JOSAM SUPER FLO GI-2000 Series Instruction Manual

SUPER FLOTM GI-2000
FULLY AUTOMATED
GREASE RECOVERY DEVICE (G.R.D.)
OPERATION, INSTALLATION
AND
MAINTENANCE MANUAL
PLEASE READ CAREFULLY
Grease Recovery Device will operate efficiently only when properly installed.
Do not discard this manual. Manual to be distributed to installer, operator and
maintenance department.
WARNING: Cancer and Reproductive Harm - www.p65warnings.ca.gov

TABLE OF CONTENTS
UNPACKING ................................................................................................................ 1
PRIOR TO INSTALLATION.......................................................................................... 2
WARNINGS .................................................................................................................. 3
PLUMBING INSTALLATION ..................................................................................... 3-5
ELECTRICAL INSTALLATION ................................................................................. 6-7
TESTING.................................................................................................................. 8-10
MAINTENANCE..................................................................................................... 10-11
CONTROLLER OPERATING PROCEDURES ...................................................... 13-21
CONTROLLER WARNINGS ...................................................................................... 12
UNIT INITIALIZATION ................................................................................................ 12
EXTERNAL CONTROLS............................................................................................ 12
ENCLOSURE CONTROLS......................................................................................... 13
MANUAL CYCLE................................................................................................... 13-14
AUTOMATIC CLEAN CYCLE ............................................................................... 15-16
CALIBRATION MODE........................................................................................... 16-17
SETTING PROBE CALIBRATION ........................................................................ 17-19
CLEAN CYCLE RATE ADJUST MODE ................................................................ 19-20
DATA LOGGING RESET ........................................................................................... 21
TROUBLESHOOTING GUIDE ................................................................................... 22
WARRANTY INFORMATION ..................................................................................... 23

1
1. UNPACKING
1.1 After unpacking the GI-2000, the unit should be checked for any damage
that may have occurred during shipment. Any damage should be reported
to Josam Company immediately.
The following items should be included with the shipment:
QTY DESCRIPTION
1 GI-2000 G.R.D.Unit
1 GI-2000 Flow Control Fitting
1 GI-2000 Manual
1 15’ Controller Cable
1 GI-2000 Controller
1 Optional Reclaim Tank
GI-2000 UNIT
DIAGRAM “A”
208/220 Wiring Conduit Port

2
2. PRIOR TO INSTALLATION
2.1 Ensure that you have selected the proper size JOSAM GI-2000 so that it
can never exceed the maximum rating of the grease recovery device to
suit your particular requirements. Size is based on Flow Rate per PDI
G-101 and ASME A112 14.4.
SIZING TABLE
Fixture-Equipment
Drain Outlet or
Trap Size
Drainage
Fixture-Unit
Value
Drainage G.P.M.
Equivalent
1-1/4” 1 7.5
1-1/2” 2 15.0
2” 3 22.5
2-1/2” 4 30.0
3” 5 37.5
4” 6 45.0
2.2 Josam Company has shipped the proper flow control fitting to suit the
GRD that you have ordered. Ensure that the proper flow control fitting is
included with your shipment. Changing the flow control fitting is not
recommended and will effect the performance and warranty of the GRD.
Model
Flow Rate
(GPM)
Grease
(Lbs.)
Inlet
Size
60305A 20 40 3”
60306A 25 50 4”
60307A 35 70 4”
60308A 50 100 4”
60309A 75 150 4”
60310A 100 200 4”
60311A 150 300 4”
60312A 200 400 5”
60313A 250 500 5”
60314A 350 700 6”
60315A 500 1000 6”
Flow Control
Rating (GPM)
Orifice
Diameter
20 1-1/8”
25 1-1/4”
35 1-1/2”
50 1-3/4”
75 2-1/8”
100 2-7/16”
150 3”
200 3-7/16”
250 3-3/4”
350 4-7/16”
500 5-1/8”

3
3. WARNINGS
3.1 The GI-2000 uses heating elements in it’s operation. Therefore, the unit
must always be filled with water. Elements damaged by electricity prior to
the unit being filled with water will void any said warranties.
3.2 To maintain optimum system performance and accuracy it is necessary to
remove solids accumulations in the tank on a regular basis.
4. PLUMBING INSTALLATION
4.1 INSTALLER
Installation is to be performed by a licensed and certified plumber.
4.2 LOCATION OF GRD
The Plumbing & Drainage Institute (PDI) recommends that the unit be
located within 25’ or as close as possible to the source of grease being
served. The unit may be set on the floor, partially recessed or completely
recessed below the floor. If the unit is to be recessed you must have
proper installation of the electrical hook-up, drain line and LCD control
unit.
4.3 CLEARANCE
When installing, ensure that there is 6” minimum clearance per side for
latch removal. Also provide a minimum of 24” on the outlet end in the
event that the heating elements require service. Provide adequate space
on the inlet side for installation or flow control servicing. When installing
the GI-2000 in a pit/vault provide a drain or a sump pump to prevent water
damage.
4.4 LEVELING THE UNIT
The GRD must be leveled from front to back and from side to side. The
unit must be leveled prior to calibration.
4.5 IMPORTANCE OF FLOW CONTROL FITTING
Extensive test reports conclude that both satisfactory operation and
maximum efficiency of the GI-2000 is dependant on the proper rate of flow
for the GI-2000 connected, which is accomplished by a flow control fitting.
As a result, a specially designed Josam flow control fitting with a 3”, 4”, 5”
or 6” threaded inlet and outlet connection, depending on size of unit, and a
threaded air intake is furnished with every GI 2000 at no extra cost. The
flow control must be installed to prevent overloading the trap and to
maintain maximum grease retention efficiency.

4
4.6 FLOW CONTROL LOCATION
The GI 2000 must be correctly installed with the flow control as per model
plumbing codes or the local authority having jurisdiction. The GI-2000
must have a flow control device properly vented at the inlet or in close to
proximity to the GRD inlet, thus allowing the fixtures to be plumbed and
vented as required without having any negative impact on the GRD or flow
control device whose purpose is integral to the performance of the GRD.
(See Diagram B)
4.7 FLOW CONTROL INSTALLATION
Install air intake vent from flow control to outside opening and fit elbow,
turned down; or intersect a common vent or vent stack without a trap as
permitted by local codes. Top of air intake from Flow Control is to be at
least 6” above flood rim level of the lowest fixture served. (See Diagram B)
It is recommended that the air intake be tied to the vent stack.
4.8 VENTED WASTE
The GI 2000 must have a vented waste (outlet), sized as required by local
code or local authority having jurisdiction for venting traps to retain water
seal and prevent siphoning. (See Diagram B)
4.9 ADDITIONAL REQUIREMENTS
When local codes require floor drains to flow through a GRD, the flow
control must be accessible and installed prior to the inlet of the unit.

5
DIAGRAM “B”
Locate the installation diagram which most closely suits your requirement.
Install the flow control and GI 2000 as shown.

6
5. ELECTRICAL INSTALLATION
5.1 INSTALLER
The electric service and connections to the GRD must be installed by a
licensed certified electrician.
5.2 POWER HOOKUP
WARNING! Do not apply power to the unit until the GI-2000 is completely
full of water. Install this unit as any GRD would normally be installed with
the exception to provide a 220/208VAC, single phase, Ground Fault
Circuit Interrupter (do not apply power to the unit at this point), 4 wire
connection hard wired to the main power box side of the GI-2000 per
code. The electrical installation must conform to NEMA-4 standards. The
main power wire cannot be run through a wall and must be wired to the
main breaker panel so that it can be viewed from the GI-2000 area.
Model
Voltage
Heating
Watts
GFI
Breaker
60306A 220/208 VAC 5000 Watts 25 AMP
60307A 220/208 VAC 7000 Watts 40 AMP
60308A 220/208 VAC 7000 Watts 40 AMP
60309A 220/208 VAC 7000 Watts 40 AMP
60310A 220/208 VAC 9000 Watts 50 AMP
For special order models not listed consult Josam Company for power
requirements.
5.3 NEMA-4 BOX
Red:.120VAC (L1)
Black:.120VAC (L2)
White:.Neutral
Green:.Ground
Wire Type:.(Type MTW) 10 Awg min.
Conduit Description:.1” Liquid-Tight Flexible Nonmetallic Conduit
Conduit Part #:.7581K44
WARNING! System power should be secured into the power inlet box
(access via heater element port as shown in Diagram A) in a method that
conforms to NEMA-4 standards. Wire nuts are used to connect the power
wires behind the power conduit port.
5.4 INITIAL POWER TEST
After confirming that the unit is level, fill the GI-2000 with water before
applying power to the unit. Verify with a current probe that L1 and L2 are
below 1 amp. Once confirmed, remove power from the unit.

7
5.5 CONTROL
Mount the GI-2000 controller conduit port on the GI-2000. The control
wiring conduit port is the round connector near the top of the GRD (See
Diagram A). Connect the controller cable to the control wiring conduit port.
Should additional length be required use Beldon type 8469 (9 Conductor
Cable). DO NOT exceed 200’.
5.6 RECLAIM TANK
The GI-2000 is shipped with a ¾” NPT brass outlet fitting. The fitting if
removed is a ¾” M.I.P. fitting. Maximum head is 15’ (maintain ¾”
connection). Appropriate material should be used to accommodate
possible temperatures of 140°F, as well as vibrations, when pumping.
When applications warrant a distance between the GI-2000 and the
reclaim tank, the piping must be sloped toward the reclaim tank. Heat
tracing may also be required. Venting is required.
5.7 RECLAIM TANK ALARM
A reclaim tank alarm is suggested and available as an optional feature
(See Diagram C). Contact Josam Company or your local Josam
representative for additional pumping requirements and available options.
DIAGRAM “C”

8
6. TESTING
6.1 G.R.D. TEST
6.1.1 Confirm that the unit is level and unlatch and lift off the large cover
on the GI-2000 tank.
6.1.2 Run water through the drain installation until the unit is full. The unit
is full when water runs out of the outlet of the GI-2000.
6.1.3 Check the system installation for leaks.
6.1.4 Turn the Auto/Manual mode switch located on the front of the
controller to the MANUAL position.
6.1.5 Apply power to the GI-2000. The controller will beep and start the
initialization procedures. Wait until the display indicates the unit is
in MANUAL mode. This will take approximately 15 seconds. Should
any part of the diagnostics fail, the unit will beep every 5 seconds
and display a message about the failure. (See Controller Operating
Section and Troubleshooting Guide for details on addressing
possible problems with the GI-2000)
6.2 HEATER TEST
6.2.1 Open the controller door by loosening the front panel screws on the
controller door.
6.2.2 Press the TEST button once and the following message will be
displayed:
PRESS UP OR DOWN KEYS UNTIL
DESIRED TEST APPEARS
6.2.3 Press UP or DOWN until the following message is displayed:
HEATER TEST (DO NOT RUN DRY): HEATER:OFF
PRESS ENTER TO CYCLE HEATER STATUS
6.2.4 Press ENTER to turn the heaters on and ensure that they are
operating. Evidence that they are operating can be detected by
heat waves and bubbles forming around the heater elements.
Pressing ENTER again will turn the heater off. Check the display
and green LED indicator to ensure proper operation. The heaters
will also automatically shut off after three minutes.
WARNING! Do not run the heater dry! Running the heater dry may
cause damage to the heaters and the unit.
6.3 PUMP TEST
6.3.1 Open the controller door by loosening the front panel screws on the
controller door.

9
6.3.2 Press the TEST button once and the following message will be
displayed:
PRESS UP OR DOWN KEYS UNTIL
DESIRED TEST APPEARS
6.3.3 Press UP or DOWN until the following message is displayed:
PUMP TEST (DO NOT RUN DRY): PUMP:OFF
PRESS ENTER TO CYCLE PUMP STATUS
6.3.4 Press ENTER to turn the pump on and observe that the pump is
running and water is being pumped into the reclaim tank.
WARNING! Do not run the pump dry! Pressing ENTER again will
turn the pump off. Check the display and green LED indicator to
ensure proper operation. The unit will also automatically turn the
pump off after 60 seconds.
6.4 TEMP SWITCH TEST
6.4.1 Open the controller door by loosening the front panel screws on the
controller door.
6.4.2 Press the TEST button once and the following message will be
displayed:
PRESS UP OR DOWN KEYS UNTIL
DESIRED TEST APPEARS
6.4.3 Press UP or DOWN until the following message is displayed:
TEMP.SWITCH TEST: SWITCH STATUS: OFF
6.4.4 Using a meter confirm change in status (continuity) when cycling
switch.
6.4.5 Press ENTER to change the switch status to ON on the LCD.
6.5 WATER LEVEL TEST
6.5.1 Open the controller door by loosening the front panel screws on the
controller door.
6.5.2 Press the TEST button once and the following message will be
displayed:
6.5.3 Press UP or DOWN until the following message is displayed:
WATER LEVEL TEST:
WATER LEVEL: XX.XINCHES
PRESS UP OR DOWN KEYS UNTIL
DESIRED TEST APPEARS

10
6.5.4 Verify that the unit reads a minimum of 4 inches of water and that
the tank level is full of water only with little to no grease. If the tank
level appears out of range run a calibration on the system (See
Section 7.5 of Controller Operating Section).
6.6 DISPLAY & LED TEST
6.6.1 Open the controller door by loosening the front panel screws on the
controller door.
6.6.2 Press the TEST button once and the following message will be
displayed:
6.6.3 Press UP or DOWN until the following message is displayed:
DISPLAY & LED TEST:
PRESS ENTER TO START TEST
6.6.4 Press ENTER to start the test. After pressing ENTER the LED will
be lit and test messages will be displayed on the LCD. Pressing UP
will revert to the previous test. Pressing DOWN will display the
following message:
END OF TEST MODE
PRESS ENTER TO EXIT.
6.7 COMPLETION OF TESTING
6.7.1 To exit test mode press Test/Cancel button twice.
6.7.2 Carefully close the control front panel and reinstall the locking
screws.
6.7.3 Turn the Auto/Manual switch on the front panel to AUTO.
6.7.4 Install the top cover of the grease trap section.
7. MAINTENANCE
7.1 To maintain optimum system performance and accuracy it is important to
remove all solids that have accumulated in the solids interceptor. Do not
allow any solids to build up in the GI-2000.
7.2 It is recommended that owner maintain a daily maintenance log to insure
GI-2000 maximum performance. See sample Maintenance Log on next
page.
PRESS UP OR DOWN KEYS UNTIL
DESIRED TEST APPEARS

11
JOSAM GI-2000 MAINTENANCE LOG PART NUMBER: ___________________
MONTH: _________________ SERIAL NUMBER: _________________
CONTACT: Chas Tevis (Phone) 215-339-5370 Ext. 212 (Fax) 800-962-3312
Date Power Mode Cycles Completed Water Level Cleaned Solids Reclaim Tank Status Initial Time Comments
Ex. Y Auto 0 – 1 – 1 – 2 4.3 Y Half Filled CT 9:00 AM Temp LED On
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31

12
CONTROLLER OPERATING PROCEDURES
1. WARNINGS
1.1 The GI-2000 uses electric heating elements in its operation. Therefore, the
unit must always be filled with water.
1.2 To maintain optimum system performance and accuracy it is necessary to
remove solids that have accumulated in the solids tank on a regular basis.
2. UNIT INITIALIZATION
2.1 When power is first applied to the GI-2000 the unit will beep and go
through internal diagnostics. Should any part of the diagnostics fail, the
unit will beep every 5 seconds and display a message about the failure.
The critical values of time/date, pump run time and calibration values are
stored in the unit and tested at start-up (See Section 6 of Installation
Section). Should one of the values be determined to be invalid, the unit
will wait for the user to re-enter the correct values.
3. EXTERNAL CONTROLS
POWER
CLEAN
CYCLE
ACTIVE
AUTO/
MANUAL
START/
STOP
Powe
r
Indicates that the unit is powered up.
Clean Cycle Active
Indicates that the clean cycle is in
operation.
Auto/Manual (Sec 5.1)
Allows the operator to select between
automatic clean cycle or manual clean
cycle operation.
Start/Stop
Allows the operator to start the manual
clean cycle operation.
LCD Displa
y
Two (2) line X 40 character
liquid crystal display provides
information and instructions to
operators and maintenance
personnel.
LED Indicators
Display the following modes of
operation:
Ø Heaters On
Ø Pump On
Ø Grease Level High
Ø Temperature Switch On
Ø Activity 1 (Spare)
Ø Activity 2 (Spare)

13
4. ENCLOSURE CONTROLS
5. MANUAL CYCLE
5.1 WHEN
5.1.1 A manual clean cycle is typically initiated when the unit is in the
MANUAL mode. A manual clean cycle is typically initiated when
personnel wish to clean the GRD regardless of grease level or
when supervision is required.
5.2 HOW
5.2.1 To begin a manual clean cycle rotate the Auto/Manual mode switch
to the MANUAL position. The following messages will be displayed
at this point.
MANUAL OPERATING MODE
DyCyc:XX WkCyc:XX MoCyc:XX YrCyc:XXX
MANUAL OPERATING MODE
PRESS “START” TO BEGIN CLEAN CYCLE
5.2.2 To initiate the manual clean cycle press the START button.
5.3 PROCESS
5.3.1 Once the START button is pressed, the clean cycle will begin, the
controller will beep and the following message will be displayed:
MANUAL CLEAN CYCLE INITIATED
X.XX MINUTES BEFORE CLEAN CYCLE STARTS
HTR
ON
TEMP
SW
ON
GREASE
LEVEL
HI
ACTIVITY
1
ACTIVITY
2
PUMP
CAL
MODE
/UP
CC
RATE
/DN
TEST
/CANCEL
ENTER
RESET
CONTROLLER LCD DISPLAY
Cal Mode/Up
Initiates the water level sensor calibration
mode and adjusts time and date values
in the time and date entry mode.
CC Rate
Initiates the clean cycle rate adjustment
mode.
Test Mode /Cancel
Initiates the test mode operation.
Enter
Mode selection and data entry switch.
Reset
Soft Reset

14
5.3.2 The user can start or stop a cycle at any time during the manual
cycle by pressing the START/STOP button. If the Start/Stop button
is pressed, the cycle will be interrupted for 60 minutes or until the
Stop/Start button is pressed. The following message will flash on
the display periodically during the clean cycle:
PRESS START/STOP BUTTON TO HALT
THE START OF THE CLEAN CYCLE
5.3.3 Once the clean cycle has started, the controller will then turn on the
heaters and the following message will be displayed:
MANUAL CLEAN CYCLE IN PROGRESS
HEATERS ON
The heaters will remain on until the grease temperature is 110°F. If
the temperature is not reached in 8 hours the controller LCD
display will indicate an error. Should this error occur, disconnect
power from the unit and call for maintenance.
5.3.4 Once the preset temperature is reached the GI-2000 will pump the
grease out and the following message will be displayed:
MANUAL CLEAN CYCLE IN PROGRESS
PUMP ON XXXX TIME REMAINING
WARNING, DO NOT RUN PUMP DRY! The GI-2000 has been
pre-calibrated at the factory, prior to shipping, to ensure a suitable
pump time and water sense level. However, if adjustments have
been made it is important to ensure that pump does not run dry!
5.3.5 After the pump cycle is complete, the controller will beep again and
the following message will be displayed:
MANUAL CLEAN CYCLE COMPLETE
5.3.6 After the indication of the cycle being complete the horn will sound
and the unit will wait for the water level in the tank to rise above the
pump inlet. The following message will be displayed:
MANUAL CYCLE COMPLETE
WAITING FOR LEVEL TO RISE
5.3.7 Once the water level rises, the unit display will cycle to the main
manual mode.
MANUAL CYCLE COMPLETE
WAITING FOR LEVEL TO RISE

15
6. AUTOMATIC CLEAN CYCLE
6.1 WHEN
6.1.1 When the unit is in the automatic mode it will perform the entire
cleaning cycle automatically when the grease quantity builds up to
the preset level.
6.2 HOW
6.2.1 To put the system in automatic mode, rotate the Auto/Manual mode
switch to the AUTO position. The following message will be
displayed:
AUTOMATIC OPERATING MODE
DyCyc:XX WkCyc:XX MoCyc:XX YrCyc:XXX
6.3 PROCESS
6.3.1 Once the grease quantity builds up to the preset level, a horn will
sound for 10 seconds to alert personnel using the sinks. The
following messages will be displayed:
AUTOMATIC CLEAN CYCLE INITIATED
X.XX MINUTES BEFORE CLEAN CYCLE START
PRESS START/STOP BUTTON TO HALT
THE START OF THE CLEAN CYCLE
6.3.2 Without any action from the user the clean cycle will automatically
begin. The user may at this time or at any point in the clean cycle
press the START/STOP button to interrupt the cycle. The cycle will
be interrupted for 60 minutes or until the Start/Stop button is
pressed again.
6.3.3 Once the clean cycle has started, the controller will then turn on the
heaters and the following message will be displayed:
AUTOMATIC CLEAN CYCLE IN PROGRESS
HEATERS ON
The heaters will remain on until the grease temperature has
reached 110°F. If the temperature is not reached within 8 hours the
controller LCD display will indicate an error. Should this error occur,
disconnect power from the unit and call for maintenance.
6.3.4 Once the tank temperature is reached the G.R.D. will pump the
grease out and the following message will be displayed:
AUTOMATIC CLEAN CYCLE IN PROGRESS
PUMP ON XXXX TIME REMAINING

16
WARNING, DO NOT RUN PUMP DRY! The system has been pre-
calibrated at the factory, prior to shipping, to ensure a suitable
pump time and water sense level. However, if adjustments have
been made it is important to ensure that pump does not run dry!
6.3.5 After the pump cycle is complete, the controller will beep again and
the following message will be displayed:
AUTOMATIC CLEAN CYCLE COMPLETE
6.3.6 After the indication of the cycle being complete the horn will sound
and the unit will wait for the water level in the tank to rise above the
pump inlet. The following message will be displayed:
6.3.7 Once the level rises, the unit display will cycle to the main
automatic mode.
7. CALIBRATION MODE
7.1 WHEN
7.1.1 The calibration mode is intended to enter the date, pump run time
and calibrate the probe for measuring grease. It is important to note
that the probe calibration senses the water level in the tank.
Therefore, the tank should not have any grease introduced into the
system. The time/date, pump run time and the probe calibration
should typically be checked at a minimum of once a year.
7.2 HOW
7.2.1 Open the controller enclosure and push the CAL/MODE button.
The following message will be displayed:
PRESS UP/DWN KEYS FOR DESIRED CAL
PRESS ENTER TO CHANGE TIME AND DATE
7.3 CANCELLING THE CALIBRATION MODE
7.3.1 Pressing the CANCEL button TWICE at any point during the
calibration mode will exit the calibration mode and revert to the
previous calibration values. Pressing the CANCEL button ONCE
will display the following message:
PRESS CANCEL KEY AGAIN TO EXIT CAL MODE
THIS LINE WILL DISPLAY AS PREVIOUS
AUTOMATIC CYCLE COMPLETE
WAITING FOR LEVEL TO RISE

17
7.4 SETTING TIME AND DATE
7.4.1 WHEN
7.4.1.1 The date and time should be set when re-calibrating the
unit.
7.4.2 HOW
7.4.2.1 After pressing the CAL/MODE button once, the user must
press enter to set the time and date.
7.4.3 PROCESS
7.4.3.1 After pressing ENTER, following the instructions as
displayed on the LCD and illustrated below:
USE UP AND DWN KEYS TO SET MIN, PRESS
ENTER WHEN SET. MIN: XX
Once ENTER is pressed from this screen the user is prompted to
confirm the time and date entry.
TIME: XX:XX DATE: XX/XX/XXXX XXX
PRESS ENTER TO SET, CANCEL TWICE EXITS
A screen is then displayed confirming your selections.
The user is then returned to the Cal/Mode menu.
7.5 SETTING PROBE CALIBRATION
7.5.1 WHEN
7.5.1.1 To adjust the electronics for level of grease sensing. It is
important to remember that the sensor calibration senses
the water level in the tank. Therefore, the tank should be
level and not have had any grease introduced into the
G.R.D. prior to performing the probe calibration
sequence.
USE UP AND DWN KEYS TO SET HOUR, PRESS
ENTER WHEN SET. HOUR: XX
USE UP AND DWN KEYS TO SET MONTH,
PRESS ENTER WHEN SET. MONTH: XX
USE UP AND DWN KEYS TO SET DAY, PRESS ENTER WHEN
SET. DAY: XX
USE UP AND DWN KEYS TO SET DAY OF WEEK, PRESS
ENTER WHEN SET. DAY OF WEEK: XXX
USE UP AND DWN KEYS TO SET LAST TWO DIGITS OF YEAR,
PRESS ENTER WHEN SET. YEAR: XXXX
USE UP AND DWN KEYS TO SET FIRST TWO DIGITS OF YEAR,
PRESS ENTER WHEN SET. YEAR: XXXX
TIME: XX:XX DATE: XX/XX/XXXX XXX
TIME AND DATE UPDATED.

18
7.5.2 HOW
7.5.2.1 Press the CAL/MODE button twice.
7.5.3 PROCESS
7.5.4.1 While in any stage of the calibration mode, pressing any
key other than enter or not pressing any key within 30
minutes will revert the system back to normal operation.
7.5.4.2 After pressing the CAL/MODE button TWICE the
following message will be displayed:
LAST CALIBRATION: XX/XX/XX PRESS ENTER TO
START, ANY OTHER KEY TO EXIT
7.5.4.3 After pressing ENTER the following message will be
displayed:
1.FILL TANK TO LEVEL ONE MARK (LOWER)
THEN PRESS ENTER TO RECORD LEVEL ONE
7.5.4.4 Prior to putting any water into the tank, ensure that the
tank is level and free of any grease or solids. After filling
the tank with water to level one of the calibration mark,
allow the water to settle and press ENTER the following
message will be displayed:
7.5.4.5 Reading level of one will take a few seconds. After the
system reads the calibration level the following message
will be displayed:
LEVEL 1 CALIBRATION
SUCCESSFUL
7.5.4.6 After a few seconds the following message will be
displayed:
2.FILL TANK TO LEVEL TWO MARK (UPPER)
THEN PRESS “ENTER” TO RECORD LEVEL TWO
7.5.4.7 After filling the tank to level two of the calibration mark,
allow the water to settle, and press ENTER the following
message will be displayed:
READING LEVEL TWO CALIBRATION
WAIT
READING LEVEL ONE CALIBRATION
WAIT
Other manuals for SUPER FLO GI-2000 Series
1
This manual suits for next models
11
Table of contents
Popular Lubrication System manuals by other brands

Clarke
Clarke CAT194 Operating & maintenance instructions

Graco
Graco Manzel HP-50 Service maintenance manual

AUTOL
AUTOL ALP01 Series installation manual

Sunshine
Sunshine YET-T1 Operation manual

DROPSA
DROPSA MiQueL EXT User Operating and Maintenance Manual

Norgren
Norgren 10-028-037 Installation & maintenance instructions