IntelliSense FG-1508 User manual

¶ Mount the detector. · Wire the detector.
¹ Test the detector.
¸ Configure the detector.
FG-1508 Glassbreak Detector and Magnetic Contact nstallation nstructions
•Align the SWITCH with the
MAGNET to determine the
correct orientation for mounting.
The glassbrea detector and
magnetic contact housings must
be aligned and mounted
parallel to each other.
Configure jumpers W1-W4 to best suit the application.
Note: Some product versions do not have all of the
jumpers.
•W1: Default: LEDs enabled. To disable LED opera-
tion, remove the jumper from position W1 on the PCB.
•W2: Default: alarm LED LATCH disabled. To enable
alarm LED LATCH, place a jumper across W2.
•W3: Default: magnetic contact enabled. To disable
the magnetic contact, place a jumper across W3.
•W4: Remove jumper for high sensitivity. Place jumper
across W4 for low sensitivity. Verify range by testing.
Note: Test the FG-1508 at least once each year.
Test the detector with the FG-701 Glassbrea
Simulator.
Activating Test Mode
1Position the simulator within 5 (1.5 m) of the
FG-1508 detector.
2Switch the FG-701 simulator to ACTIVATE and
MANual modes.
3Point the front of the simulator at the detector and
press the red start button.
The simulator buzzes, and the green LED on the
FG-1508 flashes once per second to indicate it is in
Test Mode.
If an FG-701 is not available, or if for any reason
remote activation cannot be used, use a small
screwdriver to short the test pads near location W4 on
the PCB. This activates Test Mode.
Note: In Test Mode the LED disable jumper is
overridden.
Ma e sure to replace the front cover of the FG-1508
before testing.
Testing the FG-1508
1Set the FG-701 switches to the TEST and FLEX
positions.
2Press the red start button. The simulator clic s
on and starts an eight second armed period.
3Position the FG-701 near the farthest point of the
glass and point it directly at the FG-1508. If
window coverings are present, close them fully
and hold the FG-701 behind them.
4Generate a flex signal by carefully stri ing the
glass with a cushioned tool. The FG-701 responds
by producing a burst of glassbrea audio.
If both the flex and audio are received properly, the
red alarm LED on the FG-1508 lights.
MPORTANT: Some environmental factors may
reduce the detector activation range. If you do not see
the green LED flashing after pressing the red start
button, move closer to the detector and try again.
Testing the Audio Alone
You can also use the simulator in the MANual mode
to test audio alone. The blin ing green LED on the
detector flic ers when the simulator audio is received
correctly. (See the FG-701 Operating Instructions for
additional information.) Keep in mind that this is not
a complete test.
Exiting Test Mode
After testing, exit the Test Mode using the same
procedure for activating the Test Mode. The FG-1508
also automatically exits Test Mode five minutes after
the last glassbrea simulation.
Testing the Magnetic Contact ( f Applicable)
With the magnetic contact enabled, the detector
signals an alarm to the control panel when the
normally closed magnetic contact opens. The LED
on the detector does not indicate the alarm condi-
tion. You must verify an open contact at the panel. To
reset the alarm relay, you must close the magnetic
contact (the door or window on which it is mounted
must be closed).
LEDs are enabled
Red alarm LED latches ON
when detector alarms1,2
Magnetic contact is disabled
LEDs are disabled except
during power up test
Red alarm LED lights for 5
seconds when detector alarms
Magnetic contact is
enabled
JUMPER ON OFF
W1
(LED)
W2
(LATCH)
W3
(CONTACT
DISABLE)
1Latched alarm LED does not affect timing of alarm relay.
2Reset the alarm LED by removing and restoring power, or by entering and
exiting test mode.
3Always verify range of any sensitivity settings as described in the Test section.
Low sensitivity3High sensitivity, 8 max.3
W4
(SENSITIVITY)
= Default Settings
FLEX MAN
ACTIVATE TEST
Red Start Button
FG-701 Simulator -- Top View
C
NC
V-
V+
TMPR
TMPR
Breakout tab
(both ends)
PCB wire entry
Terminal Block
W4 (Sensitivity)
W2 (Latch)
W3 (Contact Disable)
W1 (LED)
Test Pads
•Route the wire through the wire entry hole in the PCB
housing and reconnect the housing to the mounting plate.
•Cut and strip wire ends 1/4 (6.5 mm).
•Wire the detector using 18-22 AWG (1.02mm-0.64mm).
Reverse polarity connections will not damage the detector.
•Replace front cover.
For mounting guidelines, refer to the Supplemental Information on the bac page.
•Insert a screwdriver into the
slot at the end of the housing
and gently depress the front
cover latch. Lift and remove
the front cover.
•Separate the mounting plate
from the glassbrea housing.
Insert a screwdriver into the
slot at the top of the PCB,
slightly twist, and gently push
downward on the latch.
•To remove the front cover of
the magnetic contact housing,
insert a screwdriver into the
slot at the end of the bac
cover and gently depress the
front cover latch. Lift and
remove the front cover.
•Route the wires through the wiring channel.
Remove brea out tab and use an industry standard
4 (10 cm) tie wrap if a strain relief is required.
•Secure the mounting plate and magnetic contact
housing with #6 (3.5 mm) screws (not provided).
•To prepare for wiring, remove the appropriate
brea out tab in the PCB housing.
Magnet Switch
T P: Temporarily mount the detector in the intended
location and power it with a 9 V battery until testing
establishes effective range coverage. If the 9 V
battery is low, it cannot supply sufficient power, and
the detector will not operate.

FG-1508 Glassbreak Detector and Magnetic Contact Supplemental nformation
Protected glass:
Minimum size for all types is 11" (28 cm) square. Glass must be framed in
the wall of the room or mounted in a barrier of 36" (0.9 m) minimum width.
1 Protected only if both glass plates are bro en.
2 Coated glass with security films, including films for solar protection, up to 12
mil. thic may be used. Film Technologies, Inc.s GLASS GARD® GLL 1200 has
been evaluated with this product by Underwriters Laboratories Inc. at
IntelliSenses request.
3 Underwriters Laboratories of Canada has investigated the use of the coating
referred to on note 2 for plate and tempered glass.
MPORTANT: The FG-1508 must be connected to a UL listed power supply
or UL listed control unit capable of providing a minimum of four hours of
standby power.
LED ND CATORS
The two LEDs shining through the front cover indicate the detector's
operational status. The following table summarizes the LED opera-
tion when the LEDs are enabled.
REMOTE LED ENABLE/D SABLE MODE
The detectors Remote LED Enable/Disable Mode allows you to
enable or disable the detectors LEDs with the FG-701 Glassbrea
Simulator.
To enable or disable the LEDs with the FG-701:
1Position jumper W1 to OFF.
2Set the FG-701 switches to the ACTIVATE and MANual positions.
3Stand within 5 (1.5 m) of the detector, point the front of the
FG-701 at the detector, and press the red start button. The
simulator generates an audio activation signal, and the detectors
green LED flashes once per second.
4Repeat step 3, and the detectors green LED stops flashing.
5Within 2 seconds, repeat step 3 to send another activation code. If
you disable the LEDs, the detectors green LED flashes once. If
you enable the LEDs, the green LED flashes 2 to 3 times.
6Clap your hands to test the LEDs. If the LEDs are enabled, the
green LED flic ers. If the LEDs are disabled, the green LED
remains off.
MOUNT NG GU DEL NES
The FG-1508 detects framed glass bro en by an impact sufficient to
ma e a hole.
•Mount the FG-1508 (indoors only) on doors, window frames, walls,
ceilings, and in corners.
•Mount within 8 (2.4 m) of the glass.
•The detector must have a direct line-of-sight to the glass with no
obstructions.
•When wall mounting, mount the detector a minimum of 6 (1.8 m)
high.
•Do not mount within 3 (0.9 m) of forced air ducts, sirens, or bells
measuring 2 (5 cm) or more in diameter.
•If the magnetic contact is enabled, the glassbrea housing must
be mounted less than 1(2.5 cm) from the magnet housing.
Maximizing Detection
•Minimize range to the glass. Do not install beyond the maximum
specified range, even if testing indicates greater range.
•Mounting on freestanding posts and pillars is not recommended.
•The FG-1508 detects through window blinds and light drapes. For
heavy window coverings, mount the unit on the window frame
behind the covering. Test the location thoroughly whenever
window coverings are present.
•Verify all installations bac to the panel to be sure the protection
loop is intact.
Minimizing the Chance of False Alarms
•Avoid installing in rooms with high-level noise sources, such as air
compressors, bells, and power tools if those sources can be active
when the detector can signal an alarm.
•Magnetically susceptible materials affect the range of the contact
gap. If using the magnetic contact, be sure to test the gap thor-
oughly when the unit is mounted on iron or steel surfaces.
•Test false alarm immunity by activating any nown noise sources
in the room.
SELF-TESTS
The FG-1508 automatically performs a series of self-tests during
power-up. If any self-test fails, the detector signals trouble by
flashing the LEDs alternately once per second. Always return the
detector for repair if there is any indication of trouble, even if the
trouble is temporary.
SPEC F CAT ONS
Range:
8 (2.4 m) maximum
4 (1.2 m) maximum (
for lowest UL setting
)
No minimum range
Alarm relay:
Form A, 250 mA maximum
25 VDC maximum
Alarm duration:
5 seconds (unaffected by alarm LED
latching)
Tamper switch:
Cover tamper 25 mA maximum
24 VDC maximum
Power requirements:
6 - 18 VDC; 13 mA typical at
12 VDC, 22 mA max.;
AC Ripple: 4 Volts pea -to-pea at
nominal 12 VDC
Operating temperature:
14° to 122°F (-10° to 50°C)
Storage: -4° to 122° F (-20° to 50° C)
RF immunity:
30 V/m, 10 MHz - 1000 MHz
ESD immunity:
10 V, Discharges of either polarity to
exposed surfaces
Magnetic contact:
Ma e gap 0.9 (2.3 cm) minimum
Brea gap 2.2 (5.6 cm) maximum
Dimensions:
Glassbrea housing:
4-9/16 x 15/16 x 1-1/4
(11.6 cm x 2.3 cm x 3.2 cm)
Magnet housing:
4-9/16 x 15/16 x 1/2
(11.6 cm x 2.3 cm x 1.3 cm)
Weight:
2.9 oz., (81 g)
Pac aged product: 3.8 oz., (108 g)
Accessories:
FG-701 Glassbrea Simulator
Approvals/listings:
FCC / IC verified
CE
C-Tic
UL Listed
ULC Listed
FCC Notice: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference in a residential installation. This equipment generates, uses and can radiate
radio frequency energy, and if not installed and used in accordance with the instructions may cause harmful
interference to radio communications. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to radio or television reception,
which can be determined by turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures: 1) Reorient or relocate the receiving antenna, 2)
Increase the separation between the equipment and receiver, 3) Connect the equipment into an outlet on a
circuit different from that to which the receiver is connected. The installer can also consult an experienced
radio/television technician for additional suggestions, if necessary.
Normal OFF OFF
Normal, event detected Flic er OFF
Normal, event detected and Flic er Flash 5 seconds
alarm in memory
Normal, brea detected OFF ON 5 seconds
Normal, alarm latched OFF ON continuously
Power up self-test ON 1 second ON 1 second
Trouble detected Flash ON/OFF Flash OFF/ON
Low voltage Flash ON/OFF Flash ON/OFF
Test Mode Flash once per second OFF
Test Mode, event detected Flic er OFF
Test Mode, alarm Flash once per second ON 5 seconds
Condition Green LED Red LED
TH CKNESS
Type Minimum Maximum
Plate 3/32" (2 mm) 3/8" (10 mm)
Tempered 1/8" (3 mm) 3/8" (10 mm)
Laminated11/8" (3 mm) 9/16" (14 mm)
Wired 1/4" (6 mm) 1/4" (6 mm)
Coated2, 3 1/8" (3 mm) 1/4" (6 mm)
Sealed Insulating11/8" (3 mm) 1/4" (6 mm)
© 2001 IntelliSense®•IntelliSenseisaregistered trademark of Honeywell, Inc.
• All other trademarks are the properties of their respective owners. • All rights reserved.
5-051-626-01RevD
Table of contents
Other IntelliSense Security Sensor manuals

IntelliSense
IntelliSense FG-1525R Owner's manual

IntelliSense
IntelliSense FG-1025 User manual

IntelliSense
IntelliSense DUAL TEC DT-906 Series Owner's manual

IntelliSense
IntelliSense FlexGuard FG-1625SN-ADT User manual

IntelliSense
IntelliSense AWARE User manual

IntelliSense
IntelliSense FG-1008 User manual

IntelliSense
IntelliSense FG-1025 User manual
Popular Security Sensor manuals by other brands

Risco
Risco BWare BUS QUAD AM G3 quick guide

SilverSaver
SilverSaver 1000 quick start guide

Detection Systems
Detection Systems Security Escort SE88 Series user guide

Honeywell
Honeywell Miller Stopfall 7700A/YL/GP-1 User instruction manual

Eaton
Eaton Sure-Lites UX6 installation instructions

AccuLine
AccuLine 40-6700 instruction manual