Henkel LOCTITE ZETA 7760A User manual

EQUIPMENT
Operation Manual
7760A
Exposure Time
Timed
Manual
Lamp Hours
TIMER LT4H
RESET
LOCK
NiS
UP
DOWN
CURTIS
HOURS
1
10
UV
Lamp & Reflector
DescriptionP/N
98451
97328
98543
97326
Spare Components
Item No. 1329221
Warning: UV Energy is transmitted from the end of the
are required that meet ANSI Z80.3 & Z87.1 Certification.
light guide. Protective eyewear equipped with side shields
Dual Light Guide, 1M
Single Light Guide, 1.5M
Single Light Guide, 1M
Lamp On
Loctite® Zeta® 7760A UV Curing Wand System
Part Number: 1329221

TABLE OF CONTENTS
1. PLEASE OBSERVE THE FOLLOWING............................................................................3
1.1 EMPHASIZED SECTIONS...............................................................................................................................3
1.2 ITEMS SUPPLIED ..........................................................................................................................................3
1.3 FOR YOUR SAFETY......................................................................................................................................3
1.4 FIELD OF APPLICATION,(INTENDED USAGE)...............................................................................................4
2. DESCRIPTION.........................................................................................................................................4
2.1 THEORY OF OPERATION ..............................................................................................................................4
2.2 OPERATING ELEMENTS AND CONNECTIONS,REFERS TO FIGURE 1..............................................................5
3. TECHNICAL DATA.............................................................................................................................7
3.1 ENERGY REQUIREMENTS.............................................................................................................................7
3.2 DIMENSIONS...............................................................................................................................................7
3.3 UV OUTPUT CHARACTERISTICS .................................................................................................................7
4. INSTALLATION....................................................................................................................................7
4.1 SPACE REQUIREMENTS................................................................................................................................7
4.2 CONNECTIONS.............................................................................................................................................9
5. OPERATING THE UNIT..................................................................................................................9
5.1 INSERTING AND REMOVING THE LIGHT GUIDES..........................................................................................9
5.2 POWERING UP...........................................................................................................................................10
5.3 ADJUSTING DUAL WAND FOR MAXIMUM OUTPUT....................................................................................11
5.4 USING FOOT SWITCH /REMOTE DEVICE ...................................................................................................12
6. CARE AND MAINTENANCE.....................................................................................................13
6.1 REPLACING THE LAMP MODULE ...............................................................................................................13
6.2 CHANGING THE OPTICAL FILTER ..............................................................................................................15
REPLACEMENT PROCEDURE............................................................................................................................15
7. TROUBLESHOOTING....................................................................................................................16
8. DOCUMENTATION...........................................................................................................................17
8.1 WIRING DIAGRAM.....................................................................................................................................17
8.2 PIN CONNECTIONS.....................................................................................................................................18
8.3 REPLACEMENT PARTS AND ACCESSORIES.................................................................................................18
9. WARRANTY............................................................................................................................................19

1. Please Observe the Following
1.1 Emphasized Sections
WARNING!
Refers to safety regulations and required measures that protect the operator or other persons
from injury or danger to life.
Caution!
Emphasizes what must be done or avoided so that the unit or other property is not
damaged.
Notice:
Gives recommendations for better handling of the unit during operation or adjustment, as
well as for service activities.
1.2 Items Supplied
1 ZETA® 7760A UV Curing Wand System
1 UV lamp/reflector assembly (Customer to Install)
1 Pair of UV protective glasses
1 Ballast jumper wire (input voltage selection)
1 Foot switch
1 Users manual
1 Power cord
1.3 For Your Safety
For safe and successful operation of the unit, read these instructions completely. If the
instructions are not observed, the manufacturer can assume no responsibility. Be sure to
retain this manual for future reference.
WARNING!
Always wear the included UV safety glasses or glasses that conform to ANSI Z87.1/CSA
Z94.3 when operating the unit.
WARNING!
Always cover hands, face and other parts of the body that may be exposed to UV light.
WARNING!
Never look into the end of the light guide.
WARNING!
Never open the shutter mechanism without the light guide installed.
WARNING!
Never remove the cover of the unit without first switching the power off and unplugging
the power cord.
WARNING!
Damage to the power cord or the housing can result in contact with live electrical parts.
Check the power cord and housing before each use. If the power cord or unit is damaged,
do not operate.
The unit may be repaired only by a Loctite authorized service technician.

1. Please Observe the Following (continued)
Caution!
Never turn the unit on without the lamp connected to the power supply.
Caution!
The energy emitted from the end of the light guide can heat any surface that it is directed
at. Care must be taken to determine the proper offset distance and exposure time.
Caution!
Turning the lamp on and off frequently will cause the UV output of the lamp to decline at
a faster rate. It is recommended that the unit be left on during breaks and short down
times.
Caution!
Avoid making sharp bends in the light guide, as this will cause a loss of UV energy or
possibly cause permanent damage. To prevent permanent damage, the minimum bend
radii are 2.4 inches for a single light guide and 1.6 inches for a dual light guide.
Caution!
The AC requirements for the Zeta 7760A are as follows;
88 to 136 VAC and 176 to 264 VAC. When operating between 88 to 136 VAC the
enclosed jumper must be installed across pins ST4 and ST5 on the ballast.
1.4 Field of Application, (Intended Usage)
This Loctite® ZETA® 7760A UV Curing Wand System is designed for use with
Loctite® products that cure when exposed to ultraviolet light produced by the unit. The
UV energy is directed towards the product through a liquid filled light guide that is
ordered separately. Several configurations of light guides are available from Loctite,
including single and dual ended types. The unit is also capable of interfacing with an
external relay or PLC circuit.
2. Description
2.1 Theory of Operation
When the unit is switched on, electrical power is immediately supplied to the lamp and ignition
should occur within several seconds. As soon as the lamp ignites, the green “Lamp On”
indicator, located on the front panel, will come on. Also, an internal relay will change state,
making a closure across pins three and four of the footswitch connector located on the rear panel
of the unit. It will take several minutes for the lamp to reach full power. Curing takes place when
the shutter is opened, allowing UV light to be directed from the end of the liquid filled light
guide to the adhesive. The curing parameters, (UV irradiance and exposure time) must be
determined before use. The UV irradiance, (the strength of the UV light), at the adhesive surface
can be varied by adjusting the distance of the light guide from the bond. If the light guide is
moved to within ½ inch of the adhesive surface, vapors from the curing process may build up on
the end of the light guide, reducing the UV intensity. The UV exposure time required to
complete the curing process depends primarily on the UV irradiance and the properties of the
adhesive product.

The timed exposure cycle starts by momentarily engaging the footswitch or when an externally
operated device such as a PLC makes a relay closure across pins 1 and 9 of the nine pin foot
switch connection on the rear panel. In the manual operating mode, the shutter remains open for
as long as the footswitch or alternate external device maintains contact between pins 1 and 9.
The hour meter, located on the front panel, records the total hours of operation accumulated by
the lamp. UV lamps undergo a gradual reduction in UV output over time and it is recommended
that the lamp output be monitored on a regular basis. Frequently turning the power on and off
will increase the rate that the lamp degrades, therefore, it is recommended that the unit be left on
during short work breaks such as lunch and other idle periods.
Notice: The hour meter should only be reset when a new lamp is installed.
2.2 Operating Elements and Connections, refers to Figure 1
1. Power Inlet Module
Connect line cord to power inlet module.
2. Power Fuse Holder
Fuse is located in the power module.
3. Power Switch
4. Foot Switch Connection
Standard 9 pin “D” connector for foot switch or other external switch.
5. Light Guide Receptacle
Is used to retain the light guide.
6. Digital Exposure Timer
Controls the period of time the shutter is opened. The UV time is set by pressing the
buttons located directly under the LCD, which indicates seconds. The manual/timed
selector switch must be set to "timed.” When the foot switch is pressed the shutter
mechanism will open for the indicated time and the timer will begin counting down.
When the unit times out the shutter will close and the LCD will indicate the original
pre-set time interval.
7. Lamp Hour Meter
The hour meter indicates the cumulative time that the lamp has been ON.
8. Mode Selector – Manual/Timed
When set to “manual” the UV light is passed through the light guide for as long as the
foot switch is depressed. In the “timed” mode, the timer’s LCD display indicates
exposure time and the cycle is initiated when the foot switch is momentarily engaged.
9. Hour Meter Reset
Use switch to reset the hour meter to zero after installing a new lamp module.
10. Lamp On Indicator
The lamp will ignite when the main power is switched on. The green lamp on
indicator located on the front panel will light to confirm that the lamp is operating.

2. Description (continued)
7760A
Exposure Time
Timed
Manual
Lamp Hours
TIMER LT4H
RESET
LOCK
NiS
UP
DOWN
CURTIS
HOURS
1
10
UV
Lamp & Reflector
DescriptionP/N
98451
97328
98543
97326
Spare Components
Item No. 1329221
Warning: UV Energy is transmitted from the end of the
are required that meet ANSI Z80.3 & Z87.1 Certification.
light guide. Protective eyewear equipped with side shields
Dual Light Guide, 1M
Single Light Guide, 1.5M
Single Light Guide, 1M
11 1/2
8 5/16
1
2
3
Power
Switch
Fuse
Holder
Power
Inlet
Module
4
Foot
Switch
Connection
5
Light
Guide
Receptacle
6
78
9
Hour Meter Reset
Lamp: 200 Watt
Made in U.S.A. Fuse: 4 Amp, 250 V
Input: 88-136; 176-264 VAC, 47 to 440 Hz.
3.0 A @ 120 VAC, 1.7 A @ 240 VAC
Rocky Hill, Connecticut 06067, U.S.A. Zeta 7760A Wand System
Loctite® Item No: 1329221
Serial No: 4CWA-01
FOR SPARE PARTS, MANUALS, REPAIRS,
OR TECHNICAL ASSISTANCE:
Visit us at equipment.loctite.com
or call Henkel Corp. at:
USA - (1) 860-571-5174
Germany - (49) 89-9268-0
Singapore - (65) 6482-3881 Label P/N: 988341
O
Lamp On
10
Warning Manual Before Servicing Unit.
Disconnect and Refer to
Warning
FOR 88-136 VAC,JUMPER REQUIRED ACROSS PINS ST4 AND ST5 ON BALLAST.
FOR 176-264 VAC, REMOVE JUMPER FROM PINS ST4 AND ST5 ON BALLAST.
(SEE MANUAL PAGE 9)
Figure 1.

3. Technical Data
3.1 Energy Requirements
Input Voltage Ranges: 88-136 and 176-264 VAC
Operating at 88 to 136 VAC requires that the enclosed jumper be installed on
pins ST4 and ST5 on the ballast.
Input frequency Range: 47 to 440 Hz
Auxiliary Output Voltage: 27 volts DC, nominal
Main fuse located in power module in rear of unit: 4A/250V, 5mm
3.2 Dimensions
Housing: Approximately 11.5 inch W x 13.5 inch D x 8.4 inch H
Weight: Approximately 18 pounds
3.3 UV Output Characteristics
Initial UV output: 14 to 18 W/cm² typical. Exact output measurement is dependent on
the brand and calibration method of the meter used, as well as the condition and
straightness of the light guide.
7760 UV OUTPUT SPECTRUM
240 265 290 315 340 365 390 415
Wavelength (nanometers)
4. Installation
4.1 Space Requirements
A space of 12" wide x 16" deep x 9" height is required. It is important to have at least 8
inches of space behind the unit to insure proper airflow.
Caution!
Do not block the intake and exhaust fans located on the back of the housing.
Caution!
The unit should always be operated with the rubber support feet resting on a flat
surface. Do not operate with the unit resting on its side or at an angle greater than 15
degrees, front to back.

4.1 Lamp Installation
IMPORTANT NOTICE
THE CURING LAMP MUST BE INSTALLED IN THE UNIT BEFORE TURNING THE POWER
ON. THE CURING LAMP IS PACKAGED INSIDE OF THE UNIT.
Lamp Retainer Wire
Lamp
Lamp Reflector
Ceramic Hub
Anode Connection
Cathode Connection
2
1
Caution!
Do not touch the inside surface of the lamp reflector or the lamp. Contaminants from
hands will create “hot spots” and cause the module to fail prematurely. If either surface
is accidentally touched, carefully wipe with a clean, soft cloth and isopropyl alcohol.
1. Make sure the unit is unplugged.
2. Remove the four screws from the sides of the housing and lift off the cover.
3. Remove the sheet metal shroud that encases the lamp.
4. Flip the swivel section of the lamp retainer wire forward.
5. Grasp the lamp assembly by the white ceramic hub and pull back on the wire
retainer until the reflector fits into the counter bore in the lamp holding plate.
6. Flip swivel retainer back over the lamp holder.
7. Connect the power leads.
8. The power lead attachment connected to the front of the reflector should be
approximately in the twelve o’clock position. If it is not, grasp the ceramic hub of
the lamp, pull back slightly and rotate until the lead is at twelve o’clock.
9. Replace the lamp shroud and the enclosure cover.
10. Press the lamp hour reset switch to set the hour meter to zero.
Notice: Step 10 should only be done when a new lamp is installed.

4.2 Electrical Connections
WARNING! Make sure the power switch is OFF and the unit is unplugged before
proceeding.
IMPORTANT NOTICE
The enclosed jumper wire must be installed on pins ST4 and ST5 if the unit is to
be operated on line voltage between 88 and 136 VAC. Do not install the jumper
if the line voltage is between 176 and 264 VAC.
+-
L (ST1)
N (ST2)
PE (ST3)
ST4 ST5 ST6 ST7
ST4 TO ST5 JUMPER
-Plug the power cord into the power module located at the bottom left of the rear
panel.
-Plug the cord into the electrical utility outlet.
-Insert the light guide into the receptacle located on the left side of the front panel until
it is fully engaged.
-Plug the footswitch into the nine pin D-sub connector located on the rear panel. Or, if
the unit is being operated from an external processor, a closure across pins one and
nine is required to actuate the shutter. An internal relay closure connected to pins
three and four of the footswitch connector provides conformation that the curing lamp
is on. When the lamp is not on, an open circuit exists across pins three and four.
5. Operating the Unit
5.1 Inserting and Removing the Light Guides
Caution!
Be sure to remove the plastic end caps before attempting to use the light guide.
Caution!
If the free end of the light guide is secured at a fixed location, sharp bends should be
avoided since this causes a decrease in UV power.
Caution!
Never pull on the jacketing portion of the light guide. During installation or removal,
grasp the light guide on the strain relief nearest the input end of the light guide.

•Installing single light guide - Insert the large end of single light guide into the
light guide receptacle located on the left side of the front panel. Push light guide
firmly until it is fully engaged.
•Installing dual light guide – Insert the dual wand light guide into the light
receptacle and rotate the collar assembly (951639) so that the scribed line on the
collar is in the 12 o’clock position. See section 5.3 “Adjusting Wand for
Maximum Output.”
Notice:
It is important that the rotational position of a dual light guide be set to maximize the
UV output for each wand. This setting is specific to each curing unit and light guide
(See section 5.3 - Adjusting Dual Wand).
5.2 Powering Up
1. Switch the main power ON. The power switch is on the lower left of the rear panel.
2. The green LED on the front panel will come on to confirm that the lamp has
ignited.
3. Allow approximately five minutes for the lamp to reach full power.
4. Set the operating mode selector switch to “timed” or “manual”.
5. For manual operation, engage the footswitch to start UV exposure and hold until the
curing cycle is complete. When the footswitch is released the exposure will end
immediately. For timed operation use the keys directly below the numbers on the
display to set the desired exposure time. Momentarily engage the footswitch to start
the cycle. The UV exposure will begin immediately and continue until the system
times out.
6. UV exposure cycles may be initiated by using an externally operated device, such
as a PLC controlled relay in place of the footswitch.
Notice:
Avoid shutting the main power off for brief periods of time. Frequent start-ups will
cause the lamp to decay at an accelerated rate. If the system is shut down, wait a
minimum of ten minutes before restarting. Once the lamp is ignited, allow it to operate
for a minimum of fifteen minutes before turning it off.

5.3 Adjusting Dual Wand for Maximum Output
7760A
Exposure Time
Timed
Manual
Lamp Hours
TIMER LT4H
RESET
LOCK
NiS
UP
DOWN
CURTIS
HOURS
1
10
UV
Lamp & Reflector
DescriptionP/N
98451
97328
98543
97326
Spare Components
Item No. 1329221
Warning: UV Energy is transmitted from the end of the
are required that meet ANSI Z80.3 & Z87.1 Certification.
light guide. Protective eyewear equipped with side shields
Dual Light Guide, 1M
Single Light Guide, 1.5M
Single Light Guide, 1M
Adapter Hub
Scribe Line This Set Screw Locks
The Balancing Adapter
To The Units Base
This Set Screw Locks
The Light Guide In
Position After Balancing
The Output Power
Lamp On
Objective
Balance the UV output from the two light guides of Loctite® P/N 97328.
Tools Needed
3/32 inch hex key
Loctite® 7020 UV Spot Radiometer P/N: 1406024
Setup
1. Install the collar assembly (951639) over the part of the light guide receptacle that
protrudes from the front of the unit. Rotate the collar assembly until the scribed line is
in the 12 o’clock position. Use the set screw to secure the collar in that position.
2. Insert the input end of the light guide into the light guide receptacle/collar assembly
pushing it as far forward as it will go. Tighten the set screw on to the light guide.
3. Turn on electric power to unit.
4. Set shutter timer to 2 seconds.
5. Insert one of the light guide ends into a Loctite® 7020 UV Spot Radiometer P/N:
1406024.
6. Press the radiometer start switch.
7. Actuate foot switch to open shutter.
8. Release the radiometer start switch while the shutter is still opened.
9. Read and record the radiometer reading.
10. Position the second light guide in the radiometer.
11. Press the radiometer start switch.
12. Actuate foot switch to open shutter.
13. Release the radiometer start switch while the shutter is still opened.

14. Read and record the radiometer reading.
15. Compare the two radiometer readings.
16. If the lower reading is equal to or greater than 90% of the higher reading, the dual
wand is balanced.
17. If the lower reading is less than 90% of the higher reading, loosen the set screw and
rotate the collar a small increment around the light guide receptacle. Use the set screw
to secure the collar in that position.
18. Take and compare another set of radiometer readings.
19. Continue until the lower reading is equal to or greater than 90% of the higher reading.
20. Tighten set screw to lock the input end of the dual wand in position.
Notice:
If a radiometer is not available, it is recommended that Loctite Service be contacted at
1-800-LOCTITE (1-800-562-8483) to insure optimum performance when installing a
new dual ended light guide.
5.4 Using Foot Switch / Remote Device
The ZETA® 7760A can also be actuated using a foot switch or by external devices. The
foot switch connection is made up of a 9 pin D connector. Making a dry-contact relay
closure across pins 1 and 9 can actuate the unit.

6. Care and Maintenance
Notice:
It is recommended that the UV output from the lamp be monitored regularly using a
Loctite® 7020 UV Spot Radiometer P/N: 1406024. If a radiometer is not used to
monitor the lamp, it is recommended that the lamp be replaced when the hour meter,
located on the front panel, indicates 1,000 hours of operating time has elapsed. It is
normal for the lamp output to slowly decline over operating time, however, the
effective life of the lamp will decrease significantly faster if it is turned on and off
frequently.
Caution!
It is recommended that the end of the light guide be positioned no closer than ½ inch
from the Loctite® product being cured. The heat transmitted by the lamp can adversely
affect the properties of the cured product or possibly cause damaged to the part surface.
Notice:
Vapors from some products may gradually accumulate on the end of the light guide,
reducing the UV output. It should be inspected regularly and cleaned as necessary using
isopropyl alcohol and a soft, clean cloth.
Caution!
Avoid placing sharp bends in the light guide. This reduces the UV output and may
permanently damage the light guide. If the light guide is mounted on a fixture, secure it
by the metal exit fitting and not the flexible plastic section.
Caution!
Be sure to remove the plastic end caps before attempting to use the lightguide.
Notice: The intake and exhaust fan filter elements should be replaced as needed to
insure proper cooling of the power supply and UV lamp assembly. They should be
inspected routinely.
6.1 Replacing the Lamp Module
Caution!
Do not touch the inside surface of the lamp reflector or the lamp. Contaminants from
hands will create “hot spots” and cause the module to fail prematurely. If either are
accidentally touched, carefully wipe them with a clean, soft cloth and isopropyl alcohol.
1. Switch the power off and unplug the unit.
2. Allow several minutes for the lamp to cool.
2. Remove the four screws from the sides of the housing and lift off the cover.
3. Remove the sheet metal shroud that encases the lamp.
4. Disconnect the lamp leads by separating the Male/Female connector.
6. Flip the wire retainer over the lamp holder.
7. To remove the lamp module grasp the white ceramic hub and carefully pull back
and lift up at the same time.
8. Install the new lamp module and flip retainer back over the lamp holder.
9. Reconnect the power leads.

6.1 Replacing the Lamp Module (continued)
10. The bare power lead attachment to the front of the reflector should be
approximately in the twelve o’clock position. If it is not, grasp the ceramic hub of
the lamp, pull back slightly and rotate until the lead is between the two white plastic
insulators.
11. Replace the lamp shroud and the enclosure cover.
12. Press the lamp hour reset switch to set the hour meter to zero.
Notice: Step 11 should only be done when a new lamp is installed.
WARNING!
The UV lamp used in this unit contains a very small amount of mercury. Disposal of
lamps should be done in accordance with state and local regulations.
Lamp Changeover Components
Lamp Retainer Wire
Lamp
Lamp Reflector
Ceramic Hub
Anode Connection
Cathode Connection
2
1

6.2 Changing The Optical Filter
The 7760A uses a special optical filter designed to reduce heat. It is retained by two clips
and thumbscrews in front of the light receiving end of the light guide. The filter that comes
with the unit, P/N 98599 is designed to allow the use of the standard light guides. This
arrangement will transmit the wavelengths needed to obtain excellent surface cure and cure
depth in most applications. In cases where it is difficult to achieve dry to the touch surfaces,
an alternate filter, P/N 98598, is designed to improve the surface cure ability. P/N 98598
must be used with the extended range light guides.
Replacement Procedure
1. Unplug the unit and remove the cover.
2. Remove the heat shroud, exposing the lamp and optical staging.
3. Remove the light guide from the unit and turn the unit onto its front side.
4. Looking down at the light guide entrance hole, loosen the two thumbscrews securing
the filter.
5. Swing the two retaining clips out of the grooves allowing removal of the filter.
6. Set the unit back onto its feet and the remove the filter from the unit.
7. Set the unit back on its front side and insert the new filter into the hole.
8. Replace the retaining clips and thumbscrews.
Caution! Avoid touching the surfaces of the filter by holding the edges only. If the
surfaces are touched or become dirty use a clean, soft wiper and isopropyl alcohol to
remove the contaminants.

7. Troubleshooting
Type of Malfunction Possible Causes Corrections
Power does not come on. - No voltage present. Check wall circuit.
- Defective fuse. Replace fuse.
- Defective line cord Replace.
Lamp does not light. - Lamp leads are not connected.
- Defective power supply Call 800-562-8483
for technical assistance
- Lamp failure. Replace lamp module.
Power comes on but light is - Footswitch is not plugged Check footswitch
not emitted from the light- in. connection.
guide. The shutter mechanism
makes no noise when the - Defective footswitch. Call 800-562-8483
footswitch is engaged. for technical assistance
- Defective shutter mechanism. Call 800-562-8483
for technical assistance
- Faulty power supply Call 800-562-8483
or supply connection. for technical assistance
All system functions appear to - UV output has declined Replace lamp module.
be operating, but the product due to normal lamp aging.
does not cure completely.
(If a radiometer - End of light guide has an Clean light guide with
is used to monitor the UV output, accumulation of product soft cloth and isopropyl
the power is low.) or other contaminants. alcohol.
- Light guide is not fully Check light guide
inserted in receptacle. connection.
- Light guide has exceeded its Replace light guide.
usefullife.

8. Documentation
8.1 Wiring Diagram
CHASSIS AND LID
CONNECTIONS
CURTIS
HOUR
METER
RESET
SWITCH
1
34
2
SOL.
FAN
ENCL.
FAN ENCL.
FAN LAMP
FAN
NAIS CYCLE
TIMER
9
10
11
NO 6
2
NC MANUAL
TIMED
560
ohms
1N5253
0.01uf
2N6039 24 VOLTS
MC7809ACT
10k/5%/1/4W
IRF-510
270/5%/1/4W
HONEYWELL
SDP-8405-003
CML
5102H5-5V
MOUNTED ON
FRONT PANEL
9.0 VOLTS
OMRON RELAY
G5SB-14-DC9
44.4mA COIL
TO REAR PANEL
9 PIN SUB-D
CONNECTOR
PIN 4PIN 3
1N4002
TO LED ANODE
TO LED CATHODE
MODE
SELECTOR
SWITCH
LAMP
ANODE
LAMP IGNITER
POWER ENTRY MODULE - CORCOM PS0S0SS60
250V/4A
E
N
L
N
EN/O
OUTPUT PS45-27
POWER
SUPPLY
CHASSIS GROUND
PARTITION
SWITCH
INPUT
LAMP
CATHODE ---
L
PE
LAMP BALLAST
+++ N
L
N/C
PE
L
N+ -
28V SOLENOID
LEDEX
195206-131
PIN 5
FOOT SWITCH
PIN 1
REAR PANEL
9 PIN SUB-D
PIN 9
68 ohms +/-1%;
1W
1N4005
2N6039
100k
ohms
100k
ohms 1N4005
1k
+
-
2.2uf
1M
10k
ohms
10k
470k
2.2k
20k
1N914
NC
NO
91+/-5%; 5W
LM311
DELAY ADJ.
OMRON RELAY
G5SB-14-24DC
16.7 mA
1N914 1
4
87
RED DOT
RED DOT
OUTPUT ST4
ST5
INSTALL JUMPER ACROSS ST4 AND ST5
IF UNIT IS OPERATED BETWEEN 88 AND
136 VAC.
DO NOT INSTALL JUMPER IF UNIT IS
OPERATED ON VOLTAGE BETWEEN 176
AND 264 VAC.
CHASSIS
GROUND

8.2 Pin Connections
8.3 Replacement Parts and Accessories
Loctite Part Number Description
98451 Replacement Lamp Module
97326 Single End Light Guide (5 mm x 1M)
97328 Dual Ended Light Guide (3 mm x 1M)
98543 Single Light Guide (5 mm x 1.5M)
984770 Fan Filter Element, Quantity 5
97210 UV Safety Glasses
986051 Foot Switch Assembly
1406024 Loctite® 7020 UV Spot Radiometer
951639 Dual Wand Adapter Kit
1332501 Lamp Igniter
1332503 Lamp Ballast
1116644 Switching Power Supply

9. Warranty
Henkel expressly warrants that all products referred to in this Instruction Manual for the UV Curing Wand
System, (hereafter called “Product”), shall be free from defects in materials and workmanship. Liability for
Henkel shall be limited, as its option, to replacing those Products which are shown to be defective in either
materials or workmanship or to credit the purchaser the amount of the purchase price thereof (plus freight
and insurance charges paid therefore by the user). The purchaser’s sole and exclusive remedy for breach of
warranty shall be such replacement or credit.
A claim of defect in materials or workmanship in any Product shall be allowed only when it is submitted in
writing within one month after discovery of the defect or after the time the defect should reasonably have
been discovered and in any event, within 2 years after the delivery of the Products to the purchaser. This
warranty does not apply to perishable items, such as fuses and fan filters. The lamp is also warranted to
produce 50% of the minimum rated initial output at 500 hours of operating time, when measured using an
in-calibration Loctite 98720 UV Spot Radiometer, and a new Loctite 5mm x 1000mm lightguide in a fully
straight orientation. No such claim shall be allowed in respect of products which have been neglected or
improperly stored, transported, handled, installed, connected, operated, used or maintained. In the event of
unauthorized modification of the Products including, where products, parts or attachments for use in
connection with the Products are available from Henkel, the use of products, parts or attachments which are
not manufactured by Henkel, no claim shall be allowed.
No Products shall be returned to Henkel for any reason without prior written approval from Henkel.
Products shall be returned freight prepaid, in accordance with instructions from Henkel.
NO WARRANTY IS EXTENDED TO ANY EQUIPMENT WHICH HAS BEEN ALTERED, MISUSED,
NEGLECTED, OR DAMAGED BY ACCIDENT, OR IF THE SYSTEM USED TO DISPENSE ANY
LIQUID MATERIAL OTHER THAN HENKEL PRODUCTS.
EXCEPT FOR THE EXPRESS WARRANTY CONTAINED IN THIS SECTION, HENKEL MAKES NO
WARRANTY OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, WITH RESPECT TO THE
PRODUCTS.
ALL WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, AND
OTHER WARRANTIES OF WHATEVER KIND (INCLUDING AGAINST PATENT OR
TRADEMARK INFRINGEMENT) ARE HEREBY DISCLAIMED BY HENKEL AND WAIVED BY
THE PURCHASER.
THIS SECTION SETS FORTH EXCLUSIVELY ALL OF LIABILITY FOR HENKEL TO THE
PURCHASER IN CONTRACT, IN TORT OR OTHERWISE IN THE EVENT OF DEFECTIVE
PRODUCTS.
WITHOUT LIMITATION OF THE FOREGOING, TO THE FULLEST EXTENT POSSIBLE UNDER
APPLICABLE LAWS, HENKEL EXPRESSLY DISCLAIMS ANY LIABILITY WHATSOEVER FOR
ANY DAMAGES INCURRED DIRECTLY OR INDIRECTLY IN CONNECTION WITH THE SALE
OR USE OF, OR OTHERWISE IN CONNECTION WITH, THE PRODUCTS, INCLUDING, WITHOUT
LIMITATION, LOSS OF PROFITS AND SPECIAL, INDIRECT OR CONSEQUENTIAL DAMAGES,
WHETHER CAUSED BY NEGLIGENCE FROM HENKEL OR OTHERWISE.

Henkel Corporation
One Henkel Way
Rocky Hill, CT 06067-3910
Henkel Canada Corporation
2225 Meadowpine Boulevard
Mississauga, Ontario L5N 7P2
Henkel Capital, S.A. de C.V.
Calzada de la Viga s/n Fracc. Los Laureles
Loc. Tulpetlac, C.P. 55090
Ecatepac de Morelos, Edo. de México
Henkel Corporation
Automotive / Metals HQ
32100 Stephenson Hwy.
Madison Heights, MI 48071
Henkel Ltda.
Rua Karl Huller, 136 – Jd.
Canhema 09941-410
Diadema/SP, Brazil
www.loctite.com
Loctite is a trademark of Henkel Corporation, U.S.A. © Copyright 2009. Henkel Corporation
All rights reserved. Data in this operation manual is subject to change without notice.
Manual P/N: 8901957, Rev B, Date: 03/2010
Table of contents
Other Henkel Industrial Equipment manuals
Popular Industrial Equipment manuals by other brands

Maguire Products
Maguire Products WEIGH SCALE BLENDER WSBMB Series Installation operation & maintenance

Siemens
Siemens Sentron 3VA92 8-0VF30 Series operating instructions

Josoo Electric Appliance
Josoo Electric Appliance ZS-B200X instruction manual

Argal
Argal PDA 75 user manual

Graco
Graco SaniForce SDU Series Instructions - parts

Graco
Graco Therm-O-Flow 20 EasyKey instructions