Koven Bidop 3 User manual


WARNINGS & CAUTIONS
CAUTION Use the Bidop 3 under normal operating conditions, according to the instructions provided
in this operation manual and under the operating conditions stated in the product
specications.
CAUTION The Bidop 3 should only be used by skilled personnel.
CAUTION The Bidop 3 has not been evaluated for safety and compatibility in the MR environment.
The Bidop 3 has not been tested for heating or migration in the MR environment.
CAUTION Perform ultrasound testing/procedures prudently using the principle of ALARA (As Low As
Reasonably Achievable).
CAUTION Use ultrasound gel. Other materials such as baby oil or cream may produce incorrect
sounds or damage the probe.
WARNING The ultrasound gel included with purchase is non-sterile and should only be used noninvasively.
WARNING Discontinue use of ultrasound gel if an allergic reaction occurs.
CAUTION The probe transducer tip is thin and delicate. Please handle with care and use the probe
cap when not in use.
WARNING If any abnormality is found on the unit or patient, discontinue use in a manner safe to the
patient.
CAUTION Do not drop or place in an area where the unit may fall.
CAUTION Do not use simultaneously with an electrocautery debrillator, other ultrasonic device or a
mobile phone.
CAUTION Do not use in an area where a strong electromagnetic eld is present. Doing so may cause
incorrect measurements.
CAUTION Do not place near water or other liquid, where atmospheric pressure, temperature, humidity,
ventilation, sunlight, dust, salt sulfur, etc. may cause damage or where chemicals are stored
or gas may be generated.
CAUTION Protect the unit from vibration and shock during transport.
CAUTION Do not use excessive force when disconnecting cables.
CAUTION Use only designated Bidop 3 accessories.
CAUTION The Doppler box should not have direct contact with the patient or be placed in the sterile
area. The Doppler box cannot be sterilized or reprocessed.
WARNING The Doppler box and noninvasive probes should be cleaned following any patient contact.
CAUTION Do not submerge the Doppler box or probe in any liquid.
CAUTION Clean probes with a recommended probe cleaner before use.
CAUTION Do not perform any modications to the unit.
CAUTION Repairs should only be performed by authorized personnel.

TABLE OF CONTENTS
SECTION 1 Basic Operation (Vascular)
Warnings & Cautions
Introduction ................................................................................................................... 1
Features ....................................................................................................................... 1
1. Indications for Use, Contraindications, & Applications .................................. 2
1-1 Doppler Arterial & Venous Applications .........................................................…2
1-2 PPG (Photoplethysmography) Arterial & Venous Applications ........................ 2
1-3 Fetal Applications ............................................................................................. 2
1-4 Air Emboli Monitoring ....................................................................................... 2
2. Lower Extremity Doppler Arterial Waveform Pattern Analysis ....................... 3
3. Controls
3-1 Front Panel .................................................................................................... 4
3-2 Rear Panel ..................................................................................................... 5
4. Setup & Operation
4-1 Replacing the Battery ................................................................................... 6
4-2 Turning the Unit ON/OFF .............................................................................. 7
5. Basic Operation
5-1 Measuring Blood Velocity .......................................................................... 8-9
5-2 PPG: Arterial Pulse Waveform Studies ....................................................... 10
5-3 Measuring Fetal Heart Rate (FHR) ........................................................ 11-12
5-4 Fetal Heart Rate (FHR) Monitoring ........................................................13-14
5-5 Air Emboli Monitoring .................................................................................. 15
6. Mode & Menu Settings
6-1 Menu
6-1-1 Menu Operation .................................................................................. 16
6-1-2 Menu for Blood Velocity Measurement Mode .................................... 17
6-1-3 Menu for Blood Velocity Freeze Mode ............................................... 18
6-1-4 Menu for PPG (Photoplethysmography)(Measurement & Freeze) ........... 19
6-1-5 Menu for Fetal Heart Rate (FHR) Mode (Measurement & Freeze) ...... 20
6-1-6 Menu for 2.25 MHz BEEP Mode (Measurement & Freeze) ............... 21
6-2 Mode Setting Details
6-2-1 Memory - Store ................................................................................... 22
6-2-2 Memory – Read ................................................................................. 23
6-2-3 Clear ................................................................................................... 23
6-2-4 Mode (Baseline Mode) ....................................................................... 24
6-2-5 DIR (Flow Direction) ........................................................................... 24
6-2-6 Time (Time Scale) .............................................................................. 24
6-2-7 Flow (Blood Volume Flow) / Diameter (Estimated Vessel Diameter) 25
6-2-8 Others – Language ............................................................................ 25
6-2-9 Others – Unit ...................................................................................... 25
6-2-10 Others – Filter ................................................................................... 26
6-2-11 Others – Smooth .............................................................................. 26
6-2-12 Others – DISP (Display Mode) ......................................................... 26
6-2-13 Others – CAL (Calibration) ............................................................... 27
6-2-14 Others - Backlight (LCD Backlight) ................................................... 27
6-2-15 Others - Freeze ................................................................................ 27

6-2-16 Others – Auto-O (Automatic Shut O) ............................................ 28
6-2-17 Others - PRB20MHZ (20 Mhz Probe Selection) .............................. 28
Mode Settings for Optional Photplethysmography (PPG) Probe
6-2-18 AC / DC (Arterial / Venous) .............................................................. 28
6-2-19 COUNT ............................................................................................. 28
Mode Settings for Optional 2.25 MHz Probe
6-2-20 DISP (Display Mode) ........................................................................ 29
6-2-21 SOUND (Beep Sound for FHR Mode) ............................................. 29
6-2-22 UPPER (Upper Limit for FHR) ......................................................... 29
6-2-23 LOWER (Lower Limit for FHR) ........................................................ 29
6-2-24 DISP/BEEP (Display for Beep Mode, Waveform / Numerical data) ...... 30
6-2-25 LIMIT-1 (Limit for BEEP Mode) ........................................................ 30
6-2-26 LIMIT-2 (Maximum / Average for BEEP Mode) ............................... 30
6-2-26 OTHERS - UNIT (BEEP Mode Unit) ................................................ 30
7. LCD Display
7-1 Blood Velocity
7-1-1 Waveforms ......................................................................................... 31
7-1-2 Numerical Data ................................................................................... 32
7-2 Fetal Heart Rate Mode
7-2-1 Real Time FHR (Data Mode) .............................................................. 33
7-2-2 Monitoring FHR in Graph (Wave Mode) ............................................. 33
8. External Outputs
8-1 Headset ....................................................................................................... 34
8-2 Communication Port .................................................................................... 34
9. Maintenance ....................................................................................................... 35
10. Symbol List ........................................................................................................ 35
11. Operating Principles ......................................................................................... 36
12. Block Diagram ................................................................................................... 37
13. Specications .................................................................................................... 38
14. Operation, Storage & Transport Environments ............................................... 39
15. System Includes ................................................................................................ 39
16. Cleaning ............................................................................................................. 40
17. Disposal of Device ............................................................................................. 41
18. Warranty ............................................................................................................. 41
19. Safety Standards ..........................................................................................42-44
SECTION 2 Vascular Testing Instructions
1. Ankle Brachial Index ....................................................................................45-46
2. Segmental Blood Pressures ........................................................................47-48
3. Venous compression Studies .....................................................................49-50
4. PPG Toe Pressures & Toe Brachial Index ..................................................51-52
5. VenousReux ...............................................................................................53-54
TABLE OF CONTENTS

Section1
Section1
Basic Operation

Bidop3
1
Thank you for choosing the
Bidop3…
e Bidop 3 is a versatile bidirectional pocket Doppler with LCD display. It detects
arterial and venous blood flow in extremities, as well as fetal heart sounds. is guide
will help you get started using your new Doppler. Please read carefully to acquaint
yourself with the Bidop 3 operation.
FEATURES
• Bidirectional hand held Doppler with LCD
Displays real-time waveforms, numerical data and heart rate
• Microprocessor based
Various mode settings available for optimal usage
Menu is displayed on the LCD
30 velocity waveform storage capability
• Convenient probe activation button
Freezes waveform and numerical data for notation
• Interchangeable probes
2.25, 5, 8, & 10 MHz frequencies
• Automatic power “OFF”
• Optional PPG (Photoplethysmography) probe available for expanded arterial and
venous testing
NOTE
If using the Bidop 3 with the optional Smart-V-Link documentation software, please refer to the
Smart-V-Link operation manual for instructions on downloading waveforms and performing
tests with the Doppler and software.
CAUTION
Use the Bidop 3 under normal operating conditions, according to the instructions provided in
this operation manual and under the operating conditions stated in the product specications.
CAUTION
The Bidop 3 should only be used by skilled personnel.
CAUTION
The Bidop 3 has not been evaluated for safety and compatibility in the MR environment. The
Bidop 3 has not been tested for heating or migration in the MR environment.
CAUTION
Perform ultrasound testing/procedures prudently using the principle of ALARA (As Low As
Reasonably Achievable).

Bidop3
Bidop3
2
INDICATIONS FOR USE
The Bidop 3 is used for the detection of arterial and venous blood ow in extremities as well as fetal
heart sounds. It displays bidirectional velocity waveform, numerical data and fetal heart rate with heart
beat indicator.
CONTRAINDICATIONS
None known.
APPLICATIONS
(Refer to section 2 for Noninvasive Arterial & Venous testing instructions)
1-1 DOPPLER ARTERIAL & VENOUS APPLICATIONS
• Ankle Brachial Index (ABI) studies • Systolic pressures
• Peripheral vascular procedures • Blood ow velocity
• Segmental studies • Penile & digit systolic pressures
• Venous compressions
1-2 PPG (PHOTOPLETHYSMOGRAPHY) ARTERIAL & VENOUS APPLICATIONS
(Requires optional PG-21 Photoplethysmography probe, supplied separately)
• PPG toe pressures & Toe Brachial Index (TBI)
• PPG venous reux
1-3 FETAL APPLICATIONS
(Requires optional fetal probe, supplied separately)
• Evaluation of fetal heart rate and sounds throughout pregnancy except during rst
trimester, where fetal heart is not suciently developed.
1-4 AIR EMBOLI MONITORING
(Requires optional BF2M20S8A at monitoring probe, supplied separately)
• Monitoring for the detection of venous air embolism
1. INDICATIONS FOR USE, CONTRAINDICATIONS, &
APPLICATIONS

Bidop3
2. LOWER EXTREMITY DOPPLER ARTERIAL
WAVEFORM PATTERN ANALYSIS
The Doppler arterial waveforms obtained from the lower extremity may be classied into six
categories as an aid in interpretation.
TYPE TYPICAL WAVEFORM FINDINGS
0 The contour exhibits a steeply rising upslope at the onset of systole,
rapid systolic downslope, and reverse ow (below baseline). Dop-
pler sounds are loud and sharp. Normal peak forward velocity = 30
+/- 10 cm at the dorsalis pedis artery.
I The contour demonstrates a strong but diminished systolic compo-
nent and loss of reverse ow. The width of systolic pulse is broad-
ened. The Doppler sounds are diminished very little as compared
to the type “O” pattern.
II The contour shows prolongation of both the upstroke and the down-
stroke and diminished waveform amplitude (attening). Doppler
sounds are heard during systole and continuing through all or most
of diastole.
III The contour exhibits slowly rising velocity during systole, and the
amplitude of the waveform is reduced. Doppler sounds are not
sharp even during systole.
IV The amplitude of the waveform and the Doppler sounds is greatly
reduced.
V The amplitude of the waveform is extremely reduced. The contour
is hardly recognizable as an arterial blood ow waveform. The
Doppler sounds are very faint.
3

Bidop3
Bidop3
3. CONTROLS
3-1 FRONT VIEW
1) PROBE CONNECTOR To connect the probe to the Doppler
2) HEADSET CONNECTOR To connect a headset (not provided)
3) COMMUNICATION PORT To connect the Doppler to a computer
(for use with optional Smart-V-Link
software and cable)
4) VOLUME CONTROL To adjust the volume
5) LCD DISPLAY Displays real-time waveform, numerical
data, heart rate (HR) and menu for mode
settings
6) SPEAKER Outputs Doppler sound
7) SHUTTLE BUTTON Press in to enter the MENU, set mode or
execute commands
Move up or down to scroll through menu options
Displays memory data when in FREEZE mode
Turns the unit ON/OFF
8) BACK BUTTON To exit MENU mode
To go back to the previous menu
To change the display mode from WAVE to
DATA and vice-versa
4

Bidop3
3. CONTROLS
3-2 REAR VIEW
9) PROBE HOLDER For probe placement when not in use
10) STRAP HOLES To connect the carry strap
11) BATTERY COVER For battery placement
See 4-1 “Replacing the Battery” for
details.
12) PROBE Choose from 2.25, 5, 8, & 10 MHz
frequencies
13) PROBE BUTTON To turn the unit ON/OFF
To freeze the waveform & numerical data
14) PROBE CAP Protects the probe tip when not in use
5

Bidop3
Bidop3
4. SETUP & OPERATION
4-1 REPLACING THE BATTERY
1. Replace the battery when the low battery
icon appears in the lower right corner of the
LCD.
2. Turn the unit over. Press down on the
battery cover and slide the cover in the
direction of the arrow, as pictured to the
right.
3. Place a new 9-volt alkaline battery into
the unit, making sure that the positive and
negative electrodes correspond to the + and
– marks on the label inside the battery box.
6
Low Battery indicator
NOTE
If the battery is extremely low, the LCD display
will not operate and no speaker sounds will be
present.
NOTE
Use a 9-volt alkaline battery only. Use of a non-
alkaline battery may cause a shortage of power.
NOTE
If more than 1 minute is taken to replace the
battery, the mode settings will be changed back
to the default settings and all memory data will
be erased. To avoid loss of data, please have
a new battery ready before removing the old
battery.
If the message “MEMORY CLEARED PUSH
SHUTTLE BUTTON” is displayed when the unit
is turned on, all memory data has been cleared.
Press the shuttle button.

Bidop3
4. SETUP & OPERATION
4-2 TURNING THE UNIT ON / OFF
1. Install (1) 9 volt alkaline battery. See 4-1
“Replacing the Battery” for details.
2. Connect the probe to the Bidop 3 with the
alignment mark on the probe connector
pointed up (12 o’clock position).
3. Press the probe button or the shuttle button
to turn the unit ON.
4. Press and hold the probe button or the
shuttle button for longer than 2 seconds to
turn the unit OFF.
AUTOMATIC POWER OFF
During vascular operation, when the unit is in AUTO-OFF mode, the power will
automatically shut o after the following time passes
* After 15 minutes – during measurement mode
* After 2 minutes – while not in use
* After 5 minutes – while in FREEZE mode
7
NOTE
Release the button when the LCD shuts o. The
speaker will turn o after the probe (or shuttle
button) is released.
Alignment Mark
Shuttle
Button

Bidop3
Bidop3
5. BASIC OPERATION
5-1 MEASURING BLOOD VELOCITY
This section explains the fundamentals of
using the Bidop 3 to measure blood velocity.
For detailed instructions on performing
specic tests, please refer to Section 2
Vascular Testing Instructions.
1. Connect the probe to the Bidop 3 with the
alignment mark on the probe connector
pointed to the 12 o’clock position.
2. Press the probe button or shuttle button to
turn the unit ON. Turn the volume control to
the desired level.
3. Place ultrasonic gel on the probe tip or the
patients skin. Make sure there is enough gel
to cover the tip of the probe.
4. If numerical data is displayed on the LCD,
press the BACK button to display the
waveform.
8
Alignment Mark
BACK button
Shuttle
Button
Volume
Control
NOTE
If other mode settings are required, change
settings using the MENU. See 6. Mode & Menu
Settings for details.
CAUTION
Use ultrasound gel. Other materials such
as baby oil or cream may produce incorrect
sounds or damage the probe.
WARNING
The ultrasound gel included with purchase is non-
sterile and should only be used noninvasively.
WARNING
Discontinue use of ultrasound gel if an allergic
reaction occurs.

Bidop3
5. BASIC OPERATION
5-1 MEASURING BLOOD VELOCITY (cont.)
5. Place the probe on the measurement area at
an angle of approximately 45° to 60° against
ow and locate the point where maximum
Doppler sounds are heard.
6. When the waveform becomes rhythmic
and stable, wait 5 seconds or more without
moving the probe. Then, press the probe
button to FREEZE the waveform.
7. To view numerical data, press the BACK
button to change the display to DATA mode.
8. To save the waveform to view later or to
download to optional Smart-V-Link software,
see 6-2-1 MEMORY-STORE for details.
9
NOTE
If the probe button is pressed for longer than 2
seconds, the unit will be shut OFF.
BACK
button
Figure A
CAUTION
The probe transducer tip is thin and delicate.
Please handle with care and use the probe cap
when not in use.
WARNING
If any abnormality is found on the unit or
patient, discontinue use in a manner safe to the
patient.

Bidop3
Bidop3
5. BASIC OPERATION
5-2 PPG: ARTERIAL PULSE WAVEFORM STUDIES
(Requires the optional PPG probe {PG-21}, supplied separately)
Arterial pulse waveform studies by photoplethysmography are performed to determine the
presence or absence of pulsatile ow and to assess the state of perfusion in the tissue area
immediately beneath the sensor site. Use with a suitable cu and sphygmomanometer to
measure systolic pressures in the ngers and toes.
Thissection explainsthefundamentalsof obtainingaPPG arterialwaveformusingthe Bidop
3 with the PG-21 probe. For detailed instructions on performing PPG toe pressures
and calculating the Toe Brachial Index (TBI) ratio, please refer to Section 2 Vascular
Testing Instructions.
1. Connect the PG-21 probe to the Bidop 3 with the
alignment mark on the probe connector pointed
up (12 o’clock position). Press the probe button
or the shuttle button to turn the Doppler ON.
2. Press the Shuttle button to display the
MENU. Make sure MODE is set on combined
(arterial) mode. To change, scroll down with the
shuttle button until MODE is highlighted, then
press in on the shuttle button to change. Press
the BACK button to exit MENU and begin testing.
3. Ax the PPG sensor to the pad of the big toe
using double sided clear tape or the Velcro strap.
4. Press the probe button or shuttle button to turn
the unit ON.
5. The Bidop 3 automatically adjusts the gain and
basline. The PPG waveform is shown on the
LCD.
6. When the waveform becomes stable and
rhythmic, press the probe button to FREEZE it.
10
Alignment Mark
WARNING
If any abnormality is found on the unit or patient,
discontinue use in a manner safe to the patient.
Shuttle
button

Bidop3
11
5. BASIC OPERATION
5-3 MEASURING FETAL HEART RATE (FHR)(2.25 MHz probe, DATA mode)
(Required optional fetal probe, supplied separately)
This section explains the fundamentals of using
the Bidop 3 with a fetal probe to measure fetal
heart rate.
1. Connect the 2.25 MHz probe to the Bidop 3
with the alignment mark on the probe connector
pointed up (12 o’clock position).
2. Place ultrasonic gel on the top of the probe or the
patients skin. Make sure there is enough gel to
cover the top of the probe.
3. Press the probe button or shuttle button to turn
the Doppler ON. Turn the volume control to the
desired level.
4. Press in on the shuttle button to display the
MENU. Turn the shuttle button up or down to
select DISP. Make sure DATA is selected. If not,
press in on the shuttle button to select. Press the
BACK button to exit the MENU.
Alignment Mark
Shuttle
Button
Volume
Control
CAUTION
Use ultrasound gel. Other materials such as
baby oil or cream may produce incorrect sounds
or damage the probe.
WARNING
The ultrasound gel included with purchase is non-
sterile and should only be used noninvasively.
WARNING
Discontinue use of ultrasound gel if an allergic
reaction occurs.

Bidop3
Bidop3
5. Place the probe on the middle of the abdomen at a
right angle (90°) to the skin surface, and move slowly
to locate maximum Doppler heart beat sounds. The
probe needs to just lightly touch the skin.
6. When the heart rate becomes stable, press the
probe button to FREEZE it. The heart rate in
beats per minute (BPM) will be displayed on the
LCD screen. To return to monitoring mode, press
the probe button a second time.
NOTE
Be sure to verify the fetal heart rate. (Maternal
heart rate matches the maternal pulse rate).
NOTE
If stable signals are not being detected, an
asterisk (*) mark will be shown above “HR” on
the LCD screen.
NOTE
If the probe button is pressed for longer than 2
seconds, the unit will be shut OFF.
WARNING
If any abnormality is found on the unit or patient,
discontinue use in a manner safe to the patient.
5. BASIC OPERATION
5-3 MEASURING FETAL HEART RATE (FHR)(2.25 MHz probe, DATA mode)
(cont.)
12

Bidop3
5. BASIC OPERATION
5-4 FETAL HEART RATE (FHR) MONITORING (2.25 MHz Monitoring Probe, WAVE mode)
(Required optional 2.25 MHz at monitoring probe, supplied separately)
This section explains the fundamentals of using the Bidop3 with a fetal monitoring probe
to monitor fetal heart rate over a period of time.
1. Connect the 2.25 MHz fetal monitoring probe
(Model BF2M20S8C) to the Bidop3 with the
alignment mark on the probe connector pointed
up (12 o’clock position).
2. Place ultrasound gel on the probe transducer or
the patient’s skin. Make sure there is enough gel
to cover the top of the probe.
3. Press the probe button or shuttle button to turn
the Doppler ON. Turn the volume control to the
desired level.
4. Press in on the shuttle button to display the
MENU. Turn the shuttle button up or down to
select MODE. Press in on the shuttle button to
select WAVE. Press the BACK button to exit
MENU and begin monitoring.
5. Place the probe in the middle of the abdomen
at against the skin surface, and move slowly
to locate maximum Doppler heart beat sounds.
Hold probe in place or use a strap to attach the
probe to the abdomen during monitoring.
Alignment Mark
Shuttle
Button
Volume
Control
NOTE
Be sure to verify the fetal heart rate.
(Maternal heart rate matches the
maternal pulse rate).
CAUTION
Use ultrasound gel. Other materials such as
baby oil or cream may produce incorrect sounds
or damage the probe.
WARNING
The ultrasound gel included with purchase is non-
sterile and should only be used noninvasively.
WARNING
Discontinue use of ultrasound gel if an allergic
reaction occurs.
13

Bidop3
Bidop3
6. The LCD will show the monitoring waveform and
numerical heart rate. Press the probe button or
the BACK button to freeze the waveform. Up to
33 minutes of FHR monitoring can be displayed
over 4 screens (approx. 8 minutes per screen) on
the LCD. Turn the shuttle button down to show
the next screen. The “FREEZE indicator will
appear in the upper right corner of the screen.
7. To save the waveform in memory, see 6-2-1
MEMORY - STORE for details.
8. To view a previously saved waveform, see 6-2-2
MEMORY - READ for details. When viewing a
waveform from memory the meory number, e.g.
“#01” will be displayed in the upper right corner
of the LCD.
9. For additional details about the information
displayed on the LCD, see 7-2-2 Monitoring
Heart Rate in Graph (WAVE mode) for details.
WARNING
If any abnormality is found on the unit or patient,
discontinue use in a manner safe to the patient.
Memory number
Freeze indicator
Page number
Max. Min.
5. BASIC OPERATION
5-4 FETAL HEART RATE (FHR) MONITORING (2.25 MHz Monitoring Probe, WAVE mode)
(cont.)
14

Bidop3
WARNING
If any abnormality is found on the unit or patient,
discontinue use in a manner safe to the patient.
5. BASIC OPERATION
5-5 AIR EMBOLI MONITORING (2.25 MHz Monitoring Probe, BEEP mode)
(Requires optional 2.25 MHz at monitoring probe, supplied separately)
This section explains the fundamentals of using the Bidop 3 with the 2.25 MHz monitoring probe in
BEEP mode for performing air emboli monitoring.
1. Connect the 2.25 MHz monitoring probe to the Bidop
3 with the alignment mark on the probe connector
pointed up (12 o’clock position). Press the probe
button or the shuttle button to turn the Doppler ON.
2. Press the Shuttle button to display the MENU. Scroll
down with the shuttle button until DISP is highlighted.
Press in on the shuttle button to select and move the
shuttle button up or down to change to BEEP. Press in
on the shuttle button to set.
3. Use the Shuttle button to scroll to LIMIT-1 and press in
on the shuttle button to select. Move the shuttle button
up or down to change the upper blood velocity limit for
beep sounds. Press in on the shuttle button to set.
4. Use the Shuttle button to scroll to LIMIT-2 and press
the Shuttle button to select MAX or AVE. See 6-2-26
LIMIT-2 for details.
Press the BACK button to exit MENU and begin
testing.
5. Apply ultrasonic gel to the probe or the patient’s skin
and place the probe on the measurement area at a
right angle to the skin surface. Locate the point where
maximum Doppler sounds are heard.
6. The LCD will show the maximum velocity waveform. A
beep sound will be heard if the waveform exceeds the
upper blood velocity limit.
Alignment Mark
BACK button
Shuttle
Button
Volume
Control
CAUTION
Use ultrasound gel. Other materials such as
baby oil or cream may produce incorrect sounds
or damage the probe.
WARNING
The ultrasound gel included with purchase is non-
sterile and should only be used noninvasively.
WARNING
Discontinue use of ultrasound gel if an allergic
reaction occurs.
15
Table of contents
Other Koven Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Maico
Maico MA 25 Operation manual

Nasco Healthcare
Nasco Healthcare Simulaids 100-1296 instruction manual

Leica
Leica CM3050 S instruction manual

Cleaver Scientific
Cleaver Scientific multiSUB MSMINI7 instruction manual

St. Jude Medical
St. Jude Medical Optisure LDA220 user manual

Pari
Pari MONTESOL Nasal Douche Instructions for use