15
4.5.1 In the process of measurement, if ambient light
is too dark to read, press “ ”key to turn on the
backlight. The backlight will automatically turn off
after about 30 seconds.
4.5.2 During this period, pressing “ ”key will turn off
backlight.
4.5.3 In the current grade, the meter will turn on
backlight and clamp head light.
Backlight is LED with high working current.
If backlight is used often, it will shorten battery
life, so use backlight only when necessary.
Note: When battery voltage <7V, LCD displays “ ”
(low voltage) symbol. But when using the backlight,
the battery voltage will drop due to the high working
current and the“ ”symbol may display. (When the
“ ”symbol shows, accuracy of measurement is not
guaranteed). At this time, don't replace the battery.
Continue to use the meter normally without using the
backlight, and replace the battery only when the “ ”
symbol shows under normal use.
_
4.6 Automatic power-off
4.6.1 If there is no operation during any 15 minute
period after turning the machine on, the meter
will enter a suspended state and automatically
power off to save the battery. Within 1 minute
before shutdown, the buzzer will sound 3 times.
Immediatelyt before shutting down, the buzzer
will make a long sound.
4.6.2 After automatic power-off, press the “FUNC” key
to restore the meter to normal function.
4.6.3 Holding any other key except “FUNC” key and
“HOLD” more than 2 seconds when powering on
will cancel the automatic power-off function.
4.7 Measurement preparation
4.7.1 Turn the transfer switch to turn on the power.
When battery voltage is low (about <7V), LCD
displays “ ”symbol. Replace the battery
immediately.
4.7.2 “ ”symbol means that input voltage or current
should not be more than the indicated value. This
is to protect the internal line from damage.
4.7.3 Place transfer switch to the required measuring
function and range.
4.7.4 When connecting line, please connect the
common test line first, then connect charged test
line. When removing line, remove charged test
line first.
_
4.8 Current Measurement
Electric shock hazard.
Remove the probe from the meter before
measuring with current clamp.
Warning
4.8.1 Place measuring switch in position . At this
time, the meter is in the AC current measurement
state.
4.8.2 Hold the trigger, open clamp head, and clip one
lead of measurement circuit to be tested in the
clamp.
4.8.3 LCD will show readings.
Note:
1) Clamping two or more leads of circuit to be tested
simultaneously will give invalid readings.
2) To get accurate readings, place the lead to be tested
at the center of current clamp.
3)“ ” indicates that maximum input AC current is
1000A and DC current 1000A.
A~
16
4.5 Back light and clamp head light