HP 333A Service manual

OPERATING
AND
SERVICE
MANUAL
MODEL
333A/334A
DISTORTION
ANALYZER
Serial
Numbers:
333A:
1137А03146
and
greater
334A:
1140А05641
and
greater
Appendix
С,
Manual
Backdating
Changes,
adapts
this
manual
to
lower
serial
numbers.
WARNING
To
help
minimize
the
possibility
of
electrical
fire
or
shock
hazards,
do
not
expose
this
instrument
to
rain
or
excess
moisture.
Manual
Part
No.
00333-90008
Microfiche
Part
No.
00333-90058
Copyright
Hewlett-Packard
Company
19
66
P.O.
Box
301,
Loveland,
Colorado,
80537
U.S.A.
Printed:
October
1975

HEWLETT
ш
PACKARD
CERTIFICATION
Hewlett-Packard
Company
certifies
that
this
instrument
met
its
published
specifications
at
the
time
of
shipment
from
the
factory.
Hewlett-Packard
Company
further
certifies
that
its
calibration
measurements
are
traceable
to
the
United
States
National
Bureau
of
Standards,
to
the
extent
allowed
by
the
Bureau’s
calibration
facility,
and
to
the
calibration
facilities
of
other
International
Standards
Organization
members.
WARRANTY
AND
ASSISTANCE
This
Hewlett-Packard
product
is
warranted
against
defects
in
materials
and
workmanship
for
a
period
of
one
year
from
the
date
of
shipment,
except
that
in
the
case
of
certain
components,
if
any,
listed
in
Section
I
of
this
operating
manual,
the
warranty
shall
be
for
the
specified
period.
Hewlett-Packard
will,
at
its
option,
repair
or
replace
products
which
prove
to
be
defective
during
the
warranty
period
provided
they
are
returned
to
Hewlett-Packard,
and
provided
the
proper
preventive
maintenance
procedures
as
listed
in
this
manual
are
followed.
Repairs
necessitated
by
misuse
of
the
product
are
not
covered
by
this
warranty.
NO
OTHER
WARRANTIES
ARE
EXPRESSED
OR
IMPLIED,
INCLUDING
BUT
NOT
LIMITED
TO
THE
IMPLIED
WARRANTIES
OF
MERCHANTABILITY
AND
FITNESS
FOR
A
PARTICULAR
PURPOSE.
HEWLETT-PACKARD
IS
NOT
LIABLE
FOR
CONSEQUENTIAL
DAMAGES.
If
this
product
is
sold
as
part
of
a
Hewlett-Packard
integrated
instrument
system,
the
above
warranty
shall
not
be
applicable,
and
this
product
shall
be
covered
only
by
the
system
warranty.
Service
contracts
or
customer
assistance
agreements
are
available
for
Hewlett-Packard
products.
For
any
assistance,
contact
your
nearest
Hewlett-Packard
Sales
and
Service
Office.
Addresses
are
provided
at
the
back
of
this
manual.

Model
333A/334A
Section
I
SECTION
|
GENERAL
INFORMATION
1-1.
DESCRIPTION.
1-2.
The
Hewlett-Packard
Models
333A
and
334A
Distortion
Analyzers
are
solid
state
instruments
for
measuring
distortion
on
ac
voltages.
The
Models
333A
and
334A
include
two
control
loops
that
auto-
matically
tune
both
legs
of
a
bridge
circuit
which
re-
jects
the
fundamental
when
the
rejection
circuit
is
initially
set
within
the
range
of
the
loops.
The
334A
has
a
high
impedance
detector
which
operates
from
550
kHz
to
greater
than
65
MHz
and
provides
the
ca-
pability
of
monitoring
the
distortion
of
the
amplitude
modulation
on
an
rf
carrier.
1-3.
Distortion
levels
of
0.
1%
to
100%
full
scale
are
measured
in
seven
ranges
for
any
fundamental
frequen-
cy
of
5
Hz
to
600
KHz.
Harmonics
are
indicated
upto
3
MHz.
The
high
sensitivity
of
these
instruments
re-
quires
only
0.
3V
rms
for
the
100%
set
level
reference.
The
distortion
characteristics
can
be
monitored
atthe
OUTPUT
connectors
with
an
oscilloscope,
a
true
rms
voltmeter,
ог
a
wave
analyzer.
The
instruments
are
capable
of
an
isolation
voltage
of
400
volts
above
chassis
ground.
1-4.
The
voltmeter
can
be
used
separately
for
general
purpose
voltage
and
gain
measurements.
It
has
a
frequency
range
of
b
Hz
to
3
MHz
(20
Hz
to
500
kHz
for
300
uV
range)
and
а
voltage
range
of
300uV
to
300
V
rms
full
scale,
1-5.
The
AM
detector
included
in
the
Model
334A
is
a
broadband
дс
restoring
peak
detector
consisting
of
@
semiconductor
diode
and
filter
circuit.
AM
distor-
.
tion
levels
as
low
as
0.
3%
can
be
measured
on
a
3
V
to8
V
rms
carrier
modulated
30%
in
the
standard
broadcast
band.
Distortion
less
than
1%
can
be
measured
at
the
same
level
of
the
carrier
up
to
65
Mc.
Table
1-1.
MODEL
333A
DISTORTION
MEASUREMENT
RANGE
Any
fundamental
frequency,
5
Hz
to
600
kHz.
Distortion
levels
of
0.
1964-1009
are
measured
full
scale
in
7
ranges.
DISTORTION
MEASUREMENT
ACCURACY
Harmonic
measurement
accuracy
(full
scale)
Fundamental
Input
Less
Than
30
V
100%-0.
3%
0.1%
1-6.
ACCESSORY
FEATURES.
1-7.
The
accessory
available
with
the
333A
and334A
Distortion
Analyzers
is
a
voltage
divider
probe,
-hp-
Model
Ко.
10001A.
The
features
of
the
probe
are:
a.
10
megohms
shunted
by
10
pF,
giving
10:1
attenuation.
b,
DC
to
30
MHz
bandwidth.
2%
division
accuracy.
d.
600
V
peak
input.
e.
5
ns
rise-time.
1-8.
OPTION.
1-9,
Option
01іѕа
standard
-hp-
Model
333A
or
334A
with
a
special
meter
and
meter
amplifier,
compen-
sated
to
permit
response
to
VU
(volume
units)
characteristics.
1-10
INSTRUMENT
IDENTICATION.
1-11.
Hewlett-Packard
uses
a
two-section
serial
number.
The
first
section
(prefix)
identifies
a
series
of
instruments.
The
last
section
(suffix)
identifies
a
particular
instrument
within
the
series.
На
letter
is
included
with
the
serial
number,
it
identifies
the
coun-
try
in
which
the
instrument
was
manufactured.
Н
the
Serial
prefix
of
your
instrument
differs
from
the
one
on
the
title
page
of
this
manual,
a
change
sheet
will
be
supplied
to
make
this
manual
compatable
with
new
-
er
instruments
or
the
backdating
information
in
Арреп-
dix
C
will
adapt
this
manual
to
earlier
instruments.
All
correspondence
with
Hewlett-Packard
should
in-
clude
the
complete
serial
number.
Specifications
Fundamental
Input
Greater
Than
30
V
100%-0.
3%
10
На
300
ЕН?
30
Hz
300
ЕН?
10
Hz
3
MHz
10
Hz
1.2
MHz
0.1%
Elimination
Characteristics:
Fundamental
Rejection
>
80
dB
Second
Harmonic
Accuracy
for
a
fundamental
of:
5
Hz
to
20
Hz:
better
than
+1
dB
20
Hz
to
20
kHz:
better
than
«0.
6
dB
20
kHz
to
100
kHz:
better
than
-1
dB
100
kHz
to
300
kHz:
better
than
-2
dB
300
kHz
to
600
kHz:
better
than
-3
dB
1-1

Section
I
Model
333A/334A
Table
1-1.
Specifications
(Cont'd)
Distortion
Introduced
by
Instrument:
>
-70
АВ
from
5
Hz
to
200
kHz
>
-64
dB
from
200
kHz
to
600
kHz
Meter
indication
is
proportional
to
the
aver-
age
value
of
a
waveform.
FREQUENCY
CALIBRATION
ACCURACY
Better
than
+5%
from
5
Hz
to
300
kHz
Better
than:
1096
from
300
kHz
to
600
kHz
INPUT
IMPEDANCE
Distortion
Mode:
1
МО
+5%
shunted
by
«70
pF.
Voltmeter
Mode:
1
МО
+5%
shunted
by
«30
pF
(333A
only),
1
MQ
+5%
shunted
by
«35
pF
(334A
only),
1
to
300
V
ranges;
1
МО
+5%
shunted
by
«70
pF,
300
u
V
to
0.3
V
ranges.
INPUT
LEVEL
FOR
DISTORTION
MEASUREMENTS
0.3
V
rms
for
100%
set
level
(up
to
300
V
may
be
attenuated
to
set
level
reference).
The
minimum
measurable
distortion
for
floating
operation
on
the
X1
frequency
range
is
50dB
below
the
fundamental.
DC
ISOLATION
Signal
ground
may
be
+400
Vdc
from
external
chassis,
VOLTMETER
RANGE
300
рУ
to
300
V
rms
full
scale
(13
ranges),
10
dB
per
range.
VOLTMETER
FREQUENCY
RANGE
5
Hz
to
3
MHz
(300
uV
range:
20
Hz-500
kHz).
VOLTMETER
ACCURACY:
20
Hz-500
kHz
5
Hz-3
MHz
5
Hz-500
kHz
30
Hz-300
kHz
lmV-30V
1
10
Hz-i
MHz
100
V-300
V
|
10
Hz-300
kHz
300
ру
MODEL
334A
Same
as
Model
333A
except
as
indicated
below:
AM
DETECTOR
High
impedance
dc
restoring
peak
detector
with
semi-conductor
diode
operates
from
550kHz
to
greater
than
65
MHz.
Broadband
input.
Maximum
input;
40
V
p-p
ac
or
40
V
peak
transient.
.
CARRIER
FREQUENCY
550kHzto
1.6
MHz:
Distortion
introduced
by
detector
is
«
0.
3%
for
3
to
8
volt
carriers
modulated
30%.
i-2
NOISE
MEASUREMENTS
Voltmeter
residual
noise
on
the
300
pV
range:
«
25
u
V
rms
terminated
in
shielded
6000;
«30
u
V
rms
terminated
in
shielded
100
КО.
OUTPUT
For
input
frequencies
from
20
Hz
to
600
kHz,
0.1
V
rms
+
0.01
V
open
circuit
for
full
scale
meter
deflection;
0,05
V
rms
+
0.005
V
into
2
КО
for
full
scale
meter
deflection.
AUTOMATIC
NULLING
MODE
Set
Level:
At
least
0,2
V
rms.
Frequency
Ranges:
X1,
manual
null
tuned
to
less
than
3%
of
set
level;
total
frequency
hold-in
+0.
5%
about
true
manual
null.
X10
through
ХОК,
manual
null
tuned
to
less
than
1090
of
set
level;totalfrequency
hold-in
+1%
about
true
manual
null,
AUTOMATIC
NULL
ACCURACY
5
Hz
to
100
Hz:
Meter
reading
within
0
to
+3
dB
of
manual
null.
100
Hz
to
600
kHz:
Meter
reading
within
0
to
+1.5
dB
of
manual
null.
HIGH-PASS
FILTER
3
dB
point
at
400
Hz
with
18
dB
per
octave
roli
off.
60
Hz
rejection
>
40
dB.
Normally
used
only
with
fundamental
frequencies
greater
than
1
kHz.
POWER
SUPPLY
100
v/120
V/220
V/240
V
+
5%
=
10%,
48
-
66
Hz,
approximately
4
watts.
1.6
MHz
to
65
MHz:
Distortion
introduced
by
detector
is
<
1%
for
3
to
8
volts
rms
carriers
modulated
30%.
NOTE
Distortion
measurement
at
carrier
levels
as
low
as
i
volt
may
be
made
with
reduced
accuracy.
OPTION:
01
Indicating
meter
has
VU
characteristics
con-
forming
to
FCC
Requirements
for
AM,
FM,
and
TV
broadcasting.

Model
333А
/334А
Section
П
SECTION
И
INSTALLATION
2-1.
INTRODUCTION.
2-2.
This
section
contains
information
and
instruc-
tions
necessary
forthe
installation
and
shipping
of
the
Models
333A/334A
Distortion
Analyzers.
Included
are
initial
inspection
procedures,
power
and
grounding
requirements,
installation
information,
and
instruc-
tions
for
repackaging
for
shipment.
2-3.
INSPECTION.
2-4,
This
instrument
was
carefully
inspected
both
mechanically
and
electrically
before
shipment.
It
shouldbe
physically
free
of
mars
or
scratches
and
in
perfect
electrical
order
upon
receipt.
To
confirm
this,
the
instrument
should
be
inspected
for
physical
damage
intransit.
Also
checkfor
supplied
accessories,
and
test
the
electrical
performance
of
the
instrument
using
the
procedure
outlined
in
Paragraph
5-5.
If
there
is
damage
or
deficiency,
see
the
warranty
on
the
inside
front
cover
of
this
manual.
2-5.
POWER
REQUIREMENTS.
2-6.
The
Model
333A/334A
can
be
operated
from
any
ac
source
of
100
V/120
V/220 V/240
V
+
5%
-
10%,
48
-
66
Hz.
With
the
instrument
disconnected
from
the
ac
power
source,
move
the
slide
(located
on
the
rear
panel)
until
the
desired
line
voltage
appears.
Power
dissipation
is
approximately
4
watts.
2-7.
THREE-CONDUCTOR
POWER
CABLE.
2-8.
То
protect
operating
personnel,
the
National
Electrical
Manufacturers’
Association
(NEMA)
recom-
тепа
that
the
instrument
panel
and
cabinet
be
grounded.
АП
Hewlett-Packard
instruments
are
equipped
with
a
three-conductor
power
cable,
which,
when
plugged
in-
toanappropriate
receptacle,
grounds
the
instrument.
The
offset
pin
on
the
power
cable
three-prong
connec
=
tor
is
the
ground
wire.
2-9
INSTALLATION.
2-10.
The
Model
333A/334A
is
fully
transistorized;
therefore,
no
special
cooling
is
required.
However,
the
instrument
should
not
be
gperated
where
the
am-
bient
temperature
exceeds
55°
C
(131
Е).
2-11.
BENCH
INSTALLATION,
2-12.
The
Model
333A/334A
is
shipped
with
plastic
feet
and
tilt
stand
in
place,
ready
for
use
as
a
bench
instrument.
2-13.
RACK
INSTALLATION,
2-14.
The
Model
333A/334A
may
be
rack
mounted
by
using
the
5"
RackMount
Kit
(-hp-
Part
No.
5060-0775).
Instructions
for
the
conversion
are
included
with
the
kit.
The
rack
mount
for
the
Model
333A/334A
is
an
EIA
standard
width
of
19
inches.
When
mounted
in
à
rack
using
the
rack
mount
kit,
additional
support
at
the
rear
of
the
instrument
should
be
provided
if
vibra-
tion
or
similar
stress
is
likely.
2-15.
REPACKAGING
FOR
SHIPMENT.
2-16.
The
following
paragraphs
contain
a
general
guide
for
repackaging
of
the
instrument
for
shipment,
Refer
to
Paragraph
2-17
if
the
original
container
is
to
be
used;
2-18
if
it
is
not.
If
you
have
any
questions,
con-
tact
your
local
-
hp
-
Sales
and
Service
Office.
(See
Appendix
B
for
office
locations.)
NOTE
Ц
the
instrument
is
to
be
shipped
to
Hewlett-Packardfor
service
or
repair,
attachatag
to
the
instrument
identify
-
ing
the
owner
and
indicate
the
service
orrepairto
be
accomplished.
Include
the
model
number
and
full
serial
num-
ber
of
the
instrument.
In
any
corres-
pondence,
identify
the
instrument
by
model
number,
serial
number,
and
serial
number
prefix.
2-17.
If
original
container
is
to
be
used,
proceed
as
follows:
а,
Place
instrument
in
original
container
if
avail-
able.
Н
original
container
is
not
available,
a
suitable
one
can
be
purchased
from
your
nearest
-hp-
Sales
and
Service
Office.
b.
Ensurethat
container
is
well
sealed
with
strong
{аре
or
metal
bands.
2-18.
If
original
container
is
not
to
be
used,
proceed
as
follows:
a.
Wrap
instrument
in
heavy
paper
or
plastic
be-
fore
placing
in
an
inner
container.
b.
Place
packing
material
around
all
sides
of
in-
strument
and
protect
panelface
with
cardboard
strips.
c.
Place
instrument
and
inner
container
in
a
heavy
carton
or
wooden
box
and
seal
with
strong
tape
or
metal
bands.
d.
Mark
shipping
container
with
"DELICATE
INSTRUMENT,
"
"FRAGILE,
"
etc.
2-1

Section
Ш
пеене
RE
DETECTOR
MAX
INPUT
VOLTAGE
40
VOLT
P-P
AC
OR
40
VOLT
PEAK
TRARSENT
333A/2544-C-Q3854
j
334A
ONLY
LINE
switch
turns
instrument
ac
power
on.
Pilot
lamp
glows
when
instrument
is
turned
ON.
(2)Meter
indicates
distortion
or
voltage
level
of
input.
(3)
море
switch
selects
MANUAL
or
AUTOMATIC
bridge
tuning
operation.
(4)
FREQUENCY
RANGE
switch
selects
frequency
range
which
corresponds
to
fundamental
of
input
signal.
(5)COARSE
BALANCE
control
provides
coarse
ad-
justment
for
balancing
the
Wien
bridge
circuit.
(6)
FINE
BALANCE
control
provides
a
vernier
ad-
justment
for
balancing
the
Wien
bridge
circuit.
(1)
Frequency
dial
tunes
the
Wien
bridge
circuit
to
fundamental
of
input
signal.
(8)
HIGH
PASS
FILTER
switch
inserts
or
bypasses
HIGH PASS
FILTER
in
SET
LEVEL
and
DIS-
TORTION
function.
When
inserted,
filter
pro-
vides
>
40
dB
attenuation
to
50
-
60
Hz
hum
components
but
no
attenuation
to
frequencies
over
1
kHz.
OUTPUT
connectors
provide
means
of
monitor-
ing
output
of
meter
circuit.
(10)
Frequency
vernier
provides
fine
adjustment
of
frequency
dial.
METER
RANGE
switch
selects
full
scale
range
of
meter
in
percent,
dB
and
rms
volts.
Model
333A/334A
METERS
JANGE
547
LEVES
CAUTION!
(12)sENSITIVITY
selector
provides
0
-
50
dB
attenu-
ation
of
input
signal
in
10
dB
steps
in
SET
LEVEL
and
DISTORTION
functions.
(IS)SENSITIVITY
VERNIER
control
provides
fine
adjustment
of
sensitivity
setting.
Mechanical
zero
adjust
provides
mechanical
zero
adjustment
of
meter.
FUNCTION
switch
selects
type
of
operation
of
instrument.
(16)
shorting
bar
connects
circuit
ground
to
chassis
ground.
INPUT
terminals
provide
connections
for
input
signal.
(18)
NORM
RF
DET
(Model
334A
only)
selects
front
panel
INPUT
or
rear
panel
RF
INPUT
connector.
(9)
RF
INPUT
connector
(Model
334A
only)
provides
input
connection
for
AM
RF
carrier
input
signal.
FUSE
provides
current
overload
protection
for
instrument
circuits.
(21)
Line
voltage
switch
sets
instrument
to
operate
from
100
V/120
V/220
V/240
V.
(22)
Ac
power
connector
provides
input
connection
for
ac
power.
Figure
3-1.
Front
and
Rear
Panel
Description
3-0

Model
333A/334A
Section
ПІ
SECTION
ІП
OPERATING
3-1.
INTRODUCTION.
3-2.
The
Models
333A
and
334A
Distortion
Analyzers
measure
total
harmonic
distortion
of
fundamental
fre-
quencies
from
5
Hz
to
600
kHz;
harmonics
upto
3
MHz
are
included.
The
sharp
elimination
characteristics
(>80
dB),
the
low
level
of
instrument
induced
distor-
tion,
and
the
meter
accuracy
of
these
instruments
re-
sult
in
highly
accurate
measurement
of
low
level
har-
monic
distortion,
3-3.
Anrms
calibrated
voltmeter
is
inherent
in
the
333A
and
334A.
The
voltmeter
provides
a
full
scale
sensitivity
of
300
и
volts
rms
(residual
noise
«254
volts).
The
voltmeter
frequency
range
is
from
5
Hz
to
3
MHz
except
on
the
0,
0003
volt
range,
which
is
from
20
Hz
to
500
kHz.
3-4.
CONTROLS
AND
INDICATORS.
.
8-5.
Figure
3-1
illustrates
and
describes
the
function
'
of
all
front
and
rear
panel
controls,
connectors,
and
indicators,
The
description
of
each
component
is
keyed
to
the
drawing
included
within
the
figure.
3-6.GENERAL
OPERATING
INFORMATION.
3-7,
INPUT
CONNECTIONS.
3-8.
The
input
signal
can
be
connected
to
the
333A
and
334A
through
twisted
pair
leads
or
a
shielded
cable
with
banana
plug
connectors.
Keep
all
test
leads
as
short
as
possible
to
avoid
extraneous
pickup
from
stray
ac
fields.
When
measuring
Low-level
signals,
ground
loops
may
occur
causing
erroneous
readings.
Ground
loops
may
be
avoided
by
connecting
the
333A/
334A
Distortion
Analyzer
to
an
appropriate
isolation
transformer
to
break
the
chassis
ground
from
power
supply
ground.
Connect
all
other
instruments
to
one
power
strip
with
the
three-prong
connectors
as
close
as
possible.
3-9,
VOLTMETER
CHARACTERISTICS,
3-10.
The
RMS
VOLTS
markings
on
the
meter
face
are
basedonthe
ratio
betweenthe
average
and
effective
(rms)
values
of
a
pure
sine
wave.
The
ratio
of
average
to
effective
values
in
a
true
sine
wave
is
approximately
0.9to
1.
Whenthe
meter
is
used
to
measure
complex
waves,
the
voltage
indicated
may
not
be
the
rms
value
of
the
signal
applied.
This
deviation
of
meter
indica-
tion
exists
because
the
ratios
of
average
to
effective
values
are
usually
not
the
same
in
a
complex
wave
as
іп
а.віпе
wave.
The
amount
of
deviation
depends
on
the
magnitude
and
phase
relation
betweenthe
harmonics
and
fundamental
frequency
of
the
signal
applied.
Table
3-1
lists
the
deviation
of
the
meter
indication
of
a
sine
wave
partly
distorted
by
harmonics.
Ав
in-
dicated
in
the
table,
harmonic
content
of
less
than
10%
results
in
very
small
errors.
INSTRUCTIONS
Table
3-1
Effect
of
Harmonics
on
Voltage
Measurements
True
Meter
RMS
Value
Indication
Fundamental
=
100 100
Fundamental
+
10%
100
2nd
harmonic
Fundamental
+
20%
2nd
harmonic
Fundamental
+
50%
2nd
harmonic
Fundamental
+
10%
8rd
harmonic
Input
Voltage
Characteristics
100
-
102
100
-
110
96
-
104
Fundamental
+
20%
3rd
harmonic
94
-
108
Fundamental
+
50%
3rd
harmonic
90
-
116
l
NOTE
This
chart
is
universal
in
application
since
these
errors
are
inherent
in
all
average-responding
voltage-measuring
instruments.
3-11.
When
making
distortion
measurements
where
the
fundamental
frequency
is
suppressed
and
the
re-
mainder
of
the
signal
is
measured,
the
reading
ob-
tained
on
an
average
responding
meter
may
deviate
irom
the
true
total
rms
value.
When
the
residual
wave
contains
many
inharmonically
related
sinusoids,
the
maximum
error
in
the
distortion
reading
is
about
11
%
(11
%
of
the
measured
distortion)
low
for
dis-
tortion
levels
below
10
%.
EXAMPLE:
Measured
Maximum
Error
Total
Distortion
In
Meter
Indication
Distortion
2.5%
+0.
11x0.
025=
0.025+0,
0027=
0.00027
0.
0277
or
2.
8%
This
example
representsthe
maximum
possible
error,
and
in
most
cases
the
error
is
less.
In
distortion
measurements,
the
reading
of
an
average-responding
meter
is
sufficiently
close
to
the
rms
value
to
be
satisfactory
for
most
applications.
3-12.
OUTPUT
TERMINALS.
3-13,
The
OUTPUT
terminals
provide
a
0.1
V
rms
open
circuit
output
for
full
scale
meter
deflection.
These
terminals
can
be
used
to
monitor
the
output
signal
with
an
oscilloscope,
a
true
rms
voltmeter,
or
a
wave
analyzer.
The
combination
of
the
distortion
meter
and
oscilloscope
provides
more
significant
in-
3-1

Section
Ш
formation
about
the
device
under
test
than
the
expression
of
distortion
magnitude
alone.
Information
obtained
from
the
oscilloscope
pattern
is
specific
and
reveals
the
nature
of
distortion
which
sometimes
occurs
at
such
low
levels
that
it
is
difficult
to
detect
in
the
presence
of
hum
and
noise.
The
impedance
at
the
OUTPUT
terminals
is
2000
ohms,
therefore,
ca-
pacitive
loads
greater
than
50
pF
should
be
avoided
to
maintain
the
accuracy
of
meter
readings.
3-14.
OPERATING
PROCEDURES.
3-15.
INSTRUMENT
TURN-ON.
a.
Setthe
115-230
VAC
switch
to
coincide
with
the
line
voltage
in
use.
Turn
switch
to
ON
position.
Pilot
lamp
willglow,
indicating
application
of
primary
power.
3-16.
ADJUSTMENT
OF
METER
MECHANICAL
ZERO.
3-17.
The
meter
is
properly
zero-set
when
the
pointer
rests
over
the
zero
calibration
mark
and
the
instrument
is
in
its
normal
operating
enivronment
and
is
turned
off.
Zero-set
the
meter
as
follows
to
obtain
maximum
accuracy
and
mechanical
stability:
a,
Turn
instrument
on
and
allow
it
to
operate
for
at
least
20
minutes,
to
let
meter
move-
ment
reach
normal
operating
temperature.
b.
Turn
instrument
off
and
allow
30
seconds
for
all
capacitors
to
discharge.
c.
Rotate
zero
adjustment
screw
clockwise
un-
ti!
pointer
isleft
of
Zero
and
moving
upscale.
d.
Continue
rotating
screw
clockwise;
stop
when
pointer
is
exactly
at
zero.
e,
When
pointer
is
exactly
over
zero,
rotate
adjustment
screw
slightly
counterclockwise
to
relieve
tension
on
pointer
suspension.
1f
pointer
moves
offzero,
repeat
steps
c
through
е,
but
make
counterclockwise
rotation
less.
3-18.
DISTORTION
MEASUREMENT.
CAUTION?
DO
NOT
EXCEED
THE
INPUT
VOLT-
AGES
LISTED
BELOW
TO
PREVENT
DAMAGING
COMPONENTS
ON
À2
BOARD.
VOLTMETER
FUNCTION
-1V
RANGE
AND
BELOW,
AND
-DISTORTION
ANALYZER
575
FUNCTION
-
MAXIMUM
SENSITIVITY.
1.
800
V
ABOVE
100
Hz
2.
50
V
ABOVE
1
kHz
3-19.
MANUAL
MODE.
a.
Turn
instrument
on
and
mechanically
zero
meter
according
to
procedure
in
Paragraphs
3-15
and
3-16,
3-2
Model
333A/334A
Set
NORM-R.
F.
DET.
switch
to
NORM.
Set
FUNCTION
switch
to
ЗЕТ
LEVEL.
Set
MODE
switch
to
MANUAL,
If
fundamental
frequency
is
1kHz
or
greater,
set
HIGH
PASS
FILTER
switch
to
IN,
Set
SENSITIVITY
selector
to
MIN.
position,
and
rotate
VERNIER
control
maximum
counterclockwise.
NOTE
The
bandwidth
of
the
SENSI-
TIVITY
selector
is
reduced
in
the
two
extreme
CCW
positions
(positions
used
with
an
input
signal
greater
than
30
V).
Set
METER
RANGE
switch
tó
SET
LEVEL,
and
set
BALANCE
COARSE
and
FINE
con-
trols
to
center
position.
Connect
signal
to
be
measuredto
333A/334A
INPUT
terminals.
WARNING
В
REMOVE
SHORTING
STRAP
BETWEEN
FRAME
GROUND
(4)
AND
CHASSIS
GROUND
(
у
)
TERMINALS
ON
FRONT
PANEL
INPUT
TERMINALS
WHEN
MEASURING
DISTORTION
BETWEEN
TWO
POINTS
WHICH
ARE
DC
OF
FSET
FROM
GROUND
POTENTIAL.
Set
SENSITIVITY
selector
to
obtain
meter
indication
greater
than
1/3
full
scale.
Adjust
SENSITIVITY
VERNIER
for
full
scale
meter
indication
if
making
distortion
measure-
ment
in
percent;
if
making
distortion
measure-
ment
in
dB
adjust
SENSITIVITY
VERNIER
for
0
dB
meter
indication.
NOTE
If
unable
to
adjust
for
full
scale
or
0dB
indication,
(which
indicates
input
signal
is
below
0.3
volts),
set
METER
RANGE
selector
down-
scale.
Use
this
new
setting
as
the
1000
ог
0dB
SET
LEVEL
position,
thus
making
the
next
range
3090
or
-10
dB,
etc.
Set
FREQUENCY
RANGE
switch
and
frequency
dial
to
fundamental
frequency
of
input
signal.
Set
FUNCTION
switch
to
DISTORTION,
Adjust
frequency
dial
vernier
and
BALANCE
COARSE
and
FINE
controls
for
minimum

Model
333A/334A
Section
III
n.
p.
meter
indication.
Set
METERRANGE
switch
3-21.
DISTORTION
MEASUREMENT
OF
АМ
RF
down-scale
as
necessary
to
keep
meter
in-
CARRIERS
(334A
only).
dication
on
scale.
"rV
Repeat
step
m
until
no
further
reduction
in
meter
indication
can
be
obtained.
DO
NOT
EXCEED
MAXIMUM
INPUT
я
)
it
З
4
Observe
distortion
either
in
percentage
or
VOLTAGES
LISTED
ON
REAR
PANEL.
dB,
as
indicated
by
meter
deflection
and
METER
RANGE
switch
setting.
For
example,
if
meter
indicates
0.4
and
METER
RANGE
a.
Turn
instrument
on
and
mechanically
zero
setting
is
1%,
distortion
measured
is
0.4%
meter
according
to
procedure
in
Paragraphs
of
fundamental.
Similarly,
if
meter
indi-
3-15
and
3-16.
cates
-6
dB
and
METER
RANGE
setting
is
-40
dB,
distortion
measured
is
-46
dB
from
b.
Set
NORM.-R.F.
DET.
switch
to
R.
F.
DET.
fundamental.
c.
Connect
input
signalto
R.
Е.
INPUT
terminal
on
rear
panel.
SUUS
d.
Refer
to
Paragraph
3-19
for
manual
distort-
In
MANUAL
mode
the
accuracy
ion
measurement;
refer
to
Paragraph
3-20
of
distortion
measurements
is
for
automatic
distortion
measurement.
affected
by
frequency
stability
of
the
input
signal,
An
inaccuracy
——
NOTE
in
distortion
indications
occurs
when
the
frequency
drift
of
the
If
no
meter
deflection
can
be
input
signal
exceeds
the
bandwidth
obtained
with
an
RF
input,
diode
ACRI
should
be
checked.
А
spare
diode
is
located
on
the
outside
of
the
A4
shield.
of
the
rejection
curve.
If
desired,
rms
voltage
of
input
signal
can
be
measured
by
setting
FUNCTION
switch
to
3-22.
VOLTAGE
MEASUREMENT.
VOLTMETER,
and
setting
METER
RANGE
switch
to
obtain
an
on-scale
indication.
a.
Turn
instrument
on
and
mechanically
zero
meteraccordingto
procedure
in
Paragraphs
3-15
and
3-16,
3-20.
AUTOMATIC
MODE.
Ce
d.
|
Et
(
Perform
steps
athroughlof
Paragraph
3-19.
b.
Set
NORM.-R.
F.
DET.
switch
to
NORM,
i
та
Adjust
frequency
dial
vernier
and
BALANCE
c.
Set
FUNCTION
switch
to
VOLTMETER.
COARSE
and
FINE
controls
for
minimum
d.
Set
METER
RANGE
switch
to
a
range
meter
indication.
exceeding
amplitude
of
signal
to
be
mea-
sured.
When
meter
indication
is
less
than
10%
of
SET
LEVEL
indication,
set
MODE
switch
to
е,
Connect
signal
to
be
measured
to
INPUT
AUTOMATIC.
(Н
fundamental
cannot
be
terminals.
manually
nulled
below
10%
of
SET
LEVEL
indication,
automatic
mode
cannot
be
used).
f.
Set
METER
RANGE
switch
to
give
a
reading
as
close
to
full
scale
as
possible,
and
00-
Set
METER
RANGE
switch
down-scale
to
serve
meter
indication.
obtain
on-scale
meter
indication.
g.
The
dB
scale
of
the
333A/334A
is
calibrated
Observe
distortion
either
in
percentage
or
in
dBm,
such
that
0
dBm
=
1
milliwatt
dissi-
dB,
as
indicated
by
meter
deflection
and
pated
by
600
ohms.
Therefore,
a
dBm
mea-
METER
RANGE
switch
setting.
For
example,
surement
must
be
made
across
600
ohms.
if
meter
indicates
0.4
and
METER
RANGE
However,
dB
measurements
across
other
setting
is
10,
distortion
measured
is
0.
4%
impedances
can
be
converted
to
dBm
by
use
of
fundamental.
Similarly,
if
meter
indicates
of
the
Impedance
Correction
Graph
of
Fig-
-6
dB
and
METER
RANGE
setting
is
-40
dB,
ure
3-3.
For
example:
to
convert
a
-30
dB
distortion
measured
is
-46
dB
from
funda-
reading
across
200
ohms
to
dBm,
locate
the
mental.
200
ohm
impedance
line
at
the
bottom
of
the
graph.
Follow
the
impedance
line
to
the
If
desired,
rms
voltage
of
input
signal
can
be
heavy
black
line,
and
read
the
meter
correc-
measured
by
setting
FUNCTION
switch
to
tion
at
that
point.
The
correction
for
200
VOLTMETER,
and
setting
METER
RANGE
ohms
is
+5
dBm;
thus
the
corrected
reading
switch
to
obtain
an
on-scale
indication.
is
-25
dBm.
3-3

Section
Ш
3-23,
METER
INDICATION.
3-24.
The
333A/334A
meter
is
calibrated
to
indicate
in
both
dB
and
volts.
It
is
interesting
to
notethat
the
METER
RANGE
markings
differ
from
most
ac
volt-
meter
range
markings.
On
most
ac
voltmeters
(600
ohms)
0
dB
corresponds
to
the
1
volt
range.
This
is
not
true
inthe
case
of
the
333A/334A.
Since
the
instru-
ment
is
primarily
a
distortion
analyzer,
measurements
are
in
dB
(relative
measurement)
rather
than
in
dBm
(absolute
measurement).
Zero
dB
оп
the
333A/334A
corresponds
to
0.3
volt
range
rather
than
the
1
volt
range.
This
allows
210
dB
greater
dynamic
range
of
distortion
measurements.
3-25.
If
measurements
are
to
be
made
in
dBm,
10
dB
must
be
subtracted
from
the
METER
RANGE
setting.
Thus
0
dB
becomes
the
-10
dBm
range
for
absolute
power
measurements.
Zero
dBm
is
equal
to
1
milli-
watt
dissipated
by
any
impedance
andinthis
particular
case
is
600
ohms.
The
«10
DECIBELS
marking
on
the
meter
face
indicates
that
when
voltmeter
measurements
are
being
made,
the
indication
(METER
RANGE
plus
meter
indication)
is
10
dB
greater
than
when
power
(dBm)
measurements
are
being
made.
3-26.
In
short,
when
distortion
and
voltage
measure-
ments
are
being
made,
utilizethe
instrument
METER
RANGE
and
meter
scale
as
they
exist.
For
absolute
power
measurements
in
dBm,
simply
subtract
10
dB
from
the
METER
RANGE
setting.
3-27.
USE
OF
OUTPUT
TERMINALS.
3-28.
In
VOLTMETER
and
SET
LEVEL
functions,
the
333A/334A
can
be
used
as
a
low
distortion,
wide-band
amplifier.
А
portion
of
the
meter
input
(0.1
V
rms
open
circuit
forfullscale
meter
deflection
is
provided
at
the
OUTPUT
terminals.
3-29.
In
DISTORTION
function,
the
distortion
(0.1
V
rms
open
circuit
forfull
scale
deflection)
is
provided
atthe
OUTPUT
terminals
for
monitoring
purposes.
NOTE
The
INPUT
Y
terminal
and
the
OUTPUT
d
terminal
shouldnot
be
connected
directly
together
when
making
low
level
measurements.
These
terminals
are
isolated
from
each
other
by
1
ohm
which
reduces
the
effects
of
common
mode
voltages.
Model
333A
/334А
3-30.
333A/334A
WITH
OPTION
01.
3-31.
Operating
procedures
for
the
333A/334A
with
OptionOlarethe
same
asforthe
standard
instrument.
The
only
difference
between
the
standard
and
optional
instrument
is
that
the
Option
01
has
a
special
meter
and
meter
amplifier
whichis
compensated
to
respond
to
VU
(volume
unit)
characteristics.
|
3-32.
MANUAL
NULLING.
3-33.
Since
the
frequency
and
balance
controls
are
rather
sensitive
in
the
MANUAL
mode,
the
following
information
is
suppliedto
simplify
nulling
the
333A/334A
іп
the
MANUAL
mode,
When
nulling
the
3334/3344
in
the
MANUAL
mode,
connect
the
equip-
ment
as
shown
below
and
adjust
the
333A/334A
fre-
quency
and
balance
controls
for
the
waveform
shown
in
step
a
below,
Additional
waveforms
are
provided
to
simplify
nulling.
а.
www
No
harmonic
distortion,
Frequency
and
balance
adjustment
correct,
b.
7
Frequency
and
balance
control
im-
properly
adjusted,
с.
Frequency
approximately
correct;
balance
incorrect.
d.
CLLD
Balance
approximately
correct;
fre-
quency
incorrect,
e.
COX
2
Second
harmonie
predominant;
fre-
quency
and
balance
adjusted,
f.
--77---.
Second
harmonie
predominant;
fre-
quency
and
balance
adjusted;
phase
changed.
ge
7“-----”
Second
harmonic
predominant;
fre-
quency
and
balance
adjusted;
phase
changed,
һ
CXX
Third
harmonic
predominant.
i.
n
2
pee
incorrect;
meter
reading
off
Scale.
h
P
AE.
Frequency
incorrect;
meter
reading
off
scale.

Model
333А/334А
|
TEST
OSCILLATOR
BISA
луд
реак
METER
CORRECTION
-
DBM
Se
Se
2н
28
сола
FM
vem
PPAR
ИН ae
a
RICOTTA
NLL
mnm
ttbi
KG
rA
Figure
8-3.
OSCILLOSCOPE
DISTORTION.
ANALYZER
hp
3338A/3344
Lir
ру
MIN
Saat
та
Аде
и
НЕН
E
manent:
къв:
EHI
m
de
ҚЫЗЫ
рат
Impedance
Corréction
Graph.
Breseseacee
ЕН
ІШІ
Section
TIT
VES

Model
333A/334A
Section
IV
UIBISCIC
YOO
'I-p
елазъа
mee
tte
HA
REEO-9-YHES/
TEES
(339)
Q3Xxia
TVnNVW
130
|
А
УзлічО
T
|
aSvHd
Геле)
es
dWvi
са
|
|
Be
ER
39V
LIOA
EAN
Q3xi3
Nas
2
Tw
`
|
---
i
70
Гзомулув
|
(
-
Ж
`
S
[3vnNvw]
44
\
N
чалтна
O
SNINAL
C
РЭШЈО
+-
E.
85
Гепи
cL
a
„могу
-еногу
H3l
JI
dy
602v
-Sb2v
3901/48
|
U3IJITdWV
3Hd
ипочо
YILIN
|
|
i
HESS-
>
>
=
z
UOLT
Liv
32v80334
(-)
|
1SOd
EN
|
etter
strani
/-----------------р----------
назпайу
NOLLO3P^3H
1
НАМНЗА
ALIALLISN]S
9425
-|
наз
но
упма4
19
WHO
93W:
+
hy
ALIALLISNSS
{AINO
Vee")
139
NV
©
NOLLYOLSIC
O
J33A31
138
eiues-ziues
POZY-
OZY
OO
Y3LHJANOD
SUI
ДА
става
Е
«и
IS
O/d
зомуазеии
:
аму
Ei
YSLAWLIOA
HIS
O/d
48
[ANON]
|
ricco
Yen
gh
E
Ó
£s
Сі
[Nor1oNna]
39NVH
FILIN
4-0

Model
333A/334A
SECTION
THEORY
OF
4-1.
OVERALL
DESCRIPTION.
4-2.
Models
333A
and
334A
Distortion
Analyzers
in-
clude
an
impedance
converter,
a
rejection
amplifier,
a
metering
circuit,
and
a
power
supply.
The
Model
334A
also
contains
an
AM
detector,
The
impedance
converter
provides
a
low
noise
input
circuit
with
a
high
input
impedance
independent
of
source
impedance
placed
at
the
INPUT
terminals.
The
rejection
ampli-
fier
rejects
the
fundamental
frequency
of
the
input
signal
and
passes
the
remaining
frequency
components
on
to
the
metering
circuit
for
measuring
distortion.
The
metering
circuit
provides
visual
indications
of
distortion
and
voltage
levels
on
the
front
panel
meter,
M1.
The
AM
detector
(Model
334A
only)
detects
the
modulating
signal
from
the
RF
carrier
and
filters
all
RF
components
from
the
modulating
signal
before
it
is
applied
to
the
impedance
converter
circuit.
4-3.
BLOCK
DIAGRAM
DESCRIPTION.
(Refer
to
Figure
4-1)
4-4.
DISTORTION
MEASURING
OPERATION.
4-5.
For
distortion
measurement,
the
input
signalis
applied
to
the
impedance
converter
(Assembly
А2)
through
the
FUNCTION
selector,
S1,
and
the
one
megohm
attenuator.
The
one
megohm
attenuator
is
a
voltage
divider
network
which
provides
50
dB
atten-
uation
in
10
dB
steps.
The
desired
level
of
attenuation
is
selected
by
the
SENSITIVITY
selector,
62.
The
impedance
converter
provides
impedance
matching
and
unity
gain
between
the
INPUT
terminals
and
the
input
of
the
rejection
amplifier.
4-6.
Therejectionamplifier
consists
of
a
preamplifier,
a
Wien
bridge,
andabridge
amplifier,
The
SENSITIVITY
VERNIER
control,
at
the
input
of
the
preamplifier,
provides
a
set
level
signal
to
obtain
a
full
scale
read-
ing
on
the
meter
for
any
voltage
level
at
the
input
of
the
instrument.
With
the
FUNCTION
switch
inthe
SET
LEVEL
position,
a
ground
is
applied
to
the
Wien
bridge
circuit
to
allow
a
signal
reference
level
to
be
set
up
on
the
meter.
With
the
FUNCTION
switch
in
the
DIS-
TORTION
position,
the
Wien
bridge
is
used
as
an
in-
terstage
coupling
network
between
the
preamplifier
and
bridge
amplifier.
The
Wien
bridge
is
then
tuned
and
balanced
to
reject
the
fundamental
frequency
of
the
applied
input
signal,
Two
automatic
control
loops
con-
sisting
of
two
phase
detectors,
lamp
drivers,
lamps,
and
photocells
provide
fine
tuning
and
balance
in
the
AUTOMATIC
MODE.
The
remaining
frequency
com-
ponents
are
applied
to
the
bridge
amplifier
and
are
measured
as
distortion
by
the
metering
circuit.
Nega-
tive
feedback
from
the
bridge
amplifier
to
the
preamp-
lifier
narrows
the
rejection
response
of
the
Wien
bridge.
4-7,
The
output
of
the
rejection
amplifier
is
applied
to
the
metering
circuit
through
the
post-attenuator.
The
post-attenuator
is
used
to
limit
the
input
signal
Section
IV
iV
OPERATION
level
applied
to
the
metering
circuit
to
1
mV
for
full
Scale
deflection.
Sensitivity
of
the
metering
circuit
is
increased
to
300
uV
for
full
scale
deflection
onthe
.0003V
range.
The
metering
circuit
provides
a
visual
indication
of
the
distortion
level
of
the
input
signal.
In
addition
tothe
indication
provided
by
the
meter,
the
OUTPUT
terminals
provide
a
means
of
monitoring
the
distortion
components.
4-8.
DISTORTION
MEASUREMENT
IN
AM
CARRIERS.
По
4-9,
The
Model
334A
Distortion
Analyzer
contains
an
AM
detector
circuitfor
measuring
envelope
distortion
in
AM
carriers,
The
input
signal
is
applied
to
the
in-
put
of
the
AM
detector
circuit
where
the
modulating
signal
is
recovered
from
the
RF
carrier.
The
signal
is
then
applied
to
the
impedance
converter
circuit
through
the
one
megohm
attenuator
and
then
through
the
same
circuits
previously
described
in
the
distor-
tion
measuring
mode
operation.
4-10.
VOLTMETER
OPERATION.
4-11.
Inthe
voltmeter
mode
of
operation,
the
input
signal
is
applied
to
the
impedance
converter
circuit
through
the
1:1
and
1000:1
attenuator.
The
1:1
atten-
uation
ratio
is
used
onthe
0,
0003
100.
3
VOLTS
ranges,
and
the
1000:1
attenuation
ratio
is
used
in
the
1
to
300
VOLTS
ranges.
With
the
FUNCTION
switch
in
the
VOLTMETER
position,
the
output
of
the
impedance
converter
bypasses
the
rejection
amplifier
and
is
applied
to
the
metering
circuit
through
the
post-
attenuator
(METER
RANGE
switch).
Metering
circuit
sensitivity
is
increased
from
1
mV
for
full
scale
deflection
to
300
u
V
on
the
,
0003V
range,
as
it
was
in
the
distortion
measuring
operation.
The
function
of
the
post-attenuator
and
metering
circuit
is
the
same
for
voltmeter
operation
as
for
the
distortion
measuring
operation.
4-12.
DETAILED
CIRCUIT
DESCRIPTION.
4-13.
IMPEDANCE
CONVERTER
CIRCUIT.
(Refer
to
Figure
7-2)
4-14,
The
input
signal
is
applied
to
the
impedance
converter
circuit
through
the
1:1
and
1000:1
attenuator
S3R12
in
voltmeter
operation,
and
through
the
one
megohm
attenuator
S2R1
through
S2R6
in
distortion
operation.
Capacitive
dividers
S2C1
through
52С10
in
the
attenuator
keep
the
frequency
response
flat.
The
impedance
converter
is
a
low
distortion,
high
input
impedance
amplifier
circuit
with
gain
indepen-
dent
of
the
source
impedance
placed
at
the
INPUT
terminals,
4-15.
Instrument
induced
distortion
of
the
signal
being
measured
is
minimized
by
keeping
the
input
impedance
and
the
gain
of
the
impedance
converter
4-1

Section
IV
linear.
The
input
impedance
is
kept
linear
by
use
of
local
positive
feedback
from
the
source
of
A2Q1
to
the
gate
of
А201
and
to
the
protective
diodes
A2CR2
and
A2CR3.
Thus
signals
with
a
large
source
imped-
ance
can
be
measured
accurately.
Overall
induced
distortion
is
further
minimized
by
а
high
open
loop
gain
and
100%
negative
feedback.
The
high
open
loop
gain
is
achieved
by
local
positive
feedback
from
the
emitter
of
A2Q3
to
the
collector
of
A2Q2.
Overall
negative
feedback
from
the
emitter
circuit
of
A2Q4
to
the
source
of
A2Q1
results
in
unity
gain
from
the
impedance
converter.
4-16.
Thebias
points
of
thetransistors
in
the
imped-
ance
converter
are
selected
to
minimize
instrument
induced
distortion.
A2Q1,
an
extremely
low
noise,
high
impedance
field
effect
transistor,
is
the
major
component
that
makes
linearity
of
the
impedance
con-
verter
independent
of
the
signal
source
impedance.
4-17.
REJECTION
AMPLIFIER
CIRCUIT.
(Refer
to
Figures
7-3
and
7-5)
4-18.
The
rejection
amplifier
circuit
consists
of
the
preamplifier
(A3Q1
thru
A3Q3),
the
Wien
bridge
resistive
leg
and
auto
control
loop
(А501
thru
A5Q9
with
associated
lamp
and
photocell)
the
reactiveleg
and
auto
control
loop
(A5Q10
thru
A5Q18
with
associ-
ated
lamp
and
photocell),
and
the
bridge
amplifier
(A3Q4
thru
A3Q6).
4-19.
PREAMPLIFIER
CIRCUIT.
4-20.
The
signal
from
the
impedance
converter
is
applied
to
the
preamplifier,
which
is
used
during
SET
LEVEL
and
DISTORTION
measuring
operations,
Negative
feedback
from
the
junction
of
АЗВ10
and
АЗК1
is
applied
to
the
junction
of
A3R2
and
АЗС?
to
establish
the
operating
point
for
A3Ql.
Negative
feedback
from
the
emitter
of
A3Q3
is
applied
to
the
emitter
of
A3Q1
to
stabilize
the
preamplifier.
The
preamplifier,
like
the
impedance
converter,
is
designed
for
high
open
loop
gain
and
low
closed
loop
gain
to
minimize
instrument
induced
distortion,
4-21.
WIEN
BRIDGE
CIRCUIT.
4-22.
Inthe
distortion
measuring
operationthe
Wien
bridge
circuit
is
used
as
a
rejection
filter
for
the
fundamental
frequency
of
the
input
signal.
With
the
FUNCTION
selector,
S1,
inthe
DISTORTION
position,
the
Wien
bridge
is
connected
as
an
interstage
coupling
network
between
the
preamplifier
circuit
and
the
bridge
amplifier
circuit.
The
bridge
is
tuned
to
the
fundamental
frequency
of
the
input
signal
by
setting
the
FREQUENCY
RANGE
selector,
S4,
for
the
appli-
cable
frequency
range,
andtuning
the
capacitors
C4A
through
C4D.
The
bridge
circuit
is
balanced
by
ad-
justing
the
COARSE
balance
control,
R4,
andthe
FINE
balance
control,
R5.
In
the
AUTOMATIC
MODE
fine
tuning
and
balancing
are
accomplished
by
photoelectric
cells
which
are
in
the
resistive
and
reactive
legs
of
the
Wien
bridge.
The
error
signals
for
driving
the
photocells
are
derived
by
detecting
the
bridge
output
using
the
input
signal
as
a
reference.
4-23,
When
the
Wien
bridge
is
not
tuned
exactly
to
the
frequency
tobe
nulled,
a
portion
of
the
fundamental
4-2
Model
333A/334A
iNPUT
TO
BRIDGE
A
OUTPUT
OF
BRIDGE
iF
RESISTIVE
LEG
B
15
UNBALANCED
OUTPUT
OF
BRIDGE
IF
REACTIVE
LEG
С
15
DETUNED
ЗЗЗА-8-1464
Figure
4-2,
Bridge
Waveforms
frequency
will
appear
at
the
bridge
output.
The
phase
of
this
signal
depends
on
which
leg
of
the
bridge
is
not
tuned,
or on
the
relative
errors
in
tuning
if
neither
is
set
correctly.
The
magnitude
of
the
signal
is
propor-
tional
to
the
magnitude
of
the
tuning
error
of
either
or
both
legs
of
the
bridge.
l
4-24,
Figure
4-2a
shows
a
sinusoid
input
tothe
Wien
bridge.
If
the
resistive
leg
of
the
bridge
is
slightly
unbalanced,
the
output
of
the
bridge
is
very
small,
but
has
the
waveform
shown
in
Figure
4-2b
and
is in
phase
with
the
input.
As
the
resistive
leg
is
tuned,
the
signal
approaches
zero
amplitude
at
null
and
then
becomes
larger,
but
180°
out
of
phase,
if
the
null
position
is
passed,
Whenthe
resistive
leg
is
correctly
tuned
and
the
reactive
leg
is
tuned
through
null,
a
Similar
waveform
is
produced,
Figure
4-2c.
The
only
difference
is
that
the
reactive
signal
is
909
out
of
phase
with
the
resistive
signal.
4-25.
When
the
bridge
output
is
detected
using
the
input
signal
as
the
reference,
the
error
signals
in
phase
or
1809
out
of
phase
with
the
reference
develop
a
voltage
that
is
used
to
vary
the
resistance
in
the
resistive
leg
of
the
bridge,
to
tune
it
to
the
correct
null
position.
Signals
of
the
form
in
Figure
4-2c
do
not
develop
any
voltage,
as
the
resistive
detector
is
in-
sensitive
to
inputs
differing
from
the
reference
by
900.

Model
333A/334A
TYPICAL
WIEN
BRIDGE
bc
%
v
x
е
+
a
3
z
ш
be
E
ш
=
Е
=
p]
P
ГА
Section
ТУ
Шет
ТЕДІ
ОҒ
FEEDBACK
FREQUENCY
.
Figure
4-3,
Wien
Bridge
Circuit
and
Rejection
Characteristics
4-26.
In
an
independent,
but
similar,
control
loop,
the
bridge
input
signal
is
shifted
90°
and
used
as
the
reference
signal
for
the
detector.
This
detector
develops
control
voltages
to
null
the
reactive
leg
of
the
bridge,
but
is
insensitive
to
signals
of
the
form
in
Figure
4-2b,
which
are
caused
by
small
tuning
errors
of
the
resistive
branch.
4-27,
The
result
is
that
the
two
control
loops
derive
information
from
a
common
source
and
develop
two
independent
control
signals
for
nulling
the
two
legs
of
the
bridge.
These
control
voltages
are
used
to
vary
the
brilliance
of
lamps,
which
in
turn
cause
resistance
changes
in
photocells
that
form
part
of
the
Wien
bridge.
4-28.
When
the
bridge
circuit
is
tuned
and
balanced,
the
voltage
and
phase
of
the
fundamental,
which
appears
at
junction
of
the
series
reactive
leg
(S4R1,
3,
5,
7,
or
9
and
C4A/B)
and
the
shunt
reactive
leg
(SAR11,
13, 15,
17,
or
19,
and
C4C/D),
is
the
same
as
at
the
midpoint
of
the
resistive
leg
(A3R12
and
A3R14),
When
these
two
voltages
are
equal
and
in
phase,
the
fundamental
frequency
will
not
appear
at
the
drain
of
the
field
effect
transistor
A3Q4.
For
frequencies
other
than
the
fundamental,
the
reactive
|
leg
of
the
Wien
bridge
offers
various
degrees
of
attenuation
and
phase
shift
which
cause
a
voltage
at
the
output
points
of
the
bridge.
This
difference
volt-
agebetweenthe
reactive
leg
and
resistive
leg
is
amp-
lified
by
A3Q4,
A3Q5,
and
A3Q6.
Figure
4-3
illus-
trates
a
typical
Wien
bridge
circuit
and
the
rejection
characteristics
for
it.
4-29.
The
Wien
bridge
circuit
is
designed
to
cover
a
continuous
frequency
range
of
over
a
decade
for
each
position
of
the
FREQUENCY
RANGE
selector
54,
54
provides
coarse
tuning
of
the
reactive
leg
by
changing
the
bridge
circuit
constants
in
five
steps
at
1
decade
per
step.
For
the
automatic
control
loop,
the
refer-
ence
voltage
is
taken
from
R6
at
the
input
to
the
re-
jection
amplifier
and
applied
to
the
buffer
amplifier
А597.
The
reference
voltage
is
amplified
and
clipped
by
A5Q8
and
А509,
and
coupled
to
the
detector
A5Q4.
The
output
of
the
metering
circuit,
which
contains
the
fundamental
frequency
if
either
leg
of
the
bridge
is
untuned,
is
applied
to
the
buffer
amplifier
A5Q1.
It
is
amplified
by
A5Q2
and
A5Q3
and
coupled
to
the
detector
А504,
4-30.
RefertoFigure
4-4,
partial
schematic
for
detector
operation.
The
discussion
is
applicable
to
both
resistive
and
reactive
detector
circuits.
4-31.
The
signals
from
the
error
amplifier,
(A5Q2
and
A5Q3)
will
be
equal
and
of
opposite
phase,
and
will
cancel
out
each
other
when
the
detector,
A5Q4,
is
off.
However,
when
the
positive
half
of
the
reference
square
wave
gates
A5Q4
on,
the
signal
from
the
coll-
ector
of
A5Q3
will
be
shorted
to
ground.
Thus
the
signal
from
the
collector
of
A5Q2
will
be
coupled
through
the
filter
network
to
the
base
of
A5Q5.
If
the
signal
from
A5Q2
is
in
phase
with
the
reference,
the
positive
half
of
the
signal
will
be
passed,
and
if
it
is
out
of
phase,
the
negative
half
will
be
passed.
4-32,
The
normal
working
voltage
at
ASTP3
is
be-
tween
0
апа
-1volt,
The
de
output
of
the
filter
network
causes
the
voltage
at
ASTP3
to
go
in
a
positive
direc-
tion
(toward
zero)
for
in
phase
error
signals,
and
in
a
negative
direction
(toward
-1
V)for
out
of
phase
error
signals.
The
change
in
base
voltage
is
then
amplified
by
A5Q5
and
lamp
driver
А566.
This
will
change
the
brilliance
of
lamp
А6051,
which
will
vary
the
resist-
ance
of
A6V1
inthe
direction
necessary
to
balance
the
resistive
leg
of
the
bridge.
COLLECTOR
COLLECTOR
OF
Q3
OF
Q2
—
еле
с6
TP2
RIO
RII
COLLECTOR
CIRCUIT
Q9
3334/3344
-ROR
Auto
Control
Loop
Detector
Figure
4-4.
4-3

Section
IV
4-33,
Refer
to
Figure
4-5
for
the
phase
relationship
of
the
bridge
error
signal
andreference
voltage
at
the
base
of
А594.
The
shaded
portions
of
the
error
sig-
nals
(b
and
c)
indicate
that
part
of
the
error
signal
which
contributes
to
the
dc
lamp
control
voltage.
Аз
indicatedind,
any
error
signal
that
is
90°
out
of
phase
with
the
reference
does
not
affect
the
dc
lamp
control
voltage,
because
equal
amounts
of
the
positive
and
negative
portions
are
passed.
4-34.
The
operation
of
the
reactive
branch
control
loop
is
similar
to
that
of
the
resistive
branch,
The
Ref.
signalforResistive
Detector
(A5Q4)
Resistive
Error
signal
in
phase
with
reference
at
bridge
Resistive
Error
signal
180°
out
of
phase
Reactive
Error
signal
90°
out
of
phase
(Results
in
no
output
from
Re-
sistive
Detector)
Error
signal,
reactive
leg
detuned,
resistive
leg
unbalanced
333A/334A-RO
Model
333A/334A
phase
delay
circuit,
A5Q15,
A5Q16,
S4AF
and
S4C1
through
S4C5,
shifts
the
reference
voltage
909,
as
shown
in
Figure
4-5f.
This
makes
the
detector
A5Q12
sensitive
to
components
of
the
bridge
error
signal
that
are
909
out
of
phase
(g
and
h).
The
output
of
the
lamp
driver,
A5Q14,
controls
the
brilliance
of
A6DS2,
which
varies
the
resistance
of
A6V2
through
A6V5
to
tune
the
branches
of
the
reactive
leg.
Deck
AR
of
the
FREQUENCY
RANGE
switch,
S4,
switches
А5Ң56
in
parallel
with
A5R55
on
the
top
three
fre-
quency
ranges.
А6052
willbecome
brighter,
andlower
the
resistance
of
A6V2
through
A6V5,
making
variation
o
3
5°
360°
|
Lr
|
|
|
|
1
|
1809
|
Ref.
signal
for
Reactive
Detector
(A5Q12)
|
|
|
|
|
|
|
|
|
|
Reactive
Error
signal
in
phase
with
reference
at
bridge
Reactive
Error
signal
out
of
phase
Resistive
Error
signal
90°
out
of
phase
(Results
in
no
output
from
Re-
active
Detector)
Error
signal,
Reactive
leg
detuned,
Resistive
leg
unbalanced
Figure
4-5,
Reference
and
Error
Phase
Relationship

Model
333A/334A
BRIDGE
AMPLIFIER
А394
~
A3Q6
WEIN
BRIDGE
NEGATIVE
FEEDBACK
3334/334A-RO
Section
ТУ
5
50
500
5000
50K
600K
AMPLIFIER
GAIN
db
FREQUENCY
.
Figure
4-6.
Rejection
Amplifier
Block
Diagram
and
Typical
Frequency
Rejection
Characteristic
in
resistance
less
than
on
the
two
lower
ranges.
How-
ever,
less
variation
in
resistance
is
needed
to
tune
the
leg,
because
the
impedance
in
the
reactive
leg
becomes
progressively
less
as
the
higher
frequency
ranges
are
selected.
4-35.
Any
error
signal
that
is
not
an
integral
multiple
of
90°
is
the
result
of
the
reactive
leg
of
the
bridge
being
detuned,
andthe
resistive
leg
being
unbalanced.
For
example,
an
error
signal
that
is
45°
out
of
phase
(Figure
4-5e
and
j)
will
result
in
outputs
from
both
resistive
and
reactive
detectors
to
tune
the
bridge
and
reject
the
fundamental.
4-36,
When
the
FUNCTION
selector
is
set
to
the
VOLTMETER
or
SET
LEVEL
position,
the
junction
of
the
series
and
shunt
reactive
branches
ofthe
Wien
bridge
is
connected
to
circuit
groundthrough
A3R19
by
SIBF
which
disables
the
frequency
rejection
characteristic
of
the
bridge
circuit.
With
the
bridge
circuit
disabled,
the
rejection
amplifier
circuit
pro-
vides
one
dB
of
gain
for
the
fundamental
frequency
and
the
harmonics.
In
the
SET
LEVEL
operation,
this
signal
is
used
to
establish
the
SET
LEVEL
refer-
ence.
4-37.
BRIDGE
AMPLIFTER
CIRCUIT.
4-38,
The
bridge
amplifier
circuit
consists
of
three
stages
of
amplification,
A3Q4
through
A3Q6.
The
first
stage
of
amplification,
A3Q4,
is
a
field
effect
transistor
which
amplifies
the
difference
signal
between
the
gate
andthe
source.
The
field
effect
transistor
is
selected
for
minimum
noise
performance
with
the
high
impedances
of
the
Wien
bridge
circuit.
The
signal
from
the
drain
is
applied
to
the
two
stage
feedback
amplifier
A3Q5
and
A3Q6.
The
output
of
A3Q6
is
coupled
to
the
meter
circuit
by
the
post
attenuator
S3R1
through
S3R11.
Negative
feedback
from
the
out-
put
of
the
bridge
amplifier
is
applied
to
the
preamp-
lifier
circuit
to
narrow
the
frequency
rejection
characteristic.
It
can
be
noted
from
the
rejection
characteristic
(refer
to
Figure
4-3)
for
the
bridge
that
the
rejection
of
harmonic
voltages
is
not
con-
stant.
Typically
the
second
harmonic
is
attenuated
several
dB
more
than
the
third
harmonic
andthethird
more
than
the
fourth.
The
result
of
the
negative
feed-
back
is
illustrated
by
the
rejection
characteristic
Shown
in
dashed
lines
on
the
attenuation
and
phase
characteristic
of
Figure
4-3,
Figure
4-6
shows
a
simplified
block
diagram
of
the
rejection
amplifier
with
the
typical
frequency-rejection
characteristic.
Refer
to
Figure
4-7,
Bandwidth
Versus
Null
Depth
for
further
detail
on
the
rejection
characteristic.
4-39.
HIGH
PASS
FILTER.
(Refer
to
Figure
7-3).
4-40.
The
HIGH
PASS
FILTER
is
normally
used
when
the
fundamental
of
the
input
signal
is
greater
than
1
kHz.
In
the
voltmeter
mode
of
operation,
the
filter
is
not
used.
In
SET
LEVEL
and
DISTORTION
func-
tions
the
filter
presents
»
40
dB
attenuation
to
50
or
60
Hz
hum
components,
but
offers
no
attenuation
to
frequencies
over
1
kHz.
The
filter
assembly,
АТ,
consists
of
A7C1,
АТС2,
апа
А711,
The
filter
can
be
inserted
or
bypassed
by
the
HIGH
PASS
FILTER
switch,
S9.
4-41.
METER
AMPLIFIER.
(Refer
to
Figure
7-4)
4-42.
The
meter
amplifier
consists
of
the
post
atten-
uator,
the
meter
amplifier
circuit,
and
the
meter
rectifier
circuit.
4-43,
POST
ATTENUATOR.
4-44,
The
post
attenuator,
S3R1
through
S3R11,
is
a
resistive
network
which
attenuates
the
input
signal
in
10
dB
steps.
The
attenuator
is
usedin
conjunction
with
either
the
input
sensitivity
attenuator
or
the
1000:1
attenuator
to
limit
the
signal
level
to
the
meter
amplifier
to
1
mV
for
full
scale
deflection
on
all
ranges
from
1
mV
to
300
V
full
scale.
The
meter
circuit
sensitivity
is
increased
to
300
рҮ
for
full
scale
deflection
on
the
.0003V
range
by
switching
resistors
A2R29
and
A2R30
into
the
calibration
net-
work.
Resistor
A2R41
and
capacitor
A2C29
are
also
Switched
into
the
calibration
network
on
the
.0003V
range
to
extend
the
passband
of
the
amplifier.
4-45,
METER
AMPLIFIER
CIRCUIT.
4-46,
The
meter
amplifier
circuit
consists
of
a
five
stage
amplifier
circuit,
A2Q5
through
A2Q9,
which
develops
the
current
for
full
scale
meter
deflection.
Negative
dc
feedback
from
the
emitter
circuit
of
A2Q8
is
applied
to
the
base
of
А205
to
stabilize
the
dc
operating
point
of
the
meter
amplifier
circuit
and
to
minimize
the
tendency
for
de
drift
due
to
ambient
temperature
changes.
A2R51
and
A2CR8
are
electric-
ally
in
the
circuit
only
whenthe
meter
amplifier
is
over-
loaded.
When
the
voltage
on
the
emitter
of
A2Q9
4-5

Section
IV
Model
333A/334A
©,
00%
9095
eS]
EE
80%
poU
ay
70%
тине
ee
60%
A
SSE
исне
ee
LENSES
Dei
ay
DEDERE
TEE
iets
Gad
Gite!
IRR
50%
земен
касым
нөкер
касын
разни
—
=
ао
See
EE
ЕНИС
НОЕ
сарыны
40%
Z
SAO
EROT
TER
E
Ш
1]
eee
eee
eee
eee
30%
RENE атын
CINES
ШЕН
аи
e
NEM
EN]
ен
қаны
Бъз
NEN
o
RE
:
еті
RSEN
анты
е
шын
шаған
See
20%
pens
Е
DEN
E
eee
маса
[ENT
ENS
DO
оз
A
a
ИИИ
И
И
SE
SE
9%
8%
9,
6%
17%
SS
€:
e:
E
NN
П
Б
ыы
а
5%,
ИЕН
Әне
eS
ЗЕН
ЖЕН
ЕЛЕНЕ
DOE
SEND
қозды
Нас
AE
Қалта Бағын
Se
резени
erent
oo
EE
TEREA
TA
EEN
4%
ен
MASE,
EE
аара:
RR
SE
RIS
i
e
Te
RT
[ucro
E
o
d
qum
[D
pl
[E
eec]
39%,
Rss
HRS
ER
ЧЕДО
ЕРТЕ
ТЫМ
ҒА
БМИ
О
SEED
ТОЛСО
ПОГОРЕ
СОЕ
РАВНИ
[—
o
LN
ТЕ
БЕ
T
MM
ROME
Ld
2%,
pt
red
Д
кезен
Ее
Ұран
Бекет
жоға
алын
уры
ЧЕ
БН
1%
9%,
А
А
ВО
А
БА
Д-ға
8%
—
7
ST
CT
MESSEN
ENZEILUS
CANT
RT,
ESER
AT
CCA
NARS
LANL
CNA
ERT,
й
7%
Т
ОСО
aR
БЕЛШНЕ
ENEMY
ПН,
ВИС
СИ
САНО
Зан
АСИНО
БЦАА
PSN!
канға
6%
ЖОН
RR SR
ae
a
semet
СВД
кылы
акша
жөлмен
жолыма
a
INNEN
5%
SERE
КЕНЕН
ӘЛЕБЕН
ЕРИ
(ers
(ер;
aa
тақ
жесін
жадын
тарын
жн
le
.
oe
ЕНЕВ
EAA
енна
pee]
тегаш
пакта
4%
A
A
a
SE
ST
CAS
CT
URGE
ORCL
AACA
MEL!
КО
МВА
НЫС
к
EN
NN
1%
Ee
КЕННЕН
ЕЕЕ
3%
COGS)
ЕО
PE,
RM
ORME
[RS
БЕБЕ
"CREE
(as
өш
Pe
CS
жая
VE
ұз
Е
.2%
08%
09%
pepe
ud
eI
DEPT
ae
A
a
ji
07%
КЕНЕН
DE
кке
аншы
танады
роза
коне
SONSGEEE
БЕСЕЕЕБЕН
106%--
=e
eB
05%
П
БИ
Б
И
И
ee
ee
ee
ee
ee
eee
04%
LE
NY
TS
ey
Ree
ees
ears
SE
AT
Боқы».
“а
S
03%
Ae)
Sees
кке
Gee
Ge
наласы
ee
реке
dc
Ce
нее
эн
FRANK
es
жын
сакен
Бен
БЕЗІ
ИРЕ
Къси
ае
клей
0295
ТЕРЕК
ШЕН ТЕН БЕН
БЕЛЕК
ІРІЛІ
ЛЕН
БЕДЕ
ИЕ
ШЕН
eS
ee
раа
|
|
мш
Ep
iraq
БК
ШЕ
ol
vcr
ІН
ТС
Ot%
SO
SE
EE
Гета
С
Ае
EE
EE
SL
CE
OA
NT
ART
SARA
SUO
NENA!
008%
—
panman
ТЕЕ
Sissi
O
ae
a
ee
$
007%
EA
ea
ННН
TE
O
кене
eS
ae
eee
ee
a
Ses
“Ән
006%
—
BEEN
eee
КӨШІНЕН
ae
a
ОЗО екн
қанша
ee
ee
ЕНЕ»
Л
:
Б
НЕ
ЖЕРБЕН
ИНЕ
(атағы
кане
қағын
(SGA
а Е
ERGO
жалан
Ee
A
,
ee
doe
Slo
ee
ee
TL
CTE
Посао
ВАЗАРИ
ESET
004%
------------
eae
Deere
тылықты
BERE:
i
H
Aa.
ЕТЕ
BAINS
ЩЕ:
NEUEN
alqm
DRIN
qun:
mmi
SC
жене
жас
БАЕ
теке
Cpe
ERR!
l
Пані
ИДЕАЛ
a
ds
pou
qp
pec
esee
Lr
calescere
is
a=
-го
-30
-40
-50
-60
-TO
Af
DEPTH
OF
NULL,
DB
APPROXIMATE
PERCENT
BANDWIOTH-
p»
X
100
o
333A/
$334A-RO
о
1
о
Figure
4-7.
Bandwidth
Versus
Null
Depth
4-6

Model
333А/334А
becomes
abnormally
large
during
anoverload,
A2CR8
breaks
down
and
provides
a
lower
resistance
charg-
ing
path
for
А2С15,
which
reduces
the
transient
re-
covery
time
ofthe
meter
amplifier.
Negative
ac
feed-
back
is
applied
from
the
collector
circuit
of
A2Q9
to
the
emitter
circuit
of
A2Q5.
This
feedback
is
used
to
ensure
flat
frequency
response,
to
improve
linear-
ity,
and
to
reduce
the
effect
of
variationoftransistor
parameters
with
environmental
changes.
In
this
manner,
the
calibration
of
the
instrument
is
made
dependent
on
high
quality
passive
components.
4-47,
METER
RECTIFIER
CIRCUIT.
4-48.
The
meter
rectifier
is
connected
in
a
bridge
type
configuration
with
a
diode
in
each
upper
branch
and
a
dc
milliammeter
connected
across
the
midpoints
of
the
bridge.
The
simplified
meter
rectifier
is
illus-
trated
in
Figure
4-8,
The
generator
represented
by
A2Q5
through
A2Q9
with
the
internal
impedance
Ro
provides
the
meter,
M1,
with
current
for
full
scale
deflection
and
develops
a
voltage
across
the
calibra-
tion
network,
which
closes
the
ac
feedback
loop.
Capacitors
A2C27
and
A2C28
are
used
as
coupling
capacitors
for
the
ac
feedback
loop,
output
signal
to
the
OUTPUT
connector,
and
the
bridge
error
signal
to
the
input
of
the
automatic
fine
tuning
loops.
The
mechanical
inertia
of
the
meter
and
A2C26
prevents
the
meter
from
responding
to
individual
current
pulses,
Therefore,
the
meter
indication
corresponds
to
the
average
value
of
the
current
pulses
rather
than
the
peak
value.
The
meter
is
calibrated
to
indicate
the
rms
value
of
a
sine
wave.
Resistor
A2R45
im-
presses
a
fixed
bias
across
diodes
A2CR6
and
A2CR7
(biasing
them
close
to
the
barrier
voltage)
to
make
the
meter
amplifier
response
linear
to
large
variations
LonceoccuJ
A2R34
Figure
4-8.
Section
IV
in
signal
amplitude.
The
linearity
of
this
type
of
cir-
cuit
is
also
increased
by
including
the
meter
circuit
in
the
overall
feedback
loop.
4-49,
POWER
SUPPLY
CIRCUIT.
(Refer
to
Figure
7-6)
4-50.
The
power
supply
circuit
consists
of
a
+25
volt
series
regulated
supply
anda
-25
volt
series
regulated
supply
which
is
the
reference
supply
for the
+25
volt
supply.
4-51,
The
-25
volt
regulated
supply
is
ofthe
conven-
tional
seriesregulator
type.
The
amplifier
A1Q5
is
used
to
increase
the
loop gain
of
the
circuit,
thus
im-
proving
voltage
regulation.
The
positive
feedback
applied
to
the
junction
of
A1R11
and
A1R12
is
used
to
further
improve
the
line
frequency
suppression
of
the
circuit,
4-52,
The
+25
volt
regulated
supply
is of
the
conven-
tional
series
regulator
type
and
operates
the
same
as
the
-25
volt
regulated
supply.
4-53.
RF
DETECTOR
CIRCUIT.
(334A
only)
(Refer
to
Figure
7-2)
4-54,
The
RF
detector
circuit
consists
of
a
rectifier,
ААСВ1,
and
filter
circuit.
The
RF
signal
is
applied
to
the
circuit
through
the
RF
INPUT
connector
onthe
rear
panel.
The
rectifier
diode
A4CR1
recovers
the
modulating
signal
from
the
RF
carrier
and
the
fiiter
circuit
removes
any
RF
components
before
the
signal
is
applied
to
the
impedance
converter
circuit
through
the
NORM-RF
DET
switch,
87.
3334/3344-
AO
Simplified
Metering
Circuit
4-7
Other manuals for 333A
3
This manual suits for next models
1
Table of contents
Other HP Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Beta LaserMike
Beta LaserMike BenchMike Pro Addendum

Velleman
Velleman DCM100 user manual

Elster Instromet
Elster Instromet RABO G65 instruction manual

Landis+Gyr
Landis+Gyr Ultracold T550 operating instructions

London Electronics
London Electronics MAXI-INT2-L Mk.2 manual

MIB
MIB Flowmax 242i operating manual

BOS
BOS BS-DET2000BT manual

Transcell Technology
Transcell Technology TI-1680 manual

TSI Incorporated
TSI Incorporated DUSTTRAK DRX 8533 Operation and service manual

Endress+Hauser
Endress+Hauser Levelflex M FMP45 operating instructions

Topcon
Topcon BM-9A instruction manual

Phabrix
Phabrix PHQX-01E Operation manual