UE VANGUARD WirelessHART TCD50H1A User manual

WirelessHART®Toxic & Combustible Gas Detector
P/N: TCD50H1A
Installation and Maintenance Instructions
IM_TCD50-05
6/15/2017

WirelessHART Toxic & Combustible Gas Detector IM_TCD50-05
www.ueonline.com/vanguard
i
This page is intentionally left blank

WirelessHART Toxic & Combustible Gas DetectorIM_TCD50-05
www.ueonline.com/vanguard
ii
United Electric Controls provides these installation and maintenance instructions for the UE Vanguard WirelessHART toxic &
combustible gas detector solely. The end-user is responsible for maintaining the Vanguard WirelessHART toxic & combustible gas
detector in full, operational condition. All warnings must be thoroughly read, understood and followed.
Misuse of this device and failure to adhere to all warnings could result in:
• the device not performing as intended,
• risk of explosion,
• risk of losing consciousness or death.
The following symbols are used within this document:
International Hazard Warning symbol. Directions require strict adherence.
International Toxic Hazard symbol. Exposure may lead to loss of consciousness or death.
International Flammable Material Hazard symbol. Exposure to flammable materials in closed areas may lead to explosion,
loss of consciousness or death.
United Electric Controls reserves the right to change published specifications and designs without prior notice.
United Electric Controls Co.
180 Dexter Avenue
Watertown, MA
02472, USA
Telephone: +1 617 926-1000
Fax: +1 617 926-2568
Email: Tec[email protected]
Web: www.UEonline.com
8

WirelessHART Toxic & Combustible Gas Detector IM_TCD50-05
www.ueonline.com/vanguard
iii
Please read all instructional literature carefully and thoroughly before starting
Table of Contents
1.0 General
1.1 Device Overview..........................................................................1-2
1.2 Device Packaging........................................................................3
1.3 Device Approvals...................................................................... .4-5
2.0 Installation
2.1 Battery Installation......................................................................6-7
2.2 Sensor Installation...................................................................... 8-9
2.3 Sensor Placement Considerations.............................................. 9-10
2.4 Sensor Cross Sensitivity.............................................................. 10-11
2.5 Classified Area Installation..........................................................11
2.6 Field Mounting............................................................................ 12-14
2.7 Antenna Placement Considerations............................................ 15
2.8 Network Configuration................................................................ 15-18
3.0 Operation
3.1 Normal Operation....................................................................... 19
3.2 Sensor Calibration & Bump Test................................................. 20-21
3.3 Programming Mode....................................................................21-26
4.0 Field Replacement
4.1 Sensor Replacement...................................................................27
4.2 Battery Replacement.................................................................. 28
5.0 Troubleshooting Guide
5.1 Hardware.................................................................................... 29
5.2 Wireless Communication............................................................. 30
6.0 Dimensional Drawings.......................................................................31
7.0 Appendix
7.1 List of Figures.............................................................................32
7.2 List of Tables...............................................................................33
7.3 List of Graphs..............................................................................33
7.4 List of Device Warnings (French Language)...............................34-37

WirelessHART Toxic & Combustible Gas Detector
IM_TCD50-05
www.ueonline.com/vanguard
1
1.0 GENERAL
This device is for monitoring and communicating gas concentration levels for data collection or
record keeping and does not provide alarm signals.
Misuse of this device may cause explosion and/or personal injury. These instructions must be
thoroughly read and understood prior to installation of the device.
The end-user is responsible for maintaining the device in full, operational condition.
Hydrogen Sulfide (H2S) is an extremely toxic gas, and exposure may result in
loss of consciousness or death.
Methane (CH4) is a greenhouse gas and is extremely flammable. Exposure in closed areas
may cause explosion, and result in loss of consciousness or death.
Avoid dropping the device as loosening of the threaded flame-proof joints or permanent damage
may occur.
New sensors must be calibrated after installation (see Section 3.2).
Hydrogen Sulfide (H2S) sensors require a warm-up period. Do not calibrate a new Hydrogen
Sulfide (H2S) sensor immediately after installation or if the device had been powered down (see
Section 2.2).
When not in use, the device should be stored in a clean, dry area and within the temperature
range listed within the device’s environmental specifications.
1.1 Device Overview
The device is an explosion-proof and intrinsically safe, WirelessHART point gas detector used for the detection and monitoring of
harmful gases in air, in order to help maintain a safe environment. A field-interchangeable gas sensor module (see Figure 1) provides
detection of hydrogen sulfide (H2S) or methane (CH4) gas. Gas measurement readings in parts per million (ppm) or lower explosive
limit (%LEL) are communicated along with network and battery status (see Figure 2) via a local, digital display, and WirelessHART 7.2
communication protocol. The device seamlessly integrates with existing supervisory control and data acquisition (SCADA) or asset
management (AMS) systems.

WirelessHART Toxic & Combustible Gas Detector IM_TCD50-05
www.ueonline.com/vanguard
2
Heavy duty, explosion-proof
antenna
Explosion-proof (Ex d)
compartment
Integral mounting bracket
Graphic display shows gas
concentration, network and
battery status
Intrinsically Safe battery
compartment. Battery is hot
swappable in HazLoc areas
Large shrouded button allows access to
calibration mode and bump testing
Battery allows one handed,
quick-change replacement
FLEXsense™ sensor is intrinsically
safe and hot swappable in HazLoc
areas. Toxic or combustible gas
sensors automatically configure
when interchanged.
WiredHART interface for
configuration and testing
Figure 1
Figure 2

WirelessHART Toxic & Combustible Gas Detector
IM_TCD50-05
www.ueonline.com/vanguard
3
1.2 Device Packaging
The device may be shipped in one of the following two configurations. Please ensure that you have all items before proceeding with
installation and operation.
1. Base unit and mounting bracket kit with toxic or combustible FLEXsense™ sensor (Figure 3 and Figure 4).
2. Base unit and mounting bracket kit only (Figure 3 without sensor and Figure 4).
Device is shipped without the sensor or battery installed. Both must be installed prior to use. See
Section 2.0 Installation for further information.
1.2.1 Sensors
One FLEXsense™ sensor is shipped (but uninstalled) with the base unit (if ordered in that configuration). Additional, spare
FLEXsense™ sensors may be ordered and will ship separately.
1.2.2 Battery
The battery pack is shipped separately from the base unit as it contains lithium which is regulated in transportation by the U.S.
Department of Transportation, the International Air Transport Association, International Civil Aviation Organization, and European
Ground Transportation of Dangerous Goods.
FLEXsense™ Sensor
Base Unit
Mounting
Bracket Kit
Figure 3 Figure 4
Top foam for packaging Bottom foam for packaging

WirelessHART Toxic & Combustible Gas Detector IM_TCD50-05
www.ueonline.com/vanguard
4
This device is suitable for Class I, Divisions 1 & 2, Groups C and D; or non-hazardous
locations only. -20 °C (-4 °F) ≤ Tamb. ≤ 60 °C* (140°F). Type 4X and IP64**.
This product is ATEX and IECEx certified for equipment category 2. Suitable for appropriate use
in gas zone 1.
The only field replaceable parts are the battery pack and FLEXsense™ sensor. Any substitution of
components other than the battery and gas sensor module will invalidate third-party issued ap-
provals and certifications, and may impair suitability for Class I, Division 1 location.
5003390
Global Certification
IECEx ETL 16.0016
Ex db ia [ia] IIB T4 Gb
-20°C < Tamb < +60°C*
EU Certification
II 2 (2) G Ex db ia [ia] IIB T4 Gb
ITS 16 ATEX 18474
-20°C < Tamb < +60°C*
North American Certification
Class I Div. 1, Groups C & D, T4
USA: Class I, Zone 1, AEx db ia [ia] IIB T4 Gb
Enclosure Type 4X**
Canada: Class I, Zone 1, Ex db ia [ia] IIB T4 Gb
-20°C < Ta < +60°C*
*Ambient operating temperature range may vary with sensor type. See
Technical Brochure for sensor specifications.
**Enclosure type 4X and IP64 rating excludes sensor.
UE declarations and third-party issued agency certifications are available for download at: www.ueonline.com/prod_approval.
1.3 Device Approvals
5003390

WirelessHART Toxic & Combustible Gas Detector
IM_TCD50-05
www.ueonline.com/vanguard
5
Standard Issue Date Title
Conform to IEC 60079-0:2011 2011/06/22 (6th Edition) Explosive Atmospheres - Part 0: Equipment
General Requirements
Conform to IEC 60079-1:2014 2014/06/27 (7th Edition) Explosive Atmospheres - Part 1: Equipment
Protection by Flameproof Enclosures “d”
Conform to IEC 60079-11:2011 2011/06/30 (6th Edition) Explosive Atmospheres - Part 11:
Equipment protection by intrinsic safety “i”
Conform to IEC 60086-4:2014 2014/09/03 (4th Edition) Primary batteries - Part 4: Safety of lithium
batteries
Conform to IEC 61010-1 2010 Electrical equipment for measurement,
control, and laboratory use; Part 1: General
requirements
Conform to EN 55011:2009+A1:2010 2010/01/31 Industrial, scientific and medical equipment.
Radio-frequency disturbance characteristics.
Limits and methods of measurement
Conform to EN 60079-0 2011/06/22 (6th Edition) Explosive Atmospheres - Part 0: Equipment
General Requirements
Conform to EN 60079-1:2014 2014/06/27 (7th Edition) Explosive Atmospheres - Part 1: Equipment
Protection by Flameproof Enclosures “d”
Conform to EN 60079-11:2012 2012/01/31 (6th Edition) Explosive Atmospheres - Part 11:
Equipment Protection by intrinsic safety “i”
Conform to EN 61326-1:2013 2013/02/28 Electrical equipment for measurement,
control and laboratory use. EMC
requirements. General requirements
Certified to CSA C22. 2 60079-0 2011/12/01 Explosive Atmospheres Part 0: Equipment
General requirements
Certified to CSA C22.2 60079-1 2011/12/01 Explosive Atmospheres Part 11: Equipment
protection by intrinsic safety “d”
Certified to CSA C22.2 60079-11 2011/12/01 Explosive Atmospheres Part 11: Equipment
protection by intrinsic safety “i”
Certified to CSA C22.2 61010-1-12 2012/5/11 Electrical equipment for measurement,
control, and laboratory use; Part 1: General
requirements
Conform to UL 60079-1 2009/04/10 (Edition 6) UL standard for safety electrical apparatus
for explosive gas atmospheres Part 1:
Flameproof Enclosures “d”
Conform to UL 913 2013/12/06 (Edition 8) Intrinsically safe apparatus and associated
apparatus for use in Class I, II and III,
Division I Hazardous (Classified) Locations
Conform to UL 1203 2013/11/22 (Edition 5) Explosion Proof and Dust Ignition proof
electrical equipment for use in hazardous
(Classified) Locations.
Conform to UL 61010-1 2012/5/11 (Edition 3) Electrical equipment for measurement,
control, and laboratory use; Part 1: General
requirements
FCC Rule Part 15b Declaration of Conformity (DoC) &
Verification
The device has been certified in accordance with the applicable requirements of the following standards:
Table 1

WirelessHART Toxic & Combustible Gas Detector IM_TCD50-05
www.ueonline.com/vanguard
6
2.0 INSTALLATION
2.1 Battery Installation
Only battery packs provided by United Electric Controls are compatible with this device.
Battery should be removed and stored when the device is not in use.
Batteries are not rechargeable.
The battery pack contains two Lithium-Thionyl Chloride D size cells, with a nominal output voltage of 7.2V.
1. Unscrew the solid, back cover from the intrinsically safe battery compartment and set aside.
2. Remove the battery pack from its packaging.
3. Align battery pack so connection on the pack lines up with connection in the enclosure.
4. Slide the battery pack into the enclosure until it latches in place as shown in Figure 5.
5. The device is now powered and will show the word STARTING along with the hardware and software versions on the front panel
display while initial boot-up completes (see Figure 6).
STARTING
HW VER : X.XX
SW VER : X.XX
Figure 5
Figure 6

WirelessHART Toxic & Combustible Gas Detector
IM_TCD50-05
www.ueonline.com/vanguard
7
6. When hardware and software boot-up is complete, the device will prompt user to activate the installed sensor (see Figure 7)
Refer to section 2.2 for sensor activation sequence. All sensors must be activated after a battery or sensor installation.
7. Once activation is done, the front panel display will display the device name, HART long tag, and the software version, along with
an instruction to push button to show gas concentration (see Figure 8).
8. The device is now ready to be configured for deployment into a WirelessHart network. Refer to section 2.8 for network
configuration.
9. Once device configuration is complete, secure cover onto enclosure by tightening until all threads have been fully engaged to
ensure proper ingress protection.
VANGUARD
SENSOR NOT
ACTIVATED : H2S
TAG: KANBAN XXXXXXX
PUSH BUTTON TO
ACTIVATE SENSOR
Figure 7
SW VER: X.XX
VANGUARD
SENSOR: H2S
TAG : KANBAN XXXXXX
PUSH BUTTON TO
SHOW CONCENTRATION Figure 8

WirelessHART Toxic & Combustible Gas Detector IM_TCD50-05
www.ueonline.com/vanguard
8
2.2 Sensor Installation
Only FLEXsense™ sensors provided by United Electric Controls are compatible with this device.
Do not attempt to disassemble the sensor as this could damage the sensor and void hazardous location
approvals, and manufacturer warranty.
Sensors must be activated after installation.
Hydrogen Sulfide (H2S) sensors require a warm-up period and must be allowed to stabilize prior to sensor
calibration. Calibration is not recommended at temperatures above 35°C (95°F).
Sensors are packaged fully assembled and require simple installation into the base unit. Toxic (H2S) sensors are comprised of an electrochemical
sensor packaged into a stainless steel housing. Combustible (CH4) sensors are comprised of an infrared sensor packaged into a stainless steel
housing. These are consumable, field replaceable parts (see Section 4.1, Sensor Replacement).
1. Remove the protective end cap cover from the sensor housing on the base unit.
2. Remove FLEXsense™ sensor from packaging.
3. Align the locating pins (see Figure 9) on the FLEXsense™ sensor to the keyed positions on the sensor housing (see Figure 10).
4. Insert the sensor. You will feel and hear a click that ensures the sensor has been inserted correctly.
5. Push button to activate sensor.
For Methane Sensor Only
6. Once the button is pushed, the base transmitter initializes the sensor activation process. During this activation, the screen in Figure 11 will be
displayed. This activation may take up to 2 minutes.
7. If the device does not detect a starting offset concentration for the methane sensor, skip to step 9.
Figure 9 Figure 10
Locating pins
Figure 11
ACTIVATING
PLEASE
WAIT

WirelessHART Toxic & Combustible Gas Detector
IM_TCD50-05
www.ueonline.com/vanguard
9
2.3 Sensor Placement Considerations
Hydrogen Sulfide (H2S) is an extremely toxic gas, and exposure may result in loss of consciousness
or death.
Methane (CH4) is a greenhouse gas and is extremely flammable. Exposure in closed areas may
cause explosion, and result in loss of consciousness or death.
This device must be mounted vertically with the sensor pointing downward.
This device is suitable for Class I, Divisions 1 & 2, Groups C and D; or non-hazardous locations only.
-20 °C (-4 °F) ≤ Tamb. ≤ 60 °C* (140°F). Enclosure Type 4X and IP64**
8. If device detects a starting offset concentration for the methane sensor, the device will prompt the user to zero the offset. As shown in Figure
12.
A. To zero the offset, press and hold the calibration button until you see the following display in Figure 13. This takes about 10 seconds.
B. To ignore the offset, press and release the calibration button to skip the zeroing process to step 9.
For All Sensors
9. Once the sensor initialization process is complete, a “Sensor Active” message will show (see Figure 14). The message ‘Calibrate New
Sensors’ (see Figure 15) automatically follows to prompt user to perform a sensor calibration, before returning to home screen (Figure 8).
10. The sensor is self-configuring with the device and once booted up (see Section 2.8 Network Configuration) will display the proper gas con-
centration in ppm or %LEL.
NOTE: In the case of the H2S sensor, there is an associated warm up period to allow for sensor stabilization to 0ppm. Refer to table 2 below for
H2S stabilization time.
NOTE: If the sensor or battery are momentarily removed from the device, and the sensor had been powered prior to the interruption, the warm-
up period does not apply.
NOTE: To zero the methane sensor at any time, remove the sensor from the base transmitter. Wait until the display indicates a ‘Sensor Missing’
message. Reinstall the sensor and follow steps 6 through 9.
Temperature Recommended Stabilization Time
-20°C (-4°F) .5 hours
22°C (72°F) 6 hours
35°C (95°F) 24 hours
Figure 13
ZEROING DONE
SENSOR ACTIVE
SENSOR
ACTIVE
CALIBRATE
NEW
SENSORS
Figure 14 Figure 15
Figure 12
CH4 XX%.
PRESS AND HOLD
BUTTON TO ZERO SENSOR
OR PUSH TO SKIP
Table 2

WirelessHART Toxic & Combustible Gas Detector IM_TCD50-05
www.ueonline.com/vanguard
10
Chemical Abstracts Service (CAS), a division of the American Chemical Society, is the world’s authority for chemical information. Please consult
their database at www.cas.org for chemical properties of gas.
Other factors to consider include but are not limited to:
• Location – takes into account gas leak source, pressure of release, temperature, and distance to sensor.
• Air flow – prevailing winds, velocity, fans, exhaust ducts that could affect detection.
• Environmental conditions – other than air flow; such as exposure to water, snow, and dirt. Also temperature – hot and cold – and how that
may affect the gas release.
• Maintenance Accessibility – provide safe access for calibration and maintenance.
2.4 Sensor Cross Sensitivity
Depending on the type of sensor installed, the device has the ability to detect either Toxic (H2S) or Combustible (CH4) gas. The following table
and graph demonstrate the typical cross sensitivities to other gases that may be present while monitoring Hydrogen Sulfide or Methane gas.
Gas ppm Tested Signal
(as ppm H2S)
Carbon Monoxide 400 1.3
Hydrogen Sulfide 25 25.0
Ozone 5 -0.6
Nitrogen Dioxide 10 0.2
Sulfur Dioxide 20 3.5
Ethanol 200 -0.3
NO 50 15.9
Chlorine 10 -0.5
n-Heptane 500 -0.1
Ammonia 100 0.0
Methane 500 0.1
2.4.1 Typical Cross Sensitivity of Hydrogen Sulde (H2S)
Table 3
This device is ATEX and IECEx certified for equipment category 2. Suitable for appropriate use in
gas zone 1.
*Ambient operating temperature range may vary with sensor type. See Technical Brochure for sensor specifications.
**Enclosure type 4X and IP64 rating excludes sensor.
Individual applications need to be assessed for correct sensor placement. The target gas needs to be determined along with the gas density, and
potential coverage area.
For gases that are heavier than air, sensor installation is recommended near the ground and below the
potential leak source.
For gases that are lighter than air, sensor installation is recommended near the ceiling (in an enclosed
space) and above the potential leak source.

WirelessHART Toxic & Combustible Gas Detector
IM_TCD50-05
www.ueonline.com/vanguard
11
2.4.2 Typical Cross Sensitivity of Methane (CH4)
Graph 1
2.5 Classied Area Installation
This device is suitable for Class I, Divisions 1 & 2, Groups C and D; or non-hazardous locations only.
-20 °C (-4 °F) ≤ Tamb. ≤ 60 °C* (140°F). Enclosure Type 4X and IP64**.
This device is ATEX and IECEx certified for equipment category 2. Suitable for appropriate use in
gas zone 1.
Install device only into areas matching approvals on device labeling.
Do not remove the front (painted aluminum and glass) cover in hazardous area.
Install the device where shock, vibration and temperature fluctuations are minimal. Do not mount in
ambient temperatures that exceed the limits on the device labeling.
This device is for monitoring and communicating gas concentration levels for data collection or record
keeping and does not provide any alarm signals.
To prevent seizure of enclosure cover, do not remove lubricant. Threads should also be free of dirt, etc.
*Ambient operating temperature range may vary with sensor type. See Technical Brochure for sensor specifications.
**Enclosure type 4X and IP64 rating excludes sensor.
Methane
Propane
Ethane
Butane
Hexane
Ethylene

WirelessHART Toxic & Combustible Gas Detector IM_TCD50-05
www.ueonline.com/vanguard
12
2.6 Field Mounting
This device must be mounted vertically with the sensor pointing downward.
Install device where shock, vibration and temperature fluctuations are minimal. Do not mount in ambient
temperatures that exceed the limits on the device labeling.
The mounting bracket attachment on this device is factory-installed and must not be removed as it may
cause permanent damage to the antenna, the sensor assembly, and loosening of the threaded flame-proof
joints. Removal will void hazardous location approvals, and manufacturer warranty.
This device may be mounted to either a vertical or horizontal pipe, or wall (flat surface) using factory-installed mounting bracket and
included bracket hardware. The maximum pipe size for pipe mounting is 2”.
2.6.1 Vertical Pipe Mounting
1. Remove hardware from packaging (see Figure 16).
2. Place one pipe mount adaptor [1] onto desired location horizontally on pipe and slide U-bolt [2] through holes.
3. Slide mounting bracket plate [3] onto U-bolt [2] and secure with (2x) hex nuts [4]. Do not tighten.
4. Repeat step 2 with second pipe mount adaptor [1] also sliding U-bolt [2] through mounting bracket plate [3] and securing with
(2x) hex nuts [4].
5. Once position is set, tighten (4x) hex nuts using 1/2” wrench.
6. Mount device by hooking onto top pin of mounting bracket plate (see Figure 17).
7. Push attached pin [5] through bottom holes to secure device to the bracket (see Figure 18).
Figure 16 Figure 18Figure 17

WirelessHART Toxic & Combustible Gas Detector
IM_TCD50-05
www.ueonline.com/vanguard
13
2.6.2 Horizontal Pipe Mounting
1. Remove hardware from packaging (see Figure 16).
2. Place one pipe mount adaptor [1] onto desired location vertically on pipe and slide U-bolt [2] through holes.
3. Slide mounting bracket plate [3] onto U-bolt [2] and secure with (2x) hex nuts [4]. Do not tighten.
4. Repeat step 2 with second pipe mount adaptor [1] also sliding U-bolt [2] through mounting bracket plate [3] and securing with
(2x) hex nuts [4].
5. Once position is set, tighten (4x) hex nuts using 1/2” wrench.
6. Mount device by hooking onto top pin of mounting bracket plate (see Figure 19).
7. Push attached pin [5] through bottom holes to secure device to the bracket (see Figure 20).
Figure 19 Figure 20

WirelessHART Toxic & Combustible Gas Detector IM_TCD50-05
www.ueonline.com/vanguard
14
2.6.3 Wall Mounting
Wall mounting could affect the signal strength of the device
1. Remove hardware from packaging (see Figure 21).
2. Hold mounting bracket plate [3] against wall/flat surface and screw tightly into place using (4x) mounting holes (0.40” dia.).
Note: Mounting screws [6] not UE provided.
3. Mount device by hooking onto top pin of mounting bracket plate (see Figure 22).
4. Push attached pin [5] through bottom holes to secure device to the bracket (see Figure 23).
2.6.4 Device removal from mounting bracket plate
Removing the device from the mounting bracket plate for maintenance or calibration is simple.
1. Pull attached pin [5] through bottom holes using the pull ring.
2. Pull device up and unhook from mounting bracket plate [3].
Figure 21 Figure 22 Figure 23

WirelessHART Toxic & Combustible Gas Detector
IM_TCD50-05
www.ueonline.com/vanguard
15
2.7 Antenna Placement Considerations
This device complies with Part 15 of the FCC rules: may not cause harmful interference; must accept any
interference, including that which may cause undesired operation of the device.
To comply with IC RF exposure compliance requirements, the antenna used for this device must be
installed to provide a separation distance of at least 20cm from all persons.
FCC ID SJC-ETERNA2 contains IC 5853(A)-ETERNA2 CE0976
Wall mounting could affect the signal strength of the device.
The antenna supplied with the device is an omnidirectional antenna, permanently affixed to the product. In order to facilitate
communication with other WirelessHART devices as part of a wireless mesh network, it is recommended that the device be mounted
with a minimum of obstructions and noise interference. The maximum range of communication between the device and other devices
(instrumentation, repeaters, gateways) is 800 feet (244 meters).
2.8 Network Conguration
A Network ID and Join Key are required to communicate over the wireless network.
It is recommended that network configuration be performed prior to installation when applicable. The device must be configured to
communicate over the wireless network. A Network ID and Join Key are required, and must match the Network ID and Join Key of the
wireless Gateway and other wireless devices in the network.
This device sends critical gas concentration information reliably and securely via WirelessHART communication protocol, version 7.2.
Configuration can be completed using any HART handheld communicator or directly into your control system using a HART modem.
HART terminals are located inside the intrinsically safe battery compartment and are clearly marked (see Figure 24).
Configuring the gas detection device in the hazardous area requires a suitable, approved intrinsically safe
device (such as a handheld communicator, etc.). The entity parameters of the device being connected must
be compatible with those of the gas detection device.
Before connecting any device in the hazardous location, the following entity parameters must be verified:
Uo < Ui prog Lo > Li prog
Io < Ii prog Ui > Uo prog
Po < Pi prog Ii > Io prog
Co > Ci prog
Gas Detector Intrinsic Safety Entity Parameters:
Uo = 4.935V; Io = 0.179A; Po = 0.22W; Co = 1000uF; Lo = 10mH
Ui = 3.8V; Ii = 64uA; Ci = Negligible; Li = Negligible
Figure 23

WirelessHART Toxic & Combustible Gas Detector IM_TCD50-05
www.ueonline.com/vanguard
16
Connected Device Entity Parameters (will vary dependent upon device):
Ui prog = Max. voltage that can be connected at the input to the handheld or other programming device
Ii prog = Max. current that can be connected at the input to the handheld or other programming device
Pi prog = Max. power that can be connected at the input to the handheld or other programming device
Li prog = Max. inductance that can be connected at the input to the handheld or other programming device
Ci prog = Max. capacitance that can be connected at the input to the handheld or other programming device
Uo prog = Max. voltage that the handheld or other programming device can source to the device being
programmed
Io prog = Max. current that the handheld or other programming device can source to the device being programmed
2.8.1 Download Device Description (DD)
Download the latest DD at www.ueonline.com/vanguard/hart to ensure optimum communication between HART devices, gateway and
host system.
Refer to www.ueonline.com/vanguard/hart for instructions on how to download the DD into the 475 Field Communicator using the
Easy Upgrade Utility.
Note: HART Registration of this device is pending. When achieved, you will also be able to download the latest DD at
www.hartcomm.org.
2.8.2 Using a Handheld HART Communicator
1. If not previously done (from installing the battery – see Section 2.1 Battery Installation), unscrew the solid, back cover from the
intrinsically safe battery compartment and set aside.
2. Confirm battery installation or install battery (see Section 2.1 Battery Installation).
3. Push the button on the side of the device to activate the wired HART interface.
4. Connect wires from handheld HART communicator to HART terminals (see Figure 24).
5. The handheld HART communicator will search for a HART-compatible device and indicate if a connection was made.
6. Obtain the Network ID and Join Key of the wireless network you want the device to join. This information can be found under
the Settings section on the web server supporting the wireless Gateway. Consult your wireless Gateway manual for additional
information.
7. Navigating through the handheld HART communicator menu, choose Configure Network ID and enter the Network ID. You will use
WirelessHART commands 773-Write Network ID and 774-Read Network ID.
Note: by using *Quick Code 9, 2, 1 on the handheld HART communicator, you will be taken directly to the Configure Network ID
choice.
*Quick Codes may not work with all handheld HART communicators.
Figure 24
Wired HART
interface
Table of contents
Other UE Gas Detector manuals
Popular Gas Detector manuals by other brands

AIR SYSTEMS
AIR SYSTEMS CO-91 Series Calibration guide

Honeywell
Honeywell Midas user manual

Critical Environment Technologies
Critical Environment Technologies CGAS-A Series Installation & operation manual

GASTRON
GASTRON GTD-5000F instruction manual

Dräger
Dräger e-Box Polytron 5000 Assembly instructions

SST
SST GT814 Design manual