LPG Mobilift M6 Beauty User manual

1
In order to respond better to customer requirements and expectations, LPG Systems is continuously
researching ways of improving the design and quality of its products. This will explain the few minor
differences between your equipment and the item described in this guide.
≥ATTENTION
GU 1303-EN_ind B1
Edition dated 06/17
OPERATOR’S MANUAL
Mobilift M6®
Beauty
Registered trademark
Please read the complete manual carefully
before using your equipment
© LPG Systems 2017. LPG, Endermolift™ and Mobilift M6® are registered trademarks of LPG®.
Reproduction, even in part, is strictly prohibited.
Congratulations on the purchase of your Mobilift M6
Beauty
. This model represents
the fruits of numerous years of experience in the design and production of
cutaneous tissue care systems. You will fully experience the technical perfection
and reliability of LPG Systems, which is the leader in this sector.
This operator’s manual contains the operating description, basic maintenance
instructions to be performed periodically and the safety instructions.
Your device is intended for use in the care of connective tissue. It should be used
only by a professional who has attended the manufacturer’s training provided by
LPG Systems or an approved provider, if you live outside France.
If you have any doubts whatsoever concerning the operation or maintenance of
your equipment, please do not hesitate to contact LPG via the Customer Service
Department of your distributor.

2 3
Unpacking instructions to be retained.
The manufacturer reserves the right to amend the product technical specifications without prior
notice. Any reproduction – even in part – is prohibited. All the illustrations in this operator’s manual
are non-binding.
≥ATTENTION
> PACKAGE CONTENTS > TABLE OF CONTENTS
> One Mobilift M6 Beauty unit
> One main Ergolift head
> Two Ergolift chambers (Lift 20 and Lift10)
> One electrical power cord
> One operator’s manual
> One POS marketing set
1. DESCRIPTION OF THE UNIT ....................4
2. DESCRIPTION OF THE CONTROLS .................6
3. PRECAUTIONS OF USE .......................7
4. MAINTENANCE ..........................11
5. ERGOLIFT HEAD ..........................16
6. TROUBLESHOOTING .......................20
7. TECHNICAL SPECIFICATIONS ..................21
8. WARRANTY ............................22
GENERAL WARRANTY CONDITIONS .......................22
WARRANTY ACTIVATION ..................................25

4 5
The unit can only operate if it is connected to the mains by its power cord and provided the ON switch
has been activated and the green voltage light is on.
After switching on the unit, please wait a few seconds for the screen to display information.
≥ATTENTION
DESCRIPTION OF THE UNIT
1
> MOBILIFT M6® BEAUTY > MOBILIFT M6® BEAUTY (CONT’D)
DESCRIPTION OF THE UNIT
1
HOSE
USB PORT
ACCESSORIES HOLDER
FORWARD-FACING TOUCH
SCREEN
TREATMENT CHAMBER
ERGOLIFT HEAD
HANDLE
AND ACCESSORIES
HOLDER CLIP
ACCESSORIES HOLDER
POWER CORD
ON/OFF SWITCH
FILTER BOX

6 7
11Europe VIIG-H05VVF3G0,75-C13; Italy I/3/16-H05VVF3G0,75-C13; Switzerland 12G-H05VVF3G0,75-C13; UK
BS13/13-H05VVF3G0,75-C13; Japan 498GJ-VCTF3X1,25-C13; USA, Canada, Mexico N5/15-SJT3X18AWG-C13
≥ATTENTION
> CONTROL SCREEN
> ERGOLIFT HEAD
DESCRIPTION OF THE CONTROLS
2
PRECAUTIONS OF USE
3
> WARNING
> IMPORTANT SAFETY INFORMATION
All safety precautions must be observed whilst using electrical equipment. Please read all
safety notices and precautions prior to use of the equipment.
DANGER - TO MINIMIZE THE RISK OF ELECTRICAL SHOCK:
• Always disconnect the equipment from the electrical supply outlet after use and before
cleaning and maintenance.
• Check that the supply voltage of the unit indicated on the data plate complies with
the mains voltage.
• The unit must be connected by the power cord1 supplied to a grounded outlet in accordance
with current electrical standards. Electrical adapters must not be used with this equipment.
• The equipment must not be left unattended
whilst connected to the electrical supply.
• Disconnect the unit from the electrical
supply if it is not going to be used for a long
period.
• Special attention is required whilst using
the equipment with, or in the proximity of
children or disabled persons.
• Never use the unit for purposes other than
those recommended by LPG Systems. Only
use the treatment heads supplied with your
unit or those recommended by LPG.
• Never use the equipment if:
- the electrical supply cable or outlet
is damaged.
- the equipment does not function correctly.
- the equipment is damaged or has fallen
or been dropped.
- the equipment has been exposed
to excessive humidity.
In such cases, return the equipment to an
approved LPG service centre.
• Do not move the unit by pulling on the
electrical supply cable.
• Fully unwind the electrical power cord and
keep it away from warm surfaces.
• Never use the equipment if the ventilation
ports are obstructed. Ensure that the
ventilation ports are kept clear of dust or
other contaminants.
• Do not allow solid debris, liquid or other
foreign bodies to fall or be sucked into the
unit, as these could cause damage.
• Never use the equipment on a dusty, uneven
floor, or in a moist atmosphere.
• Never use the equipment in the presence of
aerosols or oxygen.
• Before disconnecting the unit from the
electrical supply, set all controls to the ‘off’
position and unplug the unit.
• It is prohibited to modify this equipment
without prior authorisation from the
manufacturer.
• This device is a stationary device and shall
not be moved when in use.
-TO MINIMIZE THE RISK OF BURNS, FIRE, ELECTRICAL SHOCK OR INJURY
TOUCH SCREEN
TREATMENT
CONTROL SETTINGS
INFORMATION
ABOUT THE TREATMENT
TREATMENT CHAMBER
ERGOLIFT HEAD
FLAPS

8 9
ATTENTION: KEEP THESE INSTRUCTIONS
Your device should be used on clean, healthy skin. It is important to read and observe the
following precautions and contraindications before using your device.
• Never touch the client and the device’s
unprotected cables or connectors
simultaneously.
• Never use the auxiliary adapter as a
treatment head or allow it to come into
direct contact with the skin.
• Only use treatment heads supplied with your
unit or recommended by LPG.
• LPG Systems will not be liable for any
inappropriate use of the equipment.
• The operator must be particularly attentive
to the sensations felt by the person
undergoing treatment.
• The operator must ensure that the
parameters (intensity, sequentiality,
differential...) are always adapted to the
cutaneous tissue being cared.
• Do not use the USB and ethernet
connections during treatment.
• Do not operate the unit in unsuitable
environmental conditions (see technical
specifications).
• The MOBILIFT M6
Beauty
uses RF energy
only for its internal function. Therefore its
RF emissions are very low and not likely to
cause any interference to nearby electronic
equipment.
• The MOBILIFT M6
Beauty
is suitable for use
in all establishments including domestic and
those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
• The MOBILIFT M6
Beauty
must be used on
a wooden, concrete or ceramic flooring. If
the floor covering is of synthetic material the
relative humidity must be at least 30%.
• The MOBILIFT M6
Beauty
must be connected
to an electrical supply equivalent in quality to
typical commercial or hospital supplies.
• If the operater of the MOBILIFT M6
Beauty
requires continuous function during power
supply network failure, it is recommended
that an alternative failsafe electrical supply
or battery power supply be provided.
• Portable or mobile RF communication
devices must not be used in close proximity
to the MOBILIFT M6
Beauty
or its cables.
• Interference may be encountered in
proximity to devices marked with the
following symbol:
> PRECAUTIONS OF USE
PRECAUTIONS OF USE
3
PRECAUTIONS OF USE
3
This unit contains programs to help the operator obtain the best anticipated results for each case undergoing
care. Under no circumstances may these programs be construed as a guarantee of successful care, which
varies, depending on the morphology, physiology and eating behaviour of each client.
≥ATTENTION
> CONTRAINDICATIONS
• Do not treat open wounds, the eyes,
intra-cavity areas, mucous membranes, the
genitals or breasts.
• Do not treat a client with an infectious
disease, a growing tumour, a phlebitis, a
wound, or an infected area.
• Do not treat a client with skin cancer, a
visible tumor, or other cancerous lesions.
Consult with a doctor in cases where the
client has a case history of tumours or is in
remission.
• Do not treat people with circulatory problems
without first consulting their doctor.
• Do not treat any swollen or inflamed
areas without seeking medical advice and
without having had training in specific LPG
techniques in this particular area.
• Do not treat client after basic medical care
without medical advice and LPG technical
training for the affected areas.
• Stop treatment immediately if the client
experiences pain and consult a doctor.
• This device should not be used on skin
rashes, herpes, inflamed or infected acne, or
vitiligo.
• To avoid bruising, exercise caution when
determining a client’s level of sensitivity and
avoid use on clients taking anti-coagulant
drugs.
• For a more detailed list of the indications
and contra-indications of ENDERMOLOGIE,
please refer to the training manuals.
• As this list is not exhaustive, always seek
medical advice in the event of doubt.
• Because of the risk of interference, it is
important that the professional ensures
the client is not equipped with a personal
medical device such as a pacemaker.
If this is the case, the professional should
obtain details about the device in question to
ensure that any interference will not affect
the correct use of the equipment.
> ELECTROMAGNETIC COMPATIBILTY

10 11
Do not use corrosive products such as acetone, trichloroethylene, or rubbing alcohol.
≥ATTENTION
MAINTENANCE
4 > IDENTIFICATION RATING PLATE
MAINTENANCE
4
> CLEANING THE UNIT
Your unit is identified by a serial number
shown on the rating plate.
The rating plate also shows the permitted
supply voltage for the unit.
If you need to contact LPG Systems because
of a technical problem, please indicate the
serial number of your MOBILIFT M6 Beauty.
This serial number provides information
on the year and month of manufacture
of your unit.
The letter indicates the year the device was manufactured. Z=2009, A=2010, B=2011, etc. The two
digits indicate the production month: 01=January; 02=February; 03=March; etc.
serial
number
voltage, frequency
and power
It is recommended that you clean your unit as often as possible, not only for reasons
of hygiene and aesthetics but also because cleaning the unit will help keep it in a good
working order and extend its useful life.
Using a moist sponge, clean the following sections:
• All the external hoods.
• Hoses.
• The electrical supply cable.
Using a cloth soaked with a small amount of alcohol-free domestic cleaning product,
clean the following sections:
• Control screen and control panel.
This icon indicates that the unit was
sold after August 13, 2006. In conformity
with the 2002/96/CE directive, it cannot
be thrown away with standard
household waste but must be disposed
of by means of recycling. When your
device reaches the end of its useful life,
it must be brought to an appropriate
recycling center or returned to your
dealer. By doing so, you help the
environment by contributing to the
conservation of natural resources and
the protection of human health.
This icon indicates that some specific
warnings or precautions associated with
this device are not on the label.
This symbol means always consult the
accompanying documents before using
your device.
This symbol means store the device
somewhere protected from inclement
weather.
This symbol indicates temperature
limits.
This symbol indicates relative humidity
limits.
This symbol means
« Danger: High Voltage » (US only)
Identification rating plates may vary. The approved one is one on your machine.
≥ATTENTION

1312
FIG. 2FIG. 1
FIG. 3 FIG. 4 FIG. 5
> INSTRUCTIONS FOR CONNECTING THE HOSE
• To connect the hose, proceed as indicated below:
• Position the locking ring in the locked position (fig. 1).
• Position the end of the hose so that the hose key is lined-up with the key slot of the treatment
head connection (fig. 2).
• Push the hose into the treatment head connection until it clicks into place.
FIG. 1 UNLOCKED LOCKED
FIG. 6 FIG. 7
FIG. 2
HOSE RETAINER KEY
LOCKING RING ORNAMENTAL RING
MAINTENANCE
4 > REPLACING THE FILTER CARTRIDGE
Your device contains a filter cartridge.
These components guarantee the efficiency of your unit and prolong its useful life.
Ensure that they are changed as soon as the command screen displays one of these
messages (fig 1-2) :
Access the ‘filter change’ menu as follows :
Select the ‘maintenance’ menu by pressing the icon indicated (fig. 3).
Select the ‘filter’ menu by pressing the icon indicated (fig. 4).
The ‘filter change’ screen indicates which filter requires to be changed (fig. 5).
ICON INDICATING A FILTER
CHANGE IS REQUIRED (FIG. 2).
WARNING MESSAGE
Never use your device without a filter.
Turn the device off before replacing your filter cartridge.
≥ATTENTION
> REPLACING THE FILTER CARTRIDGE (CONT’D)
MAINTENANCE
4
Replace the filter cartridge with a new one. (fig. 6)
Once the filter cartridge is replaced, the filter meter should be reset by pressing the icon as
indicated : (fig. 7).

1514
≥
MAINTENANCE
4
FIG. 3 FIG. 4
> TRANSPORTATION/MOVING
• Remove the accessories holder (fig. 1).
• Carry the device by the handle (fig. 2).
FIG. 2
> INSTRUCTIONS FOR DISCONNECTING THE HOSE
• To disconnect the hose, follow the same steps in the reverse order: Unlock by turning the locking
ring (fig. 3). Lift the locking ring, then remove the hose gently by pulling on the white ring (fig. 4).
MAINTENANCE
4
• Replacing flaps: once the flaps no longer treat the skin properly, they must be replaced. They
should be replaced after every 20 hours.
• A personal Endermolift Kit for use by the client is available for sale.
> REPLACING THE POWER CORD
> MAINTENANCE LOG SHEET
If the power cord of your device is damaged, please contact LPG Systems Customer Service
for a replacement.
LPG Systems Customer Service:
+33 (0)4 75 78 69 89
WHITE RING
FIG. 1
When transporting your Mobilift M6 Beauty, you should:
Never carry the device by the accessories holder.
ATTENTION
DATE NO. OF HOURS OPERATIONS PERFORMED

1716
CONNECTION
ERGOLIFT HEAD
5 > ERGOLIFT HEAD DESCRIPTION
ERGOLIFT HEAD
5
> ERGOLIFT CHAMBER
Only the Lift 20 and Lift 10 chambers can be connected to the Ergolift head.
They can be connected and disconnected by a simple push-pull action.
Lift 20
chamber
with removable
Lift 20 flaps
Lift 10
chamber
with removable
Lift 10 flaps
HOSE
HOSE CONNECTION
ERGOLIFT CHAMBER
TREATMENT CHAMBERS
CONNECTION

1918
ERGOLIFT HEAD
5 > CLEANING THE ERGOLIFT HEAD
AND ERGOLIFT CHAMBERS
FIG. 1 REMOVAL TOOLS
FIG. 2
90° ROTATION
For reasons of hygiene, the treatment heads should be cleaned after each use, using
antiseptic wipes impregnated with a bactericidal and fungicidal solution.
Special attention must be given to the cleanliness of the parts that are in contact with the
client.
Before each use, clean the flaps and Ergolift chamber:
1. Disconnect the chamber from the Ergolift treatment head. (fig. 1)
2. Remove the flaps thanks to the dedicated tool. (fig.2)
3. Thoroughly clean the Ergolift chamber, the flaps and the tool for at least one minute
with the wipes described here below. (fig.3)
4. Put the flaps back in the Ergolift chamber by following the same steps in the reverse
order (fig.4)
ERGOLIFT HEAD
5
> CLEANING THE ERGOLIFT HEAD
AND ERGOLIFT CHAMBERS (CONT’D)
FIG. 3
90° ROTATION
FIG. 4
The use of aggressive products, such as acetone, trichloroethylene, or alcohol, and abrasive sponges, ultrasound
or UV lamps is strictly prohibited. All cleaned and/or disinfected heads should be placed in the storage drawer to avoid
any confusion.
≥ATTENTION

2120
> WHAT IF I’VE PROBLEM?
TROUBLE SHOOTING
6
TECHNICAL SPECIFICATIONS
7
> TECHNICAL SPECIFICATIONS
Dimensions; length x width x height: ...................................65 x 26 x31 cm
Net weight: ..............................................................10,5 kg
Maximum set depression: .........................................35 kPa (350 mbar)
Cooling: ..........................by mechanical ventilation incorporated into the pump
Protection index: ............................................................ IP20
Electrical protection class: ......................................................1
Operating temperature: ............................................. +10°C to +30°C
Storage temperature: ............................................... -20°C to +70°C
Electrical features: .......................................100-240V / 50-60Hz / 80 VA
Operating environment:
Ambient temperature: ............................... 10 to 30° C for normal operation.
Ambient relative humidity: ............................ 30 to 75% without condensation.
Atmospheric pressure: ..........................no significant influence for operation.
Max altitude: ............................................................ 2,500 m
Unit fitted with patented treatment heads.
Device designed for uninterrupted use.
The Mobilift M6 Beauty is labeled in accordance with legislation and standards set
forth in directives EMC 2004/108/EC and low voltage 2006/95/EC.
• Is the unit properly connected to a mains plug?
• Is the mains plug live?
• Is the ON switch lit up?
• Are the filter cartridges clean and correctly placed?
• Is the hose properly connected?
•Is the hose clogged?
• Are the care chamber and flaps properly connected?
Once these checks have been carried out and if the malfunction persists, please contact
LPG Systems Customer Service or the nearest authorized dealer, indicating the model of
your unit and its serial number.
If your unit is not working properly, proceed with the following checks before calling
Customer Service:
LPG Systems Customer Service:
+33 (0)4 75 78 69 89

2322
WARRANTY
8 > GENERAL WARRANTY CONDITIONS
You have recently acquired an appliance
distributed by LPG Systems or an LPG
Systems approved distributor. It is the
purchaser/user’s responsibility to find out
from the local authorities the conditions and
professional qualifications that should be met
before using the appliance.
The purchase of this equipment implies
the legal acceptance by the purchaser/
professional user of these general warranty
conditions. If the appliance was sold to you
by an approved LPG Systems distributor, the
purchaser/user should refer to the supplier’s
warranty conditions. These may in no way
increase the undertakings made in these
present warranty conditions. The warranty
can only be implemented and is only valid if
the warranty slip has been duly filled out and
returned to LPG Systems within two weeks of
delivery, irrespective of the country. Warranty
slips that are only partially completed will be
rejected.
The appliance is guaranteed against
manufacturing flaws and defects in the raw
materials. The warranty extends for the
shorter of the following two periods: one (1)
year OR two thousand (2000) hours of use
from the invoice date. During this period, LPG
Systems undertakes to exchange or repair
free of charge, as quickly as possible, any part
that LPG Systems acknowledges as defective,
however LPG Systems does not undertake
to replace the entire appliance. Travelling
and living expenses for our technicians and
transportation costs of the appliance or parts
to and from the after-sales service workshop
are not covered by this warranty.
Replacements and repairs performed within
this warranty, with or without immobilization
of the equipment, shall not have the effect of
extending the warranty period. Replaced parts
become the property of LPG Systems or the
approved distributor. No compensation shall
be paid for loss of use. The purchaser/user is
required to allow LPG Systems the necessary
time and means to carry out all repairs and
deliveries of replacement spare parts. Failure
to comply will absolve us from our obligations
under the warranty.
The only courts of competent jurisdiction are
those within the legal district of LPG Systems’
head office regardless of any jurisdiction
clauses in any other document.
WARRANTY
8
> GENERAL WARRANTY CONDITIONS (CONT’D)
Warranty exclusions:
• Damage occurring during transportation.
Transportation of this equipment and/or
spare parts is at the recipient’s own risk.
Before taking delivery, it is the recipient’s
responsibility to verify the state of the goods
and to make a claim against the transport
company in the manner usual in the delivery
country.
• Non-observance of the installation and
operating instructions, failure to carry
out maintenance and/or negligence in
maintaining the appliance and/or its filter
cartridges, connection to a faulty electric
supply or a non-earthed electric supply or
a power supply whose voltage is different to
the one indicated on the appliance.
• If an appliance is sold before the end of the
warranty period, the warranty is transferred
to the purchaser for the remaining warranty
period, on the condition that:
i) the original invoice is provided.
ii) that the initial vendor is informed of
the sale.
• Modification, mounting of accessories or
dismantling of the equipment.
• Any operation and/or intervention not
specified in the LPG Systems Operating
Instructions and performed on the
equipment by the purchaser/user and/or any
party not approved by LPG Systems.
• Use of consumables, spare parts,
inappropriate components or parts not
supplied by LPG Systems.
• Blockage of the appliance through aspiration
of a foreign body.
• Normal wear and tear of any of the
equipment’s parts.
• Falls, impacts, lightning, fire, force majeure,
water damage and natural disasters.

2524
WARRANTY
8
WARRANTY
8
> LIMITATION AND EXONERATION OF LIABILITY
Non-compliance with the general guarantee conditions during the term of the warranty and
following expiry of the latter, may lead to LPG Systems being exonerated of any liability in
the event of damage attributed to the products supplied.
LPG Systems will not be held liable in respect of equipment/material losses or physical
accidents (i), following installation that does not comply with the statutory or regulatory
provisions of the country where the unit is to be used or (ii) which could jeopardize a
MOBILIFT M6
Beauty
unit, which was the subject of intervention work not scheduled in
the LPG Systems operator’s manual and/or carried out by the operator or a third party not
authorized by LPG Systems.
LPG Systems will be absolved of its liability in the event of use exceeding the level of
professional competence/qualification of the operator or resulting from the incorrect use of
the unit.
The personal data collected in this form are sent to LPG Systems, which is responsible for processing it in order to manage customers relations, including
after-sales service and warranties. Information with an asterisk must be provided in order for this form to be processed. According to modified French law
78-17 from January 6, 1978 relating to computers, digital files, and freedoms, also known as the “Informatique et Libertés” law, you have a right of access,
rectify, and remove personal data about you by contacting LPG Systems in writing at the following address: LPG SYSTEMS – 30, rue Dr Abel – Technoparc
de la Plaine - CS 90035 - 26902 VALENCE (FRANCE).
> WARRANTY ACTIVATION
> OR FILL IN THE FORM AND RETURN
TO LPG SYSTEMS
You can activate your warranty online by connecting to our
warranty webpage:
http://warranty.lpgsystems.com
Name*: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Address*: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Country*: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tel*: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Email: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Type of business: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Profession: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Type of unit: MOBILIFT M6 Beauty
Serial number*: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Date: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
≥ATTENTION
To validate the warranty, detach this slip and return to the address below within 15 days following delivery of the unit:
LPG SYSTEMS SA - 30, rue du Dr Abel
Technoparc de la Plaine - CS 90035 - 26902 VALENCE Cedex 9 – France

HEADQUARTERS LPG SYSTEMS S.A.
TECHNOPARC DE LA PLAINE
30, RUE DU DR. ABEL - CS 900 35 - 26902 VALENCE CEDEX 09 - FRANCE
TEL.: +33 (0)4 75 78 69 00 - FAX: +33 (0)4 75 42 80 85
INTERNATIONAL/MARKETING
2753, ROUTE DES DOLINES - BP 243
06905 SOPHIA-ANTIPOLIS CEDEX - FRANCE
TEL.: +33 (0)4 92 38 39 00 - FAX: +33 (0)4 92 96 09 65