M-system 47LPT User manual

47LPT
P. 1 / 16EM-9506-B Rev.6
VT INPUT DIGITAL PANEL METER
(4 1/2 digit, LED display type, true RMS sensing) MODEL 47LPT
OPERATING MANUAL
BEFORE USE ....
Thank you for choosing M-System. Before use, please check
contents of the package you received as outlined below.
If you have any problems or questions with the product,
please contact M-System’s Sales Office or representatives.
■PACKAGE INCLUDES:
Digital panel meter ...................................................... (1)
Engineering unit sticker label sheet ........................... (1)
■MODEL NO.
Confirm Model No. marking on the product to be exactly
what you ordered.
■OPERATING MANUAL
This manual describes necessary points of caution when
you use this product, including installation, connection,
basic maintenance procedures and detailed operations.
POINTS OF CAUTION
■CONFORMITY WITH EC DIRECTIVES
•ThisequipmentissuitableforuseinaPollutionDegree2
environment and in Measurement Category III, Installa-
tion Category II, with the maximum operating voltage of
300V.
Reinforced insulation is maintained between signal input,
outputandpowerinput.Priortoinstallation,checkthat
the insulation class of this unit satisfies the system re-
quirements.
•Altitudeupto2000meters
•The equipment must be installed such that appropriate
clearance and creepage distances are maintained to con-
form to CE requirements. Failure to observe these re-
quirements may invalidate the CE conformance.
•Inordertoenabletheoperatortoturnoffthepowerinput
immediately, install a switch or a circuit breaker accord-
ing to the relevant requirements in IEC 60947-1 and IEC
60947-3 and properly indicate it.
■POWER INPUT RATING & OPERATIONAL RANGE
Locate the power input rating marked on the product and
confirm its operational range as indicated below:
100–240VACrating:85–264V,50/60Hz,approx.6.5VA
24VDCrating:24V±10%,approx.3W
110VDCrating:85–150V,approx.3W
■GENERAL PRECAUTIONS
•Beforeyouremovetheunitormountit,turnoffthepower
supply and input signal for safety.
•Besuretoputtheterminal coveronwhile thepoweris
supplied.
■ENVIRONMENT
•Indooruse
•Whenheavydustormetalparticlesarepresentintheair,
install the unit inside proper housing with sufficient ven-
tilation.
•Donotinstalltheunitwhereitissubjectedtocontinuous
vibration.Donotsubjecttheunittophysicalimpact.
•Environmentaltemperaturemustbewithin-10to+55°C
(14to131°F)withrelativehumiditywithin30to90%RH
in order to ensure adequate life span and operation.
•Besurethattheventilationslitsarenotcoveredwithca-
bles, etc.
■REQUIREMENTS TO ENSURE IP66
•Observe the designated panel cutout size (W92 × H45
mm).
•Thewatertightpackingincludedintheproductpackage
must be placed behind the front cover.
•Both mounting brackets must be fastened tightly until
they hit the panel.
•Conrmvisuallythatthepackingisnotcontortedorex-
cessively run off the edge after installation.
■WIRING
•Make sure for safety that only qualied personnel per-
form the wiring.
•Do not install cables (power supply, input and output)
close to noise sources (high frequency line, etc.).
•Do not bind these cables together with those in which
noises are present. Do not install them in the same duct.
■AND ....
The unit is designed to function as soon as power is sup-
plied, however, a warm up for 10 minutes is required for sat-
isfying complete performance described in the data sheet.

47LPT
P. 2 / 16EM-9506-B Rev.6
COMPONENT IDENTIFICATION
Specifications
Strap
Watertight Sealing
n TOP VIEW
n FRONT VIEW n SIDE VIEW
Mounting Bracket Terminal Cover
(4) Max/Min Button
(5) Alarm/ Button
(6) Scale/ Button
(7) Shift Button
(8) Up Button
(1) Main Display
(2) Alarm Indicators
(3) Function Indicators
Alarm/ Scale/Max/Min Shift Up
Zro
HH
H
P
L
LL
Spn D/P
Tch Fnc MinMax
• COMPONENT IDENTIFICATION
No. COMPONENT FUNCTION
(1) Main display Indicates present values, setting values.
(2) Alarmindicators*1Indicate alarm status of the input signal.
LL turns on when the LL alarm is tripped.
L turns on when the L alarm is tripped.
HturnsonwhentheHalarmistripped.
HHturnsonwhentheHHalarmistripped.
Pturnsonwhennoneoftheotheralarmsistripped.
(3) Function indicators Indicate setting modes and status.
(4) Max/Minbutton Usedtoswitchthemaindisplaytoshowpresentvalues,maximumvaluesorminimumvalues.
(5) Alarm/ button Usedtomoveontothealarmsettingmode;ortoshiftthroughsettingitemsineachsetting
mode.
(6) Scale/ button Usedtomoveontothescalingsettingmode;ortoshiftthroughsettingitemsineachsetting
mode.
(7) Shiftbutton Usedtomoveontothesettingstandbystatusandshiftthroughdisplaydigitsineachsetting
item.
(8) Upbutton Usedtochangesettingvalues,toexecute/cancelForcedZero,ortosetinputsignalvaluesduring
‘Teach’ calibration.
*1.Onlythe‘P’indicatorturnsonwith‘noalarmoutput’option.Fordualalarmtype,‘LL’or‘HH’doesnotturnon.

47LPT
P. 3 / 16EM-9506-B Rev.6
INSTALLATION
TERMINAL CONNECTIONS
Connect the unit as in the diagram in the following page or refer to the connection diagram on the terminal cover.
■EXTERNAL DIMENSIONS unit: mm (inch)
92
+ 0.8
– 0
45
+ 0.6
– 0
min. 120
min. 75
Panel thickness: 1.6 to 8.0 mm
■PANEL CUTOUT unit: mm ■HOW TO MOUNT THE UNIT ON A PANEL
The watertight packing must be in place to hold the meter.
Do not remove it.
1) Insert the unit into the panel cutout.
2) Push the mounting brackets into the grooves on both
sides of the rear module, until they hit the panel’s rear
side.
Mounting BracketWatertight packing Panel
96 (3.78)
48 (1.89)
103 (4.06)
95 (3.74)
2 (.08)
47 (1.85)
86 (3.39)12.5 (.49)
98.5 (3.88)
■REAR VIEW
• No Alarm Output
■ FRONT VIEW ■SIDE VIEW
■ TOP VIEW
45 (1.77)
91.5 (3.60)
4–M3 SCREW
•†Alarm Output
12345678910 12345678910
11 12 13 14 15 16 17 18 19 20
20–M3 SCREW
TERMINAL
2–M3 SCREW
10–M3 SCREW
TERMINAL

47LPT
P. 4 / 16EM-9506-B Rev.6
■CONNECTION DIAGRAM
■TERMINAL BLOCK
•How to remove the terminal block cover
Insert the minus tip of a screwdriver into each hole at the
four corners of the cover and pull it to the direction as
indicated below to separate the terminal block cover.
9
10
POWER
U(+)
V(–)
LOAD
Inductive
Load (Coil)
Varistor or
CR Circuit
• DC Powered
LOAD
Inductive
Load (Coil)
Diode, Varistor or
CR Circuit
■ Relay Protection
• AC Powered
17
15
16
■ALARM SUFFIX CODE 1: N.O. contact, 4 points
20
18
19
Lc
(
COM
)
Lb
(NC)
La
(NO)
Hc
(
COM
)
Hb
(NC)
Ha
(NO)
17
15
16
20
18
19
Lc, LLc
LLa
La
Hc, HHc
Ha
HHa
LL OUTPUT
L OUTPUT
H OUTPUT
L OUTPUT
H OUTPUT
HH OUTPUT
9
10
POWER
U(+)
V(–)
■ALARM SUFFIX CODE 2: SPDT contact, 2 points
9
10
POWER
U(+)
V(–)
■ALARM SUFFIX CODE 0: No alarm output
+
–
7
8
DC OUTPUT
+
–
7
8
DC OUTPUT
+
–
7
8
DC OUTPUT
Note: The section enclosed by broken line is only with DC output option.
1
3
V
U
v
u
FUSE VT
LOAD
SOURCE
1
3
V
U
v
u
FUSE VT
LOAD
SOURCE
1
3
V
U
v
u
FUSE VT
LOAD
SOURCE
•How to remove the terminal block
The terminal block is separable in two pieces. Loosen two
screws on both sides of the terminal block to separate.
Be sure to turn off the power supply, input signal and
power supply to the output relays before separating the
terminal block.
Terminal Cover
Separable
Terminal Block
Terminal Block
Screw

47LPT
P. 5 / 16EM-9506-B Rev.6
SETTING PROCEDURE
POWER ON
■ GENERAL SETTING FLOWCHART
*1. Optional settings for parameters in a broken-line section.
Alarm Setting Mode is locked with no-alarm-output type.
POWER ON
*1
■INITIAL SETTING FLOWCHART
Initial Setting Complete
Measuring Started
Hold down Scale/ button for 3 seconds or more to move on to Scaling Setting Mode.
Choose an input type and display scale values.
Hold down Alarm/ button for 3 seconds or more to move on to Alarm Setting Mode.
Go through the alarm settings.
Hold down Alarm/ + Scale/ buttons at once for 3 seconds or more
to move on to Advanced Setting Mode. Go through the advanced settings.
Hold down Alarm/ button or Scale/ button for 1 second or more
to go back to Measuring Mode.
If necessary, hold down Max/Min button + Alarm/ buttons at once
for a preset time duration to enter Lockout Setting Mode.
Go through the lockout settings.
Measuring Mode
Hold down Alarm/ or Scale/ button
for ≥1 second
Hold down Scale/ button
for ≥3 seconds
Measuring
Measuring stopped*3
Mode transition
*1. Preset time can be specified with “Transition Time to Lockout Setting Mode” in Advanced Setting Mode.
*2. Alarm Setting Mode is locked with no-alarm-output type.
*3. The last measured values or status are held for DC and alarm outputs while the measuring is stopped.
(Except the analog output adjustments during the loop test and the scaling)
Scaling Setting Mode
Alarm Setting Mode
Advanced Setting Mode
Hold down Alarm/ or Scale/
button for ≥1 second
Hold down Alarm/ or Scale/ button
for ≥1 second
Hold down Max/Min + Alarm/ buttons at once
for a preset time duration*1
Hold down Alarm/ + Scale/ buttons at once
for ≥3 seconds
Lockout Setting Mode
Hold down Alarm/ or Scale/ button
for ≥1 second
Hold down Alarm/ button
for ≥3 seconds
Loop Test Output Mode
Hold down Alarm/ or Scale/ button
for ≥1 second
Hold down Alarm/ + Scale/ + Shift buttons at once
for ≥5 seconds
*2

47LPT
P. 6 / 16EM-9506-B Rev.6
■OPERATIONS IN MEASURING MODE
• Switching the main display to MAX or MIN values
PressMax/MinbuttontoswitchthemaindisplaytoMAXorMINvalues.
‘Max’or‘Min’indicatorturnsonduringtheMAX/MINdisplaymode.
PressMax/Minbuttonagainfor1secondormore,orturnthepowersupplyoffandontocanceltheMAX/MINdisplaymode.
• Forced Zero
PressUpbuttontoshiftthepresentdisplayvalueto0andtocontinuemeasuringinreferencetothispoint.
‘Zro’and‘Fnc’indicatorsturnonduringtheForcedZeromode.
PressUpbuttonagainfor1secondormoretocancelthemode.
ForcedZeroiseffectivelyusedwhen‘0’iswithinthescaledrange.
The command cannot be executed when the operation is prohibited with the lockout protection setting.
MAX/MINdisplaymodecannotbecancelledwhileintheForcedZeromode.
ForcedZerocannotbecancelledwhileintheMAX/MINdisplaymode.
ForcedZerovalueisstoredinmemoryevenwhenthepowerisremoved.
• Conrming alarm setpoints
PressAlarm/ button to confirm alarm setpoints.
Max. Value
Display
Min. Value
Display
Max/Min
button
Max/Min
button
Max/Min
button
ON
Measuring
Mode
Max ON
Min
Alarm LL Alarm L
Alarm/
button
Alarm/
button
Alarm/
button
Measuring
Mode Alarm H Alarm HH
Alarm/
button
Alarm/
button
Forced Zero
Display
Up button
Up button
1 second or longer
Measuring
Mode
ON
Zro Fnc
Display
0
Forced Zero
Canceled
Time
Up button
Indicators
1 sec
OFF ON
Forced Zero
Executed
Zro Fnc
Measuring values after Forced Zero

47LPT
P. 7 / 16EM-9506-B Rev.6
■SCALING SETTING MODE
Shift
Shift
Scale/
Alarm/
Input Type
Measuring Mode
Input Scaling Value A
Display Scaling Value A
Input Scaling Value B
Display Scaling Value B
Decimal Point Position
*1
-19999...19999
V150 / V300
*1
Teach Calibration (Zero)
Set value
Setting Parameters
Set value
Teach Calibration (Span)
*2
*2
-19999...19999
0000...1.0000
Hold down Scale/ button
for ≥3 seconds
Hold down Alarm/ or Scale/ button
for ≥1 second
Shift
Shift
Up
Up
Up (set)
Up (set)
Alarm/ Scale/
Alarm/
Alarm/ Scale/
Alarm/
Alarm/ Scale/
Alarm/
Scale/
Alarm/
Shift
Shift
Alarm/
Scale/
Alarm/
Scaling Setting Mode
Alarm/
Alarm/
Alarm/
Alarm/
Scale/
Scale/
Scale/
Scale/ A
A
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
*1. Depends upon model suffix code. Refer to the following parameter list.
*2. Refer to Teach Calibration for detailed procedures.
Analog Output 100% Adj.
UP / DOWN Shift
Alarm/
Alarm/
Analog Output 0% Adj.UP / DOWN
Alarm/
Shift
Alarm/
Alarm/
Alarm/
Analog Output Action
at Force Zero Display
DISP / SCALE Shift
■OPERATIONS IN SETTING MODES
• Main display
The main display shows the current settings while the panel meter is in the setting mode.
• Shifting through setting parameters
Inanysettingmode,pressingAlarm/ buttonshiftsoneparametertothenext.PressingScale/ button shifts one to the
previous.
• Changing parameters
PressingShiftbuttonwhileoneoftheparametersettingsisindicatedonthedisplayshiftsthepanelmeterintothesetting
standby mode. The digit to which you can apply changes starts blinking.
PressUpbuttontochangetheblinkingvalue.
PressShiftbuttontogotothenextdigit.
PressAlarm/ orScale/ button to apply the new value and go to the next or previous parameter setting.
• If you get lost...
HolddownMax/Minbuttonfor1secondormoretoreturntothemeasuringmodewithoutapplyingthelastchanges.(Those
whichhavebeenalreadyappliedbypressingAlarm/ orScale/ button are not cancelled.)

47LPT
P. 8 / 16EM-9506-B Rev.6
...
...
...
...
...
...
PARAMETER
Input type
Input scaling value A
Display scaling value A
Input scaling value B
Display scaling value B
Decimal point position 4 positions or none Decimal point position
Display value for Input Scaling Value B
Input value for Full-scale point:
Set to a specific value by using the
buttons, or by Teach Calibration
Display value for Input Scaling Value A
Input value for Zero point:
Set to a specific value by using the
buttons, or by Teach Calibration
Measuring range 0 to 300V
Measuring range 0 to 150V
INDICATOR
SF
UNCTION DEFAULT
VALUE
DISPLAY
Zro Spn
Zro Tch
Zro D/P
Spn Tch
Spn D/P
D/P
Analog output action Proportional to the display value
Fnc
Zro
Spn Fnc
Proportional to the scaling value
Analog output 0%
adjustment
Analog output 0% adjustment:
adjustable range -5 to 105%
(increasing)
(decreasing)
(increasing)
(decreasing)
Fnc
Analog output 100%
adjustment
Analog output 100% adjustment:
adjustable range -5 to 105%
Scaling setting:
The 0% and 100% of the measurement range can be set to the desired values by setting parameters.
Note: After initialization, the parameters will return to the default values.
Input Scaling Value A
Enter the desired zero value (0%). (Input scaling value A < Input scaling value B)
Input Scaling Value B
Enter the desired full scaling value (100%).
Note: For setting input scaling value A and B, proceed with the same method. The combination of inputting parameter and Teach Calibra-
tion is unable.
For product version 1.04 or eariler;
Input Scaling Value A
"0.0" is displayed as factory default setting or after initialization. For setting desired input scaling value A, calculate by using the formula
shown below. (Input scaling value A < Input scaling value B)
(Example)
Input type: 0 to 150V
Input scaling: 50 to 150
50 + (150 x (5 / 100)) = 57.5
(Zero value) (5% value of full scale of input type 0 to 150V)
Input Scaling Value B
Enter the desired full scaling value (100%).
Decimal point position: Decimal point position is specified independently from the scaling. When you set the display scaling, include
zeros for fractions (10000 to show 10.000).
Analog output: When the forced zero or low-end cutout is ON, the analog output is proportional to the display value.
Analog output 0% / 100% adjustment: Pressing [Shift] button switches the signal to increase or decrease, and then pressing [Up] button
controls it toward the desired output value. 0% value < 100% value

47LPT
P. 9 / 16EM-9506-B Rev.6
A
AB
B
Display
Zero
Full Scale
Input
Input Scaling
Display Scaling
B
AB
A
Display
Full Scale
Zero
Input
Input Scaling
Display Scaling
• Normal Scaling
The display value increases when the input signal increases.
• Inverted Scaling
The display value decreases when the input signal increases.
• ‘Teach’ Calibration
‘InputscalingvalueA’and‘InputscalingvalueB’canbeadjustedbyapplyingactualinputsignals.Thesesettingswillbelost
afteraninitialization.
How to ‘Teach’ Input Values
1) HolddownScale/ button for 3 seconds or more to enter the scaling setting mode.
2) ChooseanappropriateinputtypeandpressAlarm/ buttontomoveontoInputscalingvalueA.
3) ‘Zro’and‘Tch’indicatorsturnon.
4) Zeropoint:PressUpbuttonand‘Tch’indicatorstartsblinking.
5) Apply5%inputsignal.PressUpbuttonagainand‘Tch’indicatorsreturnstoON.
6) PressAlarm/ buttontomoveontoDisplayscalingvalueA.
7) PressAlarm/ button again to move on to Input scaling value B.
8) ‘Spn’and‘Tch’indicatorsturnon.
9) Full-scalepoint:PressUpbuttonand‘Tch’indicatorstartsblinking.
10
)Apply100%inputsignal.PressUpbuttonagainand‘Tch’indicatorsreturnstoON.
Note1:ForsettinginputscalingvalueAandB,proceedwiththesamemethod.Thecombinationofinputtingparameterand
Teach Calibration is unable.
Note2:TheTeachCalibrationisafunctiontocalibratebyapplyingactualinputsignalof5%and100%.Theerrormight
increase depending on the way of the Teaching Calibration.

47LPT
P. 10 / 16EM-9506-B Rev.6
■ALARM SETTING MODE
Shift
Shift
Alarm/
Alarm Point
Measuring Mode
LL Setpoint
LL Deadband
LL ON Delay Time
-19999...19999
2 / 4*1
0000...9999
00...99
Hold down Alarm/ button
for ≥3 seconds
Hold down Alarm/ or Scale/ button
for ≥1 second
Shift
Shift
Alarm/ Alarm/
Alarm/ Alarm/
Alarm/ Alarm/
Alarm Setting Mode
Alarm/
Shift L Setpoint
L Deadband
L ON Delay Time
-19999...19999
0000...9999
00...99
Shift
Shift
Alarm/ Alarm/
Alarm/ Alarm/
Alarm/ Alarm/
L Coil at AlarmRYEN / RYDN Shift
Alarm/ Alarm/
Shift L Trip Action
LMHI / LMLO
Alarm/ Alarm/
LL Coil at AlarmRYEN / RYDN Shift
Alarm/ Alarm/
Shift LL Trip Action
LMHI / LMLO
Alarm/ Alarm/
A
Set value
Setting Parameters
BBBBBB
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/

47LPT
P. 11 / 16EM-9506-B Rev.6
Alarm/
Alarm/
Alarm/ Alarm/
Shift HH Trip Action
HH Deadband
Main Display Flashing at Alarm
LMHI / LMLO
0000...9999
B0 / B1 / B2 / B3 / B4
Shift
Shift
Alarm/ Alarm/
Alarm/ Alarm/
*1. Alarm point is fixed at “2” when the alarm model suffix code 2 is specified.
A
BB
BB BB
HH ON Delay Time00...99 Shift
Alarm/ Alarm/
Alarm/ Alarm/
HH Coil at AlarmRYEN / RYDN Shift
Alarm/ Alarm/
Shift HH Setpoint
-19999...19999
Shift H Setpoint
H Deadband
H ON Delay Time
-19999...19999
0000...9999
00...99
Shift
Shift
Alarm/ Alarm/
Alarm/ Alarm/
Alarm/ Alarm/
H Coil at AlarmRYEN / RYDN Shift
Alarm/ Alarm/
Shift H Trip Action
LMHI / LMLO
Alarm/ Alarm/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/

47LPT
P. 12 / 16EM-9506-B Rev.6
PARAMETER
Alarm points
Quad alarm: LL, L, H, HH
Dual alarm: L, H
INDICATOR
SF
UNCTION DEFAULT
VALUE
DISPLAY
LH Fnc
LHHH LL Fnc
...
...
...
LL setpoint
LL deadband (hysteresis)
LL ON delay time LL: ON delay time (seconds)
LL: Deadband (hysteresis) value
LL: Setpoint value
LL Fnc
LL trip action LL: Hi trip
LL Fnc
LL: Lo trip
LL coil at alarm LL: Coil energized at alarm
LL Fnc
LL: Coil de-energized at alarm
D/PLL Fnc
LLP Fnc
...
...
...
L setpoint
L deadband (hysteresis)
L ON delay time L: ON delay time (seconds)
L: Deadband (hysteresis) value
L: Setpoint value
L Fnc
L trip action L: Hi trip
L Fnc
L: Lo trip
L coil at alarm L: Coil energized at alarm
L Fnc
L: Coil de-energized at alarm
D/PL Fnc
LP Fnc
...
...
...
H setpoint
H deadband (hysteresis)
H ON delay time H: ON delay time (seconds)
H: Deadband (hysteresis) value
H: Setpoint value
H Fnc
H trip action H: Hi trip
H Fnc
H: Lo trip
H coil at alarm H: Coil energized at alarm
H Fnc
H: Coil de-energized at alarm
D/PH Fnc
HP Fnc
...
...
...
HH setpoint
HH deadband (hysteresis)
HH ON delay time HH: ON delay time (seconds)
HH: Deadband (hysteresis) value
HH: Setpoint value
HH Fnc
HH trip action HH: Hi trip
HH Fnc
HH: Lo trip
HH coil at alarm HH: Coil energized at alarm
HH Fnc
HH: Coil de-energized at alarm
D/PHH Fnc
HHP Fnc
Note 1: Alarm Setting Mode is locked with no-alarm-output type.
Alarm points depend upon the model suffix code. LL and HH setpoints are usable only for quad alarm type.
Note 2: LED status: = ON, = Blinking
Note 3: Specify setpoint and deadband in the scaled range values. Alarm is disabled when ‘ – – – – ’ is specified for the setpoint.
Note 4: All alarm setpoints are disabled (reset to ‘ – – – – ’ status) when the input type has been changed. Re-setting is required.
Note 5: Decimal point is not indicated when setting deadband values.
Main display flashing
at alarm Flashing in 1.0 sec. intervals
No flashing
Fnc
Flashing in 0.1 sec. intervals
Flashing in 0.2 sec. intervals
Flashing in 0.5 sec. intervals

47LPT
P. 13 / 16EM-9506-B Rev.6
■ADVANCED SETTING MODE
Shift
Shift
Alarm/
Moving Average
Measuring Mode
Brightness
Automatic Return Time
to Measuring Mode
Transition Time to
Lockout Setting Mode
C1 / C2 / C3 / C4 / C5
00...99
00...99
00.0...99.9
Hold down Alarm/ + Scale/ button
at once for ≥3 seconds
Hold down Alarm/ or Scale/ button
for ≥1 second
Shift
Shift
Alarm/ Alarm/
Alarm/ Alarm/
Alarm/ Alarm/
Alarm/
Shift
Alarm/
Alarm/
Advanced Setting Mode
Version Indication
Display Refreshing Rate
OFF / 2 / 4 / 8 / 16 / 32 / 64
Alarm/
A
A
Set value
Setting Parameters
Shift
Scale/
Alarm/
Low End Cutout
Low End Cutout Value
OFF / ON
000...999 Shift
Alarm/ Scale/
Alarm/
Alarm/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/

47LPT
P. 14 / 16EM-9506-B Rev.6
...
...
...
PARAMETER
Moving average
Low-end cutout
Low-end cutout value
Brightness
Automatic return time to Measuring Mode
Low-end cutout: Input signal below the preset cutout value is forcibly cut to 0.
Set for the three lowest digits of the scaled range (disregarding the decimal point)
With the cutout set to OFF, the low-end cutout value setting is locked.
Automatic return time to Measuring Mode: The display goes back automatically to Measuring Mode if the front buttons are left untouched
for the specified time period while it is in one of the setting modes (except the loop test output mode).
With this value set to 0, the display must always be exited manually from the setting mode.
Transition time to Lockout Setting Mode: The display goes to Lockout Setting Mode only when the designated buttons are pressed for the
specified time duration.
*1. F00.0 = 25 msec. refreshing rate
Specify in seconds
Brightness level 1 (dark)
Low-end cutout value setting
Low-end cutout ON
Low-end cutout OFF
Moving average with 8 samples
Moving average with 4 samples
Moving average with 2 samples
Moving average with 64 samples
Moving average with 32 samples
Moving average with 16 samples
Brightness level 4
Brightness level 5 (bright)
Brightness level 3
Brightness level 2
No moving averaging
INDICATORS FUNCTION DEFAULT
VALUE
DISPLAY
Fnc
Fnc
Zro Fnc
D/P Fnc
D/P Fnc
D/P Fnc
Transition time to Lockout Setting Mode
Specify in seconds
Fnc
Version indication Version number, indication onlyN/A N/A
...
Fnc
Display refreshing rate *1Specify in seconds

47LPT
P. 15 / 16EM-9506-B Rev.6
■LOCKOUT SETTING MODE
Shift
Shift
Scale/
Alarm/
Alarm Setting Lockout
Measuring Mode
Scaling Setting Lockout
Advanced Setting Lockout
Loop Test Output Lockout
Initialization
OFF / ON
OFF / ON
OFF / ON
OFF / RESET
Hold down Alarm/ or Scale/ button
for ≥1 second
Shift
Shift
Alarm/ Scale/
Alarm/
Alarm/ Scale/
Alarm/
Alarm/ Scale/
Alarm/
Shift
Scale/
Alarm/
Lockout Setting Mode
Alarm/
OFF / ON
Alarm/
Hold down Alarm/ + Max/Min button
at once for a preset time*1
*1. Preset time can be specified with “Transition Time to Lockout Setting Mode” in Advanced Setting Mode.
A
Scale/
Scale/
Scale/
Scale/
Scale/
Scale/
Set value
Setting Parameters
A
Forced Zero Lockout
OFF / ON Shift
Alarm/ Alarm/
PARAMETER
Alarm setting lockout
Lock Alarm Setting Mode
Unlock Alarm Setting Mode
INDICATORS FUNCTION DEFAULT
VALUE
DISPLAY
D/P Fnc
D/P Fnc
D/P Fnc
D/P Fnc
D/P Fnc
Scaling setting lockout
Advanced setting lockout Unlock Advance Setting Mode
Lock Alarm Setting Mode
Unlock Scaling Setting Mode
Forced zero lockout
Lock (Disable) Forced Zero operation
Unlock (Enable) Forced Zero operation
Lock Advanced Setting Mode
Initialization
Execute Initialization
Initialization prohibited
D/P Fnc
Loop test output lockout
Lock Loop Test Output Mode
Unlock Loop Test Output Mode

47LPT
P. 16 / 16EM-9506-B Rev.6
M-SYSTEM WARRANTY
M-System warrants such new M-System product which it manufactures to be free from defects in materials and workmanship during the 36-month period following the date that such
product was originally purchased if such product has been used under normal operating conditions and properly maintained, M-System’s sole liability, and purchaser’s exclusive
remedies, under this warranty are, at M-System’s option, the repair, replacement or refund of the purchase price of any M-System product which is defective under the terms of this
warranty. To submit a claim under this warranty, the purchaser must return, at its expense, the defective M-System product to the below address together with a copy of its original
sales invoice.
THIS IS THE ONLY WARRANTY APPLICABLE TO M-SYSTEM PRODUCT AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. M-SYSTEM SHALL HAVE NO LIABILITY FOR CONSEQUENTIAL, INCIDENTAL OR
SPECIAL DAMAGES OF ANY KIND WHATSOEVER.
M-System Co., Ltd., 5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN, Phone: (06) 6659-8201, Fax: (06) 6659-8510, E-mail: info@m-system.co.jp
ERROR MESSAGES
MAIN DISPLAY
Input error, Out of the measuring range
Non-volatile memory error (reading)
*1. If the unit does not recover its function after the initialization, repairing in the factory may be required.
Non-volatile memory error (writing)
Increase/decrease the input signal until it is back within the measuring range.
While the error message is on the display, press Up button for 3 seconds or
more, go to the lockout setting mode and initialize the unit to its factory
default status.*1
Repair is needed if the display does not recover after the power is reset.
ERROR MESSAGE WHAT TO DO
Internal data error
CHARACTER SET
0 1 2 43 5 6 7 98
–
A B DC E F G IH J K L NM O P Q SR T U V W X ZY
■LOOP TEST OUTPUT MODE
Shift Loop Test Output
Measuring Mode
-19999...19999
Set value
Setting Parameters
Hold down Alarm/ + Scale/ + Shift
button for ≥5 seconds
Hold down Alarm/ or Scale/ button
for ≥1 second
Loop Test Output Mode
*1. Hold down Alarm/ or Scale/ button for ≥1 second to return to the measuring mode while setting parameters.
*1
PARAMETER
Loopt test output: Pressing [Shift] button switches the signal to increase ( ON) or decrease ( ON), and then pressing [Up]
button controls it toward the desired output value.
Alarm trip functions according to the scaling values during the loop test.
*1. The specified decimal point position is applied to the loop test output value. -19999 to 19999 when ‘No decimal fraction’ is specified.
INDICATORS
/
FUNCTION DEFAULT
VALUE
DISPLAY
Tch Fnc
Min Max
...
Loop test output Scaling value for the loop test output
(display blinking)
Zro Spn D/P N/A
Max Min
*1
LIGHTNING SURGE PROTECTION
M-System offers a series of lightning surge protectors for
protectionagainstinducedlightningsurges.Pleasecontact
M-System to choose appropriate models.
Table of contents
Other M-system Measuring Instrument manuals

M-system
M-system 73VR1100 User guide

M-system
M-system SM10 User manual

M-system
M-system 48NV User manual

M-system
M-system M Series User manual

M-system
M-system KSE-x1 User manual

M-system
M-system 47 Series User manual

M-system
M-system 73VR3100 User guide

M-system
M-system 47LT User manual

M-system
M-system R5-MS User manual

M-system
M-system R9MWTU User manual

M-system
M-system TR30 User manual

M-system
M-system 47LM User manual

M-system
M-system 47LV User manual

M-system
M-system 47LHZ User manual

M-system
M-system JB2 User manual

M-system
M-system 47DM User manual

M-system
M-system R10-BS1 User manual

M-system
M-system 71VR1 User manual

M-system
M-system MXAB User manual

M-system
M-system 43DV1-S User manual