Hidrex classic ION User manual

Bedienungsanleitung
Iontophorese-Therapiegeräte
HIDREX GS400
HIDREX PSP1000
www.hidrex.com
Gesundheit neu erleben
User Manual
M
a
n
u
a
HIDREX connect
ION

2
2019-06-06, Rev.: 1.0 User Manual
Table Of Content
Table Of Content..........................................................................................2
Read Before Use ..........................................................................................3
Your Safety Is Important.............................................................................3
Who Is The Therapy Suitable For? - Contraindications...................................4
Side Effects .................................................................................................4
Additional Important Safety Considerations ................................................5
Intended Use / Mechanism Of Action ........................................................5
Treatment Fundamentals ............................................................................6
Efficacy Of Treatment And Polarity Reversal..............................................6
Types Of Current.........................................................................................7
Installation / Treatment Setup.....................................................................9
Main System Components and Scope of Delivery......................................9
Therapy Preparations (Setup Control Unit)...............................................12
Setup for Hands and Feet Treatment........................................................13
Hints On Conducting Therapy...................................................................14
Conduct Therapy .......................................................................................15
Primary Settings And Saving Of Profiles...................................................16
Begin Treatment........................................................................................18
Important Advice for Conducting Treatments............................................20
Care And Maintenance ..............................................................................22
Special Remarks .......................................................................................22
Multi-Patient Use .......................................................................................22
Care and Purification.................................................................................22
Disinfection................................................................................................24
Reconditioning...........................................................................................24
Troubleshooting.........................................................................................24
Error Checklist...........................................................................................25
Operational Check.....................................................................................25
Shipping The Device For Repair Or Maintenance ....................................26
Applicable Regulations And Legal Requirements .................................26
Lifespan.....................................................................................................26
Symbol Legend / Manufacturer / Device Identification..............................27
Electromagnetic Compatibility (EMC) .......................................................27
Waste Management And Packaging Of Electronic Devices .......................30
Technical Data............................................................................................31
Distribution / Trading.................................................................................32

User Manual 2019-06-06, Rev.: 1.0
3
Read Before Use
Your HIDREX tap water iontophoresis device is designed to maximize
functionality and usability. System setup is easy, and the device is simple to
operate. However, the following safety instructions and legal regulations must
be observed and strictly adhered to. To ensure the proper, safe and long-term
functioning of your HIDREX iontophoresis device, the following maintenance,
care and disinfection instructions must be performed regularly.
Please read this information carefully!
Your Safety Is Important
iThis treatment system may only be powered by the HIDREX AC power
adapter (which can be identified by a HIDREX nameplate with Serial-
No.). This power supply is especially designed for this device.
iA change of the therapy device is not allowed. Do not open the device.
This therapy device has no controls inside. All service work is only to be
carried out by HIDREX GmbH.
iTo prevent burns during treatment, make sure the supplied plastic nets or
pads always cover the treatment electrodes. Avoid direct contact with the
metallic surface.
iSeveral devices may not be simultaneously used by one patient.
iPrior to treatment, remove any metallic jewelry (wedding bands, etc.)
which would otherwise be in touch with the water source (trays or pads).
Keeping such accessories on would lead to localized minor (electrical)
burns that are secondary to increased current densities.
iAvoid application of more current as can be safely and comfortably
tolerated by the patient. Always treat carefully and observe patient
reaction.
(As a rule, avoid current intensity greater than 0,2mA/cm2of active electrode area.)
iUse only genuine accessories approved by HIDREX for use with your
device. Other unapproved accessories may result in unforeseen and
dangerous behavior of your device.
Please follow the instructions for your accessories, as well.
iThe long-term effects of chronic electrical stimulation have not been
established.
iKeep the device out of the reach of children and do not leave children
unattended during therapy! There is a risk of strangulation due to the
enclosed cables.
iDo not turn the device off during the therapy. In most cases this causes a
safe but uncomfortable electric shock1.
1These shocks are unpleasant, but absolute safe.

4
2019-06-06, Rev.: 1.0 User Manual
Who Is The Therapy Suitable For? - Contraindications
In principle, the therapy is suitable for all persons from the age of 4 years. The
prerequisite for this is that the therapy is carried out under the supervision of
an adult.
From the age of 12, it is up to the legal guardians to decide whether the therapy
can also be carried out without supervision.
Regardless of age, therapy must always be performed under the supervision
of an adult if the patient is unable to understand the instruction manual and act
according to the instructions contained therein (e.g. cognitive impairment).
Users with the following conditions are considered having contraindications for
iontophoresis, so treatment should not be administered unless otherwise
directed by your physician:
iwith a cardiac pacemaker or ICD (implantable cardioverter/ defibrillator)
iin pregnancy
iwith a metal-containing intrauterine device (IUD)
iwith any metal object within the electric current path
·The electrical current path is considered as the shortest point through the body
between the two electrodes. For example, the current path for a person treating
hands would be from one hand in the tray with an electrode, up the arm, across the
chest, down the other arm and hand to the other electrode.
·This means that a user that has a metal screw in their leg or knee can do hand
treatments, but not foot treatments
iwith piercings in the contact area, which cannot be removed
ithat have within the treatment area any large skin defects / wounds (too
large to cover with petroleum jelly), potentially malignant lesions, acute
localized infections, skin eruptions, or swollen, broken, or inflamed areas
ithat have within the treatment area any impaired or absent sensation
(e.g. patients with polyneuropathies)
iApply electrical current through or across the brain, or sinuses (increased
risk of ventricular fibrillation)
iwith suspected or diagnosed heart problems or epilepsy
Side Effects
The following side effects or effects may occur for a short time after the
therapy session on the treated skin areas:
iMild dysesthesia (tingling or burning)
iShort-term skin irritation (reddening) after treatment
iErythema (skin redness, transient vesicles or blisters)
iSkin irritation or burns at the areas of electrode contact have been
reported with the use of electrical stimulators

User Manual 2019-06-06, Rev.: 1.0
5
Additional Important Safety Considerations
iPlace the treatment device on a firm level surface
iMake sure that the treatment device is at room temperature before you
power it up
iThe system should not be operated near shortwave or microwave devices.
A minimum distance of 2 meters should always be kept
iPrior to using AC wall power, check that your outlet meets the system’s
requirements of 100-240 V~ and 50-60 Hz
iUnplug the AC power adapter if a thunderstorm approaches or if you do
not intend to use the treatment system for a longer time
iThis treatment device should only be used indoors. Do not expose the
system to rain or excessive moisture
iPrior to cleaning the system, turn the device off and unplug all connectors.
For cleaning, use a soft cloth moistened with a mild cleaning agent
iDo not use kerosene, thinner, alcohol, wax remover or other solvents
iPrevent kinking of the cable and do not expose the cable to heat or
chemicals. If the cable is damaged, unplug it from the device and have it
checked by an authorized HIDREX repair center
iDo not perform iontophoresis therapy in parallel with aluminum-containing
antiperspirants in the same treatment area.
iSimultaneous connection of the patient to a ME unit for high-frequency
surgery can result in burns under the electrode surfaces of the stimulation
current unit and damage to the stimulation current unit.
Intended Use / Mechanism Of Action
Intended Use:
This tap water iontophoresis device is intended to treat Hyperhidrosis
(excessive sweating) affecting hands, feet, face, back/ neck and
underarms.
Any other use or usage beyond this scope is considered unintended
use and may have dangerous consequences.
Mechanism of Action:
During the HIDREX treatment, a current flow through the body regions that
are being treated. The water in the trays or pads mediates this current flow.
The skin areas in contact with the water will thereby secrete less sweat.
Although treatment success has been validated in numerous medical
studies, there is still no completely satisfactory scientific explanation for the
mechanism of action. Medical researchers believe that the electrical current
irritates the synapses between sweat-inducing nerves and sweat glands to

6
2019-06-06, Rev.: 1.0 User Manual
such an extent that sweat glands can no longer be stimulated. In other
words: the treatment does not affect the sweat glands directly, it only affects
the nervous input to these glands.
This effect explains why the original condition returns relatively quickly
when the treatment is discontinued.
The treatment current can be adjusted according to your individual sensitivity.
There is no risk involved as the current cannot exceed certain limits.
Treatment Fundamentals
The treatment concept of the HIDREX Tap-Water-Iontophoresis (TWI)
comprises two treatment phases:
Phase 1: The initial phase (therapy initiation) is sometimes conducted under
a doctor’s supervision. During this stage, patients learn to
administer treatments. For therapy initiation, at least three weekly
treatments of approximately 15 minutes each should be scheduled
(not more than one treatment per day). Sweat secretion will
normalize after approximately 10 treatments.
Phase 2: Long-term treatment (maintenance therapy) is necessary because
the TWI treatment effect is reversible. Depending on the severity of
the condition, maintenance therapy involves one to three weekly
sessions of approximately 15 minutes each.
Hint: It is up to the doctor to decide whether the initial therapy should be
partially or fully performed at the medical facility or at home with a loaner.
Hint: When using a special applicator, it is essential that you carry out a trial
phase. Instructions for this trial phase are enclosed with your applicator.
Efficacy Of Treatment And Polarity Reversal
In theory, the efficacy of Tap-Water-Iontophoresis (TWI) does not depend on
the direction of current flow. However, it has been clinically documented that
the area being treated with the anode (connector lA, see chapter Main
System Components) will produce slightly better results initially than the area
being treated with the cathode (connector lB).
HIDREX connect ION - Automatically Change of Polarity:
For a balanced therapy result on both treatment sides from the beginning of
the treatment, the polarity reversal function1(PWF) can be activated.
Therefore, the voltage will be gently reduced to 0V at the first half of your
treatment session, changes the polarity and rises back to your set voltage.
1The polarity change function can only be used if the therapy time is at least 10 minutes, as shorter
times would lead to ineffective treatment.

User Manual 2019-06-06, Rev.: 1.0
7
HIDREX classic ION - Manually Change of Polarity:
Despite this slight difference in efficacy in the current flow, it is recommended
to conduct all treatments in the initial phase (Phase 1) without changing
polarity until one side reaches an acceptable dryness level. Once normal or
acceptable dryness has occurred in one side of the treatment area (for
example, right hand), then it is advisable to reverse polarity and continue the
initial phase without changing polarity up to normal or acceptable dryness has
occurred in the other side of the treatment area (for example, left hand).
Once normal or acceptable dryness has occurred in both sides of the
treatment area (for example, both hands), then it is advisable to reverse
polarity for each treatment during the maintenance phase (Phase 2).
Polarity is reversed simply by changing which side of the treatment area is
treated by the anode or cathode.
For example, on the first treatment in the maintenance phase, if the right
hand was treated with the anode and the left hand was treated with the
cathode, then on the next treatment, the right hand would be treated with the
cathode, and the left hand would be treated with the anode. This sequence
would continue for every treatment.
Remark: Failure to do this does not cause any harm or side effects. Long-
term treatment results do not depend on the direction of current flow.
Types Of Current
The HIDREX devices have several options for the type of current used during
treatment of different areas:
ÖDC: Direct Current (Standard)
ÖPC: Pulsed Current
ÖVPC: Variable Pulsed Current
The classic direct current (GS) has the highest efficiency. However, the current
sensation of the GS is clearly distinct and sometimes unpleasant. Therefore,
the use of PS or VPS is highly recommended for the treatment of sensitive
skin areas (armpits, face, etc.) or children. Pulse current is significantly more
sensitive and the "feeling" of the current is significantly reduced.
The term pulse width:
The term "pulse width" used in this context defines a
parameter of the proportion of active current flow
during the therapy in percent (for example, a pulse
width of 60% means that the current flows in 60% of
the time)

8
2019-06-06, Rev.: 1.0 User Manual
Direct Current (DC):
In DC, the current (voltage) is being transmitted 100% of the time.
Öhighly efficient / distinct feeling of current (sometimes painful)
Pulsed Current (PC):
In PC, the current alternates its transmission between ON (the pulse) and
OFF (the pause) for equal lengths of time. The pulse and the pause are the
same duration1.
Ö sensitive (less feeling of current) / typically slightly less effective
Ö not recommended for treatment of feet or strong hyperhidrosis
Variable Pulsed Current (VPC):
VPC is a type of hybrid current between DC and traditional PC and allows for
the ON part of the cycle (the pulse) to run longer than the OFF part of the
cycle (the pause).
This device allows you to change the pulse-width in 10% increments (for
example, ON 60% and OFF 40% or ON 70% and OFF 30%, etc.). This allows
for more effective treatment AND a better comfort level. It is customizable for
your preferences and tolerance levels.
Övery sensitive (low current sense) / customizable effectiveness
Örecommended for all treatment areas
With increasing percentages of the
pulse-width, the length of the pulse is
increased, and the length of the
pause is decreased accordingly.
Due to the shorter pause, more
current is applied than in PC. The
chart to the right shows how
treatment efficiency increases as
pulse-width increases.
Extensive trials showed that the
efficiency of pulsed current could be
increased dramatically. The
effectiveness is not reduced at all compared to direct current when the pulse-
width is 90%.
1for example, 50 microseconds ON and then 50 microseconds OFF, and then the cycle repeats

User Manual 2019-06-06, Rev.: 1.0
9
The following overview shows you which types of current are available to you,
depending on the variant of your control unit:
Installation / Treatment Setup
To prepare for treatment, set up your therapy device according to the following
steps. Please note that the different treatment areas (hands / feet, armpits,
face, neck / back) have different set-ups. These instructions describe the
structure for the hand / foot treatment. For the set-up for other areas, please
refer to the instructions enclosed with the special applicators.
Main System Components and Scope of Delivery
I. Control Unit
jControl unit main ON/OFF switch (main power switch)
kConnector for AC adapter (12V DC)
lJacks for connecting the cable set to the treatment electrodes
II. AC Adapter
Your HIDREX therapy system is equipped with a wide range AC adapter,
which is suitable for different international mains electricity (please refer to
technical data).
This AC adapter gives the user the option to change different primary adapters
(also referred to as mains or wall plugs) as alternatives. The AC adapter of
Control Unit
Type of current
Pulse-width
Hidrex classic
ION
Direct current (DC)
Pulsed current (PC)
100
50
Hidrex connect
ION
Direct current (DC)
Pulsed current (PC)
Variable Pulsed current (VPC)
100
50
60, 70, 80, 90
j
lA
k
lB

10
2019-06-06, Rev.: 1.0 User Manual
ñ
o
m
m
your system is usually connected with the fitting plug for your region. To
change the mains plug, please refer to the following instructions:
Available Mains Plug Adapters (optional accessory):
Type-A (US) i.e. Japan, North and Central America
Type-C (EU) i.e. Europe, South America, parts of Asia
Type-G (UK) i.e. Great Britain, Malaysia, Singapore, Hong Kong
Type-I (AUS) i.e. Australia, China
Remove mains plug adapter by pressing the catch ñat the bottom side
powerful and pull the adapter towards the front.
Insert the desired mains plug adapter at the front of the AC adapter and push
it firmly until the catch ñlocks.
III. Cable Set
To connect the treatment electrodes to the control
unit, use the two connection cables ç.
AC Adapter DC Plug Mains plug adapters
¥
¦
§
ç

User Manual 2019-06-06, Rev.: 1.0
11
ä
IV. Carrying Case and Treatment Case
The case shells are also used for treatment. For better handling, the two sides
of the case äcan easily be separated by pushing the left side away from you
while pulling the right side toward you. The case must be completely open
(lying flat) before attempting to separate the sides.
V. Treatment Electrodes and Electrode Cover
You will receive plastic nets to cover the treatment
electrodes. As standard, the electrodes are made of
stainless steel, but patients with a chrome-nickel allergy
also have the option of having electrodes made of
aluminum.
VI. Optional Accessories
The optional accessories are not included and can be ordered separately if
required.
é
The ergonomic trays can be used for
treatment instead of the carrying case. The
rounded and
extended edge provides a
much more comfortable surface for the
forearms, especially when treating the
hands.
ëThe face mask is needed for the
therapy of the face and consists of 2
components. On the one hand the black
rubber electrode and on the other hand
the sponge material.

12
2019-06-06, Rev.: 1.0 User Manual
ê
The axillary applicators are needed for the
therapy of the armpits.
ìThe neck and back applicators are
needed for the therapy of the neck and
back.
í
The DUO set for simultaneous
treatment of hands and feet
consists of a
pair of adapter cables, a pair of treatment
electrodes and a pair of ergonomic trays.
Therapy Preparations (Setup Control Unit)
Place the device in a well-lit room on a firm, level surface and make sure you
can easily disconnect it from the power supply at any time. Ensure a wall outlet
is within reach and do not carry out the therapy in excessively warm and humid
rooms, e.g. a bathroom immediately after a shower.
i) Insert the AC adapter cable ¦into the jack kon the rear panel.
ii) Plug the AC adapter minto an AC wall outlet.
iii) Plug the color-coded connectors of the cable set çinto the jacks lon
the back panel to their matching colors
Make sure that the plugs are pushed firmly onto the electrodes!
(i)
(ii)
(iii)

User Manual 2019-06-06, Rev.: 1.0
13
Setup for Hands and Feet Treatment
Place one electrode1åinto each side of
the carrying case or ergonomic tray as
pictured at right.
When using the ergonomic trays é, make
sure that the pulled-down side of the tray
(palm rest) is on the side facing you.
Cover each electrode åwith one treatment
cloth æover the entire surface and insert
the two connection cables çaccording to
the colors on the connections of the
treatment electrodes å.
Make sure to forcefully push the connectors all the way onto the necks
of the treatment electrodes.
Fill both case sides äor ergonomic trays é
with warm tap water so that the skin
surfaces to be treated are well submerged.
The tops of the hand or feet should not be
covered with water, unless those areas
need to be treated.
Simultaneously Treatment of hands and feet:
For the simultaneous treatment of hands and feet, you need additional
accessories to work with the included components for hand-foot treatment.
The following items are required: 1) a set of ergonomic trays é2) the
accessory set DUO í(consisting of a pair of treatment electrodes å, plastic
nets æand DUO connection cables).
The carrying case can be used for foot treatment, and the ergonomic trays can
be used for the hands. One plastic net and one electrode should be placed
into each case side and tray, and then the sides and trays should be filled with
hand-warm tap water.
1Please remove the protective foils from the electrodes before using them!

14
2019-06-06, Rev.: 1.0 User Manual
The DUO cables ðare utilized to connect two electrodes each to one of the
grey cables as pictured below:
Now start the treatment in accordance with the instructions in the chapter
“Conduct Therapy”.
Please note that the simultaneous therapy of hands and feet does not always
achieve the optimal result. This is due to the probable different voltage
requirements and recommended pulse-width for each area. (Due to thicker
skin and less sensitivity, the feet are usually treated with higher doses than
the hands, and a higher pulse-width can usually be tolerated).
If the therapy does not have the expected effect on the feet, you should treat
the hands and feet separately.
Hints On Conducting Therapy
- Before starting treatment, thoroughly remove residues of fat-containing
care products from the skin.
- Wash the therapy area (e.g. hands) before each treatment with normal
soap and water to remove the skin's own fat and talc. Even small films of
fat can impair the flow of current and lead to local irritations.
- Do not use cream soap.
- Tap water iontophoresis leads to dry skin with frequent use and especially
at the beginning of therapy. Therefore, treat your skin immediately after
each treatment with a moisturizing care product.
- Use warm water, which is more absorbent and reduces the sensation of
electricity compared to cold water.
ð
ç
ð
ç

User Manual 2019-06-06, Rev.: 1.0
15
•LED Profile
ŒDisplay
ŽButton Dose
+
‘LED connect
•Button Dose -
’LED Pulse
•LED Polarity
Conduct Therapy
Attention: BEFORE you start the treatment and close the electrical current
circuit with your hands, feet or underarms, make sure to FIRST
turn ON the main power switch j. If this sequence is reversed,
there is also a chance that you receive a completely harmless,
but uncomfortable, electric shock!
Do not turn the device off during the therapy. In most cases this
causes a safe but uncomfortable electric shock (electric fence
effect).
v
Reduction of dosage before and during treatment
and choice of current type (DC, PC or VPC) [VPC only for Hidrex connect
ION]
w
Increase of dosage before and during treatment
and selection of profiles before therapy start
x
LED Profile - shows if memory for own settings is active
y
LED Polarity –indicates that the polarity has been changed [only for Hidrex
connect
ION]
z
Connect –indicates that the Bluetooth connection is active. [only for Hidrex
connect
ION]
{
LED Pulse –indicates that pulse current (PC) is active [only for Hidrex classic
ION]

16
2019-06-06, Rev.: 1.0 User Manual
Primary Settings And Saving Of Profiles
Your HIDREX device offers you several treatment options that you must set
before starting treatment. The following paragraph explains how you can carry
out these basic settings and save them if necessary.
1) Turn the device on:
After setting up your HIDREX therapy system as described in the chapter
"Installation/Treatment Setup", switch on the control unit with the main switch
j.
First, the main display uwill briefly show
three moving bars for about 5 seconds (self-
test of the device).
Next, the currently set therapy duration is shown in
minutes on the left and the set dose in volts is
shown flashing on the right.
2) Select desired profile:
If a profile is selected (one of the LED •lights up), the device will automatically
suggest the respective stored treatment parameters (time, dose and pulse
width). In the delivery state, recommended default values are stored for each
profile. However, you can also create your own profiles.
If no profile is selected (none of the LEDs •are lit), the device always starts
in manual mode with the settings 15 minutes and 4 volts.
Select the profile as follows:
Button
Function
Symbol / Display
Press wand hold the button to
activate the profile selection.
Press the button
w
again to
select the profile.
After 8 seconds without pressing
the button
w
you leave the
submenu.
LED
•on =
respective profile
active
LED
•off = manual mode

User Manual 2019-06-06, Rev.: 1.0
17
3) Adjusting treatment parameters:
Use the keys •and Žto set the desired treatment dose1. Each time you
press one of the pushbuttons, you increase or decrease the treatment dose
by one volt.
4) Setting the pulse width:
Select the desired pulse width as follows:
5) Polarity Reversal Function (PRF):
6) Save settings in the profile:
The selected settings are automatically saved in the active profile (the
corresponding LED •lights up) at the start of therapy.
Please note that changes voltage during the treatment are possible, but not
saved in the profile.
Hint: Once the treatment has been started, the ability to change profile or
duration is disabled. The device must be turned off briefly and then
turned back on again to change the profile or duration.
1Select the dose, within the 4-60V adjustment range, according to your doctor's prescription or refer
to the table with the factory-set treatment parameters in this manual.
Tasten
Funktion
Symbol / Anzeige
Press and hold the button
w
to
activate the pulse
-
width selection.
Press the button
w
again to select
the pulse width.
After 8 seconds without pressing the
button
Ž
you leave the submenu.
Please note that with the
HIDREX classic
ION
you can only
switch between 50% and 100%!
Example for 70% pulse
width
:
Tasten
Funktion
Symbol / Anzeige
+
Activate by simultaneously holding
the button
vand w
the polarity
change function. Note that this
function is only available with the
H
IDREX connectION.
Example for active PRF:

18
2019-06-06, Rev.: 1.0 User Manual
Begin Treatment
If you have no other treatment parameters known or given to you, the following
values are recommended as initial values for the therapy.
It is strongly recommended that you start treatment of children with reduced
parameters as indicated in the following table. If necessary, increase the
values slowly and pay attention to the patient's sensation! The treatment
should never be unpleasant or painful.
Attention: In order to avoid skin irritation by excessive dose settings in pulsed
mode (where current flow is practically imperceptible), we
recommend that you assess your individual direct current (pulse
width = 100%) thresholds for hands, feet, or armpits before you
start the first treatment session. To conduct axillary treatments, as
a rule the dose should not be raised above 15 volts, because the
sensitive skin may otherwise get burnt.
1) Starting:
Put your hands or feet in each water-filled tray and place them on the plastic
nets æ.
Attention: Avoid direct contact between skin and the metal
electrode! Do not forget to remove any jewelry!
1Beachten Sie, dass die Verwendung der Sonderapplikatoren jeweils in einer eigenen Anleitung
beschrieben wird. Diese Anleitungen müssen unbedingt beachtet werden.
Option
Adult
Child
HAND
Pulse-width:
Dose:
Time:
90%
20 volts
15 minutes
80%
10 volts
15 minutes
FOOT
Pulse-width:
Dose:
Time:
100 % = DC
30 volts
15 minutes
90%
15
volts
15 minutes
AXE
1
Pulse-width:
Dose:
Time:
50%
8 volts
15 minutes
50%
5 volts
15 minutes

User Manual 2019-06-06, Rev.: 1.0
19
2) Conduct treatment:
The device detects the closing of the circuit
("immersion monitoring") and signals the start
of the treatment. The dose in the display stops
flashing and slowly rises, starting at 4V, to the
previously set value1.
Attention: If you experience any localized pain during treatment, interrupt the
session and cover the painful area with petroleum jelly. Pressing
button •or Žwill stop the automatic increase of the dose.
3) Adjusting treatment parameters:
The treatment dosage may be increased or decreased at any time through-
out the treatment session by pressing buttons •or Ž.
The remaining therapy time cannot be changed once the treatment has
started.
4) Stop the treatment:
The treatment can be paused at any time by removing the hands or feet
easy. The treatment time is stopped (dose flashes again and shows the
selected value).
With stopped therapy, the treatment time and profile can no longer be
changed. The dose can then be changed with the •or Žkeys and the pulse
with by holding key Ž.
The treatment is continued by diving in the hands or feet in the water bath
again. As at the start of treatment, the dose value jumps to 4, stops
flashing and slowly rises again to the desired value.
5) Ending a treatment session:
The remaining treatment time is shown in minutes in the display, with the last
minute in seconds (flashing).
When the treatment time is up, the treatment dose is automatically lowered to
zero. Please keep your hands or feet in the trays until the dose display reads
"End".
Only switch off the device afterwards.
Attention: Do not turn off the device at main switch j, as long as the user is
connected to the device. Otherwise this can cause a safe but
uncomfortable electric shock, even if there are protection features
(electric fence effect).
1See also the note on "Dose Limitation" on the following pages.

20
2019-06-06, Rev.: 1.0 User Manual
Important Advice for Conducting Treatments
For your safety, your HIDREX therapy device is equipped with several
protective circuits.
Immersion monitoring
As long as you haven't closed the circuit through the skin surfaces to be
treated, the device won't deliver a therapy dose. All settings you make when
the therapy circuit is open are default settings. When the therapy circuit is
closed, the dose indicator stops flashing. Only then will the dose rise to the
previously set value and the therapy time expires.
Over-Treatment-Protection and therapy monitoring (OTP1)
The device permanently monitors the therapy circuit and the individual body
resistance of the patient. By this, problems in set-up and with the accessories
can be early recognized.
If a value that is too high is detected right at the start of therapy, the device
prevents the start of therapy.
If the device detects an increased value during
the current treatment, the display will alternate
showing the voltage and "ot". In this case you
can continue the therapy without danger.
If the value exceeds the limit for a longer period
of time, the device first reduces the dose briefly
and then slowly returns to the preset value.
If the resistance remains above the limit, the
device will stop the treatment, for your safety,
lock2the device and the display show "StoP".
Hint: If treatment does not start, "ot" or "StoP" is displayed, check the
connections and ensure that they are pushed completely. Also inspect
the electrodes for calcifications3and cleans as necessary. If you do not
notice any errors in the connections and the electrodes have been
adequately cleaned, the OTP feature is very likely to be active. In this
case, please pause the treatments for at least 2 to 3 weeks and at least
until the excessive sweating is clearly noticeable again. If this does not
help, please contact us.
1OTP = Over-Treatment-Protection
2To be able to start a new treatment a restart of the device is necessary.
3Further information on handling calcifications on the electrodes can be found in the chapter care
and purification.
This manual suits for next models
1
Table of contents
Other Hidrex Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Getinge
Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics
Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care
Pressalit Care R1100 Mounting instruction

Denas MS
Denas MS DENAS-T operating manual

bort medical
bort medical ActiveColor quick guide

AccuVein
AccuVein AV400 user manual