R9EWTU
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-6224 Rev.9 P. 3 / 21
■OPERATING MODE SETTING
(*) Factory setting
• System Conguration (SW1-1, 2)
SW1-1 SW1-2 SYSTEM CONFIGURATION
OFF OFF Three-phase / 3-wire (*)
ON OFF Single-phase / 2-wire
OFF ON Single-phase / 3-wire
ON ON ---
• Balanced or Unbalanced Load (SW1-3)
SW1-3 BALANCED / UNBALANCED
OFF Unbalanced (*)
ON Balanced
• Clamp Sensor Type (SW1-4, 5, 6)
SW1-4 SW1-5 SW1-6 CLAMP SENSOR TYPE
OFF OFF OFF CLSE-R5 (5A) (*)
ON OFF OFF CLSE-05 (50A)
OFF ON OFF CLSE-10 (100A)
OFF OFF ON CLSE-20 (200A)
ON ON OFF CLSE-40 (400A)
OFF ON ON CLSE-60 (600A)
• Conguration Mode (SW1-8)
SW1-8 CONFIGURATION MODE
OFF DIP switch setting (*)
(PC Configurator setting is invalid.)
ON PC Configurator and communication
(DIP switch setting is invalid.)
Note 1: Turn on the power supply to the unit after setting
operating mode.
Note 2: Be sure to set unused SW1-7 to OFF.
Note 3: Clamp sensor type setting is common to all circuits.
Note 4: The sensor type and other settings for individual
circuit can be performed on the PC Configurator or via
Ethernet.
■TERMINAL ASSIGNMENTS
5
14 13
1L
1K 3K
4
3L
321
12 11 10
1L 3L 1L
1K 3K
9
18 17
T2
T1 3K
8
3L
76
16 15
6ch 6ch7ch 5ch 5ch8ch 7ch8ch
6ch 6ch7ch 5ch 5ch8ch 7ch8ch
1L 3L
1K 3K
1K
1 5ch 1L Ch.5, Current input 1L 10 5ch 1K Ch.5, Current input 1K
2 5ch 3L Ch.5, Current input 3L 11 5ch 3K Ch.5, Current input 3K
3 6ch 1L Ch.6, Current input 1L 12 6ch 1K Ch.6, Current input 1K
4 6ch 3L Ch.6, Current input 3L 13 6ch 3K Ch.6, Current input 3K
5 7ch 1L Ch.7, Current input 1L 14 7ch 1K Ch.7, Current input 1K
6 7ch 3L Ch.7, Current input 3L 15 7ch 3K Ch.7, Current input 3K
7 8ch 1L Ch.8, Current input 1L 16 8ch 1K Ch.8, Current input 1K
8 8ch 3L Ch.8, Current input 3L 17 8ch 3K Ch.8, Current input 3K
9 T2 Unused 18 T1 Unused
1 NC Unused 10 NC Unused
2 1ch 1L Ch.1, Current input 1L 11 1ch 1K Ch.1, Current input 1K
3 1ch 3L Ch.1, Current input 3L 12 1ch 3K Ch.1, Current input 3K
4 2ch 1L Ch.2, Current input 1L 13 2ch 1K Ch.2, Current input 1K
5 2ch 3L Ch.2, Current input 3L 14 2ch 3K Ch.2, Current input 3K
6 3ch 1L Ch.3, Current input 1L 15 3ch 1K Ch.3, Current input 1K
7 3ch 3L Ch.3, Current input 3L 16 3ch 3K Ch.3, Current input 3K
8 4ch 1L Ch.4, Current input 1L 17 4ch 1K Ch.4, Current input 1K
9 4ch 3L Ch.4, Current input 3L 18 4ch 3K Ch.4, Current input 3K
14
56
3K
3L 1L
15
1K
16 17 18
789
3K 1K 3K
3L 1L
10
12
NC
NC 1L
11
1K
12 13
34
3ch 3ch2ch 4ch 4ch1ch 2ch1ch
3ch 3ch2ch 4ch 4ch1ch 2ch1ch
3K 1K
3L 1L 3L
• Current Input 2
4
1
P1
U(+)
5
P2
6
23
P3
V(−)
FE1
7
N
No. ID FUNCTION
1 U(+) Auxiliary power
(+)
2 V(−) Auxiliary power
(−)
3 FE1 Power ground
4 P1 Voltage input P1
5 P2 Voltage input P2
6 P3 Voltage input P3
7 N Voltage input N
• Auxiliary Power Supply, Voltage Input
• Current Input 1
No. ID FUNCTION No. ID FUNCTION
No. ID FUNCTION No. ID FUNCTION