•Thepatient’sweight,size,physicalabilityandmedical
condition.
•Thetypeoftransferandtheenvironment.
•Thecompatibilitytotheotherliftingequipmentused.
FormoreinformationseetheInvacareSlingSelection
Guide
4.2ApplyingtheSling
SlingApplicationinLyingPosition
1.Turnthepatienttotheside.
2.Folduphalfoftheslingandplaceitalongthespine
withtheloweredgejustabovekneepit.Fig.2
3.Turnthepatienttotheothersideandunfoldthesling.
4.Turnthepatientonitsbackontotheslingandensure
themid-markersarealignedwiththespine.Fig.3
PositioningofPatientArms
Thearmsofthepatientmustbeplacedinsidethesling
4.3SlingAttachmenttotheLift
Thesuspensionstrapsoftheslingareequippedwithcolor
codedloopsprovidingdifferentlengthstoplacethepatient
intodifferentpositions.Shorterstrapsattheshoulders
andlongerstrapsatthelegswillproduceamorevertical
lift,whichwillassistatransferintoasittingposition.By
lengtheningthestrapsattheshouldersandshorteningthe
strapsatthelegs,amorereclinedpositioncanbeachieved
whichismoresuitableforatransferintoalyingposition.
Matchtheloopcolorsofthecorrespondingstrapsoneach
sideoftheslingforanevenliftofthepatient.Fig.4
IMPORTANT!
Foratransferintoasittingposition,placethe
patientasverticalaspossibleinthesling
1.Attachthedesiredloopofeachstraptothe
correspondinghookonthelift:
a.With2connectionpointsattachtheshoulderstrap
beforethelegstraponeachofthehooks.Fig.5a
b.With4connectionpointsattacheachoftheshoulder
andlegstrapstoanindividualhook.Fig.5b
4.4RemovingtheSling
1.Aftercompletionofthetransfer,detachtheslingfrom
thelift.
2.Reversetheproceduredescribedin4.2Applyingthe
Sling,page5.
5Maintenance
5.1Inspection
DailyInspectionBeforeUse
Inspectiontobeperformeddaily,beforeuseandaftereach
laundering:
•Visuallyinspecttheslingandcheckallpartsforsignsof
damage,wearorpotentialfailure.
PeriodicInspection
IMPORTANT!
Inspectionsmustbeperformedbyaqualied
personwhoiswellacquaintedwiththedesign,
useandcareofslings.
Aperiodicsafetyinspectionoftheslingmustbeperformed
atleastevery6monthsunlessotherwisestatedinlocal
requirements.
Adetailedsafetyinspectionchecklistisavailablefrom
Invacareasaseparatedocument.
LOLERStatement
TheUKHealthandSafetyExecutive’sLiftingOperationsand
LiftingEquipmentRegulations1998,requireanyequipment
thatisusedintheworkplacetoliftaloadbesubjectto
safetyinspectiononasixmonthlybasis.Pleaserefertothe
HSEwebsiteforguidancewww.hse.gov.uk.
Thepersonresponsiblefortheequipmentmustensure
adherencetoLOLERregulations.
5.2CleaningandDisinfection
WARNING!
Riskofinjury
Usageofdamagedslingscancausethepatientto
fallorcauseinjurytoassistants.
–Aftereachlaundering,inspecttheslingfor
wear,tear,andloosestitching.
IMPORTANT!
Regularcleaninganddisinfectionprevents
contamination.
Cleananddisinfecttheproduct
–regularlywhileitisinuse,
–whenithasbeenincontactwithanybody
uids,
–beforeusingitforanewuser.
IMPORTANT!
Alwaysmakesurethattheproductiscompletely
driedbeforetakingitintouseagain.
CleaningInstructions
Launderanddrytheslingaccordingtothewashing
instructionsonthelabel.
Frequentuseofhightemperaturesandtumbledryingcan
reducetheservicelifeoftheproduct.
Werecommendtheuseofanordinaryhouseholdlaundry
detergent.Donotusesoftener.
Fordisinfection,selectadisinfectantthatiscompatible
withtheslingmaterialsandfollowtheinstructionsofthe
disinfectantmanufacturer.
6AfterUse
6.1StorageConditions
Theslingmustbestoredindryconditionswithamaximum
relativehumidityof50%andwithoutdirectsunlight.
6.2Reconditioning
Thisproductissuitableforreuse.Toreconditiontheproduct
foranewuser,carryoutthefollowingactions:
•Inspection
•Cleaninganddisinfection
Fordetailedinformation,see5Maintenance,page5.
Makesurethattheusermanualishandedoverwiththe
product.
Ifanydamageormalfunctionisdetected,donotreusethe
product.
6.3Disposal
Beenvironmentallyresponsibleandrecyclethisproduct
throughyourrecyclingfacilityatitsendoflife.
60124413-A5