
4
3. Checklist for Taking a Reliable Measurement
Avoid activity, eating or smoking immediately before the 
measurement.
Sit down on a back-supported chair and relax for 5 minutes. 
Keep the feet flat on the floor and do not cross your legs.
Always measure on the same arm (normally left). It is recom-
mended that doctors perform double arm measurements on a 
patients first visit in order to determine which arm to measure in 
the future. The arm with the higher blood pressure should be 
measured.
Remove close-fitting garments from the upper arm. To avoid 
constriction, shirt sleeves should not be rolled up - they do not 
interfere with the cuff if they are laid flat.
Always ensure that the correct cuff size is used (marking on the 
cuff).
Fit the cuff closely, but not too tight.
Make sure that the cuff is positioned 1-2 cm above the elbow.
The artery mark on the cuff (ca. 3 cm long bar) must lie over 
the artery which runs down the inner side of the arm.
Support your arm so it is relaxed.
Ensure that the cuff is at the same height as your heart.
4. Taking a Blood Pressure Measurement
1. Select standard (single measurement) or 
AFIB/MAM
 mode 
(automatic triple measurement): see details in chapter «2.».
2. Press the START/STOP button 1to start the measurement.
3. The cuff will now pump up automatically. Relax, do not move 
and do not tense your arm muscles until the measurement 
result is displayed. Breathe normally and do not talk.
4. When the correct pressure is reached, the pumping stops and the 
pressure falls gradually. If the required pressure was not reached, 
the device will automatically pump some more air into the cuff.
5. During the measurement, the pulse indicator BQ flashes in the 
display.
6. The result, comprising the systolic AP and the diastolic AQ blood 
pressure and the pulse rate AR is displayed. Note also the 
explanations on further display symbols in this booklet.
7. When the device has finished measuring, remove the cuff.
8. Switch off the device. (The monitor does switch off automatically 
after approx. 1 min.).
AF detection is only activated in AFIB/MAM mode.
You can stop the measurement at any time by pressing the 
START/STOP button (e.g. if you feel uneasy or an 
unpleasant pressure sensation). 
This monitor is specially tested for use in pregnancy and 
pre-eclampsia. When you detect unusual high readings in 
pregnancy, you should measure after a short while again 
(eg. 1 hour). If the reading is still too high, consult your 
doctor or gynecologist.
In pregnancy the AFIB symbol can be ignored. 
How not to store a reading
As soon as the reading is displayed press and hold the START/
STOP button 1until «M» BO is flashing. Confirm to delete the 
reading by pressing the M-button 3.
How do I evaluate my blood pressure?
The LED traffic light indicator on the left-hand side of the display 
7
 
indicates within which range the measured blood pressure lies. The 
value is either within the optimum (green), elevated (yellow) or high 
(red) range. The classification corresponds to the following ranges 
defined by international guidelines (
ESH, ESC, JSH
). Data in mmHg.
The higher value is the one that determines the evaluation. 
Example: a blood pressure value of 140/80 mmHg or a value of 
130/90 mmHg indicates «blood pressure too high».
5. Data Memory
This device automatically stores the last 99 measurement values.
Viewing the stored values
Switch the lock switch 
AO
 to «unlock» position. Press the M-button 
3
 
briefly. The display first shows «
M
» 
BO
 and «
A
», which stands for the 
average of all stored values.
Pressing the «+» AN or the «-» AM button repeatedly enables you 
to move from one stored value to another. Press the M-button 
again to exit the memory mode.
Pay attention that the maximum memory capacity of 99 
memories is not exceeded. 
When the 99 memory is full, the 
Range Systolic Diastolic Recommendation 
1. blood pressure 
too high
≥135 ≥85 Seek medical 
advice
2. blood pressure 
elevated
130 - 134 80 - 84 Self-check
3. blood pressure 
normal
<130 <80 Self-check