KERN KB Series User manual

572/573/KB/DS/FKB-BA-defsi-0651
24
GB KERN 572/573/KB/DS/FKB
Version 5.1 05/2006
Operating Instructions
Electronic Precision and Platform Balances
Table of contents:
1MODE - MENUE __________________________________________________________ 26
2Technical data_____________________________________________________________ 27
2.1 KERN 572 __________________________________________________________________ 27
2.2 KERN 573 __________________________________________________________________ 28
2.3 KERN KB __________________________________________________________________ 29
2.4 KERN FKB _________________________________________________________________ 29
2.5 KERN DS___________________________________________________________________ 30
3Fundamental information (general) ___________________________________________ 31
3.1 Intended use_________________________________________________________________ 31
3.2 Inappropriate use ____________________________________________________________ 31
3.3 Guarantee __________________________________________________________________ 31
3.4 Monitoring the test substances__________________________________________________ 31
4Fundamental safety information ______________________________________________ 32
4.1 Observe the information in the operating instructions ______________________________ 32
4.2 Staff training ________________________________________________________________ 32
5Transport and storage_______________________________________________________ 32
5.1 Acceptance check ____________________________________________________________ 32
5.2 Packaging___________________________________________________________________ 32
6Unpacking, installation and commissioning _____________________________________ 32
6.1 Place of installation, place of use ________________________________________________ 32
6.2 Unpacking __________________________________________________________________ 33
6.2.1 Installation _______________________________________________________________________ 33
6.3 Mains supply ________________________________________________________________ 33
6.4 Connecting peripheral equipment_______________________________________________ 33
6.5 Initial start-up _______________________________________________________________ 33
6.6 Adjustment _________________________________________________________________ 33
6.7 Adjusting ___________________________________________________________________ 33
6.7.1 Adjusting for verification (KERN 573) _________________________________________________ 34
6.8 Verification _________________________________________________________________ 34

572/573/KB/DS/FKB-BA-defsi-0651 25
7Operation_________________________________________________________________ 35
7.1 Display control panel _________________________________________________________ 35
7.2 Operation___________________________________________________________________ 36
7.2.1 Weighing with Tare ________________________________________________________________ 36
7.2.2 Count – Selection Preference Piece ____________________________________________________ 36
7.2.3 Percent % ________________________________________________________________________ 36
7.2.4 Formula Weighing _________________________________________________________________ 36
7.2.5 Tolerance Weighing ________________________________________________________________ 36
7.2.6 Adjust ___________________________________________________________________________ 38
7.2.7 Preselect Tare _____________________________________________________________________ 38
7.2.8 Auto Tare ________________________________________________________________________ 38
7.2.9 Speed ___________________________________________________________________________ 38
7.2.10 Auto Off_______________________________________________________________________ 38
7.2.11 Variable Factor__________________________________________________________________ 38
7.2.12 Preselect Unit___________________________________________________________________ 38
7.3 Interface RS 232 C ___________________________________________________________ 40
7.4 RS 232 C Data output via interface RS 232 C _____________________________________ 40
7.4.1 There are 4 methods for the data output via RS 232 C______________________________________ 40
7.4.2 Description of the data transfer________________________________________________________ 41
7.4.3 Numerator________________________________________________________________________ 41
7.5 Printer _____________________________________________________________________ 41
7.6 Underfloor weighing__________________________________________________________ 41
8Maintenance, upkeep, disposal________________________________________________ 42
8.1 Cleaning ____________________________________________________________________ 42
8.2 Maintenance, upkeep _________________________________________________________ 42
8.3 Disposal ____________________________________________________________________ 42
9Troubleshooting ___________________________________________________________ 42
10 Declaration of conformity__________________________________________________ 43

572/573/KB/DS/FKB-BA-defsi-0651
26
1 MODE - MENUE
MODE
CAL?
Mode
PLACE CAL
WEIGHT
Y
g
N/Y
Mode Esc
Preselect Units ?
Mode Y
b
N/Y
Mode Esc
AUTO OFF?
N/Y
Mode
weighing
REF=1
N/Y
Mode Esc
COUNT?
N/Y
Mode Y
REF=5
N/Y
Mode Esc
REF=10
N/Y
Mode Esc
REF=20
N/Y
Mode Esc
%
?
N/Y
Mode
g, kg, oz, ozt,
*, dwt, GN,
lb, ct, tLT,
tLH, tLM,
tLS, tL, t,
mo, m, b,
o,
Preselect Tare?
Mode Y
1 l 0.000g
Mode
Y N - -
Variable Factor ?
Mode Y
1 l 0.000g
Mode
Y N - -
2400Bd
N/Y
Mode Esc
PRINTER?
Mode Y
4800Bd
N/Y
Mode Esc
9600Bd
N/Y
Mode Esc
19200Bd
N/Y
Mode Esc
Numerator
N/Y
Mode Esc
AUTOPRINT
N/Y
Mode Esc
AUTOPRT PC
N/Y
Mode Esc
AUTOTARE?
N/Y
Mode
FAST 3
Mode
Manufacture Setting:
Count?: YES
REF10: YES
9600bd: YES
Preselect
Units?: g/kg
Autotare: YES
Fast: 3

572/573/KB/DS/FKB-BA-defsi-0651 27
2 Technical data
2.1 KERN 572
KERN 572-31 572-33 572-35 572-45 572-49 572-57
Readout 0,001 g 0,01 g 0,01 g 0,1 g 0,1 g 0,2 g
Weighing range 161 g 1.210 g 1.510 g 12.100 g 15.100 g 20.100 g
Taring range (subtractive) 161 g 1.210 g 1.510 g 12.100 g 15.100 g 20.100 g
Reproducibility 0,001 g 0,01 g 0,01 g 0,1 g 0,1 g 0,2g
Linearity ±0,003 g ±0,03 g ±0,03 g ±0,3 g ±0,3 g ± 0,6 g
Minimum weight for
counting parts 0,001 g 0,01 g 0,01 g 0,1 g 0,1 g 0,2 g
Test weight (included)
tolerance as per M1 50 g (F1) 500 g 500 g 5.000 g 5.000g 5.000 g
Recommended adjusting
weight F2 100g 1 kg 1 kg 10 kg 10 kg 10 kg
Verifiable No No No No No No
Stabilisation time (typical) 3 sec.
Permissible ambient temperature +10 °C ... + 40 °C
Casing (W x H x D) mm 180 x 310 x 90
Weighing plate mm Ø 106 Ø 150 Ø 150 160 x 200
Net weight kg 4
Interface Yes (RS232)

572/573/KB/DS/FKB-BA-defsi-0651
28
2.2 KERN 573
KERN 573-46 573-52
Class of accuracy II II
Readout (d) 0,1 g 0,2 g
Verification value (e) 1g 2g
Weighing range (max.) 6500 g 13000 g
Minimum load (Min) 5g 10g
Taring range (subtractive) 6500 g 13000 g
Reproducibility 0,1 g 0,2 g
Linearity ±0,2 g ±0,4 g
Minimum weight for counting parts 0,3 g 0,6 g
Test weight (included) tolerance as per M1 5000 g 5000 g
Recommended adjusting weight F2 5 kg 10 kg
Verifiable Yes Yes
Air humidity max. 80 % relative (not condensing)
Stabilization time 2 sec.
Permissible ambient temperature 0° C to + 50° C
Temperature range for verifiable applications +10 °C ... + 40 °C
Casing (W x H x D) mm 180 x 310 x 90
Weighing plate mm 160 x 200 160 x 200
Data interface Yes (RS232)
Weighing units g, kg
Vibratory filter Yes
Net weight (kg) 4

572/573/KB/DS/FKB-BA-defsi-0651 29
2.3 KERN KB
KERN KB100-3 KB1000-2 KB10000-1
Readout 0,001 g 0,01 g 0,1 g
Weighing range 101 g 1010 g 10.100 g
Taring range (subtractive) 101 g 1000 g 10.100 g
Reproducibility 0,001 g 0,01 g 0,1 g
Linearity ±0,003 g ±0,03 g ±0,3 g
Minimum weight for counting parts 0,001 g 0,01 g 0,1 g
Test weight (included) tolerance as per M1 50 g 200 g 1.000 g
Adjusting weight (not included)tolerance as per
F2 100 g F1 1000 g 10.000 g
Verifiable No No No
Stabilisation time (typical) 3 sec.
Permissible ambient temperature + 10° C … + 40° C
Casing (W x H x D) mm 165 x 230 x 85
Weighing plate mm Ø 85 130 x 130 150 x 170
Net weight kg 1,8 2,7
Data interface Yes (RS232)
2.4 KERN FKB
KERN FKB 4K0.5 FKB 8K0.1 FKB 16K0.2 FKB 36K0.5 FKB 65K1
Readout 0,05 g 0,1 g 0,2 g 0,5 g 1 g
Weighing range 4.100 g 8.100 g 16.100 g 36.100 g 65.100 g
Taring range (subtractive) 4.100 g 8.100 g 16.100 g 36.100 g 65.100 g
Reproducibility 0,05 g 0,1 g 0,2 g 0,05 g 1 g
Linearity 0,15 g 0,3 g 0,6 g 1,5 g 3 g
Minimum weight for
counting parts 0,05 g 0,1 g 0,2 g 0,5 g 1 g
Adjusting weight (not
included ) tolerance as
per F2 2 kg 5 kg 10 kg 20 kg 50 kg
Air humidity max. 80% relative (not condensing)
Verifiable No
Stabilisation time (typical) 2 sec.
Permissible ambient
temperature + 10 °C ... + 40 °C
Casing
(W x H x D) mm 350 x 390 x 120
Weighing plate mm 340 x 240
Vibratory filter Yes
Net weight kg 6,5
Data interface Yes (RS232)

572/573/KB/DS/FKB-BA-defsi-0651
30
2.5 KERN DS
KERN DS4
K0.05 S DS8
K0.1S DS8
K0.1 DS16
K0.2 DS36
K0.5 DS36
K0.5L DS65
K1 DS100
K1 DS150
K2
Readout 0,05 g 0,1 g 0,1 g 0,2 g 0,5 g 0,5 g 1g 1g 2g
Weighing range 4.100 g 8.100 g 8.100 g 16.100 g 36.100 g 36.100 g 65.100g 101.000g 151.000g
Taring range
(subtractive) 4.100 g 8.100 g 8.100 g 16.100 g 36.100 g 36.100 g 65.100g 101.000g 151.000g
Reproducibility 0,05 g 0,1 g 0,1 g 0,2 g 0,5 g 0,5 g 1g 1g 2g
Linearity 0,15 g 0,3 g 0,3 g 0,6 g 1,5 g 1,5 g 3 g 3 g 6g
Minimum weight
for counting
parts
0,05 g 0,1 g 0,1 g 0,2 g 0,5 g 0,5 g 1g 1g 2g
Adjusting weight
(not included )
tolerance as per
F2
2 kg 5 kg 5 kg 10 kg 10 kg 20 kg 50 kg 50 kg 50 kg
Verifiable No
Stabilisation
time (typical) 2 sec.
Permissible
ambient
temperature + 10 °C ... + 40 °C
Vibratory filter Yes
Casing
(W x H x D) mm
228 x 228 x 70
315 x 305 x 70
450 x 350 x 115
Weighing plate
mm
228x228
315 x 305
Net weight kg 5,5 8 12,5
Data interface Yes (RS232)

572/573/KB/DS/FKB-BA-defsi-0651 31
3 Fundamental information (general)
It is essential to read through and observe the complete operating instructions before installing and commis-
sioning.
3.1 Intended use
The balance you have acquired serves to determine the weighing value of the material to be weighed. It is
intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully
placed in the centre of the weighing plate. The weighing value can be read off after a stable weighing value
has been obtained.
3.2 Inappropriate use
Do not use the balance for dynamic weighing. In the event that small quantities are removed or added to the
material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the
balance. (Example: Slowly draining fluids from a container on the balance.)
Do not leave a permanent load on the weighing plate. This can damage the measuring equipment.
Be sure to avoid impact shock and overloading the balance in excess of the prescribed maximum load rating
(max.), minus any possible tare weight that is already present. This could cause damage to the balance.
Never operate the balance in hazardous locations. The series design is not explosion-proof.
Structural alterations may not be made to the balance. This can lead to incorrect weighing results, faults
concerning safety regulations as well as to destruction of the balance.
The balance may only be used in compliance with the described guidelines. Varying areas of applica-
tion/planned use must be approved by KERN in writing.
3.3 Guarantee
The guarantee is not valid following
- non-observation of our guidelines in the operating instructions
- use outside the described applications
- alteration to or opening of the device
- mechanical damage and damage caused by media, liquids
- natural wear and tear
- inappropriate erection or electric installation
- overloading of the measuring equipment
3.4 Monitoring the test substances
The metrology features of the balance and any possible available adjusting weight must be checked at regu-
lar intervals within the scope of quality assurance. For this purpose, the answerable user must define a suit-
able interval as well as the nature and scope of this check. Information is available on KERN’s home page
(www.kern-sohn.com) with regard to the monitoring of balance test substances and the test weights required
for this. Test weights and balances can be adjusted quickly and at a reasonable price in KERN’s accredited
DKD calibration laboratory (return to national normal).

572/573/KB/DS/FKB-BA-defsi-0651
32
4 Fundamental safety information
4.1 Observe the information in the operating instructions
Please read the operating instructions carefully before erecting and commissioning, even if you already have
experience with KERN balances.
4.2 Staff training
The device may only be operated and looked after by trained members of staff.
5 Transport and storage
5.1 Acceptance check
Please check the packaging immediately upon delivery and the device during unpacking for any visible signs
of external damage.
In the case of visible damages please obtain a signature from the bearer to serve as confirmation of dam-
age. Do not make alterations to the goods and packaging, do not remove any consignment parts. Report
the damage to the parcel service immediately in writing ( within 24 hours).
5.2 Packaging
Please retain all parts of the original packaging in case it should be necessary to return items at any time.
Only the original packaging should be used for return consignments.
Before despatch, disconnect all attached cables and loose/movable parts, remove weighing plate.
Apply any intended transport security devices. Secure all parts, e.g. glass windshield, weighing plate, power
unit etc., to prevent slipping and damage.
6 Unpacking, installation and commissioning
6.1 Place of installation, place of use
The balance is constructed in such a way that reliable weighing results can be achieved under normal appli-
cation conditions.
By selecting the correct location for your balance, you will be able to work quickly and precisely.
Therefore please observe the following at the place of installation:
- Place the balance on a firm, level surface;
- Avoid extreme heat as well as temperature fluctuation caused by installing next to a radiator or in the
direct sunlight;
- Protect the balance against direct draughts due to open windows and doors;
- Avoid jarring during weighing;
- Protect the balance against high humidity, vapours and dust;
- Do not expose the device to extreme dampness for longer periods of time. Inadmissible bedewing
(condensation of air moisture on the device) can occur if a cold device is taken into a significantly
warmer environment. In this case, please acclimatise the device for approx. 2 hours at room tem-
perature after it has been disconnected from the mains.
- Avoid static charging of the material to be weighed, weighing container and windshield.
Major display deviations (incorrect weighing results) are possible if electromagnetic fields occur as well as
due to static charging and instable power supply. It is then necessary to change the location.

572/573/KB/DS/FKB-BA-defsi-0651 33
6.2 Unpacking
Carefully remove the balance from its packaging, remove the plastic wrapping and position the balance in its
intended working location.
6.2.1 Installation
Install the balance in such a fashion that the weighing plate is absolutely horizonta
6.3 Mains supply
Electric power supply is by means of the external mains supply circuit. The printed voltage level must comply
with the local voltage.
Only use original KERN mains supply circuits. The use of other makes is subject to approval by Kern.
6.4 Connecting peripheral equipment
The balance must be disconnected from the mains before connecting or disconnecting additional equipment
(printer, PC) to or from the data interface.
Only use KERN accessories and peripheral equipment with your balance. These have been ideally coordina-
ted to your balance.
6.5 Initial start-up
A warm-up time of 15 minutes stabilises the measured values after switching on.
The accuracy of the balance depends on the local acceleration of the fall.
Please be sure to observe the information in the chapter on adjusting.
6.6 Adjustment
As the acceleration value due to gravity is not the same at every location on earth, each balance must be
coordinated – in compliance with the underlying physical weighing principle - to the existing acceleration
due to gravity at its place of location ( only if the balance has not already been adjusted to the location in the
factory). This adjustment process must be carried out during the initial start-up, after change in location and
variation of surrounding temperature. It is also recommendable to adjust the balance periodically during
weighing operation in order to obtain exact measured values.
6.7 Adjusting
Using a precision weight, the accuracy of the balance can be checked at any time and adjusted.
Warning:
With verified balances, this precision adjustment is not possible
Adjustment procedure:
Check that the surrounding conditions are stable.
A short warm-up time of about 15 minutes is recommended for stabilisation.

572/573/KB/DS/FKB-BA-defsi-0651
34
6.7.1 Adjusting for verification (KERN 573)
General Information:
Prior to the procedure of verification the balance has to be adjusted.
Remark: The adjusting is only possible when not being blocked by the adjusting switch.
The switch for this step is located at the bottom of the housing between the two turning feet.
Switch to the right Adjusting function is released.
This setting has to be selected before starting the adjusting.
Executing the adjusting according to chapter 7.2 “Operation“.
After the adjusting has been completed successfully it is nec-
essary to turn the adjusting switch to the left for blocking.
Switch to the left Adjusting function is blocked.
After the adjusting has been completed successfully the ad-
justing has to be blocked by switching to the left.
The balance is now prepared for the procedure of verification.
After the verification the adjusting switch has to be secured by
a verification mark against access.
6.8 Verification
General:
According to the EU guideline 90/384/EEC balances must be verified officially if they are to be used as fol-
lows (legally regulated area):
a) For commercial transactions if the price of goods is determined by weighing
b) For the production of medines in pharmacies as well as for analyses in the medical and pharmaceu-
tical laboratory
c) For official purposes
d) For the production of finished packages
In case of doubt, please contact your local office of weights and measures.
Verification information
An EU qualification approval is available for those balances marked as appropriate for verification in the
technical data. In the event that the balance is applied in an area subject to verification as described above,
it must be officially verified and re-verified at regular intervals.
Re-verification of a balance is carried out in compliance with the respective legal provisions of the states.
The term of verification validity for balances in Germany, for example, is normally 2 years.
The legal provisions of the country of use are to be observed.

572/573/KB/DS/FKB-BA-defsi-0651 35
7 Operation
7.1 Display control panel
Keyboard
ON / OFF
Print weighing result
in MODE: No / descending
in % and counting mode:
Form reference
in MODE: Yes / ascending
MODE configuration
(see mode structure diagram page 2)
Formula weighing
in MODE: to left
Change g-pieces
Unit changeover
in MODE: to right
Tare
in MODE: back in weighing operation
Display-Symbol Meaning
==OVERLOAD== Overload: Weighing range has been exceeded
= ============ Underload: Weighing range has been fallen short of
<< . In counting and % mode: part too light
.
Automatic tare active
PTA .
Preselect tare Tare pre-selection active
D .
Difference in % during percentage weighing
Net .
Net weight of the components during formula preparation
SUM .
Gross weight of several components during formula prepa-
ration
Æ. Balance is in counting mode and currently displaying the
weight value of the counting amount

572/573/KB/DS/FKB-BA-defsi-0651
36
7.2 Operation
7.2.1 Weighing with Tare
7.2.2 Count – Selection Preference Piece
7.2.3 Percent %
7.2.4 Formula Weighing
7.2.5 Tolerance Weighing

572/573/KB/DS/FKB-BA-defsi-0651 37

572/573/KB/DS/FKB-BA-defsi-0651
38
7.2.6 Adjust
7.2.7 Preselect Tare
7.2.8 Auto Tare
7.2.9 Speed
7.2.10 Auto Off
7.2.11 Variable Factor
7.2.12 Preselect Unit

572/573/KB/DS/FKB-BA-defsi-0651 39

572/573/KB/DS/FKB-BA-defsi-0651
40
7.3 Interface RS 232 C
Technical Data
8-bit ASCII Code
1 start bit, 8 data bits, 1 stop bits, no parity bit
Baud rate adjustable to 2400, 4800 and 9600 baud (default), 19200 baud
Sup-D- 9 pol. is necessary .
When working with an interface correct operation is secured only if the corresponding KERN-interface-
cable (max. 2m) is used.
Description of the jack
Sup-D- 9 pol. 5 1
9 6
Pin 2: Transmit data
Pin 3: Receive data
Pin 5: Signal ground
Baudrate
The Baud rate for the data transfer is adjusted with the MODE-key. The following example demonstrates
how to set the Baud rate 4800.
Select Baud rate Display
1. Press MODE-key repeatedly until "PRINTER" is displayed.
2. Press YES-key.
3. Press MODE-key repeatedly until the desired Baud rate appears (for
instance 4800 Baud).
4. Press YES-key to select 4800 Baud. The tick-mark (X) confirms the new
setting.
5. Press MODE-key repeatedly until the balance displays in grams again,
or press tare key.
PRINTER?
2400 Baud
4800 Baud
4800 Baud X
0,0 g
7.4 RS 232 C Data output via interface RS 232 C
RS 232 C Data output via interface RS 232 C
General information
As a condition for the data transfer between the balance and a peripheral device (for instance printer, PC ...)
both devise have to be set on the same interface parameter (for instance baud rate, parity ...).
7.4.1 There are 4 methods for the data output via RS 232 C
Data output via PRINT-Key
The printing process can be released by the PRINT-key. In this case the settings AUTOPRINT and AUT-
PRINT PC should be deselected.
AUTOPRINT (Data output, after having loaded the balance)
The setting AUTOPRINT is in the PRINTER-routine, and there it can be selected or deselected. When
AUTOPRINT is active the actual weighing value will be sent via the RS 232 interface when the balance has
been unloaded and then loaded after having achieved the stability.
AUTOPRINT PC (Continuous data output)
The setting AUTPRINT PC is in the PRINTER-routine, and there it can be selected or deselected. When
AUTOPRINT PC is active the actual weighing values will be sent continuously via the RS 232 interface.
Data output by transfer of remote controls
The following functions can be released by the remote controls that will be transferred as ASCII signs to the
balance.
t Tare.
w a weighing value (or unstable) is sent via RS 232 interface.
s a stable weighing value is sent via RS 232 interface.
If the balance receives the command w or s, it acts without printing delay.

572/573/KB/DS/FKB-BA-defsi-0651 41
7.4.2 Description of the data transfer
Structure of each data transfer:
Without Numerator:
Bit-Nr. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
B B B B B B B B 0 . 0 B g B B CR LF
B*
B*: = Blank or %in the range of zero point.
B, 0, ., g: = Blank or weighing value with unit, depending on the load on the weighing plate.
CR: = Carriage Return
LF: = Line Feed
With Enumerator:
Bit.Nr. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
N N N B* B B B B B B B 0 . 0 B G B B CR LF
N: = Enumerator
7.4.3 Numerator
The Numerator is found under the menu point “Printer” and can be activated and deactivated. When editing
data with the use of the print option, the range of the numerator will increase by one place.
7.5 Printer
With the serial data output RS 232 a printer can be connected.
The printout shows the weight in grams. When the counting mode is selected the number of pieces or the
weight is printed. When the percent mode is selected, the percentage or the weight will be printed. Press
The PRINT-key to print weighing results. Select the enumerator to number the weighing continuously. Turn
off the balance or use the CLEAR function to Reset the enumerator to (000).
7.6 Underfloor weighing
Objects which, because of their size or shape, cannot be put on the scale, can be weighed by means of un-
derfloor weighing.
Proceed as follows:
•Switch off the balance.
•Turn the balance over, without loading the balance plate.
•Open the cover plate on the base of the balance.
•Hang on the hook for underfloor weighing .
•Place the balance over an opening.
•Hang the item to be weighed on the hook and carry out weighing.
! CAUTION !
Take care that the hooks used for the underfloor weighing are stable enough to hold the goods
which you wish to weigh (risk of breakage).
Always make sure that there are no living beings or materials below the load that could be injured or
damaged.
! NOTE !
After completing the underfloor weighing, the opening in the floor of the balance must be closed
again (dust protection).

572/573/KB/DS/FKB-BA-defsi-0651
42
8 Maintenance, upkeep, disposal
8.1 Cleaning
Please disconnect the device from the operating voltage before cleaning.
Only use a cloth dampened with mild suds and not aggressive cleaning agents (solvents or similar). Please
ensure that fluids are not able to get into the device and rub off using a clean, soft cloth.
Loose sample residue/powder can be removed carefully using a brush or hand vacuum cleaner.
Remove any spilt material to be weighed immediately.
8.2 Maintenance, upkeep
The device may only be opened by trained service engineers authorised by KERN.
Disconnect from the mains supply before opening.
8.3 Disposal
The operating company shall dispose of the packaging and the device in compliance with the valid national
or regional law of the operating location.
9 Troubleshooting
The balance should be switched off for a short time following an interruption in the programme sequence
and disconnected from the mains supply. It is then necessary to repeat the weighing process from the be-
ginning.
Help:
Interruption Possible cause
Weight display is not illuminated. •The balance is not switched on.
•The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
•Power supply interrupted. .
The weight display changes continually •Draught/air movement
•Table/floor vibrations
•The weighing plate is in contact with foreign mat-
ter.
•Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
The weighing result is obviously incorrect •The balance display is not set to zero
•Adjustment is no longer correct.
•Great fluctuations in temperature.
•Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
Switch the balance off if other error messages should appear and then switch on again. Contact the manu-
facturer if the error message does not disappear.

572/573/KB/DS/FKB-BA-defsi-0651 43
10 Declaration of conformity
Declaration of conformity
The electronic precision and platform balances
Typ: KERN 572-31
KERN 572-33
KERN 572-35
KERN 572-45
KERN 572-49
KERN 572-57
KERN KB 100-3
KERN KB1000-2
KERN KB10000-1
KERN DS4K0.05 S
KERN DS8K0.1 + S
KERN DS16K0.2
KERN DS36K0.5 + L
KERN DS65K1
KERN DS100K1
KERN DS150K2
KERN FKB 4K0.05
KERN FKB 8K0.1
KERN FKB 16K0.2
KERN FKB 36K0.5
KERN FKB 65K1
Correspond to the production model described in the EC type-approval certificate
and to the requirements of the following EC directives:
EC EMC directive Version 89/336/ECC
Applied harmonised norms, in particular
EN 61326: 1997
+ A1: 1998
+A2:2001
EN 61000-3-2: 1995
+ A1/A2 1998
+ A14 2000
EN 61000-3-3: 1995
If a change is made to the above mentioned appliances without consulting KERN
this declaration will become invalid.
Date: 07/01/2004 Signature:
Gottl. KERN & Sohn GmbH
Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72322 Balingen-Frommern, Tel. +49-07433/9933-0, Fax +49-07433/9933-149
This manual suits for next models
29
Table of contents
Other KERN Scale manuals

KERN
KERN KERN MPB User manual

KERN
KERN FOB-N series User manual

KERN
KERN FKTF 3K0.5LM User manual

KERN
KERN KFA 600V20S User manual

KERN
KERN ALJ User manual

KERN
KERN TKFP 600V20M-A TKFP-V20-A User manual

KERN
KERN KERN MBB-M series User manual

KERN
KERN MPK Series User manual

KERN
KERN MPC M Series User manual

KERN
KERN HCD 60K-2 User manual