Jasco V-630 User manual

P/N: 0302-1935D February 2010
Model V-630/650/660/670
Spectrophotometer
Hardware/Function Manual

i
Safety Considerations
To ensure operation safety, this instrument must be operated correctly and maintained according
to a regular schedule. Carefully read to fully understand all safety precautions in this manual
before operating the instrument. Please take a moment to understand what the signal words
WARNING
!
, CAUTION, and Note mean in this manual.
(1) Safety symbols
Instruction manual symbol. If the product is marked with this symbol,
refer to the instrument manuals to protect the instrument against
damage.
WARNING
!
A WARNING
!
indicates a potentially hazardous situation which, if
not avoided, could result in serious injury or even death.
CAUTION A CAUTION indicates a potentially hazardous situation which, if not
avoided, may result in minor or moderate injury. It may also be
used to alert against damaging the equipment.
Do not proceed beyond a WARNING
!
or CAUTION notice until you
understand the hazardous conditions and have taken the
appropriate steps.
Note A Note provides additional information to help the operator achieve
optimal instrument performance.

ii
(2) Warning indication
Warnings are indicated at several locations on this instrument. Do not deface or damage the
warning indication.
Figure.1 Warning indication
1) Warning for FUSE (Fig. 2 or Fig. 3)
Only use fuses of the designated rating to protect both operator and instrument from fire and
other hazards. Whenever replacing the lamp, be sure to turn the “Power” switch OFF and
unplug the power cable from the power outlet to avoid the risk of electric shock and other
hazards.
2) Warning for GROUND (Fig. 2 or Fig. 3)
You may receive an electric shock if the instrument is not properly grounded. Correctly ground
the instrument using the grounding terminal on the power board. Do not use water supply
pipes for grounding because they are often made of nonmetals. For the same reason, do not
use gas piping for grounding either.
Warning indication
Figure 2 V-630 Rear Panel

iii
Warning indication
Figure 3 V-650/660/670 Rear Panel
(3) Warning for carrying
These instruments weigh as follows:
V-630: 15.0 kg V-650: 27.0 kg
V-660: 29.0 kg V-570: 28.0 kg
When carrying the instrument, hold the carrying handle at the bottom of the instrument
firmly (See Fig. 4 or Fig. 5).
Carrying handle (both sides)
Figure 4 V-630 side view

iv
Carrying handle (both sides)
Figure 5 V-650/660/670 side view

v
Connecting the Power Cable
WARNING!: The green/yellow ground core of the mains lead must be connected
to a ground that complies with the local electricity supply authority (or
equivalent body). The instrument is dangerous if not correctly
grounded.
240 Volt (nominal) Supply
Ground
Line
Neutral
Wire Color
Green/Yellow
Brown
Blue
115 Volt (nominal) Supply
Neutral
Wire Color
Green/Yellow
Black
White
Line
Ground
View of pin side of plug
Note: Instruments intended for operation at 115 V, 60 Hz are supplied with a mains
cable with a molded plug and socket.

vi
Regulatory Statement
CE Notice
The symbol indicates compliance of this JASCO system to the EMC (Electromagnetic
Compatibility) and Low Voltage Directives of the European Community. This symbol indicates
that this JASCO system meets the relevant basic safety and health requirements of the EC
Directive based on the following technical standards:
· EN61326-1: "Electrical equipment for measurement, control and laboratory use – EMC
requirements – Part 1: General requirements"
WARNING!: This is a Class A product. In a domestic environment this product
may cause radio interference, in which case the user may be
required to take corrective measures.
· IEC61000-3-2: "Electromagnetic compatibility (EMC) Part 3-2: Limits – Limits for harmonic
current emissions (equipment input current up to and including 16A per phase) "
· IEC61010-1: "Safety requirements for electrical equipment for measurement, control and
laboratory use – Part 1: General requirements"
· A "Declaration of Conformity" in accordance with the above standards has been made and is
on file at JASCO CORPORATION, 2967-5 Ishikawa-machi, Hachioji-shi, Tokyo 192-8537,
JAPAN.
FCC Statement (for USA only)
Federal Communications Commission Radio Frequency Interference Statement
WARNING!: This equipment generates, uses, and can radiate radio frequency
energy. If it is not installed and used in accordance with the
instruction manual, it may cause interference to radio
communications. It has been tested and found to comply with the
limits for a Class A computing device pursuant to Part 15 of FCC
Rules, which are designed to provide reasonable protection against
such interference when operated in a commercial environment.
Operation of this equipment in a residential area is likely to cause
interference, in which case the users at their own expense will be
required to take whatever measures may be required to correct the
interference.

vii
Preface
This instruction manual is your guide for using this instrument. It instructs first-time users on how
to use the instrument, and serves as a reference for experienced users.
Before using the instrument, please read this instruction manual carefully, and make sure that the
contents are fully understood. This manual should be easily accessible to the operator at all times
during instrument operation. When not using the instrument, keep this manual in a safe place. If this
instruction manual becomes lost, order a replacement from your local JASCO distributor.
Note: The operating procedure varies with the type of instrument. Operation is
described by type. Read the portion relevant to your instrument.

viii
Installation Requirements
To ensure operation safety, observe the following conditions:
(1) Do not operate the instrument under voltage fluctuations exceeding 10% of the
recommended line voltage. Otherwise, the instrument may not function properly.
(2) Frequency or spike noise in the power supply should be minimal.
(3) Ensure that the instrument is grounded.
(4) Operate the instrument in a temperature range of 15 to 30 °C.
(5) Operate the instrument in humidity below 85% (RH). JASCO recommends operating the
instrument in humidity below 60% to avoid the deterioration of the optical components due
to the condensation caused by high humidity.
(6) Operate the instrument in an atmospheric pressure range of 750 to 1060 hPa.
(7) Avoid strong magnetic fields and sources of high frequency. The instrument may not
function properly when near a strong magnetic field or high frequency source.
(8) Avoid vibration from vacuum pumps, electric motors, processing equipment and machine
tools.
(9) Avoid dust and corrosive gas. Do not install the instrument where it may be exposed to dust,
especially in locations exposed to outside air or ventilation outlets that discharge dust
particles.
(10) Do not install the instrument in a location where it may be exposed to direct sunlight.
(11) Install the instrument in a horizontal and stable position. (This includes a table or desk upon
which the instrument is installed.)
(12) Ensure that no air conditioner blows air directly onto the instrument. This may prevent
stable measurement.
(13) Install the instrument in a location that allows easy access for maintenance.
Note: The above conditions do not guarantee optimal performance of this
instrument.

ix
Servicing
Contact your local JASCO distributor for instrument servicing. In addition, contact your JASCO
distributor before moving the instrument to another location. Consumable parts should be ordered
according to part number from your local JASCO distributor. If a part number is unknown, give
your JASCO distributor the model name and serial number of your instrument.
Do not return contaminated products or parts that may constitute a health hazard to
JASCO employees.
Notices
(1) JASCO shall not be held liable, either directly or indirectly, for any consequential damage
incurred as a result of product use.
(2) Prohibitions on the use of JASCO software
Copying software for purposes other than backup
Transfer or licensing of the right to use software to a third party
Disclosure of confidential information regarding software
Modification of software
Use of software on multiple workstations, network terminals, or by other methods (not
applicable under a network licensing agreement concluded with JASCO)
(3) The contents of this manual are subject to change without notice for product improvement.
(4) This manual is considered complete and accurate at publication.
(5) This manual does not guarantee the validity of any patent rights or other rights.
(6) If a JASCO software program has failed causing an error or improper operation, this may
be caused by a conflict from another program operating on the PC. In this case, take
corrective action by uninstalling the conflicting product(s).
(7) Windows is a registered trademark of Microsoft Corporation in the United States and other
countries. In general, company names and product names are trademarks or registered
trademarks of the respective companies. However, the TM and ® marks are not used in all
cases in this manual.
(8) JASCO and the JASCO logo are registered trademarks of JASCO Corporation in Japan
and other countries.
(C) JASCO Corporation, 2010. All rights reserved. Printed in JAPAN.

x
Limited Warranty
Products sold by JASCO, unless otherwise specified, are warranted for a period of one year from
the date of shipment to be free of defects in materials and workmanship. If any defects in the
product are found during this warranty period, JASCO will repair or replace the defective part(s)
or product free of charge.
THIS WARRANTY DOES NOT APPLY TO DEFECTS RESULTING FROM THE FOLLOWING:
(1) IMPROPER OR INADEQUATE INSTALLATION
(2) IMPROPER OR INADEQUATE OPERATION, MAINTENANCE, ADJUSTMENT OR
CALIBRATION
(3) UNAUTHORIZED MODIFICATION OR MISUSE
(4) USE OF CONSUMABLE PARTS NOT SUPPLIED BY AN AUTHORIZED JASCO
DISTRIBUTOR
(5) CORROSION DUE TO THE USE OF IMPROPER SOLVENTS, SAMPLES, OR DUE TO
SURROUNDING GASES
(6) ACCIDENTS BEYOND JASCO'S CONTROL, INCLUDING NATURAL DISASTERS
This warranty does not cover the consumable parts listed below:
(1) Tungsten lamp, and other light sources
(2) Mirrors in the light source section, and cell windows
(3) Fuses, batteries, glassware, chart paper and ink
THE WARRANTY FOR ALL PARTS SUPPLIED AND REPAIRS PROVIDED UNDER THIS
WARRANTY EXPIRES ON THE WARRANTY EXPIRATION DATE OF THE ORIGINAL
PRODUCT. FOR INQUIRIES CONCERNING REPAIR SERVICE, CONTACT YOUR JASCO
DISTRIBUTOR AFTER CONFIRMING THE MODEL NAME AND SERIAL NUMBER OF YOUR
INSTRUMENT.
JASCO Corporation
2967-5, Ishikawa-machi, Hachioji-shi
Tokyo 192-8537
JAPAN

xi
Contents
Safety Considerations............................................................................................... i
Connecting the Power Cable ................................................................................... v
Regulatory Statement ............................................................................................. vi
Preface .................................................................................................................... vii
Installation Requirements..................................................................................... viii
Servicing .................................................................................................................. ix
Notices...................................................................................................................... ix
Limited Warranty ...................................................................................................... x
Contents................................................................................................................... xi
1 Unpacking and Installation............................................................................... 1
1.1 Unpacking ...............................................................................................................1
1.2 Installation and Connecting Cables......................................................................4
1.2.1 Preparation ...........................................................................................................4
1.2.2 Connecting the Cable of iRM Type.......................................................................4
1.2.3 Connecting the Cable of PC Type ........................................................................4
2 Specifications ....................................................................................................... 6
2.1 V-630 UV/VIS Spectrophotometer Specifications................................................6
2.2 V-650 UV/VIS Spectrophotometer Specifications................................................7
2.3 V-660 UV/VIS Spectrophotometer Specifications................................................8
2.4 V-670 UV/VIS/NIR Spectrophotometer Specifications.........................................9
3 General Description of Instrument.................................................................... 11
3.1 V-630 Spectrophotometer Optical System.........................................................11
3.2 V-650/660/670 Spectrophotometers Optical System .........................................13
3.3 iRM System ...........................................................................................................16
4 Names and Functions of Components ............................................................. 17
4.1 V-630 Spectrophotometer....................................................................................17
4.1.1 Overview.............................................................................................................17
4.1.2 Rear Panel..........................................................................................................18
4.2 V-650/660/670 Spectrophotometers: Overview..................................................19
4.2.1 Overview.............................................................................................................19
4.2.2 Rear Panel..........................................................................................................20
4.3 Standard Cell Holder (Model USE-753 Rectangular Cell Holder) .....................21
4.4 Intelligent Remote Module (iRM).........................................................................22
5 Maintenance........................................................................................................ 23
5.1 Cautions for Routine Operation ..........................................................................23
5.2 Cleaning the Sample Compartment ....................................................................23
5.3 Replacement of Consumables ............................................................................24
5.3.1 Replacement of Fuse..........................................................................................24

xii
5.3.2 Lamp Replacement and Adjustment of the Light Source Mirror .........................26
5.3.2.1 Lamp replacement ......................................................................................26
5.3.2.2 Adjustment of iRM type light source mirror .................................................27
5.3.2.3 Adjustment of PC type light source mirror ..................................................29
5.3.3 Consumable Parts .............................................................................................30
6 Troubleshooting ................................................................................................. 31

1
1 Unpacking and Installation
WARNING!: This section is intended for trained JASCO servicemen. If the user
attempts unpacking and installation, injury may result. Please leave
unpacking and installation to your JASCO serviceman.
CAUTION: Use this manual when you are unpacking/installing the instrument in
conjunction with a JASCO serviceman’s instructions in the case of a
problem.
1.1 Unpacking
(1) Unpacking the main unit
Take out the main unit from the carton and make sure that the serial No. displayed on the serial
No. label located on the left side of the unit and the serial No. on the inspection certificate agree.
Also ensure that the line voltage is consistent with the power requirement of the instrument.
MODEL
SERIAL No.
POWER
PROTECT
CLASS I
MADE IN JAPAN
JASCO Corpration
日本分光株式会社
_
192-8537 東京都八王子市石川町 2967-5
AC
V 50/60Hz VA
Serial Number
Line voltage, frequency
Figure 1.1 Serial No. label
(2) Unpacking the standard accessories
Remove the standard accessories from the carton and check them against the packing list. If you
find any parts missing or damaged items, contact your local JASCO distributor.
Table 1.1 List of Standard Accessories for V-630 iRM Type
Description Qty.
Comments
Intelligent remote module(iRM)
iRM touch pen
Compact flash
Card adapter
Holmium glass
Time-lag fuse
AC power cable
Allen wrench
Certificate of inspection
Instruction manual
1
1
1
1
1
2
1
1
1
1 set
Bifunctional holmium glass
holder/shielding plate
For adjustment of the light source
mirror

2
Note 1: Use of the holmium glass does not guarantee wavelength accuracy.
Note 2: Holmium glass holder can also be used as a shielding plate by rotating it
through 90 degree from its usual position.
Table 1.2 List of Standard Accessories for V-650/660/670 iRM Type
Description Qty.
Comments
Intelligent remote module(iRM)
iRM touch pen
Compact flash
Card adapter
Holmium glass
Time-lag fuse
AC power cable
Phillips screw driver
Certificate of inspection
Instruction manual
1
1
1
1
1
2
1
1
1
1 set
Bifunctional holmium glass
holder/shielding plate
For adjustment of the light source
mirror
Note 1: Use of the holmium glass does not guarantee wavelength accuracy.
Note 2: Holmium glass holder can also be used as a shielding plate by rotating it
through 90 degree from its usual position.
Table 1.3 List of Standard Accessories for V-630 PC Type
Description Qty. Comments
Holmium glass
Time-lag fuse
USB cable
AC power cable
Allen wrench
Software setup CD
Certificate of inspection
Instruction manual
1
2
1
1
1
1
1
1 set
Bifunctional holmium glass
holder/shielding plate
For adjustment of the light source
mirror
Note 1: Use of the holmium glass does not guarantee wavelength accuracy.
Note 2: Holmium glass holder can also be used as a shielding plate by rotating it
through 90 degree from its usual position.

3
Table 1.4 List of Standard Accessories for V-650/660/670 PC Type
Description Qty.
Comments
Holmium glass
Time-lag fuse
USB cable
AC power cable
Phillips screw driver
Software setup CD
Certificate of inspection
Instruction manual
1
2
1
1
1
1
1
1 set
Bifunctional holmium glass
holder/shielding plate
For adjustment of the light source
mirror
Note 1: Use of the holmium glass does not guarantee wavelength accuracy.
Note 2: Holmium glass holder can also be used as a shielding plate by rotating it
through 90 degree from its usual position.

4
1.2 Installation and Connecting Cables
Note:
Be certain to comply with the instrument installation requirements
described in this manual (see page
ⅷ
). Failure to comply with these
requirements can cause faults with the instrument.
1.2.1 Preparation
Prepare a table which can bear a weight of about 50 kg (differs according to instrument) and has
an area of 700 mm (length) × 600 mm (width) when using the iRM type.
If using a PC as well, prepare a table, which can hold a weight of about 70 kg (differs according to
instrument) and has an area of 700 mm (length) × 1500 mm (width), allowing space for a printer.
Confirm that the power switch of the main unit is off and that the power switch of the PC is off
when using a PC.
1.2.2 Connecting the Cable of iRM Type
WARNING!: You may receive an electric shock if the instrument is not properly
grounded. Correctly ground the instrument using the grounding
terminal on the power board.
CAUTION: For grounding, do not use water pipes, which are often made of
nonmetals. For safety reasons, do not use gas tubing for grounding
either.
Note: Use of a 3-pin power plug with a grounding terminal is recommended.
(1) Connect the cable from the iRM to the “iRM” connector on the rear of V-600.
(2) Connect the AC power cable to the “AC INPUT” connector on the rear of V-600.
Connect the other end of the AC power cable to an electrical outlet.
1.2.3 Connecting the Cable of PC Type
WARNING!: You may receive an electric shock if the instrument is not properly
grounded. Correctly ground the instrument using the grounding
terminal on the power board.
CAUTION: For grounding, do not use water pipes, which are often made of
nonmetals. For safety reasons, do not use gas tubing for grounding
either.
Note: Use of a 3-pin power plug with a grounding terminal is recommended.
(1) Connect the “USB” port located on the rear of V-600 and the USB port on the PC using the
USB cable.

5
(2) Connect the PC and the monitor.
(3) Connect the AC power cable to the “AC INPUT” connector on the rear of V-600. Connect
the other end of the AC power cable to an electrical outlet.
(4) Connect the AC power cables to the AC input connector of PC and monitor. Connect the
other ends of the AC power cables to an electrical outlet.

6
2 Specifications
The specifications of each model covered in this manual are given below.
2.1 V-630 UV/VIS Spectrophotometer Specifications
Optical system Single monochromator
UV/VIS region: 1200 lines/mm concave grating
Rowland off-circle arrangement
Double beam type
Light source Deuterium lamp: 190 to 350 nm
Halogen lamp: 330 to 1100 nm
Light source exchange
wavelength
Any wavelength between 330 and 350 nm can be
selected.
Detector Silicon photodiode (S1337)
Wavelength range 190 to 1100 nm
Wavelength accuracy ±0.2 nm (at 656.1 nm, when the room temperature is
stabilized.)
Wavelength repeatability ±0.1 nm
Slew speed 12000 nm/min
Spectral bandwidth 1.5 nm fixed
Photometric range 0 to 10000 %T
-2 to 3 Abs
Photometric accuracy ±0.002 Abs (0 to 0.5 Abs)
±0.003 Abs (0.5 to 1 Abs)
±0.3 %T
(Tested with NIST SRM 930D)
Photometric repeatability ±0.001 Abs (0 to 0.5 Abs)
±0.001 Abs (0.5 to 1 Abs)
Stray light 1 % (198 nm KCL 12 g/L aqueous solution)
0.04 % (220 nm NaI 10 g/L aqueous solution)
0.02 % (340 nm NaNO2 50 g/L aqueous solution)
0.02 % (370 nm NaNO2 50 g/L aqueous solution)
(10 mm cell used)
Baseline stability 0.0004 Abs/hour
(Value obtained more than two hours after turning on the
light source, when the room temperature is stabilized,
wavelength: 250 nm, and response: slow.)
Baseline flatness ±0.0006 Abs
(Value obtained after baseline correction with a
temperature variation of less than 5°C, wavelength: 200
to 1000 nm, response: medium, with smoothing
processing and wavelength scanning speed: 400 nm
/min.)
RMS noise 0.00006 Abs (0 Abs, wavelength: 500 nm, measurement
time: 60 sec, response: medium)
Power requirements 100, 115, 200, 220, 230, 240 V ±10 %, 105 VA
Dimensions and weight 216(H) ´ 486(W) ´ 441(D)mm (excluding protrusions)
15 kg
Note: Values given for wavelength accuracy, wavelength repeatability, photometric
accuracy, photometric repeatability, baseline flatness, and RMS noise are
those obtained more than one hour after the light source was turned on.

7
2.2 V-650 UV/VIS Spectrophotometer Specifications
Optical system Single monochromator
UV/VIS region: 1200 lines/mm plane grating
Czerny-Turner mount
Double beam type
Light source Deuterium lamp: 190 to 350 nm
Halogen lamp: 330 to 900 nm
Light source exchange
wavelength
Any wavelength between 330 and 350 nm can be
selected.
Detector Photomultiplier tube
Wavelength range 190 to 900 nm
Wavelength accuracy ±0.2 nm (at a spectral bandwidth of 0.5 nm, wavelength:
656.1 nm, when the room temperature is stabilized.)
Wavelength repeatability ±0.05 nm
Slew speed 12000 nm/min
Spectral bandwidth 0.1, 0.2, 0.5, 1, 2, 5, 10 nm
L2, L5, L10 nm (low stray-light mode)
M1, M2 nm (micro-cell mode)
Photometric range 0 to 10000 %T
-2 to 4 Abs
Photometric accuracy ±0.002 Abs (0 to 0.5 Abs)
±0.003 Abs (0.5 to 1 Abs)
±0.3 %T
(Tested with NIST SRM 930D)
Photometric repeatability ±0.001 Abs(0 to 0.5 Abs)
±0.001 Abs(0.5 to 1 Abs)
Stray light 1 % (198 nm KCL 12g/L aqueous solution)
0.005 %(220 nm NaI 10g/L aqueous solution)
0.005 %(340 nm NaNO2 50g/L aqueous solution)
0.005 %(370 nm NaNO2 50g/L aqueous solution)
(spectral bandwidth: L2 nm, 10 mm cell used)
Baseline stability 0.0003 Abs/hour
(value obtained more than two hours after turning on the
light source, when the room temperature is stabilized,
wavelength: 250 nm, response: slow, and spectral
bandwidth: 2nm.)
Baseline flatness ±0.0003 Abs
(value obtained after baseline correction with a
temperature variation of less than 5°C, wavelength: 200
to 850 nm, response: medium, spectral bandwidth: 2 nm,
and wavelength scanning speed: 400 nm/min with
smoothing processing)
RMS noise 0.00003 Abs (0 Abs, wavelength: 500 nm, measurement
time: 60 sec, response: medium, spectral bandwidth: 2
nm)
Power requirements 100, 115, 200, 220, 230, 240 V ±10 %, 145 VA
Dimensions and weight 270(H) ´ 460(W) ´ 602(D) mm (excluding protrusions)
27 kg
Note: Values given for wavelength accuracy, wavelength repeatability, photometric
accuracy, photometric repeatability, baseline flatness, and RMS noise are
those obtained more than one hour after the light source was turned on.
This manual suits for next models
3
Table of contents
Other Jasco Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Lutron Electronics
Lutron Electronics MHT-381SD Operation manual

National Instruments
National Instruments FD-11613 manual

YOKOGAWA
YOKOGAWA AQ6370D user manual

Foss
Foss NIRS DA1650 user manual

PCE Health and Fitness
PCE Health and Fitness PM 3 user manual

Konica Minolta
Konica Minolta MYIRO-1 instruction manual