20 www.i-sens.com www.i-sens.com 21
Step 2
While the symbol blinks, press and hold the
▲button for 3 seconds till the "ctl" appears
on the display. If the talking function is on, the
meter will say“Control solution test. Please
apply control solution onto the test strip.”
Step 1
Insert a test strip into the meter’s test strip port,
with the contact bars facing upwards. Gently
push the test strip into the port.
The symbol will blink on the screen.
If the talking function is on, the meter will say
“Blood glucose test. Please apply blood onto
the test strip.”
You may do a control solution test:
•When you want to practice the test procedure using the
control solution instead of blood
•When using the meter for the first time
•Whenever you open a new vial of test strips
•If the meter or test strips do not function properly
•If your symptoms are inconsistent with the blood glucose
test results and you feel that the meter or test strips are not
working properly
•If you drop or damage the meter
Control Solution Testing
Checking the System
You may check your meter and test strips using the
CareSensControlSolution(controlA,Band/orC).
The CareSens Control Solution contains a known
amount of glucose and is used to check that the meter
and the test strips are working properly.
The test strip vials have CareSens Control Solution
ranges printed on their labels. Compare the result
displayed on the meter to the CareSens Control
Solution range printed on the test strip vial.
Before using a new meter or a new vial of test strips,
you may conduct a control solution test following the
procedure on pages 21–23.
Notes:
•Use only the CareSens Control Solutions.
•Check the expiration dates printed on the bottle. When you
first open a control solution bottle, record the discard date
(date opened plus three (3) months) in the space provided on
the label.
•Make sure your meter, test strips and control solution are at
room temperature before testing. Control solution tests must
be done at room temperature 20-25˚C.
•Before using the control solution, shake the bottle, discard the
first few drops and wipe the tip clean.
•Close the control solution bottle tightly and store at a
temperature between 8-30˚C.