Kemot MIE0214 User manual

User Manual
Breathalyser
MIE0214
Polski 11-22
Română 23-34 Deutsch 35-46 English 1- 10


WWW.KEMOT.COM 1
MIE0214
Table of Contents
Introduction 2
Preparation for correct measurement 2
Description of parts 2
Power supply 3
Operation 3
Accessories 7
Error and troubles 8
Important notes 9
Specifications 9
Calibration and record 10
Precautions 10

Introduction
The device is intended to measure alcohol in the human breath. Measurements
obtained by this device are used in the diagnosis of alcohol intoxication.
The accuracy of this device has been established at a Blood Alcohol
Concentration of 0.008 and 0.032 % BAC. Accuracy at a Blood Alcohol
Concentration greater than 0.032% BAC has not been established.
Preparation for Correct Measurement
źBefore testing, wait at least 15 minutes after your last drinking.
źOnly use the device in a temperature10°C -40°C or 50°F -104°F.
Description of parts
WWW.KEMOT.COM
2
MIE0214

Power Supply
źSlide back the battery cover on the back of the device.
ź Insert 2 pieces AAA battery, making sure that the “+/-“symbols are correctly.
źIf the power is too low, the unit may display a battery symbol
and the device may not use correctly. You need to change a new battery.
Operation
(Test mode)
ź Insert one of the mouthpieces into the 'blow in' hole.
WWW.KEMOT.COM 3
MIE0214

WWW.KEMOT.COM
4
MIE0214
źPress and hold the TEST button until power on. Release it after power on.
źThe WAIT displays on the screen and countdown 10, 9, and 8… to zero.
źThe BLOW displays on the screen, and displays “C” with flashing. Take a deep
breath, blowing steady into the mouthpiece until beeping stops (about 6 seconds).

WWW.KEMOT.COM 5
MIE0214
źThe device displays “WAIT” and flashing the “0” on the screen.
źYour test result will be displayed on the screen after few seconds.
źYour test result will be changed the color if above alarm point.

WWW.KEMOT.COM
6
MIE0214
źIt will show error when incorrect blowing. Refer to the chapter of the “Errors
and Troubles” in detail. The device will be turned off after indicated error.

WWW.KEMOT.COM 7
MIE0214
Accessories
źMouthpieces
ź Manual

WWW.KEMOT.COM
8
MIE0214
Errors or troubles
Error
message
or troubles
Description
Remarks
No display
-Press and hold the TEST
button until power on.
-Check the battery
Auto Off
-Low battery
-Over the period of test stage
Err
-Blowing unsteady or interrupt
-Blowing less than 6 seconds
-Blowing flow too weak
-No blow during BLOW mode.
-Change temperature suddenly
-Low battery indication
-Need to change new battery
-Use alkaline battery
Alarm
sound
-During the blowing mode
-Error message indication
-Above alcohol set point
HI
-The blood alcohol content
above the display range.
-Display range:
0.20%BAC(2.00‰, 1.000mg/L)

WWW.KEMOT.COM 9
MIE0214
Important notes
źBefore testing, wait at least 15 minutes after your last drinking.
źThe blowing time must be enough until beeping stops.
źIt may show ERR (error) if none, interrupt or shorter blowing.
źIf the device changes temperature suddenly, wait some minutes to use.
źOnly use the device in a temperature 10°C -40°C or 50°F -104°F.
źRecalibration interval: 12 months or 200 measurements.
źNever store the device in a corrosive gas room.
źNever store the device under sun exposure area.
źAvoid dropping or subjecting the device to a strong shock.
Specifications
źElectrochemical sensor (Fuel Cell Sensor)
źBlowing time about 6 seconds
źBlowing flow about 10L/min
źBlowing flow interrupt detection
ź Test counter to display the operated number
ź4 digits display with status and EL backlight
źDisplay range: 0.000-0.200%BAC(2.00‰ or 1.00mg/L)
źShow “HI” if above the display range
źAccuracy: +/-0.005% at 0.05%
5 mouthpieces included
źWarm up time: 10-15 seconds
źTesting Time: 10 seconds
źRepeat test: no less than 3 minutes.
źAudible sound and change LCD color if above alarm point
źPower supply: 2 pcs AAA 1.5V alkaline battery
źBattery for standard testing is about 300 times
źLow battery indication and auto power off function
źOperation: 10°C ~ 40°C or 50°F ~ 104°F
źStore: 0ºC ~ 50ºC or 32°F~122°F
źSize: 115x60x23mm

WWW.KEMOT.COM
10
MIE0214
Calibration and record
źYou should have your alcohol tester calibrated every 12 months or 200
measurements. Failure to do so may cause your readings to be inaccurate.
Precautions
źYour blood alcohol content (BAC) may change some minutes later.
źYou should have your alcohol tester calibrated every 12 months or 200
measurements. Failure to do so may cause your readings to be inaccurate.
źDo not use this device to determine if you can drive.
źYou cannot use the test results of device in court of law.
źYou cannot use the test result of device against police's unit.
Calibration Record
Date First Used
Serial Number
#
Calibration Date
Remarks
Recalibration by

Instrukcja obsługi
Alkomat
MIE0214
Polski 11-22
Română 23-34 Deutsch 35-46 English 1- 10

WWW.KEMOT.COM
12
MIE0214

WWW.KEMOT.COM 13
MIE0214
Spis treści
Wprowadzenie 14
Przygotowanie do prawidłowego pomiaru 14
Opis elementów urządzenia 14
Zasilanie 15
Funkcje 15
Akcesoria 19
Błędy i problemy 20
Uwagi 21
Specyfikacja 21
Kalibracja urządzenia 22
Środki ostrożności 22

Wprowadzenie
Urządzenie przeznaczone jest do pomiaru zawartości alkoholu w oddechu
człowieka. Dokładność urządzenia została ustalona na stężenie alkoholu we krwi
pomiędzy 0.008 i 0.032%. Wskazania przy stężeniu alkoholu we krwi
przekraczającym 0,032% nie są dokładne.
Przygotowanie do prawidłowego pomiaru
źPrzed pomiarem należy odczekać przynajmniej 15 minut po spożyciu
alkoholu.
źUżywaj urządzenia w temperaturze 10°C ~ 40°C (50°F ~ 104°F).
Opis elementów urządzenia
WWW.KEMOT.COM
14
MIE0214

Zasilanie
źZdejmij pokrywę baterii z tyłu urządzenia.
ź Włóż dwie baterie AAA zgodnie z zaznaczoną polaryzacją “+/-“.
źJeżeli stan naładowania baterii jest niski, na wyświetlaczu pojawi się symbol
Aby zapewnić prawidłowe działanie urządzenia należy wymienić baterię
na nową.
Funkcje
(Wykonywanie pomiaru)
ź Włóż ustnik do urządzenia jak pokazano na rysunku poniżej.
WWW.KEMOT.COM 15
MIE0214
Ustnik

WWW.KEMOT.COM
16
MIE0214
źNaciśnij i przytrzymaj przycisk TEST do momentu uruchomienia urządzenia.
źNa ekranie wyświetlona zostanie informacja „WAIT” i rozpocznie się
odliczanie od 10 do 0.
źPo zakończeniu odliczania na ekranie pojawi się informacja „BLOW” oraz
zacznie migać symbol „C”. Weź głęboki wdech i dmuchaj jednostajnie w ustnik aż
urządzenie przestanie wydawać sygnał dźwiękowy (około 6 sekund).

WWW.KEMOT.COM 17
MIE0214
źNa wyświetlaczu pojawi się informacja “WAIT” i zacznie migać symbol “0”.
źWynik zostanie wyświetlony na ekranie po kilku sekundach.
źWynik testu zmieni kolor jeżeli przekroczy alarmową wartość.

WWW.KEMOT.COM
18
MIE0214
źW przypadku nieprawidłowo przeprowadzonego pomiaru na ekranie
wyświetli się symbol „Err”. Przeczytaj rozdział „Błędy i problemy” aby dowiedzieć
się więcej. Po wykryciu błędu urządzenie zostanie wyłączone.
Table of contents
Languages:
Other Kemot Test Equipment manuals
Popular Test Equipment manuals by other brands

Agilent Technologies
Agilent Technologies 86100A Release notes

Extech Instruments
Extech Instruments 412355A user guide

Softing
Softing PROFIBUS Tester 4 BC-600-PB manual

Hioki
Hioki BT3554 instruction manual

Extech Instruments
Extech Instruments BR50 user guide

Keysight
Keysight N1000A DCA-X Series quick start guide