
SECTION V - MAINTENANCE
Except for occasional tube replacement very little maintenance is
required for the model hlh. Components are operated well below
rating, and high quality components have been used throughout.
1. FACTORYADJlJSTMENTS
Only one factory adjustment is made. This is Rl32, the meter
calibration potentiometer. This adjustment is only necessary
if the panel meter is changed. To calibrate set the instrument
on the 1 or 10 milliampere range anti set the meter to the cor-
rect reading with a calibration current sufficient to deflect
the panel meter to at least 75% of full scale.
2. TROUBLESHOOTING
Refer to DR12758-C and the voltage-resistance diagram at therear
of the manual.
Follow this general procedure in trouble shooting if the fault
is not apparent from the tabulated list below:
1. Check the B+ and J?- potentials: The E+ poten,tial can be
measured between the rectifier string and R201 and should
be about 320 volts. The ripple should not exceed 3 volts
rms. The B- potential may be measured between the negative
rectifier string (RFh,5) and R2O5 and should be about 2irO
volts. The ripple again should not exceed 3 volts RMS. If
either potential is no't correct the trouble is probably due
to a defective rectifier or filter component although in
some cases the transformer may be defective. Refer to the
schematic diagram to locate the defective component.
2. SPECIFIC FAULTS
a. Instrument will not balance: If no defect has been
foundT%x%zpection suggested in l., check the volt-
age at pin 1 or 5 of Vlr. This should be 150 volts. If the
voltage is not 15Ov, check R2O3, 202, and 201 for continuity
and check Vh. If the voltage is correct, check the voltage
at pin 3 of Vl (The red dot is next to lead 1) the reading
should be 3.2 volts. If the voltage at this point has risen
to 150 volts, the filament of Vl is open and this tube should
be replaced. If this voltage checks correctly, the following
procedure should be started: Short circuit the input to
ground with the range switch on 1 microampere full scale.
This removes the negative feedback and allows the operating
points of the individual tubes to be checked. It will usually
be found that the amplifier will become very sensitive and
difficult to balance, however it is only necessary that the
voltage swing through the correct operating value to confirm
that the circuit is operating correctly. Then with the input
V-l