Human Design Medical Z1 CPAP System User manual

AN AFFILIATE OF BREAS MEDICAL AB
Human Design Medical
Z1®CPAP System
CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEM
User Guide
User Guide
English
AN AFFILIATE OF BREAS MEDICAL AB
Human Design Medical
Newton, MA 02458
Z1 User Guide v1 03012018
#40-0007 Rev. C
®

1
Contents
Welcome to Your Z1 CPAP System.................................................................2
Owner/User Responsibility......................................................................2
Healthcare Provider Responsibilities.............................................................2
Replacement Z1 Parts and Supplies .............................................................2
Terms Used in this User Guide...................................................................3
Medical Information ..............................................................................3
Z1 Warnings and Cautions ........................................................................4
Battery Warnings and Cautions....................................................................7
Getting to Know Your Z1 CPAP System ............................................................8
Package Contents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Z1 CPAP Device .................................................................................8
Power Supply (AC to DC Power Adapter) ........................................................8
Tube Adapter ...................................................................................8
CPAP Tube ......................................................................................9
End Cap ........................................................................................9
Filter ...........................................................................................9
USB Micro Cable ...............................................................................9
Heat Moisture Exchanger (HME).................................................................9
PowerShell (Optional Accessory) ................................................................9
Set Up Guide ....................................................................................10
Unpacking the Z1 Box .........................................................................10
Assembling and Powering Up..................................................................10
Using Your Z1 ...................................................................................13
Starting and Stopping Treatment ..............................................................14
Choosing Between Z-Breathe Settings .........................................................15
Programming Ramp Mode.....................................................................15
User Software..................................................................................16
Charging the Battery (PowerShell Only) ........................................................16
Care .............................................................................................17
Inspections ....................................................................................17
Cleaning ......................................................................................17
Reminder Alert (Air Leak Detection) ............................................................18
Fault Conditions ...............................................................................19
USB Firmware Upgrade ........................................................................19
Troubleshooting ...............................................................................19
Guidance and Manufacturer’s Declaration......................................................21
Technical Specifications........................................................................25
Servicing ........................................................................................27
Limited Warranty.................................................................................27
Contact Information .............................................................................28
FAA Compliance .................................................................................29
Product Registration .............................................................................31
Z1®, the HDM logo ( ), PowerShell®, NiteLog® and Z-Breathe® are registered trademarks of Human Design Medical.
All other trademarks are the property of their respective owners.
Patent Pending. © Copyright 2018, Human Design Medical. All rights reserved.

2 3
Welcome to Your Z1 CPAP System
Thank you for choosing the Z1 CPAP System from Human Design Medical– designed, engineered,
and assembled in the USA with you in mind.
The Z1 is specially designed with these features:
• Z-Breathe® Comfort Feature – Z-Breathe is designed to minimize the impact of spikes
and drops while keeping the pressure constant, thereby delivering more comfortable
therapy.
– the Z1 automatically reduces the air pressure in the mask to relieve pressure when
you exhale and to level out the pressure differences in your natural breathing pattern.
• Leak Compensation – the Z1 automatically compensates for minor air leakage
helping ensure the proper treatment pressure is maintained.
• Auto Altitude Adjustment – the Z1 adjusts to altitudes up to 8000 ft above sea level.
• Air Travel – The Z1 adjusts to an airplane’s cabin pressure and is approved for use on
airplanes.
• Upgradable Operating System– the Z1 firmware and software can be easily updated
via the USB Cable and your computer.
This User Guide will help you take advantage of the features and capabilities of this portable CPAP
device. Please read this User Guide carefully before using your Z1.
Owner/User Responsibility
The owner/user of the Z1 CPAP System is responsible for reading and understanding this User
Guide. The user is responsible for any injury or damage that results from:
• Operation of the Z1 other than in accordance with the operating instructions
contained in the User Guide supplied.
• Unauthorized maintenance or modifications to the device or attached accessories.
Healthcare Provider Responsibility
Your healthcare provider is responsible for explaining contraindications and instructions for use to
the patients/users.
Replacement Z1 Parts and Supplies
Use only genuine Z1 replacement filters, parts and supplies. Replacement filters, parts and
supplies can be purchased from retail medical supplier or at www.hdmusa.com. Use of
unauthorized filters, parts or replacement supplies can damage the device and void the warranty.
Terms Used in this User Guide
This User Guide contains special terms and icons that appear in the text to draw your attention to
specific and important information.
WARNING alerts you to possible injury or hazard.
CAUTION explains special measures for the safe and effective use of the
device.
NOTE is an informative or helpful note.
Medical Information
Indications for Use
Your Z1 CPAP System is a single user, reusable device that provides continuous positive airway
pressure (CPAP) to support treatment of adults weighing over 66 lbs (30 kg) with obstructive sleep
apnea (OSA).
Contraindications
Positive airway pressure therapy may be contraindicated in some individuals with the following
pre-existing conditions:
• Severe bullous lung disease
• Pneumothorax
• Pathologically low blood pressure
• Dehydration
• Cerebrospinal fluid leak, recent cranial surgery, or trauma.
Adverse Effects
Users should report unusual chest pain, severe headache, or increased breathlessness to
their prescribing physician. An acute upper respiratory tract infection may require temporary
discontinuation of treatment.
The following side effects may arise during the course of therapy with CPAP devices:
• Drying of the nose, mouth, or throat
• Eye irritation
• Nosebleed
• Bloating
• Skin rashes
• Ear or sinus discomfort.

4 5
Z1 Warnings and Cautions
WARNINGS:
• Read the entire User Guide before using your Z1. Consult your health care provider if
there is any part of this User Guide you do not understand.
• US Federal law restricts this device to sale by or on the order of a physician.
• Use your Z1 and its accessories only for their intended use as described in this User
Guide and as directed by your physician, clinician or health care provider.
• The Z1 is not approved for use with oxygen. Do not use supplemental oxygen or
any other type of supplementary gas with your Z1. Do not connect supplementary
oxygen or other gas to any accessory or patient interface which is connected to your
Z1.
• If you notice any unexplained changes in the performance of your Z1 or your Z1
is making unusual or harsh sounds, or if the Z1 or Power Supply are dropped or
mishandled, discontinue use and contact the Human Design Medical Service Center
at 1-855-436-8724.
• If your Z1, Power Supply, PowerShell or Battery is dropped or mishandled, or if water is
spilled into an enclosure, or if an enclosure is broken, discontinue use and contact the
Human Design Medical Service Center at 1-855-436-8724.
• A mask is not provided. Your Z1 should only be used with masks and connectors
recommended by your physician, clinician, respiratory therapist or health care
provider.
• Your Z1 is intended to be used with CPAP masks (or connectors) which have vent
holes to allow continuous flow of air out of the mask. When your Z1 is turned on
and functioning properly, new air from the Z1 flushes the exhaled air out through
the mask vent holes. However, when the Z1 is not operating properly (e.g., at low
pressure) or the Z1 is off, insufficient fresh air may be provided through the mask and
the exhaled air may be rebreathed. Rebreathing exhaled air for longer than several
minutes can, in some circumstances, lead to suffocation. This applies to most CPAP
devices. Your health care provider will recommend a mask and air tubing to suit
your needs. The Z1 supports several CPAP mask types including nasal masks, full face
masks, and nasal prongs that meet the ISO 17510-2 standard.
• Do not use your Z1 with a mask unless the Z1 is turned on. Once the mask is fitted,
confirm that air is flowing from your Z1.
• Do not block the vent hole or holes in the mask. Follow the mask manufacturer’s
instructions for proper use of the mask.
• In the event of a power failure or machine malfunction, remove the mask
immediately.
• At low pressures some exhaled gases may remain in the mask.
• Electrocution Hazard – Your Z1 is an electrical device, and as such, you should not
immerse your Z1, Power Supply, PowerShell or Battery in water. Always unplug your
Z1 and its accessories before cleaning (and clean only as instructed in this User Guide)
and make sure your Z1 and its accessories are dry before plugging back in.
• Before use, make sure the power cord and plug are in good condition and that the
cord is not damaged.
• Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturizing or
antibacterial soaps or scented oils to clean your Z1, Power Supply, PowerShell or
Battery. These solutions may create a shock hazard, cause damage and reduce the life
of these items.
• Explosion Hazard — Do not use or store your Z1, Power Supply, PowerShell or
Battery in the vicinity of flammable anesthetics or oxygen.
• Do not use your Z1 near toxic or harmful vapors.
• Do not drop your Z1 or any of its accessories or expose it to any crushing forces.
Should your Z1 be dropped and fail to operate properly, contact the Human Design
Medical Service Center at 1-855-436-8724 before using the Z1 for treatment.
• Do not insert any object into any tubing or opening of your Z1 except as described in
this User Guide.
• Do not operate your Z1 in any area which may block its air intake. Blocking the CPAP
Tube and/or air inlet of your Z1 while in operation could lead to overheating of the Z1.
• If you put the Z1 on the floor, make sure the area is free from dust and clear of
bedding, clothes or other objects that could block the air inlet or cover the Power
Supply unit.
• Only store or transport the Z1 and accessories within the storage and transport
temperature, humidity and altitude conditions specified in the Technical
Specifications section of this User Guide.
• Do not operate your Z1, Power Supply, PowerShell or Battery in an area which may
cause overheating.
• Do not cover the Power Supply, PowerShell or Battery with any object or material
which would cause them to overheat.
• Keep the power cord, Power Supply, Battery and Z1 away from hot or flammable
surfaces, gases or materials.
• Do not open your Z1 or tamper with its accessories. There are no user serviceable
parts inside. Repairs and servicing should only be performed by an authorized Human
Design Medical service agent.
• Only power the Z1 with the Human Design Medical Power Supply (AC-to-DC power
adapter) or the optional Human Design Medical PowerShell with Human Design
Medical Battery. Do not use any other AC adapters, batteries, or other power sources
to power the Z1. Do not use the Human Design Medical Power Supply, PowerShell or
Battery to power any device other than the Z1.
• Your Z1 should only be used with standard or slim style CPAP air tubing as
recommended by your physician, clinician or health care provider.

6 7
• Do not leave long lengths of air tubing around the top of your bed. The tube could
twist around your head or neck while you are sleeping.
• Do not connect or insert any device, cable or articles into the Micro USB or Micro SD
ports of your Z1 except as described in this User Guide. Connecting other devices
could result in injury to you or damage to the Z1 and void the warranty.
• Do not locate your Z1 in an unapproved place or where the air intake can be blocked
during treatment. This will lead to overheating of the Z1. If a blocked intake should
occur during operation, the unit may stop supplying pressurized air and cause a fault
condition. Refer to“Troubleshooting”in this User Guide to interpret this fault and
instructions for corrective action.
• The Z1 is designed to deliver pressures from 4 to 20 cm H2O. In the unlikely event of
certain fault conditions, pressures of up to 30 cm H2O are possible.
• Blocking the CPAP Tube while in operation could lead to overheating of the Z1.
• In the clinical environment, any personal computer that is used with your CPAP
system must be at least 5 feet (1.5 m) away from, or at least 8 feet (2.5 m) above, the
patient. It must also comply with the relevant test standard. For personal computers
the international standard is IEC 60950 or equivalent.
CAUTIONS:
• Be careful not to place your Z1 where it can be bumped or where someone is likely to
trip over the power cord.
• Make sure your Z1, Power Supply, PowerShell and Battery are kept clean, dry, and free
of dust.
• Always make sure the power cord and plugs are in good condition and the
equipment is not damaged before use.
• Make sure the area around your Z1 and accessories is dry and clean.
• During partial (below rated minimum voltage) or total power failure, therapy
pressures will not be delivered. When power is restored, your Z1 will enter standby
mode and all prior settings will be retained in memory.
• Follow the storage and cleaning instructions provided with your CPAP Tube and mask.
• The air filter should be changed at intervals specified in this User Guide.
Accumulation of filtered particles over time can reduce the efficiency and shorten the
life of any CPAP device, including the Z1. Replacement filters are available at your retail
medical supplier or at www.hdmusa.com.
• Do not wash the air filter. The air filter is not washable or reusable.
• The End Cap protects the Z1 in the event of a minor collision. Always ensure that the
End Cap is securely fitted in accordance with instructions at all times. The Z1 may not
operate if the End Cap is not properly installed.
• The temperature of the airflow for breathing produced by CPAP devices can be higher than
the temperature of the room. This is true of all CPAP devices. Caution should be exercised
when using your Z1 at room temperatures warmer than 90˚F (32˚C).
• Only operate the Z1 and accessories within the operating temperature, humidity and
altitude conditions specified in the Technical Specifications section of this User Guide.
• Only store or transport the Z1 and accessories within the storage and transport
temperature, humidity and altitude conditions specified in the Technical Specifications
section of this User Guide.
• Damage to the CPAP Tube may result in air leaks. If this occurs, stop using the damaged
CPAP Tube and replace immediately.
Battery Warnings and Cautions
The PowerShell comes with a removable Lithium Ion Battery. Careful attention must be paid to
the warnings and cautions described below relating to the Battery. Misusing the Battery may
cause the Battery to get hot, rupture, ignite and cause serious injury.
WARNINGS:
• Do not place the Battery in fire or heat the Battery.
• Do not install the Battery backwards so that the polarity is reversed.
• Do not connect the positive terminal and the negative terminal of the Battery to each other
with any metal object (such as wire).
• Do not carry or store the Battery together with necklaces, hairpins, or other metal objects.
• Do not pierce the Battery with nails, strike the Battery with a hammer, step on the Battery or
otherwise subject it to strong impacts or shocks.
• Do not solder onto the Battery.
• Do not expose the Battery to water or salt water, or allow the Battery to get wet.
• Do not disassemble or modify the Battery. The Battery contains safety and protection
devices which, if damaged, may cause the Battery to generate heat, rupture or ignite.
• Do not place the Battery on or near fires, stoves or other high-temperature locations. Do not
place the Battery in direct sunlight, or use or store the Battery inside cars in hot weather.
Doing so may cause the Battery to generate heat, rupture or ignite. Using the Battery in this
manner may also result in a loss of performance and a shortened life expectancy.
• Immediately discontinue use of the Battery if, while using, charging or storing the Battery,
the Battery emits an unusual smell, feels hot, changes color, changes shape or appears
abnormal in any other way. Contact the Human Design Medical Service Center at 1-855-
436-8724 if any of these problems are observed.
• Do not place the Battery in microwave ovens, high-pressure containers, or on induction
cookware.
• In the event that the Battery leaks and the fluid gets into one’s eye, do not rub the eye. Rinse
well with water and immediately seek medical care. If left untreated the battery fluid could
cause damage to the eye.
• Only charge the Battery with the supplied Human Design Medical PowerShell and the
supplied Human Design Medical Power Supply. Do not use any other removeable battery
charger, AC adapter, or other device to charge the Battery.

8 9
• Do not attach the Battery directly to a vehicle cigarette lighter or to the Power Supply plug.
• Do not continue charging the Battery if it does not recharge within the specified charging
time. Doing so may cause the Battery to become hot, rupture or ignite.
• Remove the Battery from the PowerShell if the Z1 is not likely to be used for some time. Do
not use the Battery to power any device other than the Z1.
• Only use or charge the Battery within the operating temperature, humidity and altitude
conditions specified in the Technical Specifications section of this User Guide.
• Only store or transport the Battery within the storage and transport temperature and
humidity conditions specified in the Technical Specifications section of this User Guide.
CAUTION:
• When the Battery is worn out, insulate the terminals with adhesive tape or similar
materials before disposal.
Getting to Know Your Z1 CPAP System
CPAP Tube
The Z1 System is provided with a 4 ft slim style CPAP Tube. The CPAP Tube connects the Z1 to your
mask. Always consult with your physician or health care provider regarding the selection of tubing
for your treatment. Follow the instructions provided with the tubing.
End Cap
The End Cap attaches to the air intake end of the Z1 and houses the Filter.
Filter
The Z1 uses a specially designed filter. The material has been specifically chosen for the Z1 and
has been designed to be much larger than other CPAP filters. The unique shape of the lter fits
into the End Cap of the Z1, which was designed to allow for more effective filter surface area.
For proper operation of the Z1, the Filter should be replaced monthly or more often if necessary.
USB Micro Cable
The USB Cable allows the Z1 to communicate with a computer to collect and view user data.
Heat Moisture Exchanger (HME)
The disposable HME works by capturing moisture as you exhale and provides it back to you
as you breathe. The HME does not use any electricity or complex moving parts and avoids the
problems encountered with electrical humidification systems.
To install the HME, connect the large diameter end of the HME to your mask and the smaller
diameter end to your CPAP Tube. Be sure both connections are fully seated and secure. Store the
HME in a sealed plastic bag between uses to extend its life. Discard when it is no longer providing
humidification.
PowerShell (Optional Accessory)
The PowerShell houses both the Z1 and the Battery which allows temporary usage of the Z1
without the Power Supply. It can be purchased at your retail medical supplier or at www.hdmusa.
com.
Z1 CPAP
Filter
Slim-style
CPAP tube
AC DC Adapter
Power Cord
Tube Adapter
USB Cable
HME
Package Contents
Z1 CPAP Base Model
Z1 CPAP Device
Power Supply & Cord
Custom USB A to Micro B Cable
Custom CPAP Tube Adapter
48” Slim Style CPAP Tube
End Cap
Filter (2- 1 installed in Z1, 1 in box)
Heat / Moisture Exchanger (HME)
Z1 Auto Getting Started Guide
Z1 Auto User Guide
PowerShell Module
Extended Life Battery
Battery Shuttle
PowerShell
PowerShell Getting Started Guide
Z1 CPAP Device
The Z1 supplies continuous positive airway pressure and can be programmed to fit the patient’s
prescriptive needs.
Power Supply (AC to DC Power Adapter)
The Power Supply provides power to the Z1. It takes AC power (voltage ranging from 100-240
volts) from a wall outlet and converts it to DC voltage to power the Z1. The Power Supply can be
used with international power sources provided the appropriate plug adapter is used.
Tube Adapter
The Tube Adapter attaches the CPAP Tube to the Z1.

10 11
Set Up Guide
Unpacking the Z1 Box
You will need to perform a few minor setup steps before using your new Z1. The instructions
below will guide you through the setup process. Some of the steps outlined below may have
been performed by your health care provider prior to delivering the Z1 to you. If you have any
questions or problems as you go through the setup process, please call Human Design Medical
Customer Service at 1-855-436-8724 for assistance. Human Design Medical wants your user
experience with the Z1 to be trouble and frustration free.
1. Carefully remove the contents of the box.
2. Some of the Z1 components have their own packaging material. Remove this material
carefully. Avoid using anything sharp when opening the packaging.
3. Carefully inspect all contents against the Package Contents list on page 8 to assure
all components are present. Keep in mind that your health care provider might have
completed some of the setup steps prior to delivery, so if a component appears to
be missing, check to see if it has already been connected. Also inspect all contents to
ensure there is no obvious or visible damage or defects. If you have any concerns or
questions, contact Human Design Medical Customer Service at 1-855-436-8724, or
your service or health care provider before use.
4. Place all the components in an area where you can easily access the Z1 for set up and
programming.
WARNING: Do not use this product if you have any concerns about the
condition of the contents or if there are any missing parts.
Z1 Base Model
Assembling and Powering Up
Tube Adapter, CPAP Tube, HME and Mask
1. Connect one end of the CPAP Tube to the round end of
the Tube Adapter then insert the other end of the Tube
Adapter into the outlet of the Z1. (Figure 2)
2. Connect the other end of the CPAP Tube to the HME
and the HME to the mask.
3. Ensure all connections are tight to avoid air leaks.
(Figure 3)
Power Supply
WARNING: The Z1 must only be used
with the Human Design Medical Power
Supply (or the optional Human Design
Medical PowerShell with Human Design
Medical Battery). DO NOT use any other
power source with the Z1.
CAUTION: Make sure the AC power cord
IS NOT plugged into an AC power outlet
before continuing to the next step.
1. Plug the AC power cord into the AC Adapter.
2. Plug the AC power cord into a wall outlet.
3. Plug the AC Adapter into the Z1. (Figure 4)
4. The Z1 will power up and cycle through a series of
startup displays.
5. After about 10 seconds, the Z1 will enter standby
mode. The display will read 0.0. (Figure 5)
NOTE: The Z1 display will be active
whenever the device is receiving power.
Figure 1
Figure 2
Filter and End Cap
Your Z1 comes with the first filter already installed. Replacement
filters may be installed as follows:
1. Remove the Filter out of the End Cap.
NOTE: To remove the End Cap, squeeze
firmly on the sides of the End Cap and pull
it away from the Z1.
2. Pull used filter out of the End Cap.
3. Remove new Filter from its packaging.
4. Insert the new Filter into the End Cap, aligning the
arched side of the Filter with the top of the End Cap.
Ensure the Filter covers all holes in the End Cap, so
that all air entering the Z1 passes through the Filter
and no air bypasses the Filter.
5. Press the End Cap onto the Z1. (Figure 1)
Figure 4
Figure 5
Figure 3
Z1 CPAP
Tube Adapter
Slim-style
CPAP tube
HME
Mask (not included)

12 13
PowerShell and Battery (Available Separately)
WARNING: The Z1 must only be used
with the optional Human Design Medical
PowerShell with Human Design Medical
Battery (or the Human Design Medical
Power Supply). DO NOT use any other
power source with the Z1.
1. Slide the Battery with the attached Shuttle into the
lower bay of the PowerShell until the Shuttle clicks into
place. (Figure 6)
NOTE: The Battery can be removed from
the PowerShell by pressing the two blue
buttons on the Battery Shuttle and sliding
the Battery out.
2. Slide the Z1 all the way into the upper bay of the
PowerShell until you hear a “click” sound, indicating the
Z1 has locked into place. (Figure 7)
CAUTION: Insert the Z1 carefully
ensuring the power barrel connector slides
into the power input on the Z1.
NOTE: The Z1 can be removed from
the PowerShell by pressing the two [ ]
symbols on the sides of the PowerShell and
sliding the Z1 out.
3. The Z1 will power up and cycle through a series of
startup displays.
4. After about 10 seconds, the Z1 will enter standby
mode. The display will read 0.0. (Figure 8)
NOTE: Remove the Battery from the PowerShell
when the Z1 is not in use, unless the Power Supply
is plugged into the PowerShell to charge the
Battery. Otherwise the Z1 and PowerShell will
drain the battery completely within 2-3 days.
Using Your Z1
LCD Display
Treatment Pressure
Display
Ramp Progress
Ramp Time
Remaining
AC Power Indicator
Battery Status Display (PowerShell Only)General Status Display
Fault
Reminder
Treatment Pressure Display
The current treatment pressure is shown on the LCD display. Once set, the prescribed treatment
pressure will be retained in the Z1’s memory even during loss of power. For example, a treatment
pressure setting of 10.0 cm H2O is displayed in Figure 9.
Ramp Progress & Ramp Time Remaining Indicators
Ramp Mode allows you to gradually acclimate to the prescribed treatment pressure. Ramp Mode
gradually increases the pressure from a low initial starting pressure to the prescribed treatment
pressure over a programmed interval of time. See “Programming Ramp Mode”.
If you program the Z1 for Ramp Mode, the Ramp Progress and Ramp Time Remaining indicators
will appear on the LCD display. For example, Figure 9 indicates that the ramp is just over half
complete, with a ramp time remaining of 10 minutes.
Battery Status Display
This icon is displayed when the optional PowerShell and Battery are connected. The level of
charge remaining on the Battery is indicated by the number of solid bars in the icon. The bars
flash when the Battery is charging. For example, the icon shown in Figure 9 indicates that the
Battery has between 35-63% of charge remaining. Make sure the Battery is fully charged before
each use. The bars will stop flashing when the charging is complete.
Number of Solid Bars Charge Remaining
1 10-35%
2 35-63%
3 63-90%
4 90-100%
AC Power Indicator
This icon is displayed when the Z1 is connected to an AC power outlet via the Power Supply.
Figure 9
Figure 6
Figure 7
PowerShell Extended
Life Battery
Figure 8

14 15
General Status Display
In Standby Mode, this display field alternates between a display of the current date, the current
function mode and the Z-Breathe setting. During treatment and while setting treatment
parameters, this display field shows the currently selected function mode (CPAP) and Z-Breathe
setting (1, 2 or 3). The General Status Display field may also give notification of certain events
which will NOT prevent prevent the Z1 from operating normally but may need to be addressed
by the user.
FAULT
“FAULT” is displayed if one of several conditions has occurred that require your attention. See Fault
Conditions (page 19) for explanations of the FAULT codes.
REMINDER
“REMINDER” is displayed if there is excessive air leakage. See Reminder Alert (page 18) for an
explanations of REMINDER conditions.
Current
Treatment
Pressure
Start/Stop
Button
CAUTION (When using the PowerShell Only): Make sure your Battery is fully
charged before commencing CPAP therapy using the PowerShell. The battery life
of a fully charged Extended Life Battery in the PowerShell is typically 8.9 hours or
longer and treatment will terminate when the charge is expended. To recharge
the Battery, see “Charging Battery”.
2. Press the Start/Stop Button a second time to stop treatment.
Figure 10
Choosing Between Z-Breathe®settings
The Z-Breathe setting of 1-3 is indicated by the number following
the CPAP designation on the LCD screen. For example CPAP-3
means the CPAP mode with a Z-Breathe setting of 3 (Figure 11) To
set the Z-Breathe setting, press and hold down the Up Arrow Button
and Down Arrow Button simultaneously while in the Standby
mode (pressure of“0.0” indicated on the display). Once you see the
Function Mode indicator flashing on the LCD screen, you can select
the desired Z-Breathe setting using the Up Arrow Button and Down
Up
Arrow
Button
Down
Arrow
Button
Ramp
Time
Ramp
Pressure
Figure 12
PRESSURE
cmH20
MINUTES
Z-Breathe Setting
Mode
Figure 11
Arrow Button. After you have made your selection, press and hold the Start/Stop Button to save it.
Your selected setting for the Z-Breathe Algorithm will be retained in the memory of your Z1 unit
after it is turned off. The currently selected Z-Breathe Algorithm setting will be displayed when the
Z1 is delivering treatment pressure.
Z-Breathe is a proprietary algorithm engineered in the Z1 to smooth out the breathing cycle
while enjoying the benefits of CPAP treatment. Typically users experience pressure swings while
breathing on a CPAP device. Pressure swings occur as a spike in pressure during exhalation and
a drop in pressure during inhalation. Z-Breathe is designed to minimize the impact of spikes and
drops while keeping the pressure constant, thereby delivering more comfortable therapy.
Pressure swings can vary between users due to lung capacity and other factors. For this reason we
have created 3 levels of Z-Breathe to accommodate individual users’preferences:
— 1 offers mild pressure swing relief
— 2 is the Default Setting and offers moderate pressure swing relief
— 3 is the most aggressive setting for pressure swing relief
Programming Ramp Mode
Some users find it comforting to acclimate to CPAP therapy each session. This can be
accomplished by programming the Ramp Mode which allows you to start with a low initial
pressure and gradually increase to the prescribed treatment pressure.
Setting Ramp Time
When the Z1 is in standby mode, press and hold the
Up Arrow Button for approximately 3 seconds until
the “00 MINUTES”value flashes. While it is flashing,
repeatedly press either the Up or Down Arrow
Buttons to increase or decrease the ramp time in 5
minute increments. (Figure 12) Set the ramp time to
“00 MINUTES”to disable the Ramp Mode.
When the desired ramp time value is reached, press
and hold the Start/Stop Button for approximately 3
seconds to save your ramp time setting (this setting
will be saved even when the Z1 is not powered). The
Z1 will return to standby mode. To return to standby
mode without saving the new setting, press and
release the Start/Stop Button momentarily instead of
holding it down.
Starting and Stopping Treatment
1. Press the Start/Stop Button on the Z1 to initiate
treatment using the pre-programmed prescription
setting. The Treatment Pressure Display will show
the actual pressure being delivered at all times. If the
Ramp Mode has been programmed, the ramp start
pressure will be displayed upon pressing the Start/
Stop Button and the indicated pressure will rise as
the Z1 ramps up to prescribed treatment pressure.
If the incorrect pressure is displayed, contact your
health care provider before commencing CPAP
therapy. (Figure 10)

16 17
NOTE: Ramp Time must be set to a minimum of 5 minutes prior to
attempting to program Ramp Start Pressure or the change to Ramp Start
Pressure will not take effect and “FAIL 2” will be shown in the General Status
Display field.
Setting Ramp Start Pressure
Press and hold the Down Arrow Button for 3 seconds until the ramp start pressure value flashes.
While flashing, repeatedly press either the Up or Down Arrow Buttons to increase or decrease the
ramp start pressure by 0.5 cm H2O increment. (Figure 12) Ramp start pressure must be lower than
prescribed pressure.
When the desired value for ramp start pressure is reached, hold down the Start/Stop Button for
approximately 3 seconds to save your setting. The Z1 will return to standby mode. To return to
standby mode without saving the new setting, press the Start/Stop Button momentarily instead
of holding it down.
User Software
Human Design Medical provides User Software for data viewing which may be downloaded from
the Human Design Medical website to any Windows computer. A separate user guide for this
software is available for download at the Human Design Medical website, www.hdmusa.com. This
software can be used to upgrade the Z1 firmware as new releases are made available.
Charging the Battery (PowerShell Only)
WARNING: Only charge the Battery with the
supplied Human Design Medical PowerShell
and the supplied Human Design Medical Power
Supply. Do not use any other battery charger, AC
adapter, or other device to charge the Battery.
With both the Z1 and the Battery fully inserted into the
PowerShell, plug the Power Supply into the PowerShell as shown
and into a power outlet. (Figure 13)
CAUTION: Make sure your Battery is fully
charged before commencing CPAP therapy using
the PowerShell. The battery life of a fully charged
Extended Life Battery in the PowerShell is typically
8.9 hours or longer, after which treatment will
terminate.
Ensure the Z1 is in standby mode (display reads 0.0). The Battery
will not charge unless the Z1 is in standby mode. The Battery
status indicator on the Z1 will flash when the Battery is charging.
The number of solid bars indicates the current level of charge.
(Figure 14)
The battery status indicator will stop flashing when the Battery Figure 14
Figure 13
NOTE: The battery will not charge if you are using your Z1 CPAP at the
same time.
Care
The Z1 System needs periodic care. It is the user’s responsibility to be aware of the following items
for care:
• Inspections
• Cleaning
• Filter Replacement
• CPAP Tube Replacement
• Reminders and Fault Conditions
Inspections
Before each use carefully inspect the Z1, CPAP Tube, Filter, and mask for signs of wear, damage or
discoloration. Contact your retail medical supplier or at www.hdmusa.com for replacement parts.
Cleaning
Regular cleaning should be performed in accordance with the instructions below.
WARNING: This device can expose you to electrocution. Do not
immerse the Z1, Power Adapter, Power Cord or Battery in water. Always
unplug these devices before cleaning (only as instructed in this User Guide)
and make sure they are dry before plugging back in.
CAUTION: Do not use bleach, chlorine, alcohol, or aromatic-based
solutions, moisturizing or antibacterial soaps or scented oils to clean the
Z1. Never attempt to clean the Power Adapter, PowerShell, Battery or
accessories with any of these solutions. These solutions may create a shock
hazard, cause damage and reduce the life of these products.
Daily
1. Remove the CPAP Tube and the mask from the Z1.
2. Wash the CPAP Tube in warm water using mild detergent.
3. Rinse thoroughly, hang, and allow to dry.
4. Before next use, reconnect the CPAP Tube and the Tube Adapter to the Z1 and mask.
NOTE: Soaking the CPAP Tube in water for longer than 10 minutes may
shorten the life of the tube.
is fully charged. Typically it will take 12 hours to fully charge the battery. After a few minutes, the
backlit display will turn off. This is to conserve power. It does not indicate that the battery is fully
charged.

18 19
The table below provides a listing of fault codes and recommended actions to be taken by the user.
Fault Code Explanation Action
21 Flashing Treatment pressure not programmed Discontinue use and contact your
health care provider or HDM.*
27 Flashing Overheating Let cool. Check to see if the fault
code goes away. Discontinue
use and contact your health care
provider or HDM.*
32 Flashing Over current Discontinue use and contact your
health care provider or HDM.*
33 Flashing Over pressure Discontinue use and contact your
health care provider or HDM.*
*Human Design Medical Service Center at 1-855-436-8724.
USB Firmware Upgrade
Z1 firmware upgrades can be downloaded via a link on support.hdmusa.com. Complete
instructions can be found on the web site. To upgrade the firmware on your Z1 connect the Z1 to
your PC with the included USB cable. The USB firmware upgrade can also be used to change the
date and time settings on the Z1 by following the instructions on the web site.
Troubleshooting
Problem Possible Causes & Solutions
No Display Power is not present. Ensure the Power Supply is
connected and the power outlet has power.
Loose connection of the Power Supply at the Z1
or loose connection of the Power Cord at the AC
Adapter. Check the connections.
Z1 is not locked into position in the PowerShell.
Slide the Z1 into the PowerShell until it clicks and
locks into place.
Battery is not locked into position in the PowerShell.
Slide the Battery into the PowerShell until the
Battery Shuttle clicks and locks into place.
Battery is not charged. Charge the Battery.
Fault Conditions
The Z1 has the ability to detect operational problems and will
display the word FAULT on the LCD screen. (Figure 15)
The LCD screen should be checked for any FAULT messages and
visible damage daily.
A fault condition is indicated when the word“FAULT” is displayed
accompanied by a two digit flashing number in place of the
pressure reading.
Figure 15
Every 90 days
The CPAP tube should be replaced every 90 days or earlier if it becomes discolored, cracked or
damaged.
CAUTION:Do not use bleach, chlorine, alcohol, or aromatic-based
solutions, moisturizing or antibacterial soaps or scented oils to clean
the CPAP tube. Never attempt to clean the CPAP tube with any of these
solutions. These solutions may create a shock hazard, cause damage and
reduce the life of the CPAP tube.
CAUTION: Prolonged exposure to sunlight may cause the CPAP Tube to
discolor and degrade
As Needed
1. Clean the exterior of the Z1 with a soft damp cloth and mild detergent and wipe
clean. Never submerge the Z1 in water. Do not allow soap or water to enter the Z1.
Filter Replacement
The Filter should be inspected weekly for holes and blockage by dirt and dust, and should be
replaced every month (or more often if necessary). The Filter is accessed by removing the End Cap
as described in “Assembling and Powering Up: Filter and End Cap”.
NOTE: The Z1 uses a specially designed Filter that is only available from
your retail medical supplier or at www.hdmusa.com.
WARNING: Do not attempt to use any air filter which is not supplied
by Human Design Medical for the Z1. The Z1 is designed to only work with
approved materials. A filter which is not approved may cause damage to
the Z1 from overheating.
CAUTION: Do not operate the Z1 without the Filter. The Z1 must
operate with a filter installed; failure to do so may damage the Z1 and allow
impurities into the respiratory system.
Reminder Alert (Air Leak Detection)
“REMINDER” is displayed on the LCD screen to indicate an excessive air leak. This could be because
you are not wearing your mask. Placing the mask on your face should remove REMINDER from the
display. If you are wearing the mask and REMINDER is displayed check the following connections
for air leaks:
• Tube Adapter to Z1
• CPAP Tube to Tube Adapter
• CPAP Tube to mask
• Mask to face (adjust position of mask and headgear to eliminate the air leak)

20 21
A power supply not designed for the Z1 is
connected to the Z1. Remove the power supply
unit and replace with a Human Design Medical
approved unit.
Treatment Will Not Start Check the display for a fault condition. Refer
to the Fault Codes table in“Fault Conditions”.
Contact the Human Design Medical Service
Center at 1-855-436-8724
Insufficient Air Flow Delivered from Z1 Air filter is dirty or clogged. Replace air filter.
Air intake is blocked. Clear blockage.
CPAP Tube is not connected properly. Check
CPAP Tube, Tube Adapter and connections to
your Z1 and your mask.
Ramp Mode is in use. Wait for air pressure to
reach prescribed treatment pressure or stop
treatment and adjust ramp time.
CPAP Tube is blocked, pinched or punctured.
Unblock CPAP Tube. Check the CPAP Tube for
punctures. Replace a damaged CPAP Tube
immediately.
Mask and/or headgear are not positioned
correctly and mask is leaking. Adjust position of
the mask and headgear until leak ceases.
Incorrect CPAP Tube selected. Do not use tubing
which was not recommended by your clinician.
Check the display for a fault condition. Refer to
fault codes in “Fault Conditions”.
Pressure Rises Inappropriately Talking, coughing or breathing in an unusual
manner. For example, avoid talking with a nasal
mask on and breathe as normally as possible.
There may be an air leak caused by improperly
connected tubing. Check for leaks.
Follow mask manufacturer’s instructions.
Mask Cushion is Vibrating Against the Skin Stop treatment. Remove and reposition the
and an Air Leak is Occurring mask and, if necessary, adjust the headgear.
Resume treatment.
Follow mask manufacturer’s instructions.
Z1 Has Been Left in a Hot Environment Allow Z1 to cool before use. Disconnect the
power cord and then reconnect it to restart the
Z1.
Electromagnetic Compatibility The Z1 and PowerShell meet all applicable
electromagnetic compatibility standards.
However, if you suspect performance of the
device is affected by interference from wireless
sources, move the Z1 and PowerShell further
away from those sources.
Guidance and Manufacturer’s Declaration
Electromagnetic Emissions & Immunity
Medical electrical equipment needs special precautions regarding EMC and needs to be installed
and put into service according to EMC information provided in this document.
The Z1 System does not have Essential Performance within the meaning of IEC 60601-1. However,
immunity testing was conducted in accordance with IEC 60601-1-2 and compliance was
confirmed as summarized in this document.
Guidance and manufacturer’s declaration - electromagnetic emissions
The Z1 CPAP System is intended for use in the electromagnetic environment specified below. The
customer or the user of the Z1 System should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment — guidance
RF emissions Group 1
CISPR 11
RF emissions Class B
CISPR 11
Harmonic emissions Class A
IEC 61000-3-2
Voltage fluctuations/ Complies
flicker emissions
IEC 61000-3-3
WARNINGS:
• The device should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the device should be observed to verify normal
operation in the configuration in which it will be used.
• The use of cables and accessories (eg, humidifiers) other than those specified for the
device is not recommended. They may result in increased emissions or decreased
immunity of the device.
The Z1 CPAP System uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
The Z1 CPAP System is suitable for use in all
establishments, including domestic establish-
ments and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes.

22 23
Guidance and manufacturer’s declaration — electromagnetic immunity
The Z1 CPAP System is intended for use in the electromagnetic environment specified below. The
customer or the user of the Z1 CPAP System should assure that it is used in such an environment.
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in
UT) for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95 % dip in
UT) for 5 sec
<5 % UT
(>95 % dip in
UT) for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95 % dip in
UT) for 5 sec
Mains power quality
should be that of a typical
commercial or hospital
environment, If the user
of the Z1 Auto System
requires continued
operation during power
mains interruptions, it is
recommended that the Z1
Auto System be powered
from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Immunity test IEC 60601
test level
Compliance level Electromagnetic
environment – guidance
NOTE: UT is the a.c. mains voltage prior to application of the test level.
.
Guidance and manufacturer’s declaration — electromagnetic immunity
The Z1 Auto System is intended for use in the electromagnetic environment specified below. The
customer or the user of the Z1 Auto System should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment —
guidance
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Z1 CPAP
System is used exceeds the applicable RF compliance level above, the Z1 CPAP System should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the Z1 CPAP System.
bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Conducted RF
IEC61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3V
3V/m
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Z1 Auto System, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d =1.2 √P
d =1.2 √P 80 MHz to 800 MHz
d =2.3 √P 80 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,ashould
be less than the compliance level in
each frequency range.b
Interference may occur in the vicinity
of equipment marked with the
following symbol:

24 25
Recommended separation distances between portable and mobile
RF communications equipment and the Z1 CPAP System
The Z1 CPAP System is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Z1 CPAP System can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the Z1 CPAP System as recommended below,
according to the maximum output power of the communications equipment.
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
WARNINGS:
Please note that mobile RF communications equipment can affect medical electrical equipment.
Wireless Quality of Service for Z1 Bluetooth Communication
with Z1 Mobile App
The Z1 and Mobile App utilize only a very small fraction of the available bandwidth and
throughput. Based on this, the Z1 and Mobile App are highly tolerant of reductions in wireless
quality of service and performance will not be affected until service is essentially lost. Wireless
connectivity is not part of the essential performance of the Z1. All functions of the Z1 can be
performed on the Z1 unit itself. The Mobile App and Z1 are tested for coexistence with other in-
band radio frequency sources. The Bluetooth hardware module is FCC listed, and complies with
FCC PART 15 SUBPART C.
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d =1.2 √P
80 MHz to 800 MHz
d =1.2 √P
800 MHz to 2,5 GHz
d =2.3 √P
Technical Specifications
30W Power Adapter Part # HD60-6010
Unit Input range 100–240V, 50–60Hz
Typical power consumption 20W (30VA)
Maximum power consumption 35W (70VA)
Operating temperature 41ºF to 95
º
F (5ºC to 35ºC)
Operating humidity 10–80% Relative Humidity, Non-Condensing
Operating altitude Sea level to 8,000 ft (2,433 m)
Operating atmospheric pressure 101 to 75.3 kPA
Storage and transport temperature -4ºF to +140ºF (-20ºC to +60ºC)
Storage and transport humidity 10–95% Relative Humidity, Non-Condensing
Storage and transport altitude No limit
Storage and transport atm. pressure No limit
Electromagnetic compatibility Compatible with IEC 60601-1-2
Aircraft use Complies with RTCA DO-160G Standard for use in
flight
IEC 60601-1 classification Class II (double insulation), Type BF
Available modes Standby, CPAP and CPAP with ramp
Deliverable pressure accuracies ±0.6 cm H2O ± 4% of the measured reading
Sound pressure level 26 dBA
Nominal dimensions (L x W x H) 6.48”X 3.30” X 2.02”(16.46 cm x 8.38 cm x 5.13 cm)
Weight 10 oz. (238 g)
Housing construction Flame retardant engineered thermoplastic
Supplemental oxygen Not for use with oxygen
Standard air filter Polyester
Air outlet The 22 mm air outlet complies with ISO 5356-1
Battery charges when Z1 not in use.
Expected Service Life Z1, PowerShell and Power Supply: 5 years
Tube Adaptor, USB Cable and End Cap: 2 years
Filter, CPAP TUbe: See Care Instructions.
Performance Specifications
Operating pressure range 4 to 20 cm H2O
Maximum single fault steady pressure 30 cm H2O
Pressure accuracy ±(0.6 cm H2O + 4% of the measured reading)

26 27
PowerShell with Extended Life Battery Specifications
Longevity (typical operating time)*
Z-Breathe Setting Treatment Pressure Operating Time
3 20 cmH2O 8.9 hours
* Results reflected were obtained via bench testing under simulated operating conditions at the
highest machine settings. Actual run time may vary due to machine settings, machine pressure
and personal breathing. Lower pressures will result in longer battery operating times.
Capacity 99.4 Watt-hours
Lifetime approximately 300 charge/discharge cycles
Shelf life Batteries are shipped with between 30% and
50% of rated capacity, providing a minimum of 6
months shelf life when stored at 25°C. If storage
temperature exceeds 25°C over a 6-month period
then shelf life will be reduced and the battery
should be recharged periodically.
Storage and transport temperature -4ºF to +140ºF (-20ºC to +60ºC)
Storage and transport humidity 10–95% Relative Humidity, Non-Condensing
Battery Technology Lithium Ion Battery Pack, 4S2P
Weight Approx. 15.5 oz (440g)
Charging Battery charges when Z1 is in standby mode.
The following symbols may appear on your device, power supply unit, or packaging.
Refer to Manual type BF equipment; Class II equipment; Start/stop;
Ingress Protection (IP) Rating; Manufacturer; Batch code; Catalog
number; Serial number; Direct current; Product disposal warning to EN50419; .
Disengagement Mechanism; Prescription only (In US. Federal law restricts these
device to sale by or on order of a physician)
Servicing
The Z1 System is intended to provide safe and reliable operation when operated in accordance
with the instructions provided by Human Design Medical. Human Design Medical recommends
that the Z1 System be inspected and serviced by an authorized Human Design Medical Service
Center if there is any sign of wear or concern with Z1 function. Otherwise, service of the Z1
generally should not be required during the warranty period (see ‘Limited Warranty’ for details) of
the device.
Limited Warranty
Human Design Medical warrants that your Human Design Medical product shall be free from
defects in material and workmanship from the date of purchase for the period specified below.
Component Warranty Period
Z1 Auto device 3 years
PowerShell 2 years
Extended Life Battery 6 months
Power Supply 2 years
If product fails to perform in accordance with product specifications, Human Design Medical will
repair or replace, at its option, the defective material or part. This warranty is only available to the
initial purchaser. It is not transferable. If the product fails under conditions of normal use, Human
Design Medical will repair or replace, at its option, the defective product or any of its components.
This limited warranty does not cover: a) any damage caused as a result of accident, improper use,
abuse, modification or alteration of the product; b) repairs carried out by any service organization
that has not been expressly authorized by Human Design Medical to perform such repairs; c)
any damage or contamination due to cigarette, pipe, cigar or other smoke; d) damage caused by
water entering the product. Warranty claims on defective product must be made by the initial
purchaser by calling Human Design Medical service support at 1-855-436-8724.
THIS WARRANTY REPLACES ALL OTHER EXPRESSED OR IMPLIED WARRANTIES, INCLUDING ANY
IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. SOME
REGIONS OR STATES DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY
LASTS, SO THE ABOVE LIMITATION MAY NOT APPLY TO YOU.
HUMAN DESIGN MEDICAL SHALL NOT BE RESPONSIBLE FOR ANY INCIDENTAL OR
CONSEQUENTIAL DAMAGES CLAIMED TO HAVE RESULTED FROM THE SALE, INSTALLATION OR
USE OF ANY HUMAN DESIGN MEDICAL PRODUCT. SOME REGIONS OR STATES DO NOT ALLOW
THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGES, SO THE
ABOVE LIMITATION MAY NOT APPLY TO YOU.
This warranty gives you specific legal rights, and you may also have other rights which vary from
region to region. For further information on your warranty rights, contact your local Human
Design Medical dealer or Human Design Medical office at 1-855-436-8724.

28 29
Contact Information
Human Design Medical / Breas
57 Chapel Street, Suite 200
Newton, MA 02458
www.hdmusa.com
Main phone (customer service): 1-855-436-8724
Service support or warranty coverage: 1-855-436-8724
FAA Compliance
To whom it may concern:
The US Department of Transportation (DOT) Final Rule,
“Nondiscrimination on the Basis of Disability in Air Travel” (73 FR
27614 which updates Title 14 CFR Part 382), effective May 13, 2009
provides important requirements for the accomodation of passengers
with respiratory assistive devices (Ventilators, Respirators and CPAP
machines).
In line with these requirements, respiratory assistive devices may
be used onboard an aircraft, without further testing by the carrier,
provided they have been tested for Electromagnetic Compatibility
(EMC) in accordance with the current version of RTCA/DO-160,
Section 21, Category M.
Human Design Medical has successfully completed testing for the
Z1 Auto System. The Z1 Auto System complies with RTCA/DO-160,
Section 21, Category M and can be considered FAA compliant.*
Some airlines may require advance notification before travel, and
devices may need to be operated by battery. Human Design Medical
recommends that customers check with their airline.
* Z1 Bluetooth must be in the off (airplane mode) position similar to
any other mobile device to maintain compliance
If you are planning on using your Z1 on an airplane, please detach this letter and carry it with you.
Human Design Medical / Breas • 57 Chapel Street Suite 200 • Newton, MA 02458 • 1-855-436-8724

30 31
This page is intentionally left blank. Register your new device.
Registering your device allows us to provide better customer
service and warranty coverage. Please register as soon as
possible.
Mail completed form to:
HDM Product Registration
Human Design Medical/
Breas
57 Chapel Street, Suite 200
Newton, MA 02458
About you.
NAME:
First MI Last
EMAIL ADDRESS:
MAILING ADDRESS:
Street or PO Box
City State Zip
PHONE NUMBER:
About the device.
WHAT DEVICE DID YOU PURCHASE?
WHERE DID YOU PURCHASE THE DEVICE? PURCHASE DATE:
____/____/____
[MM/DD/YY]
Z1 SERIAL NUMBER POWERSHELL SERIAL NUMBER
The device serial number can be found on the back of the
unit. Look for the string of numbers following
02082018

32
This page is intentionally left blank.
Table of contents
Other Human Design Medical Medical Equipment manuals
Popular Medical Equipment manuals by other brands

ADC
ADC Diagnostix 2100 Directions for use

SPORLASTIC
SPORLASTIC MALLEODYN S3 Fittings guide

Drive DeVilbiss Healthcare
Drive DeVilbiss Healthcare CHAD CH4808-L-BLUE PRODUCT INFORMATION AND INSTRUCTIONS

YOSHIDA
YOSHIDA X-era Smart Operation manual

Partial Hand Solutions
Partial Hand Solutions Titan Flex Series user guide

Maico
Maico MA 33 Operation manual

KaVo
KaVo ORTHOPANTOMOGRAPH OP 3D Pro installation manual

Durr Dental
Durr Dental XR 24 Pro Installation and operating instructions

BIOTRONIK
BIOTRONIK Galeo Pro Instructions for use

Welch Allyn
Welch Allyn KleenSpec 790 Series Directions for use

Ossur
Ossur Resolve Technical manual

GE
GE AMX-II Service manual