manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. ICON
  6. •
  7. Accessories
  8. •
  9. ICON Procon C Series User manual

ICON Procon C Series User manual

C Series
Conductivity Sensor
Operating Manual
ProCon®C Series Conductivity Sensor Operating Manual
2
Preface
Please read this manual carefully before use.
When receiving the instrument, please open the package carefully, check to ensure the instrument and accessories have not been
damaged during transportation. If any issues are found, please contact our after-sales service department or regional customer service
center. Ensure item is returned in the original packaging.
Only properly skilled authorized personnel should carry out installation, setup and operation. Ensure that the power cable is physically
separated from the power supply during the initial wiring connection or repair.
For example,
1. Apparent damage to the sensor
2. The sensor does not work properly or provides specied measurements
3. The sensor has been stored for a long time in an environment where the temperature exceeds 70°C
The sensor must be installed by licensed professionals in accordance with relevant local specications, and instructions are included in the
operation manual. Comply with the technical specications and input requirements of the controller.
Preparation before measurement
Important
The conductivity sensor has been factory calibrated. Initial on-site calibration should not be necessary.
When measuring high-purity water, avoid pollution, correctly select the constant of conductivity electrode, and preferably adopt the sealed
and owing measurement method. Otherwise, the conductivity value will increase quickly, because the carbon dioxide in the air will
become conductive carbonate ions after being dissolved in high-purity water, which will affect the measured value.
If calibration is required, please follow these steps:
4. The conductivity electrode shall be rinsed twice with deionized water [or distilled water) less than 0.5us/cm before use, and then rinsed
with the tested sample before measurement.
5. The value displayed by the instrument after measurement has been converted to the measured value at 25°C. If no compensation is
needed, unplug the temperature electrode and the instrument displays 25°C. The measured value is the conductance value of the
solution at that time.
6. The conductivity standard provides the simplest and most accurate method for calibrating conductivity sensors. Select a conductivity
standard whose concentration is close to or higher than the expected sample concentration.
7. The conductivity standard can be used to verify whether the conductivity sensor responds correctly.
Re-calibration Procedure
1. Prepare a calibrated beaker of appropriate size, which will allow you to completely immerse the conductivity sensor tip.
2. Check to determine if there are any deposits on the electrode sensing area. If the sensing surface (platinum black) is coated, clean the
sensor before continuing. Flush the sensor reaction area with distilled water.
3. Rinse the calibration beaker with some standard solutions, and then pour the selected conductivity standard solution into the calibration
beaker.
4. Immerse the electrode and ensure that the sensing area of the electrode is completely immersed in water. If there is a vent hole in the
sensing area, the electrode must be submerged under the vent hole. Shake sensor to ensure that trapped air bubbles are released from
sensing area
Important
a) Bubbles in the control area of the conductivity electrode will cause serious interference to the ion ow and will lead to erroneous
readings.
b) When measuring high-purity water, avoid exposure to CO2, correctly select the constant of conductivity electrode.
c) The recommended procedure is to adopt the sealed and owing measurement method.
5. Enter the standard value of the standard used from the menu.
6. Start calibration. Wait for the reading to stabilize, and then adjust the reading to the temperature compensation value of the conductivi-
ty standard.
NOTE:
a) Please refer to supporting instruments for specic calibration process.
b) Repeat the process to ensure that the calibration is correct.
7. Ensure sensor is ushed with DI water prior to storage.
ProCon®C Series Conductivity Sensor Operating Manual
3
Typical Installation
Intake
Output
Maintenance of conductive electrode
Cleaning of conductive electrode
1. The organic component contamination on the electrode can be cleaned with warm water containing detergent, or it can be cleaned
with alcohol.
2. Calcium and magnesium precipitates are best cleaned with 10% citric acid.
3. The bright platinum electrode can be mechanically cleaned with a soft brush. Ensure that no nicks or scratches are made on the
electrode sensor.
Do Not use a screwdriver or sharp object
to clean the electrode surface
4. The platinum electrode plated with platinum black should only be cleaned using the chemical method. The coating (platinum black) on
the electrode surface will be damaged if cleaning with a soft brush.
Warranty
Icon Process Controls warrants this product to be free from signicant deviations in material and workmanship for a period of one year
from the date of purchase. If repair is necessary and has not been the result of abuse or misuse within the warranty period, please return to
Icon Process Controls and amendment will be made without any charge. Icon Process Controls Customer Service Center will determine if
product problem is due to deviations or customer abuse. Out of warranty products will be repaired on a charge basis.
Authorization must be obtained from Icon Process Controls Customer Service Center to issue a RIR number before returning items for
any reason. When applying for authorization, please include date requiring the reason of return. Instruments must be carefully packed to
prevent damage in shipment and insured against possible damage or loss. Icon Process Controls will not be responsible for any damage
resulting from careless or insufcient packing.
Warning: Damage as a result of inadequate packaging is the User / distributor’s responsibility. Please follow the guidelines below before
transporting.
Use the original packaging material if possible, when transporting back the unit for repair. Otherwise wrap it with bubble pack and use a
corrugated box for better protection. Include a brief description of any faults suspected for the convenience of Customer Service Center, if
possible. If there are any questions, feel free to contact our Customer Service Center or distributors.
by
Phone: 905.469.9283 • Sales: sales@iconprocon.com • Support: support@iconprocon.com
Wiring
4-20mA 2-wire
Blue: mA-
Brown: mA+
4-20mA 4-wire
Transparent: 4-20mA
Black (thick): Ref
Red: Temperature
Black: Temperature
Connects directly to
ProCon®controller
Thicker
Wire
4-20mA + RS485
Output
 Red: 9-24VDC +
 Black: 9-24VDC -
Transparent: 4-20mA
Black (thick): Ref
Green: RS 485 A
White : RS 485 B
Thicker
Wire
4
22-0003
Need a great Controller?
Pair your C Series sensor with the industry’s
most innovative and standard feature rich controllers
iconprocon.com
Use with any of the following controllers — NO PREAMP REQUIRED
FIVE INPUT
SINGLE INPUT
DUAL INPUT

Other ICON Accessories manuals

ICON MicU LIve User manual

ICON

ICON MicU LIve User manual

ICON VH4 User manual

ICON

ICON VH4 User manual

ICON Ultra 4 User manual

ICON

ICON Ultra 4 User manual

ICON Uports 6 User manual

ICON

ICON Uports 6 User manual

ICON Level Pro ProScan User manual

ICON

ICON Level Pro ProScan User manual

ICON ProAudio Cube2Nano Live User manual

ICON

ICON ProAudio Cube2Nano Live User manual

ICON Utrack User manual

ICON

ICON Utrack User manual

ICON Mobile Umini VST User manual

ICON

ICON Mobile Umini VST User manual

ICON Umix1008Rack-VST User manual

ICON

ICON Umix1008Rack-VST User manual

ICON Ultrack Pro User manual

ICON

ICON Ultrack Pro User manual

ICON ProCon SSBS35 User manual

ICON

ICON ProCon SSBS35 User manual

Popular Accessories manuals by other brands

Vinotemp element VT-BC34TS owner's manual

Vinotemp

Vinotemp element VT-BC34TS owner's manual

Galvin Engineering Flowmatic TZ-IQWAVEB Product installation guide

Galvin Engineering

Galvin Engineering Flowmatic TZ-IQWAVEB Product installation guide

Simon 20000095-09 Series quick start guide

Simon

Simon 20000095-09 Series quick start guide

Crestron 1700C-BTNB-BEZEL Operations & installation guide

Crestron

Crestron 1700C-BTNB-BEZEL Operations & installation guide

Modern Portable jHub user guide

Modern Portable

Modern Portable jHub user guide

Baxtran BR40 Installation, Operation and Maintenance User’s Manual

Baxtran

Baxtran BR40 Installation, Operation and Maintenance User’s Manual

PCB Piezotronics IMI SENSORS HT642A02 Installation and operating manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS HT642A02 Installation and operating manual

ZIPPY SLW802 instruction manual

ZIPPY

ZIPPY SLW802 instruction manual

SICK WLA26 00 Series operating instructions

SICK

SICK WLA26 00 Series operating instructions

Hermann Sonda esterna Radio introduction

Hermann

Hermann Sonda esterna Radio introduction

Waterpik Waterflosser WP-150 Series manual

Waterpik

Waterpik Waterflosser WP-150 Series manual

Danfoss SASA Installation guides

Danfoss

Danfoss SASA Installation guides

IFM Electronic Efector 300 operating instructions

IFM Electronic

IFM Electronic Efector 300 operating instructions

Vuly THUNDER Assembly manual

Vuly

Vuly THUNDER Assembly manual

Thetford Tecma Silence Plus owner's manual

Thetford

Thetford Tecma Silence Plus owner's manual

Alula DWS-LL quick start guide

Alula

Alula DWS-LL quick start guide

Bpt VPM Series installation instructions

Bpt

Bpt VPM Series installation instructions

CellarPro XTSx owner's manual

CellarPro

CellarPro XTSx owner's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.