2 System Overview
11
QSetting menu
The stage under the maintenance mode includes the following menu.
For detailed step to enter each menu and parameter, refer to the description in Table 1.
zSystem setting (System Set)
zMeasurement condition setting (Meas. Set)
zOutput setting (Output Set)
Table 1 Basic setting parameters
Display Description Selection Default Reference
System Set Menu System setting Page 24
Mean.Stop Set Setting for when measurement is stopped using external input Page 27
UV Off Setting of UV lamp operation when mea-
surement is stopped Yes , No Yes Page 2 7
Delay
Setting of the amount of time that will
elapse between input and the point that
the UV lamp is turned OFF when the
"Setting of UV lamp operation" has been
set to "Yes"
1 to 120 (min) 5 Page 27
Warm Up Setting of the amount of warm-up time
when external input is reset 1 to 60 (min) 30 Page 27
No Sample Set Setting for no sample detection time Page 25
UV Off Setting of UV lamp operation during no
sample detection time Yes , No Yes Page 2 5
Delay
Setting of the amount of time that will
elapse between the point of no sample
detection and the UV lamp is turned OFF
when the "Setting of UV lamp operation"
has been set to "Yes"
1 to 120 (min) 5 Page 26
Warm Up Setting of the warm-up time when the no
sample detection has been resolved 1 to 60 (min) 30 Page 26
UV Lamp Set Setting for the UV lamp Page 27
Total Confirmation and resetting of the total UV lamp time Page 27
Meas. Set Menu Measurement condition setting Page 28
Range Measurement range selection 1000, 500.0, 100.0 (ppb) 1000 Page 28
M.AVG Set Moving average count setting Page 28
TOC TOC moving average count setting 1 to 20 (count) 10 Page 28
Temp Temperature moving average count 1 to 20 (count) 10 Page 29
Cond Conductivity moving average count 1 to 20 (count) 10 Page 29
Err. Option Set Optional setting for error Page 30
Cond Over Input conductivity error selection Yes, No Yes Page 30
Temp Over Sample temperature error selection Yes, No Yes Page 31
Output Set Menu Output setting Page 32
Alarm Set Alarm contact output setting Page 32
Alarm1 Alarm contact output setting1 Err, Hold, TOC, Cond, Temp Err Page 32
Alarm2 Alarm contact output setting2 TOC (Fixed) TOC Page 34
Analog Set Transmittance output setting Page 35
Span Span point setting Depends on the set “Range”
of measurement range. 1000 Page 36