Fault
Code Description Condition Recovery
m9 Simulated Maintenance Fault User has generated a simulated fault. Reset simulated fault.
m10 Over range.
A large concentration has been detected.
The Midas®requires an independent confirmation that
the gas hazard is gone.
Supply known clean air to the Midas®and clear this fault.
m11 User calibration expired The user specified calibration interval has elapsed. Perform zero and span calibrations. Increase span calibration period.
m12 Cartridge expires soon. Cartridge is old and will expire soon. Replace the cartridge with a new cartridge.
m13 Flow low. Midas®is no longer able to regulate flow. Check filters and pump.
m14 Interferent present. An interferent is degrading the ability of the Midas®
to detect gas. Check application.
m15 Temperature near limit. Temperature within 2° Celsius of limit. Check installation environment.
m16 Baseline fault. Sensor baseline has drifted.
Check for background gas concentration, temperature or humidity
fluctuations.
Perform zero calibration.
Replace cartridge.
m17 Inhibit timeout. Transmitter has been in inhibit mode too long. Resume monitoring or increase timeout value.
F39 Simulated Fault User has generated a simulated fault. Reset simulated fault.
F40 Sensor overdosed. Sensor has been exposed to high gas concentrations
for long periods. Replace cartridge.
F41 Baseline fault. Sensor baseline has drifted.
Check for background gas concentration, temperature or humidity
fluctuations.
Perform zero calibration.
Replace cartridge
F42 Calibration expired. Too long since last calibration. Replace or calibrate the cartridge.
F43 Cartridge expired. Cartridge is too old. Replace cartridge.
F44 Cell failure. Cartridge has failed Reflex™ check. Replace cartridge.
F45 Stabilization timeout. Cartridge has failed to stabilize.
If temperature or humidity shocks exist, precondition the cartridge.
Check for background gas concentration.
Replace cartridge.
F46 Cartridge analog failure. Electronic Failure
Gas concentration greater than full scale. Replace cartridge.
F47 Cartridge memory invalid. Checksum error. Replace cartridge.
F48 Cartridge absent. No communications. Reseat cartridge.
Replace cartridge.
F49 Cartridge wrong type. Cartridge type found to be incorrect after boot-up. Replace cartridge or press ‘✓’ accept if correct.
F80 Temperature limits exceeded. Temperature is outside limits Check installation environment.
F81 Flow fail. Flow < 70% of nominal for 15 seconds.
Check filters.
Check for kinked tubing.
Replace pump.
F82 Excessive electrical noise. Internal electronics repeatedly noisy.
Check grounding of Midas®chassis.
Check termination of cable shields.
Relo
cate the Midas®further from noise sources.
Add ferrite inductors to cables.
F83 Pyrolyzer fail. Pyrolyzer fails to heat. Check electrical connection to pyrolyzer.
Replace pyrolyzer.
F84 Misc. transmitter fault. Transmitter is defective. Service or replace Midas®.
F86 Coprocessor fail Coprocessor is damaged. Contact Honeywell Analytics Service.
F87 Pyrolyzer temp limits
exceeded Pyrolyzer temperature exceeded. Clean pyrolyzer cooling vents.
Contact Honeywell Analytics Service.
6. Diagnostics