Display
Window
Display Window
Electrode Status Display (See Page 7) - Provides electrode conductance status of the
Sensor to indicate whether electrode adjustment is warranted.
Artifact (ARTF)- Measures the level of physiological (not brain-related) and
environmental noise that the system detects. This value ranges from 0 to 100%.
The numeric value of ARTF is a moving average.
Electromyograph (EMG) - Measures the level of detected muscle activity, such as
grimacing or jaw clenching. This value ranges from 0 to 100%. The numeric value of
EMG is a moving average.
Suppression Ratio (SR) - Measures the level of EEG signal suppression. This value
ranges from 0 to 100%. The numeric value of SR is a moving average.
Patient State Index (PSI) (See Page 9) - Measures anesthetic depth. This value ranges
from 0 to 100. The numeric value of PSI is a moving average. The trend display
for PSI is color-coded, with the green band representing the range of values that is
within user-configurable alarm limits.
EEG Display (See Page 8) - Reflect electrical activity of the frontal and pre-frontal
cortices of the brain.
Density Spectral Array (DSA) (See Page 10) - Utilizes color to represent the power
levels of EEG between 0 and 30 Hz (Upper limit can be adjusted to 40 Hz).
1
2
3
4
5
6
7
SedLine is a patient-connected, 4-channel processed electroencephalograph (EEG) monitor designed specifically for intraoperative or intensive
care use. It displays electrode status, EEG waveforms and Density Spectral Array (DSA).
General knowledge of electroencephalograph (EEG) monitoring and an understanding of the features and functions of the SedLine Sedation
Monitor are prerequisites for proper use.
Do not operate the SedLine Sedation Monitor without completely reading and understanding these instructions.
Notice
Purchase or possession of this device does not carry any express or implied license to use with replacement parts which would, alone or in
combination with this device, fall within the scope of one of the relating patents.
Caution: Federal law restricts this device to sale by or on the order of a physician.
Caution: For professional use. See instructions for use for full prescribing information, including indications, contraindications, warnings,
precautions and adverse events.
For further information contact:
Masimo Corporation
40 Parker
Irvine, CA 92618, USA
Tel.: 949-297-7000
Fax.: 949-297-7001
www.masimo.com
Patents: www.masimo.com/patents.htm
SedLine, Masimo, and are registered trademarks of Masimo Corporation.
All other trademarks and registered trademarks are property of their respective owners.
Printed in the USA.
© 2015 Masimo Corporation.
EU authorized representative for Masimo Corporation:
MDSS GmbH
Schiffgraben 41
D-30175 Hannover, Germany
EC REP