Land P80P User manual

Landcal Blackbody Source
Type P80P Operating Instructions
Page 14
Land Infrared
Dronfield
Sheffield S18 6DJ
England
Telephone: (01246) 417691
Facsimile: (01246) 410585
Email: [email protected]
© Land Infrared, 2001
Land Infrared
10 Friends Lane
Newtown, PA 18940-1804
U.S.A.
Telephone: (215) 504-8000
Facsimile: (215) 504-0879
Email: [email protected]
Publication Nº: PP970311
Part Nº: 198.031
Issue B/0801
Operating
Instructions
Landcal Blackbody Source
Type P80P

Operating Instructions Landcal Blackbody Source
Type P80P
Page 15
SAFETY INFORMATION
EN 61010-1 Symbol identification
Symbol Publication Description
IEC 417, Nº 5031 Direct current
IEC 417, Nº 5032 Alternatingcurrent
IEC 417, Nº 5033 Both direct and alternating current
IEC 617-2, Nº 02-02-06 Three-phasealternatingcurrent
IEC 417, Nº 5017 Earth (ground) terminal
IEC 417, Nº 5019 Protective conductor terminal
IEC 417, Nº 5020 Frame or chassis terminal
IEC 417, Nº 5021 Equipotentiality
IEC 417, Nº 5007 On (Supply)
IEC 417, Nº 5008 Off (Supply)
IEC 417, Nº 5172 Equipmentprotectedthroughoutbydoubleinsu-
lationorreinforcedinsulation(equivalenttoClass
II of IEC 536)
ISO 3864, Nº B.3.6 Caution, risk of electric shock
ISO 3864, Nº B.3.1 Caution
BS EN 100015 Observe precautions for handling electrostatic
dischargesensitive devices
BS EN 60825: 1992 Warning, laser radiation
ThisproductcomplieswithcurrentEuropeandirectivesrelatingtoelectromagneticcompatibility
andsafety(EMCdirective89/336/EEC;Lowvoltagedirective73/23/EEC).

Landcal Blackbody Source
Type P80P Operating Instructions
Page 16
Contents
1.0 INTRODUCTION 1
2.0 DESCRIPTION 2
3.0 SPECIFICATION 3
4.0 ELECTRICALSUPPLY 4
5.0 COMMISSIONING 5
5.1 Inspectiononreceipt 5
5.2 ConnectionstomainsSupply 5
5.3 Heatingup/coolingdownofthesource 5
5.4 UsingtheRS232serialinterfaceport 5
6.0 USINGTHESOURCE 6
6.1 Introduction 6
6.2 Measuringsensor(Platinumresistancethermometer)-ifsupplied 6
6.3 Operationofthesourceatbelowambienttemperature 6
7.0 CALIBRATIONOFRADIATIONTHERMOMETERS 7
7.1 Preparation 7
7.2 Thermometercalibration 7
7.3 Accuracyofcalibration 7
7.4 Calibrationprocedures 8
8.0 MAINTENANCE 9
8.1 Incorrectoperation/failure 9
8.2 NAMAScertification 9
8.3 Storageandtransportationcase 9
9.0 EUROTHERM2216OR2416TEMPERATURECONTROLLER 10
9.1 UserGuide 10
9.2 AlteringtheSetpoint 11
9.3 AlteringtheRampRate 11
9.4 Altering the Power Limit (when applicable) 11
9.5 ° to °F conversion 11
9.6 Altering the communication address 12

Operating Instructions Landcal Blackbody Source
Type P80P
Page 17

Operating Instructions Landcal Blackbody Source
Type P80P
Page 1
1.0 INTRODUCTION
The LANDCAL blackbody source type P80P is a variable temperature, portable black body radiation source
designed for use at temperatures up to 80°C (175°F).
The source is a primary standard black body for the high precision calibration of radiation thermometers over
therange-10to80°C(15to175°F). Whenthesetpointtemperatureisreached,theoutputfromthethermometer
under test is compared with the temperature of the source as measured by an optional Platinum resistance
thermometer whose calibration is traceable to National Standards.
When used in conjunction with the Platinum resistance thermometer which is supplied complete with a UKAS
(United Kingdom Accreditation Service) calibration certificate, high precision is obtained. Alternatively the
source can be used in three other ways.
(i) If traceability to National Standards is required to a larger value of uncertainty a UKAS certificate of
calibration for the source can be supplied. The relationship between the indicated temperature on the
controllerandtheradiancetemperatureasmeasuredbyasecondarystandardradiationthermometeris
reported.
(ii) Thetemperatureofthesourcecanbemeasuredbyusingaradiationthermometeroftraceablecalibration.
This method of calibration can be described as calibration by comparison with a standard radiation
thermometer. Thismethodofcalibrationusuallyresultsinthemostaccurateaserrorsduetotemperature
gradientsandnon-blackbodyconditionsareeliminated.
(iii) IftraceabilitytoNationalStandardsisnotrequired,thesourcecanbeusedwithoutanycertification. From
previouswork thetemperature, asshown onthe controllerindication, hasbeenfoundtoagree withthe
radiancetemperaturetowithin±4K(±8°F).
Thesourceprovidesawideangletargetwhichmakesitidealforusewithbothfixedinstallationandportable,hand-
heldthermometers.
TomaketheliningupofLANDfixedinstallationradiationthermometerssimpler,anopticalbenchassemblyisoffered
asanoptionalextra. Whenthesourceisstoodontheoptionaltransportationcase,thebenchtomidtargetdimension
equalsthatoftheopticalbenchassembly.

Landcal Blackbody Source
Type P80P Operating Instructions
Page 2
2.0 DESCRIPTION
The LANDCAL blackbody source type P80P comprises a cylindrical closed end tube (cavity) approximately
160mm/6.3inlongwith aninternaldiameter of50mm(2.0in). The cavityismanufactured fromAluminiumwhich
is blackened and the closed end is angled at 120° to increase the emissivity value.
The cavity is heated or cooled using Peltier elements. The temperature is controlled by a platinum resistance
thermometerconnectedtoa3-termdigitalcontrollerhavinga±0.1°Cor±0.1°Fresolution.
An optional standard platinum resistance thermometer possessing a traceable (UKAS - United Kingdom
Accreditation Service) Certificate is inserted into the cavity from the front of the source and used to determine
the true (radiance) temperature.
Fig.1 LandcalBlackbodySourceTypeP80P CA970256

Operating Instructions Landcal Blackbody Source
Type P80P
Page 3
3.0 SPECIFICATION
Maximum working temperature: 80°C (175°F)
Recommended temperature range: -10 to 80 °C (15 to 175°F)
Stability: Withthesourcecontrollingattemperaturetheradiancetemperature
willvarybylessthan±0.5K(±1°F)overa30minuteperiod.
Heatingrate: (Ambientto75°C) 60minutes
Coolingrate: (20°Cto-10°C) 90minutes(dependentonambient)
Radiationcavity: Material: Aluminiumwithblack,hightemperaturerefractorycoating
Design: 120°cone
Innerdiameter: 50mm(2.0in)
Internallength: 155mm(6.1in)
Benchtotubecentreheight: 100mm(2.5in)
Emissivity: >0.995
Controller: Type: Eurotherm 2216 or 2416 with RS 232 serial interface
Input: Resistance thermometer
Measuring sensor Type: Platinum Resistance Thermometer (UKAS certified)
(if supplied) Length: 450mm (17.7in) plus 2m (78.7in) cable
Diameter: 6mm outer diameter, inconel sheath
Uncertainty: <±0.1°C at 50°C
Part Nº: 135.142
Electrical supply: 220/240V a.c. 50 to 60 Hz. Part No. 135.181
110/120V a.c. 50 to 60 HZ. Part No. 135.199
Powerconsumption: 0.2KVA(220/240Voperation)
Overalldimensions: Height: 185mm(7.3in)
Width: 260mm(10.2in)
Depth: 315mm(12.4in)
Weight: Nett: 11kg(24lb)
Gross: 13kg(29lb)
NOTE
The controller fitted to the furnace is configured for °C operation. If °F operation is
required, details of how to re-configure the controller can be found in the controller
OperatingInstructions.

Landcal Blackbody Source
Type P80P Operating Instructions
Page 4
4.0 ELECTRICAL SUPPLY
TheP80Pissuppliedwitharemovableelectricalsupplycable. ThecablehasanIECstyleconnectorononeendthat
matestoaninteratedpowerentrymodule.
Thecolourcodeforthecableis:
Brownlead Live
Bluelead Neutral
Green/Yellowlead Earth
Thesourcemaybeconnecteddirectlytoa5ampfusedplugandsocket.
ThewiringdiagramoftheP80PisgiveninFig.2.
Fig.2 WiringdiagramfortheLandcalBlackbodySourceTypeP80P CA970219

Operating Instructions Landcal Blackbody Source
Type P80P
Page 5
5.0 COMMISSIONING
5.1 Inspection on receipt
Physically examine all items for damage that may have occurred during transit. Check the contents against the
packing note.
If any item has been damaged in transit this should be reported to the carrier and to the supplier immediately.
Donotreturn damagedinstrumentsto thesenderas thecarrierwill notthenconsider aclaim. Savethe packing
with the damaged article for inspection by the carrier.
5.2 Connections to mains Supply
Connect the brown lead to live, blue to neutral and green/yellow to earth. When connected to the mains supply
and switched on an indicator will light. At no time should any panels be removed when this mains indicator is
on.
5.3 Heating up/cooling down of the source
The source can be operated at any temperature in the range -10°C/15°F to 80°C/175°F. Set the controller to
the required value as follows:-
(i) When the source is switched on, the fluorescent indicator panel displays the measured value (upper
display) and the set point value (lower display).
(ii) To raise or lower the setpoint value, depress the respective up/down button. After a short delay, the
setpoint will change in the required direction. Release the up/down button when the required
temperature value is reached.
NOTE
All other control parameters are factory set and locked. For correct operation, it should not
be necessary to adjust any other parameters.
5.4 Using the RS232 serial interface port
Connect the source to the personal computer (PC) as shown in Table 1.
Source. 9 pin 'D'
connector
Controller
terminal Function
PC connector
25 pin 9 pin
Terminal 2 HE Rx (receive) Terminal 2 Terminal 3
Terminal 3 HF Tx (transmit) Terminal 3 Terminal 2
Terminal 5 HD Comm Terminal 7 Terminal 5
Table 1 P80P to PC serial communications (RS232) connection schedule

Landcal Blackbody Source
Type P80P Operating Instructions
Page 6
6.0 USING THE SOURCE
6.1 Introduction
TheP80Phasbeen designedtocreateanenclosureofuniformtemperature,ideal forthecalibrationofradiation
thermometers. The cone point of the cavity is placed in the area of minimum gradients within the source. When
calibratingradiationthermometersthetargetsizerequirementsofthethermometershould,wheneverpossible,be
fulfilledbythecone. Ifthethermometerviewsthewallsofthecavityresultsofgreateruncertaintywillbeachieved.
6.2 Measuring sensor (Platinum resistance thermometer) - if supplied
ProvisionhasbeenmadetomeasurethetemperatureofthetargetblockusingaPlatinumresistancethermometer
whichcanbeinsertedfromthefrontofthesourceintothecavity. Whenplacedinthemeasuringpositionthejunction
liesinthesameplaneastheconepoint,butisapproximately40mm/1.5inbelow. Thetemperatureofthesourceas
measuredbythesensoragreeswiththeconepointradiancetemperaturetowithin±0.5K(1°F).
Theoutputfromthesensorshouldbemeasuredonanindicatorordigitalvoltmeterhavingaresolutionof0.1°C.
Thisistherecommendedwaytoobtainthetruetemperatureofthetargetcavity. Thetemperatureindicationonthe
controllershouldnotbeusedasanaccuratemeasurementoftargetcavitytemperature.
6.3 Operation of the source at below ambient temperature
TheLANDCALBlackbodySourcetypeP80PusesPeltiermodulestocoolthesourcetotemperaturesaslowas-10°C/
15°F. Howeverthemodulesareratedasonlyachievinga30°C(55°F)belowambientcoolingeffect. Thismeansthat
thefinalcoolingtemperatureachievedandthelengthoftimeittakestoachievethistemperaturewillbedependent
ontheambienttemperature. Forexampleifthesourceisoperatedinanareawithambienttemperatureof30°Cthe
minimumachievabletemperaturewillbeapproximately0°C.
Condensationmayoccurinsidethecavitywhenthesourceisusedattemperaturesbelowambientandabove0°C. This
willinnowayaffecttheoperationorperformanceofthesource. Attemperaturesbelow0°Cthiscondensationwill
freeze and form a thin layer of ice. This ice formation will not affect the performance of the source. If prolonged
operation(saylongerthan8hourscontinuous)attemperaturesbelowambientareanticipated,itisrecommendedthat
thecavityispurgedwithadryinertgassuchasNitrogentoprevententryofwatervapourintothecavity.

Operating Instructions Landcal Blackbody Source
Type P80P
Page 7
7.0 CALIBRATION OF RADIATION THERMOMETERS
7.1 Preparation
Thecontrolsettingwillusuallybethenormalworkingtemperatureofthethermometertobetested.
ForcalibrationchecksthataretraceabletoNationalStandardsthetargettemperatureisthatindicatedbythestandard
platinumresistancethermometer. Iftraceabilityisnotrequiredthesourcecanbeusedwithouttheresistancethermometer.
From previous work the temperature as shown on the control indication has been found to agree with the radiance
temperaturetowithin±2K.
TomaketheliningupofLANDfixedinstallationradiationthermometerssimpleranopticalbenchassemblyisofferedasan
optionalextra. Whenthesourceisstoodontheoptionaltransportationcase,thebenchtomidtargetdimensionequalsthat
oftheopticalbenchassembly. Portableradiationthermometersareusuallyhandheld.
Positiontheholderontheopticalbenchtoobtainthedesireddistancebetweentargetandthermometer. Byadjustingthe
verticalandtransversevernierscrewsconfirmthattheholderissightedcorrectly.
7.2 Thermometer calibration
Whensoakedconditionshavebeenobtained,placethethermometerintheholderandmeasurethethermometeroutput
onthemeasuringapparatus. Immediatelyaftermeasuretheoutputfromthestandardplatinumresistancethermometer.
Convertbothoutputsintotemperaturebyreferencetotherelevantcalibrationtablesandcompare.
7.3 Accuracy of calibration
ThesourcehasbeendesignedfortheaccuratecalibrationofLANDradiationthermometers. Theaccuracythatcanbe
achievedbyusingthesourceisdependentupon:
(i) Theuncertaintyofcalibrationandresolutionofthemeasuringresistancethermometer.
(ii) Theemissivityofthesource.
(iii) Theresolutionoftheradiationthermometerundertest.
(iv) Thetemperaturegradientspresentinthesource.
Theuncertaintyoftheresistancethermometer,specifiedonthecalibrationcertificateissuedbythecalibrationlaboratory,
willbeafunctionof:-
(i) Thecalibrationlaboratoriescapabilities.
(ii) Thetypeofresistancethermometerundertest.
(iii) Thetemperaturerangecovered.
Valuesofbetween±0.01K(±0.02°F)and±0.1K(±0.2°F)aretypicalfortheuncertainty. Avalueof±0.05K(±0.1°F)and±0.2K
(±0.4°F)shouldbespecifiedfortheresolutiondependingonthetypeofmeasuringequipmentused.
Astheemissivityofthesourceislessthan1.00,theradiancetemperaturewillbedependentonthewavelengthofthe
thermometerundertest. Forexampleasourceoperatingatatemperatureof50°C(122°F)withemissivityof1.00willshow
atemperatureof50°C(122°F)forathermometerhavingapyroelectric(wavelength=8to14µm)detector. However,asource
operatingwithemissivityof0.995at50°C(122°F)forthesamethermometer,willshowaradiancetemperatureof49.6°C
(121.3°F)forthesamethermometer.
Mosthandheldthermometersandfixedinstallationthermometersusedinconjunctionwithanindicatorhavearesolution
of±1K(±2°F). Fixedinstallationthermometers,whoseoutputismeasuredonadigitalvoltmeter,willhavearesolutionof
between±0.1K(±0.2°F)and±0.5K(±1.0°F).
Anytemperaturegradientswithinthesourcewillcauseadifferencebetweenthetemperatureasmeasuredbytheresistance
thermometerandthetrueradiancetemperatureofthesource. Avalueof±0.3K(±0.6°F)istypical.
Todeterminethebestmeasurementcapability,theuncertaintyofeachindividualmeasurementcomponentshouldbeadded
together. Typicalvalueswouldbebetween±0.5K(±1°F)and±3K(±6°F).

Landcal Blackbody Source
Type P80P Operating Instructions
Page 8
7.4 Calibration procedures
Whencalibratingradiationthermometers,itisimportanttofollowdocumentedstepbystepprocedurestoensurethat
specifiedcalibrationconditions,suchascalibrationdistance,furnacetemperatureandaperturesizearealwaysmet.
If difficulty is experienced in writing procedures, LAND Infrared would be pleased to offer guidance as to what
calibrationconditionsshouldbeadoptedforLANDproducts.

Operating Instructions Landcal Blackbody Source
Type P80P
Page 9
8.0 MAINTENANCE
8.1 Incorrect operation/failure
The source is fully tested and evaluated before supply and should give years of trouble free operation. No
regular servicing or maintenance is required. In the unlikely event of a failure, we recommend the source is
returned either directly to a LAND company, or to one of the LAND distributors for repair.
8.2 UKAS certification
To continue to carry out calibration checks which are traceable to National Standards, it will be necessary to
obtaina Certificateof Calibration. Dependingonuseage, andthe methodof calibrationemployed,the Platinum
resistance thermometer and/or the P80P source and/or the standard radiation thermometer should be returned
toLANDevery 1to3 yearsforrecertification. UKAS CertificatesofCalibration areavailablefrom LandInfrared,
U.K. Certificates of calibration traceable to N.I.S.T. are available from Land Infrared, U.S.A.
8.3 Storage and transportation case
A custom built aluminium storage and transportation case is available as an optional extra. Use of this case is
recommended.

Landcal Blackbody Source
Type P80P Operating Instructions
Page 10
9.0 EUROTHERM TEMPERATURE CONTROLLER
9.1 User Guide
Fig. 3 Eurotherm Controller - front panel controls CA970253/a
When switched on, the controller lights up, goes through a short test routine, and then displays the measured
temperature and starts to control. The output light glows or flashes as heating occurs. The Page key allows
access to parameter lists within the controller; most lists and parameters are hidden and cannot be accessed
by the operator (they contain factory-set parameters which should not be changed). A single press of the page
keydisplaysthe temperatureunits,normallyset to°C;furtherpressesrevealthelistsindicatedin thenavigation
diagram, Fig. 4. The Scroll key allows access to the parameters within a list. A single press displays the
temperature units; further presses reveal the parameters in the current list indicated in the Navigation Diagram.
Some parameters are display-only; others may be altered by the operator. To return to the Home list at any
time, press the Page and Scroll keys simultaneously, or wait for 45 seconds. The Up and Down keys are used
to alter the setpoint or other parameter values.
Fig. 4 Eurotherm Controller - navigation diagram CA970281/a
EUROTHERM
OP1
Output light Measured temperature
Setpoint temperature
Scroll
Page
Down
Up
20.0
°C/°F/K
OP
wSP
SPrr
°C/°F/K oP
OP.Hi
Home
list
Output
list
Measured
temperature; use
Down/Up to
adjust setpoint
OP is a read-only value showing the current
output power level
wSP (Working Setpoint) is only present if
the ramp rate SPrr is active
Used in factory
configuration; not
accessible to the operator
ACCS
codE
Access
list
Cms
Addr
Communications
list

Operating Instructions Landcal Blackbody Source
Type P80P
Page 11
9.2 Altering the Setpoint
(i) Press either the Down or Up key once to display the setpoint.
(ii) Use the Down or Up key to adjust the setpoint value.
The display returns to the measured temperature when no key is pressed for 0.5 seconds.
9.3 Altering the Ramp Rate
(i) Press the Scroll key until the legend SPrr (SetPoint ramp rate) is displayed.
(ii) Use the Down or Up key to adjust the ramp rate value.
The ramp rate sets the maximum rate of heating or cooling in degrees per minute. A value of OFF
cancels the ramp rate, allowing heating and cooling at the maximum rate.
9.4 Altering the Power Limit (when applicable)
(i) Press the Page key until oP (output list) is displayed.
(ii) Press the Scroll key until OP.Hi (Output High) is displayed.
(iii) Press the Down key once to display the value of OP.Hi and write down the value.
WARNING
Do not increase the value without correct calculation; the furnace elements or
wiring could burn out.
(iv) To alter the value, the Down or Up key. Do not set the value to zero; this will prevent the furnace from
heating.
9.5 °C to °F conversion
To change the controller from °C to °F operation, proceed as follows:
(i) DepressboththeUp andDown keyswhilst turningontheinstrumentswitchuntilthecontrollerdisplays
ConF.
(ii) Use the Up or Down key to change the security configuration number to 45.
(iii) Leave the display at this setting for a few seconds until PASS is displayed.
(iv) Press the Page key repeatedly until InSt Conf is displayed.
(v) Press the parameter key to display Unit and use the Up or Down keys to change from °C to °F (other
units are K and none).
(vi) Press the Page key repeatedly until Exit (E=It) is displayed.
(vii) Use the Up key to select Yes.
(viii) Leave the controller for a few moments until it reverts to the normal display.
The temperature setting will now be made in the new units selected. All temperature limits and PID
parameters are resized automatically to suit the new units.
WARNING
Do not alter any other paramters

Landcal Blackbody Source
Type P80P Operating Instructions
Page 12
9.6 Altering the communication address
(i) Press the page key until cmS is displayed.
(ii) Press the scroll key until Addr (address) is displayed.
(iii) To alter the value press the up or down key.
The display returns to the measured temperature when no key is pressed for 45 seconds.

Operating Instructions Landcal Blackbody Source
Type P80P
Page 13
Table of contents
Other Land Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

PCB Piezotronics
PCB Piezotronics J351B03 Installation and operating manual

Ametek
Ametek Reichert 7CR user guide

Micropack
Micropack FDS301 Safety & technical manual

Secomam
Secomam UviLine Connect Series operating manual

Fluke
Fluke 715 instruction sheet

BPC instruments
BPC instruments Move Operation and maintenance manual