JENCONS Zippette Digital Specifications

Operator Instructions
2 – 14
Notice D´Emploi
15 – 27
Gebrauchsanweisung
28 – 40
Manual Del Operador
41 – 53
Notes
54 – 55
Jencons Scientific Inc (USA)
800 Bursca Drive, Suite 801
Bridgeville, PA 15017
Phone: 412 257-8861
Toll Free: 800 846-9959
Fax: 412 257-8809
Email: [email protected]
Website: www.jenconsusa.com
Jencons Scientific Ltd
Cherrycourt Way Ind. Est.
Stanbridge Rd, Leighton Buzzard
LU7 4UA, ENGLAND
Phone: 01525 372010
Fax: 01525 379547
Int. Phone: +44 1525 372010
Int. Fax: +44 1525 851461
Email: [email protected]
Website: www.jencons.co.uk
Online sales: www.ecomcat.co.uk
ISO 9001:2000
FM 15114

Operator
Instructions
Zippette Digital 0-30ml manual priming
Zippette Digital 0-30ml spring-loaded

2Operator Instructions
Contents
General Safety Instruction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .This page
Restriction of Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Before Using the Zippette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Inserting the return spring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
The volume adjustment knob . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
User Calibration procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Zero reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
‘Sleep Mode’ and Low Battery indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Battery replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Maintenance/Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Sterilisation/Autoclaving . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Fluid-Path disassembly procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Tap open Tap closed
(2) Electronic
housing head
General Safety Instruction
• Never leave the Zippette on your work bench with the piston Barrel full with liquid without securing
the dispense anti-drip tap into ‘Closed’ position
• Always check that the dispense anti-drip tap is in the ‘Open’ position before attempting to dispense
• Before adjusting volume always place receiving vessel under nozzle and open tap
• Never use force

Operator Instructions 3
Restriction of Use
NEVER use the Zippette with:
• Liquids which are not compatible with PTFE, PVDF, FEP, Borosilicate glass, Alumina ceramic or might
attack Platinum-Iridium
• Hydrofluoric acid
• Liquids which contain solid particles
Temperature limits of Zippette and reagent are 15°C to 40°C
Before Using the Zippette
Check that the instrument has not been damaged in transit.
Do not grip the electronic housing head (2) when carrying or adjusting the instrument.
The Zippette Digital is fitted with a non-rechargeable Lithium battery. The battery life is greater than
30,000 three minute sessions.
The Lithium battery should be disposed of correctly, on no account should it be incinerated.
Assembly
The Zippette is packed with the dispense nozzle attached and the inlet feed tube removed. The 300mm
length of PTFE inlet tube provided should be trimmed to fit your particular reservoir. The bottom end
should be cut diagonally at a length that is close to the bottom of the container when the Zippette is
assembled to the container. Longer lengths of inlet tube are available on request.
The threaded platform base of the Zippette has a 30mm screw thread for use with the recommended
Jencons Safety amber reservoir 1.4 litre with side filler neck (Cat. No. 262-086) or with the 2.5 litre
amber reservoir with side filler neck (Cat. No. 262-103). The assembled Zippette is screwed to the reser-
voir using gentle hand torque applied to the Zippette’s threaded platform base only. Removal should also
be by means of hand torque applied to this same base.
Do not operate the piston until the unit is safely and fully mounted on the reservoir bottle!
Three adaptors are supplied to suit containers with a 38mm, 40mm or 45mm screw neck.
Operating Instructions
The control panel
Zippette Digital has four momentary action push
buttons on the display box:
1. ‘On’ – switches the unit ‘On’ and displays the
last setting
2. ‘CAL’ – to enter the calibration mode of the
instrument
3. ‘Up’ arrow – to increase the displayed volume
in calibration mode
4. ‘Down’ arrow – to decrease the displayed vol-
ume in calibration mode
Priming the Zippette
• Place a container under the Zippette’s dispense
nozzle
• Turn the volume adjustment knob until the
unit displays 30ml to enable free movement of
the piston
• Open the anti-drip tap
• Prime the unit with a few gentle up and down
strokes, taking the piston right down to its low-
est stop position and lifting it up approx. 25mm
• Repeat until a steady, bubble free flow is visible
in the dispense tube

4Operator Instructions
Inserting the return spring (when applicable)
To insert the return self-priming spring follow the following steps:
Drawing 1 Drawing 2
1. Make sure that the Zippette is completely
empty. For ease of operation, assemble the
Zippette onto an empty reservoir bottle and
turn the anti-drip tap to ‘Open’ position.
2. Turn the volume adjustment knob to set the
Zippette to the max dispense volume of 30ml
(Drawing 1).
3. Turn the fixing ring which is located at the
bottom of the Zippette’s sleeve until it’s raised
rib is aligned with the front piston sleeve slot
and remove by gently pulling down (Drawing
1). The fixing ring should stay on the barrel
protection sleeve (Drawing 2).
4. Press the two fixing levers on the back of the
piston sleeve (situated behind the LCD)
inwards (Drawing 2), and lever the back pis-
ton sleeve away from the front sleeve assem-
bly (Drawing 3).
5. Remove the front sleeve assembly from the
piston locator and move the piston out of it’s
glass barrel (Drawing 3 & 4).
6. Carefully guide the piston through the spring,
and locate the piston back into it’s glass barrel
(Drawing 5).
7. Reassemble the front Zippette piston sleeve
(Drawing 5) making sure that the cursor
locates under the stop ring (Drawing 5). Refit
the back piston sleeve and pull up the fixing
ring (Drawing 6) – turn it half circle to locate
it onto the Zippette piston sleeve (Drawing 7).
The unit is now ready for use (see operating
instructions page 3).
Volume adjustment
Front sleeve slot
Front sleeve Back sleeve
Fixing levers
Barrel
protection
sleeve
263-117
Cursor
Raised rib
Threaded platform ring
263-124
Fixing ring
263-081
Tap open position
Theaded
platform base

Operator Instructions 5
Drawing 3 Drawing 4
Drawing 5 Drawing 6 Drawing 7
Stop ring
263-103
Cursor
Return spring
263-075

The volume adjustment knob
Turning the volume adjustment knob clockwise, will increase the volume to be dispensed. Each full revo-
lution of the knob represents 2ml.
For fine volume adjustments, turn the knob a quarter turn in either direction for 0.5ml adjustment.
User Calibration procedure
Your Zippette Digital has been calibrated to it’s nominal volume of 30ml. However, due to changes in envi-
ronmental conditions and the variability in the viscosity of the media which you dispense, a recalibration
might be required. You can either recalibrate at regular intervals such as once a week or whenever you
notice that the dispensed volume is different from the volume displayed by the unit.
To fully recalibrate your Zippette Digital follow the following steps:
1. Set the Zippette Digital to the nominal volume of 30ml or any other
volume which is the most common volume you dispense.
2. Follow the common rules for calibration used in statistical quality con-
trol (ISO DIS 8655/2). Dispense five full volumes of distilled water at
20°C to establish the actual mean volume of liquid dispensed. The
attached formula will help you to follow common rules. If the gravita-
tional average result varies from the volume displayed, you should
recalibrate the Zippette.
3. Press and hold the ‘CAL’ and ‘UP’ buttons simultaneously for 5 seconds in
order to enter the calibration mode of the instrument. Immediately when
you are in the calibration mode, the word CAL flashes on the display.
4. With the word CAL flashing on the display, lower the piston to its lowest
stop position. Where your Zippette is equipped with a spring, hold the
piston at its lowest position and turn the volume adjustment knob gen-
tly anti clockwise to a position which does not allow any axial move-
ment of the piston. DO NOT USE FORCE! This is your ‘0’ position.
5. Press the ‘CAL’ button again –
the display defaults to a flashing 0.0.
6. From this ‘0’ position with the display still flashing the 0.0 volume, turn
the volume adjustment knob to display the nominal volume which was
used to establish the actual volume.
7. Once the unit displays the nominal volume, press ‘CAL’. The display stops
flashing. Using the up/down arrows, correlate the volume on the display to
the actual volume, which was established in the gravitational procedure.
8. Press the ‘CAL’ button. The display shows the word ‘end’ to indicate
that you have sucessfully completed the calibration procedure.
If you wish to terminate the calibration procedure prematurely, press the
‘On’ button at any stage. The display will show ‘Esc’ for two seconds to
indicate that you escaped the calibration procedure and the display will
revert to show the volume the Zippette is ready to dispense.
6Operator Instructions
3.
4.
5.
6.
7.
8.

Quick Zero reset
Zero reset is useful in those circumstances where the unit displays 0.0 but
the piston is not in its actual Zero position. This is for example possible
after Zippette has been disassembled for cleaning. To reset your true ‘0’
position follow these steps:
1. Press and hold the ‘CAL’ and ‘ON’ buttons simultaneously for 5 seconds in
order to enter the quick ‘0’ calibration mode of the instrument. As soon
as the display shows ‘Set’ release the buttons and the diplay flashes ‘CAL’.
2. Lower the piston to its lowest stop position. Where Zippette is
equipped with a spring, hold the piston at its lowest position. Turn the
volume setting button anti clockwise to a position which does not
allow any axial movement of the piston. This is your ‘0’ position.
3. Press the ‘CAL’ button again – the display shows the word ‘END’ to
indicate that you have successfully reset the zero position.
How to Toggle between Factory Cailbration and
Customer Calibration
Returning to the factory set calibration is easy.
A. Press and hold the ‘ON’ and ‘DOWN’ buttons simultaneously for 5
seconds until the display shows the word ‘FAC’.
To return to your user stored calibration:
B. Press and hold the ‘ON’ and ‘UP’ buttons simultaneously for 5 seconds
until the word ‘USE’ is shown on the display.
For quick check on current calibration mode of unit press ‘ON’ button
momentarily and ‘FAC’ or ‘USE’. will be displayed momentarily.
‘Sleep Mode’ and Low Battery indication
The unit is designed to operate 1700 hours before the battery requires replacement. In order to preserve
the battery life, the unit reverts to ‘sleep mode’ after being idle for 3 minutes. A light touch on the ‘ON’
button will ‘wake’ the unit up – alternatively, you can rotate the volume setting button and the unit will
display the volume on the LCD.
Whenever the battery requires replacement, the word ‘bat’ will flash on the screen for two seconds just
after waking the unit up. You can still continue to work with the unit for eight hours.
When you first power up the unit from sleep mode, the display will show which of two calibtration
modes it is using - either the factory calibration or the user calibration. ‘FAC’ indicates Factory whilst
‘USE’ indicates user calibration.
Operator Instructions 7
Calculation ( for the nominal volume)
Mean value ∑xi
x = n
xi= results of the weighings
n = number of weighings
Mean volume V = x * Z
Z = correction factor
( eg. 1.0029 µl /mg t 20°C, 1013 hPa)
Calculate the dispensed volume by taking the
temperature into account
(1 ml distilled water at 20°C = 0.997g)
1.
1B.
3.
A.
B.
2.

8Operator Instructions
Drawing 8 Drawing 9
Self tapping
screws
Battery
182-987
Positive
Negative
Battery replacement
1. To replace the battery, separate the two piston sleeve housing halves (see pages 4 & 5 – follow steps
3 & 4). On the back of the front housing half, undo the two self tapping screws (Drawing 8) using
no.0 pozi screwdriver. These screws hold the display case onto the front piston sleeve
2. Remove the display case (Drawing 9)
3. Carefully unclip the battery from it’s seating and replace (Drawing 9). Please note to insert the new
battery with the positive pole to the right looking at the back of the unit
Use only Lithium Thionyl Chloride 3.6V batteries
4. With battery inserted display will flash ‘CAL’. Press ‘ON’ button to cancel
5. Reassemble and follow quick zero reset procedure on page 7 before further use
Quick guide to button press combination
‘CAL’ + ‘UP’ enter user calibration
‘CAL’ + ‘ON’ quick zero reset
‘ON’ + ‘UP’ revert to user calibration
‘ON’ + ‘DOWN’ revert to factory settings

Operator Instructions 9
Maintenance/Cleaning
Note; All maintenance should be carried out wearing suitable eye protection and protective clothing. If
in doubt, consult your safety officer.
1. Make sure that the Zippette is completely empty and turn the anti-drip tap to ‘Open’ position.
2. Place the instrument into an empty sink together with its reservoir. Unscrew the threaded platform
base from the reservoir and lift the dispenser’s intake tube carefully out of the reservoir, whilst tap-
ping it against the reservoir’s aperture to shake off any droplets from the intake tube.
3. Hold the dispense nozzle over the aperture of the reservoir and apply gentle piston strokes in order
to return any syringe contents into the reservoir.
4. Flush out remaining syringe contents with distilled water or a suitable solvent. This will preserve the
smooth action of the piston and free action of the inlet and outlet valves. If the inlet valve does stick
and is not freed by flushing, it may be freed by gently inserting a thin rod into the inlet aperture and
gently pushing the ball off its seating. Re-check unit for operation.
5. Empty the instrument completely after cleaning, and flush through with distilled water.
Sterilisation/Autoclaving
Only the fluid path components of the Zippette Digital are suitable for chemical sterilisation or autoclav-
ing at 121°C after the normal cleaning procedure has been carried out.
Chemical sterilisation
The fluid path components of the Zippette can be soaked overnight in a dilute (1:1000) solution of
Sodium Hypochloride. (See page 10 for disassembly procedure).
For soaking the components individually, follow the fluid path disassembly instructions on the following pages.
Autoclaving
Autoclaving of fluid-path components is permis-
sible at 121 °C , 2 bar for 20 minutes, after the
normal cleaning procedure has been carried out.
Autoclaving can be carried out only in a part
assembled condition with the piston inserted into
the glass barrel (Drawing 10) follow disassembly
instruction (page 10) 1 to 3 (Drawings 11-14).
Steam sterilisation of the piston outside the glass
barrel may damage it.
Loosen the threaded platform ring (Drawing 10) by
turning it anti clockwise two full revolutions.
Place the assembly together with the PTFE inlet on
a suitably soft surface within the autoclave in
order to avoid metal contact.
Cool slowly back to ambient temperature after
autoclaving and before reassembly.
Drawing 10
Unit ready for steam
sterilisation
Threaded
platform ring

10 Operator Instructions
Fluid-Path disassembly procedure for Cleaning or Replacement
Disassembly should only be undertaken AFTER the unit has been cleaned, using the standard cleaning
procedure described on page 9. For autoclaving follow steps 1-3.
1. Undo the piston sleeve fixing ring by turning it to align with the sleeve slot and sliding it down from
the sleeve (Drawing 11 & 12).
2. Press the two fixing levers on the back of the piston sleeve inwards, and lever the back piston sleeve
away from the front sleeve assembly (Drawing 12 & 13).
3. Remove the front piston sleeve (Drawing 13 & 14).
4. Pull out the piston assembly from the glass barrel (Drawing 15).
5. Undo the threaded platform ring of the base platform (Drawing 16) and, turn the stop ring clock-
wise and pull it off the assembly (Drawing 17).
6. Carefully slide off the fixing ring and the threaded platform ring, together with the clear
Polypropylene protection sleeve (Drawing 18).
7. Take out the glass barrel from the base platform. Under the glass barrel is an encapsulated FEP ‘O’
ring which should be handled with extreme care (Drawing 18).
8. To remove the FEP dispense tube (Drawing 19): firstly turn the anti-drip tap to closed position then
lift off the dispense tube protective cover and pull out the dispense tube from its seating.
9. Reassemble the instrument following the above points in reverse order. Make sure that the FEP ‘O’
ring is well located into the base platform’s recess.
Important points to observe:
• Do not use hard tools to scrape off residue of reagent from the piston
• Take special care not to damage the piston when inserting it into the glass barrel
• Bevelled inside edge of the glass barrel must be at the top end when fitted
• Before tightening the threaded ring fully onto the threaded platform base, rotate the Polypropylene
protection sleeve so that the tooth on its base locates properly into the notch on the threaded ring
• Make sure the front end of the dispense tube is clipped into the dispense tube protection cover from
the underside
• Do not use force in assembly or disassembly
• Always wear protective clothing and eye protection when disassembling or assembling your Zippette.
Protective clothing and eye protection should also be worn when dispensing

Operator Instructions 11
Fixing ring
263-081
Front sleeve
Back sleeve
Piston
assembly
263-076
Fixing ring
263-081
Fixing levers
Drawing 14
Drawing 11 Drawing 12
Drawing 15
Drawing 13

12 Operator Instructions
Drawing 18
Drawing 16
Drawing 19
Drawing 17
Threaded
platform ring
263-124
Stop ring
263-103
Platform
ring
263-124
Borosilicate
glass barrel
263-120
Encapsulated
‘O’ ring
263-123
Zippette Digital pedestal
complete assembly
263-078
Fixing ring
263-081
Protection
sleeve
263-117
FEP dispense tube
263-069
Dispense tube
protective cover
263-070
Anti-drip tap valve

Operator Instructions 13
Problem Possible Cause Remedy
Air bubbles appear in discharge nozzle Liquid reservoir is empty Refill reservoir and prime unit
Too fast filling action Fill and dispense more slowly
Glass barrel is not sealing against FEP 'O' ring Unscrew the threaded platform ring and make sure that the
FEP 'O' ring is properly fitted into its recess
Leaking piston Clean piston. If problem persists, replace piston
Leaking discharge valve Clean unit by flushing through - if problem persists,
replace platform base
Barrel does not fill with liquid Inlet tube not fitted correctly Connect inlet tube correctly
Inlet valve stuck Free inlet valve by inserting a thin rod into the inlet aperture
Glass barrel is not sealing against the FEP 'O' ring Unscrew the threaded ring and make sure that the
FEP 'O' ring is properly fitted into its recess
Dispensing not possible Blocked dispense nozzle Disassemble the dispense nozzle and flush through
with cleaning fluid
Discharge valve stuck Clean unit by immersing platform in cleaning fluid
(taking care to protect LCD)
- if problem persists, replace platform base
Wrong dispense volume Instrument not calibrated See page 6 ‘User Calibration’
Leaking valves Clean platform base - if problem persists, replace
Liquid appears between glass barrel Glass barrel is not sealing against the Unscrew the threaded ring and make sure that the
and Polypropylene protection sleeve FEP 'O' ring FEP 'O' ring is properly fitted into its recess
Damaged FEP 'O' ring Replace FEP 'O' ring
Troubleshooting

14 Operator Instructions
Accessories
Adaptors
A range of five adaptors for fitting your dispenser to the reservoir are available.
Reservoirs
Description Cat. No.
1.4 litre 2-neck, amber Squat 262-086
2.5 litre 2-neck, amber Squat 262-103
2.5 litre 2-neck, amber Winchester 262-087
5 litre 2-neck, clear pyrex 262-088
2.5 litre single neck, clear glass, with 262-096
plastic coating, fingergrip handle
4.5 litre single neck, amber glass with plastic coating, 262-097
fingergrip handle and supplied with adaptor 222-000
Specification
Accuracy ≤±0.3% on maximum delivery and a precision ≤0.1%CV using distilled water at 20ºC.
30mm
33mm
30mm
38mm
30mm
40mm
30mm
45mm
30mm
115mm
262-140 262-129 262-128
262-130 262-084 262-085
B24/40 Angled funnel

Consignes d’utilisation 15
Sommaire
Consignes générales de sécurité . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Cette page
Restrictions d'utilisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Avant d'utiliser la Zippette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Montage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Consignes d'utilisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Introduire le ressort de rappel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Le bouton de réglage du volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Calibrage personnalisé . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Remise à zéro . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Mode de veille et indicateur de décharge de batterie . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Remplacement de la pile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Entretien/Nettoyage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Stérilisation/Stérilisation à l'autoclave . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Procédure de démontage du circuit du soluté . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Diagnostic des pannes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Accessoires . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Soupape ouverte Soupape fermée
(2) Tête d'affichage
électronique
Consignes générales de sécurité
Ne laissez jamais la Zippette sur votre plan de travail avec le cylindre plein sans avoir correctement refer-
mé la soupape anti-goutte
Vérifiez toujours que la soupape anti-goutte est sur la position 'Ouverte' avant toute utilisation
• Avant de régler le volume, placez toujours le réservoir sous le pistolet et ouvrez la soupape
• Ne forcez jamais !

16 Consignes d’utilisation
Restrictions d'utilisation
N'utilisez JAMAIS la Zippette avec :
• des liquides qui ne sont pas compatibles avec le PTFE, PVDF, FEP, le verre borosilicate, la céramique
d'alumine ou qui sont susceptibles d'attaquer le platine iridié
• De l'acide fluorhydrique
• Des liquides contenant des particules solides
La Zippette et le réactif doivent être utilisés à des températures comprises entre 15°C et 40°C
Avant d'utiliser la Zippette
Vérifiez que l'appareil n'a pas été endommagé pendant le transport.
Ne tenez pas l'appareil par la tête d'affichage électronique (2) lorsque vous réglez ou transportez
l'instrument.
La Zippette Digital est fournie avec une pile lithium non rechargeable. La durée de vie de la pile est
supérieure à 30 000 titrages de trois minutes.
La pile doit être jetée de façon appropriée, et elle ne doit en aucun cas être brûlée.
Montage
La Zippette est livrée avec le pistolet de distribution monté et le tube d'admission non monté. Ce tube
en PTFE d'une longueur de 300mm devra être coupé pour être adapté à votre réservoir personnel. Le
tube doit être coupé en diagonale à son extrémité inférieure, à une longueur proche du fond du récipi-
ent lorsque la Zippette est en place. Des tubes d'admission plus longs sont disponibles sur demande.
La base de plate-forme filetée de la Zippette possède un collet de vis de 30mm qui doit être utilisé avec
le réservoir ambré de sécurité Jencons de 1,4 litres avec un col de remplissage latéral (Cat. No 262-086)
ou avec le réservoir ambré de 2,5 litres (Cat No 262-103). La Zippette assemblée se fixe sur le réservoir
en vissant uniquement la base filetée avec précaution. Il doit être retiré en dévissant uniquement la base.
N'actionnez pas le piston tant que l'appareil n'est pas monté correctement sur le réservoir !
Trois adaptateurs sont fournis pour des récipients avec des collets de vis de 38mm, 40mm ou 45mm.
Consignes d'utilisation
Le tableau de commande
La Zippette Digital possède quatre boutons-pous-
soirs à rappel :
1. 'On' - met l'appareil sous tension et affiche les
derniers chiffres
2. 'CAL' - permet un calibrage de l'instrument
3. Flèche vers le haut - pour augmenter le volume
affiché dans le mode calibrage
4. Flèche vers le bas - pour réduire le volume
affiché dans le mode calibrage
Amorçage
• Placez un récipient sous le pistolet de distribu-
tion de la Zippette
• Tournez le bouton de réglage du volume
jusqu'à ce que l'unité affiche 30ml pour per-
mettre un mouvement libre du piston
• Ouvrez la soupape anti-gouttes
• Amorcez l'appareil en appuyant plusieurs fois, en
descendant le piston jusqu'à sa position d'arrêt
la plus basse et en le relevant d'environ 25mm
• Recommencez l'opération jusqu'à ce que le flux
s'écoule en continu et ne présente aucune bulle

Consignes d’utilisation 17
Introduire le ressort de rappel (le cas échéant)
Pour introduire le ressort de rappel à amorçage automatique, suivez les étapes suivantes :
Illustration 1 Illustration 2
1. Vérifiez que la Zippette est entièrement vide.
Pour faciliter l'opération, montez la Zippette
sur un réservoir vide et mettez la soupape
anti-gouttes sur la position "Ouverte".
2. Tournez le bouton de réglage du volume pour
régler la Zippette sur le volume de distribution
de 30ml (illustration 1).
3. Tournez la bague de fixation qui se trouve au
bas du manchon de la Zippette jusqu'à ce
que sa rainure bombée soit alignée avec la
fente du manchon du piston avant et enlevez
en tirant doucement (Illustration 1). La bague
de fixation doit rester sur le manchon de pro-
tection du cylindre (Illustration 2).
4. Enfoncez les deux leviers de fixation au dos
du manchon du piston (derrière l'écran) vers
l'intérieur (Illustration 2) et écartez le man-
chon du piston arrière du manchon avant
(Illustration 3).
5. Enlevez le manchon avant du dispositif de
positionnement du piston et sortez le piston
de son cylindre en verre (Illustrations 3 et 4).
6. Guidez avec précaution le piston au travers
du ressort et remettez le piston dans son
cylindre en verre (Illustration 5).
7. Réassemblez le manchon du piston de la
Zippette (Illustration 5) en vous assurant que
le curseur est placé au-dessous de la bague
de butée (Illustration 5). Remettez le man-
chon du piston en place et remontez la bague
de fixation (Illustration 6) - Tournez-la d'un
demi-tour pour la placer sur le manchon du
piston de la Zippette (Illustration 7). L'unité
est prête à l'emploi (voir les consignes d'utili-
sation page 3).
Réglage du volume
Fente du manchon
avant
Manchon avant Manchon
arrière
Leviers de
fixation
Manchon de
protection
de cylindre
263-117
Curseur
Nervure bombée
Bague de plate-forme
filetée 263-124
Bague de fixation
263-081
Position ouverte
de soupape
Base de
plate-forme
filetée

18 Consignes d’utilisation
Illustration 3 Illustration 4
Illustration 5 Illustration 6 Illustration 7
Bague de butée
263-103
Curseur
Ressort de rappel
263-075

Consignes d’utilisation 19
Le bouton de réglage du volume
Si vous tournez le bouton de réglage vers la droite, vous augmentez le volume distribué. Chaque tour
complet du bouton représente 2ml.
Pour effectuer un réglage précis, tournez le bouton d'un quart de tour pour obtenir un réglage de 0,5ml.
Calibrage personnalisé
Votre Zippette Digital a été étalonnée pour un volume nominal de 30 ml. Cependant, à cause du
changement des conditions environnementales et de la variabilité de la viscosité du milieu que vous dis-
tribuez, un nouveau calibrage peut s'avérer nécessaire. Vous pouvez recalibrer à des intervalles réguliers
comme une fois par semaine où à chaque fois que vous remarquez que le volume distribué est différent
du volume affiché par l'unité.
Pour recalibrer votre Zippette Digital, veuillez suivre la procédure suivante :
1. Réglez la Zippette Digital sur le volume nominal de 30 ml ou tout autre
volume qui représente le volume que vous utilisez le plus souvent.
2. Suivez les règles de calibrage utilisées dans le contrôle de qualité statis-
tique (ISO DIS 8655/2) Distribuez cinq doses complètes d'eau distillée à
20°C pour établir le volume moyen de liquide distribué. La formule ci-
jointe vous permettra de suivre des règles de base. Si le résultat moyen
gravitationnel varie par rapport au volume affiché, vous devez calibrer
de nouveau l'appareil.
3. Enfoncez simultanément les boutons 'CAL' et 'UP' pendant 5 minutes
afin d'activer le mode de calibrage de l'instrument. Dès que vous êtes
en mode calibrage, le mot CAL se met à clignoter sur l'écran.
4. Alors que le mot CAL clignote, mettez le piston dans sa position d'arrêt la
plus basse. Si votre Zippette est équipée d'un ressort, maintenez le piston
dans sa position la plus basse et tournez le bouton de réglage du volume
doucement vers la gauche, sur une position qui ne permet pas un mou-
vement axial du piston. NE FORCEZ PAS ! Ceci est votre position '0'.
5. Appuyez de nouveau sur le bouton CAL - L'affichage présente 0.0 par
défaut.
6. A partir de cette position '0', alors que l'écran affiche par intermittence
le volume 0.0, tournez le bouton de réglage du volume pour afficher le
volume nominal qui a servi à établir le volume réel.
7. Dès que l'appareil affiche le volume nominal, appuyez sur 'CAL'.
L'écran s'arrête de clignoter. A l'aide des flèches haut et bas, corrélez le
volume de l'affichage avec le volume réel qui a été établi dans la procé-
dure gravitationnelle.
8. Appuyez de nouveau sur le bouton 'CAL'. L'affichage présente le mot 'end'
pour indiquer que la procédure de calibrage a été effectuée avec succès.
Si vous souhaitez terminer la procédure de calibrage avant la fin, appuyez
sur le bouton 'On'. L'affichage présente 'Esc' pendant deux secondes pour
indiquer que vous avez quitté la procédure et l'écran retourne au volume
que la Zippette est prête à distribuer.
3.
4.
5.
6.
7.
8.
Table of contents
Languages:
Other JENCONS Dispenser manuals