F1.4 Calibration Unit
Selectable parameters: 0 ---- kg (default value)
1 ---- lb
F1.5 Gravitational Acceleration
Selectable parameters: 9.70000~9.99999. Default value =
9.79455.
F1.6 Null-point Calibration
【 】 Keeping empty the scale
Remove the weights on the weighing platform to
guarantee the scale is in the empty state. Press the
[Print] key and the meter will display [ ].
The displayed digits will reduce slowly until the
meter displays [ ]. In the end it will display
[ ] for one second, which indicates the end
of null-point calibration.
F1.7 Load-point Calibration
【 】 Loading weights
Load weights on the weighing platform to ensure
that 10% of full-scale value ≤ weight of weights
≤ full-scale value, and then press the [Print] key
to start the next step.
【 】 Entering the same weight value as that of the loaded
weights.
Entering the same weight value as that of the loaded
weights, please press the [Print] key after the scale
becomes stable, and the meter will display [ ].
After that, the displayed digits will reduce slowly until
the meter displays [ ]. In the end it will display
[ ] for one second, which indicates the end of null-
point calibration.
F1.8 Automatic Null Tracking
Selectable parameters: OFF, 1 d, 2 d, 3 d (default value)
F1.9 Automatic Reset Range at Startup
Selectable parameters: OFF, 2 %, 10 %, 20 % (default value)
F1.10 Button Reset Range
Selectable parameters: OFF, 2 %, 10 % (default value), 20 %
F1.11 Digital Filter