Mangar Camel User manual

Simple solutions for everyday independence
User Instrucons
CAMEL

2
Camel
The Camel by Mangar, is an inatable liing device
which is designed to sit up and li a fallen person
from the oor. When inated, the Camel lis the
paent into a raised seated posion ready to stand
with or without assistance.
This equipment may be used with or without a
helper as required.
The Camel is powered by a Mangar Airo Plus
Compressor and controlled using a 4-Way Hand
Control. Please refer to the separate Airo Plus
compressor instrucons before operang.
Patents Pending
WARNINGS
• This equipment MUST only be used by persons
familiar in its use.
• The medical state of a person needing to be
raised MUST be assessed before commencing a
li.
• Ensure that the surface on which the Camel is to
be placed is free from sharp objects. NOTE: Place
a blanket, groundsheet, towel or similar under
the Camel if using it on a rough surface.
• If a slide sheet has been used to transfer the
paent onto the Camel it must be removed from
under the paent before commencing a li.
• Secons 2, 3 and 4 MUST always be fully inated
in sequence from the boom up, in order to
maintain stability. NOTE: The backrest may be
parally inated before any other secon is
raised. Once the lowest secon is fully inated,
the backrest can be raised further as required.
• Do not operate the Airo compressor in the
presence of ammable anaesthec mixtures
with air, oxygen or nitrous oxide, as these
mixtures can create a combusble environment
leading to the potenal for re or even an
explosion in the presence of a spark.
Contents
Pg 3 Standard Package Contents
Pg 3 Connecng the Hand Control and Airo
Pg 4 Switching the Airo Plus ‘On’ & ‘O’
Pg 4 Posioning the paent on the Camel
Pg 4 Using the Camel
Pg 6 Aer Use
Pg 6 Maintenance
Pg 6 Roune Cleaning
Pg 6 Disinfecng
Pg 7 Specicaons
Pg 7 Warranty

Simple solutions for everyday independence
3
Camel
Standard Package Contents - g. 1
1 x Camel (A) (including carrying bag).
1 x 4-Way Hand Control (B).
1 x Airo Plus Compressor (C) with user instrucons.
Connecng the Hand Control and Airo -
g. 2
• Remove the Camel from the carrying bag,
unfasten the straps and unroll.
• Connect the four hoses from the hand control
to the corresponding numbered and coloured
connectors on the Camel (A).
• Connect the hose (B) from the hand control to
the Airo Plus compressor.
• Posion the Airo Plus compressor so that it will
not be in the way while posioning the paent.
Ensure that the hoses are not kinked.
1
2
A
C
A
B
B

4
Camel
Switching the Airo Plus ‘On’ & ‘O’ - g. 3
• To switch on, momentarily press the ‘Power’
buon (A g.3). The light (B) will come on and will
stay on while ‘Power’ is selected.
• To switch o, momentarily press the ‘Power’
buon. The baery status indicator light (C)
w i l l g o o u t a e r a p p r o x i m a t e l y 5 m i n u t e s .
NOTE: If the Airo Plus has not been used for
1 hour, it will switch itself o.
When the Airo Plus is switched on, the internal
microprocessor carries out a baery voltage check
each me air is required and if the voltage becomes
too low, the Airo Plus will automacally switch itself
o.
Posioning the paent on the Camel - g. 4
• Posion the Camel at the side of the paent so
that the paent’s head is level with the Camel
badge (A). Note: Place a blanket, groundsheet,
towel or similar under the Camel if using it on a
rough surface.
• Transfer the paent onto the Camel. The head
should lie on the Camel badge (A) and the body
should lie centrally.
Using the Camel
The following instrucons apply equally when the
Camel is being operated by a person raising themselves
without assistance, or by an assistant raising a paent.
WARNING: If a slide sheet has been used to transfer
the paent onto the Camel it must be removed from
under the paent before commencing a li.
• Press the number 1 buon on the hand control
to parally inate the backrest secon unl the
paent is in a comfortable posion - g. 5
• Press the number 2 buon to inate the boom
secon unl it becomes rigid - g. 6
3
4
6
5
A
B
A
C

Simple solutions for everyday independence
5
Camel
• Connue to press buons 3 and 4 in sequence
unl the secons become rigid, in exactly the
same manner - g. 7
• Press buon 1 to fully raise the backrest and
arrive at a sing posion - g. 8
• Whether assisted or used alone, the paent
should now be in a posion to stand or transfer
- g. 9
WARNING: In order to maintain stability, secons
must always be fully inated from the boom up,
in sequence.
Note: only the backrest (secon 1) may be parally
inated before the lower secons.
7
8
9

6
Camel
Aer Use
• To deate, disconnect each of the air hoses
from the Camel. NOTE: The inial ‘rush’ of air as
the hoses are disconnected from the Camel, can
be reduced by rst disconnecng the air hose
from the Airo Plus and then pressing each of
the buons on the hand control.
• Roll up the Camel with the hand control in the
hollow (A) - g. 10 secure the straps and place
in the carrying bag - g. 11
Maintenance
Other than roune cleaning, the Camel is
maintenance free.
We recommend that a thorough examinaon of
the Camel is carried out regularly; the frequency of
examinaon may need to be increased depending
on usage and operang environment.
If required, Mangar Internaonal can provide the
above inspecon within an annual service contract.
Please contact our Customer Service Department
for further details.
Roune Cleaning
Fully inate all secons of the Camel, sponge clean
with a non abrasive proprietary liquid cleaner or
disinfectant and dry thoroughly. Do not dry using
hot air.
Disinfecng
The Camel should be fully inated and pre-cleaned
by washing with detergent and water, thoroughly
rinsed and then disinfected using a 1:10 soluon of
household liquid bleach and water (1 part bleach
to 10 parts cold water). Leave the soluon on
the Camel for approximately 10 minutes before
thoroughly rinsing o with clean cold water. Allow
to air dry before storage. Do not dry using hot air.
A
10
11

Simple solutions for everyday independence
7
Camel
Specicaons
Maximum user weight: 320kg (50 st)
Seat height (approx): 56cm
Overall length (approx): 112cm
Overall width (approx): 70cm
Deated & rolled (in bag):
Length : 76cm
Diameter: 20cm
Weight (Inc. Hand Control): 6.5kg
Maximum air pressure: 0.35bar (5psi)
Max / Min temperature: +50°C / -15°C
112 cm
56 cm
70 cm
Camel serial number: .............................
Airo Plus serial number: .............................
Date of purchase: .............................
Invoice/Delivery note number: .............................
Mangar Internaonal Ltd, trading as Mangar Health, reserves the right to alter product specicaons and/or any of the informaon
contained within this document without noce.
Warranty
The guarantee is given against defects in material and workmanship and comprises piece, parts and labour.
The guarantee is not valid in the case of fair wear and tear, wilful damage, common neglect, misuse or
unauthorised alteraon or repair by unqualied persons.
The following statements do not aect the consumer’s statutory rights. This product is sold on the
understanding that in the event of any defect of manufacture or material appearing within the warranty
period, the defect will be reced free of charge providing that:
(a) Reasonable evidence is provided (e.g. purchase invoice, delivery note) that the claim is being made
within the warranty period.
(b) The defect is not aributable to accidental damage (either in transit or otherwise), misuse or
unauthorised repair.
Other manuals for Camel
3
Other Mangar Health Care Product manuals