KeLi TCS-JJ Series User manual

XK3101D WEIGHING INDICATOR
TCS-JJ SERIES PLASTIC COVERED STEEL
PLATFORM SCALE
USER’S MANUAL

1
PREFACE
THANK YOU FOR USING OUR XK3101D WEIGHING INDICATOR AND TCS-JJ SERIES
PLATFORM SCALE.
XK3101D IS WATERPROOF WEIGHING INDICATOR WITH LCD BACKLIGHT DISPLAY,
WATERPROOF GRADE IS IP68. WE USED THE LATEST HIGH-TECH, IT HAS THE
CHARACTERISTIC OF HIGH PRECISION, DISPLAY CELERITY, DURABLE, BEST AFTER
SALE SERVICES.
TCS-JJ SERIES PLATFORM SCALES ADOPT OUR OWN PLASTIC COVERED STEEL
TECHNIQUE, IT HAS THE CHARACTERISTIC OF WATERPROOF, CORROSION-PROOF,
LOW PRICE. IT USES OUR ORIGINAL LOADCELL PARTS (WE HAD GET NATIONAL
PATENT, PATENT NO.: 200820019951.3) AND XK3101D WATERPROOF INDICATOR,
WATERPROOF GRADE IS UP TO IP68, USEFULL TIME IS GREATLY RAISED. IT IS WIDELY
USED FOR NOT ONLY RETAIL STORES, RATION PACKING BUT ALSO FOR HIGH
HUMIDITY SURROUNDING OF FOOD AND SEAFOOD PACKING.
FOR TCS-JJ PLATFORM SCALE, WE CAN ALSO USE OUR XK3101 RED LED
WATERPROOF INDICATOR.
PLEASE READ THE INSTRUCTION MANUAL BEFORE USAGE.
1MAIN TECHNICAL INDEX
1.1 BASIC DATA
MAX.CAPACITY(kg)
15
30
60
100
150
200
300
500
600
MIN.CAPACITY
20e
TEST DIVISION e(g)
5
10
20
50
50
100
100
200
200
DISPLAY DIVISION d1(g)
2
5
10
20
20
50
50
100
100
DISPLAY DIVISION d2(g)
1
2
5
10
10
20
20
50
50
MAX. TARE
100% MAX.CAPACITY
ACCURACY
Ⅲ
1.2 PLATE SIZE AND SCALE RANGE
PLATE SIZE(cm)
SCALE RANGE(kg)
30×40
15,30,60,100
40×50
60,100,150,200
48×62
150,200,300,500,600
1.3 WORKING TEMPERATURE:-5~+35℃
1.4 STORAGE TEMPERATURE:-20~+50℃
1.5 POWER:3PCS 1# DRY BATTERIES(WITHOUT BATTERIES FROM FACTORY)OR AC
5.5V OUTPUT VOLTAGE TRANSFORMER.

2
2KEYBOARD AND CHARACTER PROMPT
2.1 KEYBOARD
ON/OFF/ZERO
:TURN SCALE ON,OFF; ZERO
TARE/B-LIGHT
:TARE, TURN BACKLIGHT ON, OFF
2.2 CHARACTER PROMPT
U3.X :SOFTWARE VERSION NUMBER.
dc x.xx: MEAN THE VOLTAGE OF BATTERY IS x.xx V .
-----:SHOW THIS AS THE WEIGHT OVER 100%FS+9E, THIS MEANS OVER LOAD.
bAtLo:THE VOLTAGE IS LOWER. YOU CAN USE A SHORT WHILE AND CHANGE BATTERY IMMEDIATELY.
THEN IT WILL SHOW oFF AND AUTO TURN OFF.
CAEnd:CALIBRATION FINISHED.
Err-0:OVERLOAD ZERO OF LOADCELL, STH. IS WRONG OF LOADCELL.
Err-1: ADDITION WEIGHT IS TOO SMALL WHEN ADJUSTING, PLEASE PUT WEIGHT AS
PROMPT.
ERR-2: MANUAL ZERO IS OVER 4% ALLOWABLE RANGE.
3 OPERATION GUIDE
3.1 PANEL
3.2 PREPARATION
3.2.1 INSERT THE POLE INTO PLATFORM BASE, TIGHTEN UP THE 3 SCREWS.
3.2.2 ADJUST FEET, WATCH LEVEL, LET THE PLATFORM IN LEVEL.
3.2.3 ADJUST THE 4 LIMIT SCREWS OF PLATFORM AND TIGHTEN THEM UP.
3.2.4 LOAD 3PCS 1# BATTERIES TO BACK OF INDICATOR OR USE AC POWER
WITH THE TRANSFORMER.
3.3 TURN ON AND TURN OFF

3
3.3.1 TURN ON:
PRESS <ON/OFF/ZERO> KEY,CURRENT CONNECTED,IT DISPLAYS SOFTWARE
VERSION (VERSION OF THIS SCALE IS U3.X )、SELF-INSPECTION 、
dc-x.xx(VOLTAGE OF BATTERY)、MAX. CAPACITY-L OR H(L MEANS DIVISION e,
H MEANS DIVISION d), ZERO,{ZERO} LAMP LIGHTING,NOW SCALE IS IN
WORKING.
3.3.2 TURN OFF:
3.3.2.1 MANUAL TURN OFF:LONG TIME PRESS<ON/OFF/ZERO>,DISPLAY oFF ,TURN
OFF.
3.3.2.2 LOW VOLTAGE OFF:WHEN VOLTAGE IS LOW, AT ZERO IT DISPLAYS bAtLo,
COULD CONTINUE USING. BUT WHEN THE VOLTAGE IS MORE LOWER, IT
SHOWS oFF, AUTO TURN OFF.
3.3.3 AUTO TURN OFF:
3.3.3.1 SAME TIME PRESS <ON/OFF/ZERO> AND <TARE/B-LIGHT>,SHOWS A-on,
SCALE IS IN AUTO OFF STATE. IT WILL BE AUTO OFF AFTER 30 MINUTES
WHEN SCALE IS AT ZERO AND NO OPERATION.
3.3.2.2 CANCEL AUTO TURN OFF :SAME TIME PRESS <ON/OFF/ZERO> AND
<TARE/B-LIGHT>,SHOWS A-oFF, CANCEL AUTO TURN OFF STATE.
ABOVE OPERATION WILL KEEP 1 SECOND AND RETURN TO WEIGHING
STATE.
3.4 ZERO
WHEN WITHOUT TARING THE WEIGHT DISPLAY IS ≤4% MAX. CAPA.,,PRESS <ON/OFF/ZERO>,
THE DISPLAY SHOWS “ 0 ”,{ZERO} LAMP LIGHTING.
3.5 WEIGHING
PUT WEIGHT TO PAN,IT SHOWS THE WEIGHT,{STABLE} LAMP LIGHTING.
3.6 TARE WEIGHING
FIRST PUT A CONTAINER ON PLATE, PRESS <TARE/B-LIGHT>,IT SHOWS 0,{ZERO}
LAMP OFF,{TARE} LAMP LIGHTING, PUT THE WEIGHT ON CONTAINER IT WLL SHOW
NET WEIGHT. REMOVE THE WEIGHT AND CONTAINER TOGETHER, IT DISPLAYS
NEGATIVE TARE, {ZERO} LAMP LIGHTING. NOW PRESS <TARE/B-LIGHT>,{TARE}
LAMP OFFRETURN TO NORMAL STATE.
3.7 BACKLIGHT
IN WEIGHING STATE, PRESS <TARE/B-LIGHT> 2 SECONDS TO TURN ON/OFF BACK
LIGHT,L-on IS ON,L-Off IS OFF. WHEN BACKLIGHT IS ON, IT WILL BE AUTO OFF AFTER
15 SECONDS AT ZERO. WHEN PUT A WEIGHT OR OPERATE KEYPAD, IT WILL BE AUTO
ON AGAIN.
3.8 BATTERY INDICATE
BATTERY INDICATE:WHEN VOLTAGE IS LOW, {BATLO} LAMP FLICKERING,PLEASE

4
CHANGE BATTERY IN TIME,WHEN THE VOLTAGE IS MORE LOWER, IT SHOWS bAtLo , oFF,
AUTO TURN OFF.
3.9 CHANGE DIVISION:
BEFORE TURN ON,FRIST KEEP PRESS <TARE/B-LIGHT>,THEN PRESS <ON/OFF/ZERO>
ONE TIME,NOW RELEASE <TARE/B-LIGHT>,AFTER TURN ON, COULD CANGE DIVISION,
THERE HAVE THREE DIVISIONS e , d1 AND d2 OF EACH CAPACITY,WHEN DISPLAY MAX.
CAPACITY –H, DISPLAY DIVISION IS d2, MAX. CAPACITY –b, DISPLAY DIVISION IS d1, MAX.
CAPACITY –L, TEST DIVISION IS e.
4 CALIBRATION
WITH THE SCALE OFF, INSERT OUR SPECIAL ADJUSTMENT TOOL INTO THE ADJUST HOLE OF
INDICATOR, PRESS <ON/OFF/ZERO> TO TURN,AFTER SELF INSPECTION, IT DISPLAYS THE
LAST MAX. CAPACITY,NOW COULD PRESS <TARE/B-LIGHT> TO ENTER NEXT STEP,OR
PRESS <ON/OFF/ZERO> TO CHOOSE MAX.CAPACITY AGAIN(UNIT:kg):15.000、30.000、60.00、
100.00、150.00、200.00、300.00、500.0、600.0、1000.0、1500.0,PRESS <TARE/B-LIGHT> TO
CHOOSE THE NEEDED CAPACITY,NOW ENTER NEXT STEP TO CHOOSE WEIGHTS OF
CALIBRATION. FOR MAX.CAPACITY 15、30、60、150、300、600、1500kg,LONG TIME PRESS
<TARE/B-LIGHT> TO ENTER 1/3 MAX.CAPACITY CALIRATION, SHORT TIME PRESS
<TARE/B-LIGHT> TO ENTER 1/2 MAX.CAPACITY CALIRATION,NOW DISPLAY FLICKING THE
NEEDED WEIGHTS VALUE THREE TIME AND RETURN INTERNAL CODE 00000, PUT THE
INDICATED WEIGHTS, IT DISPLAYS INTERNAL OF THIS WEIGHTS, AFTER STABLE, PRESS
<TARE/B-LIGHT>,IT SHOWS CAEnd,CALIBRATION FINISHED. FOR MAX.CAPACITY 100、200、
1000kg,IS 1/2 MAX.CAPACITY CALIRATION,FOR MAX.CAPACITY 500k,LONG TIME PRESS
<TARE/B-LIGHT> IS 40% MAX.CAPACITY (200kg )CALIRATION ,SHORT TIME PRESS
<TARE/B-LIGHT> IS 1/2 MAX.CAPACITY CALIRATION,PUT THE INDICATED WEIGHTS. FINAL GET
THE SPECIAL ADJUSTMENT TOOL OUT.
NOTE:
①WHEN ADJUSTING, PLEASE PUT ENOUGH WEIGHT TO ENSURE RELIABILITY.
IFTHE WEIGHT IS SO LIGHT, IT WILL SHOW Err-1 ,CALIBRATION IS
DISALLOWED.
②NO WIND, NO VIBRATING, ROOM TEMPERATURE.
③SCALE MUST BE CALIBRATED WITH PERSON WITH WHOM HAVE ELIGIBLE TO
DO THIS WORK.
5OTHER ADVICE
5.1 PLEASE TAKE BATTERIES OUT FOR LONG TIME STORAGE, TO PREVENT CORRISION FROM
LEAKAGE OF BATTERY.
5.2 PLEASE USE OUR OWN TRANSFORMER TO CONNECT AC POWER, TO PREVENT
DAMAGING AND DANGEROUS.
This manual suits for next models
1
Table of contents
Other KeLi Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Vernier
Vernier Go Direct Mini GC quick start guide

SONOPAN
SONOPAN L-50 instruction manual

PCE Health and Fitness
PCE Health and Fitness PCE-S 41 user manual

Graco
Graco ProControl 1KE Plus Instructions-parts list

Hioki
Hioki PQ3198-94 instruction manual

Emerson
Emerson Rosemount Oxymitter DR Reference manual