Interacoustics EyeSeeCam User manual

Instructions for Use – Part 1
vHIT
EyeSeeCam
Science
made
smarter
D-0101515-E – 2019/03
EN Instructions for Use
ES Manual de instrucciones
SV Bruksanvisning
FR Instructions d’utilisation
EL Οδηγίες χρήσης
DE Gebrauchsanweisung
IT Istruzioni per l'uso
PT Instruções de Utilização


Instructions for Use – EN
EyeSeeCam 1.3


Table of Contents
1Introduction ............................................................................................................................. 1
1.1 About this manual........................................................................................................ 1
1.2 Intended use ................................................................................................................ 1
1.3 Included parts .............................................................................................................. 0
1.4 Notes on safety............................................................................................................ 0
2Unpacking and installation .................................................................................................... 1
2.1 Unpacking and inspection ........................................................................................... 1
2.2 Markings ...................................................................................................................... 2
2.3 General warnings and precautions.............................................................................. 2
3Getting started and system setup......................................................................................... 5
3.1 Hardware requirements ............................................................................................... 5
3.2 Power........................................................................................................................... 5
3.2.1 Maximizing the power supply............................................................................... 5
3.2.2 Entering BIOS ...................................................................................................... 5
3.2.3 Changing BIOS setting ........................................................................................ 5
3.2.4 Windows Power Setting Setup ............................................................................ 6
3.2.5 USB mode............................................................................................................ 7
3.3 Hardware setup ........................................................................................................... 7
3.3.1 Goggle setup........................................................................................................ 7
3.3.2 IR illumination ...................................................................................................... 8
3.3.3 Focus adjustment................................................................................................. 9
3.3.4 Calibration laser Adjustment................................................................................ 9
3.4 Servicing and maintenance ....................................................................................... 10
3.5 Software setup........................................................................................................... 10
3.5.1 Installing EyeSeeCam software......................................................................... 10
3.5.2 Uninstalling the EyeSeeCam vHIT software - ................................................... 10
3.5.3 Starting EyeSeeCam from OtoAccess® 1.5...................................................... 10
3.5.4 Starting EyeSeeCam from OtoAccess® 2.0...................................................... 11
3.5.5 Backing up files in OtoAccess®......................................................................... 11
4Test procedures .................................................................................................................... 13
4.1 Opening Screen......................................................................................................... 13
4.2 Eye calibrations ......................................................................................................... 13
4.2.1 Prepare the patient ............................................................................................ 13
4.2.2 Adjust the eye/camera ....................................................................................... 14
4.2.3 Calibrate the patient........................................................................................... 15
4.3 Head calibrations....................................................................................................... 17
4.3.1 Prepare Patient .................................................................................................. 17
4.3.2 Start Head calibration ........................................................................................ 17
4.3.3 Calibration report – Head................................................................................... 18

4.3.4 Edit and Default Calibrations ............................................................................. 19
4.4 Lateral impulses ........................................................................................................ 20
4.4.1 Training guide .................................................................................................... 20
4.4.2 3D Head Modeling ............................................................................................. 21
4.4.3 Display laser during HIT..................................................................................... 22
4.4.4 Reject Noisy Eye Velocity.................................................................................. 23
4.4.5 Auto-Stop ........................................................................................................... 23
4.4.6 Peak Head Velocity Limits ................................................................................. 23
4.4.7 Set Normative Threshold Ranges...................................................................... 24
4.4.8 Default Settings.................................................................................................. 26
4.4.9 Prepare to test ................................................................................................... 27
4.4.10 Start test............................................................................................................. 28
4.4.11 Impulse counter ................................................................................................. 28
4.5 Vertical impulses- RALP/LARP ................................................................................. 30
4.5.1 Background........................................................................................................ 30
4.5.2 Eye Orientation .................................................................................................. 31
4.5.3 Head Orientation................................................................................................ 32
4.5.4 Prepare .............................................................................................................. 33
4.5.5 Direction/plane guide ......................................................................................... 33
4.5.6 Start test............................................................................................................. 34
4.5.7 Stop.................................................................................................................... 34
4.6 Suppression Head Impulse Test (SHIMP) ................................................................ 35
4.7 Nystagmus test.......................................................................................................... 36
5Reports................................................................................................................................... 37
5.1 Report Menu Options ................................................................................................ 37
5.1.1 Select and Edit Traces....................................................................................... 37
5.1.2 Saccade Numerical Data Tables ....................................................................... 39
5.1.3 Edit Saccades .................................................................................................... 40
5.1.4 3D wave ............................................................................................................. 43
5.1.5 Display ............................................................................................................... 44
5.1.6 Maximum time.................................................................................................... 45
5.1.7 Mirrored.............................................................................................................. 46
5.1.8 Denoise.............................................................................................................. 47
5.2 Generate reports ....................................................................................................... 48
5.2.1 Individual Test Reports ...................................................................................... 48
5.2.2 EyeSeeSix report ............................................................................................... 48
5.3 Printing Reports......................................................................................................... 51
5.3.1 De-identify Patient Information .......................................................................... 51
5.3.2 Print to external printer or PDF .......................................................................... 51
5.4 Export completed session data ................................................................................. 52
5.5 Video Recording ........................................................................................................ 53

6Licensing ............................................................................................................................... 55
6.1 New Systems............................................................................................................. 55
6.2 Upgrade existing systems ......................................................................................... 55
7Maintenance .......................................................................................................................... 57
7.1 General maintenance procedures ............................................................................. 57
7.2 How to clean Interacoustics Products ....................................................................... 57
7.3 Concerning repair ...................................................................................................... 57
7.4 Warranty .................................................................................................................... 58
8Technical information........................................................................................................... 59
8.1 Appendix A: Electromagnetic compatibility (EMC).................................................... 60


EyeSeeCam vHIT - Instructions for Use - EN Page 1
1Introduction
About this manual
This manual is valid for the EyeSeeCam system (revision 1.3). The product is manufactured by:
Interacoustics A/S
Audiometer Allé 1
5500 Middelfart
Denmark
Tel.: +45 6371 3555
Fax: +45 6371 3522
E-mail: info@interacoustics.com
It is the purpose of this manual to provide users of the Interacoustics EyeSeeCam module with all the
information required to carry out safe and reliable measurements.
Dizziness is a frequent symptom in ENT, Neurology, and General Medicine. Part of the clinical
examination of dizzy patients is based on the head impulse test (HIT) of the vestibulo-ocular reflex
(VOR). Head impulses are movements with small positional amplitude (10-20 degrees) but high
acceleration (3.000-6.000 degrees/s2) and high velocity (150-300 degrees/s). When performed
clinically with no quantitative analysis, the test only detects the presence of a corrective saccade as an
indirect sign of a VOR deficit.
The HIT can easily be conducted at the bedside, but it has only a moderate sensitivity (63% in experts
and 72% in non-experts). Its specificity is 78% in experts and 64% in non-experts. Sensitivity,
however, can be improved to up to 100% when quantitative data of the VOR are obtained during the
head impulse. In patients with an acute unilateral vestibular deficit the quantitative HIT is as reliable as
the caloric irrigation test. Chronic patients, however, show a pathological caloric result in only 64% of
the cases but they were reliably identified by quantitative HIT. Hence, in the acute phase of a
vestibular deficit either of these tests is sufficient, but in the chronic phase, which is typical for a
vertigo and dizziness outpatient unit, the quantitative HIT is better.
EyeSeeCam with its lightweight goggles and the integrated inertial measurement unit (IMU) provides a
measurement device for an objective and quantifiable HIT. The mobile setup with a battery-driven laptop
ensures that the HIT can be applied at the patient’s bedside and in the emergency room. With it, a doctor
can objectively and reliably assess peripheral vestibular function in a few minutes and differentiate
peripheral from central vertigo. This video-based HIT is a novel, time-saving and low-cost tool that can be
readily applied in all clinical setups, such as in vertigo outpatient units or in emergency rooms.
Intended use
The EyeSeeCam vHIT is used to present information on the performance of the balance system by
providing objective measures of eye-velocity response to head-velocity stimulus, showing the VOR
gain in the plane of rotation of the head
The system is to be used by trained personnel only such as audiologists, ENT surgeons, neurologists,
hearing healthcare professionals or personnel with a similar level of education.

EyeSeeCam vHIT - Instructions for Use - EN Page 0
The EyeSeeCam is intended for patient from 5 years and up. The patient must be physically sound as
the procedures include some physical stimulation/movement. Patient must be able the see the target
without the use of spectacles.
Contra indications
• Blindness
• Broken nose or other face/head trauma
• Problems with the cervical spine, e.g., neck trauma, acute cervical disc herniation
• Recent eye surgery
• Ptosis
• Excessive eye blinks
• Excessive eye make up
Included parts
EyeSeeCam – vHIT camera and goggle1
USB cable
EyeSeeCam vHIT Software and manuals CD
OtoAccess® Database software CD
Tape measure
Notes on safety
Our instruction manuals contain safety notes which follow the ANSI recommendations (American National
Standards Institute) for safety notes.
WARNING indicates a hazardous situation which, if not avoided, could
result in death or serious injury.
CAUTION, used with the safety alert symbol, indicates a hazardous
situation which, if not avoided, could result in minor or moderate injury.
NOTICE
NOTICE is used to address practices not related to personal injury.
1Applied part including the USB cable as according ti IEC 60601-1:2005

EyeSeeCam vHIT - Instructions for Use - EN Page 1
2Unpacking and installation
Unpacking and inspection
Check box and contents for damage
When the instrument is received please check the shipping box for rough handling and damage. If the
box is damaged it should be kept until the contents of the shipment have been checked mechanically and
electrically. If the instrument is faulty please contact your local distributor. Keep the shipping material for
the carrier’s inspection and insurance claim.
Keep carton for future shipment
The EyeSeeCam comes in its own shipping carton, which is specially designed for the EyeSeeCam.
Please keep this carton. It will be needed if the instrument has to be returned for service.
If service is required please contact your local distributor.
Inspect before connection
Prior to connecting the product it should once more be inspected for damage. All of the cabinet and the
accessories should be checked visually for scratches and missing parts.
Report immediately any faults
Any missing part or malfunction should be reported immediately to the supplier of the instrument together
with the invoice, serial number, and a detailed report of the problem. In the back of this manual you will find
a "Return Report" where you can describe the problem.
Please use "Return Report"
Please realize that if the service engineer does not know what problem to look for he may not find it, so
using the Return Report will be of great help to us and is your best guarantee that the correction of the
problem will be to your satisfaction.

EyeSeeCam vHIT - Instructions for Use - EN Page 2
Markings
The following markings can be found on the instrument:
Symbol
Explanation
Type B applied parts.
Patient applied parts that are not conductive and can be immediately
released from the patient.
Follow instructions for use
WEEE (EU-directive).
This symbol indicates that when the end-user wishes to discard this product,
it must be sent to separate collection facilities for recovery and recycling.
Failing to do so may endanger the environment.
0123
The CE-mark indicates that Interacoustics A/S meets the requirements of
Annex II of the Medical Device Directive 93/42/EEC. TÜV Product Service,
Identification No. 0123, has approved the quality system.
Year of manufacture.
This product is laser class 1. This means the maximum permissible
exposure (MPE) cannot be exceeded when viewing a laser with the naked
eye.
General warnings and precautions
This equipment is intended to be connected to other equipment thus forming a Medical Electrical System.
External equipment intended for connection to signal input, signal output or other connectors shall comply
with the relevant product standard e.g. IEC 60950-1 for IT equipment and the IEC 60601-series for
medical electrical equipment. In addition, all such combinations – Medical Electrical Systems – shall
comply with the safety requirements stated the general standard IEC 60601-1, edition 3, clause 16. Any
equipment not complying with the leakage current requirements in IEC 60601-1 shall be kept outside the
patient environment i.e. at least 1.5 m from the patient support or shall be supplied via a separation
transformer to reduce the leakage currents. Any person who connects external equipment to signal input,
signal output or other connectors has formed a Medical Electrical System and is therefore responsible for
the system to comply with the requirements. If in doubt, contact qualified medical technician or your local
representative.
A Separation Device (isolation device) is needed to isolate the equipment located outside the patient
environment from the equipment located inside the patient environment. In particular such a Separation
Device is required when a network connection is made. The requirement for the Separation Device is
defined in IEC 60601-1, edition 3, clause 16.

EyeSeeCam vHIT - Instructions for Use - EN Page 3
•The EyeSeeCam in not intended for operation in oxygen rich environment, flammable
anesthetics, flammable agents or similar
•The goggle should not be worn by patients with strong defective vision and abnormal rare blink.
Please consult a specialist in such circumstances before using the mask on these types of
patients.
•The system must be switched off before cleaning
•When used for longer time in environments temperatures above 30 °C the surface temperature of
the camera unit can exceed 43 °C (max 44 °C). If this feels uncomfortable for the patient turn off
the equipment for a while.
•The EyeSeeCam system must be serviced at least once a year. The service must include a
safety test. No parts can be serviced while in use with the patient.
•Do not use the equipment if the enclosure is not intact, but have the equipment serviced.
•Only personnel with proper training (skilled personnel) should operate the system.
•Proper use of this device depends on careful reading of this manual and all additional instructions
and labels
•Let the system obtain room temperature before turning it on. Extra time for obtaining room
temperature may be needed to avoid condensation if the system has been moved from a very
cold location to a warmer one.
•The patients or test persons should not wear any optical utilities such as glasses or contact
lenses.
•The patients or test persons should not wear makeup around the eyes

EyeSeeCam vHIT - Instructions for Use - EN Page 4

EyeSeeCam vHIT - Instructions for Use - EN Page 5
3Getting started and system setup
Hardware requirements
System requirements USB EyeSeeCam systems:
Laptop PC: One USB 2 port available
Desktop PC: One USB 2 port available
Furthermore, Microsoft .Net 2.0 Framework and Windows Installer 3.x should be available. If these
programs are not installed on your computer, you will be prompted to install them when running the
EyeSeeCam installation. Just click yes when the installer asks you, and they will be installed
automatically.
Operating System Requirements:
Windows®7 (32 & 64 bit).
Windows®8 (32 & 64 bit).
Windows® 10 (32 & 64 bit).
Power
Maximizing the power supply
Eye images in the vHIT user interface can become flickering and unstable if the power supplied to the
camera is not maximized.
These problems mainly occur in PC systems using chipset technology from INTEL.
To maximize the power supply, CPU power saving features must be disabled either in the bios setup or
via the software as described below. It is recommended to control the CPU power saving features in the
computer’s BIOS rather than via software.
Entering BIOS
BIOS setup can be entered by following the instruction on the screen shortly after the system is switched
on.
Changing BIOS setting
CPU power saving features need to be changed to have maximum power / performance by disabling
Speed Step and other power controlling features.
See example below - changes into ‘Disabled’ must be applied to the functions inside the red boxes.

EyeSeeCam vHIT - Instructions for Use - EN Page 6
Figure 3.2.3-1 Bios settings
Windows Power Setting Setup
It is also necessary to disable Windows’ own power saving features:
In control panel -> Power options properties, power manager, set power schemes to maximum
performance as shown below:
Figure 3.2.4-1 Power Option properties

EyeSeeCam vHIT - Instructions for Use - EN Page 7
USB mode
The EyeSeeCam vHIT camera must only be connected to a USB 2.0 port. If the computer only has USB
3.0 ports the USB 3.0 controller must be disabled so that the ports will work in USB 2.0 mode.
To downgrade from USB 3.0 to USB 2.0 please go to the BIOS/Config menu and select “USB”. Set the
USB 3.0 controller mode to “Disabled” and press F10 to Save and Exit the BIOS Config menu.
Figure 3.2.5-1 Bios Menu
Hardware setup
•Connect the USB cable to the computer
•Connect the other end of the USB cable to the EyeSeeCam camera
•Mount the camera on the goggle using the spherical ball-and-socket joint
•Mount the calibration laser on the bridge of the goggle
•Connect the computer power supply to a wall outlet, then connect to computer
Goggle setup
The vHIT goggle generally has one camera (monocular) fixed at the top side(s) of the mask. The
camera is held in place mechanically with a spherical ball-and-socket joint. The vHIT goggle has two
spherical sockets: one for the left and one for the right eye. The spherical joint allows the camera to be
moved from the mask and plugged into the spherical socket of the other side. This allows for
examining either the left or the right eye, depending on which eye is better suited for examination. This
feature is particularly useful in patients with either a glass eye or with a ptosis.

EyeSeeCam vHIT - Instructions for Use - EN Page 8
Figure 3.3.1-1vHIT goggle and ball and socket joint for camera. Calibration laser is placed on the bridge of the
goggle.
The vHIT goggle suppports the USB camera that is used to record the eye images. This constitutes
the major component of your USB vHIT system.
The USB camera uses infrared light (IR), which is not visible to the naked eye. The IR illumination
enables sessions to be performed in complete darkness. Although it seems to be dark, the eye images
will be displayed as clear and visible images due to the IR-illumination.
The USB vHIT goggle functions as free field of view only.
Before using the USB vHIT goggle please consider the following:
•The goggle and camera contain sensitive electronic and optical devices. Do not expose the
goggle or cameras to mechanical shock
•Do not expose the goggle to moisture (IPX 0 according to IEC 529)
•USB vHIT goggle is not to be used around explosive gas.
In addition to the camera, there is a calibration laser in the center providing the visual fixation targets
for calibration. The laser projects a luminous dot pattern in front of the head, for example, onto the
wall, if the tested subject is sitting in front of the wall, or onto the ceiling, if the tested subject is lying.
Just like the camera, this calibration laser can also be removed together with the camera. The
calibration laser is connected to the camera by means of a flexible wire.
IR illumination
Light Emitting Diodes (LEDs) are positioned at the side of the camera objective lens. The lateral LEDs
emit infrared light (IR).

EyeSeeCam vHIT - Instructions for Use - EN Page 9
Focus adjustment
The lens of the camera module allows you to focus the image. As eye sockets vary in physical
dimension from patient to patient, absolute focus requires adjustment to provide focused images – and
focused images are a precondition for stable eye tracking and noise-free data. The eye is in focus if
details of the individual patterns in the iris (iral signature) are well visible.
Figure 3.3.3-1 Focus adjustment for camera
Calibration laser Adjustment
The calibration laser, in the center of the goggle, has a knob that allows you to align the dots on the wall
by slightly rotating the knob until the 5 laser dots are aligned horizontally and vertically. Do not adjust
the calibration laser unduly. Adjust only if the dots have gone out of horizontal/vertical alignment.
Figure 3.3.4-1 Calibration laser adjustment

EyeSeeCam vHIT - Instructions for Use - EN Page 10
Servicing and maintenance
The vHIT system must be serviced at least once a year. The service must include a safety test.
Software setup
Installation of OtoAccess® – Please refer to the OtoAccess® manual and documentation.
Installation of EyeSeeCam vHIT software - Please refer to the EyeSeeCam installation CD and
instruction.
Installing EyeSeeCam software
Follow the prompts on the screen to step through the software installation.
Figure 3.5.1-1 Set up Wizard for EyeSeeCam
Uninstalling the EyeSeeCam vHIT software -
Please refer to the EyeSeeCam installation CD and instruction.
Starting EyeSeeCam from OtoAccess® 1.5
To start EyeSeeCam:
1. Double Click the OtoAccess® icon on the desktop which will open the OtoAccess® program
2. Select the instrument (in this example, EyeSeeCam) by double-clicking EyeSeeCam in the
Select Instrument box. EyeSeeCam will open in a new window.
Other manuals for EyeSeeCam
2
Table of contents
Languages: