7010 TOC ANALYZER -USER MANUAL
10
1.6 ANALYZER GENERAL HAZARDS
1.6.1 Electrical precautions and hazards
In all electrical devices that are 230 VAC (or 115 VAC opt.) powered present hazards of
electrical shock or electrocution.
To protect all the personnel involved in analyzer use and maintenance, the doors of the two
analyser enclosures are equipped with a special key for opening.
If necessary to operate inside the electrical enclosure with the analyzer powered on, please
consider that this operation must be made only by qualified personnel in accordance with
national or local regulations. Qualified personnel means a person who has been fully trained
and has professional experience to avoid electricity hazards and dangers.
Service qualified personnel will receive the special key to open the electrical enclosure.
Before to service the analyzer or parts of that electrically powered, turn off all power to avoid risk
of electrocution.
To turn off power from an electrical device is necessary to interrupt the power line using a circuit
breaker or an isolating switch to be sure that there is no power in the area being serviced.
Users and qualified maintenance personnels must proceed as follows:
pay attention to electrical shock and/or electrocutions labels placed on the analyser
always isolate power before servicing the analyser
In case of loss of power the analyser stops and will automatically restart as soon as power is
restored.
WARNING: UV lamp power supply
reaches an ignition voltage up to 3000
volts; do not operate any service
activity without removing the
instrument power cord
.
Analyser’s control circuitry is powered 24
Vdc and 12 Vdc. Inside the electrical
enclosure the lower level of protection
against direct contacts is IP2X. Analyzer
enclosures are IP54 (because of air fans
holes).
Protection against indirect contacts is
guaranteed by efficient grounding of all
isolated metal masses.
Grounding bar is located inside the
electrical enclosure, in upper left position
below the protection cover.
It will be user’s responsability to
check and garantee periodically the
perfect efficiency of analyzer’s
grounding.