HP 437B User manual

HP
437B
SERVICE
MANUAL
HP
437B
POWER
METER
POWER
REF
5.00mW
Sots
BORE
FUNCTION
‘eemUN
oy
lat
DATA
ENTRY
ON7OrF
cal
STORE
RECALL
}
#
‘
ry
Ag
00
NOT
EXCEED
|
|
PRESET
?
SENSOR
MAX
PWR
u
cOcAL
H
‘SOLN
CAL
FAC
ON/OFF
AUTORNG
SENSOR
PWR
REF
ext
beesaans
{
Scie
i
et
tg
tte
ae
mg
i
mn
te
as
aaa
HP
Part
No.
00437-90016
E1088
HEWLETT
HP
Binder
Part
No.
9282-1077
GQ
PACKARD
aZer
dH

SERVICE
MANUAL
HP
437B
POWER
METER
SERIAL
NUMBERS
This
manual
applies
to
instruments
with
serial
numbers
prefixed
2801A
and
above.
For
additional
important
information
about
serial
numbers,
see
INSTRUMENTS
COVERED
BY
THIS
MANUAL
in
Section
1.
OD
|
HEWLETT
LE
PACKARD
©
Copyright
HEWLETT-PACKARD
COMPANY
1988
1501
PAGE
MILL
ROAD,
PALO
ALTO,
CALIFORNIA,
U.S.A.
SERVICE
MANUAL
PART
NO.
00437-90016
E1088
Microfiche
Service
Manual
Part
No.
00437-90018
Operating
Manual
Part
No.
00437-90015
Microfiche
Operating
Manual
Part
No.
00437-90017
Getting
Started
Guide
Part
No.
00437-90014

HP
437B
CERTIFICATION
Hewlett-Packard
Company
certifies
that
this
product
met
its
published
specifications
at
the
time
of
shipment
from
the
factory.
Hewlett-Packard
further
certifies
that
its
calibration
measurements
are
traceable
to
the
United
States
National
Bureau
of
Standards,
to
the
extent
allowed
by
the
Bureau’s
calibration
facility,
and
to
the
calibration
facilities
of
other
International
Standards
Organization
members.
WARRANTY
This
Hewlett-Packard
instrument
product
is
warranted
against
defects
in
material
and
workmanship
for
a
period
of
one
year
from
date
of
shipment.
During
the
warranty
period,
Hewlett-Packard
Company
will,
at
its
option,
either
repair
or
replace
products
which
prove
to
be
defective.
For
warranty
service
or
repair,
this
product
must
be
returned
to
a
service
facility
designated
by
HP.
Buyer
shall
prepay
shipping
charges
to
HP
and
HP
shall
pay
shipping
charges
to
return
the
product
to
Buyer.
However,
Buyer
shall
pay
all
ship-
ping
charges,
duties,
and
taxes
for
products
returned
to
HP
from
another
country.
HP
warrants
that
its
software
and
firmware
designated
by
HP
for
use
with
an
instrument
will
execute
its
programming
instructions
when
properly
installed
on
that
instrument.
HP
does
not
warrant
that
the
operation
of
the
instrument,
or
software,
or
firmware
will
be
uninterrupted
or
error
free.
LIMITATION
OF
WARRANTY
The
foregoing
warranty
shall
not
apply
to
defects
resulting
from
improper
or
inade-
quate
maintenance
by
Buyer,
Buyer-supplied
software
or
interfacing,
unauthorized
modification
or
misuse,
operation
outside
of
the
environmental
specifications
for
the
product,
or
improper
site
preparation
or
maintenance.
NO
OTHER
WARRANTY
IS
EXPRESSED
OR
IMPLIED.
HP
SPECIFICALLY
DISCLAIMS
THE
IMPLIED
WARRANTIES
OF
MERCHANTABILITY
AND
FITNESS
FOR
A
PARTICULAR
PURPOSE.
EXCLUSIVE
REMEDIES
THE
REMEDIES
PROVIDED
HEREIN
ARE
BUYER’S
SOLE
AND
EXCLUSIVE
REMEDIES.
HP
SHALL
NOT
BE
LIABLE
FOR
ANY
DIRECT,
INDIRECT,
SPE-
CIAL,
INCIDENTAL,
OR
CONSEQUENTIAL
DAMAGES,
WHETHER
BASED
ON
CONTRACT,
TORT,
OR
ANY
OTHER
LEGAL
THEORY.
ASSISTANCE
Product
maintenance
agreements
and
other
customer
assistance
agreements
are
available
for
Hewlett-Packard
products.
For
any
assistance,
contact
your
nearest
Hewlett-Packard
Sales
and
Service
Office.
Addresses
are
provided
at
the
back
of
this
manual.
WARRANTY
5

HP
437B
HP
437B
Herstellerbescheinigung
Hiermit
wird
bescheinigt,
dass
dieses
Gerit/System
in
Uber-
einstimmung
mit
den
Bestimmungen
von
Postverfiigung
1046/84
funkenstort
ist.
Der
Deutschen
Bundespost
wurde
das
Inverkehrbringen
dieses
Gerates/Systems
angezeigt
und
die
Berechtigung
zur
Uberpriifung
der
Serie
auf
Einhaltung
der
Bestimmungen
eingeraumt.
Zusatzinformation
fiir
Mess-
und
Testgeriite:
Werden
Mess-
und
Testgerite
mit
ungeschirmten
Kabeln
und/
oder
in
offenen
Messaufbauten
verwendet,
so
ist
vom
Betreiber
sicherzustellen,
dass
die
Funk-Entstérbestimmungen
unter
Be-
triebsbedingungen
an
seiner
Grundstiicksgrenze
eingehalten
werden.
Manufacturer’s
Declaration
This
is
to
certify
that
this
product
meets
the
radio
frequency
interference
requirements
of
Directive
FTZ
1046/1984.
The
German
Bundespost
has
been
notified
that
this
equipment
was
put
into
circulation
and
has
been
granted
the
right
to
check
the
product
type
for
compliance
with
these
requirements.
Note:
If
test
and
measurement
equipment
is
operated
with
unshielded
cables
and/or
used
for
measurements
on
open
set-ups,
the
user
must
ensure
that
under
these
operating
conditions,
the
radio
frequency
interference
limits
are
met
at
the
border
of
his
premises.
6
MANUFACTURER'S
DECLARATION

Warning
ip
Safety
Considerations
This
product
and
related
documentation
must
be
reviewed
for
familiarization
with
safety
markings
and
instructions
before
operation.
This
product
is
a
Safety
Class
I
instrument
(provided
with
a
protective
earth
terminal).
Before
Applying
Power
Verify
that
the
product
is
set
to
match
the
available
line
voltage
and
the
correct
fuse
is
installed.
Safety
Earth
Ground
An
uninterruptable
safety
earth
ground
must
be
provided
from
the
main
power
source
to
the
product
input
wiring
terminals,
power
cord,
or
supplied
power
cord
set.
Any
interruption
of
the
protective
(grounding)
conductor
(inside
or
outside
the
instrument)
or
disconnecting
the
protective
earth
terminal
will
cause
a
potential
shock
hazard
that
could
result
in
personal
injury.
(Grounding
one
conductor
of
a
two
conductor
outlet
is
not
sufficient
protection.)
In
addition,
verify
that
a
common
ground
exists
between
the
unit
under
test
and
this
instrument
prior
to
energizing
either
unit.
Whenever
it
is
likely
that
the
protection
has
been
impaired,
the
instrument
must
be
made
inoperative
and
be
secured
against
any
unintended
operation.
If
this
instrument
is
to
be
energized
via
an
autotransformer
(for
voltage
reduction)
make
sure
the
common
terminal
is
connected
to
neutral
(that
is,
the
grounded
side
of
the
mains
supply).
Servicing
instructions
are
for
use
by
service-trained
personnel
only.
To
avoid
dangerous
electric
shock,
do
not
perform
any
servicing
unless
qualified
to
do
sO.
Adjustments
described
in
the
manual
are
performed
with
power
supplied
to
the
instrument
while
protective
covers
are
removed.
Energy
available
at
many
points
may,
if
contacted,
result
in
personal
injury.
Capacitors
inside
the
instrument
may
still
be
charged
even
if
the
instrument
has
been
disconnected
from
its
source
of
supply.
SAFETY
CONSIDERATIONS
7

HP
437B
Warning
7
Caution
db
For
continued
protection
against
fire
hazard.
replace
the
line
fuse(s)
only
with
250V
fuse(s)
of
the
same
current
rating
and
type
(for
example,
nomial
biow,
time
delay,
etc.).
Do
not
use
repaired
fuses
or
short
circuited
fuseholders.
Safety
Symbols
symbol
when
it is
necessary
for
the
user
to
refer
to
the
instruction
manual
(see
Table
of
Contents
for
page
references).
(
Instruction
manual
symbol:
The
product
will
be
marked
with
this
5
Indicates
hazardous
voltages.
9
Indicates
earth
(ground)
terminal.
The
WARNING
sign
denotes
a
hazard.
It
calls
attention
to
a
procedure,
practice,
or
the
like,
which,
if
not
correctly
performed
or
adhered
to,
could
result
in
personal
injury.
Do
not
proceed
beyond
a
WARNING
sign
until
the
indicated
conditions
are
fully
understood
and
met.
The
CAUTION
sign
denotes
a
hazard.
It
calls
attention
to
an
operating
procedure,
practice,
or
the
like,
which,
if
not
correctly
performed
or
adhered
to,
could
result
in
damage
to
or
destruction
of
part
or
all
of
the
product.
Do
not
proceed
beyond
a
CAUTION
sign
until
the
indicated
conditions
are
fully
understood
and
met.
8
SAFETY
CONSIDERATIONS
@

ADJUSTMENTS
5-1.
Introduction
5-2.
Safety
Considerations
Warning
Va
5-3.
Equipment
Required
5-4.
Factory
Selected
Components
This
section
contains
adjustments
and
checks
that
ensure
proper
performance
of
the
Power
Meter.
Adjustments
are
not
required
on
any
fixed
periodic
basis,
and
normally
are
performed
only
after
a
performance
test
has
indicated
that
some
parameters
are
out
of
specification.
Performance
tests
should
be
completed
after
any
repairs
that
may
have
altered
the
characteristics
of
the
instrument.
The
test
results
will
make
it
possible
to
determine
whether
or
not
adjustments
are
required.
Allow
30
minutes
for
the
Power
Meter
to
warm
up,
and
then
remove
the
top
and
bottom
covers,
for
access
to
the
test
and
adjustment
points.
To
determine
which
performance
tests
and
adjustments
to
perform
after
a
repair,
refer
to
paragraph
5-5,
Post-Repair
Adjustments.
This
section
contains
a
warning
that
must
be
followed
for
your
protection
and
to
avoid
damage
to
the
equipment
being
used.
Adjustments
described
in
this
section
are
performed
with
power
applied
to
the
instrument
and
with
protective
covers
removed.
Maintenance
should
be
performed
only
by
trained
personnel
who
are
aware
of
the
hazards
involved.
When
the
maintenance
procedure
can
be
performed
without
power,
the
power
should
be
removed.
Most
of
the
adjustment
procedures
include
a
list
of
recommended
test
equipment.
The
test
equipment
is
also
identified
on
the
test
setup
diagrams.
If
substitutions
must
be
made,
the
equipment
used
must
meet
the
critical
specification
listed
in
Table
1-3
in
Section
I.
Factory
selected
components
are
identified
on
the
schematics
and
parts
lists
by
an
asterisk(*)
which
follows
the
reference
designator.
The
nominal
value
of
the
selected
component
is
shown.
Table
5-1
lists
the
reference
designator,
the
service
sheet
where
the
component
is
shown,
the
value
range,
and
the
basis
for
selecting
a
particular
value.
ADJUSTMENTS
5-1

HP
437B
Make
adjustments
only
in
the
order
specified.
Note
4%
5-5.
Post-Repair
Table
5-2
lists
the
adjustments
related
to
repairs
or
replacement
of
any
of
the
Adjustments
assemblies.
Table
5-1.
Factory
Selected
Components
Reference
Service
Range
of
Basis
of
Selection
Designator
Sheet
Values
A4R73
A4C
8252
1.
If
the
reference
power
is
outside
the
A4VR3
with
5.11V
range
of
1.000
+/-
0.007
mW
between
OC
and
5SC,
and
if
the
A4R73,
A4VR3
combination
is
825
Q
&
5.11V
then
change
the
A4R73,
A4VR3
combination
to
1470
2
&
8.25V.
However,
if
the
'
A4R73,
A41VR3
combination
is
already
1470
2
&
8.25V,
then
a
problem
exists
elsewhere.
Zener
or
1470
Q
with
8.25V
Zener
combination
Table
5-2.
Post-Repair
Adjustments,
Tests,
and
Checks
Assembly
Related
Adjustments
or
Reference
Repaired
Performance
Test
Service
Sheet
Al
Display
None
A3B
A2
Keyboard
None
A3B
A3
Central
Processing
Unit
4-7,
4-8
A3A-A3C
A4
Analog
Assembly
4-8, 4-9,
5-6, 5-7,
5-8,
5-9
A4A-A4C
(Includes
Ref
Oscillator)
50
MHz
Reference
Oscillator
4-10,
5-8,
5-9
A4C
(Part
of
Analog
Assembly)
5-2
ADJUSTMENTS

HP
437B
5-6.
220
Hz
Frequency
Adjustment
Reference
Description
Equipment
Procedure
Service
Sheet
A4C
The
220
Hz
is
adjusted
for
maximum
Power
Meter
readout.
RANGE
POWER
METER
CALIBRATOR
POWER
METER
Figure
5-1.
220
Hz
Frequency
Adjustment
Setup
Range
Calibrator
..........c.ccccececesceteceeeceseeceenenssnaeeeees
HP
11683A
1.
Turn
on
both
the
Power
Meter
and
the
range
calibrator.
Set
the
range
calibrator
controls
as
follows:
LINE:
eiceccded
ox
dh
eede
ve
bees
vee
ee
eednatntels
oeescu
dee
eve
eee
seed
ee
ees
nana
des
ON
RANGE
\0).0.55
cavgda
vob
ea
@iaks outer
ote
aicandie
eas
ae
Sadan
coaagaaassireces
1
mW
FUNCTION
(sie
cv
sess
cities
tales
tia
syrnie
Sands
ole
a
oye
bow
daly
Saree
eclp
tele
oe
STANDBY
POLAREDY
©
si
osc
seco
Resin
diioas
sieves
liot
ei
sed
oud
vadewsekewee
NORMAL
2.
Connect
the
range
calibrator
to
the
Power
Meter
as
shown
in
Figure
5-1.
3.
Press
the
PRESET/LOCAL
key,
then
the
ENTER
key.
ADJUSTMENTS
5-3

HP
437B
4.
Press
the
ZERO
key
on
the
Power
Meter,
and
allow
time
(5
to
15
seconds)
for the
Power
Meter
zeroing
routine
to
finish.
5.
Set
the
range
calibrator’s
FUNCTION
switch
to
CALIBRATE.
6.
Press
the
Power
Meter’s
CAL
(SHIFTed
ZERO)
key.
7.
Using
(2).
(+),
Co),
or
G2,
modify
the
REF
CF
to
read
100.0%.
8.
Press
the
ENTER
key.
9.
Adjust
A4R87
(FREQ)
for
the
maximum
front
panel
reading.
5-4
ADJUSTMENTS

HP
437B
5-7.
Ranges
4
and
5
Shaper
Adjustment
Reference
Service
Sheet
A4A
Description
Ranges
4
and
5
Shaper
circuits
are
adjusted
for
proper
gain.
'
RANGE
POWER
METER
See
POWER
METER
Figure
5-2.
Ranges
4
and
5
shaper
Adjustment
Setup
Equipment
Range
Callbrator
¢
scien
cincs
sor
eel
a
tea
eaeavarco
ws
teadandeeatnes
yecte
HP
11683A
Procedure
1.
Connect
range
calibrator
to
the
Power
Meter
as
shown
in
Figure
5-2.
2.
Set
the
range
calibrator
controls
as
follows:
INE
tik
es
be
tetas
iPS
Foes
led
hee
aatenite
aelds
nasiek
deus
ba wa
Macealasegadns
ON
RANGE:
fe
scost
ib
ide
ssadaa
fica
tioia'ds
eaeeas
couse
toned
dows
dade
netic’
s
|
mW
FUNCTION
cecjekd
eyiedidage
karst
tena
aatanteedi
es
tee
nieeygeteaacit
STANDBY
POGARULY”
302i
ce-ch
ten
reek
bad
Behe
ieee
tls
PEN
has
Sy
a
Sod
NORMAL
@
3.
Press
the
Power
Meter’s
LINE
switch
to
ON.
ADJUSTMENTS
§-5

HP
437B
5-6
ADJUSTMENTS
10.
11.
12.
13.
14.
Press
the
PRESET/LOCAL
key,
then
the
ENTER
key.
Press
the
ZERO
key,
and
allow
time
(5-15
seconds)
for
the
zeroing
routine
to
finish.
Set
the
range
calibrator’s
FUNCTION
switch
to
CALIBRATE.
Press
the
Power
Meter’s
CAL
(SHIFTed
ZERO)
key.
Using
(<),
(+),
GD,
or
CO),
modify
the
REF
CF
to
read
100.0%
Press
the
ENTER
key.
Set
the
range
calibrator’s
RANGE
to
10
mW.
Adjust
A4RI11
(RNG
4
SHP)
for
a
reading
of
10.00
+0.0!1
mW
on
the
Power
Meter
display.
Set
the
range
calibrator’s
RANGE
to
100
mW.
Adjust
A4R112
(RNG
5
SHP)
for
a
reading
of
100.0
+0.1
mW
on
the
Power
Meter
display.
Repeat
steps
10
through
13
to
check
that
interaction
between
steps
has
not
caused
a
shift
in
settings.

HP
437B
@
5-8.
Power
Reference
Frequency
Oscillator
Adjustment
Note
Reference
Description
Equipment
rk
Adjustment
of
the
Power
Reference
Oscillator
frequency
may
also
affect
the
a
|
output
level
of
the
oscillator.
Thus,
after
the
frequency
is
adjusted
to
50.0
+0.5
MHz,
the
output
level
should
be
checked
as
described
in
Section
4.
A
procedure
for
adjusting
the
output
to
the
specified
level
is
provided
in
the
next
paragraph.
Service
Sheet
A4C
Variable
inductor
A4L5
is
adjusted
to
set
the
power
reference
oscillator
output
frequency
to
50.0
+0.5
MHz.
HP
4378
POWER METER
FREQUENCY
COUNTER
POWER
REF
1.00
mW
SO
MHz
Figure
5-3.
Power
Reference
Oscillator
Frequency
Adjustment
Setup
Frequency
Counter
..........
ccc
cece
cece
eee
e
een
e
een
eeees
HP
5328A
Option
031
ADJUSTMENTS
5-7

HP
437B
Procedure
5-8
ADJUSTMENTS
Connect
the
equipment
as
shown
in
Figure
5-3.
Set
up
the
counter
to
measure
frequency.
Set
the
Power
Meter
LINE
switch
to
ON.
Press
the
PRESET/LOCAL
key,
then
the
ENTER
key.
Press
the
PWR
REF
(SHIFTed
(J)
key.
Observe
the
indication
on
the
frequency
counter.
If
it
is
50.0
+0.5
MHz
no
adjustment
of
the
power
reference
oscillator
frequency
is
necessary.
If
it is
not
within
these
limits,
adjust
the
power
reference
oscillator
frequency
as
described
in
steps
6
and
7.
y
Remove
the
Power
Meter
bottom
cover.
Adjust
A4L5
(FREQ)
to
obtain
a
50.00
+0.5
MHz
indication
on
the
frequency
counter.

HP
437B
5-9.
Power
Reference
Oscillator
Level
Adjustment
Reference
Description
Note
UE
|
HP
437B
POWER
METER
POWER
REF
1.00
mW
50
MHz
es
a
=m,
QODdQdo0|O0C0O
20
Beecboo|
|
|
Service
Sheet
A4C
The
power
reference
oscillator
is
factory-adjusted
to
1.0
mW
+0.7%
using
a
special
measurement
system
accurate
to
0.5%
traceable
to
the
National
Bureau
of
Standards
and
allowing
for
a
0.2%
transfer
error.
To
ensure
maximum
accuracy
in
readjusting
the
power
reference
oscillator,
the
following
procedure
provides
step-by-step
instructions
for
using
specified
Hewlett-Packard
instruments
of
known
capability.
If
equivalent
instruments
are
used,
signal
acquisition
criteria
may
vary
and
reference
should
be
made
to
the
manufacturer’s
guidelines
for
operating
the
equipment.
The
Power
Meter
may
be
returned
to
the
nearest
HP
office
to
have
the
power
reference
oscillator
checked
and/or
adjusted.
Refer
to
Section
2,
PACKAGING.
HP
_
432A
POWER
METER
meve
DIGITAL
VOLTMETER
THERMI
STOR
MOUNT
Figure
5-4.
Power
Reference
Oscillator
Level
Adjustment
Setup
ADJUSTMENTS
5-9

HP
437B
Equipment
Procedure
5-10
Note
ADJUSTMENTS
Test
Power
Meter......0
0.0.
c
cece
cece
tee
eeaeeeeeeeeeneens
HP
432A
Thermistor
Mount...........cc
ccc
cece
cee
cececceeveeuceseugeees
HP
478A-H75
Digital
Voltmeter
(DVM)......
0.0.0...
cece
ev
cceccceeaeacuceeneneuegs
HP
3456A
1.
Set
up
the
DVM
to
measure
resistance.
Connect
the
DVM
between
the
>
9.
10.
Vrf
connector
on
the
rear
panel
of
the
test
power
meter
and
pin
|
on
the
thermistor
mount
end
of
the
test
power
meter
interconnect
cable.
Round
off
the
DVM
indication
to
two
decimal
places
and
record
this
value
as
the
internal
bridge
resistance
(R)
of
the
test
power
meter
(approximately
200
ohms).
R
(Internal
Bridge
Resistance)
Connect
the
test
power
meter
to
the
Power
Meter
as
shown
in
Figure
5-4.
Set
the
Power
Meter
LINE
switch
to
ON.
Ensure
that
the
PWR
REF
is
off.
Wait
thirty
minutes
for
the
test
power
meter
thermistor
mount
to
stabilize
before
proceeding
to
the
next
step.
4
Set
the
test
power
meter
range
switch
to
coarse
zero
and
adjust
the
@
front
panel
coarse
zero
control
to
obtain
a
zero
meter
indication.
Fine
zero
the
test
power
meter
on
the
most
sensitive
range,
then
set
the
power
meter
range
switch
to
1
mW.
rT)
Ensure
that
the
DVM
input
leads
are
isolated
from
chassis
ground
when
oF
performing
the
next
step.
Set
up
the
DVM
to
measure
microvolts.
Connect
the
positive
and
negative
input
leads,
respectively,
to
the
Vcomp
and
Vrf
connectors
on
the
rear
panel
of
the
test
power
meter.
Observe
the
indication
on
the
DVM.
If
less
than
400
microvolts,
proceed
with
the
next
step.
If
400
microvolts
or
greater,
press
and
hold
the
test
power
meter
fine
zero
switch
and
adjust
the
coarse
zero
control
so
that
the
DVM
indicates
200
microvolts
or
less.
Then
release
the
fine
zero
switch
and
proceed
to
the
next
step.
Round
off
the
DVM
indication
to
the
nearest
microvolt
and
record
this
value
as
VO.
vw

HP
437B
11.
Disconnect
the
DVM
negative
input
lead
from
the
Vrf
connector
on
the
test
power
meter
and
reconnect
it
to
chassis
ground.
12.
Press
the
PWR
REF
(SHIFTed
(7_))
key
to
turn
the
reference
oscillator
on.
13.
Record
the
indication
observed
on
the
DVM
as
Vcomp.
Vcomp
14,
Disconnect
the
DVM
negative
input
lead
from
chassis
ground
and
reconnect
it
to
the
Vrf
connector
on
the
rear
panel
of
the
test
power
meter.
The
DVM
is
now
set
up
to
measure
V1
which
represents
the
power
reference
oscillator
output
level.
15.
Calculate
the
value
of
V1
equal
to
|
mW
from
the
following
equation:
Vi
—
Yo
=
Voomp
—
V
(Vcomp)?
—
(10-8)
(4R)
(EFFECTIVE
EFFICIENCY)
Where:
Vo
=
previously
recorded
value
Vcomp
=
previously
recorded
value
10-3
=
1
milliwatt
R
=
previously
recorded
value
EFFECTIVE
value
for
thermistor
mount
at
50
MHz
EFFICIENCY
(traceable
to
the
National
Bureau
of
Standards).
16.
Remove
the
Power
Meter’s
top
cover
and
adjust
A4R114
(LEVEL)
until
the
DVM
indicates
the
calculated
value
of
V1.
ADJUSTMENTS
5-1]

HP
437B
Typical
Calculations
5-12
ADJUSTMENTS
ACCURACY:
DVM
Measurements:
(Vcomp)
+0.018%
(V;
—
Vo)
+0.023%
(R)
+0.03%
Math
Assumptions:
+0.01%
EFFECTIVE
EFFICIENCY
CAL
(NBS):
+0.5%
MISMATCH
UNCERTAINTY:
+0.1%
(Source
and
Mount
SWR
<1.05)
<+0.7%
MATH
ASSUMPTIONS:
ae
2Vcomp
(V1
—
Vo)
+
Vo?
—
Vi?
RF
“
(4R)
(EFFECTIVE
EFFICIENCY)
Assume:
Vp*—
V,2
=
(V;
—
Vo)?
=
(V;
=Vg)
7
oe
-V,2
+
2V,
=
Vo?
Want:
Vo2
—
V,?
«error
=
(V,2
+
2V,
Vo
—
Vo2)
—
(Vo?
—
V2)
if
2V9
(Vi
—
Vo)
<
2Veomp
(Vi
—
Vo)
i-e.,
Vo
<
Veomp:,
error
is
negligible.
Vcomp
~
4
volts.
If
Vo
<
400
xV,
error
is
<
0.01%.
(Typically
Vp
can
be
set
to
<50
nV).
Derivation
of
Formula
for
V,;
—
Vo
eee:
2Vcomp
(V,
—
Vo)
+
Vo?
—
Vi?
RF
“
(4R)
(EFFECTIVE
EFFICIENCY)
Desired
Ppp
=
1
mW
=
1073
2Vcomp
(V;
—
Vo)
+
Vo?
—
Vi?
=
10-3
SS
(4R)
(EFFECTIVE
EFFICIENCY)
Let
(4R)
(EFFECTIVE
EFFICIENCY)
(1078)
=
K
Substitute
—
(V,
—
Vo)?
for
Vo?
—
V,?
(see
Math
Assumptions
under
Accuracy).
Then
0
=
(V,—
Vo)?
—
2Vcomp
(V;
—
Vo)
+
K
or
V;)
—
Vo
=
Vcomp
—
V_
(Vcomp)?
—
K

SERVICE
8-1.
General
Information
8-2.
Service
Sheets
8-5.
Safety
Considerations
This
section
contains
information
required
for
servicing
the
Power
Meter.
It
includes
block
diagrams
with
theory
and
troubleshooting,
schematic
diagrams,
and
schematic
diagram
notes.
The
foldout
pages
in
the
last
part
of
this
section
are
called
the
service
sheets
(SS).
They
contain
block
diagrams,
schematic
diagrams,
supplemental
diagrams,
and
associated
information.
8-3.
Block
Diagrams
The
block
diagram
and
related
service
information
is
found
on
Service
Sheet
BD1.
BDI
is
an
overall
block
diagram
that
shows
major
functional
sections
of
the
Power
Meter.
It
serves
as
an
index
to
the
more
detailed
information
on
the
succeeding
service
sheets
and
is
the
starting
point
for
most
trouble
shooting
procedures.
8-4.
Schematic
Diagrams
Service
Sheet
|
is
a
detailed
block
diagram
and
mechanical
assembly
drawing
of
the
Al
Display
Assembly
and
the
A2
Keyboard
Assembly.
These
assemblies
are
replaced
as
assemblies
only,
and
SS1
is
provided
to
show
interconnects
between
these
assemblies
and
the
A3
Digital
board.
The
schematic
diagrams
and
their
related
information
are
presented
in
the
additional
seven
Service
Sheets
A3a,
A3b,
A3c,
A3d,
and
Ada,
A4b,
Adc.
These
diagrams,
in
functional
groupings,
are
designed
to
aid
in
understanding
the
principles
of
operation
and
to
aid
in
troubleshooting
the
Power
Meter.
Before
applying
power,
verify
that
the
instrument
is
properly
set
to
operate
from
the
available
line
voltage
and
that
the
correct
fuse
is
installed.
An
uninterrupted
safety
earth
ground
must
be
provided
from
the
main
power
source
to
the
instrument
input
wiring
terminals,
power
cord,
or
supplied
power
cord
set.
SERVICE
8-1

HP
437B
8-2
Warning
ie
Caution
mm
SERVICE
8-6.
Warnings
and
Cautions
Pay
attention
to
the
WARNINGS
and
CAUTIONS.
They
must
he
followed
for
your
protection
and
to
avoid
damage
to
the
equipment.
Maintenance
described
herein
must
be
performed
with
power
supplied
to
the
instrument
and
with
the
protective
covers
removed.
Such
maintenance
should
be
performed
only
by
service
trained
personnel
who
are
aware
of
the
hazards
involved
(for
example,
fire
and
electrical
shock).
Where
maintenance
can
be
performed
without
power
applied,
the
power
should
be
removed.
Any
interruption
of
the
protective
(grounding)
conductor
(inside
or
outside
of
the
instrument),
or
disconnection
of
the
protective
earth
terminal
will
create
a
potential
shock
hazard
and
could
result
in
personal
injury.
Grounding
one
conductor
of
a
two-conductor
outlet
is
not
sufficient.
Whenever
it
is
likely
that
the
protection
has
been
impaired,
the
instrument
must
be
made
inoperative
(i.e.,
secured
against
unintended
operation).
If
the
instrument
is
to
be
energized
via
an
autotransformer
(for
voltage
reduction),
make
sure
the
common
terminal
is
connected
to
neutral
(that
is,
the
grounded
side
of
the
mains
supply).
Make
sure
that
the
250
volt
normal-blow
fuses
with
the
specified
current
rating
are
used
for
replacement.
Do
not
use
repaired
fuses
or
short-
circuited
fuse
holders.
To
do
so
could
create
a
shock
or
fire
hazard.
Some
printed
circuit
boards
contain
devices
that
may
be
damaged
if
the
board
is
removed
or
installed
while
the
power
is
on.
Verify
that
the
LINE
switch
is
OFF
or
that
the
power
cord
is
unplugged
before
you
remove
or
install
a
printed
circuit
board.
After
removing
devices
from
sockets,
store
the
devices
with
the
pins
in
conductive
foam.
This
will
prevent
accidental
damage
from
a
static
discharge.
Table of contents
Other HP Measuring Instrument manuals