Failure to comply with the precautions below could lead to injury or death.
WARNING
Use the Correct Power Supply
Make sure that the power supply voltage matches the AQ7277’s rated supply voltage and that it
does not exceed the maximum voltage range specified for the power cord.
Check the Power Supply Capacity
Make sure that the power supply capacity sufficiently meets the AQ7277’s maximum power
consumption (or current capacity) before you connect the power cord.
Do Not Look at the Laser Beam
Do not look directly or indirectly into the laser beam or at a specular reflection of the beam without
protective equipment. The laser beam may cause blindness or damage to your eyes.
Do Not Operate in an Explosive Atmosphere
Do not operate the instrument in the presence of flammable gasses or vapors. Operation in such
an environment constitutes a safety hazard.
Do Not Operate in an Environment with Flammable or Explosive Gasses, Steam, or Dust
Do not operate the instrument in an environment with flammable or explosive gases, steam, or
dust (dangerous places). Operation in such an environment constitutes a safety hazard.
Do Not Remove the Case
Only qualified YOKOGAWA personnel should remove the instrument’s covers. Opening the case
is dangerous, because some areas inside the instrument have high voltages.
Remove Connections When Carrying or Moving the Instrument
When carrying or moving the instrument, remove all cords (including the power cord) and cables.
CAUTION
When Using the Same Wavelength That Is Used in Real Communication to Make
Measurements
If communication light is present in the optical fiber that you want to measure, the communication
may be affected. Take appropriate precautions to avoid communication interference.
In addition, to make accurate measurements with the AQ7277, take the measurement environment
(such as the presence of communication light) into consideration.
When Using a Wavelength (1625/1650 mm) Different from That Used in Real Communication
to Make Measurements
When there is communication light in the fiber under measurement, use a light beam with a
different wavelength to make measurements.
If the device connected to the system under measurement does not have a filter for blocking
1625 nm and 1650 nm laser beams or depending on the lightfastness power rating or the
attenuation characteristics of the blocking filter, the light pulses that the AQ7277 emits may
damage the device. Be sure to install an appropriate blocking filter and check that the device’s
ratings are adequate before making measurements.
When Using SC Angled Physical Contact Connectors
When the AQ7277 has SC angled physical contact connectors, connect optical fibers that have
the same type of connectors.
The SC angled physical contact connector’s ferrule tip is angle-polished. Using a different type of
connector may damage the connector end face.