Hopkins Newborn & Mom Pulse OX 594020 Troubleshooting guide


User Manual for the Hopkins®Newborn & Mom Pulse Ox
Instructions to User
Dear Hopkins Customer,
Thank you very much for purchasing our product. Please
read the manual very carefully before using this device.
Failure to follow these instructions can cause abnormal
readings or damage to the oximeter.
No part of this manual may be photocopied, reproduced or
translated into another language without the prior written
consent. We reserve the right to improve and amend it at
any time without prior notice.
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions: (1)
this device may not cause harmful interference, and (2)
this device must accept any interference received,
including interference that may cause undesired operation.
Version of the Manual: Ver 1.2
Revised date: September, 2014
Hopkins Medical Products
All rights reserved.
I

User Manual for the Hopkins®Newborn & Mom Pulse Ox
Warnings!
•Check the device, sensors and probes to make
sure that there is no visible damage that may
affect user’s safety or measurement performance
with regard to sensors and clips. It is
recommended that the device should be inspected
once a week. When there is obvious damage, stop
using the device.
•The device should be carefully monitored if in
use when ambient temperature exceeds 98.6°F,
37ºC. Injury may occur due to overheating of the
sensor
•Discomfort may occur if the oximeter is used on
the same finger for an extended period of time,
especially for poor microcirculation patients.
•It is recommended that the oximeter should not
be applied to the same location for longer than 2
hours. If any abnormal condition is found, please
change the position of oximeter.
•Avoid placing the device on the same limb which
is wrapped with a cuff for blood pressure
measurement or during venous infusion.
•DO NOT clip this device on edema or tender
tissue.
II

User Manual for the Hopkins®Newborn & Mom Pulse Ox
•The light (the infrared light is invisible) emitted
by the device is harmful to the eyes, so service
technicians and patients should not stare at the
light.
•Please follow local regulations when disposing of
the oximeter, its accessories and batteries.
Warnings forAdult Fingertip Operation
•The finger should be inserted into the sensor in the
correct position.
•The patient’s finger should remain still during
measurement.
•Do not put wet finger directly into sensor.
•Do not let anything block the emitting light from
device.
•Vigorous exercise and electrosurgical device
interference may affect the measuring accuracy.
•Using enamel or other makeup on the nail may
affect the accuracy of measurement; long
fingernails may cause failure of measurement or
inaccurate measurement results.
•Keep the oximeter away from dust, vibration,
corrosive substances, explosive materials, high
temperature and moisture.
III

User Manual for the Hopkins®Newborn & Mom Pulse Ox
Important!
•Existence of high intensive light sources, such as
fluorescence light, ruby lamp, infrared heater or
strong sunshine, etc. may cause inaccurate
measurement. Please put an opaque cover on the
sensor or change the patient’s location.
•If the first reading shows unstable waveform, the
measurement may be false. In order to obtain a
more accurate measurement, simply wait for the
waveform to stabilize or remove the sensor and
take a second reading.
IV

User Manual for the Hopkins®Newborn & Mom Pulse Ox
Content
1 Overview ....................................................................... 1
1.1 Appearance............................................................. 1
1.2 Name and Model.................................................... 2
1.3 Intended Use........................................................... 2
1.4 Structure and Conformation................................ 2
1.5 Features ...............................................................2-3
2 Battery Installation...................................................... 3
3 Operation...................................................................... 4
3.1 Direct measurement............................................4-6
3.2 Alarm and alarm silence....................................... 7
3.3 Menu Screen........................................................8-9
3.4 Hopkins®Y-Probe Sensor..............................10-11
3.5 Proper application of Hopkins®Y-probe .......... 10
3.6 Safe Operation in a Neonatal Incubator........... 17
4. Technical Specifications.......................................17-19
5. Accessories ................................................................. 20
6. Maintenance & Replacement Parts ...................20-22
V

User Manual for the Hopkins®Newborn & Mom Pulse Ox
6.1 Oximeter Maintenance...................................20-21
6.2 Cleaning and Disinfecting Instruction............... 22
6.3 Replacement Parts..........................................22-23
7. Troubleshooting......................................................... 24
Appendix........................................................................ 25
AKey of Symbols....................................................... 25
B SpO2Common Knowledge...............................25-30
C Electromagnetic Interference Data.................31-38
VI

User Manual for the Hopkins®Newborn & Mom Pulse Ox
1 Overview
1.1 Appearance
Figure 1 Front/Rear View
1

User Manual for the Hopkins®Newborn & Mom Pulse Ox
1.2 Name and Model
Name: Hopkins®Newborn and Mom Pulse OX Kit
Model: #716266
1.3 Intended Use
This Fingertip Oximeter is intended for measuring the
pulse rate and functional oxygen saturation (SpO2)
through a patient’s finger. It is applicable for
spot-checking SpO2and pulse rate of adult and pediatric
patients in homes and clinics.
1.4 Structure and Conformation
It consists of main unit and photoelectric sensor, and an
additional sensor port.
1.5 Features
• External Hopkins®Newborn Y-probe sensor
#592018
• Large true color OLED display of SpO2, PR Pulse
Bar, PI & Plethysmogram
• Innovative 4 directions display
• Automatic power on/off
• Audible & visible alarm function
2

User Manual for the Hopkins®Newborn & Mom Pulse Ox
• Shift parameter display between PR and PI
• 2AAA alkaline batteries with low power consumption
• Low battery voltage indication
2 Battery Installation
Figure 2 Battery Installation
1. Refer to Figure 2; insert two AAA size batteries into
the battery compartment properly.
2. Replace the cover.
Please make sure that the batteries are correctly
installed, incorrect installation may cause the
device not to work. Remove batteries from unit
for storage or prolonged periods of
non-use to avoid battery corrosion.
3

User Manual for the Hopkins®Newborn & Mom Pulse Ox
Operation
3.1 Direct Measurement
1. Open the clip as shown in Figure 3.
Figure 3 Put finger into the Oximeter
2. Put finger into the rubber cushions of the clip (make
sure the finger is in the correct position with OLED
display facing up), and then clip the finger.
3. The device will power on automatically in 2 seconds,
and start to display software version number.
4. The user can read the values and view the waveform
on the display screen as seen in figures 4A-4F
below.
Shift parameter display between PR and PI during
measurement:
Press the Display Key, the screen will tilt 90º. At this time,
PR will shift to PI automatically (as shown in 4B). When
shown as 4B, the display will shift to 4C automatically
after 20 seconds without operation. Press the key again
and the screen will change to another position (as shown
4

User Manual for the Hopkins®Newborn & Mom Pulse Ox
in 4D). Shifting parameter display between PR and PI is
under the condition of the screen tilting to a side, as
shown in 4B and 4E. After 20 seconds without operation,
the PI will automatically shift to PR, as shown in 4C and
4F.
Figure4A Figure4B
Figure4C Figure4D
Figure4E Figure4F
5

User Manual for the Hopkins®Newborn & Mom Pulse Ox
Screen Description:
“%SpO2”: SpO2symbol; “99”: SpO2 value;
“PR”: Pulse rate icon; “65”: Pulse rate value;
“”: Pulse beat symbol;
“”: Pulse intensity histogram.
“PI%”: Perfusion index icon; “1.4”: Perfusion
index value;
6

User Manual for the Hopkins®Newborn & Mom Pulse Ox
3.2 Alarm
When measuring, if SpO2value or pulse rate value
exceeds the preset alarm limit, the device will alarm
automatically and the value which exceeds limit on the
screen will flash. The detailed information refers to
chapter 4.
To silence the alarm or re-set the pulse ox
1. Press Display Key to mute. If this alarm event
continues, the pulse ox will resume alarm sound
automatically 2 minutes later if the condition persists.
2. Remove the finger from the pulse ox or SpO2
probe.
3. The alarm will automatically silence when SpO2
and pulse rate return to normal ranges
7

User Manual for the Hopkins®Newborn & Mom Pulse Ox
3.3 Menu Screen
Figure 5 Menu Screen
Press and hold the display key to enter the menu screen.
Menu screen description
:
1. “SpO2alm Lo
”
:SpO2alarm Lower limit.The
user can modify the value of 85~99, the step is
“1”, the default is 90.
2. “PR alm Hi”:Pulse Rate alarm Upper limit. The
user can modify the value of 100~240, the step is
“5”, the default is 120.
3. “PR alm Lo”:Pulse Rate alarm Lower limit. The
8

User Manual for the Hopkins®Newborn & Mom Pulse Ox
user can modify the value of 30~60, the step is
“1”, the default is 50.
4. “Pulse beep”:Pulse beep button. When SpO2
value (90~99) changes, the pitch tone changes
accordingly.
5. “Save, exit menu”:Press and hold this item to
store and exit from the setup menu then enter the
display screen.
6. “Restore default”:Restore default setting. Refer
to Figure 5 for each default value.
On setup menu screen:
1. Press the Display Key to choose the setting item
2. Press and hold the Display Key to activate the
setting item, then press it to modify the setting
parameter
3. Next, press and hold the Display Key to confirm
the modification and exit from this setting item.
4. Move the setting item to “Save, exit menu”, and
press and hold the Display Key to store the
modification and exit from the setup menu.
9

User Manual for the Hopkins®Newborn & Mom Pulse Ox
3.4 Hopkins®Y-Probe Sensor
1. Connect the Hopkins®Y-Probe Sensor to SpO2
sensor connector as shown below. Make sure the side
with “Arrow” faces upward.
Figure 6 Probe Connection
Important! Note: When the Hopkins®Y-Probe Sensor
is connected properly, the built-in finger clip sensor
will be disabled. The measurement is detected from the
external SpO2probe.
2. The finger should be inserted into the SpO2probe in
the correct position with the light emitting diode
“LED” on top of the measuring location.
3. The oximeter will power on automatically 2 seconds
10

User Manual for the Hopkins®Newborn & Mom Pulse Ox
later, displaying the software version number.
4. Other operation is similar to chapter 3.1 direct
measurement.
3.5 ProperApplication of Hopkins Y-Probe
Your Newborn and Mom Pulse OX Kit, Model:
#716266 comes with a package of 10 hook-and-loop
non-stick sensor wraps which are provided to secure
the Hopkins®Adjustable Y-Probe Sensor for accurate
monitoring of pulse and blood oxygen levels in the
patient. The recommended application site for infant
and pediatric patients is on the finger, foot, or hand.
The recommended application for adult patients is on
the finger.
Using the Hopkins®Y-Probe wrap (#592012)
A. FingerApplication Adult or Pediatric Use:
1. Choose a well perfused (large blood supply) site
that fits in the sensor. The index finger is
recommended for adult patients.
2. Remove nail polish or artificial fingernails.
3. Place the sensor to the patient with the light
source (LED) to the fingernail. The photo-detector
(PD) sensor should be placed on the opposite side so
both sensors are lined up and facing each other.
11

User Manual for the Hopkins®Newborn & Mom Pulse Ox
4. Secure the sensor with the Hopkins® Adjustable
Y-Probe Sensor Latex Free Belt and tighten gently.
5. Apply a Hopkins® Y-Probe wrap on top to
prevent ambient light interference. A secure wrap with
the probe sensor centered in the middle of it will
minimize any effects from ambient light sources.
Other wraps can be used to secure lead wires to
minimize the movement of the sensor itself.
IMPORTANT! LED Sensors are light sensitive. Use
Hopkins® Y-Probe wrap on top of the belt to better
secure the probe.
B. Infant Hand or Foot:
1. Attach the sensor to a well perfused fleshy
portion of the infant’s hand or foot.
2. Place the sensor to the patient with the light
12

User Manual for the Hopkins®Newborn & Mom Pulse Ox
source (LED) on the top of the hand or foot.The
photo-detector (PD) sensor should be placed on the
opposite side so both sensors are lined up and facing
each other.
4. Secure the sensor with the Hopkins® Adjustable
Y-Probe Sensor Latex Free Belt and tighten gently.
5. Apply a Hopkins® Y-Probe wrap on top to
prevent ambient light interference. A secure wrap with
the probe sensor centered in the middle of it will
minimize any effects from ambient light sources.
Other wraps can be used to secure lead wires to
minimize the movement of the sensor itself.
IMPORTANT! LED Sensors are light sensitive. Use
Hopkins® Y-Probe wrap on top of the Adjustable
Y-Probe Sensor with belt to better secure the probe.
Product Care
13
Table of contents
Other Hopkins Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Weyer
Weyer Variotherm REA-KCE Instructions for use

SIMS GRASEBY
SIMS GRASEBY MR10 Technical & service manual

Covidien
Covidien Nellcor D-YSE Instructions for use

BBC Bircher
BBC Bircher CareMat A01T L869 Translation of the original instructions

Ambu
Ambu aBox 2 Instructions for use

Gima
Gima GIMANORD Operative manual

Braun
Braun Aesculap Instructions for use/Technical description

Limacorporate
Limacorporate Physica knee system Instructions for use

Perdelle
Perdelle Swan Assembly instructions

ORTHOSERVICE RO+TEN
ORTHOSERVICE RO+TEN Go up manual

Thuasne
Thuasne ActionReliever Fittings guide

Lafayette Instrument
Lafayette Instrument 83015A user manual