7
ENG
8. Place stethoscope head so that it was located in depres-
sion, somewhat higher than elbow bend (Fig. 12). If You
use the device with built-in stethoscope, to reach better
auscultation of pulse on full hand, the cuffmay be turned
round the hand approximately by quarter of circle (60-90˚)
in such way that stethoscope head would be positioned on
inner side of hand (more close to the body).
Fig. 12
9. Pressure shall be measured in sitting or lying position of human. IN SITTING
POSITION WATCH THAT PART OF HAND WITH CUFF WOULD BE LOCATED AT THE
LEVEL OF HEART, AND HAND WOULD BE FREELY LOCATED ON THE TABLE AND
WOULD NOT MOVE
.
MEASUREMENT PROCEDURE
1.
Insert metal tubes of stethoscope into ears. Close air valve on air pump,
having turned it clockwise. Squeezing air pump, pump up the cuff, listen-
ing to pulse by stethoscope. After You stopped to listen to pulse, pump
up the cuffby 30 mm Hg more
.
2.
Slowly opening air valve, turning it counterclockwise, bleed down pressure
in the cuff. Watch that pressure in the cuffwould fall down at speed 2-4 mm
Hg per second. This is necessary to get accurate result.
3.
As soon as You hear weak pulse beats, memorize manometer reading. This
is Your systolic (upper) arterial pressure
.
4. Pressure in the cuffis continuing falling down at the same speed (2-4 mm Hg
per second). You are continuing to listen to pulse. Sounds which You hear will
change. Unlike the first beats, they will become softer, resembling rustling.
At the moment when You practically stop seize pulse, memorize manometer
reading. This is Your diastolic (lower) arterial pressure.
CARE, STORAGE, REPAIR AND DISPOSAL
1. Keep this Device from exposure to higher humidity, direct sunlight, shocks.
2. Do not keep and use this Device near heating installations and open fire.
3. Keep the Device clean and protect it from dust.
4. Contact of device with aggressive solutions is not allowed.
5. Protect the arm cuffand air tubes from contacting on sharp things.
6.
This Device does not contain special controls to adjust the measurement
accuracy. Independent opening of manometer is prohibited. Repair the
Device only in authorized organizations.
7. Service life of device is indicated in section TECHNICAL SPECIFICATIONS.
Service life is determined from the moment of commodity delivery to the
customer. On expiration of the warranted service life apply from time to
time to authorized repair organizations to check the technical condition of
the Device.