RECALLING VALUES FROM MEMORY
This monitor can be used by 2 individuals. Each user can store up to 90
measurements.
1. Press and release either the M1 or M2 button. The unit will first display the
average of the last 3 stored measurements.
2. Every new press of the M1 or M2 button will recall a previous reading. The
latest reading will be recalled first.
3. To stop recalling readings from memory, press the POWER button.
CLEARING VALUES FROM MEMORY
From the power display o, press and hold down either the M1 or M2 button
until the display shows CLr. This indicates that all measurements have been
erased.
CARE, MAINTENANCE, AND CLEANING
• Clean the blood pressure monitor body and cu carefully with a slightly damp,
soft cloth. Do not press. Do not wash cu or use chemical cleaner on it. Never
use thinner, alcohol, or petrol (gasoline) as cleaner.
• Leaky batteries can damage the unit. Remove the batteries when the unit will
not be used for a long time.
• Follow local ordinances and recycling instructions regarding disposal or
recycling of the device and device components, including batteries.
• If the unit is stored near freezing, allow it to acclimate to room temperature
before use.
• This blood pressure monitor is not field-serviceable. You should not use any
tool to open the device nor should you attempt to adjust anything inside the
device. If you have any problems with this device, please contact HoMedics
Consumer Relations (contact information can be found on the Warranty page).
• Do not immerse the unit in water as this will result in damage to the unit.
• Do not subject the monitor or cu to extreme temperatures, humidity,
moisture, or direct sunlight. Protect from dust.
• Do not fold the cu and tubing tightly.
• Do not disassemble the monitor or cu. If in need of repair, refer to the
Warranty section of this manual.
• Do not subject the monitor to extreme shocks (do not drop on floor).
• Do not inflate the cu unless wrapped around arm.
• Do not wrap the cu around body parts other than your arm.
• Do not drop or insert any object into any opening or hose.
• To avoid accidental strangulation, keep this product away from children.
Do not drape tube around neck.
• This monitor may not meet its performance specifications if stored outside
these temperature and humidity ranges:
Storage/Transportation Environment Operating Environment
Temperature: -13°F ~ 158°F (-25°C ~ 70°C)
Humidity: less than 93% RH
Temperature: 41°F ~ 104°F (5°C ~ 40°C)
Humidity: 15% ~ 93% RH
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below,
and should only be used in such environments:
Immunity test IEC 8080 1
test level
Compliance
level
Electromagnetic environment- guidance
Conducted
RF
IEC 61000-
4-6
Radiated
RF
IEC 61000-
4-3
3 Vrms
150 kHz to
80MHz
3 V/m
80 MHz to
2.5 GHz
3 Vrms
10 V/m
Portable and mobile RF communications equipment should be
used no closer to any part of the device, including cables,
than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2
d = 1.2 8O MHz to 800 MHz
d = 2.3 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site surveya, should be less than the
compliance level in each frequency rangeb.
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from
structures, objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the device.
b Over the frequency range 150 kHz to 8O MHz, field strength should be less than 10 V/m.
WARRANTY
LIMITED 5-YEAR WARRANTY
HoMedics sells its products with the intent that they are free of defects in
manufacture and workmanship for a period of 5 years from the date of original
purchase, except as noted below. HoMedics warrants that its products will be free
of defects in material and workmanship under normal use and service. This blood
pressure monitor meets the simulated measurement cycles test requirement per
EN1060-3, part 8.10. This warranty extends only to consumers and does not extend
to Retailers.
To obtain warranty service on your HoMedics product, contact a Consumer Relations
representative by telephone at 1-800-466-3342 for assistance. Please make sure to
have the model number of the product available.
HoMedics does not authorize anyone, including but not limited to Retailers,
the subsequent consumer purchaser of the product from a Retailer, or remote
purchasers, to obligate HoMedics in any way beyond the terms set forth herein.
This warranty does not cover damage caused by misuse or abuse; accident; the
attachment of any unauthorized accessory; alteration to the product; improper
installation; unauthorized repairs or modifications; improper use of electrical/power
supply; loss of power; dropped product; malfunction or damage of an operating part
from failure to provide manufacturers recommended maintenance; transportation
damage; theft; neglect; vandalism; or environmental conditions; loss of use during the
period the product is at a repair facility or otherwise awaiting parts or repair; or any
other conditions whatsoever that are beyond the control of HoMedics.
This warranty is eective only if the product is purchased and operated in the
country in which the product is purchased. A product that requires modifications or
adoption to enable it to operate in any other country than the country for which it was
designed, manufactured, approved, and/or authorized, or repair of products damaged
by these modifications is not covered under this warranty.
THE WARRANTY PROVIDED HEREIN SHALL BE THE SOLE AND EXCLUSIVE
WARRANTY. THERE SHALL BE NO OTHER WARRANTIES EXPRESS OR IMPLIED
INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS OR ANY
OTHER OBLIGATION ON THE PART OF THE COMPANY WITH RESPECT TO PRODUCTS
COVERED BY THIS WARRANTY. HOMEDICS SHALL HAVE NO LIABILITY FOR ANY
INCIDENTAL, CONSEQUENTIAL, OR SPECIAL DAMAGES. IN NO EVENT SHALL THIS
WARRANTY REQUIRE MORE THAN THE REPAIR OR REPLACEMENT OF ANY PART OR
PARTS THAT ARE FOUND TO BE DEFECTIVE WITHIN THE EFFECTIVE PERIOD OF THE
WARRANTY. NO REFUNDS WILL BE GIVEN. IF REPLACEMENT PARTS FOR DEFECTIVE
MATERIALS ARE NOT AVAILABLE, HOMEDICS RESERVES THE RIGHT TO MAKE
PRODUCT SUBSTITUTIONS IN LIEU OF REPAIR OR REPLACEMENT.
This warranty does not extend to the purchase of opened, used, repaired,
repackaged, and/or resealed products, including but not limited to sale of such
products on internet auction sites and/or sales of such products by surplus or bulk
resellers. Any and all warranties or guarantees shall immediately cease and terminate
as to any products or parts thereof that are repaired, replaced, altered, or modified,
without the prior express and written consent of HoMedics.
This warranty provides you with specific legal rights. You may have additional rights
that may vary from state to state. Because of individual state regulations, some of the
above limitations and exclusions may not apply to you.
©2018 HoMedics, LLC. All rights reserved. HoMedics is a registered trademark of HoMedics, LLC.
Distributed by Homedics, LLC, 3000 N Pontiac Trail, Commerce Township, MI 48390
IB-BPAO300
P/N: 72-B68MN-HM11 VER: 001
Printed in China
FEDERAL COMMUNICATIONS COMMISSION COMPLIANCE
STATEMENT
Changes or modifications to this equipment not expressly approved by the
manufacturer could void the user’s authority to operate the equipment. This
equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed
to provide reasonable protection against harmful interference in a residential
installation. This equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may
cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment o and
on, the user is encouraged to try to correct the interference by one or more of
the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit dierent from that to which
the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
ELECTROMAGNETIC COMPATIBILITY (EMC)
Guidance and manufacturer’s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below,
and should only be used in such environments:
Phenomenon Professional
healthcare facility
environment a)
Home healthcare
environment a)
Conducted and
radiated RF
emissions
a) CISPR 11
Group 1 Class B
Harmonic distortion Not applicable
Voltage fluctuations
and flickering Not applicable
a) The equipment is suitable for use in Home Health Environments and Professional Health
Care Environments limited to patient rooms and respiratory treatment facilities in hospital
or clinics. The more restrictive acceptance limits of Group 1 Class B (CISPR 11) have been
considered and applied. The equipment is suitable for use in the mentioned environments
when directly connected to the Public Mains Network.
b) The test is not applicable in this environment unless the ME EQUIPMENT and ME SYSTEM
used will be connected to the PUBLIC MAINS NETWORK and the power input is otherwise
within the scope of the Basic EMC standard.
Guidance and manufacturer’s declaration – Electromagnetic immunity – Enclosure port
Phenomenon Basic EMC standard
or test method Immunity test levels
Professional
healthcare facility
environment
Home healthcare
environment
Electrostatic
discharge IEC 61000-4-2 + 8kV contact
+ 2kV, + 4kV +, + 8kV, + 15kV air
Radiated RF EM
fields IEC 61000-4-3 a) 10 V/m b) 80MHz
- 2.7 GHz 80% AM
at 1kHz
Proximity fields
from RF wireless
communicaions
equipment
IEC 61000-4-3 COMPLIANT
NOTE: Further information about distances to
be maintained between portable and mobile
RF communications equipment (transmitters)
and the AVITA BP68 can requested from
AVITA using the contact information provided
in this manual. However, it is advisable
to keep the electromechanical aerosol
equipment at an adequate distance of, at
least, 0.5 m from mobile phones or other RF
communications transmitters to minimise
possible interference.
Rated power
frequency magnetic
fields
IEC 61000-4-8 30 A/m c)
50 Hz or 60 Hz
a) The equipment is suitable for use in Home Health Environments and Professional Health
Care Environments limited to patient rooms and respiratory treatment facilities in hospital or
clinics. The more restrictive IMMUNITY acceptance limits have been considered and applied.
b) Before modulation is applied.
c) This test level assumes a minimum distance of at least 15 cm between the ME EQUIPMENT
or ME SYSTEM and sources of power frequency magnetic fields.
Recommended separation distances between portable and mobile RF
communication equipment and the device.
The device is intended for use in an electromagnetic environment where radiated
RF disturbances are under control. User can help prevent electromagnetic
interference by keeping the device at a minimum distance from portable and
mobile RF communications equipment (transmitters). Below table details the
maximum output power of transmitter :
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d = 1.2
80 MHz to 800 MHz
d = 1.2
800 MHz to 2.5 GHz
d = 2.3
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation ls aected by absorption and reflection from
structures, objects, and people.
TROUBLESHOOTING
If any abnormality arises during use, please check the following points.
Problem Probable Cause Recommended Action
Nothing appears in the
display even when the
power is turned on.
Batteries are drained. Replace all batteries with
new ones.
Batteries are not correctly
aligned with terminals. Reinsert batteries in the
correct position.
Low battery symbol
appears. Batteries are drained. Replace all batteries with
new ones.
In colder temperatures,
batteries have weaker
electrical charges.
Warm up the batteries, or use
the device in a warmer setting.
Device operation time is
inconsistent. Dierent battery brands
have dierent life spans. Use alkaline batteries and
replace all batteries at the
same time with the same
brand of batteries.
No reading after
measurement. Batteries are drained. Replace all batteries with
new ones.
Suspicious blood pressure
results. Cu positioned improperly. Adjust patient and arm cu to
measure.
Blood pressure naturally
varies throughout the day. Rest a while, relax, and
measure again.
Suspicious heart rate
results. Body movement during
device use. Refrain from moving during
measurement.
Measurement shortly after
exercise or exposure to the
outdoors.
Do not take measurements
after exercise or coming back
from the outdoors.
Power switches o
automatically. System design. Push the POWER button again,
and then begin measure.
During measuring, air
reinflates. It could be a normal action.
If the user’s blood pressure
is higher than the initial
pressure value, the device
automatically pumps to a
higher pressure by 40mmHg
each time.
Relax, and try to take a
measurement again.
The arm cu is not fastened
properly. Check that the arm cu is
fastened properly and retake
the measurement.
Note: If the unit still does not work, contact HoMedics Consumer Relations.
Under no circumstance should you disassemble or attempt to repair the unit by
yourself. Contact information for HoMedics Consumer Relations can be found on
the Warranty page.
SPECIFICATIONS
• Measuring range: Blood Pressure: 30-280mmHg
Pulse Rate: 40-199 beats/min
• Calibration accuracy: Blood Pressure: 30-280mmHg
Pulse Rate: 40-199 beats/min
• Operating environment: 10°C~40°C (50°F~104°F) with relative humidity up to
85% (non-condensing)
• Storage/Transportation environment: -20°C~+50°C (-4°F~+122°F) with
relative humidity up to 85% (non-condensing)
• Power source: 4 x 1.5 V AAA batteries
• Weight: approx. 226g (without batteries)
• Dimensions: approx. 95 x 130 x 45 (W x H x D)
• Cu circumference (M Size): approx. 9” to 17” (23 cm to 43 cm)
Note: These specifications are subject to change without notice.
ERROR CODES
Err Code Meaning Corrective Action
Err 0 Could not detect pulse. Too much
body movement. Reduce movement and retry measurement.
Err 1 Leakage in cu pressure/inflation
too low. The arm cu is not fastened properly.
Reapply the cu, and take a measurement
again.
Err 2 Pressure fault, could not detect
pulse. Too much body movement. Rest a while, relax, and retry measurement.
Err 3 Could not detect pulse during
deflation. Too much body
movement.
The arm cu is not fastened properly.
Reapply the cu, and take a measurement
again.
Err Memory error. Take o batteries to reboot the device, then
take another measurement.
Low batteries. Replace all batteries with new ones.
POTENTIAL FOR ELECTROMAGNETIC INTERFERENCE
To avoid inaccurate results caused by electromagnetic interference between
electrical and electronic equipment, do not use the device near a cell phone
or microwave oven. For most wireless communication
devices, it is recommended to maintain a distance of 10.8
feet (3.3m) in order to avoid electromagnetic interference.
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions: (1)
this device may not cause harmful interference, and
(2) this device must accept any interference received,
including interference that may cause undesired operation.