Kanmed BABYBED BB101 User manual

User Manual
Art no: BB100-401F-070 ver 3 Eng
2017-03-01
Manufactured by:
Kanmed AB
www.kanmed.se
Distributed by:
This manual is valid for all BabyBeds that has the Red Stoppers in the side walls and
also valid for beds that has been upgraded to hydraulic tilt release and/or new side
walls with Red Stoppers.
Caution
Please read this manual carefully.
Incorrect use may cause patient
or user injury.

User Manual BB100-400 /070
Page 2 of 20
Contents
1 GENERAL DESCRIPTION / INTENDED USE .........................................................................................3
2 WARNINGS ................................................................................................................................................3
3 DESCRIPTION OF SYMBOLS...................................................................................................................4
4 GENERAL INSTRUCTIONS FOR USE .....................................................................................................5
5 CLEANING ................................................................................................................................................10
6 REGULAR SAFETY CHECKS BEFORE USE ..........................................................................................13
7 TROUBLE SHOOTING.............................................................................................................................13
8 TECHNICAL DATA ...................................................................................................................................14
9 TECHNICAL SECTION - SERVICE AND MAINTENANCE ..................................................................15
10 SPARE PARTS/ACCESSORIES............................................................................................................17
11 WARRANTY STATEMENT.....................................................................................................................18
12 EMC GUIDE-LINES ...............................................................................................................................18
13 DISPOSAL ..............................................................................................................................................18
BB101Standard without rails BB100 Standard with rails BB400 Twin Version with rails

User Manual BB100-400 /070
Page 3 of 20
1 Intended Use
Intended Use
The intended use of the Kanmed BabyBed is to provide an ideal ergonomic and safe Bed for
premature and new born babies in a hospital environment. It is also intended to be used with
the Kanmed BabyWarmer systems BW3 or BW-50.
It is intended to be used by hospital personnel or the baby’s caretaker according to the
instructions given in this manual and according to your established clinical routines. Only
persons that have been instructed in the use of the Kanmed BabyBed shall be allowed to use
the product.
2 Warnings
•Do not use the Kanmed BabyBed with babies/small children that are able to move so
that they might fall/climb out of the Bed.
•Ensure that the side walls are locked in the highest position when the baby is
unattended.
•Do not move the Bed around in the high position. Always use the lowest position.
•Do not use the Side Walls to move the Bed –always use depending of model the Side
Rails or front handle.
•Do not leave a baby unattended with the Side Walls down.
•Respect the weight limits.
•Check that the Bed is not wobbly meaning that all wheels and bolts are tightened.
•Ensure that there are no people or objects in the way when changing height.
•Remove the Tent Pole if the Bed is used without a tent.
•The Tent Pole is only to be used for a tent.
•If children, relatives or patients, or patients with special needs are at risk of pressing /
playing with the Foot Pedal, then disconnect the power. Make sure that the Kanmed
BabyWarmer still has power if connected.
•Keep safe distance to other products that can move up or down.
•Max mattress thickness is 40mm
•Only persons that have been instructed in the use of the Kanmed BabyBed shall be
allowed to use the product.

User Manual BB100-400 /070
Page 4 of 20
•
3 Description of Symbols
Read the user manual before use.
Models available:
BB100
Standard BabyBed
BB101
Same but without rails
BB400
BabyBed for Twins
BB401
Same but without rails
BB300
Standard BabyBed fixed height
BB301
Same but without rails
BB400F
BabyBed for twins with fixed height
BB401F
Same but without rails
Maximum safe working loads permitted:
Bed Tray, all versions: Baby(ies) weight 10 kg total + Water Mattress(es) Gel Mattresses
or other mattresses, total 10kg
Storage Box Lid under Bed Platform: 2kg
Storage Box under the Bed Platform: 2kg
Side Rails around the Bed: Max 3,5 kg on each Rail. Max totally 10 kg on all Rails.
Shelf: 5kg. Rear mount kit poles: 3 kg each
Front handle: 5 kg (models without side rails)
Symbol on Foot Pedal for raising and lowering the Bed Platform (not for
BB300 and BB400F)
Symbol for tilt function, found on the Bed Tray Label
Fulfils MDD 93/42 EEC
Earth potential equalization connection point.
IP 30
Degree of protection (against ingress of water and foreign objects)
Protection grade Type B (Body) (not for BB300 / BB400F
When the BabyBed reached end of life, it should be returned to the
distributor for recycling in accordance with the EU 2002/96/EC (WEEE)
directive if applicable.
Year of manufacture.
Maximum minutes
nonstop use of the
electrical Pillar
Total weight of bed when loaded with max permitted weights
Risk of hand injury

User Manual BB100-400 /070
Page 5 of 20
4 Instructions for Use
4.1 First cleaning
The Kanmed BabyBed and accessories must be cleaned according to your hospital routines.
Please refer to section 5 Cleaning before use with a patient.
See section 5 for mounting of side walls, etc.
4.2 Check stability
Make sure that all bolts and screws are properly tightened and that the Bed feels stable.
4.3 Lowering the Side Walls
The Side Walls and the Front Wall are lowered in the following way.
Use both hands and take hold of the upper edges.
Pull straight up (5mm, ¼ inch) –pull towards you (2 cm, 1 inch) –push down gently.
Use both hands to pull the Side Walls up again –push a little forward and then lower it.
Ensure that they are securely in place.
Ensure that all Side Walls are moving correctly and in good condition.
NOTE: Check that nothing gets trapped when you lower and raise the Side Walls.
NOTE: Removing the Side Walls –see section 5 Cleaning
4.4 Moving the Bed around
Always use the Side Rails or the front handle for moving the Bed.
Warning: Use a low Bed height when moving or transporting, especially with a baby inside.
NOTE:
•Do not use the Side Walls for moving the Bed.
•Do not forget to unplug any cables before moving.
•Use Wheel Brakes when Bed is not to be moved.

User Manual BB100-400 /070
Page 6 of 20
4.5 Connecting the BabyBed to the mains (not for BB300/301 and BB400F/401F)
The BabyBed must be connected to an
earthed power outlet using a grounded cable.
Route the cable so that no one can get caught
or fall because of it.
NOTE: The only way to break the power
is to either disconnect the cable at the
Pillar base or at the wall socket.
Place the Bed so that the cable
can easily be disconnected.
WARNING: To avoid the risk of
electrical shock, this equipment must
only be connected to a power outlet
that has protective earth.
4.6 Adjusting height (not for BB300/301 and BB400F/401F)
The height is adjustable by pressing the Foot Pedal
at the wheel base.
Connect power and run the Bed up and down
and ensure that it moves smoothly without
any unusual noise.
The motor can be heard which is an
indication/warning that the Bed Platform is now moving.
Foot pedals can be mounted on the left or right side or on
both sides. Front pedal is also available.
WARNING:
When changing height, make sure that no person, cable or other items prevents the
movement and that no cable is jammed or wrapped between the Pillar and the Wheelbase or
somewhere else.
If children, relatives or patients are at risk of playing with the foot pedal, then disconnect the
power. Make sure that the Kanmed BabyWarmer still has power.
Do not block the movements when activating the electrical height adjustment.

User Manual BB100-400 /070
Page 7 of 20
4.7 Tilting
When tilting the Bed, hold the front Rail or Handle firmly by both hands and press the release
button under the right Corner.
At the same time, press the Bed down or pull up till the desired tilt is achieved. When
releasing the button the bed should stay in the desired position/angle.
Max tilt angle can be limited by a technician. See technical section.
A kit is available to upgrade the tilt function to + and –17 degrees, Trendelenburg position.
Art no BB173.
4.8 Tent Pole, Tent and Mattress
Mount the Tent Pole in the rear bracket.
Turn it to a 90 degree angle from the Bed and
press it down so the screws can pass in the groves.
The Tent has a channel inside.
Thread the Tent Pole into the channel in order to
secure it and place the Tent sides on the outside
of the Side Walls.
Place the Mattress in the Bed.
Other mattress types and homemade tents
may void Kanmed legal responsibility.
NOTE: The tent pole has 2 height positions.
Please make sure you use the right tent pole.
Tilt release button on models
without side rails
Tilt release button on models with
side rails
Stop screw position B100/101/300/301
Stop screw position Twin bed versions

User Manual BB100-400 /070
Page 8 of 20
4.9 Baby Tray
The Baby Tray may be fixed to the Bed Platform or be loose, depending on your order.
Having the Bed Tray loose allows you to lift it out for easy cleaning and for cleaning of the
Bed Platform as well.
In case you want the Baby Tray permanently fixed to the Bed Platform Kanmed can supply a
special tape for this purpose.
4.10 Storage Boxes
Storage Boxes with or without lid can be
mounted under the Bed Platform.
They can be swung around for easy access.
Swing them back in after use.
Respect the load limits written on the
Lid and in the Storage Boxes
4.11 Kanmed BabyWarmer
The Kanmed BabyWarmer Control Unit can be hung on
the Side Rails. A special handle (800-0057) can replace
the existing handle and will reduce
the risk of the unit accidentally falling of the rail.
See www.kanmed.se or contact your distributor.
The Heating Pad cable is preferably thread through
one of the holes in the Bed Tray
before connecting to the Control Unit.

User Manual BB100-400 /070
Page 9 of 20
4.12 Other accessories
Kanmed accessories are mounted and used according to the instruction that comes with the
parts.
Respect weight limits indicated in Section 3
Adding other accessories needs careful consideration regarding weight limits, suitability,
attaching to the Bed, etc.
Kanmed can only be held responsible for accessories delivered by Kanmed.
For all Kanmed accessories see www.kanmed.se or contact your distributor.
4.12.1 BabyWarmer power cable
The Kanmed BabyWarmer is connected to mains power by its own cable or by a special Short
Cable (Order Number 800-0502) that is connected at the upper base at the Pillar –se
picture-
See Kanmed BabyWarmer user manual for further information about the use of the Kanmed
BabyWarmer.
4.12.2 Rear mount Kit
All beds can be equipped with a rear mount kit that
will hold the Baby Warmer Control Unit securely.
Tubes are standard 25 diameter stainless steel tubes.
Ordering numbers are found in Section 10
Max weight on each pole: 3 kg
4.12.3 Optional shelf
A shelf can be ordered instead of the Storage Boxes
Art no: 800-0530
4.12.4 Changes / upgrades Side Rails
It is possible to exchange the bed corners on beds that has side rails (BB100, 300, 400,
400F) so that the Side Wall Guides are no longer needed. Art no 800-0110
It is also possible to exchange bed corners on the beds with side rails so that the side rails
are removed making the bed less wide.
Art no BB-108 for BB100 and BB300.
Art no BB-109 for BB400 and Bb400F
It is of course also possible to upgrade a bed without side rails to have side rails.
Please contact your supplier for details and prices.

User Manual BB100-400 /070
Page 10 of 20
5 Cleaning
General information
The Bed and accessories should be carefully cleaned between each patient. Follow the
routines set forth by the person responsible for cleaning and disinfection in your hospital and
department.
Inspect all parts for wear and strength before and after cleaning.
Below is a list of cleaning and disinfection agents recommended by Kanmed. Using other
agents and/or exaggerated cleaning may reduce the lifetime of some or all parts. Contact
Kanmed with the product specification if in doubt.
The Side Walls are made of PET (eco-friendly) and will withstand all kind of cleaning and
disinfection fluids.
Do not use bleach and solvents.
Cleaning: Use soap water
Disinfection: The Kanmed BabyBed withstands most common surface disinfectants used in
hospitals. Disinfectant wipes with Virkon®, Diversey Limpiador Clorado, Clinell wipes, Chlor
Clean wipes, Meliseptol®rapid, Dax yt, Terralin®,Dsimozon®pur , perform®and similar.
If in doubt try on a small surface or contact your distributor or Kanmed.
SPECIFIC INFORMATION
5.1 Cleaning of the Bed
Use well wrung out cloths.
NOTE: Kanmed BabyBed must under no circumstances be showered with any fluid.
Do not clean it in disinfection chambers.
5.2 The Baby Tray
The Baby Tray is normally loose. It can therefore be lifted out for easy cleaning and for
cleaning of the Bed Platform as well.
5.3 Cleaning Side Walls
Removing Side Walls.
Press the red stoppers inwards to unlock the side
wall and lift it up.
Press the red stopper outwards to lock the Side Wall
Do not wash the Side Walls in a dishwasher.
Don’t scrub the Side Walls with a scratchy material.

User Manual BB100-400 /070
Page 11 of 20
5.4 Cleaning Side Wall Guides.
The Side Wall Guides can easily be removed and cleaned.
Side Wall Guides are marked left L and right R
NOTE: Beds without side rails have no Side Wall Guides. It is possible to exchange the bed
corners on beds that has side rails so that the Side Wall Guides are no longer needed
5.5 Putting Side Walls back in place.
5.5.1 Beds with Side Wall Guides:
Insert the side wall guides in the corners. Observe L left and R right.
Place them as in the picture
Press the pin on the guide firmly into the hole in the corner.
Insert the Side Wall in the groves, see arrow,
and slide them down.
The folded upper edge of the side wall
shall face outwards.
Side walls will easily slide / click in place.
If the Side Walls are not sliding easily you may add
very little grease at the Side Wall edges.
Insert the side wall guides as shown in the picture

User Manual BB100-400 /070
Page 12 of 20
5.5.2 Beds without Side Wall Guides
Slide the side walls in place and press the
red stoppers outwards.
Folded upper edge of the side wall shall face outwards.
Side wall corner for beds without side rails
that eliminates side wall guides
5.6 Storage Boxes and Lids
The Storage Boxes and Lids are cleaned and disinfected same way as the rest of the Bed.
5.7 Mattress
Use a surface cleaning and disinfecting agents recommended by Kanmed above.
5.8 Tent
See label for washing instructions.
In case of contamination the Tent may be washed at 90°C but this may result in shrinking
and an earlier wear out.
5.9 Side wall Corner
If your side wall corner
do not have the small hook as
indicated at the arrow then contact your distributor.
NOTE: The groves on the rear Side Wall
are intended for fixation of tubes.

User Manual BB100-400 /070
Page 13 of 20
6 Systematic safety checks before use
General
Ensure that all parts are in good condition and do not show discoloring or cracks. Change
parts that are not in good condition.
Additional annual maintenance is required. See technical section
Recommended
controls
before use
Executed
by
Time
schedule
Additional information
Function of Side Walls
User
Before
every use
Make sure that they move up and
down easily and that they lock in the
upper position.
Tilting
User
Before
every use
Make sure it moves freely and lock
secure in any angle desired.
Cables and connectors
User
Every day
Make sure that they are in good
condition.
Stability and integrity
User
Every day
Ensure that the Bed is in good
condition and stable and that all
functions works correctly
Kanmed BabyWarmer
User
See Kanmed BabyWarmer user
manual
Wheels / brakes
User
Before
every use
7 Trouble shooting
Symptom
Possible cause
Action
The lifting column does not
move as wanted.
(not for BB300,301 and
BB400F, 401F)
No mains
Check the cables.
Check power outlet
Something obstructing the
movement
Check that nothing
blocks
Electrical connection broken
or not properly connected
Technical service
Foot Pedal faulty or cable not
connected
Technical Service
You cannot tilt the Bed
properly. It gets stuck in
utmost positions.
It does not keep the position
Control is out of order or
needs adjustment
Technical service
Something is obstructing it
Check that nothing
blocks it
Bed feels unstable
Loose screws or wheels
Technical service
Height adjustments makes
unusual noises
Technical service
Side Walls are hard to move.
Not fitted correctly
Check for correct fitting
Add grease
Side Walls are stuck
Guides turned the wrong way
around
See section 5.5
Side wall are not firm, rattles
Side wall guides wrongly
mounted
Mount correctly

User Manual BB100-400 /070
Page 14 of 20
8 Technical data
BB100 / 101 Standard
BB300 / 301 Fixed height
BB400 / 401 Twin versions
BB400F / 401F Fixed height
Outer dimensions
With side rails
L 85cm, W 63cm
Without side rails
L 83cm, W 53cm
With side rails
L 95cm W 80 cm
Without side rails
L 93 cm W 70 cm
Bed Tray
Inner dimensions
L690mm W465mm H90 mm
Made of PP
L765mm W640mm H90mm
Made of PP
Height from floor
of Bed Tray
bottom
Min 78 cm Max 108 cm. Fixed height is 80cm
Ordering 9,5 cm double wheels lowers all beds about 4 cm.
Weight
About 46 kg
Fixed height about 36 kg
About 52kg
Fixed height about 42 kg
Wheel base
64 cm X 87 cm
Wheels
Diameter 125mm with brakes or 9,5 cm double wheels with brakes
Tilt
About 17 degree. Option ±17 degree
Mattress
Soft foam coated
Tent
Washable 60°C (90°C)
Storage Box
L320mm W330mm H120mm, Lid 320x330mm
Colors
Unisex light blue for all plastic details. Wheels and metal parts light grey
Voltage
BB100/400
230V 50Hz / 60Hz or 115V 60Hz, 150VA
Fuses
BB100/400
2xT6,3A 230V. Only replace with same type and rating
Power outlet on Pillar BB100/101 and BB400/401: Max 300W
Protection BB100/101/400/401: Class I
Products is CE-marked.
Conforms to UL 60601-1
Environmental data
In use: +10 to + 40ºC, <85% RH
Transport: -20 to +60ºC, non-condensing.
Storage: +10 to +40ºC, non-condensing.
Atmospheric pressure: 70 to 106 kPa
Modifications
WARNING
Any modifications on Kanmed BabyBed including homemade
accessories may void Kanmed’s responsibilities totally and are not
allowed without written approval of Kanmed.
Creating Systems
Any person connecting the Kanmed BabyBed to any other systems
mains socket may have created a “system” according to IEC 60601-1
section 16, and must be qualified to evaluate any consequences that
may harm staff, patient and equipment.
Expected Lifetime
Kanmed warrants a safe lifetime of 10 years from first day of use. This
is under the condition that the Kanmed BabyBed has been serviced
according to the user and or service manual and that the unit has not
been modified or changed in any way or for any reason.

User Manual BB100-400 /070
Page 15 of 20
9 Technical Section - Service and Maintenance
NOTE: All spare parts used for replacement or repairs shall be supplied by Kanmed or its
authorized distributor. The use of “non Kanmed permitted parts” will automatically reassign
all safety responsibilities to the hospital. Only qualified personnel, biomed engineer or
equivalent, shall attempt repairs. Do not attempt any service with a patient in the bed.
9.1 Annual checks / maintenance
Check the following at least once a year:
•That cables and connectors free of defects.
•Check the Side Walls for cracks or damages.
•Check the Corners –no cracks in the plastic, screws tightened? Note, special plastic
screws.
•Check the Side Wall Guides if needed for the model at hand
•Check that the Side Walls lower and raise without problems and is easily lifted away.
•Check that the Foot Pedal for Pillar height control responds without delay, that the
raising and lowering function is working correctly and that it makes no unusual noise
when operated and that its cable is undamaged.
•Check that the tilt function works correctly. It should not make any noises.
•Check that the wire to the Gas Spring and tilt handle is undamaged.
•Check that the wheels are tightened to the stand and not damaged in any way, that
they roll without noise, and that the wheel locking function works properly.
•Check that all screws are properly tightened, especially the screws connecting the
height adjusting Pillar to the wheel base, the screws on top of the Pillar, the screws
holding the Bed Platform and Gas Spring.
•Check that Foot Pedal Label is not worn
•Check that the Bed Tray Labels are not worn.
Service
NOTE
Always unplug the power before performing repairs on the BabyBed.
Always perform a complete function test after repairs.
9.2 Adjusting the release of the tilt function / Gas Spring
Hydraulic Tilt release
If the bed has mechanical tilt release then refer to version
1 of this manual for service information.
Available on www.kanmed.se.
It is a simple operation to upgrade
to hydraulic tilt release–contact your distributor.
By pressing the button, you activate the release pin
located at the top of the gas spring inside the release head.
If the tilt is not moving at all or not firmly blocked by the gas
spring then you must adjust the function.
Release the gas piston from the bracket on the pillar
and release the locking nut at the release head (see picture)
Tilt is not moving: Turn the gas spring one turn in –
insert gas piston in the bracket and push in the bolt.
Test that it now moves freely and locks hard when
the button is released. If not add another turn.
The tilt is not locking: Turn the gas spring one turn out
and test.
Make a final check. Then tighten the bolts and the locking nut.

User Manual BB100-400 /070
Page 16 of 20
9.2.1 Adjusting the tilt limit.
The tilt limit can be limited by moving down the Gas Spring holder bracket on the Pillar. Make
sure to tighten it properly.
A kit is available to modify the tilt range to be able to tilt the bed ± 17 degrees
9.3 Changing or adjusting a Side Rail or Corner
See spare part list for part numbers. The Side Rail Corners are screwed from the underside
with two (2) special plastic screws at each Corner. The holes in the Bed are oval allowing a
tight adjustment of the Side Rails and Side Walls. It is advisable to remove the screws at 2
Corners to fit new Corners or Side Rails.
Do not use any other screws than Kanmed part no 700-0880 because they will not have
enough holding power. A screw can be screwed in and out about 6 times in the same plastic
hole.
9.4 Adding, removing or adjusting Storage Boxes or Shelves
Loosen the screws to the Storage Box Holder. (If necessary remove the black decorative
plastic stripe.) Move it to the new position and tighten the screws properly.
To remove or add an extra Box Holder: Raise the Pillar more that 5 cm. Press down the black
plastic frame that goes around the bottom of the Pillar. Now you can remove the bolt pins or
add bolt pins.
When adding Storage Boxes ensure that the Front Side Wall can be lowered sufficiently.
Cut the decorative black plastic stripes to fit. New stripes can be ordered at Kanmed.
9.5 Changing the Foot Pedal / electrical connections
Lay the Bed carefully on its side. Unplug the pedal from the Pillar. Unscrew the Foot Pedal
and replace it. Raise the Bed and check functions and that up / down arrows are indicating
correct.
9.5 1 Adding an extra foot pedal or front foot pedal.
Please contact kanmed for necessary instructions and parts.
9.6 Changing the Pillar
Raise the Pillar electrically about 5 cm if possible.
Unplug power.
Empty the Bed and remove side walls and the Baby Tray.
Remove Storage Boxes. Press the plastic frame at the bottom of the Pillar down. Peel away
the plastic strips so that you can remove the metal pins that hold accessories to the Pillar. If
Pillar could not move then you will have to order new metal pins.
Remove Gas Spring from the Pillar as well as its bracket.
Loosen the bolts that hold the Bed Platform. These two bolts are lock tightened and loosening
them requires quality tools and strength. Observe the earth cable
Loosen the bolts on top of the Pillar to remove the Bed bracket. Observe the earth cable on
the underside.
Unplug the height control.
Turn the Pillar/Bed upside down.
Unscrew the earth potential stud connections. (Older models only)
Remove the earth protection cable under the Bed. (Older models only)
Loosen screws under the wheel base and remove the foot pedal.
Unscrew the Pillar. Observe the earth cable
To assemble –perform instructions in reverse order.
Make a full function check and an electrical safety test to ensure that the Bed Platform is
properly earthed.

User Manual BB100-400 /070
Page 17 of 20
10 Spare parts/Accessories
NOTE: Many other accessories and spare parts are available. Please check with your local
supplier or on the Kanmed webpage www.kanmed.se Please contact your supplier for a
price list.
Part no
Description
Remark
BB106
Rear mount kit standard bed
Complete kit incl holder for BW3
BB107
Rear mount kit twin beds
Complete kit incl holder for BW3
BB108
Conversion kit
Converts BB100/300 to BB101/301
BB109
Conversion kit
Converts BB400/400F to BB401/401F
800-0077
Side Wall left/right BB100/BB300
29 cm height.
800-0076
Side Wall front BB100/BB300
29 cm height.
800-0010
Side Wall rear BB100/BB300
29 cm height, Grooves for tubing
800-0101
Side Wall left/right BB101/BB301
25 cm height.
800-0100
Side Wall front BB101/BB301
25 cm height.
800-0105
Side Wall rear BB101/BB301
25 cm height, Grooves for tubing
800-0079
Side Wall left/right BB400/400F
29 cm height.
800-0078
Side Wall front BB400/400F
29 cm height.
800-0042
Side Wall rear BB400/400F
29 cm height, Grooves for tubing
800-0107
Side Wall left/right BB401/401F
25 cm height.
800-0108
Side Wall front BB401/401F
25 cm height.
800-0109
Side Wall rear BB401/401F
25 cm height, Grooves for tubing
800-0074
Side Wall Guide Left side
Set of 4
800-0073
Side Wall Guide Right side
Set of 4
800-0072
Red side wall stopper
Set of 4
800-0023
Side Rail front/rear BB100/BB300
Length 40cm
800-0022
Side Rail sides BB100/BB300
Length 62 cm
800-0044
Side Rail front/rear BB400/BB400F
Length 57,7 cm
800-0045
Side Rail sides BB400/BB400F
Length 72,3 cm
800-0011
Corner handle for Side Rail beds
With side wall guides
800-0110
Corner handle BB100-400 (May 2017)
With side rails. No guides
800-0098
Corner BB101-401
Without side rails
700-0880
Screws for Corner Bracket
Use only these special screws
BB155
Mattress BB100/BB300 68 cm x 46 cm
Plastic covered foam
BB450
Mattress BB400/BB400F 76 cm x 64 cm
Plastic covered foam
800-0004
Baby Tray BB100/BB300
800-0041
Baby Tray BB400/BB400F
800-0002
Storage Box
800-0013
Storage Box holder
Includes set of bolt pin and screws
800-0001
Storage Box lid
BB171
Bottle holder
Not yet available
800-0065
Bracket for DIN rails (fits the Side Rails)
Not yet available
800-0020
Tent Pole for all beds
Two height positions
800-0057
Bracket/handle for BW3
800-0502
Special power cable for BabyWarmer
Length 80-100cm
800-0003
Wheel top Cover
BB160
Tent for BB100/101 / BB300/301
BB460
Tent for BB400/401 /400F/401F
800-0501
Foot Pedal with cable
800-0029
Bolt pin for accessories
(that are to be fixed on the Pillar)
Dia 7 mm, L 30 mm Screw M5.
Including the screws. Set of 2
800-0503
Label Bed tray
800-0501
Label Foot pedal
BB103
Trendelenburg kit ±17 degrees
New gas spring and top plate

User Manual BB100-400 /070
Page 18 of 20
11 Warranty Statement
Kanmed warrants the purchaser that the Kanmed BabyBed is free from defects in materials
and workmanship for a period of 5 years from the date of delivery. The warranty does not
apply to plastic parts, mattresses, cables and tents.
The sole obligation of Kanmed with respect to any such defect is limited to the repair with
new or re-manufactured parts or, at the discretion of Kanmed, replacement of the equipment
or refunding of the purchase price.
This warranty shall not apply if the product has been modified, adjusted or repaired other
than by Kanmed or by organizations authorized by Kanmed or modified, adjusted or repaired
not in accordance with written instructions provided by Kanmed. Neither shall the warranty
apply if the equipment has been subject to misuse, negligence or accident.
These warranties are made on the condition that prompt notification of a defect is given to
Kanmed or its authorized dealers within the warranty period.
Kanmed shall have the sole right to determine whether a defect exists.
Kanmed shall not in any case be liable for special or consequential damages arising from the
breach of warranty, breach of contract, negligence or any other legal theory.
12 EMC Guide Lines
The BabyBed fulfils the applicable requirements of IEC 60601-1-2:2007 Electromagnetic
compatibility (EMC).
The BabyBed shall be used in a hospital environment, but not close to HF surgical equipment
or MR cameras.
WARNING: Use of accessories and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
WARNING: Portable RF communications equipment (e.g. mobile phones, radio transmitters,
antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the BabyBed, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
NOTE: The EMISSIONS characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which
CISPR 11 class B is normally required) this equipment might not offer adequate protection to
radio-frequency communication services. The user might need to take mitigation measures,
such as relocating or re-orienting the equipment.
For further information about EMC compliance with the specified emissions and immunity
standard, test levels, etc. please contact Kanmed.
13 Disposal
When the Kanmed BabyBed have reached end of life, it should be recycled in
accordance with the EU 2002/96/EC (WEEE) directive if applicable.

User Manual BB100-400 /070
Page 19 of 20

User Manual BB100-400 /070
Page 20 of 20
Kanmed AB
Gårdsfogdevägen 18B
SE-168 67 BROMMA –Stockholm
Sweden
Telephone +46 8 56 48 06 30
Telefax +46 8 56 48 06 39
E-Mail: in[email protected]
Web page: www.kanmed.se
Distributed by:
This manual suits for next models
2
Table of contents
Other Kanmed Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Thuli Tables
Thuli Tables 300 Stationary owner's manual

Boston Scientific
Boston Scientific EPT-1000 XP Operator's manual

DeVillbiss Air Power Company
DeVillbiss Air Power Company 525DS Service manual

Mastercare
Mastercare Master Turner Medium user guide

Pro-tec
Pro-tec optimax 2010 Operation manual

VitalBase
VitalBase Royal user manual