IKA RC 5 control User manual

IKA RC 5 control
Operating instructions EN

2
Device setup
Fig. 1
1
2
3
4
3
5
6
4
3
3
9
10
11
12
13
14
7
8
Item Designation
1 Filling opening lid
2 Wireless Controller (WiCo)
3 Handle
4 Venting grid
5RS 232 port
6USB port
7External temperature sensor socket
8Multifunction port
9Pump connector IN
10 Pump connector OUT
11 Power switch
12 Power socket
13 Backflow
14 Overflow

3
Contents
EN
Source language: German
Page
Device setup............................................................................................................................................. 2
EU Declaration of Conformity................................................................................................................... 4
Note for USA (FCC)................................................................................................................................... 4
Note for Canada (IC)................................................................................................................................. 4
Explication of warning symbols ............................................................................................................... 4
Safety instructions ................................................................................................................................... 5
General information����������������������������������������������������������������������������������������������������������������������������������������� 5
Fluids ��������������������������������������������������������������������������������������������������������������������������������������������������������������� 6
Refrigerant������������������������������������������������������������������������������������������������������������������������������������������������������� 6
Battery (Battery pack RB 1 for WiCo)��������������������������������������������������������������������������������������������������������������� 6
Disposal instructions ���������������������������������������������������������������������������������������������������������������������������������������� 5
Intended use ............................................................................................................................................ 7
Use������������������������������������������������������������������������������������������������������������������������������������������������������������������ 7
Range of use���������������������������������������������������������������������������������������������������������������������������������������������������� 7
Wireless remote control ����������������������������������������������������������������������������������������������������������������������������������� 7
Unpacking................................................................................................................................................ 8
Unpacking ������������������������������������������������������������������������������������������������������������������������������������������������������� 8
Delivery scope�������������������������������������������������������������������������������������������������������������������������������������������������� 8
Preparations............................................................................................................................................. 8
Setting up�������������������������������������������������������������������������������������������������������������������������������������������������������� 8
Connecting the tubings������������������������������������������������������������������������������������������������������������������������������������ 8
Inserting battery into the WiCo ����������������������������������������������������������������������������������������������������������������������� 9
Mounting the WiCo to the station������������������������������������������������������������������������������������������������������������������� 9
Filling and draining������������������������������������������������������������������������������������������������������������������������������������������� 9
Fluids (Standard information for IKA fluid) ����������������������������������������������������������������������������������������������������� 11
Moving the device ����������������������������������������������������������������������������������������������������������������������������������������� 12
Charging the Battery Pack RB 1 (rechargeable battery)������������������������������������������������������������������������������������ 12
Changing battery to WiCo����������������������������������������������������������������������������������������������������������������������������� 12
WiCo - holder WH 10������������������������������������������������������������������������������������������������������������������������������������ 12
Operator panel and display .................................................................................................................... 13
station���������������������������������������������������������������������������������������������������������������������������������������������������������� 13
WiCo ������������������������������������������������������������������������������������������������������������������������������������������������������������ 13
Commissioning and operating................................................................................................................ 14
Useful information ................................................................................................................................. 15
Working with WiCo................................................................................................................................ 15
Working screen at the time of delivery ����������������������������������������������������������������������������������������������������������� 15
Explanation of symbols on the working screen ����������������������������������������������������������������������������������������������� 15
Menu navigation and structure ���������������������������������������������������������������������������������������������������������������������� 16
Menu (Details) ����������������������������������������������������������������������������������������������������������������������������������������������� 18
Interface and output .............................................................................................................................. 23
Maintenance and cleaning ..................................................................................................................... 26
Error codes............................................................................................................................................. 27
Accessories............................................................................................................................................. 28
Technical data ........................................................................................................................................ 29
Warranty ................................................................................................................................................ 31
Pump performance curve ....................................................................................................................... 31

4
EU Declaration of Conformity
Note for USA (FCC)
This equipment complies with Part 15 of the FCC rules� Any changes or modifications not expressly approved by the
Manufacturer could void the user's authority to operate the equipment� This device complies with Part 15 of the FCC
rules subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept all interference received, including interference that may cause undesired operation�
NOTE:
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of
the FCC Rules� These limits are designed to provide reasonable protection against harmful interference when the equip-
ment is operated in a commercial environment�
This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance
with the instruction manual, may cause harmful interference to radio communications� Operation of this equipment in a
residential area is likely to cause harmful interference in which case the user will be required to correct the interference
at his own expense�
Note for Canada (IC)
This device complies with Industry Canada license-exempt RSS standard(s)� Operation is subject to the following two
conditions:
(1) This device may not cause interference
(2) This device must accept any interference, including interference that may cause undesired operation of the device�
This device complies with Health Canada’s Safety Code 6/IC RSS-210� The installer of this device should ensure that RF radiation
is not emitted in excess of the Health Canada’s requirement�
Explication of warning symbols
Indicates an (extremly) hazardous situation, which, if not avoided, will result in death,
serious injury�
Indicates a hazardous situation, which, if not avoided, can result in death, serious
injury�
Indicates a potentially hazardous situation, which, if not avoided, can result in injury�
Indicates practices which, if not avoided, can result in equipment damage�
Indicates the risk of fire or explosion!
Indicates crushing risk to fingers/hand�
Warning: flammable material!
WARNING
CAUTION
NOTICE
DANGER
WARNING
WARNING
We declare under our sole responsibility that this product corresponds to the regulations 2014/35/EU, 2006/42/EC, 2004/108/
EC and 2011/65/EU and conforms with the standards or other normative documents: EN 61010-1, EN 61010-2-010,
EN 61326-1, EN 60529, EN ISO 12100 and DIN 12876-1�
Bluetooth®module:
Directive: 2014/53/EU Standards: EN 300328, EN 301489-17, EN 301489-1, EN 60950-1
A copy of the complete EU Declaration of Conformity can be requested at sales@ika�com�

5
Safety instructions
General information:
• Read the operating instructions completely before
starting up and follow the safety instructions.
• Keep the operating instructions in a place where it can be
accessed by everyone�
• Ensure that only trained staff work with the device�
• Follow the safety instructions, guidelines, occupational
health, safety and accident prevention regulations�
• Set up the device in a spacious area on an even, stable,
clean, non-slip, dry and fireproof surface�
• The device has been constructed to the requirements of
EU (European Union) and EFTA (European Free Trade As-
sociation) countries�
• Do not operate the device in explosive atmospheres�
• Protect the device and accessories from bumping and
impacting�
• Check the device and accessories for damage before
each use� Do not use damaged components�
• Safe operation is only guaranteed with the accessories
described in the ”Accessories” section�
•
The device must only be operated with the original power
cable�
• The socket for the power cable must be easily accessible�
• Socket must be earthed (protective ground contact)�
• The voltage stated on the type plate must correspond to
the power voltage�
• The device can only be disconnected from the power sup-
ply by pulling out the power plug or the connector plug�
• Disconnect the power plug before attaching or chang-
ing any accessories�
• Disconnect the power plug before cleaning, mainte-
nance and transportation of the unit�
•
The device must only be opened by trained specialists,
even during repair� The device must be unplugged from
the power supply before opening� Live parts inside the
device may still be live for some time after unplugging
from the power supply�
Coverings or parts that can be re-
moved from the device must be
put back on the device again to
ensure safe operation, for example to keep foreign ob-
jects and liquids, etc� from getting into the device�
• The device may only be used as prescribed and as de-
scribed in these operating instructions� This includes op-
eration by instructed specialist personnel�
• When using critical or hazardous materials in your pro-
cesses, IKA recommends to use additional appropriate
measures to ensure safety in the experiment� For ex-
ample, users can implement comprehensive monitoring
equipment�
• Process pathogenic material only in closed vessels under
a suitable fume hood� Please contact IKA application
support if you have any question�
If the power switch is not within
reach when device is operating, an
EMERGENCY STOP switch that
can be easily accessed must be installed in the work area�
• A laboratory circulator refrigerates and circulates fluid
according to specified parameters� This involves hazards
due to low temperatures and general hazards due to the
device of electrical energy� The user safety can not be
ensured simply with design requirements on the part of
the device� Further hazard sources may arise due to the
type of tempering fluid, e�g� by exceeding or undercut-
ting certain temperature thresholds or by the breakage
of the container and reaction with the carrier fluid� It is
not possible to consider all eventualities� They remain
largely subject to the judgment and responsibility of the
operator� For this reason, it may become necessary for
user to take other precautionary safety measures�
• Insucient ventilation may result in the formation of ex-
plosive mixtures� Only use the device in well ventilated
areas�
• When device is used for external circulation, extra pre-
caution must be taken for cold fluid leakage due to
damaged hose�
- Use suitable hoses for connection�
- Secure hoses and tubes against slippage and avoid
kinks�
- Check hoses, tubes and bath at regular intervals for
possible material fatigue (cracks/leaks)�
Do not start up the device if:
- It is damaged or leaking
- Cable (not only supply cable) is
damaged�
• After a power failure during operation, the device may
start automatically (depending on operating mode)�
Because of the heavy weight of
the device, at least two person are
needed for carrying the device�
• Transport the device with care (see section “Moving the
device“)�
• Do not transport or empty the bath while it is still cold�
Check the temperature of the bath fluid before draining�
• Always empty the bath before moving the unit�
Always empty the bath if the unit
is unused for long time�
WARNING
Disposal of the device:
• The device must be disposed of in accordance with na-
tional or local regulations
CAUTION
WARNING
NOTICE
NOTICE

6
Fluids:
Only use the fluids, which fulfill the
requirements for safety, health and
device compatibility� Be aware of
the chemical hazards that may be associated with the
bath fluid used� Observe all safety warning for the fluids�
• Depending on the bath fluid used and the type of opera-
tion or toxic can arise� Ensure suitable extraction�
• Do not use any fluid which may cause dangerous reac-
tions during processing�
• Only use recommended bath fluid� Only use non-acid
and non corroding fluid�
Never operate the device without
sucient refrigeration carrier flu-
id! Check the fluid level detection
at a regular basis (see the section “Filling and draining”)�
• Continuous monitoring of the bath and the filling level of
the bath fluid is required�
• To ensure a sucient fluid circulation, the viscosity of
the bath fluid must not exceed of 50 mm2/s at the low-
est operating temperature�
• Untreated tap water is not recommended� It is recom-
mended to use distilled water or high purity water (ion
exchangers) and add 0,1 g soda (sodium carbonate
Na2CO3) /liter, to reduce corrosive properties�
Don’t use following fluids:
- Untreated tap water
- Acids or bases
- Solutions with halides: chlorides, fluorides, bromides,
iodides or sulfur
- Bleach (Sodium Hypochlorite)
- Solution with chromates or chromium salts
- Glycerine
- Ferrous water�
Battery pack RB 1 (for WiCo):
If during operation the battery
pack RB 1 (rechargeable battery)
becomes fully discharged, the de-
vice (station) will continue to run or is shut down depend-
ing on the value settings for “Time Out”, “Safe Speed”
and “Safe Temperature”� If the device is set so that it con-
tinues to run when the battery of the WiCo is fully dis-
charged, the only means of switching the station off are
the “safe STOP”, “ON/OFF” key and the “Power switch”!
Please note the following
safety instructions for the
battery pack RB 1 (recharge-
able battery):
•
Keep the battery pack out of reach of children at all times�
• Store the battery pack in a cool, dry place�
• Never throw the battery pack into a fire� Keep it away from
direct sunlight and temperatures above 60 °C� High tem-
peratures will damage the battery pack and render it unus-
able� Temperatures above 100 °C may cause it to explode�
• Never throw the battery pack into water or expose it to
moisture� Water may lead to a short-circuit, causing the
battery pack to explode�
•
Do not deform or crush the battery pack or damage it
in any other way� This can cause battery fluid to leak
and/or the battery pack to explode�
• When not in use, keep battery packs away from pa-
perclips, coins, keys, nails, screws or other small metal
objects which could cause the contacts to be bridged�
Short-circuiting may result in an explosion�
• Explosion of a battery pack may release battery fluid and
cause a fire�
• The lithium polymer battery pack must only be used and
charged in IKA products designed for use with this bat-
tery pack�
• When the battery pack is inserted it should slide in easily
and without resistance� Do not force it�
• If the battery pack is removed for an extended period of
time, store it in a sealed plastic bag to prevent short-circu-
iting due to moisture or contact with metal components�
• The operating temperature range of the battery pack is
from 0 °C to + 45 °C� Note that the battery pack capac-
ity will be reduced at temperatures below 20 °C�
•
Only the rechargeable battery types recommended in the
technical data may be used in the device!
Do not charge batteries that have leaked or that
are discolored, deformed or damaged in any
other way�
NOTICE
NOTICE
NOTICE
NOTICE
CAUTION
Refrigerant:
The device is not allowed to be
used in an ATEX (Atmosphere EX-
losible) zone�
Flammable refrigerant used� Do
not use mechanical devices or oth-
er means to accelerate the defrost-
ing process, other than those rec-
ommended by the manufacturer�
Do not damage the refrigerant cir-
cuit�
• Set the device in well-ventilated room and
keep all venti-
lation openings in the enclosure or, in the structure for
building, clear of obstruction�
WARNING
WARNING
WARNING
• Flammable refrigerant used� Only the manufacturer can
open the device and
repair it� The third parties are not
allowed to open the device!
• The device with a sealed refrigerating system with less
than 150 g refrigerant� It can be located in any laboratory�

7
Intended use
Use:
Use RC (Refrigerated Circulators) for cooling and circulating
fluids�
Range of use:
Indoor environments similar to that a laboratory of re-
search, teaching, trade or industry�
Wireless remote control:
Before using the wireless link between the WiCo and the
laboratory device, first check whether your region is in-
cluded in the radio communication approval for the de-
vice� If it is not, remote control can also be performed
using a USB cable�
The safety of the user cannot be guaranteed:
- If the device is operated with accessories that are not
supplied or recommended by the IKA�
- If the device is operated improperly or in contrary to the
IKA specifications�
- If the device or the printed circuit board are modified by
third parties�
Disposal instructions:
•
The device, packaging, accessories and batteries must be
disposed of in accordance with national regulations�
• When disposing of the battery pack, please tape over the
contacts with adhesive tape to prevent short-circuiting due
to moisture or contact with metal components� Short-circu-
iting may result in an explosion�
• Do not throw used battery packs into your household
waste� Dispose of them properly in accordance with statu-
tory regulations�
End users are obliged by law to return all used
disposable and rechargeable batteries� Throwing
them into the household waste is prohibited� Dis-
posable/rechargeable batteries containing harmful sub-
stances are marked with this symbol to indicate that they
may not be disposed of as household waste�

8
Fig. 4
Union nut
Stopper
Adapter
England
Adapter
Australia
Adapter
Europe, Switzerland
Fig. 3
Adapter
USA, China
Fig. 2
Unpacking
Unpacking:
- Please unpack the device carefully�
-
Any damage should be notified immediately to the shipping
agent (post oce, railway network or logistics company)�
Delivery scope:
- RC 5 control with WiCo
- Power cables
-
Hose olive
NW 8 (2 pieces) see Fig� 2
-
Hose olive
NW 12 (2 pieces) see Fig� 2
- WiCo holder WH 10
- OS 1�0 power supply (for WiCo)see Fig� 3
- Temperature sensor Pt 100�30
- USB 2�0 cable (micro A – micro B)
- USB 2�0 cabel (A – micro B)
- Plastic cap (for “Overflow” connector)
- Plastic cap (for “Backflow” connector)
-
User guide
- Warranty card�
Preparations
Setting up:
- Place the unit on an even, stable, clean, nonslip, dry and
fireproof surface�
- Keep at least 20 cm of open space on the front and rear
side�
- The place for installation should be large enough and
provide sucient air ventilation to ensure the room does
not warm up excessively because of the heat from de-
vice radiates to the environment�
- Do not set up the device in the immediate vicinity of
heat sources and do not expose to sun light�
- Cooling machine, pump motor and electronics produce
intrinsic heat that is dissipated via the venting grids (4)!
Never cover these venting grids!
Note: After setting up the device, wait at least one hour
before starting the operation to avoid the damage to the
cooling system�
Connecting the tubings:
-
Unscrew the union nuts and stoppers using a wrench
(AF 19) from the pump connector IN and OUT �
-
Connect the hoses for circulating the external system to
the pump connection M 16 x 1 for IN and OUT directly
or with the olives�

9
External
system
IKA®
Calorimeter
Fig. 6
Model:WiCo CBC 5
LPS
IP40
Model:WiCo CBC 5
LPS
IP40
Model:WiCo CBC 5
LPS
IP40
Note: notch upward!
Model:WiCo CBC 5
LPS
IP40
Model: WiCo CBC 5
LPS
IP 40
Fig. 7
Note: If the WiCo need be permanently attached to the
station, we strictly recommend to fasten the unlocking
button with the integrated screw (turn counter clock wise)�
IKA
®
CBC 5 control
High
Low
Fluid level
R
Fig. 8
Fig. 9
Union nut
Olives
Fig. 5
Note: When the external system is not necessary, please
seal the pump connector IN and OUT with the existing
union nuts and stoppers�
- Screw the hose olives to the pump connector IN and
OUT with union nuts� Slide the hoses (NW 12) onto the
olives� The hoses must be secured with suitable clamps�
- Via a hose to the “Overflow” connection, overflowing
fluid can be directed into a suitable vessel� The vessel
should be positioned lower than the “Overflow” con-
nection
In case of closed external system, if there is in any case
no risk of exceeding the maximum fluid level, you can
close the “Overflow” connection with the included
overflow cap� This minimizes the energy loss and pro-
tects, in case of long-term applications with low tem-
peratures and high humidity, against unwanted water
input and the associated volume increase�
Check fluid heat expansion!
- Connect the “Backflow” connection to IKA calorimeter
with a suitable hose�
When IKA calorimeter is not connected, please close the
“Backflow” connection with included backflow cap�
Inserting battery into the WiCo:
Mounting the WiCo to the station:
Filling and draining:
- Before filling the fluid into the bath, open the venting
grid as indicated in following figure�

10
Note: Pay attention to the fluid level information�
Fluid level information on the WiCo display:
— Low level
— High level
Fluid level information on the fluid level indicator:
Red (Low Level)
Red (High Level)
Fig. 11
Fig. 12
Fig. 10
Drain valve
Drain port
- Check and make sure that the drain valve is closed (ro-
tate clockwise to the stop position, see Fig� 10)�
Note: Please note information in section “Commissioning
and operating”�
- Connect the power plug and turn on the device with
power switch�
- The low level icon ( ) appears on the display of the
WiCo� Meanwhile, the bottom LED segment of fluid
level indicator lights up in red color to indicate the low
fluid level�
- Open the filling opening lid and remove the stopper
from the filling opening� Then, fill fluid to the bath�
- To drain the fluid from the bath, connect a hose to the
drain port and turn the drain valve in counter clockwise
direction with a straight screwdriver�
Warning: Do not empty the bath while it is still cold,
there is a risk of freezing�

11
Fluid (Standard information for IKA fluid):
IKA Designation
Operating temperature
range for open bath
application
(°C)
Operating temperature
range for closed bath
applications
(°C)
Safety temperature
(°C)
Flash point
(°C)
CF�EG28�N10�80�8
-10 ��� 80 -10 … 80 90 115
CF�EG39�N20�80�16
-20 … 80 -20 … 80 90 115
CF�EG44�N25�80�19
-25 … 80 -25 … 80 90 115
CF�EG48�N30�80�22
-30 … 80 -30 … 80 90 115
UF�Si�N30�150�10LV
-30 … 130 -30 … 150 145 >170
HF�Si�20�200�50
20 … 200 20 … 200 255 >280
HF�Si�20�250�50A
20 … 200 20 … 250 255 >280
H
2
O (Water)
5 … 95 5 … 95 - -
Customized 1
Customized 2
Check the suitability of the fluid according to your application�
Nomenclature for IKA fluid:
CF�EG28�N10�80�8 --
(1) Classification:
HF: Heating Fluid
CF: Cooling Fluid
UF: Universal Fluid
(2) Chemical composition:
Si:Silicone oil
EG:Ethylene Glycol
(3) Temperature range: (
Minimum temperature� Maximum temperature)
N: Negative Temperature
(4) Viscosiy:
Viscosiy at 25 °C for Heating Fluid (HF)
Viscosity at -20 °C for Cooling Fluid (CF)
Viscosity at 25 °C for Universal Fluid (UF)
Dynamic viscosity [mPa�s] is a product of kinematic viscosity [mm²/s] and density [kg/m³] of the fluid divided by 1000�
(5) Additional information:
A: Oil Additives
LV: Low Viscosity
Note: for open bath application!
Note: Tap water may be unsuitable for operation because the calcium carbonate content may cause calcification�
High purity water (from icon exchangers) and distilled or bi-distilled water are unsuitable for operation due to corrosive
properties of these media� High purity water and distillates are suitable as a medium after adding 0�1 g soda (Na2CO3,
sodium carbonate) per liter of water�
Note: The temperature limit values are adjustable in accordance with the fluid used�
(5) Addition information
(4) Viscosity (8
mm²/s
)
(3) Temperature range (-10 – 80 °C)
(2) Chemical composition (Ethylene Glycol 28%)
(1) Classification (Cooling fluid)

12
1
Fig. 14
WiCo – holder WH 10
WiCo –holder WH 10
Fig. 15
WiCo –holder WH 10
H 44 Boss head clamp
H 38 Holding rod
IKA
®
RC 5 control
IKA
®
RC 5 control
IKA
®
RC 5 control
max. 10˚
Fig. 13
Moving the device:
Empty all fluid in the bath before move device from one
place to other place�
The device must be lifted by two persons with the upper
handles�
It can also be moved on flat surface by lifting and pushing
the front of the device� It is easy to move the device with
the help of the rolling of the wheels�
The angle of inclination should never more than 10 º at
any direction when move the device!
Note:
The device must not be moved during operation�
After moving the device, you must wait for at least one
hour before restarting the unit�
Charging the Battery Pack RB 1 (rechargeable battery):
The battery pack in the WiCo can be charged by any of the following means:
- On the station
- Via a USB cable at the PC or station
- Via an OS 1�0 power supply unit�
Replacing battery to WiCo:
Notice: Please comply with the relevant safety instructions in the "Safety instructions" section for the RB 1 battery pack!
WiCo – holder WH 10:

13
Operator panel and display
Item Designation Function
A Fluid level indicator: Indicate the bath fluid level (when the fluid level is too low or too high, the
bottom or the top LED segment change into red color)�
B LED bar: Display different status of the circulator with different color�
C Contacts: Communicate and charge the WiCo�
D safe STOP key: Safely stop the working of the circulator in emergency�
E Bluetooth®searching key: Search the WiCo when the Bluetooth®is active�
F Bluetooth®LED: Bluetooth®indicator�
WiCo:
Fig. 17
R
P
L
U
T
V
M
N
OS
High
Low
Fluid level
Fig. 16
A
F
D
E
B C
Item Designation Function
L On/Off
button
:Switch the WiCo on and off�
M “int/ext”
button
:
Switch between the internal and external temperature display and control�
N “Timer/Pump”
button
:
Switch between the timer and pump display�
O Graph
button
:Display time/temperature graph�
P Rotating/pressing knob: Set the temperature value�
Start/Stop the cooling function�
R Rotating/pressing knob: Navigation, selecting and changing the settings in the menu�
Set the pump speed value�
Start/Stop the pump function�
S Display: Screen�
T “Back”
button
: Return to the previous menu level�
U "Menu"
button
:Press it once: main menu is displayed�
Press it a second time: back to the working screen�
V Lock
button
: Lock/Unlock the knobs and
buttons
�
station:

14
Commissioning and operating
Note:
Before commissioning, make sure that the device has
not been moved for one hour!
Check whether the voltage given on the type plate cor-
responds to the available power voltage�
The socket used must be earthed
(fitted with earth contact)�
If these conditions have been met, the machine is ready
for operation when the power plug is plugged in�
If these conditions are not met, safe operation is not guar-
anteed and the machine could be damaged�
Observe the ambient conditions (temperature, humidity,
etc�) listed under “Technical Data”�
After switching on the power switch of the station, the
screen of the WiCo displays the device designation and
the software version after a beep�
After several seconds, screen display the information of
WiCo�
IKA®
Circulator
RC Control
Version 1.2.008
Then the working screen appears and the device is ready
for operation�
Information
Internal Max: 40 °C
Internal Min: 5 °C
External Max: 40 °C
External Min: 5 °C
Fluid: H2O (Water)
Change the temperature setting with left knob (P)�
Change the pump speed setting with the right knob (R)�
In standby status, activate the refrigerating function by
pressing the left knob (P)� The pump and fan starts to run
at once� The refrigerating function may delay to start at
maximum 2 minutes�
In working status, press the right knob (R) to stop opera-
tion of the pump� The refrigerating function and the pump
stops�
Note:
In standby status, press the right knob (R) to start
the pump function, the
refrigerating
function will not be
activated�
In working status, press the left knob (P) to stop the
refrig-
erating
function, the pump keep running�
When the WiCo is switched on without being connected
to the station, the green LED bar (B) and the green Blue-
tooth®LED (F) on station lights up�
The control elements of the WiCo allow locked by press-
ing the lock button (V) about 2 seconds, so no accidental
changes during operation are possible (lock symbol
appears in the display)�
By pressing lock button (V)
about 2 seconds
again, the
control elements are released (lock symbol disappears
from the display)�
Note: In an emergency, the device function can be turned
off by pressing the “safe STOP” button (D) at the front of
station� The LED bar (B) changed into red color�
A message appears in the display indicates that the station
was forced off� To restart, press the power switch off and
on again�
If the Bluetooth®function of the WiCo is activated, the
Bluetooth®icon appears on the screen and the user
can search the WiCo by pressing the searching button (E)�
Then, a beep is heard�
Fig. 18
Fig. 19
Fig. 20
A
0.00
int °C
ext °C
int 5.00 1000
set °C Pump rpm

15
Useful information
The station is controlled via a WiCo� If the WiCo is
mounted on the station, data is exchanged between the
station and WiCo via the contacts (
C
)� The screen of the
WiCo displays the home icon �
The WiCo is equipped with a USB socket (Universal Serial
Bus) with which the WiCo can be connected with station,
the USB icon appears on the screen�
If the WiCo is not connected with station via a USB
cable, the data exchange between the station and the
WiCo via Bluetooth®� In this case, the Bluetooth®icon
is displayed �
Depending on the structure of the building, the WiCo can
be operated at a distance up to 15 m from the station,
using the Bluetooth®connection�
The WiCo could be either installed on the
station
or put
on a safe place where is accessed easily by the user during
operation�
If the WiCo is mounted on the station, the battery is
charged through the contacts (C)�
The battery could also be charged via the USB port on the
WiCo (See “Charging the Battery Pack RB 1“ in “Prepara-
tions“ section)�
Working with WiCo
Working screen at the time of delivery:
Fig. 21
Explanation of symbols on the working screen:
The symbols displayed change depending on the status and
settings of the WiCo and the station� The screen below
shows the most significant symbols on the working screen�
Bluetooth®:
This symbol means the station and the WiCo are com-
municating via Bluetooth®�
The symbol no longer appears if no Bluetooth®communi-
cation is being performed�
99 Device number:
This figure appears when the “Device Number“ is activat-
ed in menu option “Display“�
The figure can be set from 0 to 99�
See section “Display“ under “Menu (Details)“�
Lock:
This symbol means that the function of the keys and the
rotary knobs for controlling the WiCo are disabled�
The symbol no longer appears if the functions are enabled
once again by pressing the key button a second time�
Temperature Sensor:
This symbol appears when the external temperature sen-
sor is connected�
AOperating Mode:
This symbol indicates the operating mode currently selected
(A, B, C, D)�
USB:
This symbol means the WiCo is communicating or the
battery is charged via a USB cable� The symbol no longer
appears if no USB cable is being used for communicating
with the station�
Home:
This symbol means that the WiCo is connected to the
station and is communicating with the station via the
contacts�
The symbol no longer appears if the WiCo is removed
from the station�
A
0.00
int °C
ext °C
int 5.00 1000
set °C Pump rpm
Fig. 22
A
99
10.00
int °C
ext °C
int 10.00 1000
set °C Pump rpm
PC
10.00
Note: The wireless symbol appears only when the station
is switched on�

16
Battery pack:
This symbol indicates the charging status of the RB 1 bat-
tery pack within the WiCo�
The charging symbol appears if the WiCo
- Is connected to a PC via a USB cable
- Is connected to a station via a USB cable
- Is connected to the power supply unit OS 1�0 via a USB
cable
- Is connected to the station via the charger contacts�
Refrigerating
This symbol indicates that the refrigerating function is active�
indicates active refrigerating process�
Fluid level:
This symbol indicates fluid level�
The red symbol means the fluid is above the maximum
fluid level� Excess fluid should be drained out�
The red symbol means the fluid is below the minimum
fluid level� Please add fluid�
Pump:
This symbol indicates that the pump is activated�
Warning:
This symbol indicates that warning is active�
PC PC Control:
This symbol means that either the station or the WiCo
is connected to a computer and is being controlled from
the computer�
PR Program Control:
This symbol indicates that the WiCo is controlled by a pro-
gram (see “PROGRAMS”)�
Menu navigation and structure:
Menu navigation:
Press the "Menu" button (U)�
Select the menu by turning the right rotating/pressing knob (R) to the right or left�
Open the menu item by pressing the right rotating/pressing knob (
R
)�
Turn the rotating/pressing knob (
R
) to select the desired menu option and edit
the values or settings�
Press the rotating/pressing knob (
R
) to get into sub menu items to active/inactive
to switch settings or to confirm settings ("OK")�
Press the "Back" button (T) to a setting or cancel to return to the previous
menu�
Press the "Menu" button (U) to return directly to the working screen�
Note: If you press the “Menu“ button (U), the system skips directly back to the working screen�
If you press the ”Back” button (T), the system skips back to the previous screen�
Menu navigation: Press the “Menu“ button (U) and rotate rotating/pressing knob (R)�
Press the ”Back” button (T) or the ”Menu” button (U)�
Menu
A
0.00
int °C
ext °C
int 5.00 1000
set °C Pump rpm
> Tempering > Control
Control Mode
Internal (int)
External (ext)
Menu
BackBackBackBack
TEMPERING
MENU Tempering
Control
Limits
Temperature Sensor
> Tempering
Control
Control Mode
Control Parameters
Fluids
Fig. 24
Note: When the tempering or pump functions are active, the menu is locked� On the display, the active menu option is
highlighted in yellow� The active status of a menu item is marked with a check (√)�
Control elements for
menu navigation
Fig. 23
R
U
T

17
Menu structure:
TEMPERING Control Control Mode
Fluids
Control Parameters
Internal (int) activated
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
activated
activated
activated
40 °C
5°C
60 °C
± 0.0 K
3.0, 20.0, 2.0, 3s, +2.00, -1.00
3.0, 20.0, 2.0, 3s, +2.00, -1.00
-
Automatic
Internal
Internal
Internal (Kp, Ti, Td, Ts, Prop_Bp, Prop_Bn)
2 Point Calibration
2 Point Calibration
Manual
Eternal (ext)
Eternal (ext)
I T (int - ext)I
Calibration
Reset Calibration
Maximum
Minimum
Eternal
Eternal
Eternal (Kp, Ti, Td, Ts, Prop_Bp, Prop_Bn)
3 Point Calibration
3 Point Calibration
Information (Kp, Ti, Td, Ts, Prop_Bp, Prop_Bn)
Limits
Temperature Sensor
Notification Hysteresis
Sound Infinite
Beep
ON
CF.EG28.N10.80.8
CF.EG39.N20.80.16
CF.EG44.N25.80.19
CF.EG48.N30.80.22
UF.Si.N30.150.10LV
HF�Si�20�200�50
HF�Si�20�250�50A
H2O (Water)
Customized 1
Customized 2
PUMP Limits Speed
Pressure
Maximum
Maximum
Minimum
4600 rpm
1000 rpm
0.7 bar
TIMER Set
Display
00:00 [hh:mm]
-
MODE A
C
B
D
-
-
-
activated
GRAPH Automatic
Manual Axis Assignment Set Temperature
Actual Internal Temp.
Actual External Temp.
X - Axis
Y - Axis
15 min
30 min
60 min
120 min
Axis Scaling
Maximum
Minimum
-
-
-
activated
activated
activated
activated
activated
40 ºC
5 ºC
PROGRAMS Program 1
...
Program 10
-
-
-
MODULES M1 - Valve
M2 - Output
M3 - Input Ext. Standby
Alarm
Switch
Error
Warning
ON
ON
Default Status
Default Status
Default Status
Open
Open
Open
Closed
Closed
Closed
ON
Default Status Open
Start
Closed
Stop
Refill
-
-
-
-
-
-
-
-
activated
activated
activated
activated
-
Level 2
Level 4
OUTGAS Speed
Interval
1000 rpm
10 s
SAFETY Time out
Password
Set
Speed
Temperature 30.00 °C
000
1000 rpm
00:30 [mm:ss]
DISPLAY Main Screen Pump
Timer
Device number
-
0
activated
Factory Settings
Menu
SETTING Languages
Display
Sound
Factory Settings
Communication
Information
English
Background
Brightness
Key Tone
Device name
Tempering
Safety
Intro
Bluetooth
Pump
Software
Firemware Update Info
Firmware Update Info
Black
White
Deutsch
... -
activated
activated
Circulator WiCo
-
-
-
yes
yes
yes
yes
yes
yes
-
activaed
Standard Mode
Battery Mode
100%
40%

18
Menu (Details):
TEMPERING:
1. Control:
1) Control Mode:
Internal (int): The temperature is regulated according to
the internal temperature sensor�
Extern (ext): The temperature is regulated according to the
external temperature sensor�
2) Control parameters:
Automatic: The optimal control parameters for PID tem-
perature control are determined automatically� This is the
recommended mode�
Selecting "Automatic" allows you to set the dynamics for
temperature control�
Ts must be adjusted to match the response characteristic
(total of all time constants) of the closed loop controlled
system, so that the control variable can deliver a uniform
and measurable change in the control deviation� Ts values
that are too small or too large can lead to instability of the
controller�
Prop_Bp: Proportional Band Maximum�
Prop_Bn: Proportional Band Minimum�
The Proportional Band is the range below (Prop_Bp) and
above (Prop_Bn) the set value in which the control output
value is calculated via the difference between the actual
and the set value and the PID parameters�
Examples of non-optimal settings:
3) Fluids:
Under the option "Fluids", a variety of fluids can be selected�
The selected fluid limits the setting range of the target
temperature� See table in the section “Fluids (Standard in-
formation for IKA fluid)“�
The maximum and minimum temperature values of the
selected fluid can be set within these limitations�
2. Limits:
Under the option "External (ext)", the maximum and mini-
mum temperature for external temperature control can be
set�
Under "l Δ T (int - ext) l", the maximum difference be-
tween the internal and external temperature can be set�
The function limits the absolute temperature difference
between the external system and internal bath during the
heating up and cooling down process� This can protects
sensitive device like the glass reactors from thermal shock
(e�g� glass break)�
°C
t
Fig. 26
Manual: The control parameters for PID temperature con-
trol can be set manually�
"Manual" should only be used in the event of special tem-
perature control requirements�
When "Manual" is selected, the following parameters can
be set for "Internal (int)" and "External (ext)" temperature
control:
Kp: Proportional coecient
The proportional coecient Kp is the controller amplifica-
tion and determines how strongly the control deviation
(the difference between the target temperature and actual
temperature) directly affects the control variable (on-time
of the cooling)� Kp-values that are too large can lead to
the controller overshooting�
Ti: Integral time
The integral time Ti (s) is the correction time and deter-
mines how strongly the duration of the control deviation
affects the control variable� Ti compensates for an existing
control deviation� A high Ti means a smaller and slower
effect on the control variable� Ti-values that are too small
can lead to instability of the controller�
Td: Differential time
The differential time Td (s) is the derivative time and de-
termines how strongly the rate of change of the control
deviation affects the control variable� Td compensates for
rapid control deviations� A high Td means a smaller and
slower effect on the control variable� Td-values that are
too large can lead to instability of the controller�
Ts: Sampling time
The sampling time Ts (s) is the time interval over which the
control deviation is determined and the respective control
variable (dependent on Kp, Ti and Td) is calculated�
Fig. 25 (cooling curve in “Automatic” mode)
°C
t
Kp too high
°C
t
Ti too low
°C
t
Kp too low
Ti too high
°C
t
Td too high

19
3. Temperature sensor:
1) Calibration:
The internal and external temperature measurement can be calibrated and adjusted�
You can select 2-point calibration or 3-point calibration for internal and external measurement�
Calibration proceeding (example: 2-point calibration):
> Tempering
Temperature Sensor
Calibration
Reset Calibration
> Temperature Sensor
Calibration
Internal
External
> Calibration
Internal
2 Point Calibration
3 Point Calibration
> Calibration > Internal
2 Point Calibration
Temperature sensor calibration
at
Step: 1 / 2
5.00 C
OK
> Calibration > Internal
2 Point Calibration
Temperature sensor calibration
at
Step: 1 / 2
30.00 C
OK
> Calibration > Internal
2 Point Calibration
Temperature sensor calibration
at
Step: 1 / 2
30.00 C
OK
> Calibration > Internal
2 Point Calibration
Target Temp: 30.00 C
Actual Temp: 25.00 C
Please wait...
Step: 1 / 2
> Calibration > Internal
2 Point Calibration
Target Temp: 30.00 C
Actual Temp: 30.00 C
Step: 1 / 2
OK
> Calibration > Internal
2 Point Calibration
Step: 1 / 2
Enter the temperature from the
measuring device
5.00 C
OK
> Calibration > Internal
2 Point Calibration
Step: 1 / 2
Enter the temperature from the
measuring device
32.00 C
OK
> Calibration > Internal
2 Point Calibration
Step: 1 / 2
Enter the temperature from the
measuring device
32.00 C
OK
> Calibration > Internal
2 Point Calibration
Sensor calibrated successfully
OK
Fig. 27
2) Reset Calibration:
By “Reset Calibration”, the calibration value for the internal or the external temperature sensor will be deleted�
4. Notification:
1) ON:
This menu option allows you to activate/deactivate the “Notification“ function� A green check mark and a beep inform-
ing when the target value (hysteresis) has been reached�
2) Hysteresis:
This menu option allows you to set the hysteresis from 0 to ±2�0 K�
3) Sound:
Infinite: Beep until you press the “Back” button�
Beep: Single beep�
PUMP
Limits:
In “Limits“ option, the user is allowed to set the maximum
and minimum speed to the pump, and also the maximum
pressure�
TIMER
1. Set:
The user can set a target time (duration)�
When device functions are started normally, this time is
displayed on the working screen� The device functions
stop automatically once this time has expired� The running
time is then displayed again on the display�
Note: To deactivate the target time specification, set the
target time to 00:00�
2. Display:
Activate the timer display on the main display (working
screen)�

20
MODE
1. Operating Mode A:
After power-on/power failure no automatic restart of
functions�
2. Operating Mode B:
After power-on/power failure automatic restart of func-
tions, depending on previous settings�
3. Operating mode C:
Set values (set in A or B) cannot be changed�
After power-on/power failure automatic restart of func-
tions, depending on previous settings�
4. Operating mode D:
Confirmation request for set value changes, if functions
are active� After power-on/power failure no automatic re-
start of functions�
DISPLAY
1. Main Screen:
In the “Main Screen“ menu option, you can specify what
information will be displayed on the screen�
2. Device Number:
In the “Device number“ menu option, you are allowed to
edit the device number from 0 to 99 that will be shown
on the main screen�
When using multiple units, it may be helpful to identify
WiCo and station�
Note: Labeling device number to the station�
GRAPH
In this menu, you can set the options for the time-temper-
ature diagram�
1. Automatic:
The scaling of the temperature axis (Y-axis) is automatically
determined, depending on the target temperature and the
actual internal and external temperature�
The time axis (X-axis) is permanently scaled to 30 minutes�
2. Manual:
1) Axis Assignment:
The temperature values to be displayed can be selected�
2) Axis scaling:
Scaling of the time (X) and temperature axis (Y) can be
selected or set�
PROGRAMS
Under programs, 10 user-defined temperature-time pro-
files can be created� A program can consist of up to 10
segments�
Once a program has been selected, the following options
are available:
1. Start:
Starts the program upon request of loop mode�
1) Infinite loop: Upon completion of the last segment,
the program continues with the first segment until the
user ends the program by stopping a device function�
2) Loop Count: Indicates the total number of loop cycles
until program end�
Note: At the end of the program all device functions are
switched off�
2. Edit:
Edit/change the selected program�
1) Seg No.: Segment number�
2) Ctrl. Sensor (int/ext): Determines whether control is
through the internal (int) or external (ext) temperature
sensor�
3) Temp.: Target temperature�
4) Ctrl.Mode (Time / +/- x.x K): In “Ctrl.Mode Time”
the target values and settings of the segment are valid
for the duration indicated in the column “Time hh:mm”�
Afterwards, the next program segment is automatically
executed�
In “Ctrl.Mode +/- x.xx K”, the hysteresis (tolerance) of
the actual temperature to the target temperature is set
(e�g� +/- 0�1 K)� The target values and settings of the seg-
ment are valid until the actual temperature reaches the
target temperature +/- hysteresis for the first time�
Afterwards, the next program segment is automatically
executed�
5) Pump rpm: Target speed of pump�
6) M1 (ON/OFF): MODULES M1-Valve:
OFF: M1 valve in initial state
ON: M1 valve in inverted initial state�
Note: The initial state of the M1 valve is defined in “MOD-
ULES M1-Valve” in “Default Status” as “Open” or “Closed”�
7) M2 (ON/OFF): MODULES M2-Output switch
OFF: M2-output switch in initial state�
ON: M2-output switch in inverted initial state�
Note: The initial state of the M2-Output switch is defined
in “MODULES M2-Output” in “Default Status” as “Open”
or “Closed”�
Other manuals for RC 5 control
1
Table of contents
Other IKA Controllers manuals