LANGER EMV-Technik P331-2 User manual

2021.01.24 User manual P331-2_PS_GM.doc
User manual
P331-2 set
ESD generator
(IEC 61000-4-2)
Copyright © January 2017
LANGER EMV-Technik GmbH

- 2 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
Table of Contents Page
1Declaration of Conformity...........................................................................................3
2General Information ....................................................................................................4
2.1 Storage of the User Manual................................................................................................4
2.2 Reading and Understanding the User Manual ....................................................................4
2.3 Local Safety and Accident Prevention Regulations.............................................................4
2.4 Images ...............................................................................................................................4
2.5 Limitation of Liability...........................................................................................................4
2.6 Errors and Omissions.........................................................................................................4
2.7 Copyright............................................................................................................................4
3Scope of delivery.........................................................................................................5
4Technical Parameters .................................................................................................6
4.1 P331-2 ESD generator .......................................................................................................6
4.2 BPS 203 Burst Power Station.............................................................................................6
4.3 SM 02-01 Shunt..................................................................................................................7
5Safety............................................................................................................................8
5.1 Labels and Signs................................................................................................................8
5.2 Intended Use......................................................................................................................8
5.3 Reasonably foreseeable Misuse.........................................................................................8
5.4 Staff Requisition .................................................................................................................9
5.5 Safety Instructions..............................................................................................................9
6P331-2 ESD Generator (IEC 61000-4-2)....................................................................10
6.1 Design and Function of the P331-2 probe ........................................................................10
6.2 Characteristics..................................................................................................................11
7Operational Notes......................................................................................................12
8System Set-Up ...........................................................................................................13
9Verifying the Waveform.............................................................................................15
10 Warranty..................................................................................................................16

- 3 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
1 Declaration of Conformity
Manufacturer:
Langer EMV-Technik GmbH
Nöthnitzer Hang 31
01728 Bannewitz
Germany
Langer EMV-Technik GmbH herewith declares that the
P331-2 set, ESD generator
with P331-2, BPS 203
conforms with the following relevant regulations:
-EMC Directive 2014/30/EU
-Low-Voltage Directive 2014/35/EU
-Restriction of certain Hazardous Substances 2011/65/EU
The following applicable standards were used to implement the requirements specified by the
aforementioned directives:
-EN 61000-6-1:2007-10 (EMC)
-EN 61000-6-3:2011-09 (EMC)
-EN 61010-1:2011-07 (Safety)
-DIN EN 50581:2013-02 (Restriction of hazardous substances)
Name of the person authorized to compile the technical file:
Gunter Langer
Bannewitz, 2020-03-02
Signature:
_________________
G. Langer, Managing Director

- 4 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
2 General Information
2.1 Storage of the User Manual
This user manual enables the safe and efficient use of the P331-2 set. It must be kept close at
hand and accessible to the user.
2.2 Reading and Understanding the User Manual
Read the user manual carefully, observe the safety information (Chapter 5) and follow the
instructions given in this manual before putting the device into service.
2.3 Local Safety and Accident Prevention Regulations
The local accident prevention and general safety regulations also apply to ensure that the
P331-2 set is used for its intended purpose.
2.4 Images
Figures have been included in this user manual to assist the reader's understanding but may differ
from the device's actual version.
2.5 Limitation of Liability
In the following cases, Langer EMV-Technik GmbH can assume no liability for damage to property
and personal injury if:
- The information given in this user manual has not been observed.
- P331-2 set was operated by staff not qualified in the field of EMC.
- P331-2 set was subjected to unauthorized modifications or technical changes.
- P331-2 set was not used according to its intended purpose.
- Spare parts or accessories were used that had not been approved by
Langer EMV-Technik GmbH.
The actual scope of delivery may deviate from the illustrations and texts in this user manual due to
the customization of orders or due to technical changes and innovations.
2.6 Errors and Omissions
The information in this manual has been carefully checked and is believed to be accurate;
however, the Langer EMV-Technik GmbH assumes no responsibility for any clerical, typographical,
or proofreading errors, or omissions.
2.7 Copyright
The content of this user manual is protected by copyright law and may only be used in connection
with the P331-2 set. This user manual may not be used for any other purpose without the prior
written approval of Langer EMV-Technik GmbH.

- 5 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
3 Scope of delivery
Item Designation Type Parameter Pcs.
01 ESD Generator (probe) P331-2 1
02 Burst Power Station BPS 203 1
03 Control software BPS 203-Client 1
04 Shunt SM 02-01 0.1 R 1
05 Control cable FBK 12P 1 m 1
06 High-voltage cable HV FI-FI 1 m 1
07 USB cable type A-B USB-AB 1
08 Measuring cable SMA-SMB 1 m 1
09 Power supply unit 12 V / 1 A 1
10 System case P331 case 1
11 User manual 1
12 Quick guide 1
Important: The scope of delivery may deviate depending on the respective order.

- 6 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
4 Technical Parameters
4.1 P331-2 ESD generator
Frequency range
0.1 Hz –10 Hz
Voltage range
100 V –6 kV
Pulse form
0.7 / 60 ns
Energy storage capacity
150 pF
Internal resistance
330
Weight
0.25 kg
Sizes (L x W x H)
(79 x 41 x 40) mm
Table 1: P331-2 technical parameters
4.2 BPS 203 Burst Power Station
Frequency range
0.1 Hz –30 Hz
Output voltage
± (0.1 ... 6) kV
Supply voltage
12 V / 1 A DC
Interface
USB
Control software
BPS 203-Client
Weight
350 g
Sizes (L x W x H)
(175 x 122 x 61) mm
Table 2: BPS 203 technical parameters

- 7 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
4.3 SM 02-01 Shunt
Frequency range
DC –3 GHz
Input resistance
0.1 Ω
Output resistance
50 Ω
Correction factor
26 dB
Single pulse capacity 5/50 ns, 1/60 ns
360 A
Table 3: SM 02-01 technical parameters
Burst sequences IEC 61000-4-4
5/50 ns
Burst repetition
300 ms
Impulses per burst
100
Pulse frequency
500 kHz
Current
150 A
Table 4: SM 02-01 continuous power rating

- 8 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
5 Safety
5.1 Labels and Signs
General warning sign
Warning; Electricity
Prohibition sign; No access for
people with active implanted
cardiac devices.
Table 5: Safety signs
Safety instructions in this user manual are marked by symbols (Table 5). Observe the safety
precautions and act cautiously to avoid accidents as well as personal and material damages.
5.2 Intended Use
The P331-2 set is used for conducted coupling of ESD pulses into ICs. The P331-2 probe sizing
orientates itself by mechanisms of the ESD coupling into electronic assemblies (according to IEC
61000-4-2 / HMM
1
). The BPS 203 burst power station supplies and controls the probe.
The P331-2 probe and BPS 203 burst power station are built according to their specified use
therefore they should be used only for the following purposes:
- Injection of ESD pulses into IC pins or balls with P331-2 powered by BPS 203.
- Control of the P331-2 via BPS 203-Client or DLL.
- The P331-2 set must be used in conjunction with the ICE1 set from Langer EMV-Technik GmbH.
Any use beyond these specifications is considered contrary to the intended use.
5.3 Reasonably foreseeable Misuse
Danger resulting from misuse!
Misuse of the P331-2 set can lead to dangerous situations!
-Use of the product outside of the given specifications.
-Modification or changing of the product without consent of Langer EMV-Technik GmbH.
-Operating the product with a technical fault.
1
HMM –Human Metal Model
Warning!

- 9 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
5.4 Staff Requisition
Only qualified staff with training, knowledge, and experience in the field of EMC is allowed to
operate the P331-2 set.
Persons whose ability to perform is influenced or impaired by alcohol, drugs, or pharmaceuticals,
are not allowed to operate the P331-2 set.
Certain functions may only be carried out by qualified personnel of Langer EMV-Technik GmbH.
5.5 Safety Instructions
Danger resulting from Electricity!
Risk of injury by electrocution!
Only connect the high-voltage cable to the P331-2 probe before operation.
Don’t touch the probe tip of a P331-2 probe while it is in operation.
-If insulation is damaged, the power supply has to be disconnected immediately.
-Replace damaged parts with undamaged parts before operation. Contact Langer EMV-
Technik GmbH for proper replacements.
-Protect live parts from moisture to avoid short circuits.
Never leave a Langer EMV-Technik GmbH product unattended while this is in operation.
Danger resulting from electromagnetic fields!
Risk of affecting a cardiac device!
Persons with a cardiac device, such as a pacemaker, are not allowed to work on or approach the
P331-2 set while it is in operation.
Warning; Electricity
Prohibition; No
access for people
with active implanted
cardiac devices

- 10 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
6 P331-2 ESD Generator (IEC 61000-4-2)
The probe is used to generate standard ESD pulses according to IEC 61000-4-2 for ESD injection
into the device under test via conductors (Figure 2: P331-2 pulse form).
The P331-2 probe allows the user to couple ESD into IC pins via conductors according to the
standard IEC 61000-4-2 both directly and indirectly via coupling networks (standard). Coupling
networks are used for coupling into interface connections or special high-speed interfaces such as
USB, LVDS, Ethernet, etc. Inductive or capacitive couplers are suitable coupling networks
(Information: Langer EMV-Technik GmbH).
6.1 Design and Function of the P331-2 Probe
Figure 1: Description of the P331-2 probe
The pin contact is the P331-2 probe's high-voltage (HV) output that is used to inject the ESD pulse
into the test IC.
The test pulse is generated in the probe through a high-voltage switch and the coupling networks
that are required by the standard (Figure 2). The high voltage that is needed for the pulse
generation is generated in the BPS 203 and led to the HV port of the P331-2 probe via a high-
voltage cable. The BPS 203 controls the P331-2 probe. The signals are led to the control cable
port via a control cable. The pulse/contact LED indicates when an ESD pulse is triggered and the
device under test is contacted. The LED lights up green as soon as there is a galvanic connection
between the pin contact and the device under test. A red light signals the triggered pulses.
The Power LED signals the P331-2 probe’s power supply. The probe's GND contact area ensures
low impedance, all-over contact with the GND 25 ground plane. Magnets that are integrated in the
probe hold it on the ground plane.

- 11 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
6.2 Characteristics
Figure 2: P331-2 pulse form
Figure 3: P331-2 equivalent circuit diagram
The ESD pulse is characterized by its current characteristic which is shown in Figure 2.
Figure 3 shows the equivalent circuit diagram of the P331-2 probe. Both are in accordance with
the standard IEC 64000-4-2. Please refer to Table 6 for the respective waveform parameters.
HV [kV]
I (max) [A]
+/- 10%
I (30 ns) [A]
+/- 30%
I (60 ns) [A]
+/- 30%
0.5
1.86
1
0.5
1
3.72
2
1
2
7.5
4
2
4
15
8
4
6
22.5
12
6
Table 6: Waveform parameters
The respective short-circuit peak current can be calculated on the basis of the generator voltage
UVG when the probe is in operation.
IP = UVG ·K where: K = 3.7 A / kV
The equation reveals that the probe supplies 3.7 A per kV of the generator voltage.
Note: The pulse form is only guaranteed if the P331-2 probe is operated at a minimum voltage of
200 Volt.

- 12 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
7 Operational Notes
-The test set-up should always be operated via a filtered power supply.
-Attention! Functional near fields and interference emissions may occur when operating
EMC test set-ups. The user is responsible for taking measures to prevent any interference
to the correct function of products outside the EMC environment of the test set-up (in
particular through radiated interference).
This can be achieved by:
-observing an appropriate safety distance,
-use of shielded or shielding rooms.
-The disturbances that are injected into the ICs can destroy (latch-up) the device under test if their
intensity is too high. Protect the device under test by:
-increasing the disturbance gradually and stopping when a functional fault occurs,
-interrupting the power supply to the device under test in the event of a latch-up.
-Attention! Make sure that internal functional faults are visible from outside. The device
under test may be destroyed due to an increase in the injection intensity if the faults are
not visible outside. Take the following measures as necessary:
-monitoring of representative signals in the device under test,
-special test software,
-visible reaction of the device under test to inputs (reaction test of the device under test).
We cannot assume any liability for the destruction of devices under test!

- 13 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
8 System Set-Up
Figure 4: System set-up with P331-2 probe and the ICE1 set
The components marked with an asterisk (*) are not included in the scope of delivery.
Figure 4 shows the set-up of the IC test system with the ICE1 set
2
(Fehler! Verweisquelle konnte
nicht gefunden werden.) and the P331-2 set. The BPS 203 burst power station generates a high
voltage and supplies this to the P331-2's HV input via the HV FI-FI 1 m cable. In addition, the BPS
203 also controls the P331-2 via the FBK 12P 1m control cable. The PC in turn controls the
BPS 203 via the USB-AB cable. The BPS 203-Client software is installed on the PC.
The ESD current pulse is generated from the high voltage in the P331-2 probe (Figure 3). The
current pulse (Figure 2) flows into the test IC when contact to the pin is made.
The test IC is mounted on a special test board
3
. The test board is inserted into the GND 25
ground plane and connected to the CB 0708 connection board.
The ground plane and the connection board are integral parts of the ICE1 set. The evaluation of
signals from the test IC may require external devices such as an oscilloscope or special test
hardware (Figure 5).
2
For details of the ICE1 set please see the appropriate user manual.
3
For manufacturing of the test board: "Guideline IC EFT immunity", Langer EMV-Technik GmbH

- 14 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
Figure 5: Test set-up with the P331-2 set and ICE1 set
The tasks and devices listed in the table below are described in their respective manuals:
Tasks and Devices
Manuals
Instructions for the development of the
test board
Test process
Guide line IC EFT immunity (Langer EMV-
Technik GmbH)
GND 25 ground plane
CB 0708 connection board
OA 4005 oscilloscope adapter
TH 22 probe head holder
Monitoring and controlling the test IC
ICE1 set user manual
Table 7

- 15 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
9 Verifying the Waveform
The SM 02-01 shunt can be used to verify the waveform of the current pulse. The shunt has a
bandwidth of 3 GHz and can be loaded with a maximum pulse current of 180 A in the single-pulse
mode of the BPS 203.
The shunt is inserted into the GNDA 02 ground adapter (Figure 6). The SMA output is connected
to the 50R input of an oscilloscope with a corresponding bandwidth. The oscilloscope's attenuator
is set to 26 dB (x20). 1 V at the display corresponds to a current of 1 A in the probe. When using
an oscilloscope with a bandwidth > 3 GHz, please note that this is limited to 3 GHz.
The waveform has to be verified prior to every major measuring job. Provided the waveform does
not deviate from the given parameters, the P331-2 probe only has to be calibrated every two years
by Langer EMV-Technik GmbH.
Figure 6: Measurement set-up with P331-2 and SM 02-01 to verify the pulse form

- 16 -
LANGER
EMV-Technik
DE-01728 Bannewitz
mail@langer-emv.de
www.langer-emv.com
P331-2 set
10 Warranty
Langer EMV-Technik GmbH will remedy any fault due to defective material or defective
manufacture, either by repair or by delivery of spare parts, during the statutory warranty period.
This warranty is only granted on condition that:
- the information and instructions in the user manual have been observed.
The warranty will be forfeited if:
- an unauthorized repair is performed on the product,
- the product is modified,
- the product is not used according to its intended purpose.
This document may not be copied, reproduced or electronically processed, either in its entirety or in part, without the
prior written permission of Langer EMV-Technik GmbH. The management of Langer EMV-Technik GmbH assumes no
liability for damage that may arise from using this printed information.
LANGER
Noethnitzer Hang 31
Tel.: +49 351/430093-0
EMV-Technik GmbH
DE-01728 Bannewitz
Fax: +49 351/430093-22
www.langer-emv.com
Table of contents