LW Scientific i4 User manual

Camera Attachments
A
K
B
P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046
P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046
Instruction Manual
E Brightness Control
F Focus Friction Control
G Coarse Focus
H Fine Focus
I On/Off Switch
(located on rear)
J X/Y Axis Stage Controls
K Base Condenser
L Filter Holder
M Substage Iris Diaphragm
N Substage Abbe Condenser
O Stage
P Slide Holder
Q Objectives
R Nosepiece
S Eyepieces
A Diopter Adjustment
B Binocular Head
C Head Retention Screw
D Stage Stop Lever
(located on left side)
Specifications
Maintenance
Ask your
authorized LWS
dealer about
additional
accessories
Upgrades:
Recommended
i4 Infinity
i4 Infinity
Introduction
The LW Scientific I-4 microscope features exceptional optical quality with an infinity optical system, like found on the
most expensive laboratory microscopes, but at a fraction of the price. The ergonomic narrow design allows users to
rest their arms flat on the table and easily operate all controls. The binocular eyetubes can also be rotated upward to
achieve an additional 2 inches of height for taller users. LED lighting produces “daylight” color, with a cool temperature
and long life. Comfort, durability, dependability, and superior imaging make the I-4 a great value for labs, physicians,
vets, and universities. The I-4 microscope is intended for use as a biological microscope in a professional environment in
accordance with the guidelines set forth in this operations manual.
2
4
5
3
1Always cover your microscope with the dust cover when not in use. When cleaning the lenses, use lens paper or a
cotton swab dipped in lens cleaning solution.
Excess oil should be cleaned off your 100x objective and stage at once. An alcohol pad is best for removing oil from
the stage and on the other metal parts, but is not recommended for use on the lenses. Use lens cleaning solution and
lens paper to clean off your objectives.
Dust in the nosepiece or ocular tubes should be blown out using filtered air. Canned air dusters work well for this job.
Whenever you remove an objective, we recommend that you place it back into the original plastic shipping vial until
ready to be placed back on the microscope. SCREW THE OBJECTIVE SECURELY INTO THE CAP OF THE HOLDER - DO NOT
DROP OBJECTIVE LOOSELY INTO CONTAINER.
To keep your microscope in top condition for years, LW Scientific recommends that you have the microscope
professionally serviced once a year.
Warning: The 40x objective is not sealed for oil immersion. Damage to the 40x objective due to oil
immersion is not covered under warranty.
Head
Binocular (Seidentopf)
Inclined 30°, rotates 360°
Diopter adjustment
10X/18 wide-field eyepieces
10X/20 High-Point super-wide eyepieces available
Trinocular available
Nosepiece
Reverse quadruple hole multiple ball bearing
Illumination
Moveable Abbe Condenser (NA 1.25)
Iris Diaphragm
Blue and green filters
LED light
Construction
Rugged alloy
Stage
Mechanical stage (140 mm x 130 mm)
– coaxial drive controls
Range of traverse is 78 mm x 50 mm
Acid and reagent resistant finish
Objectives
Infinity Semi-Plan, Infinity Plan Achromat Objectives
4X N.A. 0.10
10X N.A. 0.25
40XR N.A. 0.65
100XR N.A. 1.25 (oil immersion)
50X oil Plan Infinity available
Adjustment Controls
Eyepiece: Interpupillary distance adjustment 48-75 mm
Stage Controls: Knobs allow movement of slide on X-Y axis
Etched vernier scales
Coarse Adjustment: Range of 30 mm
Fine Adjustment: Graduation of 2μm
Variable Light Adjustment
Dimensions and Weight
Weight: 16 lbs / 7.3 kg
Height: 14.1" / 267 mm
Length: 10.7" / 267.5 mm
Width: 6.1" / 152.5 mm
Shipping weight: 17 lbs/ 8kg
Size N.A. Mag. Field of View
4X 0.10 40X 4.5mm
10X 0.25 100X 1.8mm
20X 0.40 200X 0.9mm
40XR 0.65 400X 0.45mm
50XR 0.95 0.36mm500X
60XR 0.85 0.3mm600X
100XR 1.25 0.18mm1000X
Objectives: The following numbers are
based on use with 10x/18 eyepieces.
I
D
C
G
H
F
J
E
L
M
N
O
S
PQ
R
Lumin
Epi-Module Objectives
Phase and Dark Field
Operating Environment
Indoor Use Only
Ambient temperature: 5˚ to 40˚C (41˚ to 104˚F)
MKT-7.5.3-L-127 | Rev 2
P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046
Unpacking and Setup
Lamp Replacement
Assembly
Operation
LW Scientific packs each I-4 Microscope with utmost care. Examine the outer
and inner containers for any visual damage. Retain all of the packing
material until you have examined and tested your new microscope. If there
is damage, please contact the shipping company, as our warranty does not
cover shipping damage. If you are uncertain who the shipper is, contact the distributor from whom you purchased the
microscope. Please retain all packaging material for future use. Carefully unpack** your I-4 Microscope using the
following checklist for all the parts and accessories:
2
4
5
3
1Remove the body of the microscope and place it on a sturdy, dust-free surface. Remove the plastic plugs in the
nosepiece. Install the objectives in such a way that when you turn the nosepiece clockwise, you are moving from the
4x, to 10x to 40x and finally the 100x objective.
Remove the microscope head from the Styrofoam carton and pull off the protective covers from the eyepiece tubes
and head mount. Insert the head mount into the upper arm of the body, taking notice that the “groove” in the flange
beneath the head must line up with a “tab” in the mounting ring on the scope (turn the head toward the side to line
up). The “groove” system is a safety feature to keep the head from falling off if the retention screw is loosened. Once
the head seats, then tighten the head retention screw to secure the head in place. Note: Do not over-tighten.
Remove plastic covers and packing material covering the stage, the condenser and lower light assembly
Insert the 10x eyepieces.
Attach the power adapter.
2
3
4
5
6
7
8
9
10
11
1Once you have assembled all the parts and allowed your microscope to come to room temperature, plug the power
adapter into the microscope and then into the appropriate AC outlet. Note: excess cold can fog lenses and cause lamp to
fail.
Turn the light on using the black on/off switch on the rear of the scope. Next adjust the light intensity using the brightness
control wheel located on the right side of the scope.
In order to become acquainted with the controls, choose a specimen slide with which you are familiar. For example, an old
hematology slide or a commercially prepared slide. Place the slide into the slide holder by pushing back on the thumb
guard to open the slide finger. The slide finger closes slowly to eliminate the possibility of chipping the corner of your slide
when it closes.
Move the slide to the center of the stage, by turning the stage control knobs, located just below the stage. These knobs
allow you to move the slide on the X-Y axis (forward/backward and left/right).
The sub-stage iris should then be set to match the aperture of the objective for maximum resolution under each objective
power. There are numbers on the iris ring to show the correct setting for each objective power. You should begin with the 4x
or 10x objective. Only use the iris wide open when under the 100x oil objective. Closing down the iris on smaller objective
powers will improve resolution, contrast, and depth of field.
Place the filter of your choice onto the lower light assembly. Note that many customers prefer to use the blue filter for
routine use, or no filter at all.
Once you are comfortably seated, look into the oculars and move the eyepiece tubes together or apart until you see only
one complete circle of light. You have now adjusted your interpupillary distance. The I-4 binocular eyetubes can also be
rotated completely from low position to top position, which raises the eyepieces nearly 2 inches higher for tall users.
Using the 4x or 10x objectives and the coarse and fine adjustment knobs, bring the specimen into focus. Now, move the 40x
objective into place. You will feel a “clicking” action when the objective is seated properly. Again, adjust focus for best
image. You should also adjust the iris diaphragm (as described in step 5) for the best contrast and resolution.
Diopter Adjustment: Since you are using a binocular microscope, you have to adjust the normal difference in vision
between your two eyes. This is a simple but critical adjustment! Close your left eye and look into the right ocular with
your right eye. Adjust the focus to give you the best image. Now look at the ocular tube on the left. You will see that
the left ocular tube has a built-in adjustment ring. Now close your right eye and look with your left eye into the left
ocular. Using the diopter adjustment ring on the left ocular tube, adjust the focus until you see a clear, focused field.
Friction Adjustment: With repeated use and wear, the stage may drift downward out of focus. If this happens, you need only
to tighten the friction control ring (located on the right side of the microscope between the coarse adjustment and the
body of the microscope). If the coarse focus is hard to turn, you may choose to loosen the friction adjustment. There is a
black plastic friction wrench in your packaging that will engage the friction control ring to help you turn it.
Stage Stop Lever: To help prevent the stage from hitting the objectives, the I-4 Microscope is equipped with an adjustable
stage stop. Rotate the 100X oil objective into place, and put a slide into the slide holder. Slowly raise the stage, stopping
when the slide makes contact with the objective. Now, turn the stage stop lever in a clockwise direction toward you to lock
the stage from going any higher. The stage stop lever is located on the left side of the microscope between the coarse
adjustment and the body of the microscope.
Power
1 - Microscope body with Abbe condenser
2 - 10x eyepieces
1 - Binocular head (Seidentopf style)
1 - Friction adjustment wrench (for coarse focus)
2 - Filters (blue & green)
4 - Objectives 4X, 10XR, 40XR, 100XR (oil)
1 - 3 prong power cord
1 - Warranty card
1 - Immersion oil
1 - Dust cover
**Note - Some parts may be packed in the
outer recesses of the styrofoam blocks
Ensure that the power switch is in the OFF position and that the power adapter is removed. Remove the magnetic base
condenser to expose the LED light. Unscrew the black top to remove protective glass cover and set aside. Raise the
stage as high as possible to allow better access to the LED assembly. *BE CAREFUL WHEN RAISING THE STAGE NOT TO
DAMAGE THE OBJECTIVE. Next, take a small Philip’s Head screwdriver and remove the 2 screws that secure the LED
light.
Set them aside in a secure place. Using a soldering iron, gently melt the solder on the positive and negative
connections and detach the wires. Note which color wire correspnds to the positive and negative wires. Gently use a
small flat head screwdriver to remove the LED disk from the heat conductive tape. Once removed, discard the old LED
light. Take the new LED light and place it onto the adhesive tape, just as the old LED light was placed. Be sure to keep
the wire connections exposed as they will need to be connected to the new LED light.
Once the new LED light is secured to the tape and centered, solder the two wires back to the corresponding positive
and negative contact pads on the LED light. Use the two small Philip’s Head screws to secure in place. Allow 5 minutes
for the solder to set. Plug the power adapter into the back of the microscope, place the base condenser back on the
microscope, and turn the microscope on to test the LED light connection. If connection is good, screw the protective
glass cover back onto the LED assembly, and lastly, place the magnetic base condenser back into position.
If you suspect faulty electronics, call LW Scientific’s technical service department at 800-726-7345.
Power:
Input: AC 100-240V / 50-60 Hz
Lamp: S-LED W1
Fuse: T250V 500mA
AB C NOTE:
A: If using a trinocular head, pull out knob when
using a camera.
B: i4 Trinocular C-mount tube.
C: i4 Trinocular eyepiece tube.

Other LW Scientific Microscope manuals

LW Scientific Z4 User manual

LW Scientific

LW Scientific Z4 User manual

LW Scientific OB2MF User manual

LW Scientific

LW Scientific OB2MF User manual

LW Scientific M1-BHP4 User manual

LW Scientific

LW Scientific M1-BHP4 User manual

LW Scientific M5 LabScope User manual

LW Scientific

LW Scientific M5 LabScope User manual

LW Scientific Student Scope OEM-SF3 User manual

LW Scientific

LW Scientific Student Scope OEM-SF3 User manual

LW Scientific Advanced Scope OEM-SAF3 User manual

LW Scientific

LW Scientific Advanced Scope OEM-SAF3 User manual

LW Scientific M2-EP User manual

LW Scientific

LW Scientific M2-EP User manual

LW Scientific M2-P-SE User manual

LW Scientific

LW Scientific M2-P-SE User manual

LW Scientific OB2MF User manual

LW Scientific

LW Scientific OB2MF User manual

LW Scientific Z2BI User manual

LW Scientific

LW Scientific Z2BI User manual

LW Scientific Revelation III LED Owner's manual

LW Scientific

LW Scientific Revelation III LED Owner's manual

LW Scientific Revelation III LED User manual

LW Scientific

LW Scientific Revelation III LED User manual

LW Scientific i5 Series User manual

LW Scientific

LW Scientific i5 Series User manual

LW Scientific Vision VIS12 User manual

LW Scientific

LW Scientific Vision VIS12 User manual

LW Scientific Advanced User manual

LW Scientific

LW Scientific Advanced User manual

LW Scientific LW-101 User manual

LW Scientific

LW Scientific LW-101 User manual

LW Scientific Mi5 User manual

LW Scientific

LW Scientific Mi5 User manual

LW Scientific Mi5 User manual

LW Scientific

LW Scientific Mi5 User manual

LW Scientific PFA-12 User manual

LW Scientific

LW Scientific PFA-12 User manual

LW Scientific MiniVID WiFi User manual

LW Scientific

LW Scientific MiniVID WiFi User manual

LW Scientific Revelation III LED User manual

LW Scientific

LW Scientific Revelation III LED User manual

LW Scientific VFA-12 User manual

LW Scientific

LW Scientific VFA-12 User manual

LW Scientific RV30A-DH User manual

LW Scientific

LW Scientific RV30A-DH User manual

LW Scientific OB2MF Owner's manual

LW Scientific

LW Scientific OB2MF Owner's manual

Popular Microscope manuals by other brands

Fein Optic RB20-HD user manual

Fein Optic

Fein Optic RB20-HD user manual

Olympus BX-URA2 instruction manual

Olympus

Olympus BX-URA2 instruction manual

Etaluma LS560 Startup guide

Etaluma

Etaluma LS560 Startup guide

Olympus DP74 instructions

Olympus

Olympus DP74 instructions

Olympus CHA-P instruction manual

Olympus

Olympus CHA-P instruction manual

OPTO-EDU MDJ A13.2602 instruction manual

OPTO-EDU

OPTO-EDU MDJ A13.2602 instruction manual

PCE Instruments PCE-LCM 50 user manual

PCE Instruments

PCE Instruments PCE-LCM 50 user manual

Tecnai T12 Operating procedures

Tecnai

Tecnai T12 Operating procedures

ThermoFisher Scientific Continuµm manual

ThermoFisher Scientific

ThermoFisher Scientific Continuµm manual

National Geographic 80-10201 instruction manual

National Geographic

National Geographic 80-10201 instruction manual

Hitachi McDATA Sphereon 4500 instruction manual

Hitachi

Hitachi McDATA Sphereon 4500 instruction manual

Ken A Vision Comprehensive T-1901C instruction manual

Ken A Vision

Ken A Vision Comprehensive T-1901C instruction manual

Stalwart STM-3040 Series instruction manual

Stalwart

Stalwart STM-3040 Series instruction manual

National DC5-420TH Instructions for use

National

National DC5-420TH Instructions for use

Bresser ADL-601F instruction manual

Bresser

Bresser ADL-601F instruction manual

Celestron kids Quick setup guide

Celestron

Celestron kids Quick setup guide

Omano OM139 instruction manual

Omano

Omano OM139 instruction manual

National 106 instructions

National

National 106 instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.