manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. MATE
  6. •
  7. Test Equipment
  8. •
  9. MATE Pilot User manual

MATE Pilot User manual

The Mate Logo is a registered trademark of Mate Precision Tooling Inc.
Mate Precision Tooling • 1295 Lund Boulevard • Anoka, Minnesota 55303 USA • Phone 763.421.0230 • 800.328.4492 • Fax 763.421.0285 • 800.541.0285
Dimensions in Inches(millimeters) mate.com
Mate Pilot™ Turret Calibration System
Installation and Operation Instructions
Verification Mode
Adjustment Handle
Tensioning Knob
T-Bar
Battery Cover
Tri-color indicator light
Key Slot
Mate Pilot™ Turret
Calibration System
These instructions show the
operation of the Mate Pilot
Calibration System in two
modes.
• Verification Mode—
Used to confirm the precise
concentric and angular
alignment of your turret to
maintain high quality piece
parts, extend tool life and
increase productivity.
• Alignment Mode—Used
to restore the concentric
and angular alignment of
each station with the same
or better precision as the
initial machine installation.
Achieving precise alignment
improves piece part quality,
extends tool life, and
increases productivity.
Verification Mode
1. Rotate the turret until the station to be verified is in the tool change position and remove punch guide and/or die.
2. Install the lower calibration instrument into the die holder and tighten the clamp screw(s) as you would for a die.
3. Install the upper calibration instrument into upper holder, gently lowering the instrument until the lock pawl (see Figure 1) rests on the top of
the turret key in the machine. Caution: Do not allow the upper instrument to drop onto the turret key or through the turret bore as this may
damage the instrument and/or the turret bore.
4. Look into the turret gap to ensure the threaded end of the length adjustment handle is not engaged with the lower calibration instrument.
5. Rotate the turret until the station to be verified is positioned under the ram. Note: The turret must be locked in place. For auto-index
stations, the auto-index must be engaged.
6. Lower the upper instrument gently, by depressing the release button and lowering the instrument using the adjustment handle, until the
threaded end of the adjustment handle rests on the top of the lower instrument. Caution: Do not allow the upper instrument to drop onto the
turret key or through the turret bore as this may damage the instrument and/or the turret bore.
7. Reach into the machine and rotate the T-bar on the adjustment handle until the red light just turns on, then continue rotating with the
tensioning knob (see Photo 2) until the interlocking teeth of the upper and lower instruments are engaged and the tensioning knob starts to
click. Do not use the T-bar when any of the lights are on as this may result in a false verification.
8. View the color of the indicator light. If the light is obscured, then view the reflection of the light on the handle.
• Red: angularity and concentricity is not confirmed. Perform an alignment as described on pages 2 and 3.
• Yellow: angularity and concentricity is within 0.0012(0.030)
• Green: angularity and concentricity is within 0.0003(0.008) - ideal for 0.048(1.20) material or less.
9. Loosen the adjustment handle until the threaded end of the adjustment handle is fully disengaged from the lower instrument.
10. Lift the upper instrument gently, using the length adjustment handle, until the lock pawl rests on the top surface of the turret key.
11 Look into turret gap to ensure the threaded end of the length adjustment handle is not engaged with the lower instrument.
12. Rotate the turret until the station that has been verified is in the tool change position.
13. Note: If you did not previously use the alignment mode, then skip this step. Tighten any screws that were previously loosened during
alignment in accordance to the punch press manufacturer’s torque specifications.
14. Repeat steps 5 to 11 above to ensure the tool holder did not move during tightening.
15. Remove the upper and lower instruments.
Photo 2: Tighten tensioning
knob on adjustment handle
until knob starts to click.
Photo 1: Install upper and
lower instruments. Rotate
station under machine ram.
Photo 3: View Light Color
Figure 2:
Engaged, but not aligned
Angularity and concentricity within 0.0012(0.030)
Angularity and concentricity within 0.0003(0.008)
Note: Angularity and concentricity within 0.0003(0.008) -
Green Indicator Light - is recommended when punching
materials with a thicknesses of 0.048(1.20) or less.
Release Button
Upper Instrument
Lower Instrument
Interlocking Teeth
Key Slot
Threaded End of
Adjustment Handle
Sensor Contact Surface
Sensors
Figure 1:
Retaining Clip
Lock Pawl
Pilot is a trademark of Mate Precision Tooling Inc.
The Mate Logo is a registered trademark of Mate Precision Tooling Inc.
Mate Precision Tooling • 1295 Lund Boulevard • Anoka, Minnesota 55303 USA • Phone 763.421.0230 • 800.328.4492 • Fax 763.421.0285 • 800.541.0285
Dimensions in Inches(millimeters) mate.com
Mate Pilot™ Turret Calibration System
Installation and Operation Instructions
Alignment Mode for Machines with Fixed Upper and Moveable Lower Holders
Alignment Mode—For thick turret punch press with fixed upper holder and moveable lower holder.
Used to restore the concentric and angular alignment of each station with the same or better precision as the
initial machine installation. Achieving precise alignment improves piece part quality, extends tool life, and
increases productivity.
Install the Instruments into the Punch Press
1. Rotate the turret until the station to be aligned is in the tool change position, and remove punch holder and/or die as applicable.
(Tip: Tooling in adjacent stations should also be removed to provide an adequate work space.)
2. Inspect the upper turret bore for damage. Pay particular attention to the turret bore keys. The upper calibration instrument should
slide freely in the upper turret bore. Repair as required, prior to alignment.
3. Loosen the screw(s) that hold die holder in place. See photo 1. (Tip: Any thread-retaining compound used in previous installations
should be removed.)
4. Tighten the screw(s) mentioned above until just snug. The die holder will need to slide during the alignment process.
5. Install the lower calibration instrument into the die holder and tighten the clamp screw(s) as you would for a die. See photo 2
6. Install the upper calibration instrument into upper holder, gently lowering the instrument until the lock pawl (see figure 1 on page 1)
rests on the top of the turret key in the turret bore. Caution: Do not allow the upper instrument to drop onto the turret key or through
the turret as this may damage the instrument and/or the turret bore. See photo 3.
7. Look into the turret gap to ensure the threaded end of the length adjustment handle is not engaged with the lower calibration
instrument. See photo 4.
Align Lower Holder
1. Rotate the turret until the station to be aligned is under the ram. Note: The turret must be locked in place. For auto-index stations,
the auto-index must be engaged.
2. Lower the upper instrument gently by depressing the release button and lowering the instrument using the adjustment handle, until
the threaded end of the adjustment handle rests on the top of the lower instrument. Caution: Do not allow the upper instrument to
drop through the turret bore as this may damage the instrument and/or the turret bore.
3. Reach into the machine and tighten the adjustment handle, using the T-bar, until the interlocking teeth of the upper and lower
instruments are fully engaged. The indicator light will change color from red, to yellow, and then to green. See photo 5. During this
process the lower tool holder will be adjusted into precise concentric and angular alignment with the upper tool holder. Caution: Do
not use any device other than the T-bar to tighten the adjustment handle. Excessive torque may damage the machine and/or the
calibration instrument.
4. Loosen the adjustment handle until the threaded end of the adjustment handle is fully disengaged from the lower instrument.
5. Lift the upper instrument gently, using the length adjustment handle, until the lock pawl rests on the top surface of the turret key.
6. Look into turret gap to ensure the threaded end of the length adjustment handle is not engaged with the lower calibration
instrument. See photo 4.
7. Rotate the turret until the station that has been aligned is in the tool change position. Caution: Do not rotate the turret with the two
halves of the calibration instrument tightened together.
8. Tighten the die holder retaining screws until snug.
9. Proceed to step 5 of the Verification Mode Procedure on the previous page.
*The Mate Pilot™ Thick Turret Calibration System is not recommended for use in a punch press with an eccentric turret.
Photo 1: Loosen Die holder
Photo 5: Align the station,
using the T-bar to tighten the
adjustment handle until the
light goes green.
Photo 2: Install Lower Instrument
Photo 3: Install Upper Instrument
Photo 4: Check disengagement
Patent Pending
The Mate Logo is a registered trademark of Mate Precision Tooling Inc.
Mate Precision Tooling • 1295 Lund Boulevard • Anoka, Minnesota 55303 USA • Phone 763.421.0230 • 800.328.4492 • Fax 763.421.0285 • 800.541.0285
Dimensions in Inches(millimeters) mate.com
Mate Pilot™ Turret Calibration System
Installation and Operation Instructions
Alignment Mode for Machines with Movable Upper and Lower Holders
Alignment Mode—For thick turret punch press with removable upper and lower holders.
Used to restore the concentric and angular alignment of each station with the same or better precision as the
initial machine installation. Achieving precise alignment improves piece part quality, extends tool life, and
increases productivity.
Install the Instruments into the Punch Press
1. Rotate the turret until the station to be aligned is in the tool change position, then remove punch holder and/or die as applicable.
(Tip: Tooling in adjacent stations should also be removed to provide more work space.)
2. Inspect the upper tool holder for damage. Pay particular attention to the tool holder alignment keys. The upper calibration instrument
should slide freely in the upper tool holder. Repair as required, prior to alignment.
3. Loosen the screw(s) that hold the upper and lower tool holders in place. See photo 1. (Tip: Any thread retaining compound used in
previous installations should be removed.)
4. Tighten the screw(s) mentioned above until just snug. The upper and lower tool holders will need to slide during the alignment
process.
5. Install the lower calibration instrument into the die holder and tighten the clamp screw(s) as you would for a die.
6. Install the upper calibration instrument into upper holder, gently lowering the instrument until the lock pawl (see Figure 1 on page 1)
rests on the top of the turret key in the turret bore. Caution: Do not allow the upper instrument to drop onto the turret key or through
the turret bore as this may damage the instrument and/or the machine. See photo 2.
7. Look into the turret gap to ensure the threaded end of the length adjustment handle is not engaged with the lower calibration
instrument. See photo 3.
Align Upper and Lower Holder
1. Rotate the turret until the station to be aligned is under the ram. Note: The turret must be locked in place. For auto-index stations,
the auto-index pins must be engaged.
2. Lower the upper instrument gently by depressing the release button and lowering the instrument using the adjustment handle, until
the threaded end of the adjustment handle rests on the top of the lower instrument. Caution: Do not allow the upper instrument to
drop through the turret bore as this may damage the instrument and/or the turret bore.
3. Reach into the machine and tighten the adjustment handle using the T-bar, until the interlocking teeth of the upper and lower
instruments are fully engaged. The indicator light will change color from red, to yellow, and then to green (see photo 4). During this
process the upper and lower tool holders will be adjusted into precise concentric and angular alignment. Caution: Do not use any
device, other than the T-bar, to tighten the adjustment handle. Excessive torque may damage the machine and/or the calibration
instruments.
Align the Station to the Punch Press:
1. Attach the alignment bar to the upper calibration instrument using the M4 screw provided.
2. Clamp an approximately 24”(600mm) x 48”(1200mm) piece of mild steel 0.100(2.50) thick or thicker into the punch press work
holder clamps. See photo 5.
3. Mount a dial-test-indicator (DTI) with a magnetic base in a suitable position so the probe can be moved along the length of the
alignment bar. See photo 5.
4. Use the manual jog mode on the machine control to move the sheet metal with the DTI attached, so that the probe of the DTI can
move along the full length of the alignment bar. The total indicator reading (TIR) should be <0.001(0.02) when the tool holder is
correctly aligned. See photo 5.
5. Use the adjustment bar to rotate the calibration instruments (and the upper and lower tool holders) until the total indicator reading
(TIR) of <0.001(0.02) is achieved. You may need to move the sheet with the DTI back and forth several times until the desired
reading is achieved. TIP: If it is difficult to rotate the calibration instruments, slightly loosen one or more of the retaining screws.
6. Tighten all the accessible retaining screws. See photo 6.
7. Recheck the alignment (repeat steps 4 to 6) to confirm the tool holders did not move during tightening. If the TIR of <0.001(0.02) is
achieved, then proceed to the next step. If not, repeat steps 4 to 6.
8. Remove the alignment bar.
9. Lift the upper instrument gently, using the length adjustment handle, until the lock pawl rests on the top surface of the turret key.
10. Look into turret gap to ensure the threaded end of the length adjustment handle is not engaged with the lower calibration
instrument.
11. Rotate the turret until the station that has been aligned is returned to the tool change position. Caution: Do not rotate the turret with
the two halves of the calibration instrument tightened together.
12. Tighten any retaining screws not previously tightened in step 6 above.
13. Proceed to step 5 of the Verification Mode Procedure.
Photo 1: Loosen upper and
lower tool holder retaining
screws
Photo 3: Check disengagement
Photo 4: Align the station,
using the T-bar to tighten the
adjustment handle until the
light goes green.
Photo 5: Draw the probe of the
dial-test-indicator along the
length of the alignment bar.
Photo 2: Install upper and
lower instruments
Photo 6: Tighten the upper and
lower tool holder retaining
screws.
The Mate Logo is a registered trademark of Mate Precision Tooling Inc.
Mate Precision Tooling • 1295 Lund Boulevard • Anoka, Minnesota 55303 USA • Phone 763.421.0230 • 800.328.4492 • Fax 763.421.0285 • 800.541.0285
Dimensions in Inches(millimeters) mate.com
LIT00366 Rev C
Mate Pilot™ Turret Calibration System
Installation and Operation Instructions
Storage Instructions
1. Separate the upper from the lower by unscrewing the adjustment handle.
2. Remove the batteries if storage for longer than three (3) months is expected. See battery installation instructions.
3. Coat the lower with a spray lubricant/preservative or light machine oil. Examples: LPS 1, WD-40.
4. Wipe off any excess lubricant/preservative from the sensor contact surface.
5. Re-assemble the lower and upper aligning the external orientation slots, and tighten the handle until the light glows green.
6. Coat the outside surface of the upper instrument with the lubricant/preservative.
7. Loosen the handle and position the upper and lower with the interlocking teeth disengaged to prevent inadvertent
activation of the indicator light during storage. See photo at right.
8. Tighten the handle using the tensioning knob until a click is felt.
Battery Removal and Installation
1. Separate the upper from the lower by unscrewing the adjustment handle.
2. Remove the clip which secures the adjustment handle to the upper with a screwdriver. Be careful not to
damage the sensors.
3. Remove handle from the upper assembly.
4. Unscrew the button head screws on the cover. Use 2mm hex wench for 1-1/4” B station. Use 2.5mm hex
wrench for 2” C, 3-1/2” D, and 4-1/2” E Stations. Remove the cover and set aside.
5. Remove old batteries.
6. Install new batteries paying attention to polarity.—Locate the negative (-) of the battery against the spring in
the battery pack first. Then compress the spring with the battery, and slide battery into the battery pack until
the positive (+) end of the battery touches the end contact.
7. Test the operation of the batteries. Gently lower the upper instrument on to the lower instrument until the
interlocking teeth are fully engaged. If the light fails to activate, repeat steps 5 and 6.
8. Reinstall the cover, using the button head screws and wrench used in step 4.
9. Reinstall the adjustment handle. Install the handle through the upper instrument until the tip protrudes
through the underside. Replace the clip into the groove on the adjustment handle. Use needle nose pliers
to guide the ears of the clip into the slot until is snaps into position.
Removal and Reinstallation of Adjustment Handle.
• Remove the clip using a flat bladed
screw driver to pry out the retaining
clip. Take care not to damage the
sensors located around the handle
• Replace the clip using needle nose
pliers to guide the ears of the clip
into the slot on the handle until it
snaps into position.
• Pry the first battery out by
placing the tip of a flat
bladed screwdriver under
the top lip of the battery
holder and on top of the
flat of the battery.
• Pry the second battery out
by placing the tip of a flat
bladed screwdriver behind
the body of the battery.
Battery Removal 2” C, 3-1/2” D and 4-1/2” E Stations
Preparation for Use
1. Separate the upper from the lower by unscrewing the adjustment handle.
2. Ensure that the serial number of both the upper and lower instruments are identical.
3. Wipe off any lubrication/preservative oil from the unit with a soft cloth paying close attention to the interlocking teeth. Compressed air may be
used; care should be taken not to blow air in the sensor area. Always blow from the inside to the outside.
4. Install the batteries (if applicable). See Battery Removal and Installation.
5. Re-assemble the upper and lower by tightening with the tensioning knob.
6. Check the color of the tri-color indicator light.
• No light—check battery installation. Replace battery if required. See section below.
• Red or Yellow—interlocking teeth may be dirty. Clean interlocking teeth and recheck.
• Green—calibration system is operational and ready form immediate use.
Note: If you cannot get the light to turn green by using the tension knob only and the parts are clean, the unit may be out of adjustment. Contact
Mate Precision Tooling for immediate assistance.
Pilot is a trademark of Mate Precision Tooling Inc.

Other MATE Test Equipment manuals

MATE DYNOGRIP VISE User manual

MATE

MATE DYNOGRIP VISE User manual

MATE DYNOLOCK BASE User manual

MATE

MATE DYNOLOCK BASE User manual

MATE DYNOMOUNT User manual

MATE

MATE DYNOMOUNT User manual

Popular Test Equipment manuals by other brands

Telequipment D52 manual

Telequipment

Telequipment D52 manual

PCE Health and Fitness PCE-LES 300 user manual

PCE Health and Fitness

PCE Health and Fitness PCE-LES 300 user manual

Hillstone HAC240 operating manual

Hillstone

Hillstone HAC240 operating manual

Crystal GauGeCaLXP Service Instruction Sheet

Crystal

Crystal GauGeCaLXP Service Instruction Sheet

VDO CTC II installation instructions

VDO

VDO CTC II installation instructions

Tektronix RSA5100B series instructions

Tektronix

Tektronix RSA5100B series instructions

YOKOGAWA 90020B user manual

YOKOGAWA

YOKOGAWA 90020B user manual

Hella Gutmann mega macs PC quick start guide

Hella Gutmann

Hella Gutmann mega macs PC quick start guide

JDS Uniphase IP Video Test Option HST-3000 Testing Guide

JDS Uniphase

JDS Uniphase IP Video Test Option HST-3000 Testing Guide

XPOtool 61185 user manual

XPOtool

XPOtool 61185 user manual

CLEAN Instruments PH30 Operation manuals

CLEAN Instruments

CLEAN Instruments PH30 Operation manuals

Sonel TKF-13 Operation manuals

Sonel

Sonel TKF-13 Operation manuals

Fluke Fiber OneShot PRO user manual

Fluke

Fluke Fiber OneShot PRO user manual

Tektronix TDS1001B Security instructions

Tektronix

Tektronix TDS1001B Security instructions

Emec ECDI Series operating manual

Emec

Emec ECDI Series operating manual

Monarch Nova-Strobe bax instruction manual

Monarch

Monarch Nova-Strobe bax instruction manual

SADT HARTIP 4100 Operation manual

SADT

SADT HARTIP 4100 Operation manual

Fluke FEV-CON-TSL Safety information

Fluke

Fluke FEV-CON-TSL Safety information

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.