3.1 GENERAL
CURRENT RANGE:
0.1A - 400A a.c. RMS
OUTPUT VOLTAGE LEVELS:
10mV/A: 1mV a.c. per 0.1 Amp a.c.
1mV/A: 1mV a.c. per 1 Amp a.c.
WORKING VOLTAGE:
600V CAT.Ⅱper IEC 1010 - 1
OPERATING TEMPERATURE:
0℃to 50℃80 relative humidity
STORAGE TEMPERATURE: -20℃to 60℃
TYPE OF SENSING:
induction coil sensing for AC current
MAXIMUM OUTPUT IMPEDANCE: 45
MAXIMUM JAW OPENING: 29mm
MAXIMUM CONDUCTOR SIZE: 28mm
SIZE: 64(W)×143(L)×30(D)
WEIGHT:185g
4
3.2 ELECTRICAL SPECIFICATION
Frequency Response: 50-60Hz
4. PRECAUTION AND PREPARATIONS
FOR MEASUREMENT
Dot not apply the voltage to the output
plugs.
Do not use or store this instrument in a
high temperature or high humidity
environment and do not store the unit
indirect sunlight.
Do not measure current before the unit
is not combined with DMM.
If the instrument is used near noise
generating equipment, be aware that
output voltage may become unstable or
get more errors.
5
This instrument must not be used
on
uninsulated conductors at a voltage
greater than 250V ac rms. or 250V
dc
5. AC CURRENT MEASUREMENT
Set the DMM at ACV function and a
relevant range.
Connect the plug of transducer with a
DMM and combined firmly.
Press the trigger to open transformer
jaws and clamp one conductor only,
making sure that the jaws is firmly
closed around the conductor, then read
the reading from the digital display of
DMM.
6. MAINTENANCE
to keep the instrument clean, wipe the case
with a damp cloth and detergent, do not use
abrasives or solvents.
Any adjustment, maintenance and repair
shall be conducted by a service personnel.
6