AP 720S Semi-Automated Instrument
Note: The assumed starting point is the instrument state after
completion of a previous run and following routine maintenance and
shutdown. (The alternate state would be the interim between two
successive runs of the instrument). The wash buffer and deionized
water tubes will be on a clean surface following completion of the
drain function and the power will be off.
Prepare Serum Samples
1. Centrifuge all serum samples.
2. Transfer each serum sample into an appropriately labeled clean
 plastic storage tube with lid.
Load Sample / Reagent Rack
3. Load the required number of empty Sample Cups into the
 Sample / Reagent Rack. Start at position #1 and do not leave
 any empty positions.
4. Transfer the correct volume of each prepared sample into the
 sample cups:
  600µl serum for a 36-allergen panel
  490µl serum for a 20-allergen panel
Note: The red Sample Fill Line on the Sample / Reagent Rack can be
used as a guide to check that there is a fill volume of approximately
600µl in the Sample Cups.
5. Load the Antibody Reservoir (BLUE line) onto the Sample /
 Reagent Rack.
6. Remove the Antibody Reagent bottle from the kit and gently
 invert prior to use.
7. Fill the Antibody Reservoir with the required amount of reagent
 as listed in the Instruction Manual.
8. Install the Sample / Reagent Rack on the instrument aligning
 sample position #1 with the red indicator line and lowering the
 rack so that the Center and Positioning Pin locate into the holes
 at the bottom of the rack.
Load Pette Rack
9. Remove the required number of OPTIGEN Pettes from the kit
 box and label with sample identification. (For a given run all
 pettes must be of the same panel type, see Step 21 below).
 Reseal the plastic bag and return the kit to the refrigerator.
10. Attach a Pette Tip to each Pette Body.
11. Place the Pette Rack on the Pette Stand. (Always use the Pette
 Stand when loading or removing Pettes.)
12. Load the Pettes into the Pette Rack. Always start at position
 #1 and do not leave any empty positions. Check that each Pette
 Body is correctly secured in the Pette Rack. Ensure the number
 of Pettes matches the number of samples.
13. Remove the Pette Rack from the Pette Stand.
14. Install the Pette Rack on the instrument by lowering it onto
 the center guide bar with pette cartridge #1 positioned front
 and center so that the rack is properly aligned with the
 positioning pin.
15. Close the door of the instrument.
Load Wash Buffer and Deionized Water,
and Empty Waste
16. Prepare the Wash Buffer according to the instructions in the
 OPTIGEN package insert.
17. Fill the Wash Buffer Bottle with the required amount of Wash
 Buffer as listed in the Instruction Manual.
 Secure the yellow cap on the Wash Buffer Bottle and ensure the
 weighted end of the Wash Buffer tube is located at the bottom
 of the bottle.
18. Ensure there is sufficient (1L) deionized water in the Deionized
 Water Bottle and fill if needed. Secure the white cap on the
 Deionized Water Bottle and ensure the weighted end of the
 Deionized Water tube is located at the bottom of the bottle.
19. Empty the Waste Bottle by pressing the quick release connectors
 to detach the tubes, unscrew the cap, and then empty the waste
 liquid. Reconnect the Waste Bottle by screwing the cap back on
 and tightening. Attach the quick release connectors to the cap,
 matching the color-coded labels.
Doc. No. 0822 - ENG Rev. 00
www.hcdiagnostics.com
Start Processing
20. Turn on the AP 720S. After initializing, the [Main Menu] screen
 will be displayed.
21. Select the panel type (Panel 20 or Panel 36). Press “Next” and
 move to the [Check] screen.
22. Ensure there is sufficient Antibody Reagent in the Reagent
 Reservoir, sufficient Wash Buffer in the Wash Buffer Bottle, and
 sufficient Deionized Water in the Deionized Water Bottle, and
 check that the Waste Bottle is empty.
23. Press “Start” to begin automated processing. (There are 8
 processes in all to complete the Auto-Run).
Load Photoreagent Mixture
24. At the end of Process #7, the [Change Photo Rgt] will appear,
 the AP 720S will pause and an alarm will sound. Press “Alarm Off”.
25. Prepare Photoreagent Mixture by mixing an equal amount of
 Photoreagent AB and CD. Note: Allow Photoreagent AB and CD
 to come to room temperature prior to mixing. Once mixed, the
 Photoreagent should be used immediately for best results.
26. Fill the Photoreagent Reservoir (RED line) with the required
 amount of Photoreagent Mixture as listed in the Instruction
 Manual.
27. Open the door. Remove the Antibody Reservoir and load the
 Photoreagent Reservoir in the Sample / Reagent Rack. Close
 the door.
Continue Processing
28. Press “Continue” on the [Change Photo Rgt] screen.
29. Press “Yes” on the [Check] screen to start Process #8
 (Photoreagent Aspiration).
30. As soon as the first pette is filled with Photoreagent Mixture,
 set a timer for 10 minutes.
31. At the end of Process #8, an alarm will sound. Press “Alarm Off”
 on [Process End] screen.
Unload and Read Pettes
32. Open the door, remove the Pette Rack from the instrument and
 set on the Pette Stand.
33. Insert a black top plug into each Pette.
34. Unload the pettes one by one from the Pette Rack.
35. Remove the Pette Tip from each Pette.
36. Insert a white bottom plug into each Pette.
37. Load the Pettes in order into the Pette Cassette Tray.
38. When the timer reads 10 minutes load the tray in the CLA-1
 Luminometer and read the test results.
Completion of Operation
39. Remove the Sample / Reagent Rack and discard all used sample
 cups.
40. To run a second set of pettes immediately following the first set,
 press “End” on the [Process End] screen to return to the [Main
 Menu] screen.
41. Alternatively, to prepare for shutdown, press “Rinse” on the
 [Process End] screen.
42. Unscrew the caps on the Wash Buffer and Deionized Water
 bottles. Pull out the tube from the Wash Buffer Bottle and place
 it in the Deionized Water Bottle along with the Deionized Water
 tube. Press “Ok” in the [Rinse] screen.
43. On completion of the rinse remove the Wash Buffer and Deionized
 Water tubes from the Deionized Water bottle. Place Wash Buffer
 and Deionized Water tube on a clean surface. Press “Ok” on the
 [Drain] screen.
44. After the drain function is finished the [Main Menu] screen
 appears. Turn off instrument.
45. Rinse Reagent Reservoirs with deionized water.
Doc. No. 0822 - ENG Rev. 00
www.hcdiagnostics.com
QUICK START TM
AP 720S Semi-Automated Instrument (continued)
[Main Menu]
[Panel Type]
[Check]
[Change Photo Rgt]
Elapsed Time 0:0:15
[Check]
Add Photo Rgt?
[Process End]
Elapsed Time 0:0:15
Antibody,
Wash Buffer,
Water.Waste
[Panel 36]
Assay Panel 36
Next
Back
Start
Alarm Off
Yes
Alarm Off
RinseEnd
No
Continue
[Rinse]
Please change tube
to Water Bottle.
[Drain]
Please pulll out tube
from Water Bottle.
Back Ok
End Ok
36 20
Maintenance
07/ 2/23
14:25:21

Other Hitachi Medical Equipment manuals

Hitachi S3ESEL User manual

Hitachi

Hitachi S3ESEL User manual

Hitachi EUP-VV531 User manual

Hitachi

Hitachi EUP-VV531 User manual

Hitachi S12 User manual

Hitachi

Hitachi S12 User manual

Hitachi EUP-C516 User manual

Hitachi

Hitachi EUP-C516 User manual

Hitachi VC34A Probe User manual

Hitachi

Hitachi VC34A Probe User manual

Hitachi S31KP User manual

Hitachi

Hitachi S31KP User manual

Hitachi EUP-CC531S User manual

Hitachi

Hitachi EUP-CC531S User manual

Hitachi L34 User manual

Hitachi

Hitachi L34 User manual

Hitachi S211 User manual

Hitachi

Hitachi S211 User manual

Hitachi EUP-B712 User manual

Hitachi

Hitachi EUP-B712 User manual

Hitachi EUP-C524 User manual

Hitachi

Hitachi EUP-C524 User manual

Hitachi C252 User manual

Hitachi

Hitachi C252 User manual

Hitachi VC35 Probe User manual

Hitachi

Hitachi VC35 Probe User manual

Hitachi CC41R1 User manual

Hitachi

Hitachi CC41R1 User manual

Hitachi EUP-C715 User manual

Hitachi

Hitachi EUP-C715 User manual

Hitachi S121 User manual

Hitachi

Hitachi S121 User manual

Hitachi EUP-LV74 User manual

Hitachi

Hitachi EUP-LV74 User manual

Hitachi UST-5710-7.5 User manual

Hitachi

Hitachi UST-5710-7.5 User manual

Hitachi EUP-L74M User manual

Hitachi

Hitachi EUP-L74M User manual

Hitachi UST-5412 User manual

Hitachi

Hitachi UST-5412 User manual

Hitachi L43K User manual

Hitachi

Hitachi L43K User manual

Hitachi OPTIGEN AP 720STM User manual

Hitachi

Hitachi OPTIGEN AP 720STM User manual

Hitachi EUP-L73S User manual

Hitachi

Hitachi EUP-L73S User manual

Hitachi UST-5417 User manual

Hitachi

Hitachi UST-5417 User manual

Popular Medical Equipment manuals by other brands

Mobility Research Therapy Mouse quick start guide

Mobility Research

Mobility Research Therapy Mouse quick start guide

Prytime Medical STAAR Assemble, Disassemble & Troubleshooting Instructions

Prytime Medical

Prytime Medical STAAR Assemble, Disassemble & Troubleshooting Instructions

AtriCure cryoICE BOX Quick reference guide

AtriCure

AtriCure cryoICE BOX Quick reference guide

mychway SM-7205N user manual

mychway

mychway SM-7205N user manual

Vermeiren D100 instruction manual

Vermeiren

Vermeiren D100 instruction manual

Hill Laboratories HF54 Plus Instructions for use

Hill Laboratories

Hill Laboratories HF54 Plus Instructions for use

bort medical Generation 114 490 G Instructions for use

bort medical

bort medical Generation 114 490 G Instructions for use

Neurotech MediStim XP instruction manual

Neurotech

Neurotech MediStim XP instruction manual

Bard QuickLink 70310QCA1 quick start guide

Bard

Bard QuickLink 70310QCA1 quick start guide

Dr. Mach LED 6MC user manual

Dr. Mach

Dr. Mach LED 6MC user manual

Beijing Choice Electronic Technology MD300C22 Operator's manual

Beijing Choice Electronic Technology

Beijing Choice Electronic Technology MD300C22 Operator's manual

Topcon ALADDIN HW3.0 user manual

Topcon

Topcon ALADDIN HW3.0 user manual

Stryker Atlas 660Z Maintenance manual

Stryker

Stryker Atlas 660Z Maintenance manual

Rimba 880 manual

Rimba

Rimba 880 manual

SurgiVet V1030 Operation guide

SurgiVet

SurgiVet V1030 Operation guide

Medisoft Ergocard CPX Hardware user manual

Medisoft

Medisoft Ergocard CPX Hardware user manual

Dorel Juvenile Group Safety 1st Advanced Solutions IH198 user guide

Dorel Juvenile Group

Dorel Juvenile Group Safety 1st Advanced Solutions IH198 user guide

Cardinal Health Kangaroo Connect Enteral Feeding Pump Directions for use

Cardinal Health

Cardinal Health Kangaroo Connect Enteral Feeding Pump Directions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.