Check the package
containing the electrodes and make sure the gel is intact; if the
package is visibly dented or damaged, do not use the product.
Put the plugs of the electrode cables into the sockets of the Hea-
ly. Place the electrodes onto the desired area.
Important note on using the adhesive electrodes:
•Never place electrodes in the region of the carotid artery
•Do not place the electrodes on open wounds
•Before use clean the application area and if necessary shave
the area
•Choose an area free from hair, skin lesions or scars, folds in the
skin, or bony protuberances and which is not in the vicinity of
metallic implants.
•Clean and dry the application area before placing the elec-
trode on it
•Avoid the use of solvents as these could cause skin irritation
and reduce the adherence of electrodes.
•If using alcohol to clean the area, make sure that it has comple-
tely dried before applying the electrodes.
Figure 3.1.5.3: Scan serial
number screen
Figure 3.1.5.4: Connect screen

EN | 73
T. +49 30 54905949-1 | www.healy.eu | info@healy.de
•Connect the electrode to the cable of the device.
•Check the functioning of the system before beginning the
procedure.
•If the procedure is interrupted, check the functioning of the
system again before recommencing the session.
•Remove the liner and apply the electrode to the skin surface.
•Before switching on the device, check the adherence of the
electrode to the skin surface.
WARNINGS:
•A reduced contact between the skin and the electrode may
cause the electric current to concentrate in one point and
cause burning.
•Interrupt the application if there is discomfort or a burning
sensation.
• After the session, turn o the device, disconnect the cable
from the electrode and gently remove the electrode from the
skin by lifting it up by one corner.
•Rapid removal of the electrode could cause skin traumas.
•Do not remove by pulling the connection part of the cable.
•After use, reposition the electrodes in the plastic liner and put
them back in the pouch.
•Electrodes may be reused: Reuse only for the same person.
• Verify the adhesion of the electrode before use. When electro-
des lose adhesion, gently rub one or two drops of water onto
the gel surface. This may extend the usability. If there is no im-
provement, replace with new electrode.
•Keep the packet closed in order to prevent the unused electro-
des from becoming dehydrated.
•Do not reposition the elec-
trode during use.
4. Operating the Healy
You should become familiar
with these frequently used
operations:
• switch On-O (please refer
to chapter 3.1.5 and 4.4)
•select application (please re-
fer to chapter 4.1)
•start application (please re-
fer to chapter 4.1)
•connect the electrodes
(please refer to chapter 3.1.6)
4.1 Starting the appli-
cation using the smart-
phone
After the installation of the
App, pairing and connection
of the electrodes, the Healy
is ready for operation. In the
Figure 4 .1.1: Overview interface

74 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
After selecting a program category the program page with the
icons of the dierent programs is loaded (see gure 4.1.2).
Figure 4.1.3: Information page
for the placing the electrodes
Figure 4.1.4: Application page Figure 4.1.5: Application page
with white time indicator –
electrical circuit is closed
overview screen you can select which program category you
want to use (see gure 4.1.1).
Figure 4.1.2: Program page

EN | 75
T. +49 30 54905949-1 | www.healy.eu | info@healy.de
4.3 Starting the application without the smartphone
The device always stores the last application program. By pressing the
on-switch of the Healy hardware,
the last program that was sent
from the smartphone to the device
will start with an intensity of 15%.
4.4 Stopping the
application
You can see the current pro-
gress of the program running
on the device by using the App.
•To stop the application select
one of the following:
• briey press the on/o-switch
(pause mode)
•tap the icon in the application
mode in the app (see gure
4.4.1)
• hold the on/o-switch for
about 10 seconds until device
hardware resets
•in emergency situations it is
possible to remove the elec-
trodes or disconnect the elec-
trode cables
After you select a program from the program page screen the in-
formation for placing the electrodes is displayed (see gure 4.1.3).
After selecting “Start” on the
information page the applica-
tion is started and the applica-
tion page is loaded (see gure
4.1.4). In the application page
you can set the intensity of the
application by using the plus
and minus buttons.
When the electrical circuit
is closed the circle, which is
an indicator for the time, will
switch from black to white (see
gure 4.1.5).
4.2 Handling of the Healy
hardware
During the application you
should use the clip on the
back of the device to attach
the Healy to your clothing. The
electrodes should be axed
as shown in the App (please
refer to chapter 10). Route the
electrode cables to the Healy
so that you feel comfortable.
Figure 4.1.6: Application page
with black time indicator –
electrical circuit is open
Figure 4.4.1: Pause screen.
By pressing the “X” in the
pause screen you cancel the
program and get back to the
overview interface.

76 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
5. Cleaning, disinfection and maintenance
5.1 General
Maintenance activities increase the lifespan and safety of the
Healy. Cleaning and disinfection reduces the risks of cross conta-
mination and infections. Please read this section carefully.
5.2 Cleaning
Switch o the Healy before cleaning. For cleaning, commercial
cleaning products may be used. The surfaces can be wiped with
a damp sponge or cloth. Do not use too much liquid and make
sure that no liquid penetrates beneath the covers or enclosures.
Caution:
Do not use cleaning agents with solvents (e.g. acetone, nail poli-
sh, cleaners) or abrasive cleaners as they will damage the surface.
5.3 Storage between applications
After the application unplug all cables and electrodes and put
them into the storage case and store in a cool and dry location.
Refer to the safety notes for environmental storage conditions.
5.4 Maintenance
There are no serviceable parts by the users. In a case of damage,
please contact the manufacturer. The following maintenance
activities are recommended and should be considered for safe
working.
Frequency By whom Maintenance activity
Prior to each use User Check housing, cables and
electrodes for damages.
In case of damage, contact
the manufacturer or an
authorized dealer
5.5 Troubleshooting
In the following you nd typical warnings or messages that are
related to the app or hardware of the device. If you face an event
that is not listed here or a problem still exists, please contact the
manufacturer.
Event or message Cause Actions required
No connection from
the smartphone to
the device
Device is
switched o
Switch on the device
and restart the App
No connection from
the smartphone to
the device
Bluetooth in
your smart-
phone is
deactivated.
Open connectivity
settings and activate
Bluetooth
Repeated short
blinking of right LED
Batteries are
low
Charge the device

EN | 77
T. +49 30 54905949-1 | www.healy.eu | info@healy.de
Event or message Cause Actions required
Device is not
responding
No Stimulation
output
Undened
error/EM dis-
turbance
Hold the on-switch
for 10 seconds. This
triggers a hardware
reset.
Undened
error/EM dis-
turbance
Contact your autho-
rized dealer or the
manufacturer Hea-
ly GmbH
6. System Components
The Healy is a set of electrodes and a charging cable.
Reference
number
Description Quantity
0006 Healy hardware device 1
108-4063 Press button adhesive electrodes
round Ø 32 mm
4
108-4062 Connection cable for electrodes 96
cm – press button on 2 mm
2
Reference
number
Description Quantity
108-4061 Bracelet electrode (black, carbon) 2
108-4004 Ear electrodes (pair) 1
108-4096 Felts 20
108-4068 Charging cable USB 0.15 m 1
7. Technical Data
7.1 General
Protection against in-
gress of water and parti-
culate matters
IP 42, Protected against particu-
late matter, larger than 1mm. Pro-
tection against dripping water at
an angle of 15°.

78 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
7.2 Electrical parameters
7.2.1 Power source requirements
The Healy can be connected to any power source with the sup-
plied USB-charging cable and the following specications:
Supply voltage range 4.4 V – 5.5 V DC
Output current Max. 2500 mA
Frequency DC
Socket type USB- Type A
Power type High powered 500 mA or higher
Insulation type Double or reinforced insulation
Compliance To IEC60950 or IEC60601-1
7.2.2 Healy
Input voltage See 7. 2.1
Input Current 20 mA
Protection class Internally powered
Battery capacity 145 mAh
Battery charging time 60 min – 90 min
Charging cycles Life
expectancy
500 times, more than 80% of initial
capacity at 20°C
Output voltage -10 V to +10 V
Output current Max. 4 mA (voltage control) / ±2,6 mA
(current control)
Accuracy voltage ±0.1 V with ±10 %
Accuracy current ±20 µA with ±10 %
Output frequency DC up to 1 MHz
Accuracy current ±20 µA with ±10 %
Output frequency –
Stimulation
DC up to 1 MHz
7.3 EC-Declaration of Conformity Under the Directive
2004/108/EC of the council regarding
electromagnetic compatibility
This is an electrical device for application on humans.
Type: Healy Year of manufacture: 01-2019 – 12-2019
The product is designed and manufactured in accordance with
the following directives:

EN | 79
T. +49 30 54905949-1 | www.healy.eu | info@healy.de
2014/30/EU Electromagnetic compatibility
2014/35/EU Low Voltage Directive
2011/65/EU ROHS - Directive
under the sole responsibility of:
Company: Healy GmbH|Darritzer Str. 6|16818 Kränzlin – Germany
The device complies with the following harmonized norms:
EN 61000-3-
EN 55014
EN 61558
EN 60335
The technical documentation with risk analysis is completely
available.
The instructions for use accompanying the product are available
in the respective national language.
Kränzlin/13.03.2019
Place/date Signature CEO
7.4 Mechanical parameters
Dimensions 55 x 57 x 13 mm
Weight 32 g
Cable length charging cables 0.15 m
Cable length electrode cables 96 cm (REF: 108-4062)
Output socket 2 x 2 mm pin
Input socket USB-Type B micro
7.5 Environmental conditions
7.5.1 Operation and storage
Temperature Range 5°C to 40°C / 41°F to 104°F
Relative humidity 15 % to 93 %
Atmospheric pressure 700 hPa to 1060 hPa
8. Disposal
The expected service life is 5 years. Ask your local authority for
recycling advice.
9. Used Symbols and Icons
The following icons are used on the device, the packaging and
the labelling:

80 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
Use only inside
Electronic waste – do not
dispose. Recycle acc. to local
regulations or contact your
authorized dealer or the
manufacturer.
Recycling symbol
SGS US CAN certication
Mark
In conformity with European
legislation Keep away from heat
Bluetooth Read instructions for use
before use
Symbol for Bluetooth Wire-
less connection
Caution: Read instructions
for use before use. Follow
safety advices in this manual.
Temperature limitations Symbol for Battery
Follow the instructions for use Best before Symbol for on- and o-switch
Keep dry Hardware reference number Symbol for USB-Connection
One patient use Plus icon for positive elec-
trode socket
Minus icon for negative elec-
trode socket
REF 0006

EN | 81
T. +49 30 54905949-1 | www.healy.eu | info@healy.de
Symbol for generator operation – stimulation output
The following symbols are used in the Healy – App
Icon for over-
view interface
Icon for the
trademark Healy
Battery status of
the Healy hard-
ware
Icon for the blue-
tooth signal
strength between
the Healy App and
Healy hardware
Pictograph for the placement of bracelet electro-
des. Appears after selection of the program
Note: the background colour may vary, depending
on the selected and congured program parameter
Button for increasing
the intensity in percen-
tage (current or voltage)
Button for decreasing the intensity in
percentage (current or voltage)
Icon for
“Meridians”
program.
Symbol with informa-
tion about actual inten-
sity of the application
Symbol for scanning and connecting
with the Healy hardware, appears
after starting the app and during
establishing the Bluetooth connection
Button for
scrolling to the
rst page with
programs

82 | EN Healy GmbH |Schloss Kränzlin |Darritzer Str. 6 |16818 Kränzlin – Germany
Button for jumping two
pages backwards
Button for jumping two pages
forward
Button for scrolling to the
last page with programs
Button for moving one
page back
Button for moving one page
forwards
Note: The background colour may vary,
depending on the selected and con-
gured program parameter
10. Revision status
Revision Date Remark
021th October 2016 First edition
11. Attachment instruction for use “Electrodes” REF: 108-
4061 bracelet electrode (black, carbon)
11.1 Description
The bracelets are made of conductive fabric and hydratable
material or conductive elastic fabric.
11.2 Instructions for use
•Check that the packaging is not damaged. If it is damaged or
has visible defects, do not use the product.
•Connect the bracelet of the equipment selected, with the
connection cable.
• When the session is complete, switch o the equipment and
disconnect the cable.
11.3 Warnings
•Do not apply the bracelet onto the skin in the presence of abra-
sions, erythema or wounds.
11.4 Product life
The life of the product is a direct result of the amount of use and
the conditions of maintenance and storage of the product itself.
As soon as any signs of material deterioration appear the pro-
duct must be eliminated.

EN | 83
T. +49 30 54905949-1 | www.healy.eu | info@healy.de
12. Attachment instruction for use “Electrodes” REF:
108-4004 ear electrodes (pair)
12.1 Description
Only use the ear electrodes in combination with the supplied ad-
hesive felts. In people with very dry skin, problems may occur due
to insucient electrical contact in individual cases. Therefore,
always ensure that the felts are soaked with physiological saline.
In the application with ear electrodes the intensity should be
set that the current is slightly perceptible. A too high current
intensity can cause dizziness and nausea. During the applica-
tion mood changes can occur, such as a feeling of heaviness or
lightness. In such cases, you should continue the application for
a few minutes until you feel normal again.
12.2 Care and storage of the electrodes
Add one to two splashes of a mild detergent to warm water.
Clean the electrode with this water and a soft cloth (or sponge)
thoroughly and rinse o after that. Do not immerse the elec-
trode in the liquid. For storage put the electrode back into the
packaging until next use and store it in a dry, clean place.
12.3 Service life of the product
The service life of the product depends on the functionality
which is in turn dependent on the number of applications, the
state of maintenance and the condition.
Upon recognition of any damages, the product shall not be used
anymore and shall be disposed.
12.4 Specications
Connectors: 2 mm connector
12.5 Important note on using the ear electrodes
•Prior to the application, the electrodes must be equipped with
the supplied self-adhesive felts!
•The felts must be well-moistened throughout the entire appli-
cation.
•In case of any kind of discomfort, such as stinging, burning,
or pain stop the application or reduce the intensity immedia-
tely until no unpleasant symptoms remain. If necessary, check
the contacts and improve the conductivity by remoistening
the felts..
Healy GmbH
Darritzer Str. 6
16818 Kränzlin, Germany

Healy GmbH | Schloss Kränzlin | Darritzer Str. 6 | 16818 Kränzlin – Germany
T. +49 30 54905949-1 | www.healy.eu | info@healy.de
Frequencies for Life
Other manuals for Wellness
1
Table of contents
Other Healy Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Dynatron
Dynatron Dynatronics ibox Operator's manual

Pentax
Pentax 9106 operating manual

BLADDERSCAN
BLADDERSCAN BVI 6100 Operation & maintenance manual

Etac
Etac Nova Standard ampu manual

Contec Medical Systems Co.
Contec Medical Systems Co. EKG6012 user manual

Cardiac Science
Cardiac Science POWERHEART G3 PLUS 9390E AED Quick reference guide