Note: To gain better access to the sensitivity switch on Mk1 Water
Level Meters, you may want to unscrew the two screws holding
the circuit board to the Sonalert, in order to move the circuit
board out of the way.
Back of Mk1 Water Level Meter Faceplate
Sensitivity
Switch
Molex
Connector
Instructions
For Mk1 Water Level Meters (Molex Tape Connection)
1. Remove the battery.
2. Unscrew the small brass screw on the side of the sensitivity
knob. Remove the sensitivity knob.
3. Unscrew and remove the faceplate.
4. Disconnect the Molex connector that connects the circuit
board to the tape.
5. Unsolder the wires at the connections to the sensitivity
switch; note the order in which they are connected.
6. Unscrew the nut holding the sensitivity switch. Remove the
switch from the faceplate.
7. Use the blue resistor to make a connection between the
centre and right prongs on the switch. Solder in place. See
photos at top right.
8. Install the new sensitivity switch through the faceplate.
Position the three connection prongs opposite the battery
tray and slightly towards the centre of the reel. Secure
using the nut.
9. Strip the ends of the wires removed from the sensitivity
switch by 3 - 4 mm, if required. Attach the wires to the
correct positions by soldering:
The two red wires on the bottom of the circuit board are
connected to the first two connections on the sensitivity
switch. The black wire on the bottom of the circuit board is
connected to the third prong on the sensitivity switch. The
black wire closest to the top of the circuit board is attached
to the prong on the top of the sensitivity switch. The other
black wire connected to the top of the sensitivity switch is
connected to the negative terminal on the battery tray. Use
the diagram and photo below for reference.
Remove
Screws for
Easier Access
Note: You do not require the resistor marked “3092” when replacing
the sensitivity switch for a Mk1 Water Level Meter.
10. Reconnect the tape to the circuit board using the Molex
connector.
11. Replace the battery and test the Water Level Meter to
ensure the connections are correct.
12. Fasten the sensitivity knob using the brass screw
13. Replace the faceplate, with three screws.
11. Reconnect the tape to the circuit board. Press down on
the white terminals on the circuit board and insert the
tape leads. Release the terminals and the leads should be
secured. The lead on the bottom of the tape (numbers
facing up, may be black mark on lead) is inserted into the
terminal with a white square below it on the circuit board.
12. Replace the battery and test the Water Level Meter to
ensure the connections are correct.
13. Fasten the sensitivity knob using the brass screw
14. Replace the faceplate, with three screws.