MedSource MS-74011 User manual

Copyright
Our company owns all rights of this unpublished work and intends to maintain it as a
confidential work. We may also seek to maintain this work as an unpublished copyright. This
publication is to be used solely for the purpose of reference or operation of our device. No
part of this work can be disseminated for other purposes.
In the event of inadvertent or deliberate publication, we intend to enforce its rights to this work
under copyright laws as a published work. Those having access to this work may not copy,
use, or disclose the information in this work unless expressly authorized by our company.
All information contained in this publication is believed to be correct. We shall not be liable for
errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this material. The information this publication refers to is
protected by copyrights or patents and does not convey any license under the patent rights
of our company, nor the rights of others. We do not assume any liability arising out of any
infringements of patents or other rights of third parties.
Content of the manual is subject to change without prior notice.
ALL RIGHTS RESERVED
Bluetooth®and the Bluetooth®Logo are registered trademarks of Bluetooth®SIG, Inc.
Ver2.0
Issue date: January 18, 2011

Contents
CHAPTER 1 Introduction ................................................................................ 1
1.1 Brief Introduction ....................................................................................................... 1
1.2 Safety Information ..................................................................................................... 1
1.3 Electromagnetism Interference .................................................................................. 3
CHAPTER 2 General Description ................................................................. 4
2.1 Appearance ............................................................................................................... 4
2.2 Explanation of Symbols ............................................................................................... 6
2.3 Power Supply .............................................................................................................. 7
2.4 Product Features ....................................................................................................... 8
CHAPTER 3 SpO2Probe Installation ........................................................... 8
CHAPTER 4 ID, Date and Time Setting ........................................................ 9
4.1 Date & Time Settings .................................................................................................. 9
4.2 User ID Setting ............................................................................................................ 9
CHAPTER 5 Take a Measurement ................................................................. 10
CHAPTER 6 Other Settings .......................................................................... 12
6.1 Alarm Setting ........................................................................................... 12
6.2 Data Management .................................................................................. 14
6.3 System Setting ........................................................................................ 16
CHAPTER 7 Maintain and Repair ................................................................ 18
7.1 Maintenance ............................................................................................ 18
7.2 Safety Checks ......................................................................................... 18
7.3 Calibration and Verification ......................................................................19
7.4 Trouble Shooting ..................................................................................... 20
7.5 Warranty and Repair ............................................................................... 20
7.6 Storage and Transportation ..................................................................... 22
APPENDIX A Specifications .......................................................................... 23
APPENDIX B Statement of Manufacturer .................................................... 25

MS-74011
MS-74011

- 2 -
● When connecting this device to other peripherals, make sure that you are qualified to
operate this device. Any peripheral must be certified according to the protocol of IEC 950
and IEC 601-1-1. Any input/output device should follow the protocol of IEC 601-1-1.
● Considering the probe is a sensitive device, please strictly follow the probe application
instructions.
● The malfunction of the probe may cause inaccurate data which serves as a foundation to
treat patients, so make sure to pay more attention to the probe and inspect it usually.
● The worn-out data cables may cause inaccurate data which is used as a reference to treat
patients, so please pay attention to the data cable and check it frequently.
● Do not touch the AC adapter with wet hands. Otherwise, you may suffer electric shock.
● The disposable accessories should not be cycled.
● Clean the probe with an H2O solution and a neutral detergent. Do not submerge the probe.
Do not use the probe in autoclave (sterilizer).
● This device is intended for use by persons trained in professional health care. The operator
must be thoroughly familiar with the information in this manual before using the device.
● Before cleaning or disinfecting the probe, unplug it from the pulse oximeter to prevent
probe or pulse oximeter from being damaged, and to protect user under safety situation.
● To avoid an electrical hazard, never immerse the unit in any liquid or attempt to clean
it with liquid cleaning agents. Always disconnect the device from the AC adapter before
performing cleaning of maintenance.
● Alarm must be set up according to different situations of individual patient. Make sure that
chime sound can be activated when alarm function begins to work.
Cautions:
Notes:
● Application of this device may influence the measuring accuracy in the background of
electromagnetic areas such as electro-surgery environment.
● SpO2measurements may be adversely affected in the presence of high ambient light.
Shield the probe area (with a surgical towel, for example) if necessary.
● Dyes introduced into the bloodstream, such as methylene blue, indocyanine green, indigo
carmine, and fluorescein, may adversely affect the accuracy of the SpO2reading.
● Any condition that restricts blood flow, such as use of a blood pressure cuff or extremes in
systemic vascular resistance, may cause a failure to determine accurate pulse rate and
SpO2readings.
● Remove fingernail polish or artificial fingernails before applying SpO2probes. Fingernail
polish or artificial fingernails may lead to inaccurate SpO2readings.
● Hazards arising from software errors have been minimized. Hazard analysis conforms
to meet ISO14971: 2000 and EN60601-1-4: 1996. Significant levels of dysfunctional
hemoglogins, such as carboxyhemoglogin or methhemoglobin, will spawn an affection of
the accuracy of the SpO2measurement.
● Optical cross-talk can occur when two or more probes are located in adjoining area. It can
be eliminated by covering each site with opaque material. Optical cross-talk may adversely
affect the accuracy of the SpO2readings.
PULSE OXIMETER

- 3 -
● Obstructions or dirt on the probe’s red light or detector may cause a probe failure. Make
sure there are no obstructions and the probe is clean.
● The AC adapter and accessories used with the device should be complied with the
requirement of IEC60601-1.
● For routine equipment maintenance, please refer to the service procedures at the
associated section as indicated in the manual.
● As to the other concerns for attention, please carefully look through the specific chapter in
this instruction.
● Federal law restricts this device to sale by or on the order of a physician.
This pulse oximeter is designed and tested in compliance with the EMC standard, complying
with the international standard for the EMC of the electronic medical device - IEC 60601-1-2.
However, because of the proliferation of radio frequency transmitting equipment and
other sources of electrical noise in the health-care and home environments (e.g. cellular
phones, mobile two-way radios, electrical appliances), it is possible that high levels of
such interference due to close proximity or strength of a source, may result in disruption of
performance of this device.
This apparatus complies with the IEC 60601-1-2 international standard. The requirements of
this international standard are: CISPR11, GROP1, and CLASS B.
1.3 Electromagnetism Interference
INSTRUCTION MANUAL

- 4 -
CHAPTER 2 General Description
The pulse oximeter adopts 2.8 inch TFT screen, which can display the SpO2, pulse rate and
other indication parameters, such as time, ID number, pulse amplitude bar and battery power
status, alarm limits and the connections of probes, etc.
Description of Fig.2-1:
(1) SpO2socket: For connecting the SpO2probe with the pulse oximeter. Also used to
connect the USB cable for data transmission to PC.
(2) Alarm • lamp: When SpO2or/and PR alarm occurs, it flashes (the color of the lamp is yellow).
(3) Up • button: Press this button to increase the value by one increment. Or press it down
and hold it to continuously increase the value.
(4) Down •button: Press this button to decrease the value by one decrement. Or press it
down and hold it to continually decrease the value.
(5) Menu/Select • button: For entering main menu, or confirming the selection/setting.
(6) Alarm inhibit/off • button: Press this button to inhibit alarm sound for 120 seconds. And
long press the button, power off the alarm.
(7) Return/Lock •button: On the measuring screen, it serves as Lock button; On the menu
and sub-menu screen, it serves as Return button.
(8) Power switch: Press and hold it for 3 seconds to power the device on, and for about 4
seconds to turn the device off.
(9) Charge indicator: During the pulse oximeter is being charged, the lamp is flashing; when
it is charged to full, the lamp is lighted without flashing; and the lamp is not lighted when
it is not being charged.
(10) Adapter socket: For connecting the power adapter.
(11) USB socket: Designed to update the software of the device and only serves engineers.
(12) Metal electrodes: For charging by charger stand.
2.1 Appearance
Fig. 2-1
PULSE OXIMETER
(2)
(6)
(11)
(5)
(4)
(3)
(7)
(8)
(9)
(10)
(12)
(1)
Pulse Oximeter

- 5 -
1
Measure...
SpO2
bpm
bpm
PR
060
075
115
99
2
90
%
Menu Lock
1
2
Menu Lock
Measure...
115
Fig.2-2 Digital display Fig.2-3 Wave display
Note:
Only the Alarm inhibition/off button and Power switch button will be active in the status that
the keyboard is locked.
Measurement screen:
♦Status bar 1: The status of the pulse oximeter is shown on the bar.
♦Status bar 2: If the measured SpO2or PR value exceeds the alarm limits, there will be
the corresponding information.
♦SpO2:SpO2area of display
● It shows the oxygen saturation level of functional hemoglobin during normal measurement.
● The color of the SpO2value is red when the SpO2is outside the alarm limits.
● It shows two dashes throughout probe off and finger out conditions.
♦99: SpO2high alarm limit indicator
♦90: SpO2low alarm limit indicator
♦Pulse amplitude bar.
It indicates the dynamic pulse amplitude and rate. As the detected pulse becomes
stronger, more bars are illumined with each pulse. The reverse is true for weak pulses.
♦75: PR high alarm limit indicator
♦60: PR low alarm limit indicator
♦PR: PR area of display
● It shows the pulse rate in beats per minute during normal measurement.
● The color of the PR value is yellow when the PR is outside the alarm limits.
● It shows three dashes throughout probe off and finger out conditions.
♦15:00: The current time.
INSTRUCTION MANUAL

- 6 -
2.2 Explanation of Symbols
Symbol Symbol
IPX1
Explanation
Type BF applied part Protected against dripping water
Attention, consult
ACCOMPANYING
DOCUMENTS
Audible alarm off/inhibition
It indicates alarm inhibition
when the countdown of 120s
displayed
ID indication
Audible alarm on
Pulse Beep on
The adapter is connected Battery power indication
Pulse Beep off
Keyboard is unlocked
Note: The Bluetooth®data transmission is reserved.
Keyboard is locked
USB cable is connected USB cable disconnected
SpO2probe is inserted SpO2probe off
Explanation
Bluetooth®activated Bluetooth®deactivated
PULSE OXIMETER
Alarm inhibition on/off
Prevent from rain
Date of Manufacture
Power on/off
Serial number
Manufacturer’s information
SN

- 7 -
Fig.2-4
2.3 Power Supply
2.3.1 Powered by batteries
2.3.2 Charging batteries through adapter
Batteries Installation :
1) Open the battery cover: Rotate the fixing screw slightly
in the rear panel to the position which is marked with “ ”
and then push the cover as indicated by arrowhead, as
shown in Fig.2-4.
2) Install 3 AA batteries lightly as indicated by the polarity
signs in battery compartment.
Note: Make sure the polarities of the batteries are correct.
3) Close battery housing cover
Close the battery housing cover and rotate the screw to the
position which is marked with . And the batteries are locked.
Battery life and replacement
There are five statuses of the indicator: the centre with 4 bars(full), 3 bars, 2 bars, 1 bar, and
empty with the frame in red. That the frame of indicator become red means little of battery
capacity remains. You should replace the batteries with new ones in time. Or else, the
indicator displays with a red frame constantly until battery capacity reaches critical condition
(the battery voltage is 3.5V±0.1V) at which time the unit shuts down.
Cautions:
● Be sure to install batteries with correct polarities.
● Only the approved batteries are recommended to be used.
● Do not use batteries not specified for this unit.
● Do not dispose of batteries in fire.
● If battery fluid gets on your skin or clothing, rinse with plenty of clean water immediately.
● Remove the batteries from this unit when you are not going to use it for a long period of
time (approximately one month).
● Do not use batteries of different types together.
● Do not use new and used batteries together.
● Dispose of batteries in accordance with the local ordinances and regulations.
To charge, please ensure the installed batteries are Ni-MH ones. Connect the adapter
with the pulse oximeter and the wall outlet, and then press and hold the power switch for 3
seconds to power the device on.
A prompt window appears enquiring you “Are the batteries Ni-MH ones?”. Select “Yes” to
charge them by wall outlet or “No” to abandon charge.
Warning!
Do not use any other AC adapter that not specified by our company.
Note: Please charge over 8 hours at the first time, or may reduce the battery life.
INSTRUCTION MANUAL

- 8 -
2.3.3 Charging batteries by charger stand
The device can also be charged by charger stand optional with the pulse oximeter. For more
information, refer to “Charger Stand Instruction Manual”.
Warning!
DO NOT charge when the non-rechargeable batteries are installed, otherwise damages or
injuries may be caused to the devices or users.
Note: During battery charge, the “Power Auto” and “Brightness” item can not be accessed.
2.3.4 Powered by AC power
While the device without batteries, it can be powered by connecting to external power source
through AC-DC adapter and can be used with patient.
● Simple to use and easy to operate.
● Compact, light in weight and convenient to carry.
● TFT display with adjustable backlight displays SpO2, pulse rate, pulse amplitude bar, etc.
● Up to 127 patients’ ID and 72-hour records storage.
● Visual and three-level audible alarms, battery power low alarm.
● Adjustable pulse beep and backlight.
● Data transfer to PC for review by USB cable
● Powered by three AA alkaline/ Ni-MH batteries.
● Battery charge function with adapter or charge stand (optional).
● Suitable for adult and pediatric (greater than 3 years of age) users.
2.4 Product Features
CHAPTER 3 SpO2Probe Installation
The SpO2Probe is shown in the following figure:
Insert the SpO2probe to the socket, as shown in Fig.3-1.
Then indicator is shown on the display . If the SpO2probe
is disconnected from the unit, the indicator appears and
the prompt “Probe Off !!!” appears in the status column.
The socket is also used for USB cable connected with the
PC for data transmission. For more information on data
transmission and data management, refer to corresponding
software instruction manual.
Fig.3-1
PULSE OXIMETER

- 9 -
User ID Setting
Select Return
User ID
001
CHAPTER 4 ID, Date and Time Setting
Fig.4-1 Fig.4-2 Fig.4-3
Fig.4-4
Always set date and time before using the unit for the first time. Set different ID
numbers for different users.
Check date and time are correct before using the unit, and reset them if necessary. Date and
time are important indicators when a measurement is taken.
4.1 Date & Time Settings
Set the correct time according to the following steps:
4.2 User ID Setting
Date and Time
Date: 20
09
09
17
12
49
17
02
-
-
-
-
Time:
Select Return
Alarm Setting
Data Manage
Date and Time
System Setting
MENU
Select Return
User ID Setting
Date and Time
Date: 20 09
17
12
49
17
02
-
-
-
-
Time:
Select Return
1) Press the power switch for 3 seconds to power on the pulse oximeter and then press the
menu button to enter the main menu, refer to the Fig.4-1.
2) Press the Down button to select “Date and Time” item, and then press the Menu button to
enter the time setup screen, refer to Fig.4-2.
Pick different sub-items to set and press the Select button to highlight it and then using the
Up or Down button to adjust the value. At last, press the Select button to confirm.
Set other sub-items of data and time according to the “Year setting” illustrated in the above
Fig.4-2 & 4-3.
Enter the “User ID Setting” from the main menu screen, refer
to Fig.4-4. Press the “Select” button to make the User ID
number highlighted, and then press the Up or Down button
to increase or decrease the ID number, and then press
the Select button to confirm your settings. The range of ID
number is: 001-127.
Note: User ID cannot be changed and you may get the
prompt information saying that “Cannot change!!!”, if you
attempt to change it when taking a measurement.
INSTRUCTION MANUAL

- 10 -
◆Only use the SpO2probes provided by manufacturer for SpO2measurements. Other
SpO2probes may cause improper performance.
◆Inaccurate measurements may be caused by:
◆Do not use the SpO2probe with exposed optical components.
◆Excessive patient movement may cause inaccurate measurements.
◆Tissue damage can be caused by incorrect application or use of the probe, for example
by wrapping the probe too tightly. Inspect the probe site to ensure skin integrity and correct
positioning and adhesion of the probe. More frequently inspection should be taken depend
on different patients if necessary.
Fig.5-1 Placement of the probe
CHAPTER 5 Take a Measurement
Before taking a measurement:
Note: If there is no finger in the probe, “Finger Out!!” will be shown.
Warnings!
◆The measurement would not be performed if the following instances come across in operation:
• Shock
• Low temperature of hand
• Have taken vascular activity medicine
• Anemia
• carboxyhemoglobin
• methemoglobin
• methylene blue
• Indigo carmine
● Select the suitable probe in terms of type and dimension.
● Plug probe into SpO2socket on top panel of pulse oximeter.
● Clip the patient finger to the rational position of the probe as the illustration as Fig.5-1.
• Incorrect application or use of probe
• Significant levels of dysfunctional hemoglobins (such as carboxyhemoglobin or methemoglobin)
• Intravascular dyes such as indocyanine green or methylene blue
• Exposure to excessive illumination, such as surgical lamps (especially ones with a xenon
light source), bilirubin lamps, fluorescent lights, infrared heating lamps, or direct sunlight
• High-frequency electro surgical interference and defibrillators
• Venous pulsations
PULSE OXIMETER

- 11 -
◆Loss of pulse signal can occur in any of the following situations:
• Placement of a probe on an extremity with a blood pressure cuff, arterial catheter, or
intravascular line.
• The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia
• There is arterial occlusion proximal to the probe
• The patient is in cardiac arrest or shock
• The probe is too tight
• There is excessive illumination from light sources such as a surgical lamp, a bilirubin
lamp, or sunlight
• A blood pressure cuff is inflated on the same extremity as the one to which an SpO2
probe is attached
Note: SpO2probe should be kept from the light source, e.g. radial lamp or infrared lamp.
INSTRUCTION MANUAL

- 12 -
CHAPTER 6 Other Settings
From the main menu, select and enter the “Alarm Setting” screen, refer to Fig.6-1.
1) SpO2alarm setup
To set SpO2alarm high limit, in the “Alarm Setting” screen, press the Up or Down button to
select the “High Limit (SpO2)”. And then press the Select button to highlight the value. Press
the Up or the Down button to adjust the value, and at last press the Select button to finish the
setting of SpO2high limit. The range of SpO2high limit is 71%~100%.
Set the low limit for SpO2as the above steps of SpO2high limit settings. The range of SpO2
low limit is 70%~99%.
2) PR alarm setting
Also, the limits settings of PR are performed as similarly as the SpO2limits. The range of
high limit is 31bpm-235bpm. And the range of low limit is 30bpm-234bpm.
3) Alarm on/off
In the Alarm Setting screen, press the Up or Down button to move the cursor to the
Alarm(SpO2) or Alarm(PR) sub-item, and press the Select button to highlight it. Press the
Up or Down button to select ON or OFF(refer to Fig.6-2). And then press the Select button to
confirm your settings. If you set the SpO2and/or PR alarm off, the corresponding limits on the
measurement display is crossed with “×”.
6.1 Alarm Setting
Alarm Setting
Alarm
ON
ON
093
070
ON
230
075
Low Limit
High Limit
Select Return
Fig.6-2
Fig.6-1
Alarm Setting
Alarm OFF
093
070
ON
230
075
Low Limit
High Limit
Select Return
PULSE OXIMETER

AUDIBLE ALARM INHIBITION/OFF:
Short pressing the button to silence the audible alarm for 2 minutes, the audible alarm
indicator will be displayed as , together with the countdown from 120s to 0s. Short pressing
it again, you can cancel alarm inhibition; Long pressing the button, the audible alarm
indicator will be displayed as which indicates the audible alarm off. And then long pressing
the button, you can activate the audible alarm again.
Notes:
1. The alarm sound will go on until alarm disappears or is turned off.
2. After silencing the alarm, the corresponding indicator will indicate this.
3. The power low alarm: the corresponding indication lamp will be flashing with a red frame.
Assignment of priority:
Paramter SpO2PR /
/
/
/
Yellow
Flashing with yellow
0.5Hz
Di - Di - Di Di
20 s
Probe off/Finger out
5 s
PR too high/low
Red
Flashing with yellow
1.5Hz
Di- Di – Di ----- Di - Di
3 s
SpO2too high/low
High
item property Medium Low
Value
Lamp Frequency
Audiblesound
Alarm cycle
Alarm info
Alarm lamp
- 13 -
ALARM PRIORITY:
There are three-level priorities for selection.
High priority: indicates the patient is in a dangerous situation.
Medium priority: indicates the warnings should be paid attention to.
Low priority: indicates the technical alarm caused by the device itself.
Alarms of the pulse oximeter include technical and physiological alarms. All the three priorities
are divided by built-in module and can not be changed by user.
Warnings!
When an alarm occurs, check patients’ conditions immediately.
● Check the parameter which is alarming.
● Check patient’s condition.
● Search for the source of the alarm.
● Make the alarm mute if necessary.
● Check the alarm when no warning.
INSTRUCTION MANUAL

- 14 -
6.2 Data Management
From the main menu screen, select and enter the “Data Manage” screen, refer to Fig.6-3.
Press the Up or Down button to select the sub-item to set, and then press the Select button
to confirm or Return button to return to the previous screen.
Notes:
● The user may get the prompt information saying that “No records” when entering the data
manage interface if there is no stored record in the device.
● The device does not support cycle storage, when the memory is full, a prompt window of
“Space Expire!” appears. To record the newest measurements, you should delete the old
ones for free space when no measurement is being taken.
Fig.6-3
SpO2Trend
PR Trend
Clear Data
Data Manage
Select Return
Data Review
6.2.1 Data Review
Pick ”Data Review” sub-item and enter by pressing the Select button. The screen as in
Fig.6-4 will appear.
Pressing the Select button again, a dialog box will pop up, refer to Fig.6-5. After selecting an
ID number, press the Up or Down button to select “Delete” or “Review” and then press the
Select button to confirm. You can delete or review all the records saved under the ID.
Fig.6-4 Fig.6-5
Data Review
03/17 17:42:43 072098
098
098
098
098
072
071
076
072
03/17 17:42:39
03/17 17:42:35
03/17 17:42:31
03/17 17:42:27
Select Return
Data Review
03/ 72
098
72
72
69
069
03/
03/
03/
03/17 17:42:27
Select Return
User ID
Review
006
Delete
PULSE OXIMETER

Fig.6-10
Data Manage
Select Return
Data Review
Deleting Data?
Yes No
- 15 -
6.2.2 SpO2Trend
6.2.3 PR Trend
6.2.4 Clear Data
Pick “SpO2Trend” sub-item and enter by pressing the Select button. The screen as in Fig.6-6
will appear.
Press the Select button again, a dialog box will pop up, refer to Fig.6-7. After selecting a
ID number, press the Up or Down button to select “Delete” or “Review” and then press the
Select button to confirm, you can delete or review all the SpO2trends saved under the ID.
The operation is referred to 6.2.2 SpO2Trend.
Pick ”Clear Data” sub-item and enter by pressing the Select button. A dialog box will pop up,
refer to Fig.6-10. Press the Up or Down button to pick " Yes " or " No " and then press the
Select button to determine whether delete all the records.
Note: Pay attention to the data deletion as the deleted data cannot be restored.
Fig.6-6 Fig.6-7
SpO2Trend
SpO2%1/1
Time(s)
Select Return
20
0
40
60
80
100
SpO2Trend
SpO2%1/1
Time(s)
Select Return
20
0
40
60
80
100
User ID
Review
006
Delete
Fig.6-8 Fig.6-9
PR Trend
PR 1/1
Time(s)
bpm
Select Return
50
0
100
150
200
250
PR Trend
PR
1/1
Time(s)
bpm
Select Return
0
50
100
150
200
250
User ID
Review
006
Delete
INSTRUCTION MANUAL

- 16 -
6.2.5 In the data management screen, the upper half screen will display the maximum,
minimum and average values of SpO2 and PR for the current ID. Refer to Fig.6-11.
Fig.6-11
Measure...
006 Min
93
065 073 087
97 099
Avg Max
6.3 System Setting
From the main menu, select and enter the System Setting screen refer to Fig.6-12&6-13.
Press the Up or Down button to pick the different sub-items and press the Select button to
enter.
Pick the Brightness sub-item and enter it from the System Setting screen, and set the
brightness level and brightness time.
Brightness level: 1-7
Bright time: ON, 15, 30, 45, 60, 75, 90,105,120 seconds.
Fig.6-12 Fig.6-13
Blue Tooth
Beep Setting
Display Mode
Keyboard Lock
System Setting
Select Return
Brightness
About
Default Setup
System Setting
Select Return
Power Auto
6.3.1 Brightness
6.3.2 Bluetooth(Reserved)
6.3.4 Display mode
6.3.3 Beep setting
Note: The brightness cannot be set in the charging process.
Enter the Beep setting screen from the System Setting screen, and set the beep level and on/off.
Beep level: 1-7
Beep(switch): on or off.
Enter the Display Mode screen from the System Setting screen, and select the display mode
for measurement screen from Wave and Digital.
Select ‘Wave”, SpO2plethysmogram will be displayed on the lower screen of the
measurement screen.
PULSE OXIMETER

- 17 -
Enter the Keyboard lock screen from the System Setting screen, and set the lock function on/
off. If set to on, long press the Lock button, the Menu/select, Up and Down, Alarm inhibition/off
buttons on front panel will be deactivated, except the power switch and lock buttons.
To unlock the keyboard, long press the Lock button again.
6.3.5 Keyboard lock
6.3.6 Power Auto
Enter the Power Auto screen from the System Setting screen, and set the auto power-off time
and the auto power-off on/off.
Auto power-off time: Set how long it takes when no operation before the device
automatically power off.
Time: 1,2,3,4,5…15 minutes.
Auto: Set the auto power-off function on or off.
Note: The auto power off time cannot be set in the charging process.
6.3.7 About
Enter the About screen from the System Setting screen, the version of software is displayed.
6.3.8 Default Setup
Enter the Default Setup screen from the System Setting screen, and you can restore the
default configurations.
Note: The user cannot restore the factory defaults configurations when measuring.
Default configuration includes:
ID: 001
SpO2low limit: 90% ; SpO2high limit: 100% SpO2alarm: on
PR low limit: 60 bpm; PR high limit: 100bpm PR alarm: on
Brightness Level: 3 Bright time: On
Bluetooth: Off
Beep level: 4 Beep(switch): on
Display(mode): Wave
Keyboard lock: On
Auto power-off time: 3 minutes Auto(power-off): on
Note: After changing the batteries, the settings may get back to the default settings.
INSTRUCTION MANUAL

- 18 -
CHAPTER 7 Maintain and Repair
Use only the substances approved by us and methods listed in this chapter to clean
or disinfect your equipment. Warranty does not cover damage caused by unapproved
substances or methods.
We make no claims regarding the efficacy of the listed chemicals or methods as a means
for controlling infection. For the method to control infection, consult your hospital’s Infection
Control Officer or Epidemiologist. Keep your equipment and accessories free of dust and dirt.
To avoid damage to the equipment, follow these rules:
● Always dilute according to the manufacturer’s instructions or use lowest possible
concentration.
● Do not immerse any part of the equipment into liquid.
● Do not pour liquid onto the equipment or accessories.
● Do not allow liquid to enter the case.
● Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners (such
as acetone or acetone-based cleaners).
7.1 Maintenance
7.2 Safety Checks
Caution:
If you spill liquid on the equipment or accessories, contact us or your service personnel.
Note:
To clean or disinfect reusable accessories, refer to the instructions delivered with the
accessories.
Before every use, or after your pulse oximeter has been used for 6 to 12 months, or whenever
your pulse oximeter is repaired or upgraded, a thorough inspection should be performed by
qualified service personnel to ensure the reliability. Follow these guidelines when inspecting
the equipment:
● Make sure that the environment and power supply meet the requirements.
● Inspect the equipment and its accessories for mechanical damage.
● Make sure that only specified accessories are applied.
● Inspect if the alarm system functions correctly.
● Make sure that the batteries meet the performance requirements.
● Make sure that the pulse oximeter is in good working condition.
In case of any damage or abnormity, do not use the pulse oximeter. Contact your hospital’s
biomedical engineers or your service personnel immediately.
Cleaning
Your equipment should be cleaned on a regular basis. If there is heavy pollution or lots
of dust and sand in your place, the equipment should be cleaned more frequently. Before
cleaning the equipment, consult your hospital’s regulations for cleaning the equipment.
Recommended cleaning agents are:
● Mild soap (diluted)
● Ammonia (diluted)
● Sodium hypochlorite bleach (diluted)
● Hydrogen peroxide (3%)
● Ethanol (70%)
● Isopropanol (70%)
PULSE OXIMETER
Table of contents
Other MedSource Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Drive DeVilbiss Healthcare
Drive DeVilbiss Healthcare 1025DS Service manual

Otto Bock
Otto Bock 6A53 Instructions for use

Siemens
Siemens LITHOSTAR Multiline Adjustment instructions

Olive
Olive OLV-B1 user manual

EMS
EMS AIR-FLOW handy 2+ Operation instructions

Medline
Medline Alterra MAXX Operation and maintenance manual