Radical-7 Safety Information, Warnings and Cautions
www.masimo.com 16 Masimo
•Low arterial oxygen saturation.
•Excessive ambient or environmental noise.
•Improper sensor placement.
WARNING: Wireless communication of alarms to a secondary monitoring station should not
be relied upon as a primary alarm.
WARNING: A functional tester cannot be used to assess the accuracy of the Radical-7.
CAUTION: If using Radical-7 during full body irradiation, keep the sensor out of the radiation
field. If the sensor is exposed to the radiation, the reading might be inaccurate or the device
might read zero for the duration of the active irradiation period.
CAUTION: When patients are undergoing photodynamic therapy they may be sensitive to
light sources. Pulse oximetry may be used only under careful clinical supervision for short
time periods to minimize interference with photodynamic therapy.
CAUTION: The device must be configured to match your local power line frequency to allow
for the cancelation of noise introduced by fluorescent lights and other sources.
CAUTION: High ambient light sources such as surgical lights (especially those with a xenon
light source), bilirubin lamps, fluorescent lights, infrared heating lamps, and direct sunlight
can interfere with the performance of the sensor.
CAUTION: To prevent interference from ambient light, ensure that the sensor is properly
applied, and cover the sensor site with opaque material, if required. Failure to take this
precaution in high ambient light conditions may result in inaccurate measurements.
CAUTION: For home use, ensure that the Radical-7 alarm can be heard from other rooms in
the house, especially when noisy appliances such as vacuum cleaners, dishwashers, clothes
dryers, televisions, or radios are operating.
CAUTION: When Silence Duration is set to All Mute or All Mute with Reminder on Radical-7,
there will be no audible alarms on Radical-7 or Patient SafetyNet; however, there will be
visual alarms displayed on Radical-7 and Patient SafetyNet view.
CAUTION: If the Low Perfusion message is frequently displayed, find a better perfused
monitoring site. In the interim, assess the patient and, if indicated, verify oxygenation status
through other means.
CAUTION: To minimize radio interference, other electrical equipment that emits radio
frequency transmissions should not be in close proximity to Radical-7.
CAUTION: Do not place the Radical-7 on electrical equipment that may affect the device,
preventing it from working properly.
CAUTION: Failure to charge Radical-7 promptly after a Low Battery alarm may result in the
device shutting down.
CAUTION: To minimize electromagnetic interference, only use a SatShare cable that has a
ferrite bead installed.
CAUTION: During SatShare operation, alarms may be muted on the Radical-7. Use the
multi-parameter monitor for audible alarms during SatShare operation.
CAUTION: Ensure Radical Docking Station is connected to AC power source when charging
Radical-7.
CAUTION: Do not connect to an electrical outlet controlled by a wall switch or dimmer.