IMR EX440 User manual

R1
Environmental Equipment, Inc.
IMR EX440 Portable SMART Sensor Gas Detector
User Manual
Read this manual carefully before using this device.
(727)328-2818 / (800)RING-IMR • Fax: (727)328-2826
www.imrusa.com • email: inf[email protected]om

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 2of 21
Table of Contents
Table of Contents.................................................................................................................................................... 2
1. Introduction ..................................................................................................................................................... 4
2. Main Features and Specification ..................................................................................................................... 4
2.1 Main features............................................................................................................................................ 4
2.2 Specification............................................................................................................................................. 5
3. Structure and Function..................................................................................................................................... 6
3.1 Appearance............................................................................................................................................... 6
3.2 Display ..................................................................................................................................................... 7
3.3 Button Function........................................................................................................................................ 7
4. Operation ......................................................................................................................................................... 8
4.1 Power On.................................................................................................................................................. 8
4.2 Power Off................................................................................................................................................. 8
4.3 Stress Test ................................................................................................................................................ 8
4.4 Menu......................................................................................................................................................... 9
4.4.1 Basic Menu ....................................................................................................................................... 9
4.4.2 Advanced Menu.............................................................................................................................. 10
4.5 Check Status........................................................................................................................................... 10
4.6 Zero Translation..................................................................................................................................... 11
4.7 Calibration and Alarm Setup.................................................................................................................. 11
4.7.1 Zero Calibration.............................................................................................................................. 12
4.7.2 Span Calibration.............................................................................................................................. 13
4.7.3 Mixed Gas Calibration.................................................................................................................... 15
4.7.4 Alarm Level Setup.......................................................................................................................... 16
4.7.5 Channel Configuration.................................................................................................................... 17
4.7.6 Password Modification ................................................................................................................... 17
5. Battery ........................................................................................................................................................... 18
6. Sensor Replacement ...................................................................................................................................... 18
7. Common Faults and Solutions....................................................................................................................... 18
8. Warnings........................................................................................................................................................ 19

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 3of 21
9. Warranty........................................................................................................................................................ 20
10. Contact Information................................................................................................................................... 21

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 4of 21
1. Introduction
EX440 portable multi gas detector is an intrinsically safe instrument that can continuously detect the
concentration of combustible gases and toxic gases. It is designed for use in explosion proof, toxic gas leak
rescue, underground pipeline or mines and other hazardous locations. It effectively ensures the safety of staff
life without being invasive, and prevents loss of production equipment.
The detector uses natural diffusion as the method to detect gas. It is constructed with high strength engineering
plastics and composite with antiskid rubber housing. The detector is comfortable to hold, is water-proof, and
dust-proof.
2. Main Features and Specification
2.1 Main features
Adjustable alarm level, high alarm and low alarm
Adjustable calibration point
High concentration protection for LEL gas sensor
LEL sensor fault self-check
Low battery indication
Real-time clock
Smart interchangeable modular sensor
Auto calibration
Two level three types alarm(visual, audible, vibrate)
Data communication to PC
STEL and TWA alarm for toxic gas
Password management, to prevent unauthorized configuration
Intrinsically safe

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 5of 21
2.2 Specification
Testing method: Natural diffusion
Target gas: see appendix table 1
Response: CO, H2S, CH4 T90<30s;
Accuracy:CH4:±5% FS;
CO: absolute error:±5 PPM
Relative error: ±10%;
H2S:±5 ppm
Alarm method: Visual, audible, vibrate alarm
Working temperature: -20℃~50℃
Working humidity: <95%RH no dews
Power voltage: DC3.7V 2200mAh (Lithium battery)
Charging time: ≤4h;
Working time: ≥20h continuously (non-alarm status)
Ex classification: ExiaⅡCT4 Ga/ExibD21T4;
IP rate: IP66 ;
Dimension: 147mm×76mm×37mm(L×W×H);
Weight: about 350g.

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 6of 21
3. Structure and Function
3.1 Appearance
Front Calibration cover
Back
No.
Name
1
Alarm light
2
Buzzer
3
Sensor(s)
4
Button
5
LCD screen
6
Air intake
7
Calibration cover
8
Alligator clip
9
Label(s)
10
Charging contacts

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 7of 21
3.2 Display
3.3 Button Function
Button
Function
Power on, press for more than 3s
Cancel setup, press once
Calibration, in power off status, press and at the same time for more than 5s
Power off, press for more than 3s
Increase selected value, press
Enter selected item of setup, press
Enter setup menu, press and a the same time for more than 3s
Check the detector’s status, including temperature, time, STEL and TWA①, maximum
value ②, press
See next/different items, press
Confirm the set parameter, press
Turn off alarm sound and vibration in alarm status, press
Note: ①This function only available for toxic gases
②Oxygen gas has maximum expose value and minimum expose value
No.
Function
1
Calibration indicator
2
Gas name
3
High/Low alarm
4
Sound on/off
5
Battery voltage
6
Gas concentration
7
TWA(STEL)alarm
8
Gas unit of measure
1
2
3
5
6
7
8
4

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 8of 21
4. Operation
4.1 Power On
In power off status, press for more than 3s, the detector will be power on, and then enter self-test interface
as shown in below picture.
Self-test:
1) Power on beep, to test the buzzer
2) Vibration and light indication
3) After self-test completed, the detector will then cycle through the STEL and TWA set points for each
sensor
4) After warm up complete, the detector will enter working status, see as below picture.
4.2 Power Off
In working status, press and hold this button, the screen will show “shutting down…”, the buzzer will give
out sound, after 3s, and the detector will be power off.
4.3 Stress Test
Every day before you use the detector preform a “stress test” to make sure that the detector is functioning
correctly.
Test method: Power on, apply with target gas (gas concentration is higher than the detector’s high alarm), if
detector’s reading is correct, then it could be used. If detector displays error is out of bounds, please recalibrate
it. If detector does not have any response or display, please contact IMR.

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 9of 21
4.4 Menu
4.4.1 Basic Menu
In working status, press and to enter the menu setup interface, press to find different items, press
to enter selected item. After enter a certain item, press to change the figure, press to confirm the
figure, press can quit the item without saving.
Menu
Sub menu
Description
DATE/TIME
back
select or increase the value
save and exit
ALARM MODE
back
select
save and exit
RECORDGAP
back
change the figure
save and exit
CAUTION MESSAGE
back
change the figure
save and exit
BATTERY MESSAGE
back
save and exit
BACKLIGHT TIME
back
change the figure
save and exit

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 10 of 21
4.4.2 Advanced Menu
In the basic menu, press and at the same time twice, the detector will require password to enter next
interface. Press to increase the figure, press to confirm the input, after input password, press to
enter advanced menu, then press to change to see different items, press to enter selected item. The
advanced menu setup is shown as below.
Menu
Sub menu
Description
STAFF DEPLOY
back
change the setup
save and exit
ZERO CALABRATION
back
zero translation
DATA UPLOAD
back
select
save and exit
PERIPHIAL CONFIG
back
select
save and exit
POWER MANAGEMENT
back
select
save and exit
LANGUAGE SELECT
back
select
save and exit
Note: Default password for the advanced menu is: 0000
4.5 Check Status
In working status, press button once, the screen will in turn display current temperature, time, STEL①,
TWA①, maximum concentration and minimum concentration since power on etc.
Note: ①This function is only available for toxic gases.
②This function is only available for Oxygen.

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 11 of 21
4.6 Zero Translation
If the detector’s reading in clean air is not zero, then you can use this function to reset the reading to zero.
In working status, press and at the same time for more than 1s, the detector will require password input
to enter next interface, after input password, the detector will enter advanced menu, choose the function of
, then press button to do a zero calibration, the detector will indicate if this succeed or not, if zero
calibration is successful, there will be a “√” mark at the right side of the related gas name, otherwise, there will
be a “×” mark.
4.7 Calibration and Alarm Setup
Enter the setup interface.
In power off status, press and for more than 3s, the detector will power on and do a self-test, and then
the password will be required to continue, see picture below.
Press to change the figure (from 0 to 9). Press to change to the next digit. After last figure is completed
press to confirm the input. The detector will go into the next interface as shown below if the password was
correct.
Press to move the cursor. The icon will be black if is currently selected. Then press to enter the
selection.

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 12 of 21
4.7.1 Zero Calibration
Press to move the cursor to and then press there will be a countdown as shown below.
The detector will then go to the auto-zero calibration interface after counting down finishes. You can also press
twice to enter the auto-zero calibration interface directly, seen below.
If zero calibration is successful, there will be a ” √” mark at the right side of the related gas name, if it is not
successful, there will be a “×” mark, seen below.
In the auto-zero calibration interface, you can press to enter manual zero calibration as shown below.

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 13 of 21
The screen will display the zero point A/D value of each gas sensor channel, after the A/D values are stable.
Press and the detector will execute zero calibration. If successful there will be a ”√” mark otherwise it will
display an ”×”.
In the auto-zero calibration interface you can press to start the count down again, and after the countdown
finishes. The detector will enter span calibration. If you press twice before the countdown finishes the
detector will go to mixed gases calibration.
4.7.2 Span Calibration
The interface shown below will be displayed when you enter span calibration. It displays the calibration point.
If you do not want to do calibrate for this gas channel press and it will change to next gas channel.
If you need to change the calibration point; then press and the screen below will be displayed.
Press to move the selection. Press to change the figure. Press continuously to the last digit to
confirm the input and save; the detector will go to the next interface (waiting for gas input)

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 14 of 21
If there is gas input then the detector will enter auto span calibration as shown below.
While preforming an auto span calibration, you can press to go into manual span calibration as shown
below.
Press to confirm, and then if manual auto span calibration succeeds, the display below will be shown.
If manual calibration fails, the display below will be shown.

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 15 of 21
After span calibration is completed, the next gas channel can be calibrated. Follow the above instruction and
finish calibration for all of the gas channels. Finally, after all calibration is finished you will see the display
below for example.
4.7.3 Mixed Gas Calibration
During the count down after zero calibration has finished press then you will be in the mixed gas
calibration interface, shown below:
Press to change the selection. Press to select or unselect. A black dot means the sensor is enabled, and
no dot means the sensor is disabled. After gas channel selection is done press continuously to confirm the
selection. The detector will go to the next interface and will wait for gas input, shown below.
O2
C
O
NH3
CH4

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 16 of 21
If there is gas input. The corresponding gas channel will start auto calibration. After calibration is completed the
result will be shown below.
While waiting for gas input. If you press the interface will start a countdown. Then the interface will
change to span calibration screen. If you press again before the countdown finishes. It will again return
back to the mixed gas calibration interface.
While waiting for gas input, if you press the detector will return to working status, and the working
interface. The selected gas channel displays as normal, the unselected gas channels will display only.
4.7.4 Alarm Level Setup
Press to move the selection to . Then press to enter. Then the screen below will be shown.
Press to move the selection onto the numbers below “H”(High alarm), “L”(Low alarm), ST(STEL
alarm),TW(TWA alarm). The corresponding number will begin to flash. Press and the screen below will be
shown.
19.5
23.5

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 17 of 21
Press to change the figures, press to move the selection. After the last bit is set press and the value
will be saved. The detector will then enter next alarm setup.
4.7.5 Channel Configuration
Press to move the selection to . Then press to confirm and enter the next interface shown below.
Press to move the selection. The corresponding channel will start to flash. Press to select or unselect. A
black dot means the sensor is enabled, and no dot means the sensor is disabled. Continue pressing to save
all channels’ configuration and exit. After the detector returns to the working interface, the unselected gas
channel(s) will display only. The enabled gas channel(s) will display readings normally.
4.7.6 Password Modification
Press to move the selection to . Then press to confirm and enter. The screen below will be
displayed.
Press to change the password. Press to move the selection. After the last digit is entered press . The
new password will be saved and the detector will return to working status.

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 18 of 21
5. Battery
When there is low battery indication show on the screen, or the detector will not power on due to low voltage
charge the battery.
In power off situation, connect the charger head with AC 100~230V power. Then connect the charging cable
with the detector. The detector will power on automatically and display as charging status. Battery charging is
completed when the battery indication icon is full and not changing. Then you can unplug the charger and
disconnect the detector. Then the detector can be put into use.
Warning: It is not possible to power on the detector when charging the battery while powered off. Please do not
charge the battery in the field. In case the plugging/unplugging of the charger head may cause sparks and then
lead to fire or explosion. Charging the battery while the detector is powered on will have a reduced charging
speed.
6. Sensor Replacement
The detector uses smart modular sensors. Be aware of the sensor’s life and replace with new sensor regularly. It
is suggested to calibrate the sensor every half year in order to maintain accuracy.
7. Common Faults and Solutions
Faults
Possible reason
Solution
Unable to power on
Battery voltage too low
Charge the battery
System halted
Contact IMR
Circuit problem
Contact IMR
No response to target gas
Response delay
Wait for response
Circuit problem
Contact IMR
Reading’s not accurate
Sensor overdue
Contact IMR or replace the sensor
Long time no calibration
Calibrate the sensor
Time error
Battery voltage run out
Charge the battery and then set the
time
EMI
Set the time again
Zero translation is not available to
use
Sensor drift too much
Calibrate the sensor or replace the
sensor
Screen display“-0”
Sensor drift
Zero translation

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 19 of 21
8. Warnings
Do not drop the detector from high locations.
You may not be able to power on the detector in high concentration gas environments.
Please operate the detector only as outlined in this operation manual. Otherwise the detection results may
not be accurate. The detector may be damaged if used outside of the recommended guidelines.
The detector should not be used in environments that contain corrosive gases (such as High concentrations
of chlorine gas). Do not use in other harsh environment (such as over-high/low temperature, over-high
humidity, magnetic and intense daylight).
If the detector needs to be cleaned use a soft cloth dipped in water to wipe the detector gently. Do not use
corrosive solvents or abrasive substances to clean the detector. It may cause damage the detector.
It is prohibited to disassemble, replace or repair the detector privately.
It is prohibited to charge the detector or download data from the detector in hazardous area.
In order to ensure the detector’s accuracy. Calibrate the detector in regularly. The calibration frequency
should be less than one year.
Any application or operation errors beyond this manual. Contact IMR for further information.
The third sensor channel can be set as a Bias voltage channel; sensors with high power consumption should
not be used in this sensor channel.

®
Environmental Equipment, Inc.
IMR EX440 User Manual
3632 Central Ave. St. Petersburg, FL 33711 –Phone: 727-328-2818 / 800-RING-IMR –Fax: 727-328-2826 –E-mail: info@imrusa.com –Web: www.imrusa.com
THE PRIDE OF THE COMPANY –MADE IN THE USA
Page 20 of 21
9. Warranty
IMR Environmental Equipment, Inc. states the following:
IMR‚ as manufacturer hereby grants the following worldwide IMR warranty for an IMR analyzer purchased
from an authorized dealer.
1. The IMR warranty shall entitle every IMR customer to demand a free replacement or repair of the defective
parts from any IMR dealer authorized for the respective IMR unit.
2. The IMR warranty shall be granted on the factory new unit and shall commence on the date of the delivery
of the original IMR unit to the customer. It shall last for a period of twelve months regardless of the type
and the intensity of use and regardless of any change of owner, which may occur during this warranty
period.
3. The IMR warranty shall refer to absence of faults with respect to the state of the art nature of the sold unit in
terms of material and finish. The warranty for all parts fitted during the twelve-month warranty period shall
end with the unit warranty.
4. After the establishment of a material or production fault by IMR or the authorized IMR dealer, the faults
will be eliminated by means of free repair or replacement. Replaced parts shall become the property of IMR.
5. No warranty claims may be made for maintenance and setting work, cleaning or other utility materials
required for the function of the unit and other wear parts unless they have a direct bearing on work
performed under the warranty.
6. The terms and conditions for the acknowledgement of this warranty shall be the presentation of the fully
completed warranty card, which must contain the confirmation from the authorized IMR dealer on its
delivery and, if applicable, the prescribed maintenance work.
7. The IMR warranty shall only be applicable if
7.1 The analyzer has been maintained in accordance with the instructions issued by the manufacturers and
the operating instructions by an authorized IMR dealer.
7.2 Only original IMR spare parts have been used for any repairs.
7.3 The unit has been used properly, the operating instructions observed and the unit has not been used for a
purpose other than the one for which it has been designed.
7.4 The IMR unit has been left in its original design and meets the original IMR specifications.
7.5 The fault is not due to external influences or use for a purpose other than the one for which it has been
designed.
7.6 Exclusively authorized IMR dealers have made repairs to the IMR unit.
7.7 The IMR unit has been sent to an authorized IMR dealer immediately after the fault was discovered.
8. Warranty time for the analyzer, including electrochemical sensors is 12 months.
Table of contents
Other IMR Gas Detector manuals
Popular Gas Detector manuals by other brands

GASTRON
GASTRON GPD-100 instruction manual

International Gas Detectors
International Gas Detectors POR-MULT user manual

Honeywell
Honeywell Touchpoint 4 quick start guide

RKI Instruments
RKI Instruments 35-3010RK-07 manual

Oceanus
Oceanus OC-903 operating manual

LEGRAND
LEGRAND Tynetec ZXT812 quick start guide